The Uncertainty of Flow Meter Measurement. Mr.Surachart Changpanich Flowlab & Service Co.,Ltd.

Similar documents
FLOW MEASUREMENT VERIFYING YOUR VALUE Small Volume Provers

Uncertainty in Gas Measurement, How does it affect me? Mark Johnson AGIT Gas Speak Colloquium 2016 PROVIDING CONFIDENCE IN MEASUREMENT QUALITY

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES

Analytical Measurement Uncertainty APHL Quality Management System (QMS) Competency Guidelines

A Case for Periodic Calibration or Verification of RTDs

What is measurement uncertainty?

ALASTAIR MCLACHLAN Applications Cameron

NEW VSL Conditioning plate!

Introduction to the evaluation of uncertainty

The Uncertainty of a Waterdraw Calibration vs. Gravimetric Calibration on Small Volume Provers

CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS. A.Gnanavelu

Uncertainty of Coriolis-Based LACT Units

NORTH SEA FLOW MEASUREMENT WORKSHOP 2004 In. St Andrews, Scotland

Guidelines for the Selection and Operation of Provers with Micro Motion ELITE Coriolis Flow Meters

Flow Management Devices, LLC An ISO 9000:2008 Certified Company

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015

How to do a Gage R&R when you can t do a Gage R&R

Process Gas Flow Measurement and Performance Evaluation API TECHNICAL REPORT 2571 SECOND EDITION DRAFT, FEBRUARY 2018

57:020 Mechanics of Fluids and Transfer Processes OVERVIEW OF UNCERTAINTY ANALYSIS

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

Stage 4 G-03: CONDUCTIVITY OF SOLUTIONS CP: USP. May 2015 INTRODUCTION

Measurement And Uncertainty

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

Fine Tuning Coriolis Flow Meter Calibrations Utilizing Piece- Wise Linearization

ITTC Recommended Procedures

ik () uk () Today s menu Last lecture Some definitions Repeatability of sensing elements

Application of Density Meters to Liquid Hydrocarbon Measurement

UNCERTAINTIES FOR PRESSURE-TIME EFFICIENCY MEASUREMENTS

Probability & Statistics: Introduction. Robert Leishman Mark Colton ME 363 Spring 2011

Why is knowledge on measurement uncertainty so important in setting policies on energy efficiency? Rainer Stamminger & Christoforos Spiliotopoulos

Real-Time Optimization (RTO)

Custody Transfer Measurement and Calibration Round Robin Testing for Natural Gas with Coriolis Meters

Smart DP Flow Metering A summary of Established Technology and Latest Innovation. Gary Fish

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY

FINAL DRAFT: October 25, 2010 API TECHNICAL REPORT FUEL GAS MEASUREMENT. FINAL DRAFT: October 25, 2010 Page 0

VALIDATION OF ANALYTICAL METHODS. Presented by Dr. A. Suneetha Dept. of Pharm. Analysis Hindu College of Pharmacy

Traceability of on-machine measurements under a wide range of working conditions

Process Control and Instrumentation Prof. D. Sarkar Department of Chemical Engineering Indian Institute of Technology, Kharagpur

Precision and Accuracy Assessing your Calorimeter

Precision and Accuracy Assessing your Calorimeter

Measurement of fluid flow in closed conduits

ALLOWAY METHOD OUTLINE

EAS 535 Laboratory Exercise Weather Station Setup and Verification

FUNDAMENTAL CONCEPTS IN MEASUREMENT & EXPERIMENTATION (continued) Measurement Errors and Uncertainty:

Review of CMC Template. Yoshiya Terao (NMIJ Fluid Flow Div.) WGFF Workshop, June 2012 Abridged for APMP/TCFF meeting, Nov. 2012

New Methodologies for Instrument Set point and Loop Uncertainty Calculations

Guide to the Expression of Uncertainty in Measurement (GUM)- An Overview

Experiment (4): Flow measurement

Heat Exchangers for Condensation and Evaporation Applications Operating in a Low Pressure Atmosphere

Experimental Uncertainty (Error) and Data Analysis

Model 2300XP PSL & Process-Particle Wafer Deposition System

Statistical inference

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

ents & Uncertainties Significant Figures 1.005, Round best to the experimental want to meters and use 300 m 2. significant figures because of

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

1.4 System Accuracy* B. G. LIPTÁK (1982, 1995, 2003) DEFINITIONS OF TERMS

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Measurements & Instrumentation

