application note LED measurement radiant power measurement

Size: px
Start display at page:

Download "application note LED measurement radiant power measurement"

Transcription

1 application note LED measurement radiant power measurement

2 Contents 1 Introduction Setup Introduction Measurement setup Integrating spheres LED measurement procedure CIE 127 & Admesy calibration procedure Overview Spectral and spatial distribution Remarks for measurement procedure References

3 1 Introduction This document is intended to clarify the procedure for accurate measurements of LED and SSL (Solid State Lighting) products by means of an integrating sphere. In perspective of CIE 127: 2007 recommendations, the entire procedure including calibration, absorption correction and led measurements is listed. Special attention is given to the enhancements of the calibration procedure with an Admesy tungsten broadband reference standard compared to the conventional calibration procedure. At last a number of general remarks concerning LED measurements are provided to ensure accurate and repeatable LED measurements. These measurements can be carried out using Admesy s Iliad software application. Latest Iliad Version can be downloaded from the Admesy website for free. 3

4 2 Setup 2.1 Introduction A typical LED measurement procedure with integrating sphere consists of three steps: calibrating the test setup, absorption correction for device under test (DUT) and actual measurements. Special attention is given to enhancements of the sphere setup calibration procedure with Admesy products compared to conventional calibration procedures as defined by CIE 127. Details are discussed in chapter CIE 127 & Admesy calibration procedure. 2.2 Measurement setup In order to conduct fast and accurate LED and SSL measurements, an integrating sphere setup as shown in figure 1 is recommended (CIE 127: 2007)]. An integrating sphere (1) connected to a spectrometer (2) by a fiber-optic flex cable (3) to measure a LED (4). For absorption correction an additional correction connecting an auxiliary lamp (5) may be necessary. Various reference standard lamps (not shown in the scheme below) are needed for spectral and luminous flux calibration as defined by CIE 127. This Admesy test setup is developed for luminous flux (lm) measurements and provides information regarding X, Y, Z parameters as well as x,y, u,v, CCT, dominant and peak wavelength, CRI and spectral distribution Fig 1 Integrating sphere test setup for LED measurements. 4

5 2.3 Integrating spheres An integrating sphere is used to spatially distribute the unidirectional radiation emitted by the LED into homogeneous radiation. Depending on properties of the LED, an appropriate sphere size has to be chosen. Large spheres typically have better light distribution compared to small spheres due to the relative small geometrical errors objects in the sphere cause. While small spheres have the advantage of a relatively high throughput, resulting in an increased signal-noise ratio. Table 1 provides an overview of Admesy spheres and typical applications. Model Diameter Auxiliary lamp Application Type of DUT 1 75mm No Production Standard single LEDs 2 150mm Yes Production Laboratory Standard single LEDs High power LEDs 3 250mm Yes Laboratory, CIE 127 compliant for 2π measurements Standard single LEDs High power LEDs Small LED modules Table 1 Admesy integrating spheres and applications. 5

6 3 LED measurement procedure The scheme below shows three steps of a typical LED measurement procedure, starting with sphere calibration procedure according to CIE 127 (left column) in two steps. The Admesy procedure with only one step is shown in the right column. The proposed adjustments in calibration procedures are described in detail in chapter 4. Followed by DUT and fixture absorption by means of an auxiliary lamp according to CIE 127 procedure. At last the actual LED measurement procedure. Procedure overview CIE 127: 2007 procedure Admesy procedure CIE 127 calibration: step 1 Admesy calibration Spectral wavelength calibration with tungsten broadband reference standard. Spectral wavelength calibration with Admesy tungsten broadband reference standard. Compensated for enhanced blue region spectral response (blue filter) and spatial distribution (cosine corrector). CIE 127 calibration: step 2 Absolute luminous flux calibration with LED reference standard (compensate for spectral and spatial distribution of tungsten). Absorption correction: step 1 Applicable to sphere with auxiliary port only (150 and 250 mm). Measure spectral radiant distribution without objects attached to sphere. Absorption correction: step 2 Applicable to sphere with auxiliary port only (150 and 250 mm). Measure spectral radiant distribution with test LED (turned off) attached to sphere. Start LED measurement Measurement of LEDs Fig 2 Overview of complete LED measurement procedure including calibration procedure. Fig 2 Overview of complete LED measurement procedure including calibration procedure. 6

7 4 CIE 127 & Admesy calibration procedure 4.1 Overview CIE 127 LED sphere setup calibration procedure describes two steps: Spectral wavelength calibration for spectral response of sphere. Absolute luminous flux calibration to compensate for poor blue spectral response of tungsten reference standard and typical spatial distribution of LED. Admesy enhanced calibration procedure in one step: Spectral calibration with blue enhanced broadband reference standard (poor blue region compensation) and cosine corrector (spatial LED distribution). 4.2 Spectral and spatial distribution CIE 127 requires the second step with a LED reference standard to calibrate the setup for LEDs based on their typical spectral and spatial distribution. Due to the wide range of variations in LEDs, CIE 127 states that a large amount of LED reference standards should be maintained. For each type of LED to be measured, a corresponding LED reference standard is required for calibration. Concerning the spectral deviations, the spectral distribution of the tungsten reference standard (fig 3) shows relatively low spectral power in the blue region compared to the longer wavelengths. In the shorter wavelength region, the signal noise ratio therefore is non optimal. Measuring white and blue LEDs which have typically high luminous flux in this region (fig 4) with a poor signal noise ratio may result in inaccurate measurement results. Second reason for calibrating with a LED reference standard is because of typical spatial distribution of LEDs. Fig 3 Spectral distribution of broadband tungsten reference standard used for CIE 127 calibration step 1. Fig 4 Spectral distribution of a (white) LED reference standard used for CIE 127 calibration step 2. Fig 5 Spectral distribution of Admesy enhanced broadband tungsten reference standard. To overcome all inconveniences in both calibration procedure as well as time and cost efficient LED measurements, Admesy has developed an enhanced halogen reference standard with blue filter and cosine corrector. This reference standard allows integrating both calibration steps of the CIE 127 recommendation. Because of the integrated blue light balancing filter, a more balanced spectral distribution is created (fig 5) causing a better signal noise ratio in the short wavelength region. Thus, allowing accurate measurements of white and blue LEDs. In addition, a cosine corrector ensures optimal Lambertian distribution to match the spatial distribution of a wide range of LEDs. Instead of maintaining multiple reference standards, one Admesy reference standard is sufficient for calibrating integrating sphere setups. 7

