GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

Similar documents
MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Types of Analyzers: Quadrupole: mass filter -part1

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis

Tendenze nell innovazione della strumentazione in spettrometria di massa:

TANDEM MASS SPECTROSCOPY

Mass Spectrometry. What is Mass Spectrometry?

Chemistry Instrumental Analysis Lecture 35. Chem 4631

Lecture 8: Mass Spectrometry

Mass Spectrometry in MCAL

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

Chemistry 311: Topic 3 - Mass Spectrometry

Lecture 8: Mass Spectrometry

MASS SPECTROMETRY. Topics

Translational Biomarker Core

MS Goals and Applications. MS Goals and Applications

Mass spectrometry gas phase transfer and instrumentation

Introduction to the Q Trap LC/MS/MS System

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Mass spectrometry of proteins, peptides and other analytes: principles and principal methods. Matt Renfrow January 11, 2008

Analysis of Polar Metabolites using Mass Spectrometry

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile

Mass spectrometry.

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

Time-of-Flight Mass Analyzers

MS/MS .LQGVRI0606([SHULPHQWV

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS)

MS Goals and Applications. MS Goals and Applications

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap

This is the total charge on an ion divided by the elementary charge (e).

Harris: Quantitative Chemical Analysis, Eight Edition

Protein analysis using mass spectrometry

Mass Spectrometry. Fundamental LC-MS. Mass Analysers

Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass Spectrometer

Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY:

Molecular Mass Spectrometry

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

CHAPTER D3 TOF ION OPTICS

Quattro Micro - How does it work?

Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry

1. The range of frequencies that a measurement is sensitive to is called the frequency

Mass Analysers for LC MS

ICPMS Doherty Lecture 1

Ion traps. Quadrupole (3D) traps. Linear traps

What is a mass spectrum?

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Molecular Mass Spectrometry

AQA Chemistry A-Level : Atomic Structure

Introduction to LC-MS

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4

PC 235: Mass Spectrometry and Proteomics

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

BIOINF 4399B Computational Proteomics and Metabolomics

Tandem MS = MS / MS. ESI-MS give information on the mass of a molecule but none on the structure

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

TOMAHAQ Method Construction

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Mass spectrometry: forming ions, to identifying proteins and their modifications Stephen Barnes, PhD

Mass Spectrometry. Electron Ionization and Chemical Ionization

Mass spectrometry: forming ions, to identifying proteins and their modifications Stephen Barnes, PhD

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April

PARTICLE ACCELERATORS

Introduction to GC/MS

1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading:

VIII. QUADRUPOLE ION TRAPS: AN ION TRAPPING INST. A. BASIC DESIGN (REF: JMS, 1997, 32, 351)

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

Quadrupole Storage Mass Spectrometry

BIOINF 4120 Bioinformatics 2 - Structures and Systems - Oliver Kohlbacher Summer Systems Biology Exp. Methods

Collision Cross Section: Ideal elastic hard sphere collision:

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Ch 13. Basics of Mass Spectrometry (I) : Principles & Ionization Sources

Mass Spectrometry Course

Ch. 22 Mass Spectrometry (MS)

Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid. Chromatography-Tandem Mass Spectrometry

Overview of NETCHEM MSc & PhD courses: mass spectrometry in EFSC

1 Mass Analysis. Gene Hart-Smith and Stephen J. Blanksby

Thermo Finnigan LTQ. Specifications

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Beam Diagnostics Lecture 3. Measuring Complex Accelerator Parameters Uli Raich CERN AB-BI

Ionization Techniques Part IV

Introduction to Mass Spectrometry and Proteomics

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative

LC-MS Based Metabolomics

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Selecting and planning for the right mass spectrometer

Mass Spectrometry. General Principles

Choosing the metabolomics platform

CHROMATOGRAPHY AND MASS SPECTROMETER

Space Charge Effects in Linear. Quadrupole Ion Traps

Transcription:

DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015

High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary pumps Inlet Ion Source Mass Filter Detector Data System Sample Plate Target HPLC GC Solids probe MALDI ESI IonSpray FAB LSIMS EI/CI TOF Quadrupole Ion Trap FTMS Microch plate Electron Mult. Hybrid Detec. PC s UNIX Mac

Mass Analyzers Quadrupole Analyzer (Q) Ion Trap Mass Analyzer (QIT) Time-of-Flight Analyzer (TOF) Orbitraps(Orbi)

Electromagnetism Force or F = q 2 x q 1 4πεr 2 or q 2 x V 1 -V 2 d q 2 Lens at Voltage V 1 r q 2 d q 1 charge Lens at Voltage V 2

Electromagnetism In mass spectrometry we work with charges and lenses 0V +200V + Same situation as +1000V +1200V -200V 0V It s the difference that matters

