Design and Validation of an Automated Hydrometers Calibration System

Similar documents
The uncertainty evaluation for the hydrometer calibration at NMISA. Presenter: Bongani Ndlovu Date: 09/10/2018

Publication V. M. Heinonen and S. Sillanpää, The effect of density gradients on hydrometers, Measurement Science and Technology 14, (2003).

EURAMET Project 858: Hydrostatic weighing exchange of experiences. Project outline

EUROMET Project 702 EUROMET.M.D-K4

Liquid density determination by using a solid density standard

Air Density Determination Using 1 kg Buoyancy Mass Comparison(III)

Traceable Mass Determination and Uncertainty Calculation

Portuguese Institute for Quality. Contents

Measuring Repeatability, Linearity & Sensitivity Drift - P.3

National Physical Laboratory New Delhi

Automated volume measurement for weihts using acoustic volumeter

Metrological Characterization of a Primary Vickers Hardness Standard Machine - NIS Egypt

Water Vapour Effects in Mass Measurement

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT

Alignment of the Measurement Scale Mark during Immersion Hydrometer Calibration Using an Image Processing System

hapter 13 Archimedes Up-thrust

Fluid Mechanics. Chapter 14. Modified by P. Lam 6_7_2012

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L

Chapter 15 - Fluid Mechanics Thursday, March 24 th

EVALUATION OF ADSORPTION EFFECT BY USING SILICON SPHERE AND SILICON SURFACE ARTEFACTS

Measuring Viscosity. Advanced Higher Physics Investigation. Jakub Srsen SCN: Boroughmuir High School Center Number:

Interlaboratory Comparison & Proficiency Testing Program

Chapter 9. Solids and Fluids. 1. Introduction. 2. Fluids at Rest. 3. Fluid Motion

Thermodynamic Properties of Low-GWP Refrigerant for Centrifugal Chiller

Final report Comparison Identifier: APMP.SIM.M.P-K1c Other designation: NIST-NPLI/Pressure/2

Chapter 12. Fluid Mechanics. A. The density ρ of a substance of uniform composition is defined as its mass M divided by its volume V.

A laser metroscope for the calibration of setting rings

Fluid Mechanics Introduction

CERTIFICATE OF ACCREDITATION

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

CHAPTER 28 PRESSURE IN FLUIDS

Interlaboratory Comparison & Proficiency Testing Program

MASS AND VOLUME COMPARISONS AT MIKES

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.

Interlaboratory Comparison & Proficiency Testing Program

EFFECT OF COVARIANCES IN A DENSITY MEASUREMENT OF SILICON SAMPLES BY HYDROSTATIC WEIGHING

COMPARISON OF INMETRO AND INTI HUMIDITY STANDARDS. Final Report

General Accreditation Guidance. Liquid-in-glass thermometers selection, use and calibration

Accuracy Calibration Certificate

Physics 201 Chapter 13 Lecture 1

storage tank, or the hull of a ship at rest, is subjected to fluid pressure distributed over its surface.

CERTIFICATE OF ACCREDITATION

CALIBRATION CERTIFICATE # 503

Fluid Mechanics. If deformation is small, the stress in a body is proportional to the corresponding

The Uncertainty of a Waterdraw Calibration vs. Gravimetric Calibration on Small Volume Provers

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty

TOPICS. Density. Pressure. Variation of Pressure with Depth. Pressure Measurements. Buoyant Forces-Archimedes Principle

PHYSICS HYDROSTATICS FORM 5

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

Fluid Mechanics-61341

Meteorology 432. Barometry Spring 2013

Evaluation of the Effect of Relative Humidity of Air on the Coefficients of Critical Flow Venturi Nozzles

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

Introduction. Measurement of temperature is generally considered to be one of the simplest and most accurate measurements performed in engineering.

ME-B41 Lab 1: Hydrostatics. Experimental Procedures

SIM SIM.M.D-K3. provide. SIM.M.M-K5 for. A set of. Institute LACOMET LATU INTI. Country Costa Rica Uruguay Argentina Chile México Canada Brazil

CHAPTER 13. Liquids FLUIDS FLUIDS. Gases. Density! Bulk modulus! Compressibility. To begin with... some important definitions...