8 5 Remarks for measurement procedure High speed measurements (i.e. production settings) are technically speaking not as sophisticated compared to laboratory measurements. To ensure accuracy and repeatability of production measurements as recommended by CIE 127: 2007, a number of remarks are listed in this chapter with a brief explanation. Control sphere for dust and dirt. Control calibration (date) of spectrometer. Control calibration (date) of reference standard: due to maximum lifetime. Control stabilization time of reference standard before calibration. Control power supplies (current, voltage, AC/ DC, waveform). Control measurement parameters: i.e. measurement integration time, resolution. Control ambient temperature and humidity of test environment. CIE 127: 2007 states an ambient temperature of 25ºC (77F) unless otherwise specified. Control ambient lighting (no ambient light should enter the sphere). Control LED position during measurement (see examples below) Fig 6 Undesired situation (left) light beam is affected by obstructions. Desired situation (right) light beam is not affected by any obstructions. Fig 7 Generally undesired situation (left) where light from LED leaks outside the sphere. Desired situation (right) where all light emitted from LED enters the sphere. Fig 8 Components of LED and/or fixture may absorb light 8

9 6 References CIE, Publication No Measurements of LEDs, Central Bureau of the CIE, Vienna, Austria

10 Admesy B.V. Sleestraat CA Ittervoort The Netherlands T +31 (0) F +31 (0) info@admesy.com The material in this document is subject to change. No rights can be derived from the content of this document. All rights reserved. No part of this document may be reproduced, stored in a database or retrieval system, or published in any form or way, electronically, mechanically, by print, photo print, microfilm or any other means without prior written permission from the publisher. Version /

Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products

Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products White Paper Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products 1. Introduction The LED industry is growing rapidly and this naturally brings up an important need for reliable

More information

Measurement method for the proficiency testing program

Measurement method for the proficiency testing program APLAC T088 Appendix Measurement method for the proficiency testing program Introductions This measurement method is prepared for use by the APLAC Proficiency Testing Program Photometric measurement of

More information

Photometric Test Report

Photometric Test Report Photometric Test Report Relevant Standards RIES LM-79-2008 RANSI C82.77:2014 Prepared For RAB LIGHTING INC 170 Ludlow Ave, PO BOX 970, Northvale, NJ 07647-2305 USA Crystal Wu 13482123342, Crystal.Wu@rabweb.com

More information

TeraLED. Thermal and Radiometric Characterization of LEDs. Technical Specification.

TeraLED. Thermal and Radiometric Characterization of LEDs. Technical Specification. TeraLED Thermal and Radiometric Characterization of LEDs Technical Specification M E C H A N I C A L A N A L Y S I S TeraLED HARDWARE OPTIONS Depending on the dimensions and power level of the LED or LED

More information

Practical Optical Measurements of OLED Panels for Lighting Applications

Practical Optical Measurements of OLED Panels for Lighting Applications Practical Optical Measurements of OLED Panels for Lighting Applications TOKI KAWABATA KONICA MINOLTA, INC. JAPAN YOSHI OHNO National Institute of Standards and Technology U.S.A. * This work was conducted

More information

NVLAP LAB CODE LM Test Report. For REMPHOS TECHNOLOGIES LLC. (Brand Name: ) 90 Holten St, Danvers, MA 01923,USA.

NVLAP LAB CODE LM Test Report. For REMPHOS TECHNOLOGIES LLC. (Brand Name: ) 90 Holten St, Danvers, MA 01923,USA. LM-79-08 Test Report For REMPHOS TECHNOLOGIES LLC (Brand Name: ) 90 Holten St, Danvers, MA 01923,USA LED Lamp Model name(s): RPT-LEDBIAX-1500LM-4000K Representative (Tested) Model: RPT-LEDBIAX-1500LM-4000K

More information

Photometric Test Report

Photometric Test Report Photometric Test Report Relevant Standards IES LM-79-2008 ANSI C82.77-2002 UL1598-2008/ UL1993-2012 Prepared For Keystone Technologies 1390 Welsh Road,North Wales, PA 19454 Stephanie DeLaurentis, 267-656-2272,

More information

OLLA White Paper on the Necessity of Luminous Efficacy Measurement Standardisation of OLED Light Sources

OLLA White Paper on the Necessity of Luminous Efficacy Measurement Standardisation of OLED Light Sources IST- 004607 OLLA on the Necessity of Luminous Efficacy Measurement Standardisation of OLED Light Sources OLLA Editors OLLA WP5 members Version 3.0 Date of issue 25 September 2007 Dissemination Level Public

More information

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS ABSTRACT LEDs are used in many applications, ranging from signalling to ambient lighting to displays. Specifications and tolerances are used

More information

Impact of New SSL Standards on Goniospectroradiometric Measurements

Impact of New SSL Standards on Goniospectroradiometric Measurements Impact of New SSL Standards on Goniospectroradiometric Measurements By Dr. Ðenan Konjhodžić, Application Engineer Presented by Justin Blanke, International Sales Manager Outline Goniophotometry and goniophotometer

