High Precision Mass Measurement in Automation Tatjana Ivanova 1, Jānis Rudzītis 2

Size: px
Start display at page:

Download "High Precision Mass Measurement in Automation Tatjana Ivanova 1, Jānis Rudzītis 2"

Transcription

1 Solid State Phenomena Vol. 164 (2010) pp Online available since 2010/Jun/30 at (2010) Trans Tech Publications, Switzerland doi: / High Precision Mass Measurement in Automation Tatjana Ivanova 1, Jānis Rudzītis 2 1 Riga Technical University, Ezermalas iela 6, Riga, Latvia 2 Riga Technical University, Ezermalas iela 6, Riga, Latvia puchkova@inbox.lv, aria@latnet.lv Keywords: mass comparator, high precision measurement accuracy, factors influencing accuracy Abstract. High-precision mass measurement equipment is required in some areas of science and technology. Physics, chemistry, pharmaceutics and high precision mechanics are common examples. In metrology, high-precision scales are used for verification and calibration of lower precision mass measurement equipment (weights and scales). Mass comparators are the most accurate mass measurement instruments available today. It is a special type of electronic scales designed to compare mass of two weights. They can be automatic or manual, with various measurement ranges and accuracy classes. This article discusses principles of operation of mass comparators and practice of high-precision mass measurement. There are special computer programs that can be used in conjunction with these instruments, which may significantly improve measurement accuracy (when mass comparator is controlled remotely) as well as simplify calculations and reporting procedures. This article describes one of these programs ScalesNet32 which can be used with mass comparators produced by Sartorius (Germany). Introduction Mass measurement is practiced widely both in industry and in trade and commerce. Mass is defined as the quantity of matter of an object. However, the measurement of mass is mostly carried out by weighing, using a variety of balances. Reference weights are classified into categories depending upon their precision, material and type of construction. There are seven classes defined in the OIML Recommendation 111, namely classes E1, E2, F1, F2, M1, M2 and M3. The weights are made from a metal or metal alloy. Generally platinum, iridium (E1), stainless steel (E2, F1, F2), brass or plated bronze (F2, M1), cast iron (M2, M3) are used. For example, maximum permissible error of 1 kg weight is 0.5 mg for class E1 and 500 mg for M3. Typical price of E1 weight set (1 mg 1 kg) is 3160, and only 690 for class F2. Today the high-precision weights (classes E1-F2) are calibrated mostly using mass comparators. These devices compare mass of reference weight (A) with mass of test weight (B). The construction of these balances is essentially based on the electromagnetic force restoration scales. Mass comparators are available with capacities in the range from micrograms to kilograms with excellent repeatability and linearity. Electromagnetic Force Restoration Scales The most accurate electronic balances are based on electromagnetic force restoration (EMFR), also called electromagnetic force compensation. These weigh scale sensors comprise a parallel guidance mechanism that ensures the accurate introduction of the object to be weighed, one or more levers, and an electromagnetic system (magnet and coil) that assumes the role of the weight in a two-pan scale balance. The triangular knife edges of a pan balance are replaced by flexible bearings. Changing the ratio of the levers allows forces smaller than 1 N to balance much bigger ones. Today it is common to have a system with one, two, or even three levers, depending on the load range. For greater convenience of use, modern EMFR balances integrate a Roberval mechanism like that in industrial scales. The balance must hold its parallelogram shape to maintain a state of equilibrium regardless of where the load is placed on the platform. The balance must retain a perfect All rights reserved. No part of contents of this paper may be reproduced or transmitted in any form or by any means without the written permission of TTP, (ID: , Pennsylvania State University, University Park, United States of America-15/06/14,07:15:53)

2 20 Mechatronic Systems and Materials: Mechatronic Systems and Robotics parallelogram shape so that the system remains in equilibrium no matter where the load is placed on the platform. Equilibrium is maintained by a control system incorporating an optical position sensor. When the coil force compensates the gravitational force exerted by an unknown mass, optical sensor detects a stable predefined zero position that indicates a state of equilibrium. The electric current that measures the applied mass is digitized, temperature compensated, filtered, and finally displayed. Weigh scale sensors based on the EMFR principle are typically times more accurate than strain-gauge-based industrial load cells and other devices based on the measurement of deformation. Signal strength also compares favorably with strain gauge output. An ideal EMFR load cell has no spring constant, shows no deformation under load, and self-compensates for the thermal expansion of the levers. In reality, some elastic deformation due to the load must reproduce within a few nanometers in order to guarantee the required precision, and the coil must typically be positioned to a tolerance within 10 nm (~3/1000 mil) to eliminate the effects of the unwanted but residual spring constant of the lever system. These sensors typically offer a displayed resolution of between and 3 million digits (1 million digits = 1 ppm), depending strongly on the application. In the analytical market segment, resolutions of up to 50 million digits are required as well as sub-ppm precision (e.g., 10 µg increments over a 200 g range). Some mass comparators, called window comparators, even have resolutions of 1 billion digits over a limited range. Mass Comparators Mass comparator is a special type of electronic scales used for mass measurement by comparison of two masses: known (reference weight A) and unknown (test weight B). Unknown mass of test weight B is being compared with known mass of reference weight A. Both weights are measured multiple times and then mass difference is calculated. Comparators are most often used in measuring laboratories for calibration of weights and masses classes E1, E2, F1, F2, M1 according to OIML R111. Main part of the comparator is high-precision measuring system which is separated from electronic circuits. Such solution guarantees elimination of temperature influence (warming of electronic elements) on the indications of comparator. Comparator is equipped with big graphic backlit display with user-friendly menu. In order to ensure proper accuracy of measurements, a half-automatic system of external calibration with calibration weight has been applied. Comparators have an interface for communication with computer. Mass comparators (m.c.) are usually based upon EMFR sensors (see above). This type of sensor provides the highest accuracy of measurement. Overall construction of m.c. must be as rigid as possible to avoid any mechanical instability. Comparator is highly influenced by external conditions such as temperature, breeze and vibrations. For the purpose of proper conditions, it is necessary to use the comparator in a room with very precise temperature control (laminar air-conditioning) and place of usage should bee from any vibrations. Electronic part of m.c. is separated from the mechanical (and electromagnetic) part. There are two types of m.c.: manual and automatic. In manual comparators, operator has to change weights (A and B) manually during measurement cycles. Automatic comparators have special device used to change weights. This greatly eliminates influence of operator and improves accuracy and repeatability. Also, automatic comparators may be controlled remotely using computer. For example, it is possible to place weights into comparator during working day, and perform measurement automatically at night when there is less interference (vibrations, etc).

