CHAPTER III EXPERIMENTAL SYSTEM

Size: px
Start display at page:

Download "CHAPTER III EXPERIMENTAL SYSTEM"

Transcription

1 CHAPTER III EXPERIMENTAL SYSTEM 3.1 Introduction The basic design and implementation of laser induced Raman scattering systems is, in general, relatively simple. Laser light from a single laser source is focused to a point (small cross-sectional area) in a flame/sample to induce Raman scattering activity. The light collection optics are positioned 90 to the laser beam and on the same plane as the direction of propagation to improve the spatial resolution of the system. The added benefit of this arrangement is the minimization of Mie scattering interferences, i.e. scattering from large particles other than molecular species. The collection optics are as large as possible since Raman scattering is relatively weak. A spectrograph with high efficiency and throughput (low f#) is used to disperse the collected light into its component wavelengths, or Raman signals, while providing excellent spatial imaging resolution. Filters are used at the spectrograph entrance plane to block out Rayleigh scattered and stray light from interfering with the measurement of the Raman signals. An intensified charged coupled device (CCD) camera simultaneously measures the intensity and spatial distribution of all the Raman signals. The first part of this work deals with Raman spectroscopy performed in steady atmospheric flames to develop the Raman bandshape fitting technique for use as a temperature probe in practical combustion devices. These laboratory experiments are used to identify the response of the UV Raman system and validate the accuracy of the temperature measurements. The validated Raman bandshape technique is then used to measure temperature during testing of a sub-scale rocket-engine-like test article. 3.2 Laser Source The H 2 Stokes Raman scattered signals are induced by a Lambda-Physik COMPex 150T narrowband KrF excimer laser that is tunable from cm -1 ( nm). Typically, when the laser is operated in the narrow bandwidth mode, a small fraction of the emitted light is broadband (resulting from unpolarized amplifier spontaneous emission) (Grünefeld et al. 1996). The ratio of the intensity of the narrowband component of the laser 37

2 output to the total output of the laser is defined as the locking efficiency, η. To increase η, also the degree of linear polarization which is directly proportional to η, and locking repeatability from shot-to-shot, the laser is modified to operate in a single-pass configuration (Grünefeld et al. 1996). A fraction of the laser beam is sent through an etalon to monitor the laser locking efficiency in real-time. The locking efficiency for the single-pass configuration used in the present work is ~85%. The broadband output spectrum, obtained by collecting the Rayleigh scattered light as shown in Fig. 3-1, is the asymmetrical emission spectrum of the KrF excimer which has a full spectral bandwidth of ~250cm -1 and a peak emission, ν B, at ~ cm -1. Since the broadband laser linewidth is much broader than the spectrometer/detector linewidth (~6.75cm -1 ) and the spectrometer entrance slit is extremely small, the instrumental linewidth for the broadband laser emission, ν B = 50cm -1, is determined by comparing the Rayleigh spectrum with a calculated profile using a 80%Gaussian/20%Lorentzian lineshape distribution function. The asymmetry in the broadband KrF emission spectrum is clearly visible. A sample spectrum of the Rayleigh scattered light from the narrowband laser is shown in Fig The narrowband Gaussian/Lorentzian lineshape fitted to the spectrum reveals asymmetry and broadening induced by the incomplete locking efficiency of the laser. Therefore instead of the single narrowband 80%Gaussian/20%Lorentzian profile, a superposition of two 80%Gaussian/20%Lorentzian functions using the narrowband and broadband laser bandwidths using the broadband laser center frequency of peak intensity scaled by the locking efficiency is used to fit the experimental Rayleigh spectrum with sufficient accuracy. The importance of this signal mixing process, the lineshape distribution functions, and their implications on accurately fitting the experimental spectra are discussed in Chapter Four. Additionally, fitting the narrowband Rayleigh spectrum using the method described in Chapter Four reveals the bandwidth of the narrowband laser, ν L = 0.8cm -1 (0.005nm), and verifies the locking efficiency with the value obtained from the etalon, ~85%. The output of the KrF laser has enough photon energy to excite several rotational transitions of the A 2 Σ(v =3) X 2 Π(v =0) band of the OH molecule. The resulting fluorescence via the (3 1) band of strong emission lines between cm -1 interfere with the H 2 O and H 2 Raman signals. Fluorescence interferences can be significantly reduced by properly tuning the narrowband laser emission away from electronically accessible molecular transitions 38

3 Broadband KrF Laser Linewidth 80%Gaussian+20%Lorentzian Profile Measured Intensity 1 Normalized Intensity (arb. units) Frequency (cm -1 ) Fig. 3-1 Rayleigh spectrum of the broadband KrF excimer laser compared to 80%Gaussian/20%Lorentzian lineshape distribution with Γ = 50 cm

4 Narrowband KrF Laser Linewidth Narrowband (80%Gaussian/20%Lorentzian) Mixed Narrow and Broadband (80%Gaussian/20%Lorentzian) Profiles Measured Intensity 1 Normalized Intensity (arb. units) Frequency (cm -1 ) Fig. 3-2 Rayleigh spectrum of the narrowband KrF excimer laser compared to narrowband 80%Gaussian/20%Lorentzian and a superposition of the narrowband 80%Gaussian/20%Lorentzian and broadband 80%Gaussian/20%Lorentzian profiles scaled by the locking efficiency (η = 85%). The narrowband laser bandwidth, ν L = 0.8cm -1, is determined using the convolution method described in Chapter 4 which includes the spectrometer linewidth and slit function. 40

5 (Andresen et al and Wehrmeyer et al. 1992). Therefore, an OH excitation spectrum is obtained to determine the optimum narrowband laser emission frequency, ν L. For the excitation spectrum the spectrograph is centered at cm -1 with a spectral coverage of 1175cm -1 to record the OH fluorescence emission from the (3 1) band, which is also used as the calibration source to characterize the frequency response of the spectrograph and intensified CCD (Charge Coupled Device) detector. A near stoichiometric H 2 -air flame is probed with the laser by tuning the laser, in cm -1 increments, through its locking range. The resulting excitation spectrum is shown in Fig The 3 0 OH transition frequencies pumped by the laser are listed in Table 3-1 along with the associated 3 1 fluorescence frequencies (Diecke and Crosswhite 1962). Tuning the laser to cm -1 ( nm), a frequency close to the peak emission frequency of the KrF excimer, maximizes the laser emission intensity and the Raman scattered signal strengths while minimizing OH fluorescence (shown by the arrow in Fig. 3-3). 3.3 UV Raman Light Collection System A schematic of the experimental UV Raman system is shown in Fig The line integrated Raman scattered light is collected at 90 to the laser beam using a 50 mm diameter f/4 lens and relayed by another lens through a Laser Optik 0 dielectric thin-film notch filter into a 0.5m spectrograph (Acton Research model 500M). The thin-film notch filter is used at the spectrograph entrance to prevent Rayleigh and Mie scattered light from interfering with the Raman signals. An absorption/transmission scan was performed to characterize the frequency response of the thin-film notch filter which is non-uniform over the H 2 Stokes frequency range. A high-efficiency 3600 groove/mm holographic grating (56cm -1 /mm dispersion at cm -1 ) is used to disperse the Raman signals onto an intensified CCD array with a 576 x 384 pixel chip consisting of22µm x 22µm pixels. For the H 2 /N 2 gas mixtures and the H 2 -air flames the photomultiplier gain, or the detected number of counts on the CCD, is ~30counts/e- and ~100counts/e-, respectively. Maximum spectral resolution is achieved without significantly attenuating the Raman signals by setting the spectrograph entrance slit to 37.5µm, which is equivalent to the spatial resolution of one pixel considering the pixel size and intensifier minification ratio (~1.65). The resolution of the spectrograph, ν spect, is 6.75cm -1 (0.05nm) determined from measurements of emission lines from a UV Hg-vapor lamp. Combined with the dispersion of the grating, the spectrograph entrance slit produces and additional rectangular 41

