Plastic Scintillation and dryness evaporation: a new procedure for Alpha/Beta determination in water samples

Similar documents
APPLICATION OF A FREE PARAMETER MODEL TO PLASTIC SCINTILLATION SAMPLES

Synthesis of Plastic Scintillation microspheres: composition evaluation and alpha/beta discrimination capabilities

PLASTIC SCINTILLATORS: A POWERFUL TOOL TO REDUCE MIXED WASTE.

Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer

Basics of a/b Discrimination for Liquid Scintillation Counting

Application Note. Abstract. Introduction. Experimental

Gross alpha and beta measurements by LSC in waters: advantages, problems and limitations. Maurizio Forte Arpa Lombardia

Radiochemical Sensor for Continuous and Remote Liquid Effluents Monitoring

GEMMA RAURET, J S MESTRES and J F GARCIA

Journal of Environmental Radioactivity

Northwest University, Mafikeng, Centre of Applied Radiation Science & Technology (CARST), South Africa.

LSC for Quality Control of 99m TC Eluate from 99 Mo- 99m Tc Generator

A New Measurement Method For Tritium Rapid Detection by LSC. IM.CAEP Xu Qinghua

COMPARATIVE STUDY BETWEEN EXTERNAL STANDARD METHOD AND INTERNAL STANDARD METHOD FOR LOW-LEVEL TRITIUM MEASUREMENTS

Gross Alpha-Gross Beta Analysis in Water by Liquid Scintillation Counting (LSC)

Determination of 210 Pb and 210 Po in Water Samples

LIQUID SCINTILLATION COUNTERS. {Beta Counters}

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106.

NICKEL-63/59 IN WATER

Primary Standardisation of I-125

RESULTS YOU CAN COUNT ON

Liquid Scintillation Counting Performance Using Glass Vials in the Wallac 1220 QuantulusTM

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER

Hyuncheol Kim*, Yoonhee Jung, Wanno Lee, Guen-Sik Choi, Kun Ho Chung, Mun Ja Kang

Acronyms, Abbreviations, and Symbols Foreword to the First Edition Foreword to the Second Edition Preface to the First Edition Preface to the Second

TECHNETIUM-99 IN WATER

Depth Distribution of H-3, C-14 and Co-60 in Decommissioning of the Biological Shielding Concrete of KRR-2

Simultaneous measurement of 89 Sr and 90 Sr in aqueous samples by liquid scintillation counting using the spectrum unfolding method

Practical Approaches using TDCR Measurements and Alpha/Beta Separation

ISO Water quality Strontium 90 and strontium 89 Test methods using liquid scintillation counting or proportional counting

Hidex 300 SL Automatic TDCR Liquid Scintillation Counter

Liquid Scintillation Counter

Working Copy. tlrq /rb I 'B-16. Determination of Gross Alpha/Beta Activity. Revision 0. Laboratory Manag erllqao/rso. Effective Date. lnitials.

TECHNETIUM-99 IN WATER

URANIUM AND THORIUM-ISOTOPES RADIOCHEMICAL SEPARATION AND QUANTIFICATION OF NORM SAMPLES BY ALPHA-SPECTROMETRY

Implementation of ISO/IEC in a low level liquid s cintillation tritium laboratory

1/5 Specifications Hidex 300 SLL automatic TDCR liquid scintillation counter

Liquid Scintillation at NPL: Practical Applications

LSC-BASED APPROACH FOR RADON IN SOIL GAS MEASUREMENTS. The Marzeev Institute of Hygiene and Medical Ecology 50, Popudrenko str., Kiev 02094, Ukraine.

