Calibration of temperature sensors within Length Standard Section of NMIJ

Similar documents
CERTIFICATE OF CALIBRATION

EVALUATION OF THE ACCURACY OF A THERMOMETRIC RESISTANCE BRIDGE TO DIFFERENT VALUES OF ATMOSPHERIC PRESSURE

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

Certificate of Accreditation to ISO/IEC 17025:2005

Proper Platinum Resistance Thermometer Calibration Uncertainty Analysis

BUREAU INTERNATIONAL DES POIDS ET MESURES

A Case for Periodic Calibration or Verification of RTDs

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon

ECNG3032 Control and Instrumentation I

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators

Calibration of Temperature Block Calibrators

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Temperature measurement

Study on a thermal diffusivity standard for the laser flash method measurements 1

European Association of National Metrology Institutes

Argon Triple-Point Device for Calibration of SPRTs

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN

PRECISION MEASUREMENT OF SELF-HEATING EFFECT OF RESISTANCE THERMOMETERS INSTALLED ON MATERIAL ARTIFACTS

473 Dew Point Hygrometer

Temperature Measurement

BUREAU INTERNATIONAL DES POIDS ET MESURES

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

473 Dew Point Hygrometer

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer

g/cm³ DAkkS Calibration Laboratory for Temperature and Density High-precision calibration for temperature and density

Maximum Performance From Dry-Well Thermometer Calibrators. Mark Finch Product Manager EMEA Fluke Calibration

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

Ajchara Charoensook, Chaiwat Jassadajin. National Institute of Metrology Thailand. Henry Chen, Brian Ricketts and Leigh Johnson

Combined method for establishment and dissemination of the international temperature scale

Approximation of ITS-90 with High Temperature Thermometers up to 1085 C

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity

Towards SI traceable humidity measurements with radiosondes

ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS

How do you really know what the temperature is? Michael de Podesta

EVALUATION OF THE UNCERTAINTY IN THE REALIZATION OF THE PLTS-2000 AT NMi

MEASUREMENTS OF ISOBARIC HEAT CAPACITY OF LIQUID PENTAFLUOROETHANE (R125)

Designing a Thermostat Worksheet

Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements

473-SHX Dew Point Mirror

INTERNAL DOCUMENT No Errors introduced by the standard resistance in bridge thermometry measurements. T J P Gwilliam & S B Keene

An Investigation of the Magneto-Resistance and Quantum Phase Transitions of Cryogenic Sensors

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Abstract. 1. Introduction

Temperature. 3

Thermometry at Low and Ultra-low Temperatures

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

REPORT to the CCT on Comparison COOMET. T-K3 (COOMET theme No. 285/RU-а/03) Final Report

Final Report August 2010

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

Traceability of temperature measurements in Estonia

International Temperature Scale of 1990 From Wikipedia, the free encyclopedia

IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH

CERTIFICATE OF ACCREDITATION

SIMPLE LIQUID NITROGEN COMPARATOR MODEL 461 User Maintenance Manual/Handbook

Introduction to Blackbody Sources

Good morning everyone, and welcome again to MSLs World Metrology Day celebrations.

RADIOMETRIC COMPARISON BETWEEN A NATIONAL LABORATORY AND AN INDUSTRIAL LABORATORY

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

Critical parameters of

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

CCT-K2.5: NRC/NMIJ/INRIM comparison of capsuletype standard platinum resistance thermometers from 13.8 K to K

Received: 17 December 2009 / Accepted: 11 March 2010 / Published online: 10 April 2010 Springer Science+Business Media, LLC 2010

PGA309 Programmable Sensor Signal Conditioner. Bringing Linearity to a Non-Linear Pressure Sensor World

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

Temperature Measurements

Final Report for the APMP.T-K4 (Draft B on October 27, 2011)

SEN TRONIC AG 1 A 6 6 / "

THM80X Series Feature Application. Temperature & Humidity / Flow Measuring Specialist