6 1 0.9 Q 1 (11) Q 2 (11) 0.8 Q 2 (10) Intensity (arb units) Q 1 (10) R 1 (14) P2 (7) P 1 (8) P 2 (8) R 2 (15) P 1 (9) Frequency (cm -1 ) Fig. 3-3 OH excitation spectrum showing the (3 0) band of the A 2 Σ - Χ 2 Π transition The spectrograph is set to cm -1. The arrow indicates the optimum operating frequency of the laser for the Raman flame spectra, ν L = cm -1 ( nm). 42

7 Table 3-1. OH absorption and resulting fluorescence lines excited by the KrF laser from Diecke and Crosswhite (1962). (3 0) Absorption Lines (3 1) Fluorescence Lines Designation Frequency (cm -1 ) Trans. Prob. Designation Frequency (cm -1 ) Q 1 (10) 40, R 1 (9) 37, Q 1 (10) 36, P 1 (11) 36, Q 21 (10) 40, R 2 (9) 37, Q 2 (10) 36,826.7 P 2 (11) 36, R 1 (14) 40, R 1 (14) 36, Q 1 (15) 36, P 1 (16) 35, R 21 (14) 40, R 2 (14) 36, Q 2 (15) 36,377.3 P 2 (16) 35, P 2 (7) 40, R 2 (5) 37, Q 2 (6) 37, P 2 (7) 36, Q 12 (10) 40, R 1 (9) 37, Q 1 (10) 36, P 1 (11) 36, Q 2 (10) 40, R 2 (9) 37, Q 2 (10) 36, P 2 (11) 36, R 2 (14) 40, R 2 (14) 36, Q 2 (15) 36,377.3 P 2 (16) 35, P 1 (8) 40, R 1 (6) 37, Q 1 (7) 37, P 1 (8) 36, Q 1 (11) 40, R 1 (10) 37, Q 1 (11) 36, P 1 (12) 36, Q 21 (11) 40, R 2 (10) 37, Q 2 (11) 36,

8 P 2 (12) 36, P 12 (8) 40, R 1 (6) 37, Q 1 (7) 37, P 1 (8) 36, P 2 (8) 40, R 2 (6) 37, Q 2 (7) 37,006.4 P 2 (8) 36, R 1 (15) 40, R 1 (15) 36, Q 1 (16) 36,293 P 1 (17) 35, R 21 (15) 40, R 2 (15) 36, Q 2 (16) 36, P 2 (17) 35, Q 12 (11) 40, R 1 (10) 37, Q 1 (11) 36, P 1 (12) 36, Q 2 (11) 40, R 2 (10) 37, Q 2 (11) 36, P 2 (12) 36, R 2 (15) 40, R 2 (15) 36, Q 2 (16) 36, P 2 (17) 35, P 1 (9) 40, R 1 (7) 37,287.3 Q 1 (8) 37, P 1 (9) 36,

9 TO COMPUTER DIGITAL CAMERA WITH GATED INTENSIFIER BEAM DUMP 0.5 m IMAGING SPECTROGRAPH 50 mm DIA., f/4 SINGLET LENSES 37.5 µm 14 mm TRIGGER FOR LASER AND CAMERA INTENSIFIER 0.17 mm DIA. BEAM WAIST H 2 -AIR FLAME FOCUSING LENS SINGLE-PASS TUNABLE KRF EXCIMER LASER Fig Schematic of experimental UV Raman system. 45

10 lineshape function, ν slit, of ~2.1cm -1. The combined resolution of the spectrograph/ccd system, ν spect-slit, with a slit width of 37.5µm is ~7.1cm -1 (~0.055nm) as determined from measurements of emission lines from a UV Hg-vapor lamp. To increase the Raman scattered signal, the 5mm x 22mm laser beam is focused to a ~0.17mm x 0.75mm profile in the sample volume with the smallest dimension aligned with the spectrograph entrance slit width. The probe length or spatial resolution of the Raman system, as determined by the magnification ratio of the collection system (1:1) and the CCD intensifier minification ratio, is 8.64mm in the direction of beam propagation. The extremely narrow entrance slit width required to provide the necessary spectral resolution does limit the amount of light striking the CCD per laser pulse; however, integration of the Raman scattered light from the 8.64mm probe length significantly increases the total measured signal strengths and allows for statistically reliable single-pulse and time-averaged measurements. 3.4 Detector Response The intensified CCD camera is an excellent spectroscopy tool. However, a significant problem arises when trying to assign each camera pixel a specific frequency because the resolution of the spectrograph is not adequate to fully disperse the signal and the 22µm pixel width is large compared to the sharp atomic or fluorescence lines used for calibration. Typically, the peak centers of intensity of the 32 OH (3 1) fluorescence lines used to calibrate the spectrograph/ccd are between 1-2 pixels in width, as illustrated in Fig Thus for calibration of the spectrograph/ccd, the peak centers of the OH fluorescence lines between cm -1 are fitted using Lorentzian profiles with weighted line centers about the mean pixel location and a linear regression analysis to generate a 6 th order polynomial calibration curve. The error in the calibration fit for any one frequency measurement is equivalent to the variance of the fit, σ 2 = ±0.3cm -1, which is ~4µm on the image or about one fifth of a pixel (Tellinghuisen 1993). Spatial smearing of the experimental spectra occurs because of the finite size of the pixels, geometric response of the CCD array, and from the microchannel plate intensifier. These effects along with the focal length of the spectrograph and dispersive capability of the grating limit the resolving power of the collection system. Thus to provide the maximum spectral resolution without significantly attenuating the Raman signals, the spectrograph entrance slit is 46

11 1 OH Calibration Normalized Intensity (arb. units) Pixel # Fig A 40 pixel region of a typical (3 1) OH fluorescence spectrum used for calibration of the spectrograph/ccd. The OH fluorescence line illustrated is Q 2 (15) at cm -1 resulting from laser excitation of R 2 (14) at cm -1. The frequency scale is ~2cm -1 per pixel. 47

