Urban Legends in Gasmetering
|
|
- Jody Bryant
- 5 years ago
- Views:
Transcription
1 Urban Legends in Gasmetering Jos van der Grinten Scientist High Pressure Gas Department
2 Enquiry three statements A national standards institute can disseminate the m³ accurately without re-calibration. Yes / No A calibration lab can measure accurately without re-calibration. Yes / No Your gasmeter can measure accurately without re-calibration. Yes / No 8-9 June
3 Contents Urban legends Definitions and terminology Cases Calibration on air, application with natural gas Forward curve adjustment used for reverse application No recalibration required when using two meters in series Calibration up to 70% extrapolation to 100% Z configurations 50 bar calibration for 60 bar operation How to deal with statements Conclusions and recommendations 8-9 June
4 Urban legends Broodje Aap. De folklore van de post-industriële samenleving. (Urban legends. The folklore of the post industrial society.) Characteristics: Fictional, told as true Sources are missing No exact data, names, locations Sensational or bizarre story Latent fears or prejudices Pseudoscientific nonsense Alternative facts / fact-free journalism Ethel Portnoy ( ) 1979 Photo: Hans van Dijk - Anefo 8-9 June
5 Definitions and terminology 1 Step 1: Calibration à deviation ± uncertainty reference value Instrument value GUM deviation Uncertainty (k=2) (Guide to the expression of Uncertainty in Measurement, 1993) Uncertainty the range of values that can reasonably be attributed to the measurand a measure for the amount of missing information 8-9 June
6 Definitions and terminology 2 Calibration results on certificate Flowrate [m³/h] Deviation [%] Uncertainty [%] Validity of certificate expires after After the end of the re-calibration period Accident with the meter Modifications made to the meter (parameter list part of the certificate) 8-9 June
7 Definitions and terminology 3 Step 2: conformity assessment à compliance with regulations, e.g. tolerances ( 95% confidence) L K? J K? L tol tol + Test / verification: conforming or not Inspection: non-conforming or not 8-9 June
8 Definitions and terminology 4 Gas meters MID, annex IV (MI 002) Normative document OIML R 137 TSO contracts Class Q min 3% Q < Q t * 2% * 3% 6% 2% 4% 1% 2% Q t Q 1.5% Q max * 1% * 1.5% 3% 1% 2% 0.5% 1% 0.3% 0.3% WME 0.6% 0.4% 0.2% 0.3% 0.3% In-service requirements à national law TSO: high-volume à financial risk à high accuracy 8-9 June
9 Calibrate on air, use on natural gas? year < % 8-9 June
10 Calibrate on air, use on natural gas? year < % 8-9 June
11 Calibrate on air, use on natural gas? year 2016 Meter T Lab A 41 barg As Found Lab A 41 barg polyn. powers -2, -0.2, -0.33, 0 Lab B atm. air As Found Lab B atm. air polyn. powers -2, -0.2, -0.33, 0 error [%] % Flow [m³/h] 8-9 June
12 Calibrate forward Adjust reverse Claim The adjustment and verification of a meter carried out for one direction is sufficient and valid for a usage in both flow directions, if the meter run including input and output pipes is carried out as described in the type examination certificate for a bidirectional usage. In this case the meter provides results which fulfill the MPE independent from the flow direction. Claim is for USM of a specific brand Refers to legal metrology No restrictions to adjustment method 8-9 June
13 Calibrate forward Adjust reverse Error [%] Forward, adj.fact = Reverse, adj.fact. = Error [%] Forward, adj.fact = Reverse, adj.fact. = year year Flowrate [m³/h] Flowrate [m³/h] USMs as found fw & rev Adjust factors 0.80% diff. in % diff. in 2013 < 0.05% diff. in 2016! Error [%] year 2016 Forward, adj.fact = Reverse, adj.fact. = Flowrate [m³/h] 8-9 June
14 Calibrate forward Adjust reverse Newer meters perform better Method only demonstrated for linear adjustments Reckon with at least 0.1% difference between forward and reverse No check for polynomial or multi-point adjustments 8-9 June
15 In series no recalibration? In Situ comparison showed 0,4% difference 10D 5D 10D 5D Slave Manufacturer A Master Manufacturer B In reality average shift = 1%! 8-9 June
16 Extrapolation What to do if you cannot reach Q max? Take volume 30 of PTB Testing Instructions, p 46 Calibrate up to 70% Q max Low pressure gas or air curve in full range must be available Shape of the curve between 0.2 Q max and highest Q Extrapolate in Reynolds domain until 100% Q max, Curve <½ MPE For legal metrology purposes!! Intended for turbine meters,... also used for USMs 8-9 June