COMBINED LSC-BASED METHOD FOR RADON IN AIR MEASUREMENTS

SLAC Radioanalysis Laboratory

TECHNETIUM-99 IN SOIL

SEQUENTIAL DETERMINATION OF AMERICIUM, PLUTONIUM AND URANIUM IN LIQUID EFFLUENTS FROM NUCLEAR POWER PLANTS

Large area scintillation detector for dosimetric stand with improved light collection

Background measurements in the underground labs: Gran Sasso, Modane and Boulby

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France

EVALUATION OF HIGH-PURITY SYNTHETIC SILICA VIALS IN ACTIVE AND PASSIVE VIAL HOLDERS FOR LIQUID SCINTILLATION COUNTING OF BENZENE

Čerenkov counting and liquid scintillation counting of 36 Cl

BALANCED WINDOW METHOD IN 14 C LIQUID SCINTILLATION COUNTING. P TheodÛrsson 1 S Ingvarsdottir G I Gudjonsson

IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES

Quality Assurance. Purity control. Polycrystalline Ingots

TDCR in a nutshell. P. Cassette, Laboratoire National Henri Becquerel, France

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Outline. Why Are We Concerned? More Concerns

Revised material in Section 4, Analytical Chemistry

The measurement of tritium in environmental water samples

Field Method for Gross Alpha- And Beta- Emitting Radionuclide Detection in Environmental Aqueous Solutions

LS6500 Scintillation Counting Systems

Measurement of tritium with plastic scintillators in large vials of a low background LSC -an organic waste-less method-

Understanding the contribution of naturally occurring radionuclides to the measured radioactivity in AWE Environmental Samples

NEW DEVELOPMENTS OF AUTORADIOGRAPHY TECHNIQUE TO IMPROVE ALPHA AND BETA MEASUREMENTS FOR DECOMMISSIONING FACILITIES

Júlio Takehiro Marumo. Nuclear and Energy Research Institute, IPEN CNEN/SP, Brazil

Simultaneous determination of radium and uranium activities in natural water samples using liquid scintillation counting

Time-Resolved Liquid Scintillation Counting

TIME-RESOLVED LIQUID SCINTILLATION COUNTING

Fundamentals of Radionuclide Metrology

DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion

Hidex LSCs. LSC 2013 Barcelona

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

Unit 2. Instrumentation. Experts Teaching from Practical Experience

RADIOLOGICAL CHARACTERIZATION Laboratory Procedures

2100TR Liquid Scintillation Counter

COMBINED LSC BASED METHOD FOR RADON IN AIR MEASUREMENTS

GUIDE TO LABORATORY SURVEYS. Introduction

A Brief Overview of Radiation and Analytical Water Testing for Radiological Contaminants.

Part 1: Water quality Radium-226. Test method using liquid scintillation counting

Application Note. Understanding Performance Specifications for Low Background Alpha Beta Counters. FOM What Is It and Is It Useful?

Validation of detection efficiency of alpha particles in commercial liquid scintillation counters

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method

Radiometric Measurements of Environmental Radioactivity

Counting Solutions LSC Technical Tips from Packard

Development of Continuous Inflow Tritium in Water Measurement Technology by using Electrolysis and Plastic Scintillator

Determination of H-3 and C-14 in the frame of decommissioning projects at the Paul Scherrer Institute

The LSC Approach to Radon Counting in Air and Water

BUREAU INTERNATIONAL DES POIDS ET MESURES

*Corresponding author,

Clearance Monitoring. Chris Goddard.

1.1. This is a method for the separation and measurement of 228 Ra in water via its beta emitting 228 Ac daughter.

Measurement of the carbon 14 activity at natural level in air samples

METHOD TEST METHODS FOR MEASURING RADIONUCLIDE EMISSION FROM STATIONARY SOURCES

COMPARISON OF TWO DIFFERENT METHODS FOR DETERMINATION OF BIOGENIC FRACTION IN LIQUID FUELS

Calibration & Use of a Capintec CAPTUS 3000 Portable Thyroid Uptake System for Iodine-125 Bioassay Measurements Todd W.