12 set to 37.5µm, which is equivalent to the spatial resolution of one pixel based on the pixel size and intensifier minification ratio. The spectrograph instrumental linewidth, including the slit function and the smearing on the detector, is measured by recording the lineshape of an isolated and intense 3 1 OH fluorescence line in the spectral region of interest as shown in Fig Unlike the typical fluorescence spectrum shown in Fig. 3-5, the selected OH spectra for the linewidth calibration have well defined peak centers that are sharp (<0.025cm -1 or less than the 6.75cm -1 resolving power of the spectrograph) and contained within a few pixels. Fluorescence spectral shapes are typically dominated by radiation damping or predissociative processes and for instruments with moderate to weak spectral resolution are adequately described by Lorentzian profiles. The OH spectrum in Fig. 3-6, upon de- convolution of the slit function and laser bandwidth as described by methods in Chapter Four, confirms the measured spectrograph instrumental linewidth, ν spect, of 6.75cm Burners Temperature measurements are taken in dilute H 2 -N 2 mixtures issuing from a gas heater jet (Sylvania Series III model , 12.5mm diameter) and in atmospheric pressure H 2 -air flames of various rich equivalence ratios produced in an adiabatic Hencken burner (12.5mm diameter). The contour fit Raman derived temperatures are compared to measurements made with Pt6%Rhodium-Pt13%Rhodium bare uncoated thermocouples. Radiation corrections are made using methods suggested by Kaskan (1957). Fuel- and air-flow rates for the conditions probed are listed in Table 3-2 and are measured with Teledyne-Hastings hot-wire mass flow meters. The flow meters have accuracies of ±1% of their full scale. Based on calibrations using laminar flow elements, the flow rates used in the experiments are accurate to within ±1%, and the predicted adiabatic flame temperatures, based on the combined relative error of the mass flow meters, are accurate to within ±1.5%. 48

13 Normalized Intensity (arb. units) OH Fluorescence Linewidth Gaussian Lorentzian Measured Intensity Frequency (cm -1 ) Fig Measurement of the instrumental linewidth of the spectrograph/ccd using the Q 2 (16) 3 1 OH fluorescence line resulting from excitation of the R 2 (15) 3 0 OH absorption line. The spectrograph instrumental linewidth, ν spect, is 6.75cm -1 after de-convolution of the slit function and laser bandwidth. 49

14 Table 3-2. Flow-rates for the laboratory Raman measurements in the H 2 /N 2 mixtures and H 2 -air flames. Flow Rates (L/min) T adiabatic T thermocouple Mixture H 2 N 2 Air (K) (K) H 2 /N H 2 /N H 2 -air H 2 -air

15 3.6 Marshall Space Flight Center Tests The configuration of the Raman system for the hot-fire tests of the sub-scale rocketengine like test article at NASA Marshall Space Flight Center Test Stand 115 is very similar to the laboratory system previously described. An identical Lambda-Physik COMPex 150T narrowband KrF excimer laser is used to induce the H 2 Stokes Raman scattered signals. However, the laser is operated in the standard configuration and not in single-pass mode to maintain minimal divergence of the laser beam (~0.2mrad for the conventional mode vs. ~1mrad for the single-pass mode). The better coherence is required to transmit the beam over the long distance from the instrumentation building to the test article without considerable beam expansion. This allows for tighter focusing of the beam in the sample volume, thus maintaining a high energy density for Raman scattering activity. The narrowband/broadband linewidth parameters and locking efficiency are comparable to those in the single-pass mode. For the hotfire test, the laser is operated at 10Hz and a pulse energy of ~425mJ at the exit plane of the laser cavity with only ~225mJ reaching the test article due to losses from optical components along the path length. Also to prevent damaging the test article windows, the laser beam is focused with a 2m lens. The Raman scattered light is collected with a 50mm diameter f/5 air-spaced doublet, used for its excellent spatial imaging properties, and relayed by an identical lens through the same Laser Optik 0 dielectric thin-film notch filter and Acton Research model 500M 0.5m spectrograph equipped with the 3600 groove/mm holographic grating and the same ICCD as used in the laboratory experiments. The test article for the hot-fire test consists of an injector plate assembly, GH 2 /GO 2 hotgas torch igniter, choked flow convergent-divergent section, rapid expansion nozzle plate, and several spacers, thermocouple sections, calorimeter rack, or annular windowed sections. Schematics of this modular combustion test article (MCTA) showing the thermocouple, calorimeter, and shadowgraphy/raman test configurations are illustrated in Fig The combustion chamber of the MCTA is 10cm in diameter and 48cm in length. The 7 element coaxial injector is shown in Fig Each injector element is of co-annular design consisting of an inner flow of LO X shrouded by LH 2. Table 3-3 lists the test conditions for the hot-fire tests of the MCTA. Figure 3-9 shows a hot-fire test of the MCTA with a windowed section aft of the injector and torch igniter for shadowgraph imaging. Shadowgraphy measurements are performed at the injector face and at several downstream axial positions to analyze injector spray 51

16 performance with thermocouple and calorimeter measurements to obtain combustion temperatures and thermal efficiency for comparison with computational fluid dynamic (CFD) calculations. Raman temperature measurements, for validation of the contour-fit temperature measurement technique for practical combustion devices and comparison with thermocouple measurements and CFD predictions, are performed in the combustion chamber ~40cm downstream of the injector face to prevent liquid-phase interferences and to ensure complete mixing of the individual injector element spray cores and chemical/thermal equilibrium. 52

17 Co-Axial 7 element Injector Thermocouple Racks Nozzle Plate GH 2 /GO 2 Torch Igniter Calorimeter Choked Flow Section Windowed Section Fig. 3-7 Cross-section schematic of the modular combustion test article (MCTA) in various test configurations. The top schematic shows the MCTA with the thermocouple rack installed for spatial temperature measurements, the middle schematic shows the calorimeter installed for combustion efficiency measurements, and the bottom schematic shows the windowed section installed for shadowgraphy and Raman measurements. 53

18 Fig. 3-8 View of the 7 element co-axial injector looking through the exhaust nozzle down the combustion chamber. The injector is composed of two annular flows, the inner injector post supplies LO X, while the outer shroud supplies LH 2. 54

19 Table 3-3. MCTA hot-fire test conditions. a Combustion Chamber Conditions Nozzle a Ignition c Mainstage 1 d Mainstage 2 d Exhaust Plume MR b Total Flow Rate (lbm/s) P c (psig) >>14.7 T c (K) ρ (g/cc) V (m/s) - ~15 ~15 ~300 N H2 (molecules/cm 3 ) - ~1 x ~1 x ~2 x [H 2 ] (%) - ~0.9 ~0.9 - Laser Fluence (GW/cm 2 ) - ~1 ~1 ~5 Raman Signal Gain e Choked flow at the throat exit. Listed values are approximations 10cm downstream of the throat in the rapidly expanding exhaust plume. b MR is the oxidizer to fuel mass ratio. Stoichiometric is 8:1. c The chamber conditions for ignitions. d During the test sequence the MCTA is operated at two different mainstage combustion conditions. e Raman signal gain based on equilibrium calculations of in-chamber combustion conditions and compared with laboratory experiments of N H2 = , T = 293K, and a laser fluence of ~10 (GW/cm2). For a 2%/98% H 2 /N 2 mixture at P = 1800psi and a N H2 = with ~0.1 (Gw/cm2) laser fluence the expected gain in Raman signal is ~12. LH 2 inlet T = ~40K LO x inlet T = ~110K Test Duration = ~20-42s 55