17 Extrapolation Result of a recalibration Cal. Lab Often No low-pressure or air curve available (e.g. recalibration) Base extrapolations on smooth curves Higher order polynomials will be unstable outside the calibrated range 8-9 June
18 Extrapolation Turbine gasmeter, 42 bar, natural gas 0.00 As Found As Left Adjusted Fit Error [%] Flow [m 3 /h ] 8-9 June
19 Extrapolation G16000 Turbine gasmeter deviation e (%) Calibration data polyn. powers -0.2, -0.33, -2, add. unc = 0.10% polyn. powers 1, 2, 3, add. unc = 0.15% polyn. powers 1, 2, 3, 4, add. unc = 0.12% 0.10 B-spline, 3rd order, 4 knots, add. unc = 0.04% WME Non-dimensional flowrate Q/Qmax 8-9 June
20 Z configurations Parallel use Serial comparison Points of attention Flow disturbances and flow conditioning 1 Noise 2 (100 db(a)) Pulsations caused by closed side branches June
21 Z configurations Separate calibration of two meters 1 0.5% difference with or without side branch à Recommendation: test the entire configuration Example Hilko den Hollander, EFMWS 2015, Noordwijk Calibration of a complete Z skid and its functionality 8-9 June
22 Pressure dependency 1 Claim 50 bar are cheaper 60 bar So even when operating conditions are 60 bar we order a 50 bar calibration There will be not that much difference? Pressure dependency 8-9 June
23 Pressure dependency Meter U Lab A 48 barg As Left Lab A 60 barg As left 0.5 error [%] % / bar Flow [m³/h] 8-9 June
24 Pressure dependency USM (clean without corrections) % / bar 0.20 Deviation [%] bara as found 41 bara as found 31 bara uncorrected 41 bara Chord B failure bara initial 5 years ago Flow velocity [m/s] 8-9 June
25 How to deal with statements? No recalibration necessary? Only for meters on the scrap yard Change or damage to the meter à recalibrate End of recalibration period à recalibrate Innovative declarations issued by notified bodies Made on request of manufacturers So ask the manufacturer for Magnitude of the change à additional uncertainty Systematic proof, not just one case Publications in the open literature Results of studies and tests preferably conducted by an independent third party 8-9 June
26 Conclusions and recommendations Metrological principle Change anything on the meter and the calibration certificate will become void Legal metrology Some modifications are possible without jeopardizing the performance with respect to legal tolerances Use acceptance criteria smaller than legal tolerances However legal metrology tolerances are too wide for industrial applications (TSOs) Systematic testing and reporting required 8-9 June
27 Conclusions and recommendations Cases Δe Air calibration 0.9% High-pressure air, not documented yet How to scale the meter curve? Meter principle? Forward calibration reverse adjustment > 0.1% Newer meters perform better No recalibration with two meters in series ~ 1% Extrapolation > 0.1% Z configuration > 0.5% Pressure dependency + or % / bar 8-9 June
28 Conclusions and recommendations Currently Urban legends have always been there Alternative facts / fact-free reporting Future Well documented test results à Transparency Transparency à Confidence Confidence à Trust 8-9 June
29 Enquiry change of opinion A national standards institute can disseminate the m³ accurately without re-calibration. Did you change your opinion? A calibration lab can measure accurately without re-calibration. Did you change your opinion? Your gasmeter can measure accurately without re-calibration. Did you change your opinion? Yes / No Yes / No Yes / No 8-9 June
30 Questions? Schrödinger s gasmeter m³ The only way to know is to re-calibrate!! 8-9 June
31 Physikalisch-Technische Bundesanstalt Braunschweig and Berlin Bundesallee Braunschweig Germany Jos van der Grinten Telephone:
The new Closed Loop pigsar and its projected CMC. Jos van der Grinten, Detlef Vieth, Bodo Mickan
The new Closed Loop pigsar and its projected CMC Jos van der Grinten, Detlef Vieth, Bodo Mickan Introduction pigsar Calibration facilities Closed Loop pigsar (CLP) Calibration model Uncertainty analysis
More informationAdjustment procedures and their corresponding uncertainties
Adjustment procedures and their corresponding uncertainties Jos van der Grinten Chief Metrologist, NMi EuroLoop Contents! Introduction! Currently applied adjustment procedures! Residuals! Uncertainties!