More information

Main Applications MR 16 Track lighting Downlight

Main Applications MR 16 Track lighting Downlight ProLight PDSB-7FxL-BC0P 7W COB Light-Engine LEDs Technical Datasheet Version: 1.4 Features High flux density of lighting source Good color uniformity RoHS compliant Energy Star binning structure, warm

More information

Report No.: HZ p. (Lumens) Stabilization Time CRI (K)

Report No.: HZ p. (Lumens) Stabilization Time CRI (K) Test Summary Sample Tested: ML2LA23LTRBNI827 Luminous Efficacy Total Luminous Flux Power (Lumens /Watt) (Lumens) (Watts) Power Factor 73.6 1737.0 23.61 0.9872 CCT Stabilization Time CRI (K) (Light & Power)

More information

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006 LM-79-08 Test Report For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006 Architectural Flood and Spot Luminaires Model name(s): FLS20U40B/N FLS20U45B/N FLS20U50B/N FLS20U40W/N FLS20U45W/N

More information

Photometric Test Report. Keystone Technologies

Photometric Test Report. Keystone Technologies Photometric Test Report Relevant Standards IES LM-79-2008 ANSI C82.77-10-2014 UL1598-2008 Prepared For Keystone Technologies 1390 Welsh Road,North Wales, PA 19454 Stephanie DeLaurentis, 267-656-2272, sdelaurentis@keystoneballast.com

More information

Photometric Test Report. Keystone Technologies

Photometric Test Report. Keystone Technologies Photometric Test Report Relevant Standards IES LM-79-2008 ANSI C82.77-10-2014 UL1598-2008 Prepared For Keystone Technologies 1390 Welsh Road,North Wales, PA 19454 Stephanie DeLaurentis, 267-656-2272, sdelaurentis@keystoneballast.com

More information

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006 LM-79-08 Test Report For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006 Architectural Flood and Spot Luminaires Model name(s): FLS20U40B FLS20U45B FLS20U50B FLS20U40W FLS20U45W FLS20U50W

More information

Main Applications MR 16 Track lighting Candle lighting

Main Applications MR 16 Track lighting Candle lighting ProLight PDSA-6FxL-xC1x 6W COB Light-Engine LEDs Technical Datasheet Version: 1.3 Features High flux density of lighting source Good color uniformity RoHS compliant Energy Star binning structure, neutral

More information

NVLAP LAB CODE LM Test Report. For LED PANEL LIGHTING CO.,LTD. (Brand Name: N/A)

NVLAP LAB CODE LM Test Report. For LED PANEL LIGHTING CO.,LTD. (Brand Name: N/A) LM-79-08 Test Report For LED PANEL LIGHTING CO.,LTD. (Brand Name: N/A) 7 F, Jinchangda Industrial Park,zhangkengjing,GuanLan, Baoao,Shenzhen,Guangdong,China 2x2 Luminaires for Ambient Lighting of Interior

More information

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK)

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK) LM-79-08 Test Report For AOK LED LIGHT COMPANY LIMITED (Brand Name: AOK) Building 1, St George's Science and Technology Industrial Park, Shajin Street, Shenzhen, Guangdong Province, China Zip 518104 Outdoor

More information

TeraLED. Thermal and Radiometric Characterization of LEDs. Technical Specification.

TeraLED. Thermal and Radiometric Characterization of LEDs. Technical Specification. TeraLED Thermal and Radiometric Characterization of LEDs Technical Specification M e c h a n i c a l a n a l y s i s TeraLED hardware options Depending on the dimensions and power level of the LED or LED

More information

LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl350u1

LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl350u1 LM-79-08 Test Report For Simkar Corporation (Brand Name: ) LED Luminaire Model name(s):salho2xl350u1 Test & Report By: Garman Mo Engineer: Garman Mo Date: Feb.20,2017 Review By: Manager: Tommy Liang Note:

More information

NVLAP LAB CODE LM Test Report. For. LED Wav es,inc. (Brand Name: LED Wav es)

NVLAP LAB CODE LM Test Report. For. LED Wav es,inc. (Brand Name: LED Wav es) LM-79-08 Test Report For LED Wav es,inc (Brand Name: LED Wav es) 4100 1st Ave,The Esquire Building,Brooklyn, NY 11232 2x2 Luminaires for Ambient Lighting of Interior Commercial Spaces Model name(s): SL2-2X2-40W-11

More information

Photometric Test Report

Photometric Test Report Photometric Test Report Relevant Standards IES LM-79-2008 ANSI C82.77-2002 UL1598-2008/ UL1993-2012 Prepared For DONGBU LIGHTEC CO. LTD 14, Saneop-ro 104beon-gil, Ojeong-gu, Bucheon-si, Gyeongg-do, Korea

More information

Improvement Uncertainty of Total luminous Flux Measurements by Determining Some Correction Factors

Improvement Uncertainty of Total luminous Flux Measurements by Determining Some Correction Factors ISSN: 2347-325 Volume 3 Number 6 (June-205) pp. 264-274 www.ijcrar.com Improvement Uncertainty of Total luminous Flux Measurements by Determining Some Correction Factors Manal A. Haridy* Photometry and

More information

NVLAP LAB CODE LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl5s50u1

NVLAP LAB CODE LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl5s50u1 LM-79-08 Test Report For Simkar Corporation (Brand Name: ) LED Luminaire Model name(s):salho2xl5s50u1 Test & Report By: Garman Mo Engineer: Garman Mo Date: Dec.05,2016 Review By: Manager: Tommy Liang Note:

More information

Practical Approaches to Uncertainty Analysis for Light Measurement Systems Using Integrating Spheres and Spectroradiometers

Practical Approaches to Uncertainty Analysis for Light Measurement Systems Using Integrating Spheres and Spectroradiometers Practical Approaches to Uncertainty Analysis for Light Measurement Systems Using Integrating Spheres and Spectroradiometers Dan Scharpf Labsphere, Inc. CORM 2015 Annual Conference Boulder, CO May 12-15,

More information

NVLAP LAB CODE LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl4s50u1

NVLAP LAB CODE LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl4s50u1 LM-79-08 Test Report For Simkar Corporation (Brand Name: ) LED Luminaire Model name(s):salho2xl4s50u1 Test & Report By: Garman Mo Engineer: Garman Mo Date: Mar. 01,2017 Review By: Manager: Tommy Liang

More information

EVERLIGHT ELECTRONICS CO.,LTD. Technical Data Sheet High Power LED 1W (Preliminary)

EVERLIGHT ELECTRONICS CO.,LTD. Technical Data Sheet High Power LED 1W (Preliminary) Features Feature of the device: small package with high efficiency Typical color temperature: 3500 K. Typical view angle: 120. Typical light flux output: 33 lm @ 350mA ESD protection. Soldering methods:

More information

LM Test Report. for. Forest Lighting Corporate. LED Tube. Model: FL-T8U BN

LM Test Report. for. Forest Lighting Corporate. LED Tube. Model: FL-T8U BN NVLAP LAB CODE 201071-0 LM-79-08 Test Report for Forest Lighting Corporate Address: 2252 Northwest Parkway SE, Suite D, Marietta, GA 30067 LED Tube Model: FL-T8U450-15-BN Laboratory: Leading Testing Laboratories

More information

LM Test Report. For. Beyond LED Technology. (Brand Name: Beyond LED) 2725 Mountain Industrial Blvd. Suite A-1 Tucker, GA 30084

LM Test Report. For. Beyond LED Technology. (Brand Name: Beyond LED) 2725 Mountain Industrial Blvd. Suite A-1 Tucker, GA 30084 LM-79-08 Test Report For Beyond LED Technology (Brand Name: Beyond LED) 2725 Mountain Industrial Blvd. Suite A-1 Tucker, GA 30084 ZOHO Series LED Flood Light Model name(s): BLTFLD50W5K Representative (Tested)

More information

Main Applications Par lighting LED Bulb Ceiling lighting Spot lighting Down lighting

Main Applications Par lighting LED Bulb Ceiling lighting Spot lighting Down lighting ProLight PACB-5FxL-BC2P 5W COB Light-Engine LEDs Technical Datasheet Version: 1.2 Features High flux density of lighting source Good color uniformity RoHS compliant Energy Star binning structure, neutral

More information

TEST REPORT : Report Reference No... : V2.0. Tested by (name + signature)... : Fair Deng.

TEST REPORT : Report Reference No... : V2.0. Tested by (name + signature)... : Fair Deng. TEST REPORT Report Reference No.... : 4323305.50 V2.0 Tested by (name + signature)... : Fair Deng. Approved by (name + signature)... : Jason Zhang.. Date of issue... Number of pages. Testing Laboratory...

More information

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006 LM-79-08 Test Report For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006 Architectural Flood and Spot Luminaires Model name(s): FLS40U40B FLS40U45B FLS40U50B FLS40U40W FLS40U45W FLS40U50W

More information

PHOTOMETRIC TESTING & EVALUATION TO IES LM Sample Tested LN130-A30 LED HIGH BAY. Prepared for: Janie Chung

PHOTOMETRIC TESTING & EVALUATION TO IES LM Sample Tested LN130-A30 LED HIGH BAY. Prepared for: Janie Chung PHOTOMETRIC TESTING & EVALUATION TO IES LM-79-08 Sample Tested LN130-A30 LED HIGH BAY Prepared for: Janie Chung KMW Inc. 183-6 Youngcheon-ro, Dongtan-myun Hwasung-si, Gyeonggi-do Korea 445-813 Phone: 253-826-7636

More information

Main Applications High specification down lighting LED Bulb Ceiling lighting Spot lighting Low-Bay lighting

Main Applications High specification down lighting LED Bulb Ceiling lighting Spot lighting Low-Bay lighting ProLight PABA-10FxL-x76x 10W COB Light-Engine LEDs Technical Datasheet Version: 1.3 Features High flux density of lighting source Good color uniformity RoHS compliant Energy Star binning structure, neutral

More information

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD LM-79-08 Test Report For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD (Brand Name: THAILIGHT) Sanhui Ind. Area, Cunwei, Hengli, Dongguan, China. Architectural Flood and Spot Luminaires Model name(s):

More information

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD LM-79-08 Test Report For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD (Brand Name: THAILIGHT) Sanhui Ind. Area, Cunwei, Hengli, Dongguan, China. Architectural Flood and Spot Luminaires Model name(s):

More information

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT8N20-A

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT8N20-A LM-79-08 Test Report for MLS CO., LTD NO.1 MULINSEN RD XIAOLAN TOWNSHIP ZHONGSHAN City, GUANGDONG Linear Ambient Luminaires Model: MT8N20-A Laboratory: Leading Testing Laboratories NVLAP CODE: 200960-0

More information

Main Applications High specification down lighting High profile architectural lighting Street lighting Stage lighting High-Bay & Low-Bay lighting

Main Applications High specification down lighting High profile architectural lighting Street lighting Stage lighting High-Bay & Low-Bay lighting ProLight PABB-100FxL-xAAx 100W COB Light-Engine LEDs Technical Datasheet Version: 1.2 Features High flux density of lighting source Good color uniformity RoHS compliant Energy Star binning structure, neutral