20 Fig. 3-9 Hot-fire test of the MCTA. Windowed section is installed just aft of the injector to provide shadowgraphy measurements during injection. 56

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006 Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006 1. Examples of advantages and disadvantages with laser-based combustion diagnostic techniques: + Nonintrusive + High

More information

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Section I Q1. Answer (i) (b) (ii) (d) (iii) (c) (iv) (c) (v) (a) (vi) (b) (vii) (b) (viii) (a) (ix)

More information

Raman spectroscopy of molecule densities in Hydrogen-Air mixture premixed gas ignited by spark discharge

Raman spectroscopy of molecule densities in Hydrogen-Air mixture premixed gas ignited by spark discharge Proc. 2012 Joint Electrostatics Conference 1 Raman spectroscopy of molecule densities in Hydrogen-Air mixture premixed gas ignited by spark discharge Ayumi Kumamoto, Ryo Ono, Tetsuji Oda Dept. of Advanced

More information

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source 3rd International EUVL Symposium NOVEMBER 1-4, 2004 Miyazaki, Japan Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source H. Tanaka, A. Matsumoto, K. Akinaga, A. Takahashi

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION doi:10.1038/nature10721 Experimental Methods The experiment was performed at the AMO scientific instrument 31 at the LCLS XFEL at the SLAC National Accelerator Laboratory. The nominal electron bunch charge

More information

25 Instruments for Optical Spectrometry

25 Instruments for Optical Spectrometry 25 Instruments for Optical Spectrometry 25A INSTRUMENT COMPONENTS (1) source of radiant energy (2) wavelength selector (3) sample container (4) detector (5) signal processor and readout (a) (b) (c) Fig.

More information

Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes

Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes M. W. Teague*, Tonya Felix*, M. K. Hudson, and R. Shanks *Department of Chemistry, Hendrix College, Conway, AR

More information

Atomization. In Flame Emission

Atomization. In Flame Emission FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic by its monatomic particles in gaseous state in the

More information

APPLICATION OF SPONTANEOUS RAMAN SCATTERING TO THE FLOWFIELD IN A SCRAMJET COMBUSTOR T. Sander and T. Sattelmayer Lehrstuhl für Thermodynamik,

APPLICATION OF SPONTANEOUS RAMAN SCATTERING TO THE FLOWFIELD IN A SCRAMJET COMBUSTOR T. Sander and T. Sattelmayer Lehrstuhl für Thermodynamik, APPLICATION OF SPONTANEOUS RAMAN SCATTERING TO THE FLOWFIELD IN A SCRAMJET COMBUSTOR T. Sander and T. Sattelmayer Lehrstuhl für Thermodynamik, TU-München, D-85747, Garching, Germany Introduction The weight

More information

Introduction to laser-based combustion diagnostics

Introduction to laser-based combustion diagnostics Introduction to laser-based combustion diagnostics (Lecture 1b) Lecture prepared for course in laser-based combustion diagnostics by Per-Erik Bengtsson and Joakim Bood Division of Combustion Physics at

More information

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency. Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light

More information

Fluorescence tracer technique for simultaneous temperature and equivalence ratio measurements in Diesel jets

Fluorescence tracer technique for simultaneous temperature and equivalence ratio measurements in Diesel jets Renewable energies Eco-friendly production Innovative transport Eco-efficient processes Sustainable resources Fluorescence tracer technique for simultaneous temperature and equivalence ratio measurements

More information

Abstract... I. Acknowledgements... III. Table of Content... V. List of Tables... VIII. List of Figures... IX

Abstract... I. Acknowledgements... III. Table of Content... V. List of Tables... VIII. List of Figures... IX Abstract... I Acknowledgements... III Table of Content... V List of Tables... VIII List of Figures... IX Chapter One IR-VUV Photoionization Spectroscopy 1.1 Introduction... 1 1.2 Vacuum-Ultraviolet-Ionization

More information

requency generation spectroscopy Rahul N

requency generation spectroscopy Rahul N requency generation spectroscopy Rahul N 2-11-2013 Sum frequency generation spectroscopy Sum frequency generation spectroscopy (SFG) is a technique used to analyze surfaces and interfaces. SFG was first

More information

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber CH217 Fundamentals of Analytical Chemistry Module Leader: Dr. Alison Willows Electromagnetic spectrum Properties of electromagnetic radiation Many properties of electromagnetic radiation can be described

More information

Fundamental Mechanisms, Predictive Modeling, and Novel Aerospace Applications of Plasma Assisted Combustion

Fundamental Mechanisms, Predictive Modeling, and Novel Aerospace Applications of Plasma Assisted Combustion Fundamental Mechanisms, Predictive Modeling, and Novel Aerospace Applications of Plasma Assisted Combustion Walter R. Lempert, Igor V. Adamovich, J. William Rich, Jeffrey A. Sutton Department of Mechanical

More information

X-Rays From Laser Plasmas

X-Rays From Laser Plasmas X-Rays From Laser Plasmas Generation and Applications I. C. E. TURCU CLRC Rutherford Appleton Laboratory, UK and J. B. DANCE JOHN WILEY & SONS Chichester New York Weinheim Brisbane Singapore Toronto Contents

More information

Inhomogeneous Mixing Behavior of Recirculated Exhaust Gas in a Lean Premixed Flame

Inhomogeneous Mixing Behavior of Recirculated Exhaust Gas in a Lean Premixed Flame Inhomogeneous Mixing Behavior of Recirculated Exhaust Gas in a Lean Premixed Flame 2nd Japan-China Joint Seminar July 11, 2016, Gifu University, Japan Masaharu Komiyama Department of Mechanical Engineering

More information

Resonance Raman measurements utilizing a deep UV source

Resonance Raman measurements utilizing a deep UV source Resonance Raman measurements utilizing a deep UV source Adam Willitsford a, C. Todd Chadwick b, Hans Hallen b, and C. Russell Philbrick a a The Pennsylvania State University, Department of Electrical Engineering,

More information

FLAME PHOTOMETRY AIM INTRODUCTION

FLAME PHOTOMETRY AIM INTRODUCTION FLAME PHOTOMETRY AIM INTRODUCTION Atomic spectroscopy is based on the absorption, emission or fluorescence process of light by atoms or elementary ions. Information for atomic scale is obtained in two

More information

LASER BASED DIAGNOSTIC SYSTEM FOR SPRAY MEASUREMENTS

LASER BASED DIAGNOSTIC SYSTEM FOR SPRAY MEASUREMENTS Journal of KONES Powertrain and Transport, Vol. 22, No. 3 2015 LASER BASED DIAGNOSTIC SYSTEM FOR SPRAY MEASUREMENTS Łukasz Jan Kapusta, Piotr Jaworski, Andrzej Teodorczyk Warsaw University of Technology,

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES Chemistry 524--Final Exam--Keiderling May 4, 2011 3:30 -?? pm -- 4286 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted. No open books or

More information

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at 2001 Spectrometers Instrument Machinery Movies from this presentation can be access at http://www.shsu.edu/~chm_tgc/sounds/sound.html Chp20: 1 Optical Instruments Instrument Components Components of various

More information

Comments to Atkins: Physical chemistry, 7th edition.