Electromagnetism In mass spectrometry we work with charges and lenses 0V -200V + Same situation as +1000V +800V -200V -400V It s the difference that matters

Electromagnetism Work or energy W= q 2 x V 1 -V 2 d x d Lens at Voltage V 1 q 2 d Lens at Voltage V 2 q is often expressed in multiples of e (charge on an electron) i.e energy is expressed in ev (electron volts)

Electromagnetism V Work or energy W= 1 -V q 2 2 x x d d 0V -200V + Potential energy 300 ev If q is 1 charge then: Loses 200 ev of potential energy Gains 200 ev of kinetic energy 0 ev

Time-of-flight Mass Analyzer Analyser performs the equivalent of the 100m sprint for ions Start line Known Distance (L) Finish line Ions are given the same amount of energy through a pulse (Energy is proportional to charge and the applied potential. E=zeV, z is number of charges, e is the amount of charge on an electron, V is volts) Ions then move at the speed determined by their mass (E=0.5mv 2, v for velocity (L/t)) i.e. velocity goes down as mass goes up Distance to finish line is established and stopwatch (t) is accurate to 1 nanosecond or better So. zev 2 mv 2 which can be turned into m z t 2eV L 2

ToF: Problems Starting position is not always the same! A) Temporal: Ions of the same mass can form at different times but have the same KE. B) Spatial: Ions of the same mass form having the same KE but in different parts of the source; C) Kinetic Energy: Ions of the same mass can form with different KEs. D) Direction: There is also the problem that when an ion forms, it may do so with an initial velocity in the opposite direction.

Ion formation problem solution: Time Lag Focussing/Delayed extraction 0 V. + + + Ions of same mass, different velocities Step 1: No applied electric field. Ions spread out. 0 V. 20 kv. + + + 0 V. Step 2: Field applied. Slow ions accelerated more than fast ones. + 20 + kv. + Step 3: Slow ions catch up with faster ones. 0 V.

Improving resolution: The reflectron Reflectron is an electrostatic hill for the ions to try and overcome 21.30 kv In reality they fail and are turned around Ions that have a higher velocity will have more kinetic energy and will climb the hill further before being turned around.

Improvement of resolution. 100 90 80 4700 Linear Spec #1 MC=>SM3=>MC=>MC[BP = 1571.8, 2191] 2191.4 Continuous extraction 70 % Intensity 60 50 40 30 Resolution = 500 Delayed extraction 20 10 0 1560 1566 1572 1578 1584 1590 Mass (m/z) Reflector 100 90 80 4700 Linear Spec #1 MC=>SM3=>MC=>MC[BP = 1570.7, 1617] 1616.7 Linear mode % Intensity 70 60 50 40 Resolution = 3,500 + Delayed extraction % Intensity % Intensity 30 20 10 0 1560 1566 1572 1578 1584 1590 Mass (m/z) 4700 Reflector Spec #1 MC=>MC[BP = 1570.7, 2642] 100 2642.4 90 80 70 Resolution = 10,000 60 50 40 30 20 10 0 1560 1566 1572 1578 1584 1590 Mass (m/z) 100 90 80 70 60 50 40 30 20 10 4700 Reflector Spec #1 MC=>MC=>MC[BP = 1570.7, 1751] m/z = 1570.68 Resolution = 20,000 1751.0 Reflector Reflector mode Delayed extraction + Reflector Reflector mode + Delayed extraction + Reflector 0 1560 1566 1572 1578 1584 1590 Mass (m/z)

Quadrupole Mass Analyser: Description -[U+Vcos( t)] -(U+ 2r y y z x x +[U+Vcos( t)] +(U+Vcos t Consists of Four Metal Rods Opposite Rods have same voltage and adjacent rods have opposite voltage The voltage is a combination of a constant potential (U) and an alternating potential (V)

Quadrupole Mass Analyser: Description Voltage cos( t) t 3 0V t 2 t 4 U V t 1 t t 1 2 t t 3 4 Taken from Roman Korner: quadrupole mass analyzers. BM58/BMP58 Course 1998.

Quadrupole Mass Analyser: Description Ions are given a small amount of kinetic energy (1-2 ev) to allow them to traverse the quadrupole to the detector (assuming they have stable trajectories)

Quadrupole Mass Analyser: Scanning voltage Complete mass scan time

Quadrupole Mass Analyser: Scanning products of fib a m+2h collsion gas =6.8e-4, ce=40 SM100602 1 (3.073) 100 767.92 767.73 768.04 Daughters of 769ES+ 4.72e7 644.28 766.62 % 0 185.80 273.38 444.37 388.01 459.89 643.85 643.35 642.92 766.49 756.87 645.59 904.59 903.60 902.73 777.73 1076.57 1074.90 1077.44 1074.46 905.58 1078.62 906.33 1262.09 1349.29 200 400 600 800 1000 1200 1400 m/z

Quadrupole Mass Analyser Used to be one of the cheapest mass spectrometers on the market Very good for precursor ion scanning and selective reaction monitoring

-[U+Vcos( t)] -(U+ Vcos t 2r y +[U+Vcos( t)] y z x x +(U+Vcos t

Quadrupole Ion Traps: Description

Quadrupole Ion Traps: Description Taken from: Web Site of Chemistry Department Purdue University,Indiana, USA RF voltage is applied to ring electrode to produce a three-dimensional quadrupole electric field for trapping ions.