Radio-HPLC Flow Cells

TRITIUM DETERMINATION IN WATER

Development of a rapid solvent extraction apparatus for aqueous chemistry of the heaviest elements

A Rapid Procedure for Screening Transuranium Nuclides in Urine Using Actinide Resin and Low

Maintaining analytical quality The challenges of decommissioning analysis

Counting Solutions LSC Technical Tips from PerkinElmer

CIEMAT/NIST STANDARDIZATION METHOD EXTENDED TO ANODE OUTPUTS FOR BETA AND ELECTRON-CAPTURE NUCLIDES J. F. ORTIZ

Transcription:

a new procedure for Alpha/Beta determination in water samples Bouchra Aboudou, Hector Bagán, Alex Tarancón Sanz, Gemma Rauret, Jose Francisco García Departmento de Química Analitica de la Universidad de Barcelona, Spain LSC 2010 Advances in Liquid Scintillation Spectrometry Paris, France, 6-10 September 2010

Introduction - Gross Alpha/Beta are an important parameters to monitorize the radioactivity in water Depending on the level, specific analysis are required. - Procedures to determine Gross Alpha/Beta activity.. Proportional counters: sample evaporated to dryness and deposited on a planchet. - Well established - Low detection efficiency. Liquid Scintillation Counting: sample mixed with scintillation cocktail (PSA). - High detection efficiency - Amount of sample? Mixed waste production.

Introduction - Plastic Scintillation is an alternative method to liquid scintillation.. In general, sample solution is added to a vial filled up with plastic microspheres (PSm) - High detection efficiency (except for low beta emitters) No limited amount of sample No mixed waste production. - New procedure for Gross Alpha/Beta determination by Plastic Scintillation:. Water sample evaporated to dryness on the PSm inside the counting vial. Measurement in a Scintillation counter (PSA).

Objective To evaluate the capability of the new procedure based on Plastic Scintillation and dryness evaporation to the determination of gross Alpha/Beta parameters in aqueous samples. (Partially achieved)

Experimental Apparatus and reagents. - Detector Quantulus liquid scintillation (EG&G Wallac) - Poliethylene vials (20 ml) from Packard Instruments Co. - Plastic Scintillation microspheres (Detec-Rad) diameters (120-180, 400-500 and 630-800 µm) - Active solutions: 90 Sr/ 90 Y, 14 C and 241 Am Sample preparation procedure: 20 ml scintillation vial 2 g of PSm 1-20 ml of solution (1 ml) Evaporate to dryness at 40ºC Measurement 40ºC

Optimum conditions determination. - Liquid Scintillation vs Plastic Scintillation vs Plastic Scintillation+ dryness evaporation - Infuence of the PSm diameter. - Influence of the beta emitter energy. - Influence of sample volume

Liquid Scintillation vs Plastic Scintillation vs Plastic Scintillation+ dryness evaporation. Conditions: PSm (630-800 µm), 90 Sr/ 90 Y 241 Am 241 Am LS -> PS: higher distance lower energy shorter pulse PSA shifts to lower values PS -> PS+dryness : intermediate.

Liquid Scintillation vs Plastic Scintillation vs Plastic Scintillation+ dryness evaporation. Conditions: PSm (630-800 µm), 90 Sr/ 90 Y 241 Am 241 Am 1.8 1.6 1.4 1.2 CL CP CP evap 1 Eff (%) 0.8 0.6-0.2 0 200 400 600 800 1000 1200 channels Detection Efficiency: PS 27.6 PS+dryness evaporation 99.3 % 0.4 0.2 0 LS -> PS: higher distance lower energy shorter pulse PSA shifts to lower values PS -> PS+dryness : intermediate.

Liquid Scintillation vs Plastic Scintillation vs Plastic Scintillation+ dryness evaporation. Conditions: PSm (630-800 µm), 90 Sr/ 90 Y 241 Am 90 Sr/ 90 Y LS -> PS: higher distance lower energy shorter pulse PSA shifts to lower values PS -> PS+dryness : intermediate.