3 Solid State Phenomena Vol Fig. 1. Typical manual mass comparator Comparator is equipped with esthetic weighing chamber with glass anti-draught protection. All elements of the weighing chamber are manufactured from glass or metal which minimizes the influence of electrostatic charges on weighing result. High-precision mass measurement requires estimation of air density because buoyancy of weight may influence observed mass value. In fact, air density is part of definition of conventional mass. The most accurate mass comparators have vacuum camera in order to completely eliminate influence of air movements and density. These high-end devices can be found only in the highest-rated international laboratories. Mass comparators have very high resolution (millions and billions of discrete values). The very basic idea of comparator is that difference between masses of two items (DM=M1-M2) may be measured much better than their masses as such (M1, M2). If M1 is known, then M2=M1-DM. For example, measured masses of two weights are 975 and 977 mg. In reality, mass of first weight is 1001 mg (known from previous calibration). Unknown mass of second weight may be calculated: 1001+( )=1003 mg. Of course, this is very simplified example. In real life, several measurement cycles are performed in order to estimate mass difference and assure its consistency. Usual measurement cycle is ABBA (where A is reference weight and B is test weight). Result of the first cycle is discarded. Results of three or more consecutive cycles are used for calculations. The whole procedure (four or more cycles) may take substantial time (40 minutes and more). Results of ABBA cycles must meet several requirements. For example, there must be no significant difference between results of separate cycles and previous calibration of comparator. Otherwise, result of the whole measurement (several cycles) is discarded and procedure must be started over again. Sometimes it takes many hours to obtain acceptable result that meets all the criteria. Result of successful measurement is always corrected using parameters such as air density, weight density, previous calibration and adjustment result, etc. The final result is conventional mass value of test weight. All these checks and calculations are performed automatically by the controller of mass comparator or computer running special software (see below). For example, typical mass comparator Sartorius CC310 has measurement range up to 200 g, resolution 0.01 mg, standard deviation is 0.01 mg. It allows calibration of weights up to class E1 (highest). Professional analytic scales Sartorius LA230S have range up to 230 g, resolution 0.1 mg, standard deviation 0.1 mg.

4 22 Mechatronic Systems and Materials: Mechatronic Systems and Robotics Fig. 2. Typical automatic mass comparator Comparators are used mostly for high-precision measurement. Their users are national and international mass measurement laboratories and other similar organizations. This is a very small market for such complex devices. Therefore, prices are very high. Typical comparator may cost more than a new car. Two main vendors of comparators are "Sartorius" (Germany) and "Mettler Toledo" (Switzerland). For example, Sartorius is making comparators at their headquarters in Göttingen, Germany. There is small factory where complex equipment is manufactured piece-bypiece. Software Comparators may be connected to computer using various interfaces (RS232, RS485, USB). Special software expands capabilities of comparators. It allows remote control of comparator as well as simplified calculations, reporting and database features. For example, exact mass values of reference weight may be stored in database. Operator only selects required reference weight from the list. All tests on measurement results are performed entirely by software, making impossible reporting of invalid results. Air measurement instruments (thermometer, barometer, hygrometer) may be connected to the same computer for measurement of air density (this is needed for calculation of the conventional mass). For example, comparators made by Sartorius can be used in conjunction with ScalesNet32 software developed by "Maro Elektronik" (Germany). This software provides the following capabilities: calibration of customer weights, dissemination of Mass Scale, Calibration of reference weights, calibration and adjustment of weighing instruments (comparators), recording and graphic plotting of external parameters (air temperature, etc). When the weight is calibrated, the data that must be recorded includes the work order or customer, as well as weight-specific data such as serial number, manufacturer, design, characteristics, etc., to ensure unambiguous identification of the weight. This identification data is stored in a database, which makes it possible to generate a history of the weight at any time. The mass comparators used in calibration of weights are monitored by ScalesNet32, and must be calibrated and adjusted at specified intervals. The results of these procedures are also stored in a database. Furthermore, ScalesNet32 uses connected sensors to monitor reference weights and climate stations. The user is