More informationCalibration and verification: Two procedures having comparable objectives and results
Calibration and verification: Two procedures having comparable objectives and results KLAUS-DIETER SOMMER, Landesamt für Messund Eichwesen Thüringen (LMET), Germany SAMUEL E. CHAPPELL, Consultant, Formerly
More informationTRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS
TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,
More informationERROR LIMITS AND MEASUREMENT UNCERTAINTY IN LEGAL METROLOGY
ERROR LIMITS AND MEASUREMENT UNCERTAINTY IN LEGAL METROLOGY K.-D. Sommer 1, M. Kochsiek and W. Schulz 1 Landesamt fuer Mess- und Eichwesen Thueringen, German Phsikalisch-Technische Bundesanstalt, German
More informationAPPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY
APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. EVALUATION OF MEASUREMENT UNCERTAINTY POLICY... 5
More informationAPPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT
APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. ESTIMATION OF UNCERTAINTY OF MEASUREMENT POLICY...
More informationMeasuring Instruments Directive (MID) MID/EN14154 Short Overview
Measuring Instruments Directive (MID) MID/EN14154 Short Overview STARTING POSITION Approval vs. Type examination In the past, country specific approvals were needed to sell measuring instruments in EU
More informationCertification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60
Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60 Bernd Meißner Physikalisch-Technische Bundesanstalt Weighing Instruments Laboratory Braunschweig Abstract
More informationTraceability of on-machine measurements under a wide range of working conditions
Traceability of on-machine measurements under a wide range of working conditions Frank Keller, Matthias Franke, Norbert Gerwien, Lisa Groos, Christian Held, Klaus Wendt Traceable in-process dimensional
More informationINTERNATIONAL OIML D 25 DOCUMENT. Vortex meters used in measuring systems for fluids ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL OIML D 25 DOCUMENT Edition 1996 (E) Vortex meters used in measuring systems for fluids Compteurs à vortex utilisés dans les ensembles de mesurage de fluides OIML D 25 Edition 1996 (E) ORGANISATION
More informationTechnical Protocol of the CIPM Key Comparison CCAUV.V-K5
Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 2017-03-06 revised 2018-02-13 (changed schedule) Task and Purpose of the Comparison According to the rules set up by the CIPM MRA the consultative
More informationNEW VSL Conditioning plate!
Calibration of Prover Tanks using a Coriolis Mass Flow Meter as the Master Meter. Is this possible for Pipe Prover calibration too? Erik Smits (fsmits@vsl.nl) 2 nd European Flow Measurement Workshop Sintra,
More information2.1. Accuracy, n- how close the indication of the thermometer is to the true value.
AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria
More informationTechnical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1
Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Updated at 12-December-2012 Task and Purpose of the Comparison According to the rules set up by the CIPM
More informationFundamentals of Mass Determination
Fundamentals of Mass Determination Michael Borys Roman Schwartz Arthur Reichmuth Roland Nater Fundamentals of Mass Determination 123 Michael Borys Fachlabor 1.41 Physikalisch-Technische Bundesanstalt Bundesallee
More informationTraceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE
Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results David MILDE Traceability Validation Measurement Uncerainty Outline Introduction to 3 pillars Recommended
More informationAppendix B1. Reports of SMU
Appendix B1 Reports of SMU Euromet Project 600 Comparison of Surface Roughness Standards /13 EUROMET SUPPLEMENTARY COMPARISON SURFACE TEXTURE Project No. 600 Final Report Elaborated by: M. Szmicskova Bratislava,
More informationDetermination of particle size distribution Single particle light interaction methods. Part 4:
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 21501-4 Second edition 2018-05 Determination of particle size distribution Single particle light interaction methods Part 4: Light scattering airborne
More informationTest report CVN Project number Page 2 of 2
Project number 604545 Page 2 of 2 Tests : Results : The diaphragm gas meters as specified in annex 2 were tested for compliance with the European Standard BS EN 1359 : 1999 "Gas meters - Diaphragm gas
More informationMeasurement & Uncertainty - Basic concept and its application
Measurement & Uncertainty - Basic concept and its application Feb. 7. 2012 JongOh CHOI (choijongoh@kriss.re.kr) Knowledge and measurement When you can measure what you are speaking about, and express it
More informationA Unified Approach to Uncertainty for Quality Improvement
A Unified Approach to Uncertainty for Quality Improvement J E Muelaner 1, M Chappell 2, P S Keogh 1 1 Department of Mechanical Engineering, University of Bath, UK 2 MCS, Cam, Gloucester, UK Abstract To
More informationTITLE OF THE CD (English): New OIML Document
OIML TC 3/SC 5 1CD - New OIML Document COMMITTEE DRAFT OIML CD1 Date: 27 April 2009 Reference number: OIML/TC 3/SC 5/N1 Supersedes document: Not applicable OIML TC 3 / SC 5 Project p2 Title: The role of
More information) where the inlet pressure ( P 1
Cone DP Meter Calibration Issues Authors: Casey Hodges, Charles Britton, William Johansen & Richard Steven CEESI, 54043 Weld County Road 37, Nunn, CO 80648 Email: chodges@ceesi.com, Telephone: US 970-897-2711
More informationMeasurement Uncertainty Knowing the Unknown
Measurement Uncertainty Knowing the Unknown Anish Shah FOUNDER and Chief Metrology Officer at Metrologized, LLC, a company founded in 2014 to further the metrology knowledge-base within the manufacturing
More informationCalibration of Temperature Block Calibrators
European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)
More informationWhy is knowledge on measurement uncertainty so important in setting policies on energy efficiency? Rainer Stamminger & Christoforos Spiliotopoulos
Why is knowledge on measurement uncertainty so important in setting policies on energy efficiency? Rainer Stamminger & Christoforos Spiliotopoulos Agenda What is the problem? Differentiating targets Sources
More informationProvläsningsexemplar / Preview. First edition
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 12764 First edition 2017-08 Measurement of fluid flow in closed conduits Flowrate measurement by means of vortex shedding flowmeters inserted in
More informationUncertainty of Measurement A Concept
Uncertainty of Measurement A Concept Shweta V. Matey 1, Dr. Nitin K. Mandavgade 2, and Dr. R.R. Lakhe 3 1 (Asst. Professor, Dept. of Mechanical Engg, Lokmanya Tilak College of Engineering, Mumbai University,
More informationGuide to the Expression of Uncertainty in Measurement (GUM) and its supplemental guides
National Physical Laboratory Guide to the Expression of Uncertainty in Measurement (GUM) and its supplemental guides Maurice Cox National Physical Laboratory, UK maurice.cox@npl.co.uk http://www.npl.co.uk/ssfm/index.html
More informationProvläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition
INTERNATIONAL STANDARD ISO 17123-4 Second edition 2012-06-01 Optics and optical instruments Field procedures for testing geodetic and surveying instruments Part 4: Electro-optical distance meters (EDM
