CRT 16 Quick Reference Guide

Size: px
Start display at page:

Download "CRT 16 Quick Reference Guide"

Transcription

1 CRT 16 Quick Reference Guide Service Version Rev 03/30/01 Main Tests/Analyzer Option & Modes: Tests: Measured: Calculated: Na, K, Cl, TCO 2, Gluc, BUN (Urea), Creat, Hct Osmolality, BUN/Creat Ratio, Anion Gap Options: Analyzer Configurations: Stand-Alone (w/ or w/out Printer) Auto-Sampler (Cups-Only, or Primary Tube) (non-urine) Dilutor Option (Auto-Sampler plus Urine Diluting Option) Modes: Sample Types: Whole Blood (WB), Serum/Plasma, Urine, CSF (no MicroSample Option) Slope Limits: Na: 8.8 to 11.6 (1,7) K: 8.7 to 11.7 (5,8) Cl: 8.5 to 13.0 (1,4,5) TCO 2 : 8.0 to 14.5 (1,2,3,5) Gluc: 5.0 to 65.0 (1,2,3,7) BUN: 8.8 to 13.1 (1,3,7) Creat: 30.0 to (1,2,3,7) Hct: 5.0 to 20.0 (1,6,7) Notes: 1 = Long Life Electrode 2 = Int'l Ref Electr ode 3 = Membrane Cap; Push-On 4 = Membrane Cap; Screw-On 5 = Uses Washer 6 = Impedence (Air Detector) 7 = No Washer 8 = Mid-Range Life Flow Limits: Calibration: (seconds) Analysis: (seconds) Offset Limits: Slope/Intercept Used. Analysis Time/Rate: Analysis Time: 90 sec {published specification; not including Probe flush at start} 95 sec {full analysis time w/ "normal" Gluc/Creat range samples} {Extra rinse "F" cycles added for high Gluc/Creat increasing times) Maximum Time: 125 sec (full "F" rinse cycles added) Analysis Rate: 40 samples/hour (WB/Serum/Plasma) {Best Case: normal range samples; based on 90 sec time} 15 samples/hour (Urine Mode)

2 Quick Reference (CRT 16): Continued Sample Volume: WB, Serum/Plasma, CSF: Urine (before dilution) 385 ul. 100 ul. Dilution Ratios: Urine Dilution (final concentrations): All Tests (except BUN, Creat): 1:6 (one cup needed) BUN/Creat: 1:36 (two cups needed) Software Version: 3.13 (English & German only) 03/2001 (Italian/Japanese/French: Pending) Septum Positions: (6 position septum) Electrode/Flowcell Positions: Top: Std C Top: TCO 2 {Flowcell & Electrode} Std B Mixer Std A Reference D-Solution Cl {Main Flowcell} L-Solution BUN {Main Flowcell} Bottom: Flush Gluc {Main Flowcell} (non-septum) Air (outside) Creat {Main Flowcell} (non-septum) Sample (Probe out) K {Main Flowcell} Na {Main Flowcell} Bottom: ADET/Hct Fluids Details: Std A and B: Used to calibrate all electrodes except TCO 2.. Varying analytes concentrations included. (Preservative added.) Std C: Used to calibrate TCO 2; ; also in flowpath during idling and also used for some wash-out. Required for good stability of Gluc & Creat. (Preservative added.) Flush: Used to rinse flowpath during/after analysis cycles; may use more than 1 flush sequence if Gluc and/or Creat sample concentrations are high. H-Solution: Connects to Mixer; is an acid solution for conversion of HCO3- in sample to CO2 gas for transport into TCO2 electrode. Reference Solution: Connects to reference electrode; 2 M KCl. D-Solution: In optional 'Dilutor Pack'; urine diluent solution. L-Solution: In optional 'Dilutor Pack'; urine one-point cal standard.