5 Solid State Phenomena Vol informed when a particular reference weight, climate station or mass comparator is due for calibration; if calibration is not performed within a defined period, the particular device is blocked by ScalesNet32 from further use. The software consists of several components using the client-server model. ScalesServer is server part, using SQL database (MS SQL server) for centralized data storage. ScalesDesk and ScalesMass are clients. ScalesMass must be connected directly to comparators. All parts communicate over IP network. Calibration protocol is the output of ScalesNet32. This document contains all essential information regarding reference and test weights, measurement cycles, result testing and calculations: class, nominal mass, conventional mass and uncertainty, vendor, ID, density of reference weight. This weight must be traceable up to international prototype of kilogram; the same information for test weight (the weight being calibrated); results of three (or more) separate ABBA measurement cycles; results of conventional mass value calculation of test weight and its uncertainty. The procedure is performed automatically (controlled from laptop running ScalesMass component of ScalesNet32). At the same time, measurement of air density is performed. After that, the program calculates conventional and true mass values of "B", as well as its uncertainty. All intermediate results are included into protocol. Calculation algorithms comply with OIML Recommendation 111. These are explained in ScalesNet32 manual too. Unfortunately, ScalesNet32 has no means to evaluate long-term instability of scales and weight. Probably this is because the program has been designed for near-to-ideal conditions of use. In real life, long-term instability of measurement process may substantially increase overall uncertainty of results. However, existing recommendations and methods pay little attention to this fact. For example, calibration of the same set of weights every month over one year period may produce significant dispersion of results even if these weights and scales were handled very carefully and all their calibrations were successful. This is unclear issue which has to be investigated further. Usage of Mass Comparators Mass comparators are extremely sensitive to both electrical and mechanical environmental conditions. This is a price for improved convenience and accuracy. This fact is often overlooked by users and vendors of mass comparators. Expected accuracy of comparator may only be reached in nearly ideal conditions. For example, it is very hard to achieve class E1 performance even if comparator has been designed for this. Mechanical influences are important for any type of mass measurement system (mechanical or electronic). The laboratory must be built on stable ground. The building must be massive and stable (thick walls). Mass measurement must be carried out on the ground floor. There must be no large moving masses around (like trucks and trains). High-quality air conditioning is essential. Mass comparators are very susceptible to electromagnetic and electric influence. Strong magnetic fields may directly interact with electromagnetic system of the comparator. Poor quality of electrical power (harmonics, voltage instability, etc) degrades accuracy of the whole EMFR part of device. According to operation manuals, Sartorius mass comparators may be operated over wide range of acceptable voltage and frequency of grid power. However, we have discovered that stability of measurements is tightly correlated with stability of grid voltage. In fact, it can be predicted using the cheapest digital multimeter bought at marketplace. If voltage is jumping between volts, don't start measurement. It will fail. Wait until voltage becomes more stable. Of course, this effect may be easily eliminated using UPS with double conversion (also called "true on-line"). These UPS provide the best quality of output voltage regardless what they receive from grid. Other laboratories also found that proper grounding of mass comparators are very important.

6 24 Mechatronic Systems and Materials: Mechatronic Systems and Robotics Summary Use of comparators for high-accuracy mass measurement and specific requirements for their accurate work are described. Influence of environmental factors on measurement accuracy is emphasized. References [1] International Recommendation OIML R Edition 2004 (E), Weights of Classes E1, E2, F1, F2, M1, M1-2, M2, M2-3 and M3, Part 1: Metrological and technical requirements, International Organization of Legal Metrology, Paris (2004) [2] Traceability of Measurements, ILAC G2:1994, International Laboratory Accreditation Cooperation (ILAC) (1996) [3] Premium Accuracy for Weighing and Mass Comparison, Publication W 1064-e04116, Sartorius mechatronics, year of publication unknown, available online at [4] The Latvian law: On uniformity of measurements ( Par mērījumu vienotību ), initial edition 27 February 1997, current edition 21 July (2006) [5] Kochsiek M., Gläser M., Comprehensive Mass metrology, Mass Comparators, Physikalisch- Technische Bundesanstalt, Braunschweig (Germany) (1984)

7 Mechatronic Systems and Materials: Mechatronic Systems and Robotics / High Precision Mass Measurement in Automation /

TRACEABILITY AND CAPABILITY CONTROL OF MASS MEASUREMENT EQUIPMENT AND DRIFT STATISTICAL ANALYSIS OF NATIONAL MASS STANDARDS IN LATVIA

TRACEABILITY AND CAPABILITY CONTROL OF MASS MEASUREMENT EQUIPMENT AND DRIFT STATISTICAL ANALYSIS OF NATIONAL MASS STANDARDS IN LATVIA Traceability and capability control of mass measurement equipment and drift statistical analysis of national mass in atvia TRACEABIITY AND CAPABIITY CONTRO OF MA MEAUREMENT EQUIPMENT AND DRIFT TATITICA

More information

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Introduction Humidity generators are mechanical devices that provide stable temperature and humidity