More informationMetrological Characterization of a Primary Vickers Hardness Standard Machine - NIS Egypt
MAPAN Metrological - Journal of Metrology Characterization Society of a India, Primary Vol. Vickers 5, No. Hardness, 00; pp. Standard -7 Machine - NIS Egypt ORIGINAL ARTICLE Metrological Characterization
More informationMeasurement & Uncertainty - Concept and its application
Measurement & Uncertainty - Concept and its application June. 20 2012 JongOh Choi (choijongoh@kriss.re.kr) I don t know what I don t know. If we don t express what we know in the form of numbers, we really
More informationAchieving traceability of industrial computed tomography
Achieving traceability of industrial computed tomography Markus Bartscher, Marko Neukamm, Uwe Hilpert, Ulrich Neuschaefer-Rube, Frank Härtig, Karin Kniel, Karsten Ehrig*, Andreas Staude*, Jürgen Goebbels*
More informationOIML R 141 RECOMMENDATION. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL RECOMMENDATION OIML R 141 Edition 2008 (E) Procedure for calibration and verification of the main characteristics of thermographic instruments Procédure pour l'étalonnage et la vérification
More informationMEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration
MEASUREMENT PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE UNCERTAINTY OCTOBER/NOVEMBER 2012 Prepared by for ENAO Assessor Calibration B SCOPE Introduction House Rules Overview 17025 & 15189 MU Approaches
More informationWithdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST
DEUTSCHER KALIBRIERDIENST Guideline DKD-R 5-7 Calibration of Climatic Chambers Edition 07/2004 English translation 02/2009 Page 1 of 31 Published by the Accreditation Body of the Deutscher Kalibrierdienst
More informationDecision rules applied to conformity assessment
Technical Report No. 01/ 2017 January 2017 Decision rules applied to conformity assessment Technical Report 1 Impressum EUROLAB Technical Report 1/2017 Decision rules applied to conformity assessment Authors:
More informationRevision of CMM performance specifications
Educational Articles: Article nr. 3 Revision of CMM performance specifications Premise The working group (WG) Specifications is one of several working groups of ia.cmm, the International Association of
More informationEstimated Sound Power Level of the ENERCON E-92 Operational Mode I (Data Sheet)
Estimated Sound Power Level E-92 Page 1 of 3 Estimated Sound Power Level of the ENERCON E-92 Operational Mode I (Data Sheet) Imprint Editor: ENERCON GmbH Dreekamp 5 26605 Aurich Germany Telephone: 04941-927-0
More informationOIML R 50-3 RECOMMENDATION. Edition 2014 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL RECOMMENDATION OIML R 50-3 Edition 2014 (E) Continuous totalizing automatic weighing instruments (belt weighers). Part 3: Test report format Instruments de pesage totalisateurs continus à
More informationTechnical Procedure for General Laboratory Equipment
Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all
More information+ + ( indicator ) + (display, printer, memory)
CERTIFICATION OF STRAIN GAUGE LOAD CELL FAMILIES B. Meissner Physikalisch-Technische Bundesanstalt Weighing Instruments Laboratory 38116 Braunschweig, Germany Abstract: Testing and certification of load
More informationThis document is a preview generated by EVS
TECHNICAL SPECIFICATION SPÉCIFICATION TECHNIQUE TECHNISCHE SPEZIFIKATION CEN ISO/TS 15530-3 December 2007 ICS 17.040.30 English Version Geometrical product specifications (GPS) - Coordinate measuring machines
More informationRECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS
The Fifth International Proficiency Testing Conference Timisoara, Romania () th 8 th September, 0 RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS Adriana Vâlcu, Sterică Baicu National Institute of Metrology,
More informationFORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO).
FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO). Jorge C. Torres-Guzmán*, Amritlal Sawla**, Daniel Ramírez-Ahedo* * Centro Nacional de Metrología, México. **Physikalisch-Technische
More informationPart 5: Total stations
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 17123-5 Third edition 2018-02 Optics and optical instruments Field procedures for testing geodetic and surveying instruments Part 5: Total stations
More informationRecent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar
Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) Dr. Steven D. Phillips Precision Engineering Division National Institute
More informationOA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents
UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC 17025 Table of contents 1 GENERAL... 2 2 THE STANDARD SIST EN ISO/IEC 17025... 2 3 SA'S POLICY IN IMPLEMENTING
More informationMeasurement uncertainty and legal limits in analytical measurements
UNCERTAINTY Measurement uncertainty and legal limits in analytical measurements MIRELLA BUZOIANU, Reference Materials Laboratory, National Institute of Metrology (INM), Bucharest, Romania 1 Introduction
More informationOIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air
INTERNATIONAL DOCUMENT OIML D 28 Edition 2004 (E) Conventional value of the result of weighing in air Valeur conventionnelle du résultat des pesées dans l'air OIML D 28 Edition 2004 (E) ORGANISATION INTERNATIONALE
More informationGuidelines on the Calibration of Static Torque Measuring Devices
European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET/cg-14/v.01 Previously EA-10/14 July 2007 Calibration Guide EURAMET/cg-14/v.01
More informationIntercomparison of Measurements of the Thermophysical Properties of Polymethyl Methacrylate 1
International Journal of Thermophysics, Vol. 25, No. 5, September 2004 ( 2004) Intercomparison of Measurements of the Thermophysical Properties of Polymethyl Methacrylate 1 S. Rudtsch 2 and U. Hammerschmidt
More informationE C t y p e - e x a m i n a t i o n C e r t i f i c a t e
E C t y p e - e x a m i n a t i o n C e r t i f i c a t e Page 1 of 1 Issued by Manufacturer NMi Certin B.V., designated and notified by the Netherlands to perform tasks with respect to conformity modules
More informationEA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE
Publication Reference EA-10/14 EA Guidelines on the Calibration of Static Torque Measuring Devices PURPOSE This document has been produced by EA to improve harmonisation in determining the calibration
More informationMETHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone
METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT Scott Crone North American Sales Manager AMETEK Calibration Instruments 8600 Somerset Drive Largo, Florida 33773 USA INTRODUCTION Like any other piece of equipment,
More informationMetallic materials Brinell hardness test. Part 3: Calibration of reference blocks
INTERNATIONAL STANDARD ISO 6506-3 Third edition 2014-10-01 Metallic materials Brinell hardness test Part 3: Calibration of reference blocks Matériaux métalliques Essai de dureté Brinell Partie 3: Étalonnage
More informationASIA PACIFIC METROLOGY PROGRAME
APMP.M.M-K6 (SUB) multiples mass key comparison ASIA PACIFIC METROLOGY PROGRAME Final Report of APMP Comparison of Mass standards APMP.M.M-K6 (SUB) multiples mass key comparison Coordinated by: Vietnam
More informationESTIMATION OF UNCERTAINTY IN HARDNESS MEASUREMENT OF RUBBER AND OTHER ELASTOPLASTIC MATERIALS
XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17, 006, Rio de Janeiro, Brazil ESTIMATION OF UNCERTAINTY IN HARDNESS MEASUREMENT OF RUBBER AND OTHER ELASTOPLASTIC MATERIALS
More informationLocal time scale at PTB UTC(PTB) Ali Al-masoudi, Stephan Falke, Sebastian Häfner, Stefan Vogt, Christian Lisdat
Local time scale at PTB UTC(PTB) Ali Al-masoudi, Stephan Falke, Sebastian Häfner, Stefan Vogt, Christian Lisdat Outline Time scales History of atomic clock at PTB Local time scale UTC(PTB) Time transfer
More informationE v a l u a t i o n C e r t i f i c a t e
E v a l u a t i o n C e r t i f i c a t e Page 1 of 1 Issued by In accordance with Producer WELMEC guide 8.8 General and Administrative Aspects of the Voluntary System of Modular Evaluation of Measuring
More informationQUESTIONS & ANSWERS 1
QUESTIONS & ANSWERS 1 ECO PASSPORT by OEKO-TEX The textile and clothing industries are changing with great speed. Industry associations and NGOs are prodding all elements of the supply chain to adhere
More informationALASTAIR MCLACHLAN Applications Cameron
ALASTAIR MCLACHLAN Applications Cameron Have worked in the oil and gas flow measurement industry for 27 years primarily in custody transfer measurement using ultrasonic meters. Joined Cameron Caldon Ultrasonics
More informationAN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS
WHITE PAPER AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS The Energy Conservatory Minneapolis, MN Introduction: The purpose of this document is to explain the details of the accuracy specifications
More informationPTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION
IMEKO 22 nd TC3, 15 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION R. Kumme
More informationSurface Roughness - Standards and Uncertainty R. Krüger-Sehm and L. Koenders