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty

Verification for Non-Automatic Weighing Instruments (NAWI)

2.4 The ASME measurement-uncertainty formulation

IMPORTANT FEATURES OF ESTIMATION & INTERPRETATION OF MEASUREMENT UNCERTAINTY. B. Bhattacharya, K. Ramani, P.H. Bhave

Appendix B: Accuracy, Precision and Uncertainty

Traffic Forecasting Risk: Study Update 2004

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

) where the inlet pressure ( P 1

Potential Fields Pool Equipment (PFPE) Focus Group. James Kinsey Randy Herr Dan Fornari

Inspection of Welds in Small Pipes by NDT Ultrasonic Phased-Array techniques. Detection, positioning, sizing and classification of defects.

Doubt-Free Uncertainty In Measurement

Technical Paper 3 Verifying Particle Counter Calibration: Size Verification

ITS. One of the highest resolutions for particle sizing and counting. Multisizer 3 COULTER COUNTER SCIENCE

ALIGNMENT OF THE MSGC BARREL SUPPORT STRUCTURE

Fuel Consumption Measurement Uncertainty How can I evaluate the quality of my measurement results?

Procedure for Uncertainty Determination for Calibration Consoles

Lecture 2. Estimating Single Population Parameters 8-1

Paper # # Verification of Coriolis Flow

Manual of Petroleum Measurement Standards Chapter 5 Metering

ACCURACY AND PRECISION

Chapter 8 Statistical Quality Control, 7th Edition by Douglas C. Montgomery. Copyright (c) 2013 John Wiley & Sons, Inc.

Statistical Analysis of Engineering Data The Bare Bones Edition. Precision, Bias, Accuracy, Measures of Precision, Propagation of Error

PHYS 212 PAGE 1 OF 6 ERROR ANALYSIS EXPERIMENTAL ERROR

DYNAMIC LOAD COMPENSATION

Chem 321 Lecture 4 - Experimental Errors and Statistics 9/5/13

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

Measurements and Data Analysis

White Paper. Perform Conductivity Measurements In Compliance with USP <645>

CORIOLIS MASS FLOW METERS FOR GAS AND LIQUID MEASUREMENT. Michael Keilty (Presented by Hubbard Donald) Endress + Hauser Flowtec AG

Vocabulary of Metrology

CE 321 Sample Laboratory Report Packet

TI 226H Calibration of Flow Check Devices using Positive Displacement Flow Meter

Instrument types and performance characteristics

Source: Chapter 5: Errors in Chemical Analyses

A Unified Approach to Uncertainty for Quality Improvement

Energy Forecasting Customers: Analysing end users requirements Dec 3rd, 2013 Carlos Alberto Castaño, PhD Head of R&D

Reference Guide. Science Reference 9/25/ Copyright 1996 Gary Lewis Revisions 2007 by John Pratte

Urban Legends in Gasmetering

PRACTICAL CONTROL OF DYNAMIC METERING SYSTEM AND DATA ACQUISITION

Rocky Mountain Measurement Seminar Denver August 20, 2015 Coriolis Meters

Transcription:

The Uncertainty of Flow Meter Measurement By Mr.Surachart Changpanich Flowlab & Service Co.,Ltd. Email : surachart.flowlab@outlook.com Mobile Phone : 081-8617005 Line ID : Surachart_59 NSC-TISI-TIS 17025 Calibration 0126 1

Introduction Whenever a dynamics measurement of fluid flow is made, the value obtained is simply the best estimate that can be obtained of the flow rate or quantity. In practice, the flow rate or quantity could be slightly greater or less than this value, the uncertainty characterizing the range of values within which the flow rate or quantity is expected to lie that could reasonably be attributed to the measurement, with a specified confidence level. 2

Calibration The process of using a reference standard to determine a calibration factor.calibration adjusts the output of meter to bring it to a value which is within the specified accuracy tolerance.this process is normally conducted by the meter manufacturer. Purpose of Flow Meter Calibration Improve measurement accuracy Product is valuable Checks the condition of equipment Trading Follow to regulation 3

Meter Factor = Standard value Meter s indicated value Meter Accuracy Factor Meter Factor Compensation Meter s value Standard value METERED VALUE X MF = PRECISE VALUE (Simplified) 4

Repeatability means just what it say - how nearly the same reading a meter will provide for a given flow condition. % Repeatability = [(MF.max / MF.min) 1] x 100 5