Quadrupole Ion Traps Taken from: Web Site of Chemistry Department. Purdue University, Indiana, USA

2D Ion Trap Mass Analyzer LTQ

A Little Bit About Orthogonal ToFs DETECTOR Important Factors in on-axis MALDI-ToF: TOF Resolution and mass accuracy depend on knowing exact distance, time and energies given to the ions SOURCE Ions form in the same direction as ToF analyser and so the source conditions are important factors in mass accuracy and resolution e.g. Starting position of ions Starting kinetic energy and direction of ions

A Little Bit About Orthogonal ToFs Collisional Cooling of ions (usually in a quadrupole) before they enter the ToF region helps further improves ToF sensitivity and Resolution. Cooling allows better positioning of ion packet into pusher region of the ToF

A Little Bit About Orthogonal ToFs DETECTOR TOF Reducing the ions velocity here Allows better focusing of ions here SOURCE Collisional cooling of ions increases mass accuracy and resolution

The Orbitrap Positive ion moving very fast + Positive ion attracted and has its path bent - + Attraction of rod is high, Velocity of ion is high.. Ion makes an orbit around rod Very high negative potential Similar to a satellite around the earth

LTQ Orbitrap Hybrid Mass Spectrometer Finnigan LTQ Linear Ion Trap API Ion source Linear Ion Trap C-Trap Differential pumping Orbitrap Differential pumping

LTQ Orbitrap Operation Principle 1. Ions are stored in the Linear Trap 2.. are axially ejected 3.. and trapped in the C-trap 4.. they are squeezed into a small cloud and injected into the Orbitrap 5.. where they are electrostatically trapped, while rotating around the central electrode and performing axial oscillation The oscillating ions induce an image current into the two outer halves of the orbitrap, which can be detected using a differential amplifier Ions of only one mass generate a sine wave signal

Detection:Fourier Transform Samples are typically not just 1 ion but several and on top there will be ions with several different m/z values. Typical image current for a simplish mixture looks like this: Time How do we deduce the individual frequencies?

Detection:Fourier Transform Luckily waves do not affect each other and so within the messy image current, the waves are present intact. Fourier transform is a mathmatical technique which can deduce the frquencies present.

Resolution The longer you measure, the better the spectrum.more datapoints for final deduction

Frequencies and Masses The axial oscillation frequency follows the formula Where w = oscillation frequency k = instrumental constant m/z =. well, we have seen this before k m / z Many ions in the Orbitrap generate a complex signal whose frequencies are determined using a Fourier Transformation

Collision induced dissociation Mixture MS 1 Collision cell MS 2 Accelerate ions into collision cell Potential energy -10 ev -50 ev Delta Kinetic energy 40 ev 0 ev

0 50 75 100 125 150 175 200 225 250 275 300 325 350 375 400 425 450 475 500 525 550 575 600 m/z Collision induced dissociation Mixture MS 1 Collision cell MS 2 sm207803 17 (0.950) Cm (11:20) 100 TOF MS ES+ 7.02e3 sm117502 18 (0.781) Cm (6:19) 100 TOF MSMS 556.30ES+ 1.64e3 sm117501 1 (0.043) Cm (1:23) 100 TOF MSMS 556.30ES+ 708 % % % 0 m/z 500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 0 50 75 100 125 150 175 200 225 250 275 300 325 350 375 400 425 450 475 500 525 550 575 600 m/z A mass spectrum of a mixture Isolation of an ion In our case a protonated peptide A mass spectrum of the fragments Produced by the ion

Q Exactive Beam CID fragmentation mass selection Remove neutrals D30 high resolution MS and MS/MS

Q- ToFs

Orbitrap Fusion CID Ion trap CID fragmentation mass selection Beam CID fragmentation Redesigned instrument capable of: 15 MSMS HCD (orbitrap) per second 20 MSMS CID (ion trap) per second

Mass Analyzers Quadrupole Analyzer (Q) Low (1 amu) resolution, fast, (relatively) cheap Ion Trap Mass Analyzer (QIT) Fair resolution, all-in-one mass analyzer Time-of-Flight Analyzer (TOF) Good resolution, exact mass, fast, no upper m/z limit, costly Orbitraps(Orbi) High resolution, exact mass, costly