More information

Photometric Test Report

Photometric Test Report Photometric Test Report Relevant Standards IES LM-79-2008 ANSI C82.77-2002 UL1598-2008/ UL1993-2012 Prepared For MaxLite Inc 10 York Avenue, West Caldwell, NJ 07006 Ramesh Raghavan, 973-244-7300 x146,

More information

Standard-Tech Co. Ltd Testing Center

Standard-Tech Co. Ltd Testing Center Test Datasheet Test Location: [ ] 1. B802, No.11 Caipin Road, Guangzhou Science City, Guangzhou, [ ] Guangdong, China 2. R108, 1 st Floor No.69 GuangPu West Road, Guangzhou Science City, Guangzhou, Guangdong,

More information

LedEngin, Inc. High Luminous Efficacy White Power LedFlex Emitter LZ4-00CW15. Key Features. Typical Applications. Description

LedEngin, Inc. High Luminous Efficacy White Power LedFlex Emitter LZ4-00CW15. Key Features. Typical Applications. Description High Luminous Efficacy White Power LedFlex Emitter LZ4-00CW15 Key Features High Luminous Efficacy 15W White LED Unique package design with ceramic substrate, integrated glass lens and Flex PCB New industry

More information

Integrating Spheres in Molecular Spectrophotometry

Integrating Spheres in Molecular Spectrophotometry Integrating Spheres in Molecular Spectrophotometry Theory and Practice 2012 Perkin Elmer 2012 Perkin Elmer General Sphere Theory 3 Integrating Spheres Types of Sphere Measurements Total Reflectance (Specular

More information

Wet Locations Use 71400A 120 W PF V 2.9A 93% 144 pcs. Philips 130 LM/W 5000K Hrs. 5 Years. 24-1/4 Lbs 665*530*120 1 PCS

Wet Locations Use 71400A 120 W PF V 2.9A 93% 144 pcs. Philips 130 LM/W 5000K Hrs. 5 Years. 24-1/4 Lbs 665*530*120 1 PCS 71400A GAS STATION Canopy Luminaire Wet Locations Use QPL ID # PL9D2UU8ZA41 OVERALL LAMP PARAMETERS LED DRIVER LED LIFESPAN & ENVIRONMENT SAFETY&EMC OTHERS Model: Input Voltage Input Current Input Power

More information

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT5Z02-6

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT5Z02-6 LM-79-08 Test Report for MLS CO., LTD NO.1 MULINSEN RD XIAOLAN TOWNSHIP ZHONGSHAN City, GUANGDONG Linear Ambient Luminaires Model: MT5Z02-6 Laboratory: Leading Testing Laboratories NVLAP CODE: 200960-0

More information

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT5Z02-9

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT5Z02-9 LM-79-08 Test Report for MLS CO., LTD NO.1 MULINSEN RD XIAOLAN TOWNSHIP ZHONGSHAN City, GUANGDONG Linear Ambient Luminaires Model: MT5Z02-9 Laboratory: Leading Testing Laboratories NVLAP CODE: 200960-0

More information

NVLAP LAB CODE LM Test Report. For. Shenzhen Guanke Technologies Co. Ltd. (Brand Name: )

NVLAP LAB CODE LM Test Report. For. Shenzhen Guanke Technologies Co. Ltd. (Brand Name: ) LM-79-08 Test Report For Shenzhen Guanke Technologies Co. Ltd (Brand Name: ) 12th/13th floor of 3th building, Tongfuyu industrial area, heshuikou community, Gongming town 518106 Shenzhen, China. Replacement

More information

NIST Role in Supporting Solid State Lighting Initiative

NIST Role in Supporting Solid State Lighting Initiative CORM 2007 Annual Meeting, Gaithersburg, May 8-11, 2007 NIST Role in Supporting Solid State Lighting Initiative Yoshi Ohno, Ph.D. Optical Technology Division National Institute of Standards and Technology

More information

PLWK1595Cx8 COB Series

PLWK1595Cx8 COB Series PLWK1595Cx8 COB Series Product Datasheet Spot Light Down Light Track Light General Lighting Cylinder Light PAR Lamp Introduction The Plessey COB series provides high luminous flux paired with excellent

More information

Morris Products 53 Carey Road, Queensbury, NY 12804

Morris Products 53 Carey Road, Queensbury, NY 12804 Catalog Number 71622 UPC Number 60198671622 Description LED Ultra Thin Recessed Canopy Fixture 40W Voltage: 120/208/240/277 Shown with optional recessed mounting Bezel/Trim Cat# 71620 QPL ID # P29AE57D

More information

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK)

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK) LM-79-08 Test Report For AOK LED LIGHT COMPANY LIMITED (Brand Name: AOK) Building 1, St George's Science and Technology Industrial Park, Shajin Street, Shenzhen, Guangdong Province, China Zip 518104 Outdoor

More information

Main Applications High specification down lighting LED Bulb Ceiling lighting Spot lighting Low-Bay lighting

Main Applications High specification down lighting LED Bulb Ceiling lighting Spot lighting Low-Bay lighting ProLight PABS-6FxL-x63x 6W COB Light-Engine LEDs Technical Datasheet Version: 1.1 Features High flux densit of lighting source Good color uniformit RoHS compliant Energ Star binning structure, neutral

More information

Cat# LED High Bay

Cat# LED High Bay Cat# 71403 LED High Bay Model: 71403 Input Voltage Input Current Input Power Power Factor 100-277VAC 50/60HZ 1.7A Max 144 W PF 0.92 OVERALL LAMP Luminance 17371 LM PARAMETERS Luminous Efficiency 118 LM/W