More information

Traceable Mass Determination and Uncertainty Calculation

Traceable Mass Determination and Uncertainty Calculation Traceable Mass Determination and Uncertainty Calculation Falko Hilbrunner, Thomas Fehling, Siego Mühlich, Mario Schreiber Sartorius Lab Instruments GmbH Weender Landstrasse 94-108 Goettingen, Germany ABSTRACT

More information

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment

More information

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air INTERNATIONAL DOCUMENT OIML D 28 Edition 2004 (E) Conventional value of the result of weighing in air Valeur conventionnelle du résultat des pesées dans l'air OIML D 28 Edition 2004 (E) ORGANISATION INTERNATIONALE

More information

+ + ( indicator ) + (display, printer, memory)

+ + ( indicator ) + (display, printer, memory) CERTIFICATION OF STRAIN GAUGE LOAD CELL FAMILIES B. Meissner Physikalisch-Technische Bundesanstalt Weighing Instruments Laboratory 38116 Braunschweig, Germany Abstract: Testing and certification of load

More information

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60 Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60 Bernd Meißner Physikalisch-Technische Bundesanstalt Weighing Instruments Laboratory Braunschweig Abstract

More information

White Paper. Perform Conductivity Measurements In Compliance with USP <645>

White Paper. Perform Conductivity Measurements In Compliance with USP <645> Perform Conductivity Measurements In Compliance with USP Water is the most widely used substance, raw material, or ingredient in the production, processing and formulation of compendial articles.

More information

610,- KERN ADB. Analytical balances. Quick-Finder Analytical balances. Large glass draught shield with 3 sliding doors

610,- KERN ADB. Analytical balances. Quick-Finder Analytical balances. Large glass draught shield with 3 sliding doors Quick-Finder Readout [d] mg Weighing [Max] g Price excl. of VAT Page 0,01 101 ABT 100-5NM 1980,- 39 1 0,01 0,1 42 120 ABT 120-5DNM 2120,- 39 1 0,01 0,1 82 220 ALJ 200-5DA 1750,- 37 1 0,01 0,1 82 220 ABT

More information

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1. 076-00-067-ACC-S Mettler Toledo, LLC 900 Polaris Parkway Columbus, OH 0.800.METTLER Customer SAMPLE Accuracy Calibration Certificate Company: Address: City: Zip / Postal: State / Province: Mettler Toledo

More information

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION IMEKO 22 nd TC3, 15 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION R. Kumme

More information

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE Publication Reference EA-10/14 EA Guidelines on the Calibration of Static Torque Measuring Devices PURPOSE This document has been produced by EA to improve harmonisation in determining the calibration

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion

More information

Accuracy Calibration Certificate

Accuracy Calibration Certificate gl06-0-067-acc Mettler Toledo Service Business Unit Industrial 900 Polaris Parkway Columbus, OH 0-800-- Services Provided By: Customer Accuracy Calibration Certificate Company: Sample Company Test Address:

More information

473-SHX Dew Point Mirror

473-SHX Dew Point Mirror Humidity and Temperature Reference Hygrometer For Temperature up to 125 C Precise and stable chilled mirror dew point mirror technology High temperature optical components High temperature sample fan Cable

More information

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0 076-0-067-ACC-USL-S Mettler Toledo, LLC 900 Polaris Parkway Columbus, OH 0.800.METTLER Customer Accredited by the American Association for Laboratory Accreditation (ALA) CALIBRATION CERT #788.0 SAMPLE

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Laboratory locations: Schedule of Accreditation 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Unit 1G Kilroot Business Park Carrickfergus Co Antrim BT38 7PR Contact: Mr R E Hughes Tel:

More information

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT Michael Borys, Michael Gläser, Michael Mecke Physikalisch-Technische Bundesanstalt (PTB), Germany ABSTRACT Spheres made of a silicon

More information

THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY

THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY D. Peschel 1, D. Mauersberger 1, D. Schwind 2, U. Kolwinski 2 1 Solid mechanics department, PTB, Germany 2 Gassmann Theiss Messtechnik

More information

Problem 1 in the section "self-thinking" Team «MG 12»

Problem 1 in the section self-thinking Team «MG 12» Problem 1 in the section "self-thinking" Team «MG 12» 1 Problem 1 in the section "self-thinking" Calculate the error in the measurement of mass of object by balance, taking into account external factors.

More information

The Mechatronics Design for Measuring Fluid Friction Losses in Pipe Flows Rıza Gurbuz

The Mechatronics Design for Measuring Fluid Friction Losses in Pipe Flows Rıza Gurbuz Solid State Phenomena Vol. 113 (2006) pp 603-608 Online available since 2006/Jun/15 at www.scientific.net (2006) Trans Tech Publications, Switzerland doi:10.4028/www.scientific.net/ssp.113.603 The Mechatronics

More information

Automated volume measurement for weihts using acoustic volumeter

Automated volume measurement for weihts using acoustic volumeter IMEKO 20th TC3, 3rd TC16 and 1st TC22 International Conference Cultivating metrological knowledge 27th to 30th November, 2007. Merida, Mexico. Automated volume measurement for weihts using acoustic volumeter

More information

Automatic Comparison of Weights and Mass Standards

Automatic Comparison of Weights and Mass Standards Automatic Comparison of Weights and Mass Standards Sławomir Janas, Michał Solecki RADWAG Balances and Scales Tadeusz Szumiata Kazimierz Pulaski University of Technology and Humanities in Radom Automatic