Comments to Atkins: Physical chemistry, 7th edition. Comments to Atkins: Physical chemistry, 7th edition. Chapter 16: p. 483, Eq. (16.1). The definition that the wave number is the inverse of the wave length should be used. That is much smarter. p. 483-484.

More information

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Bedanta Kr. Deka, D. Thakuria, H. Bora and S. Banerjee # Department of Physicis, B. Borooah College, Ulubari,

More information

Determination of bandwidth and beamwidth of a Rhodamine 6G dye laser using different optical setups

Determination of bandwidth and beamwidth of a Rhodamine 6G dye laser using different optical setups Practicum Spectroscopy HS 2 Determination of bandwidth and beamwidth of a Rhodamine 6G dye laser using different optical setups DLA Jorge Ferreiro, study degree Chemistry, 5 th semester, fjorge@student.ethz.ch

More information

Development of optimized Raman Spectroscopy setup for species detection in flames

Development of optimized Raman Spectroscopy setup for species detection in flames Development of optimized Raman Spectroscopy setup for species detection in flames Henrik Johansson Thesis submitted for the degree of Bachelor of Science Project duration: 2 months Supervised by Christian

More information

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept. Spectrum of Electromagnetic Radiation Electromagnetic radiation is light. Different energy light interacts with different motions in molecules. CHEM*344 Chemical Instrumentation Topic 7 Spectrometry Radiofrequency

More information

novaa 800 D Atomic Absorption Spectrometer

novaa 800 D Atomic Absorption Spectrometer Technical Data Atomic Absorption Spectrometer Cpt : +27 (0) 21 905 0476 Jhb : +27 (0) 11 794 Dbn : +27 (0) 31 266 2454 1/7 General The is a compact atomic absorption spectrometer with deuterium background

More information

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT.

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT. CHEM 322 Name Exam 3 Spring 2013 Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT. Warm-up (3 points each). 1. In Raman Spectroscopy, molecules are promoted

More information

A few Experimental methods for optical spectroscopy Classical methods Modern methods. Remember class #1 Generating fast LASER pulses

A few Experimental methods for optical spectroscopy Classical methods Modern methods. Remember class #1 Generating fast LASER pulses A few Experimental methods for optical spectroscopy Classical methods Modern methods Shorter class Remember class #1 Generating fast LASER pulses, 2017 Uwe Burghaus, Fargo, ND, USA W. Demtröder, Laser

More information

Diagnósticos em Plasmas

Diagnósticos em Plasmas Tecnologia a Plasma para o Processamento de Materiais Diagnósticos em Plasmas Diagnósticos Ópticos João Santos Sousa, nº50901 Semestre Inverno 2004/2005 21 de Janeiro de 2005, 9h-10h, sala F8 Contents

More information

Richard Miles and Arthur Dogariu. Mechanical and Aerospace Engineering Princeton University, Princeton, NJ 08540, USA

Richard Miles and Arthur Dogariu. Mechanical and Aerospace Engineering Princeton University, Princeton, NJ 08540, USA Richard Miles and Arthur Dogariu Mechanical and Aerospace Engineering Princeton University, Princeton, NJ 08540, USA Workshop on Oxygen Plasma Kinetics Sept 20, 2016 Financial support: ONR and MetroLaser

More information

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES Chemistry 524--Final Exam--Keiderling Dec. 12, 2002 --4-8 pm -- 238 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted plus one 8.5 x 11 sheet

More information

Chapter 24 Photonics Question 1 Question 2 Question 3 Question 4 Question 5

Chapter 24 Photonics Question 1 Question 2 Question 3 Question 4 Question 5 Chapter 24 Photonics Data throughout this chapter: e = 1.6 10 19 C; h = 6.63 10 34 Js (or 4.14 10 15 ev s); m e = 9.1 10 31 kg; c = 3.0 10 8 m s 1 Question 1 Visible light has a range of photons with wavelengths

More information

Electromagnetic Radiation and Scientific Instruments. PTYS April 1, 2008

Electromagnetic Radiation and Scientific Instruments. PTYS April 1, 2008 Electromagnetic Radiation and Scientific Instruments PTYS 206-2 April 1, 2008 Announcements Deep Impact 6 PM Wednesday Night Pizza, no beer Watch at home if you can t watch here. It will be discussed in

More information

high temp ( K) Chapter 20: Atomic Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment

More information

EUV lithography and Source Technology

EUV lithography and Source Technology EUV lithography and Source Technology History and Present Akira Endo Hilase Project 22. September 2017 EXTATIC, Prague Optical wavelength and EUV (Extreme Ultraviolet) VIS 13.5nm 92eV Characteristics of

More information

Two-dimensional concentration and temperature measurements in extended flames of industrial burners using PLIF

Two-dimensional concentration and temperature measurements in extended flames of industrial burners using PLIF Two-dimensional concentration and temperature measurements in extended flames of industrial burners using PLIF D. Müller a*, W. Triebel a, A. Bochmann a, G. Schmidl a, D. Eckardt a, A. Burkert a, J. Röper

More information

Skoog Chapter 6 Introduction to Spectrometric Methods

Skoog Chapter 6 Introduction to Spectrometric Methods Skoog Chapter 6 Introduction to Spectrometric Methods General Properties of Electromagnetic Radiation (EM) Wave Properties of EM Quantum Mechanical Properties of EM Quantitative Aspects of Spectrochemical

More information

LIBSlab ANALYZERS ANALYZERS

LIBSlab ANALYZERS ANALYZERS ANALYZERS ANALYZERS Chemical multi-elemental analysis with LIBS in modular benchtop design LIBSlab LIBSpector compact sample chamber for the LIBS analysis of solid, liquid and gaseous samples. Sample chamber

More information

Characterisation & Use of Array Spectrometers

Characterisation & Use of Array Spectrometers Characterisation & Use of Array Spectrometers Mike Shaw, Optical Technologies & Scientific Computing Team, National Physical Laboratory, Teddington Middlesex, UK 1 Overview Basic design and features of

More information

PHY410 Optics Exam #3

PHY410 Optics Exam #3 PHY410 Optics Exam #3 NAME: 1 2 Multiple Choice Section - 5 pts each 1. A continuous He-Ne laser beam (632.8 nm) is chopped, using a spinning aperture, into 500 nanosecond pulses. Compute the resultant

More information

Wolfgang Demtroder. Laser Spectroscopy. Basic Concepts and Instrumentation. Second Enlarged Edition With 644 Figures and 91 Problems.