More information

TECHNICAL GUIDE. Integrating Sphere Radiometry and Photometry

TECHNICAL GUIDE. Integrating Sphere Radiometry and Photometry TECHNICAL GUIDE Integrating Sphere Radiometry and Photometry sales@labsphere.com www.labsphere.com TABLE OF CONTENTS 1.0 Introduction to Sphere Measurements 1 2.0 Terms and Units 2 3.0 The Science of

More information

NVLAP LAB CODE LM Test Report. For CE INNOVATIONS LTD. (Brand Name: N/A) 911 Denison St Markham, ON L3R 3K4 Canada

NVLAP LAB CODE LM Test Report. For CE INNOVATIONS LTD. (Brand Name: N/A) 911 Denison St Markham, ON L3R 3K4 Canada LM-79-08 Test Report For CE INNOVATIONS LTD (Brand Name: N/A) 911 Denison St Markham, ON L3R 3K4 Canada Four-Foot Linear Replacement Lamps Model name(s): IR-T8-15W4FT16503500K IR-T8-15W4FT16504000K IR-T8-15W4FT16505000K

More information

MODEL NAME : LLDMWW0-15K*0*A

MODEL NAME : LLDMWW0-15K*0*A Eagle Eye (Outdoor LED Module) Application Note MODEL NAME : LLDMWW0-15K*0*A RoHS Compliant HALOGEN FREE (30)-4022 1 / 16 CONTENTS 1. Product Description 3/16 2. Specifications 2.1. Chromaticity Bins for

More information

LM-79 Report TEST REPORT NO. L523A

LM-79 Report TEST REPORT NO. L523A LM-79 Report TEST REPORT NO. L523A Test Result Summary Luminaire Manufacturer Pioneer Engineering Inc. Luminaire Model Number S ST-0246-5L-5K-120 Input Voltage RMS (V) 120.03 Input Current RMS (A) 0.28

More information

Cree XLamp CXA2 Studio LED

Cree XLamp CXA2 Studio LED Cree XLamp CXA2 Studio LED Product family data sheet CLD-DS100 Rev 2A Product Description The XLamp CXA2 Studio LED is optimized for motion picture, studio and photography applications, delivering high

More information

LM Test Report. For MaxLite Inc. LED Commercial Down lights Model: RR82540W

LM Test Report. For MaxLite Inc. LED Commercial Down lights Model: RR82540W LM-79-08 Test Report For MaxLite Inc. 12 York Ave West Caldwell NJ 07006, USA LED Commercial Down lights Model: RR82540W Laboratory: Leading Testing Laboratories NVLAP CODE: 200960-0 Tel: +86-571-56680806

More information

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING)

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING) LM-79-08 Test Report For Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING) Rm 1902, Easey Comm Bldg,253-261 Hennessy Rd Wanchai, HONG KONG Four-Foot Linear Replacement Lamps Model name(s): HC-T8-4FT-22W-ID(X,Clear)

More information

Stray Light Rejection in Array Spectrometers

Stray Light Rejection in Array Spectrometers Stray Light Rejection in Array Spectrometers Mike Shaw, Optical Technologies & Scientific Computing Team, National Physical Laboratory, Teddington, Middlesex, UK 1 Overview Basic optical design of an array

More information

EAHP1215WA2. Features. Description. Applications

EAHP1215WA2. Features. Description. Applications Features LM-80 Certified High Power COB & High CRI LED Multi-Chip Solution Dimension:12 mm x 15 mm x 1.6 mm Main Parameters: Luminous Flux, Forward Voltage, Chromaticity and Color Rendering Index RoHS

More information

YJ-BC-270H-G01 High CRI LED

YJ-BC-270H-G01 High CRI LED YJ-BC-270H-G01 PRODUCT: HIGH CRI CHIP ON BOARD LED FEATURES: 27 mm 27 mm x 0.5 mm chip-on-board LED Φ19 mm light emission surface 120 emission angle 95 min Ra DESCRIPTION Yuji LED s BC270H series high

More information

EVERLIGHT ELECTRONICS CO.,LTD.

EVERLIGHT ELECTRONICS CO.,LTD. Technical Data Sheet High Power LED 0.5W (Preliminary) Features Feature of the device: small package with high efficiency Typical view angle: 120 Typical color temperature: 3500 K. ESD protection. Soldering

More information

LED Measurement Instrumentation

LED Measurement Instrumentation R.Young LED Measurement Instrumentation Optronic Laboratories, Inc. ABSTRACT The production and use of LEDs is increasing rapidly. They are being used in applications previously occupied by traditional

More information

LUSTRON V3 LUSTROUS GREEN TECHNOLOGY OF LIGHTINGS LUSTRON V3. High Power Solid-State LED Light Source

LUSTRON V3 LUSTROUS GREEN TECHNOLOGY OF LIGHTINGS LUSTRON V3. High Power Solid-State LED Light Source Technical Datasheet LS45 High Power Solid-State LED Light Source Introduction For a brighter solid-state light source, Lustrous Technology is proud to release the new. Ideal for your high concentration

More information

Main Applications Par lighting LED Bulb Ceiling lighting Spot lighting Down lighting

Main Applications Par lighting LED Bulb Ceiling lighting Spot lighting Down lighting ProLight PACE-21FxL-xC3P 21W COB Light-Engine LEDs Technical Datasheet Version: 1.4 Features High flux densit of lighting source Good color uniformit RoHS compliant Energ Star binning structure, neutral

More information

RoHS. Dimming: PWM/VR/0-10V QPL ID # PL21D5Q1B95E. Cat# W LED FlatPanel Slipfitter Mount. EPA Rating 2.