Surface Roughness - Standards and Uncertainty R. Krüger-Sehm and L. Koenders Physikalisch-Technische Bundesanstalt, Braunschweig, Germany Surfaces Generation - Grinding - Honing - Lapping Characterisation
More informationMechanics and Acoustics
info sheet PHYSIKALISCH-TECHNISCHE BUNDESANSTALT DIVISION 1 Mechanics and Acoustics Mechanics and Acoustics How heavy, how strong, how fast, how loud? Division 1 is your reference for the precision measurement
More informationUncertainties associated with the use of a sound level meter
NPL REPORT DQL-AC 002 Uncertainties associated with the use of a sound level meter Richard Payne April 2004 April 2004 ABSTRACT Uncertainties associated with the use of a sound level meter Richard Payne
More informationCALIBRATION REPORT FOR THERMOHYDROMETER
CALIBRATION REPORT FOR THERMOHYDROMETER SAMPLE CUSTOMER S/N: XXXXX Report Number: Y205188 Page 1 of 5 Report No. Y205188 ICL CALIBRATION LABORATORIES, INC. ISO/IEC 17025 and ANSI/NCSL Z540-1 accredited
More informationINVESTIGATION OF TRANSFER STANDARDS IN THE HIGHEST RANGE UP TO 50 MN WITHIN EMRP PROJECT SIB 63 Falk Tegtmeier 1, Michael Wagner 2, Rolf Kumme 3
XXI IMEKO World Congress Measurement in Research and Industry August 30 September 4, 2015, Prague, Czech Republic INVESTIGATION OF TRANSFER STANDARDS IN THE HIGHEST RANGE UP TO 50 MN WITHIN EMRP PROJECT
More informationECO PASSPORT by OEKO-TEX
QUESTIONS & ANSWERS ECO PASSPORT by OEKO-TEX The textile and clothing industries are changing with great speed. Industry associations and NGOs are prodding all elements of the supply chain to adhere to
More informationINTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR
INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR M. Araya 1, D. Almeida 2 1 National Laboratory of Temperature of Chile, LCPNT, Santiago, Chile 2 Ecuadorian Standardization
More informationCertificate of Calibration Sound Level Calibrator
NATacoustic Acoustic Caliration & Testing Laoratory Level, 48A Elizaeth Street., Surry Hills NSW 200 AUSTRALIA Ph: (02) 828 0570 email: service@natacoustic.com.au wesite: www.natacoustic.com.au A division
More informationPhysikalisch-Technische Bundesanstalt
Physikalisch-Technische Bundesanstalt COOMET Supplementary Comparison COOMET.L-S18 (COOMET project: 673/UA-a/15) Involute gear measurement standards Final report September 2017 Coordinator: Physikalisch-Technische
More informationGuideline DKD-R 4-2. Sheet 3. Edition 7/2011.
Guideline DKD-R 4- Sheet 3 Calibration of instruments and standards for Calibration of standards with periodic profiles in horizontal direction by means of stylus instruments Edition 7/0 http://dx.doi.org/0.7795/550.07069en
More informationCorrect Traceability in Flexible and Complex Dimensional Metrology. Uncertainty in coordinate metrology: - Needs - Standards.
Measurement service and metrological development on request in the Technology Park of Gijón (Asturias) Here is the laboratory of ISM3D, 0 m below the ground level Correct Traceability in Flexible and Complex
More informationMorehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:
Morehouse 1 Morehouse Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA 17403 PH: 717-843-0081 web: www.mhforce.com sales: edlane@mhforce.com 2 This presentation will cover the calibration
More informationCERTIFICATE OF CALIBRATION
Page 1 of 10 CALIBRATION OF Sound Level Meter: Brüel & Kjær Type 3050-A-040 No: 3050-100751 Id: - Microphone: Brüel & Kjær Type 4189 No: 2621142 Preamplifier: Brüel & Kjær Type 2669 No: 2679073 Supplied
More informationCommissioning of the Beta Secondary Standard (BSS2)
Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 7500-1 Third edition 2004-08-15 Metallic materials Verification of static uniaxial testing machines Part 1: Tension/compression testing machines Verification and calibration
More informationA basic introduction to reference materials. POPs Strategy
A basic introduction to reference materials POPs Strategy 2009-2010+ A tutorial 16 September 2009 Angelique Botha R&D metrologist Contents Why do we need reference materials? comparability of results metrological
More informationQuality assurance for sensors at the Deutscher Wetterdienst (DWD)
Paper submitted to ICAWS 2017: Topic 3 Sustainability of the measurements: Calibration, intercomparisons, laboratory and field performance tests, quality assurance and control assessment for traceable
More informationGuidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018
Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018 Guidelines on the Calibration of Automatic Instruments for