Exam 3--PHYS 101-WWP--Fall Chapters 8, 9, & 10

Chapter 9 Solids and Fluids. Elasticity Archimedes Principle Bernoulli s Equation

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER

Eric G. Paterson. Spring 2005

Certificate of Accreditation

DROP-WEIGHT SYSTEM FOR DYNAMIC PRESSURE CALIBRATION

CHAPTER 10- GRAVITATION

A Method for Verifying Traceability in Effective Area for High Pressure Oil Piston-Cylinders

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

Energy Considerations

Guide for volume determination within the scope of reference measurement procedures in medical reference measurement laboratories, Part 1

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

Towards SI traceable humidity measurements with radiosondes

Chapter 9: Solids and Fluids

Chapter 15. m. The symbolic equation for mass density is: ρ= m V. Table of Densities

Physics 107 HOMEWORK ASSIGNMENT #9

Mass per unit volume Mass of 1 m 3. Units of density = kg m 3 Density is a scalar. mass volume. density = m V. rho. 1m x 1m x 1m

Objective: To enable the students to describe, state and derive the terms and expressions relevant in carrying out experiment 5 meaningfully.

Liquids CHAPTER 13 FLUIDS FLUIDS. Gases. Density! Bulk modulus! Compressibility. To begin with... some important definitions...

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

CHARACTERISTIC OF FLUIDS. A fluid is defined as a substance that deforms continuously when acted on by a shearing stress at any magnitude.

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Metrology and Environment

MEASUREMENT OF WBGT INDEX IN AXIAL FLOW AS THE MAIN THERMODYNAMIC PARAMETERS VARY G.

Upthrust and Archimedes Principle

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory

Loadcell Calibration - Evaluation of Uncertainties

MS20 Laboratory Seawater Salinity and Density

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

Viscosity * Desmond Schipper Andrew R. Barron. 1 Introduction

CERTIFICATE OF CALIBRATION

SIM Regional Comparison on. The Calibration of Internal and External Diameter Standards SIM.L-K FINAL REPORT July 2012

Final Report on Comparison of Hydraulic Pressure Balance Effective Area Determination in the Range up to 80 MPa

Chapter 2 Hydrostatics Buoyancy, Floatation and Stability

The Fundamentals of Moisture Calibration

Archimedes Principle

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Lesson 6 Matter. Introduction: Connecting Your Learning

Supplementary Comparison

Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge

EURAMET Project no Inter-comparison of a 1000 L proving tank

Al-Saudia Virtual Academy Pakistan Online Tuition Online Tutor Pakistan

Transcription:

Design and Validation of an Automated Hydrometers Calibration System M.H. Mohamed 1, M.A. Bayoumi, Alaaeldin A. Eltawil 1, A. E. Abuelezz 3 Dept. of Mass, Density and Pressure, National Institute for Standards, Giza, Egypt 1 Dept. of Mech. Engineering, Faculty of Engineering, Al-Azhar University, Cairo, Egypt Dept. of Force & Material Metrology, National Institute for Standard, Giza, Egypt 3 Abstract: iquid density measurement is an important activity in many laboratories. Usually hydrometers are used to measure the density of liquids in the range from 600 kg/m 3 up to 1850 kg/m 3. For the purpose of metrological traceability of Hydrometers with the international density reference standards a calibration system has been designed, constructed and validated at the National Institute for Standards, NIS, Egypt. he system covers the range from 500 kg/m 3 up to 000 kg/m 3. he system is automatically operated for the intended measurements and calibrations using hydrostatic weighing (Cuckow s method). Distilled water was used as a reference liquid. Finite element analysis using (ABAQUS) software for 3D has been used to check the stress, deformation, rigidity, stability as well, to optimize the design of the system. A laser light source and two linear encoders are used to position the standard liquid surface at the hydrometer scale mark during calibration process. For accurate measurements of scale mark position, an image processing technique combined with a high resolution CCD camera for capture the reading during calibration is used. ab VIEW program is designed to processing the reading from measurement instruments in a fully automated mode. he calibration results present an expanded uncertainty (with a coverage factor of k= ) less than 100 ppm. he apparatus s performance has been validated by comparing the output results of the system versus aboratoire National de Metrologie et d Essais NE calibration certificates of 3 standard hydrometers. Key words: iquid density, Hydrometers Calibrations, Cuckow s method, Hydrostatic weighting system I. INRODUCION A hydrometer is a glass instrument consisting of a bulb has a mass at its bottom and extended by a stem of small diameter, on which scale readings are marked [1] as shown in Figure 1. he operation principle operation of the hydrometer is based on the Archimedes principle [] where a solid body suspended in a fluid will be buoyed up by a force equal to the weight of the fluid displaced by the submerged part of the suspended body. Apart from the simple construction of the hydrometers its calibration needs a lot of effort where two methods are normally used. he comparison method and the Cuckow s method.he National Institute for Standards Egypt (NIS) nowadays provides a calibration service for hydrometers in the range from 600 kg/m 3 up to 1850 kg/m 3. he calibrated hydrometers are used as working standards to calibrate customer s hydrometers and provide traceability to thousands of hydrometers used for density measurements in the distilling, petroleum, chemical, dairy industries. At present, NIS working standard hydrometers has been calibrated at aboratoire National de Metrologie et d Essais, NE, France [3]. he NIS perform calibrations for the customers hydrometers by the comparison method [4] using several liquids. Mixing of these liquids to reach any point of calibration located at the hydrometer stem. he mixing process is difficult and hazard especially in the range above distilled water. In addition to the personal errors, changing in the environmental temperature during the calibration process, the meniscus which appear around the stem while the standard and under test hydrometers in freely floating in the liquid, a set of fragile, breakable standard hydrometers used to carry out the comparison method against under test hydrometers. As a consequence of the stated previous reasons, calibrations cannot be made exactly at any calibration point located at the hydrometer stem and large uncertainties of measurements are reached. A method in which only one liquid covers the whole calibration range from 500 kg/m 3 to 000 kg/m 3 was developed initially at the NP, UK [5] and suitably adapted by us in India. My colleagues, at the NP, India, developed an open-scale hydrometer with a very small range but with a scale interval of 0.01 kg/m 3 and also designed and Copyright to IJIRSE www.ijirset.com 1735