The Low Temperature Physics of Thin Films Superconducting Tin and Monolayer Graphene

INTER-COMPARISONON TEMPERATURE MEASUREMENT USING FLUKE THERMOMETER MODEL 1502A

Exercise 1: Thermocouple Characteristics

Topic 1.2 Measurement and Uncertainties Uncertainty and error in measurement. Random Errors

Temperature and Length. Pieter Greeff October 2013

Infrared Thermometer Calibration A Complete Guide

Unconventional Double-Current Circuit for Deflection and Temperature Simultaneous Measurement

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Section 7. Temperature Measurement

5.5. Calibration and Test Procedures

The kelvin: new definition, selected mises-en-pratique and direct link with SMD activities

Sensors and Actuators Sensors Physics

TEMPERATURE COEFFICIENTS OF TOPCON RADIATION THERMOMETERS

Metrology and Environment

For the National Voluntary Laboratory Accreditation Program

Temperature measurement

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

CALYS 80P. Field pressure calibrator with high accuracy 0.01%

Uncertainty sources of reference measurement procedures for enzymes

Cryogenic Instrumentation I Thermometry OUTLINE Thermometry Pt (pure metal) Temperature Ranges of Thermometer Application Typical Resistive Thermal

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany

Transcription:

Calibration of temperature sensors within Length Standard Section of NMIJ Dr. Akiko Hirai Length Standard Section National Metrology Institute of Japan (NMIJ) 1

Contents Temperature measurement system with PRT and thermometry bridge Comparative calibration of temperature sensors within Length Standard Section Calibration system Calibration procedures Example of calibration result Uncertainty analysis of temperature sensors 2

Temperature measurement by Pt resistance thermometers and precision thermometry bridge Group of SPRTs (100 ) Standard resistor (100 ) Thermometry bridge PRTs (100 ) 3

Temperature measurement by Pt resistance thermometer and resistance thermometry bridge R 1 R 3 R1 R 3 R 2 R 4 R 2 R 4 4

Temperature measurement by Pt resistance thermometer and resistance thermometry bridge R 1 R 3 R R 1 3 R x R s R x R s R R x s R R 1 3 PRT Standard resistor When R x R s 100 1 x 10-4 1 ppm 0.25 mk 5

Traceability of temperature sensor National primary standard, NMIJ Triple point of water Quantum Hall effect resistance standard Accredited calibration laboratory Triple point of water Melting point of Gallium Standard resistor 1 SPRT PRT Thermometry bridge Standard resistor 100 Length standard section, NMIJ 6

Calibration system of temperature sensors within Length Standard Section Calibrate PRT sensors with reference to SPRT Required calibration uncertainty of PRT is several mk Just a few degree of temperature range around 20 is enough Linear approximation of relationship between temperature and resistance is satisfactory 7

History of calibration system of temperature sensors within Length Standard Section Cu blocks for comparative calibration of different types of temperature sensors 8

History of calibration system of temperature sensors within Length Standard Section 1st Generation Dewar flasks (1997~2001) 2nd Generation Thermal insulation box (2002~2005) 3rd Generation Temp. controlled Cu block (2006~Present) 9

Calibration system of temperature sensors, 1G Double Dewar flasks Change temperature by surrounding heating pads and refrigerants Temperature change was too fast Caused temperature gradient Heating pad Double Dewar flasks Refrigerant 10

Calibration system of temperature sensors, 2G Cupper block in double Dewar flasks set in thermal insulation box Change temperature by air conditioner of laboratory Temperature change was slow enough Too slow! (a few weeks in total!) Double Dewar flasks in thermal insulation box 11

Calibration system of temperature sensors, 3G Cu block in Cu chamber Actively control temperature of block by Peltier controller Temperature was set stable at each calibration point SPRT PRT Cu chamber Water Insulator Cu block Cu block in Cu chamber Peltier device 12

Prevention of heat penetration Heat penetration through leads Better to lay leads in Cu chamber 13