More informationMetallic materials Brinell hardness test. Part 2: Verification and calibration of testing machines
INTERNATIONAL STANDARD ISO 6506-2 Third edition 2014-10-01 Metallic materials Brinell hardness test Part 2: Verification and calibration of testing machines Matériaux métalliques Essai de dureté Brinell
More informationOperator's Manual. optris BR 400. Calibration source
Operator's Manual optris BR 400 Calibration source Optris GmbH Ferdinand-Buisson-Str. 14 13127 Berlin Germany Tel.: +49 30 500 197-0 Fax: +49 30 500 197-10 E-mail: info@optris.de Internet: www.optris.de
More informationFigures. Quality Management in the Automotive Industry. Capability of Measurement Processes
Verband der Automobilindustrie Quality Management in the Automotive Industry 5 Capability of Measurement Processes Capability of Measuring Systems Capability of Measurement Processes Expanded Measurement
More informationTest Report EN (2008)
Test Report EN 62233 (2008) Product Name and address of the applicant Name and address of the manufacturer Model Rating Brand name Heating elements for building into floors Advanced Heating Technologies
More informationSAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)
SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February
More informationHigh pressure comparisons between seven European National Laboratories
J4/2007 High pressure comparisons between seven European National Laboratories Range 50 MPa to 500 MPa Report on EUROMET Project 881 Markku Rantanen, Sari Semenoja, Måns Ackerholm, Antoine Condereys, Zdenek
More informationMethod of Measuring Machinery Sound Within an Equipment Space
ANSI/AHRI Standard 575 (Formerly ARI Standard 575) 2008 Standard for Method of Measuring Machinery Sound Within an Equipment Space Price $15.00 (M) $30.00 (NM) Printed in U.S.A. 8Copyright 1994, by Air-Conditioning
More informationMeeting future needs for Metrological Traceability A physicist s view
Meeting future needs for Metrological Traceability Dr L R Pendrill IUPAP C2 SUNAMCO Secretary Presentation at Emerging Issues in Metrological Traceability IAEA HQ, Vienna, 17 February 2004 SP, Measurement
More informationTurbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas
Turbine Meter TRZ 0 PRODUCT INFORMATION Reliable Measurement of Gas TURBINE METER TRZ 0 Method of operation, Construction Method of operation The TRZ 0 turbine meter is a flow meter suitable for gas measurement
More informationA61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012
Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability
More informationCOMMON MISTAKES IN DIMENSIONAL CALIBRATION METHODS
COMMON MISTAKES IN DIMENSIONAL CALIBRATION METHODS AN EDUCATIONAL PRESENTATION FROM THE LEADING MANUFACTURER OF METROLOGY INSTRUMENTS EDUCATION Bulletin EDU-14002A-H No. 2183 Mitutoyo Institute of Metrology
More informationMetallic materials Knoop hardness test. Part 1: Test method. Matériaux métalliques Essai de dureté Knoop Partie 1: Méthode d'essai
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 4545-1 Second edition 2017-12 Metallic materials Knoop hardness test Part 1: Test method Matériaux métalliques Essai de dureté Knoop Partie 1:
More informationComparison protocol EURAMET project
Comparison protocol EURAMET project Bilateral three-phase AC and energy comparison at frequency 1. Introduction This bilateral comparison was triggered by MIKES need to get support for their planned extension
More informationINTERNATIONAL JOURNAL OF ADVANCED RESEARCH IN ENGINEERING AND TECHNOLOGY (IJARET)
INTERNATIONAL JOURNAL OF ADVANCED RESEARCH IN ENGINEERING AND TECHNOLOGY (IJARET) International Journal of Advanced Research in Engineering and Technology (IJARET), ISSN ISSN 0976-6480 (Print) ISSN 0976-6499
More informationSpa & Wellness for Pipettes
epservices for premium performance Spa & Wellness for Pipettes Pipette Performance Plans Treat your Pipettes Well with Premium Services from the Experts! Total approach philosophy For over 60 years Eppendorf
More informationFinal report on CCQM-K36.1 with erratum
Final report on CCQM-K36.1 with erratum 5 October 2010 Ref 1107-03 HDJ Report no. DFM-2008-R07 Summary A follow-up comparison to CCQM-K36 has been carried out November 2007 to February 2008, designated
More informationAccuracy Calibration Certificate
gl06-0-067-acc Mettler Toledo Service Business Unit Industrial 900 Polaris Parkway Columbus, OH 0-800-- Services Provided By: Customer Accuracy Calibration Certificate Company: Sample Company Test Address:
More information