fabricated a thermostatic bath. he bath was examined for constancy and uniformity of temperature, throughout the working space. he hydrometer is calibrated by submerging it to the required level and finding its apparent mass in the liquid. he density of the liquid is simultaneously determined with the help of the open-scale hydrometer.o get rid of these problems a fully automated hydrostatic weighing system working in the range from 500 kg/m 3 up to 000 kg/m 3 is design and constructed for the calibration of hydrometers via Cuckow s method [5-8]. Fig.1. A freely floating hydrometer II. CONSRUCION OF HE DESIGNED HYDROMEER CAIBRAION SYSEM Figure shows the system that has been designed in NIS-Egypt for hydrometers calibrations. he system consists of main cabinet (1) which carries the high sensitive electronic digital balance () with maximum capacity 300 g and resolution 0.1 mg. iquid temperature controller (3),commercially available which is used for controlling the temperature of the water jacket with stability better than ± 0.05 K 1 3 Fig.. he constructed hydrometer calibration system Figure 3 shows the mechanical lifter (4) which has been designed to be rigid enough to carry the weighing cell which is consisting of two coaxial cylinders (5) having in between the jacket water. Copyright to IJIRSE www.ijirset.com 1736

Figure 4 shows a system for ambient condition measurements consists of barometer (6) with resolution 0.1 Pa to measure atmospheric pressure and thermo-hygrometer (7) with temperature resolution 0.001 ºC and relative humidity with resolution 0.01 %. his system is needed for air density calculations [9-11] at the time of weighing the hydrometer in the air, while digital temperature indicator (8) with prop sensor P100 for measuring the temperature of the reference liquid with resolution 0.001 ºC. Figure 5 shows a laser light source (9) fixed on a mechanical lifter for the movement up and down via controlled motor and two linear encoders to position the liquid surface at the calibration point and a CCD camera (10) with resolution (1600H 100V) fixed on the same mechanical lifter of the laser light source and in the perpendicular direction. An imaging processing technique [1] is used to indicate the scale marks to be calibrated. 5 4 7 8 9 6 10 Fig.3. Reference iquid ifter Fig.4. he system for ambient condition measurements Fig.5. ASER &CCD Camera ifter III. VAIDAION OF HE HYDROMEER CAIBRAION SYSEM he validation has been performed, by comparing the results of calibration of 3 standard hydrometers, calibrated and certified by NE France with the calibration results, of the same hydrometers, obtained using the constructed calibration system at HCS-NIS, based on Cuckow's method. a. CUCKOW S MEHOD Cuckow derived a method for a calibration laboratory to determine the density ρ represented by a given scale mark on the hydrometer stem. Force balance equations are written for a hydrometer in three conditions [13]: 1. When floating freely in a liquid of unknown density (ρ ),. When weighed in air and 3. When weighed while immersed to the same level as case 1, but in a liquid of known density. By solving the three equations simultaneously, one can derive an equation for ρ : 1 0 a O DCos / g a O O / g m V 1 3 s s s 0 (1) Copyright to IJIRSE www.ijirset.com 1737