RoHS. Dimming: PWM/VR/0-10V QPL ID # PL21D5Q1B95E. Cat# W LED FlatPanel Slipfitter Mount.   EPA Rating 2. Cat# 71893 380W LED FlatPanel Slipfitter Mount OVERALL LAMP PARAMETERS LED DRIVER LED Input Voltage Input Current Input Power Power Factor Luminance Luminous Efficiency CRI Beam Angle Main Structure Output

More information

NVLAP LAB CODE LM Test Report. For IGT LIGHTING INC. (Brand Name: IG T ) 3755 Lincoln St. Suite B, Riverside, CA 92503

NVLAP LAB CODE LM Test Report. For IGT LIGHTING INC. (Brand Name: IG T ) 3755 Lincoln St. Suite B, Riverside, CA 92503 Report No.: GZE150726-B LM-79-08 Test Report For IGT LIGHTING INC. (Brand Name: IG T ) 3755 Lincoln St. Suite B, Riverside, CA 92503 Four-Foot Linear Replacement Lamps Model name(s): IGTTU-1501850F-4BC

More information

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Report Number: Model: PL06510-001A KR8-00L-15D-40K-SSGC-FF w KR8-80L-10V-FR Date: 07/31/2015 Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Prepared

More information

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law Phys316 Exploration 2: Verifying Stefan-Boltzmann Relationship Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law Where A is the effective radiating area,

More information

March 26, Title: TEMPO 21 Report. Prepared for: Sviazinvest, OJSC. Prepared by: Cree Durham Technology Center (DTC) Ticket Number: T

March 26, Title: TEMPO 21 Report. Prepared for: Sviazinvest, OJSC. Prepared by: Cree Durham Technology Center (DTC) Ticket Number: T March 26, 2012 Title: TEMPO 21 Report Prepared for: Sviazinvest, OJSC Prepared by: Cree Durham Technology Center (DTC) Ticket Number: 10806-T Co NVLAP lab code 500070-0 The accreditation of the Cree Durham

More information

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report NVLAP Lab Code 500077-0 Report Number: Model: PL09245-001A SPAR38-1503025TD-12DE26-1 Date: 9/14/2016 Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report Prepared

More information

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Report Number: Model: PL05401-001 RULHT2MEC40K-12xxx Date: 02/11/2015 Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Prepared For: Bernard Krauska

More information

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report NVLAP Lab Code 500089-0 Report Number: Model: Date: RESTL-2014-0031 BXSPWA02MC-US 01/28/2014 Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Prepared

More information

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report NVLAP Lab Code 500089-0 Report Number: Model: Date: RESTL-2014-0032 BXSPWA03MG-US 01/28/2014 Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Prepared

More information

LED Entryway Light RoHS

LED Entryway Light RoHS 72508 LED Entryway Light RoHS QPL ID # PSMN8JUU Model: 72508 Input Voltage 120-277 VAC 60HZ Input Current 0.5A Max @120V Input Power 25W Power Factor PF 0.95 OVERALL LAMP PARAMETERS Luminance 3,121 LM

More information

Everlight EZHVD 1W LED Series

Everlight EZHVD 1W LED Series Everlight Introduction The High Voltage (HV) LED is a surface-mount high voltage component featuring compact size, high brightness, and DC input voltage that more closely resembles the V rms of AC mains

More information

Fundamental Concepts of Radiometry p. 1 Electromagnetic Radiation p. 1 Terminology Conventions p. 3 Wavelength Notations and Solid Angle p.

Fundamental Concepts of Radiometry p. 1 Electromagnetic Radiation p. 1 Terminology Conventions p. 3 Wavelength Notations and Solid Angle p. Preface p. xiii Fundamental Concepts of Radiometry p. 1 Electromagnetic Radiation p. 1 Terminology Conventions p. 3 Wavelength Notations and Solid Angle p. 4 Fundamental Definitions p. 7 Lambertian Radiators

More information

CBM-360 LEDs. CBM-360 Product Datasheet. Features: Table of Contents. Applications

CBM-360 LEDs. CBM-360 Product Datasheet. Features: Table of Contents. Applications CBM-360 LEDs Features: Extremely high optical output: Over 5,000 lumens from a single package (white) Table of Contents Technology Overview...2 Test Specifications...2 White Binning Structure...3 Chromaticity

More information

Received. Revised record

Received. Revised record Received MASS PRODUCTION PRELIMINARY CUSTOMER DESIGN DEVICE NO. : DHE-0000991 PAGE : 12 Revised record REV. DESCRIPTION RELEASE DATE 1 New spec 2011.03.09 2 Change Format 2011.03.31 3 1. Change Color Binning

More information

IST OLLA White Paper on the Necessity of Luminous Efficacy Measurement Standardisation of OLED Light Sources

IST OLLA White Paper on the Necessity of Luminous Efficacy Measurement Standardisation of OLED Light Sources IST- 004607 OLLA White Paper on the Necessity of Luminous Efficacy Measurement Standardisation of OLED Light Sources Public White Paper Updated version March 2008 Address for comments and suggestions:

More information

Cree XLamp XP-E2 Torch LEDs

Cree XLamp XP-E2 Torch LEDs Cree XLamp XP-E2 Torch LEDs Product family data sheet CLD-DS101 Rev 2H Product Description The XLamp XP-E2 Torch LED provides lumen output similar to the XLamp XP-G LED but in the smaller optical source

More information

SHP-TS GroLux. SHP-TS 600W GroLux PRODUCT OVERVIEW

SHP-TS GroLux. SHP-TS 600W GroLux PRODUCT OVERVIEW Patented arc tube delivers highest photosynthetic efficiency of any lamp in the world Performance maintained at an exceptional level due to improved structure of outer bulb Optimised for the highest Phytolumens

More information

NO LONGER ORDERABLE LUXEON S. Best punch and easy integration. For NEW designs consider these replacement products: ILLUMINATION