Wolfgang Demtroder. Laser Spectroscopy. Basic Concepts and Instrumentation. Second Enlarged Edition With 644 Figures and 91 Problems. Wolfgang Demtroder Laser Spectroscopy Basic Concepts and Instrumentation Second Enlarged Edition With 644 Figures and 91 Problems Springer Contents 1. Introduction 1 2. Absorption and Emission of Light

More information

Microwave Enhanced Combustion and New Methods for Combustion Diagnostics

Microwave Enhanced Combustion and New Methods for Combustion Diagnostics Microwave Enhanced Combustion and New Methods for Combustion Diagnostics Richard Miles, Michael Shneider, Sohail Zaidi,Arthur D ogariu James Michael, Tat Loon Chng, Chris Limbach Mathew Edw ards 2011 Plasm

More information

Advanced Spectroscopy Laboratory

Advanced Spectroscopy Laboratory Advanced Spectroscopy Laboratory - Raman Spectroscopy - Emission Spectroscopy - Absorption Spectroscopy - Raman Microscopy - Hyperspectral Imaging Spectroscopy FERGIELAB TM Raman Spectroscopy Absorption

More information

Problem Solving. radians. 180 radians Stars & Elementary Astrophysics: Introduction Press F1 for Help 41. f s. picture. equation.

Problem Solving. radians. 180 radians Stars & Elementary Astrophysics: Introduction Press F1 for Help 41. f s. picture. equation. Problem Solving picture θ f = 10 m s =1 cm equation rearrange numbers with units θ factors to change units s θ = = f sinθ fθ = s / cm 10 m f 1 m 100 cm check dimensions 1 3 π 180 radians = 10 60 arcmin

More information

Excimer Lasers Currently best UV laser sources Consist two atom types which repel each other eg nobel gas and halide or oxide which normally do not

Excimer Lasers Currently best UV laser sources Consist two atom types which repel each other eg nobel gas and halide or oxide which normally do not Excimer Lasers Currently best UV laser sources Consist two atom types which repel each other eg nobel gas and halide or oxide which normally do not bond But when excited/ionized these atoms attract Bound

More information

Introduction to FT-IR Spectroscopy

Introduction to FT-IR Spectroscopy Introduction to FT-IR Spectroscopy An FT-IR Spectrometer is an instrument which acquires broadband NIR to FIR spectra. Unlike a dispersive instrument, i.e. grating monochromator or spectrograph, an FT-IR

More information

Laser Physics OXFORD UNIVERSITY PRESS SIMON HOOKER COLIN WEBB. and. Department of Physics, University of Oxford

Laser Physics OXFORD UNIVERSITY PRESS SIMON HOOKER COLIN WEBB. and. Department of Physics, University of Oxford Laser Physics SIMON HOOKER and COLIN WEBB Department of Physics, University of Oxford OXFORD UNIVERSITY PRESS Contents 1 Introduction 1.1 The laser 1.2 Electromagnetic radiation in a closed cavity 1.2.1

More information

Lecture 0. NC State University

Lecture 0. NC State University Chemistry 736 Lecture 0 Overview NC State University Overview of Spectroscopy Electronic states and energies Transitions between states Absorption and emission Electronic spectroscopy Instrumentation Concepts

More information

Atomic Emission Spectroscopy

Atomic Emission Spectroscopy Atomic Emission Spectroscopy Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

NTUA. Physics Department, National Technical University of Athens. June IPRD10,, Siena, Italy S. Maltezos SIENA ITALY

NTUA. Physics Department, National Technical University of Athens. June IPRD10,, Siena, Italy S. Maltezos SIENA ITALY NTUA SIENA ITALY ATMOSPHERIC MONITORING FOR VERY HIGH ENERGY GAMMA RAY CHERENKOV TELESCOPES BASED ON HSRL: DEVELOPMENT OF HIGH ACCURACY NON-INVASIVE ETALON CHARACTERIZATION TECHNIQUES S. Maltezos,, E.

More information

Plasma Spectroscopy in ISTTOK

Plasma Spectroscopy in ISTTOK Plasma Spectroscopy in ISTTOK J. Figueiredo 1, R. B. Gomes 1, T. Pereira 1, H. Fernandes 1, A. Sharakovski 2 1 Associação EURATOM/IST, Centro de Fusão Nuclear, IST, 1049-001 Lisboa, Portugal 2 Association

More information

Single Emitter Detection with Fluorescence and Extinction Spectroscopy

Single Emitter Detection with Fluorescence and Extinction Spectroscopy Single Emitter Detection with Fluorescence and Extinction Spectroscopy Michael Krall Elements of Nanophotonics Associated Seminar Recent Progress in Nanooptics & Photonics May 07, 2009 Outline Single molecule

More information

2101 Atomic Spectroscopy

2101 Atomic Spectroscopy 2101 Atomic Spectroscopy Atomic identification Atomic spectroscopy refers to the absorption and emission of ultraviolet to visible light by atoms and monoatomic ions. It is best used to analyze metals.

More information

Survey on Laser Spectroscopic Techniques for Condensed Matter

Survey on Laser Spectroscopic Techniques for Condensed Matter Survey on Laser Spectroscopic Techniques for Condensed Matter Coherent Radiation Sources for Small Laboratories CW: Tunability: IR Visible Linewidth: 1 Hz Power: μw 10W Pulsed: Tunabality: THz Soft X-ray

More information

Stimulated Raman scattering of XeCl 70 ns laser pulses in silica fibres

Stimulated Raman scattering of XeCl 70 ns laser pulses in silica fibres J. Opt. A: Pure Appl. Opt. 1 (1999) 725 729. Printed in the UK PII: S1464-4258(99)00367-0 Stimulated Raman scattering of XeCl 70 ns laser pulses in silica fibres Nikolai Minkovski, Ivan Divliansky, Ivan

More information

Lifecycle of laser-produced air sparks

Lifecycle of laser-produced air sparks Lifecycle of laser-produced air sparks S. S. Harilal, 1 B. E. Brumfield, and M. C. Phillips Pacific Northwest National Laboratory, P.O. Box 999, Richland, Washington 99352, USA We investigated the lifecycle

More information

Fabry-Perot Interferometer for atmospheric monitoring useful for EAS detection E.Fokitis 1, K. Patrinos 1, Z. Nikitaki 1

Fabry-Perot Interferometer for atmospheric monitoring useful for EAS detection E.Fokitis 1, K. Patrinos 1, Z. Nikitaki 1 Fabry-Perot Interferometer for atmospheric monitoring useful for EAS detection E.Fokitis 1, K. Patrinos 1, Z. Nikitaki 1 ABSTRACT A piezotunable Fabry-Perot interferometer is studied as a candidate Doppler

More information

AN INTRODUCTION TO ATOMIC SPECTROSCOPY

AN INTRODUCTION TO ATOMIC SPECTROSCOPY AN INTRODUCTION TO ATOMIC SPECTROSCOPY Atomic spectroscopy deals with the absorption, emission, or fluorescence by atom or elementary ions. Two regions of the spectrum yield atomic information- the UV-visible