When calibrating a hydrometer for use in liquids that are less dense than standard liquid, a stainless steel weight is added to sink the hydrometer. he weight is a ring of known mass and volume placed over the stem, resting on the top of the bulb. It is submerged in the standard liquid during the hydrometer calibration as equation (1). Where ρ : Density of the distilled water, : Volumetric thermal expansion coefficient of the glass from which the hydrometer is made, : emperature of the distilled water, 0 : Reference temperature, ρ a : Density of the air at the time the hydrometer was weighed in air, : Calibration coefficient for the balance, O : he balance reading for the hydrometer weighing in air, D : Diameter of the hydrometer stem at the scale mark to be calibrated, : Surface tension of the distilled water, θ : Contact angle of the distilled water on the hydrometer stem, g : ocal acceleration due to gravity, O 3 : Output from the balance for the hydrostatic weighing, : Surface tension of the distilled water, θ : Contact angle of the distilled water on the hydrometer stem, m s : Mass of the stainless steel weight added to sink a hydrometer, V s : Volume of the stainless steel weight, s : Volumetric thermal expansion coefficient of the stainless steel weight. o calculate hydrometer corrections for a particular liquid under test. he values of A and B are the output the calibration. C A B Cos () A 1 0 a O / g O / g m V 1 a O 3 s s s 0 R (3) Where: R is the hydrometer reading at the calibrated scale mark D B A R a O g (4) Copyright to IJIRSE www.ijirset.com 1738

b. MEASUREMENS AND RESUS he three standard hydrometers, A6, A17 and A43, which used for the validation process in this research, were carefully selected so that the comparison range cover different densities at the full range of the newly constructed calibration system. he obtained results as well as the certified results of NE are given in able 1 for density range 0.700 0.70 g/cm 3 ( A6), in able for density range 0.897 0.917 g/cm 3 ( A17) and in able 3 for density range 1.494 1.54 g/cm 3 ( A43). able 1 shows the calibration results at (0.707, 0.711 and 0.715 g/cm 3 ) calibration points in between (NE) and (HCS- NIS) for the hydrometer A6.he maximum expanded uncertainty is 3.0 10-5 for (HCS-NIS) and it is the same for (NE).he maximum difference in the corrections between (HCS-NIS) and (NE) is (-0.000007 g/cm 3 ) at calibration point (0.711 g/cm 3 ) able 1.he results of the hydrometer A6 he Range of Density 0.700 0.70 resolution 0.0001 Serial no. HCS-NIS Results NE Results A6 Calibrated Points Corrections Expanded Uncertainty 0.707-0.000053 3.0 10-5 0.711-0.000047 3.0 10-5 0.715-0.000079 3.0 10-5 0.707-0.000051 3.0 10-5 0.711-0.000040 3.0 10-5 0.715-0.000073 3.0 10-5 able shows the calibration results at (0.904, 0.908 and 0.91 g/cm 3 ) calibration points in between (NE) and (HCS- NIS) for the hydrometer A17.he maximum expanded uncertainty is 3.0 10-5 for (HCS-NIS) and it is the same for (NE).he maximum difference in corrections between (HCS-NIS) and (NE) is (0.000008 g/cm 3 ) at calibration point (0.91 g/cm 3 ). able.he results of the hydrometer A17 he Range of Density 0.897 0.917 Resolution 0.0001 Serial no. A17 Calibrated Points HCS-NIS Results NE Results Corrections Expanded Uncertainty 0.904 0.000016.9 10-5 0.908 0.000009.9 10-5 0.91 0.000040 3.0 10-5 0.904 0.000011 3.0 10-5 0.908 0.000004 3.0 10-5 0.91 0.00003 3.0 10-5 able 3 show the calibration results at (1.496, 1.509 and 1.5 g/cm 3 ) calibration points in between (NE) and (HCS- NIS) for the hydrometer A43.he maximum expanded uncertainty is 5.5 10-5 for (HCS-NIS) and he maximum expanded uncertainty is 4.0 10-5 for (NE).he maximum difference in corrections between (HCS-NIS) and (NE) is (0.000008 g/cm 3 ) at calibration point (1.509 g/cm 3 ). Copyright to IJIRSE www.ijirset.com 1739