GOOD REPEATABILITY DOES NOT NECESSARILY MEAN GOOD ACCURACY 6

POOR REPEATABILITY MEANS POOR ACCURACY 7

GOOD REPEATABILITY TO BE ADJUSTED FOR THE RESULT IS GOOD ACCURACY 8

AFTER ADJUSTMENT THE ACCURACY IS GOOD 9

An important point in this discussion is that the meter can insure accurate measurement, even with marginally performing meter. The reliability which can be represented to measurement parameter following: Linearity is meter accuracy within a Rangeability expresses the flow range over which a meter operates while meeting a stated accuracy tolerance. It is often stated as Turndown maximum flow divided by minimum flow over the range as 250-2500 LPM, the range ability is 10:1. % Linearity = ½ x [(MF.max - MF.min)] x 100 Avg.MF 10

Stability: The indication is when the meter factors maintain stability over time by continuously having values within a set tolerance. METER FACTOR SEQUENCE / METER FACTOR STABILITY 11

STABILITY ESSENTIAL TO ACHIEVE A HIGH CONFIDENCE LEVEL BEST INDICATION THAT THE METERING SYSTEM IS UNAFFECTED BY RANDOM ERRORS A STATISTICAL RECORD OF METER PROVINGS IS THE BEST MEANS TO ESTABLISH THE LEVEL OF SYSTEM UNCERTAINTY FOR A METERING INSTALLATION 12

Accuracy VS Uncertainty $$$$$ Accuracy : The closeness in the agreement between the result of a measurement and the true value of the measurement. The quantitative expression of accuracy is in terms of uncertainty. Good accuracy implies small random and systematic errors. Uncertainty : The parameter, associated with the results of a measurement, that characterizes the dispersion of the values that could reasonably be attributed to the measurement. 13

Evaluation of the uncertainty in a measurement process The first stage in an uncertainty evaluation is to define the measurement process. For the measurement of flow rate, it will be normally be necessary to combine the values of a number of input quantities to obtain a value for output. The definition of the process should include the enumeration of all the relevant input quantities. The sources of introducing uncertainty in a measurement process can be divided arbitrarily into the following categories: a) Calibration uncertainty; b) Data acquisition uncertainty; c) Data processing uncertainty; d) Uncertainty due to methods; e) Others 14

The effecting cost of inaccuracy Application for pipe line transfer or marine loading which throughput of 500 m 3 /h. The cost per m 3 = 26,890 THB. (HSD _5 Nov 15) Value of each 0.XX % of bias. Example 1. For 0.15% of bias of Marine loading for 8 h/day ; a) THB 161,340/Day b) THB 58,889,100/Year 2. For 0.10% of bias of marine loading for 8 h/day ; a) THB 107,560/Day b) THB 39,259,400/Year The deviation 0.05 % which to be save: THB. 19,629,700 per year. 15

The sources of uncertainty can be classified as following; Random uncertainty : Type A evaluation Systematic uncertainty : Type B evaluation Random uncertainty The sources of errors that cause a variation in output reading even when the input parameter has no changed. Type A evaluations of uncertainty are those using statistical methods, specifically, those that use the spread of a number of measurement. An estimate characterizing the range of values within which the true measured value of a quantity lies and how frequently the reading does lie within this range per confidence level is 95% Evaluation: Custody measurement starts with verification or proving a meter to a repeatability of 0.05% of 3 consecutive runs. The evaluation is an uncertainty of +/- 0.073% at 95% confidence level. 16

Accurate Measurement The accuracy of measurement is +/- 0.1% repeatable 20 out of 20. 100 % confidence level. Zero bias 17

Custody Measurement +/- 0.1% repeatable 19 out of 20 95 % confidence level. Zero bias 18

Repeatable occur a bias error of measurement. 0.4% +/- 0.1% repeatable 19 out of 20. 95 % confidence level. 0.4% bias error error 19

Poor repeatability of measurement. +/- 0.3% repeatable 19 out of 20. 95 % confidence level. Zero bias 20

21

22

Random Uncertainty Statement 23

Systematic uncertainty Type B evaluations of uncertainty require a knowledge of the probability distribution associated with the source of error that relate with systematic within this range per confidence level such as; a) Uncertainty due to methods; Tank Prover, Master Meter, SVP. b) Data acquisition uncertainty; Flow Meter, Temperature, Pressure, Density, and Electronic Batch controller. c) Data processing uncertainty; Product Calculation. d) Others 24

25

Q/A; Thank you 26