More information

Ultra-narrow-band tunable laserline notch filter

Ultra-narrow-band tunable laserline notch filter Appl Phys B (2009) 95: 597 601 DOI 10.1007/s00340-009-3447-6 Ultra-narrow-band tunable laserline notch filter C. Moser F. Havermeyer Received: 5 December 2008 / Revised version: 2 February 2009 / Published

More information

Astronomy 203 practice final examination

Astronomy 203 practice final examination Astronomy 203 practice final examination Fall 1999 If this were a real, in-class examination, you would be reminded here of the exam rules, which are as follows: You may consult only one page of formulas

More information

Spectroscopy Problem Set February 22, 2018

Spectroscopy Problem Set February 22, 2018 Spectroscopy Problem Set February, 018 4 3 5 1 6 7 8 1. In the diagram above which of the following represent vibrational relaxations? 1. Which of the following represent an absorbance? 3. Which of following

More information

Atomic Absorption Spectrometer ZEEnit P series

Atomic Absorption Spectrometer ZEEnit P series Atomic Absorption Spectrometer ZEEnit P series Technical Data ZEEnit series Update 07/2014 OBue 1/ 5 ZEEnit P series Variable high-end AA Spectrometer with Deuterium and Zeeman Background Correction with

More information

Application of IR Raman Spectroscopy

Application of IR Raman Spectroscopy Application of IR Raman Spectroscopy 3 IR regions Structure and Functional Group Absorption IR Reflection IR Photoacoustic IR IR Emission Micro 10-1 Mid-IR Mid-IR absorption Samples Placed in cell (salt)

More information

Diffuse reflection BBSFG optical layout

Diffuse reflection BBSFG optical layout Diffuse reflection BBSFG optical layout Figure 1 shows the optical layout of the broad bandwidth sum frequency generation (BBSFG) system. A Nd:YVO 4 laser (a, Spectra-Physics MillenniaVs) pumps the Ti:Sapphire

More information

Lecture 10. Lidar Effective Cross-Section vs. Convolution

Lecture 10. Lidar Effective Cross-Section vs. Convolution Lecture 10. Lidar Effective Cross-Section vs. Convolution q Introduction q Convolution in Lineshape Determination -- Voigt Lineshape (Lorentzian Gaussian) q Effective Cross Section for Single Isotope --

More information

High-Spectral-Resolution Two-photon Pump Polarization Spectroscopy Probe (TPP-PSP) Technique for Measurements of Atomic Hydrogen

High-Spectral-Resolution Two-photon Pump Polarization Spectroscopy Probe (TPP-PSP) Technique for Measurements of Atomic Hydrogen High-Spectral-Resolution Two-photon Pump Polarization Spectroscopy Probe (TPP-PSP) Technique for Measurements of Atomic Hydrogen Aman Satija, Aizaz H. Bhuiyan and Robert P. Lucht. School of Mechanical

More information

In this paper a LIF method is used in which the effect of quenching is avoided simply by exciting the

In this paper a LIF method is used in which the effect of quenching is avoided simply by exciting the Laser-induced fluorescence with tunable excimer lasers as a possible method for instantaneous temperature field measurements at high pressures: checks with an atmospheric flame Peter Andresen, A. Bath,

More information

Planar Laser-Induced Iodine Fluorescence Technique for Flow Visualization and Quantitative Measurements in Rarefied Flows

Planar Laser-Induced Iodine Fluorescence Technique for Flow Visualization and Quantitative Measurements in Rarefied Flows Planar Laser-Induced Iodine Fluorescence Technique for Flow Visualization and Quantitative Measurements in Rarefied Flows Professor James McDaniel*, Eric Cecil*, Erin Reed* and Josh Codoni* Professor Iain

More information

n ( λ ) is observed. Further, the bandgap of the ZnTe semiconductor is

n ( λ ) is observed. Further, the bandgap of the ZnTe semiconductor is Optical Spectroscopy Lennon O Naraigh, 0000 Date of Submission: 0 th May 004 Abstract: This experiment is an exercise in the principles and practice of optical spectroscopy. The continuous emission spectrum

More information

Use of a High-Resolution Overview Spectrometer for the Visible Range in the TEXTOR Boundary Plasma

Use of a High-Resolution Overview Spectrometer for the Visible Range in the TEXTOR Boundary Plasma Use of a High-Resolution Overview Spectrometer for the Visible Range in the TEXTOR Boundary Plasma Sebastijan BREZINSEK, Albrecht POSPIESZCZYK, Gennadij SERGIENKO, Philippe MERTENS and Ulrich SAMM Institut

More information

of a molecule possessing a vibrational energy hν is

of a molecule possessing a vibrational energy hν is Question 1 (a) A pulse N:YAG laser is to be employe in a Rayleigh scattering experiment to etermine gas temperature. The laser can be use at 532 nm (secon harmonic), 355 nm (thir harmonic), or 266 nm (fourth

More information

High Resolution Optical Spectroscopy

High Resolution Optical Spectroscopy PHYS 3719 High Resolution Optical Spectroscopy Introduction This experiment will allow you to learn a specific optical technique with applications over a wide variety of phenomena. You will use a commercial

More information

Chem 155 Midterm Exam Page 1 of 10 Spring 2010 Terrill

Chem 155 Midterm Exam Page 1 of 10 Spring 2010 Terrill Chem 155 Midterm Exam Page 1 of 10 ame Signature 1. Mercury (Hg) is is believed to be hazardous to human neurological health at extremely low concentrations. Fortunately EPA Method 45.7 cold vapor atomic

More information

Design of a High Spectral Resolution Lidar for Atmospheric Monitoring in EAS Detection Experiments

Design of a High Spectral Resolution Lidar for Atmospheric Monitoring in EAS Detection Experiments Nuclear Physics B (Proc. Suppl.) 190 (2009) 261 265 www.elsevierphysics.com Design of a High Spectral Resolution Lidar for Atmospheric Monitoring in EAS Detection Experiments E. Fokitis a, P. Fetfatzis

More information

AS 101: Day Lab #2 Summer Spectroscopy

AS 101: Day Lab #2 Summer Spectroscopy Spectroscopy Goals To see light dispersed into its constituent colors To study how temperature, light intensity, and light color are related To see spectral lines from different elements in emission and

More information

Remote Raman & Fluorescence Capabilities for Chemical Detection at University of Hawaii

Remote Raman & Fluorescence Capabilities for Chemical Detection at University of Hawaii Remote Raman & Fluorescence Capabilities for Chemical Detection at University of Hawaii Anupam Misra Shiv Sharma HIGP, SOEST, Univ. of Hawaii * Daytime Operation * Long Detection Range (no sample collection)

More information

ON THE ACCURACY OF SCALAR DISSIPATION MEASUREMENTS BY LASER RAYLEIGH SCATERING.