able 3. he results of the hydrometer A43 he Range of Density 1.494 1.54 Resolution 0.000 Serial no. HCS-NIS Results NE Results A43 Calibrated Points Corrections Expanded Uncertainty 1.496 0.000040 5.3 10-5 1.509 0.000043 5.4 10-5 1.5 0.000041 5.5 10-5 1.496 0.00004 4.0 10-5 1.509 0.000051 4.0 10-5 1.5 0.000046 4.0 10-5 IV. VAIDAION OF OBAINED RESUS BEWEEN (HCS-NIS) AND (NE) o validate the results obtained from the HCS-NIS, with previous results obtained from NE have been carried out. he formula for the normalized error (E n -value) was used [14] Value ( ab) Value (Re f ) E n (5) U( ab) 95 U(Re f ) 95 Where: Value (ab): is the value reported by the participating laboratory; Value (Ref): is the reference value for the measured; U (ab) 95: is the uncertainty reported by the participating laboratory; U (Ref) 95: is the uncertainty of the reference value Results are judged as following: -1 E n 1 is satisfactory OR E n > 1 or E n < -1 is unsatisfactory From the comparison for the range of density measured. he relative error between HCS-NIS and NE is not more than 0.019%.By substituting in equation (5) the normalized error (E n ) values of the measurements are 0.17, 0.11 and 0.6 in the hydrometers A6, A17 and A43 respectively. From this comparison with calibration certificates, HCS-NIS has the validation to measure the density and calibration the hydrometers in the range from 500 kg/m 3 up to 000 kg/m 3. V. CONCUSION A fully automated hydrometer calibration system at the (NIS) Egypt, has been designed, constructed and validated with the objective to be used for liquid density measurements and hydrometers calibrations in the range from 500 kg/m 3 up to 000 kg/m 3.he apparatus s performance has been validated by comparing the output results versus NE s calibration certificates. he calibration results present an expanded uncertainty (K=) less than 100 ppm. Copyright to IJIRSE www.ijirset.com 1740

REFERENCES [1] John D.Wright,V.E, NIS Calibration Service for Hydrometers, special publication, 008 [] Stott V and A E Dunstan Hydrometers, in the Science of Petroleum, (ondon, Oxford University Press), vol. 3, PP. 3-7, 1938 [3] Rolande Aguirre, calibration certificates, NE, France, Numbers From: F031808 / 6 to F031808 / 40, 006 [4] S. V. Gupta, Practical density measurement and hydrometry, Series in Measurement Science and echnology, IoP Publishing, PP. 183-195, 00 [5] orefice S, Heinonen M and Madec Bilateral comparisons of hydrometer calibrations between the IMGC-NE and the IMGC-MIKES, Metrologia, vol. 37, PP.141 7, 000 [6] Y. J. ee, K. H. Chang, J. C. Chon and C. Y. Oh, Automatic alignment method for calibration of hydrometers, Metrologia, vol. 41, 004. [7] Salvatore orefice and Andrea Malengo, Calibration of hydrometers, Meas. Sci. echnol, vol. 17, PP. 560 566, 006 [8] F. W. Cuckow, A new method of high accuracy for the calibration of reference standard hydrometers, J.S.C.I, 1949. [9] A Picard, R S Davis, M G aser and K Fujii, Revised formula for the density of moist air (CIPM-007), Metrologia, 45, PP. 149 155, 008. [10] Picard A and Fang H, hree methods of determining the density of moist air during mass comparisons, Metrologia, PP. 39 31 40, 00 [11] Picard A and Fang H, Methods to determine the density of moist air, IEEE. rans. Instrum. Meas, vol. 5, PP. 504 7, 003 [1] S. orefice, A. Malengo, An image processing approach to calibration of hydrometers, Metrologia, 41, 004 [13] Jesus Aguilera, John D Wright and Vern E Bean, Hydrometer calibration by hydrostatic weighing with automated liquid surface positioning, IoP Publishing,vol. 19,008 [14] Ilya Kuselman and Ales Fajgelj, Key metrological issues in proficiency testing-response to metrological compatibility-a key issue in future accreditation, Springer, 011. Copyright to IJIRSE www.ijirset.com 1741