More information

Technical Note Development of a New, High Sensitivity 2000 kg Mechanical Balance

Technical Note Development of a New, High Sensitivity 2000 kg Mechanical Balance Technical Note Development of a New, High Sensitivity 2000 kg Mechanical Balance Jian Wang Division of Mechanics and Acoustics, National Institute of Metrology, Beijing 100029, China; wjian@nim.ac.cn;

More information

White Paper. Moisture Analyzer Routine Testing

White Paper. Moisture Analyzer Routine Testing Moisture Analyzer Routine Testing This white paper describes the influences and sources of error which may be present when conducting moisture analyses. It discusses the routine tests which are necessary

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

Guidelines on the Calibration of Static Torque Measuring Devices

Guidelines on the Calibration of Static Torque Measuring Devices European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET/cg-14/v.01 Previously EA-10/14 July 2007 Calibration Guide EURAMET/cg-14/v.01

More information

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure DATASHEET Pascal ET Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure - electrical signals (ma, mv, V, ) - temperature (TC and

More information

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory Bülent ÖZGÜR 1,*, Okhan GANİOĞLU 1, Nasser Al-Qahtani 2, Faisal Al-Qahtani 2 1 TÜBİTAK, National

More information

A Research on High-Precision Strain Measurement Based on FBG with Temperature Compensation Zi Wang a, Xiang Zhang b, Yuegang Tan c, Tianliang Li d

A Research on High-Precision Strain Measurement Based on FBG with Temperature Compensation Zi Wang a, Xiang Zhang b, Yuegang Tan c, Tianliang Li d Advanced Materials Research Submitted: 214-1-31 ISSN: 1662-8985, Vol 183, pp 121-126 Accepted: 214-11-3 doi:1428/wwwscientificnet/amr183121 Online: 215-1-12 215 Trans Tech Publications, Switzerland A Research

More information

Compact Titration. Compact Titrator G20. One Click Titration Simple & Dependable

Compact Titration. Compact Titrator G20. One Click Titration Simple & Dependable Compact Titration Compact Titrator G20 One Click Titration Simple & Dependable Simplicity The Intelligent Helper for Simple Titrations One Click Titration Intuitive method adaptation Accurate results simply

More information

Model 2300XP PSL & Process-Particle Wafer Deposition System

Model 2300XP PSL & Process-Particle Wafer Deposition System Model 2300XP PSL & Process-Particle Wafer Deposition System Deposit PSL spheres on wafers to create NISTtraceable PSL size standards for - calibrating wafer inspection systems - providing fab-wide and

More information

Air Density Determination Using 1 kg Buoyancy Mass Comparison(III)

Air Density Determination Using 1 kg Buoyancy Mass Comparison(III) Air Density Determination Using 1 kg Buoyancy Mass Comparison(III) Jin-Wan CHUNG, Kyung-Ho CHANG, Woo-Gab LEE, Kwang-Pyo KIM Division of Physical Metrology, Korea Research Institute of Standards and Science(KRISS)

More information

1. Invent Yourself: Physics Precise Weighting

1. Invent Yourself: Physics Precise Weighting 1. Invent Yourself: Physics Precise Weighting Tibor Basletid Požar 1. Invent yourself: Physics Topic: precise weighting. Study the physical effects that influence precise weighting of solid objects with

More information

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software V5.2 Plutonium and Uranium Isotopic Analysis Software Advanced Isotopic Ratio Analysis Software for HPGe Gamma-Ray Spectra Analyzes Pu, and a wide variety of heterogeneous samples containing Pu, Am, U,

More information

THE NEW WATT BALANCE EXPERIMENT AT METAS

THE NEW WATT BALANCE EXPERIMENT AT METAS XX IMEKO World Congress Metrology for Green Growth September 9 14, 2012, Busan, Republic of Korea THE NEW WATT BALANCE EXPERIMENT AT METAS H. Baumann, A. Eichenberger, B. Jeckelmann, Ch. Beguin +, R. Clavel,

More information

Weighted Stability Index (WSI) Metric Model Mike Libassi Intel Corp 8/11/99

Weighted Stability Index (WSI) Metric Model Mike Libassi Intel Corp 8/11/99 Weighted Stability Index (WSI) Metric Model Mike Libassi Intel Corp 8/11/99 Abstract Methods, such as McCabe's Cyclomatic Complexity, have proven that complexity is a reliable predictor of defects. Although

More information

Designed to go where you work

Designed to go where you work Designed to go where you work Finally, a digital substitute for your mercury-in-glass thermometers! Accurate and repeatable to ± 0.05 C over its full range, the 55A/55A Stik Thermometer is the new gold

More information

Loadcell Calibration - Evaluation of Uncertainties

Loadcell Calibration - Evaluation of Uncertainties Loadcell Calibration - Evaluation of Uncertainties This document is a revision of the evaluation issued in June 2015. Novatech s calibration laboratory is not UKAS accredited due to the high number of

More information

Verification for Non-Automatic Weighing Instruments (NAWI)

Verification for Non-Automatic Weighing Instruments (NAWI) Verification for Non-Automatic Weighing Instruments (NAWI) Every instrument should be accurate and correct to be permitted for use in any commercial transaction, particularly at the Point Of Sale. An instrument