NO LONGER ORDERABLE LUXEON S. Best punch and easy integration. For NEW designs consider these replacement products: ILLUMINATION For NEW designs consider these replacement products: LUXEON CoB Core Range LUXEON CoB Core Range High Density Matrix Platform light engines LUXEON S Best punch and easy integration LUXEON S is a ridgid

More information

Cree XLamp CXB3050 LED

Cree XLamp CXB3050 LED Cree XLamp CXB3050 LED Product family data sheet CLD-DS124 Rev 3A Product Description The XLamp CXB3050 LED Array is a member of the second generation of the CXA family that delivers up to 30% higher efficacy

More information

Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR

Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR Application note Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR Introduction Applications that use spectrometers to measure the light energy of radiant sources require an irradiance-calibration,

More information

Uncertainty determination of correlated color temperature for high intensity discharge lamps.

Uncertainty determination of correlated color temperature for high intensity discharge lamps. Uncertainty determination of correlated color temperature for high intensity discharge lamps A.B. El-Bialy 1, M.M. El-Ganainy 2 and E.M. El-Moghazy 3 1 University College for Woman for Art, science and

More information

Novel Flux Calibration Source. CORM 2007 David C. Gross May 10, 2007

Novel Flux Calibration Source. CORM 2007 David C. Gross May 10, 2007 Novel Flux Calibration Source CORM 2007 David C. Gross May 10, 2007 LED Calorimetry Absolute Radiometry Design Goals Proof of Concept - Design and Results Total Spectral Flux Calibration Absolute Radiometry

More information

LM Test Report. For Elec-Tech International Co., Ltd. T8 LED TUBE Model: XX(XX:41-50)

LM Test Report. For Elec-Tech International Co., Ltd. T8 LED TUBE Model: XX(XX:41-50) LM-79-08 Test Report For Elec-Tech International Co., Ltd NO.1 JINFENG ROAD,TANGJIAWAN TOWN,XIANGZHOU DISTRICT,ZHUHAI CITY,GUANGDONG PROVINCE T8 LED TUBE Model: 542991XX(XX:41-50) Laboratory: Leading Testing

More information

LM Test Report. For Elec-Tech International Co., Ltd. T8 LED TUBE Model: XX(XX:41-50)

LM Test Report. For Elec-Tech International Co., Ltd. T8 LED TUBE Model: XX(XX:41-50) LM-79-08 Test Report For Elec-Tech International Co., Ltd NO.1 JINFENG ROAD,TANGJIAWAN TOWN,XIANGZHOU DISTRICT,ZHUHAI CITY,GUANGDONG PROVINCE T8 LED TUBE Model: 542961XX(XX:41-50) Laboratory: Leading Testing

More information

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report Report Number: Model: PL08473-001A SA21-16027MDFD-12DE26-1 Date: 5/18/2016 Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report Prepared For: Jonathan Vollers

More information

Integrating Sphere Spectroradiometer System for LED LPCE-2(LMS-8000A) Brochure. Global Office of Lisun Electronics Inc.

Integrating Sphere Spectroradiometer System for LED LPCE-2(LMS-8000A) Brochure. Global Office of Lisun Electronics Inc. Integrating Sphere Spectroradiometer System for LED LPCE-2(LMS-8000A) Brochure Global Office of Lisun Electronics Inc. http://www.lisungroup.com Lisun Group (Hong Kong) Limited Add: Room C, 15/F Hua Chiao

More information

Received. Revised record

Received. Revised record Received MASS PRODUCTION PRELIMINARY CUSTOMER DESIGN DEVICE NO. : DHE-0000991 PAGE : 12 Revised record REV. DESCRIPTION RELEASE DATE 1 New spec 2011.03.09 2 Change Format 2011.03.31 3 1. Change Color Binning

More information

SPEC NO: DSAJ6327 REV NO: V.1 DATE: JUL/31/2009 PAGE: 1 OF 8 APPROVED: WYNEC CHECKED:

SPEC NO: DSAJ6327 REV NO: V.1 DATE: JUL/31/2009 PAGE: 1 OF 8 APPROVED: WYNEC CHECKED: XPower PRELIMINARY SPEC KA-8070-2 SERIES Features ATTENTION OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC DISCHARGE SENSITIVE DEVICES High efficient lightsource. Designed for high current operation. Low

More information

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING)

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING) LM-79-08 Test Report For Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING) Rm 1902, Easey Comm Bldg,253-261 Hennessy Rd Wanchai, HONG KONG Four-Foot Linear Replacement Lamps Model name(s): HC-T8-4FT-13W-ID(X,

More information

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report NVLAP Lab Code 500077-0 Report Number: Model: PL06803-001A BPAR30S-0803040C-12DE26-1C100 Date: 9/26/2015 Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report Prepared

More information

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Report Number: Model: PL09285-001 ARE-EDG-5M-xx-06-E-UL-xx-700-xxxx-40K Date: 09/13/2016 Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Prepared For:

More information

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Report Number: Model: PL09294-001A ARE-EDG-3M-xx-06-E-UL-xx-700-xxxx Date: 09/14/2016 Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Prepared For:

More information

SPEC NO: DSAJ6328 REV NO: V.1 DATE: JUL/31/2009 PAGE: 1 OF 9 APPROVED: WYNEC CHECKED:

SPEC NO: DSAJ6328 REV NO: V.1 DATE: JUL/31/2009 PAGE: 1 OF 9 APPROVED: WYNEC CHECKED: XPower PRELIMINARY SPEC KADS-8070-2 SERIES Features ATTENTION OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC DISCHARGE SENSITIVE DEVICES Super high flux output and high luminance. Designed for high current

More information