Liquid Scintillation vs Plastic Scintillation vs Plastic Scintillation+ dryness evaporation. Conditions: PSm (630-800 µm), 90 Sr/ 90 Y 241 Am 90 Sr/ 90 Y 0.7 0.6 CL CP CP evap 0.5 Eff (%) 0.4 0.3 0.2 0.1 0 0 200 400 600 800 1000 1200 channels Detection Efficiency: PS 163.2 PS+dryness evaporation 190.5 % LS -> PS: higher distance lower energy shorter pulse PSA shifts to lower values PS -> PS+dryness : intermediate.

Liquid Scintillation vs Plastic Scintillation vs Plastic Scintillation+ dryness evaporation. Conditions: PSm (630-800 µm), 90 Sr/ 90 Y 241 Am LS -> PS: higher distance lower energy shorter pulse PSA shifts to lower values PS -> PS+dryness : intermediate.

PS+ dryness evaporation. Influence of the PSm diameter Conditions: PSm (120-180, 400-500, 630-800 µm), 90 Sr/ 90 Y 241 Am 241 Am PSm diameter increase: total surface decreases available surface increases? thinner recovery layer - higher energy longer pulses - PSA shifts to higher values

PS+ dryness evaporation. Influence of the PSm diameter Conditions: PSm (120-180, 400-500, 630-800 µm), 90 Sr/ 90 Y 241 Am 241 Am 1.2 1 120-180 400-500 630-800 With solution Evaporated Eff (%) 0.8 0.6 0.4 Diameter Eff±STD Eff±STD 120-180 D1: 59,6 ± 0,9 D1: 101,5 ± 1,3 400-500 D3: 41,1 ± 0,7 D3: 101,7 ± 1,3 0.2 630-800 D4: 27,6 ± 0,8 D4: 99,3 ± 1,3 0-0.2 0 200 400 600 800 1000 1200 channels PSm diameter increase: total surface decreases available surface increases? thinner recovery layer - higher energy longer pulses - PSA shifts to higher values

PS+ dryness evaporation. Influence of the PSm diameter Conditions: PSm (120-180, 400-500, 630-800 µm), 90 Sr/ 90 Y 241 Am 90 Sr/ 90 Y PSm diameter increase: total surface decreases available surface increases? thinner recovery layer - higher energy longer pulses - PSA shifts to higher values

PS+ dryness evaporation. Influence of the PSm diameter Conditions: PSm (120-180, 400-500, 630-800 µm), 90 Sr/ 90 Y 241 Am 90 Sr/ 90 Y Eff (%) 0.7 0.6 0.5 0.4 0.3 120-180 400-500 630-800 With solution Evaporated Diameter Eff±STD Eff±STD 120-180 D1: 178,2 ± 3,4 D1: 191 ± 6,8 400-500 D3: 173,5 ± 2,6 D3: 192,7 ± 2,5 0.2 630-800 D4: 163,2 ± 2,3 D4: 190,5 ± 2,9 0.1 0 0 200 400 600 800 1000 1200 channels PSm diameter increase: total surface decreases available surface increases? thinner recovery layer - higher energy longer pulses - PSA shifts to higher values

PS+ dryness evaporation. Influence of the PSm diameter Conditions: PSm (120-180, 400-500, 630-800 µm), 90 Sr/ 90 Y 241 Am Proposed diameter: higher (630-800 µm)

PS+ dryness evaporation. Influence of beta emitter energy Conditions: PSm (120-180, 400-500, 630-800 µm), 14 C (156 kev) - 90 Sr/ 90 Y ( 2.27 MeV) 14 C 90 Sr/ 90 Y PSm diameter increase: total surface decreases available surface increases? thinner recovery layer - higher energy longer pulses - PSA shifts to higher values