Calibration procedures -measurement - Attach SPRT and PRT probes in/on Cu block Set Cu block in temperaturecontrolled Cu chamber Change temperature of Cu block and keep it for a while Repeat at several temperature 14

Calibration procedures - analysis - Get resistance ratios of SPRT and PRT, to standard resistor Convert resistance ratio of SPRT into temperature Find coefficients of linear model between resistance ratio and temperature of PRT by least-square algorithm 15

Example of calibration - stepwise change of temp.- 1.084000 1.083000 SPRT PRT_Ch0 1.082000 1.081000 Use 40 data from each plateau Resistance ratio 1.080000 1.079000 1.078000 1.077000 1.076000 1.075000 1.074000 2009/2/27 12:00:00 2009/2/28 0:00:00 2009/2/28 12:00:00 Date, Time 2009/3/1 0:00:00 2009/3/1 12:00:00 2009/3/2 0:00:00 Profile of resistance ratio of SPRT and PRT throughout calibration 16

Example of calibration - convert into temp. - 21 Around 19.0, 19.5, 20.0, 20.5, 21.0 20.5 Temp. SPRT 20 19.5 19 18.5 2009/2/27 12:00:00 2009/2/28 0:00:00 2009/2/28 12:00:00 2009/3/1 0:00:00 2009/3/1 12:00:00 2009/3/2 0:00:00 Date, Time Temperature change of SPRT 17

Example of calibration - resistance ratio vs. temp. - 21.0 20.5 Temp. SPRT 20.0 19.5 T ar b 19.0 18.5 1.074000 1.076000 1.078000 1.080000 1.082000 1.084000 Resistance ratio of PRT_Ch0 Relationship between resistance ratio and temperature 18

Example of calibration -residual error - 0.4 0.3 0.2 Residual error [mk] 0.1 0-0.1-0.2-0.3-0.4 18.5 19 19.5 20 20.5 21 Temperature [K] Residual error after linear approximation 19

Uncertainty factors in calibration of PRT Reference SPRT Calibration at TPW Calibration at melting point of Ga Interpolation by ITS-90 Self heating Standard resistor Calibration Temperature dependency Thermometry bridge Repeatability Complement Linearity Calibration procedure Residual error of linear approximation Repeatability Temperature uniformity of Cu block 20

Evaluation of each uncertainty factor Reference SPRT Calibration at TPW 0.5 mk Provided in calibration certificate Calibration at melting point of Ga 1.0 mk Provided in calibration certificate Interpolation by ITS-90 1.0 mk Estimate propagation from TPW and MP of Ga Self heating 0.25 mk Evaluate by changing measurement current 21

Evaluation of each uncertainty factor (Cont'd) Standard resistor Calibration 5.5 x 10-6 0.01 mk Calculate from calibration certificate Temperature dependency 0.6 x 2 ppm/ =1.2 ppm 0.3 mk Calculate from specification 22

Evaluation of each uncertainty factor (Cont'd) Thermometry bridge Error sources: ADC, amplifier, switches, and stray conductance in the electronic circuits Repeatability 1.5 ppm 0.4 mk Evaluate by repeating reconnection of standard resistors Complement 5.5 ppm / 3 0.8 mk Evaluate by exchanging Rs and Rx Linearity 6 ppm / 3 0.9 mk Evaluate from check data by manufacturer 23

Evaluation of complement of thermometry bridge R 1 R 3 R A Standard resistor A R B Standard resistor B 24

Evaluation of complement of thermometry bridge (Cont'd) Assume complement error of the bridge as. RA r and RB R RA r Therefore, RB RB 1 1 r RA RA r 2 2 B 1 1 2 25

Evaluation of complement of thermometry bridge (Cont'd) Standard resistor I: 100.000739 Standard resistor II: 99.99948 Standard resistor III: 99.99984 Resistance ratio A B I II III I 0.999980 0.999985 II 1.000010 1.000001 III 1.000005 0.999990 26