More information

FAVIMAT (AI)ROBOT2 FAVIGRAPH. Automatic Single-Fibre Testers

FAVIMAT (AI)ROBOT2 FAVIGRAPH. Automatic Single-Fibre Testers FAVIMAT (AI)ROBOT2 FAVIGRAPH Automatic Single-Fibre Testers Single-Fibre Testers FAVIMAT and FAVIGRAPH The FAVIMAT testing methods Testing and evaluation of a wide range of fibre properties are essential

More information

MEASURING THE COMPLEXITY AND IMPACT OF DESIGN CHANGES

MEASURING THE COMPLEXITY AND IMPACT OF DESIGN CHANGES Presentation Paper Bio Return to Main Menu P R E S E N T A T I O N W4 Wednesday, March 8, 2000 11:00AM MEASURING THE COMPLEXITY AND IMPACT OF DESIGN CHANGES Mike Libassi Intel Corporation International

More information

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012 Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability

More information

1551A Ex/1552A Ex Intrinsically Safe Stik Thermometer. The new gold standard of industrial temperature calibration

1551A Ex/1552A Ex Intrinsically Safe Stik Thermometer. The new gold standard of industrial temperature calibration 1551A Ex/155A Ex Intrinsically Safe Stik Thermometer The new gold standard of industrial temperature calibration Designed to go where you work Finally, a digital substitute for your mercury-in-glass thermometers!

More information

OIML R 141 RECOMMENDATION. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 141 RECOMMENDATION. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 141 Edition 2008 (E) Procedure for calibration and verification of the main characteristics of thermographic instruments Procédure pour l'étalonnage et la vérification

More information

The New Quintix. Intuitive application guideance. Maximum accuracy at all time. Easy cleaning. Plug & Work connectivity to printer or computer

The New Quintix. Intuitive application guideance. Maximum accuracy at all time. Easy cleaning. Plug & Work connectivity to printer or computer Te New Quintix. Convenient Weiging. Intuitive application guideance Laboratory Balances Maximum accuracy at all time Easy cleaning Plug & Work connectivity to printer or computer AC Adapter Sartorius AC

More information

OIML R 76-2 RECOMMENDATION. Edition 2007 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Non-automatic weighing instruments

OIML R 76-2 RECOMMENDATION. Edition 2007 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Non-automatic weighing instruments INTERNATIONAL RECOMMENDATION OIML R 76-2 Edition 2007 (E) Non-automatic weighing instruments Part 2: Test report format Instruments de pesage à fonctionnement non automatique Partie 2: Format du rapport

More information

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Jin Wan Chung 1, Shih Mean Lee, Sungjun Lee 1 1 Korea Research Institute of Standards

More information

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

STATUS OF THE WIGOS DEMONSTRATION PROJECTS STATUS OF THE WIGOS DEMONSTRATION PROJECTS Demonstration Project Morocco Strengthening Moroccan RIC Capacities (Submitted by Rabia Merrouchi, National Meteorological Service of Morocco (DMN)) February

More information

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors 1. Scope... 2 2. Introduction... 2 3. Selection of thermocouples

More information

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement Publication Reference EA-10/11 EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical PURPOSE This document has been produced by EA as a means of giving advice for calibrating

More information

Infrared Absorption Measurement of Carbon Concentration Down to 1x10 14 /cm 3 In CZ Silicon

Infrared Absorption Measurement of Carbon Concentration Down to 1x10 14 /cm 3 In CZ Silicon Solid State Phenomena Vols. 18-19 (25) pp 621-626 Online available since 25/Dec/15 at www.scientific.net (25) Trans Tech Publications, Switzerland doi:1.428/www.scientific.net/ssp.18-19.621 Infrared Absorption

More information

Understanding Weighing A Grocer s Guide to Metrology

Understanding Weighing A Grocer s Guide to Metrology Understanding Weighing A Grocer s Guide to Metrology Supermark Backroom Service Counter Fresh Produce kg Introduction Metrology is a term you might not be familiar with, but it s something that impacts

More information

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas Turbine Meter TRZ 0 PRODUCT INFORMATION Reliable Measurement of Gas TURBINE METER TRZ 0 Method of operation, Construction Method of operation The TRZ 0 turbine meter is a flow meter suitable for gas measurement

More information

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY Proceedings of the 6th International Conference on Mechanics and Materials in Design, Editors: J.F. Silva Gomes & S.A. Meguid, P.Delgada/Azores, 26-30 July 2015 PAPER REF: 5376 STATISTICAL PROCESS CONTROL

More information

After publication, this Indian Standard will supersede the following Indian Standards:

After publication, this Indian Standard will supersede the following Indian Standards: For Comments Only Doc: PGD 26(1110)W Draft Indian Standard SPECIFICATION FOR NON-AUTOMATIC WEIGHING INSTRUMENTS PART 1 TERMINOLOGY AND DEFINITIONS ICS No. 01.040.17; 17.100 Not to be reproduced without

More information

ME6504 METROLOGY AND MEASUREMENTS

ME6504 METROLOGY AND MEASUREMENTS ME6504 METROLOGY AND MEASUREMENTS Lecture by, M.E., Ph.D. Associate Professor/ Mechanical E.G.S. Pillay Engineering College, Nagapattinam SYLLABUS 06-Jul-18 2 UNIT I BASICS OF METROLOGY Introduction to