ON THE ACCURACY OF SCALAR DISSIPATION MEASUREMENTS BY LASER RAYLEIGH SCATERING. ON THE ACCURACY OF SCALAR DISSIPATION MEASUREMENTS BY LASER RAYLEIGH SCATERING. P.Ferrão, M.V Heitor and R. Salles Instituto Superior Técnico Mechanical Engineering Department Technical University of Lisbon

More information

Chem Homework Set Answers

Chem Homework Set Answers Chem 310 th 4 Homework Set Answers 1. Cyclohexanone has a strong infrared absorption peak at a wavelength of 5.86 µm. (a) Convert the wavelength to wavenumber.!6!1 8* = 1/8 = (1/5.86 µm)(1 µm/10 m)(1 m/100

More information

Air Force Research Laboratory

Air Force Research Laboratory Air Force Research Laboratory Air Force Research Laboratory Plasma Excited Oxygen Effects on Combustion and Perspectives on Applications to High-Speed Propulsion Date: 10 November 2011 Integrity Service

More information

First observations of the second solar spectrum with spatial resolution at the Lunette Jean Rösch

First observations of the second solar spectrum with spatial resolution at the Lunette Jean Rösch First observations of the second solar spectrum with spatial resolution at the Lunette Jean Rösch Malherbe, J.-M., Moity, J., Arnaud, J., Roudier, Th., July 2006 The experiment setup in spectroscopic mode

More information

Lecture 28. Aerosol Lidar (4) HSRL for Aerosol Measurements

Lecture 28. Aerosol Lidar (4) HSRL for Aerosol Measurements Lecture 28. Aerosol Lidar (4) HSRL for Aerosol Measurements Review of single- and multi-channel aerosol lidars Principle of High Spectral Resolution Lidar (HSRL) HSRL instrumentation University of Wisconsin

More information

Stand-off species detection through picosecond Raman LIDAR with temporal fluorescence suppression and telescope characterisation

Stand-off species detection through picosecond Raman LIDAR with temporal fluorescence suppression and telescope characterisation May 2009 Stand-off species detection through picosecond Raman LIDAR with temporal fluorescence suppression and telescope characterisation Master of Science Thesis Martin Levenius Division of Combustion

More information

OPTI 511L Fall A. Demonstrate frequency doubling of a YAG laser (1064 nm -> 532 nm).

OPTI 511L Fall A. Demonstrate frequency doubling of a YAG laser (1064 nm -> 532 nm). R.J. Jones Optical Sciences OPTI 511L Fall 2017 Experiment 3: Second Harmonic Generation (SHG) (1 week lab) In this experiment we produce 0.53 µm (green) light by frequency doubling of a 1.06 µm (infrared)

More information

Analytical Spectroscopy Review

Analytical Spectroscopy Review Analytical Spectroscopy Review λ = wavelength ν = frequency V = velocity = ν x λ = 2.998 x 10 8 m/sec = c (in a vacuum) ν is determined by source and does not change as wave propogates, but V can change

More information

Chapter 2. Time-Resolved Raman Spectroscopy. Technical aspects and instrumentation

Chapter 2. Time-Resolved Raman Spectroscopy. Technical aspects and instrumentation Chapter 2. Time-Resolved Raman Spectroscopy Technical aspects and instrumentation 20 2.1 Photon dispersion in scattering media When measuring from a small focal volume in a transparent sample or a surface

More information

Underwater Raman Sensor for Detecting High Explosives and Homemade Explosives (HMEs)

Underwater Raman Sensor for Detecting High Explosives and Homemade Explosives (HMEs) Underwater Raman Sensor for Detecting High Explosives and Homemade Explosives (HMEs) Shiv K. Sharma* Hawaii Institute of Geophysics & Planetology, University of Hawaii (UH) at Manoa, Honolulu, Hawaii,

More information

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters ) September 17, 2018 Reference literature (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters 13-14 ) Reference.: https://slideplayer.com/slide/8354408/ Spectroscopy Usual Wavelength Type of Quantum

More information

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Atomic Absorption Spectrophotometry Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Defination In analytical chemistry, Atomic absorption spectroscopy is a

More information

Prediction of Transient Deflector Plate Temperature During Rocket Plume Impingment and its Validation through Experiments

Prediction of Transient Deflector Plate Temperature During Rocket Plume Impingment and its Validation through Experiments Prediction of Transient Deflector Plate Temperature During Rocket Plume Impingment and its Validation through Experiments PRIYA KAMESH KAMATCHI*, VISHNU POOVIAH*, PISHARADY.J.C ** *Department of Mechanical

More information

Chapter-4 Stimulated emission devices LASERS

Chapter-4 Stimulated emission devices LASERS Semiconductor Laser Diodes Chapter-4 Stimulated emission devices LASERS The Road Ahead Lasers Basic Principles Applications Gas Lasers Semiconductor Lasers Semiconductor Lasers in Optical Networks Improvement

More information

UV Vibrational Raman Spectroscopy Flame Diagnostics System ABSTRACT

UV Vibrational Raman Spectroscopy Flame Diagnostics System ABSTRACT UV Vibrational Raman Spectroscopy Flame Diagnostics System Yeshayahou Levy 1 and Liana Kartvelishvili 2 Faculty of Aerospace Engineering, Technion Haifa, 32000, Israel 1. E-mail: levyy@aerodyne.technion.ac.il

More information

CHAPTER 7 SUMMARY OF THE PRESENT WORK AND SUGGESTIONS FOR FUTURE WORK

CHAPTER 7 SUMMARY OF THE PRESENT WORK AND SUGGESTIONS FOR FUTURE WORK 161 CHAPTER 7 SUMMARY OF THE PRESENT WORK AND SUGGESTIONS FOR FUTURE WORK 7.1 SUMMARY OF THE PRESENT WORK Nonlinear optical materials are required in a wide range of important applications, such as optical

More information

LIDAR. Natali Kuzkova Ph.D. seminar February 24, 2015

LIDAR. Natali Kuzkova Ph.D. seminar February 24, 2015 LIDAR Natali Kuzkova Ph.D. seminar February 24, 2015 What is LIDAR? Lidar (Light Detection And Ranging) is an optical remote sensing technology that measures distance by illuminating a target with a laser

More information

University of Cyprus. Reflectance and Diffuse Spectroscopy

University of Cyprus. Reflectance and Diffuse Spectroscopy University of Cyprus Biomedical Imaging and Applied Optics Reflectance and Diffuse Spectroscopy Spectroscopy What is it? from the Greek: spectro = color + scope = look at or observe = measuring/recording

More information

Light Emission. Today s Topics. Excitation/De-Excitation 10/26/2008. Excitation Emission Spectra Incandescence

Light Emission. Today s Topics. Excitation/De-Excitation 10/26/2008. Excitation Emission Spectra Incandescence Light Emission Excitation Emission Spectra Incandescence Absorption Spectra Today s Topics Excitation/De-Excitation Electron raised to higher energy level Electron emits photon when it drops back down

More information

Assessment of the Azimuthal Homogeneity of the Neutral Gas in a Hall Effect Thruster using Electron Beam Fluorescence

Assessment of the Azimuthal Homogeneity of the Neutral Gas in a Hall Effect Thruster using Electron Beam Fluorescence Assessment of the Azimuthal Homogeneity of the Neutral Gas in a Hall Effect Thruster using Electron Beam Fluorescence IEPC-2015-91059 / ISTS-2015-b-91059 Presented at Joint Conference of 30th International

More information