PS+ dryness evaporation. Influence of beta emitter energy Conditions: PSm (120-180, 400-500, 630-800 µm), 14 C (156 kev) - 90 Sr/ 90 Y (2.27 MeV) 14 0.35 C 90 Sr/ 90 Y 0.3 0.25 0.2 120-180 400-500 630-800 Eff (%) 0.7 0.6 0.5 0.4 0.3 0.2 0.1 120-180 400-500 630-800 Eff (%) 0.15 0.1 0.05 0-0.05 0 200 400 600 800 1000 1200 channels With solution 0 0 200 400 600 800 1000 1200 channels Evaporated Diameter Eff±STD Eff±STD 120-180 D1: 39,7 ± 0,8 D1: 80,3 ± 1,1 400-500 D3: 25,4 ± 0,6 D3: 78,5 ± 0,7 630-800 D4: 18,3 ± 0,6 D4: 78,3 ± 1,1 Low-Medium energy beta emitters can be detected with high efficiency

PS+ dryness evaporation. Influence of the sample volume Conditions: PSm (630-800 µm) Solution volumes: 1, 5, 20 ml. Calibration procedures: Comprimise PSA: 90 Double PSA: 120 - no 90 Sr/ 90 Y at α spectrum (quantification 241 Am) 60 - no 241 Am at β spectrum region (quantification 90 Sr/ 90 Y) Compromise PSA 90Sr/ 90 Y 241Am Double PSA 90Sr/ 90 Y 241Am Volume Eff β Eff α 1 ml 161.98±2.32 84.27±1.84 5 ml 163.21±2.74 76.38±0.73 20 ml Volume Eff β Eff α 1 ml 110.92±2.20 47.35±2.42 5 ml 117.43±4.23 41.38±1.00 20 ml Detection efficiency is not constant with sample volume change (material deposited) -> calibration in progress.

Results and discussion sample quantification (exploratory results) PS+ dryness evaporation. Conditions: PSm (630-800 µm) Calibration: Constant Efficiency Volumes: 5 ml. Drinking water (Fontvella) (3 Replicates). Act Sr Act Am Prop compromise Double PSA Err Sr Err Am Err Sr Err Am 17.91 38.23 1:1-28.92±4.30-5.20±3.28-3.53±1.61-2.17±3.02 36.54 38.48 2:1-15.45±2.01-9.59±3.87 4.62±4.12-1.02±3.60 18.14 71.67 1:2-49.52±5.62-5.44±3.06-9.12±3.33-1.53±2.16 75.10 38.64 4:1-9.66±2.37-19.83±1.50 5.11±4.40-5.67±3.08 18.13 154.56 1:4-106.46±5.85-3.33±0.95-36.10±8.32-1.85±3.90 Exploratory Relative Errors in Gross Alpha/Beta Compromise PSA > Double PSA Exploratory Relative Errors - Double PSA < 12% (except proportion 1:4)

Conclusions - A new procedure for Alpha/Beta determination in water samples is defined based on the use of plastic scintillator microsphere and dryness evaporation. - Evaporated water samples on PSm produce:. pulses of intermediate energy between LS and PS. pulses of intermediate duration between LS and PS. spillover vs PSA distribution shifted to lower values - Changes in the PSm diameter produce slight changes in spillover vs PSA distribution. - Low energy beta emitters are detected with high efficiency. - Detection efficiency depends on the material deposited (calibration is required) - Exploratory results using using constant efficiency and double PSA allow Gross Alpha / Beta activity determination with relative errors < 12 %

Acknowlegments. Ministerio de Ciencia e Innovación (Spain) for Financial support (CTM2008-01147). Plastic Scintillation and dryness evaporation: a new procedure for Alpha/Beta determination in water samples Bouchra Aboudou, Hector Bagán, Alex Tarancón Sanz, Gemma Rauret, Jose Francisco García Departmento de Química Analitica de la Universidad de Barcelona, Spain