More information

792A AC/DC Transfer Standard

792A AC/DC Transfer Standard 92A AC/DC Transfer Standard Technical Data Support for your most demanding ac measurement requirements ppm total uncertainty Traceable to national standards range: 2 mv to 00 V Frequency range: Hz to 1

More information

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks INTERNATIONAL STANDARD ISO 6506-3 Third edition 2014-10-01 Metallic materials Brinell hardness test Part 3: Calibration of reference blocks Matériaux métalliques Essai de dureté Brinell Partie 3: Étalonnage

More information

792A AC/DC Transfer Standard

792A AC/DC Transfer Standard 92A AC/DC Transfer Standard Technical Data Support for your most demanding ac measurement requirements ppm total uncertainty Traceable to national standards range: 2 mv to 00 V Frequency range: Hz to 1

More information

The Vaisala AUTOSONDE AS41 OPERATIONAL EFFICIENCY AND RELIABILITY TO A TOTALLY NEW LEVEL.

The Vaisala AUTOSONDE AS41 OPERATIONAL EFFICIENCY AND RELIABILITY TO A TOTALLY NEW LEVEL. The Vaisala AUTOSONDE AS41 OPERATIONAL EFFICIENCY AND RELIABILITY TO A TOTALLY NEW LEVEL. Weather Data Benefit For Society The four most important things about weather prediction are quality, reliability,

More information

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE Michael Bair and Pierre Delajoud 2002 DEC 10 Revised 2004 MAR 26 FORWARD FPG8601 is a pressure standard designed

More information

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018 Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018 Guidelines on the Calibration of Automatic Instruments for

More information

INC 331 Industrial Process Measurement. Instrument Characteristics

INC 331 Industrial Process Measurement. Instrument Characteristics INC 331 Industrial Process Measurement Instrument Characteristics Introduction Measurement is the experimental process of acquiring any quantitative information. When doing a measurement, we compare the

More information

THERMOCOUPLE CHARACTERISTICS TRAINER

THERMOCOUPLE CHARACTERISTICS TRAINER THERMOCOUPLE CHARACTERISTICS TRAINER (Model No : ) User Manual Version 2.0 Technical Clarification /Suggestion : / Technical Support Division, Vi Microsystems Pvt. Ltd., Plot No :75,Electronics Estate,

More information

Measuring and Analyzing the Moment of Friction in Double-Row Ball Bearings Robert Korbut 1,a, Jerzy Nachimowicz 1,b

Measuring and Analyzing the Moment of Friction in Double-Row Ball Bearings Robert Korbut 1,a, Jerzy Nachimowicz 1,b Solid State Phenomena Online: 2013-03-18 ISSN: 1662-9779, Vol. 199, pp 149-154 doi:10.4028/www.scientific.net/ssp.199.149 2013 Trans Tech Publications, Switzerland Measuring and Analyzing the Moment of

More information

Article for the 28 th Sensing Forum

Article for the 28 th Sensing Forum Article for the 28 th Sensing Forum Investigation of the Basic Performance of Analytical Balances Proposal of a microbalance performance evaluation method that includes the installation environment Presented

More information

INTERNATIONAL OIML R 61-2 RECOMMENDATION

INTERNATIONAL OIML R 61-2 RECOMMENDATION INTERNATIONAL OIML R 61-2 RECOMMENDATION Edition 2004 (E) Automatic gravimetric filling instruments Part 2: Test report format Doseuses pondérales à fonctionnement automatique Partie 2: Format du rapport

More information

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,

More information

Air velocity & IAQ measuring instruments

Air velocity & IAQ measuring instruments Data sheet dp Air velocity & IAQ measuring instruments - air velocity & IAQ measuring instrument - dp air velocity & IAQ measuring instrument incl. differential pressure C CO Intuitive: clearly structured

More information

Introduction To Metrology

Introduction To Metrology Introduction To Metrology Meaning of Metrology Metrology is the science of measurement. Metrology may be divided depending upon the quantity to be measured like metrology of length, metrology of time.

More information

PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY

PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY WHITEPAPER ENGLISH PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY PARTICLE MEASUREMENT Particle measurement in cleanrooms. WP1508006-0100-EN, V1R0, 2015-08 PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY

More information

Air velocity & IAQ measuring instruments

Air velocity & IAQ measuring instruments Data sheet dp Air velocity & IAQ measuring instruments - air velocity & IAQ measuring instrument - dp air velocity & IAQ measuring instrument incl. differential pressure C CO Intuitive: clearly structured

More information

Specific Accreditation Guidance

Specific Accreditation Guidance Specific Accreditation Guidance Materials Characterisation of Metallic Items by X-Ray Fluorescence and Atomic (Arc/Spark) Emission Techniques August 2018 Copyright National Association of Testing Authorities,

More information

Calibration of Temperature Block Calibrators

Calibration of Temperature Block Calibrators European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)

More information

ELG4112. Electromechanical Systems and Mechatronics

ELG4112. Electromechanical Systems and Mechatronics ELG4112 Electromechanical Systems and Mechatronics 1 Introduction Based on Electromechanical Systems, Electric Machines, and Applied Mechatronics Electromechanical systems integrate the following: Electromechanical