Evaluation of complement of thermometry bridge (Cont'd) Combination (II / I) x (I / II) - 1 Complement error Average Standard deviation of 10 measurements -1.1 x 10-5 1.6 x 10-6 (III / I) x (I / III) - 1 (II / III) x (III / II) - 1-1.0 x 10-5 -9.2 x 10-6 1.3 x 10-6 1.9 x 10-6 5 6 1.1 10 / 2 5.5 10 5.5 ppm / = 3.2 ppm 3 Corresponds to 0.8 mk 27

Another uncertainty factor of thermometry bridge The error associated with the digits of binary and decimal divider stages in bridge. 1st digit, 0~9 R 1 2nd digit, 0~9 3rd digit, 0~9 Variation in contact resistance 'Sawtooth' shaped error 28

'Sawtooth' error of thermometry bridge Measured resistance ratio 1.081 1.080 1.079 Error from difference of contact resistances between '7' and '8' r Actual Ideal 19.6 19.8 20.0 20.2 Temperature [ ] 29

Measured resistance ratio of SPRT 1.081 1.080 1.079 'Sawtooth' error of thermometry bridge (Cont'd) Ideal One is over 1.08, the other is under 1.08 Actual r r r 1.079 1.080 1.081 Measured resistance ratio of PRT Both are in the same range 30

Resistance ratio 'Sawtooth' error of thermometry bridge (Cont'd) Discontinuous change occurs when the ratio acrosses 1.08 1.09 1.088 1.086 1.084 1.082 SPRT PRT_Ch0 Residual error 4.0 2.0 0.0-2.0-4.0 Residual error [mk] 1.08-6.0 1.078 2005/8/2 22:00 2005/8/3 4:00 2005/8/3 10:00 2005/8/3 16:00 Date, Time 2005/8/3 22:00-8.0 2005/8/4 4:00 2005/8/4 10:00 Residual error of linear approximation obtained from mixed data with different digits. 31

Evaluation of each uncertainty factor (Cont'd) Calibration procedure Residual error of linear approximation 0.2 mk Evaluate by data Repeatability Temperature uniformity of Cu block 0.3 mk Evaluate by repeating measurement while changing PRT position Set control limit of uncertainty of calibration procedure as 1.5 mk in standard deviation. 32

Uncertainty budget Category Factor u(xi) ci ui SPRT TPW 0.5 mk 1 0.5 mk Melting point of Ga 1.0 mk 1 1.0 mk ITS-90 1.0 mk 1 1.0 mk Self heating 0.25 mk 1 0.25 mk Standard resistor Calibration 5.5 x 10-6 2.5 K/ 0.01 mk Temperature dependence 0.6 0.5 mk / 0.3 mk Bridge Repeatability 1.5 ppm 0.25 mk / ppm 0.4 mk Complement 5.5 ppm / 3 0.25 mk / ppm 0.8 mk Linearity 6 ppm / 3 0.25 mk / ppm 0.9 mk Calibration Residual error 1.5 mk 1 1.5 mk Repeatability Uniformity Combined standard uncertainty 2.5 mk 33

Other uncertainty factors when using PRT Secular change of PRT Self heating of PRT Contact to the artefact Temperature distribution Temporal change of temperature 34

Conclusion Calibration system and procedures of temperature sensors within Length Standards Section are introduced. Just a few degree of temperature range around 20 is enough Linear model is enough Uniformity and slow change of temperature of Cu block are the most important Uncertainty analysis Evaluate uncertainty of thermometry bridge 35

References Evaluation of thermometry bridge R. Walker, "Automatic linearity calibration in a resistance thermometry b ridge," Int. J. Thermophys. 32, 215 (2011). D.R. White, "A method for calibrating resistance thermometry bridges," Success in the 21st century depends on modern metrology, 1-2, 471 (1997). 36

Thank you for your kind attention. 37