More information

PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS

PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS XIX IMEKO World Congress Fundamental and Applied Metrology September 611, 009, Lisbon, Portugal PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS Adriana Vâlcu 1, Sterică Baicu 1 National Institute

More information

OIML R 50-3 RECOMMENDATION. Edition 2014 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 50-3 RECOMMENDATION. Edition 2014 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 50-3 Edition 2014 (E) Continuous totalizing automatic weighing instruments (belt weighers). Part 3: Test report format Instruments de pesage totalisateurs continus à

More information

Strain Measurement Techniques for Composite Coupon Testing

Strain Measurement Techniques for Composite Coupon Testing Strain Measurement Techniques for Composite Coupon Testing Introduction Characterization of the properties of anisotropic and inhomogeneous composite materials for use in demanding structural applications

More information

Glossary Innovative Measurement Solutions

Glossary Innovative Measurement Solutions Glossary GLOSSARY OF TERMS FOR TRANSDUCERS, LOAD CELLS AND WEIGH MODULES This purpose of this document is to provide a comprehensive, alphabetical list of terms and definitions commonly employed in the

More information

Compact Titration Compact Titrator G20. One Click Titration Simple & Dependable

Compact Titration Compact Titrator G20. One Click Titration Simple & Dependable Compact Titration Compact Titrator G20 One Click Titration Simple & Dependable Simplicity The Intelligent Helper for Simple Titrations One Click Titration Intuitive method adaptation Accurate results simply

More information

The Elzone II Particle Count and Size Analyzer

The Elzone II Particle Count and Size Analyzer The Elzone II 5390 Particle Count and Size Analyzer Elzone II Overview of Basic Theory Operates using the Electrical Sensing Zone (ESZ) Principle, also known as the Coulter Principle. First commercially

More information

Edgetech Instruments Inc.

Edgetech Instruments Inc. 2015_07_ETI OVERVIEW Rev 3 MADE IN THE USA Edgetech Instruments (ETI) is a USA company based in Massachusetts. We have over 50 years of expertise in the design and manufacture of a comprehensive range

More information

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014 Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares

More information

Calibration and verification: Two procedures having comparable objectives and results

Calibration and verification: Two procedures having comparable objectives and results Calibration and verification: Two procedures having comparable objectives and results KLAUS-DIETER SOMMER, Landesamt für Messund Eichwesen Thüringen (LMET), Germany SAMUEL E. CHAPPELL, Consultant, Formerly

More information

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION IMEKO 22 nd TC3, 12 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES

More information

6 Securing of the delivery and acceptance

6 Securing of the delivery and acceptance 6 Securing of the delivery and acceptance Sector/Field Ox activities 1 Building Energy consumption 2 Spaces 3 Envelope, walls, roof, base floors, windows and doors Total energy consumption by types of

More information

Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge

Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge 2090TN05 Revised 2004 March 26 Technical Note FPG8601 is a pressure standard designed to cover the range from

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of:. Av. Rio Consulado # 2614, Col. San Juan de Aragón

More information

Design of Intelligent Meteorological System Based on MEMS Wang Hao 1,a, Nie Meng 2,b, Huang Qing an 3,c

Design of Intelligent Meteorological System Based on MEMS Wang Hao 1,a, Nie Meng 2,b, Huang Qing an 3,c Key Engineering Materials Online: 2014-04-09 ISSN: 1662-9795, Vols. 609-610, pp 801-806 doi:10.4028/www.scientific.net/kem.609-610.801 2014 Trans Tech Publications, Switzerland Design of Intelligent Meteorological

More information

OIML R 51-2 RECOMMENDATION. Edition 2006 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Automatic catchweighing instruments

OIML R 51-2 RECOMMENDATION. Edition 2006 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Automatic catchweighing instruments INTERNATIONAL RECOMMENDATION OIML R 51-2 Edition 2006 (E) Automatic catchweighing instruments Part 2: Test report format Instruments de pesage trieurs-étiqueteurs à fonctionnement automatique Partie 2:

More information

Temperature. 3

Temperature. 3 Temperature In 1848, Sir William Thomson (Lord Kelvin) stated the zero principle of dynamics. This principle enabled him to define thermodynamic temperature and to establish an objective method of measuring

More information

Chilled Mirror Hygrometry

Chilled Mirror Hygrometry Chilled Mirror Hygrometry Precision instruments for critical measurement and control applications Chilled Mirror Hygrometry Wherever precise, fundamental measurements are required in your laboratory or

More information

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

Magnetic Field Characteristics of the Flat Type Electro-Magnetic Barrel Finishing Machine O. Sugiura

Magnetic Field Characteristics of the Flat Type Electro-Magnetic Barrel Finishing Machine O. Sugiura Key Engineering Materials Online: 2004-02-15 ISSN: 1662-9795, Vols. 257-258, pp 515-520 doi:10.4028/www.scientific.net/kem.257-258.515 2004 Trans Tech Publications, Switzerland Magnetic Field Characteristics

More information

VMS-GeoMil. Background

VMS-GeoMil. Background Background When using a drilling rig for cone penetration testing, a mechanical clamp can be mounted to the drilling head (by means of a special transition piece). The depth than can be achieved depends

More information