3 Quick Reference (CRT 16): Continued Linear Ranges: WB/Serum/Plasma: CSF: Urine: Na: 80 to 200 mmol/l n/a 10 to 300 mmol/l K: 1 to 10 mmol/l n/a 10 to 300 mmol/l Cl: 60 to 200 mmol/l 60 to 200 mmol/l 20 to 300 mmol/l TCO2: 5 to 50 mmol/l n/a n/a Gluc: 25 to 500 mg/dl 25 to 500 mg/dl n/a BUN: 3 to 100 mg/dl n/a 100 to 1500 mg/dl Creat: 0.2 to 20 mg/dl n/a 10 to 400 mg/dl Hct: * 15 to 65 % n/a n/a * (Whole Blood Only) Pump/Valve Requirements: Pumps: Always 2 Pumps Valves: Only Waste Valve; No Pinch Valve Assembly required (for any configuration). Dependency Rules & Information: { Hct Measurement: Requires calibrated (no errors) Na+ channel for Hct reporting. { BUN Measurement: Requires calibrated (no errors) Na+ and K+ channels for BUN reporting. No separate NH4+ slope shown/required (as with CRT 12/14). { TCO2 Measurement: Requires calibrated (no errors) Hct channel for TCO2 correction. Assume: 12 < Hct % < 70 (for whole blood samples) Hct < 12 (acceptable for serum/plasma samples) Proper "Sample Type" must be selected to match sample or "Mismatch Error". { Standby Mode: NOT available on CRT 16. { Passwords: Password options for analysis available (like STP/Ultra) (47 unique passwords). If enabled, password required after pressing ANALYZE. Special "supervisor" password; Only "supervisor" password accesses Setup Menu. "TEXT" key ENABLES/DISABLES password protection (from Password menu). { Dialysate "Sample Type" not available on "Sample Type Key"; use 'Serum/Plasma' for dialysate selection. Updates: Current policy: Updates to CRT 16 will only be allowed from CRT 12/14 units. Updates will require that units have current "ADC FAM Chip PEEL Device (U37; ver >2.1) and "ANP ROM Chip" (U50; ver > 3.0) on the Main Controller Board. Chips must be replaced on all units with older version chips to operate as a CRT 16.

4 Calibration Information: 2 Point Calibration: (Turned off in Standby ) Fixed: Done automatically each TWO (2) Hours. Manual: At any time by pressing Calibrate key. 1 Point Calibration: Done with EACH sample analysis. Standby Mode: Manual entry and exit (via sub-menu option). All calibrations turned off in Standby Mode. Idle pumping done only. Sample & QC Storage: Stat Mode: Tray (Autosampler) Mode: Serum: Urine: Holds last 80 samples (FIFO). Holds one tray (40 cups/samples). 12 Urines/Tray Max. QC Analysis: Recommended: One (1) Level Start of Each Shift/Every 8 Hours ('Ideal': "Nova Chemistry Control + Creat" 2 levels, plus Hct Controls 2 levels) Chemistry Controls Pus Creatinine: (Store FROZEN; Good for 14 days open-vial refrigerated) Multipack: Levels 1 & 2 only Level 3: Optional QC (sold separate; high range) QC Files/Libraries: Seven (7) QC Files Plus Proficiency on all CRT: Files 1-7: For QC data storage: ( Open QC files: Can program w/ any QC matl) File 0: Proficiency samples & Performance Check Solution only.

5 CRT 16 Creatinine Electrode Theory Relevent Chemical Reaction: Creatinine Amidohydrolase (Enzyme) Creatinine + H 2 O > Creatine Creatinine Amidohydrolase (Enzyme) Creatine + H 2 O > Sarcosine + Urea Sarcosine Oxidase (Enzyme) Sarcosine + O 2 + H 2 O > Formadehyde + Glycine + H = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = V H 2 O > 2 H+ + O e- [2 e-] = Proportional to amount of Creatinine. = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = Creatrinine Creat Membrane > H2O2 Electrode Creatrinine Creatrinine > > H2O2 H2O2 2 e-

6 Quick Reference (CRT 16): Continued CRT 16 Electrode/Standards Information: Rev: 02/29/96 Electrode Standards Used in Approximate Delta mv Calibration/Slope (between standards) Na Stds A & B 14 K Stds A & B 21 Cl Stds A & B 13 TCO2 Stds A & C 20 Gluc Stds A & B 175 BUN Stds A & B 25 Creat Stds A & B > 400 Hct/ADET Stds A & B > 300

7 General Maintenance/QC: CRT 16 Maintenance Summary Page 1 of 6 Rev 03/30/01 1. Clean sample inlet port/base of Septum. (AS NEEDED) 2. Flowpath Conditioning. (AS NEEDED) (Use serum/plasma/whole blood: Automated Cycle) a.) Done daily for low-mid volume users. b.) Not required for higher volume users. 3. Control Analysis. (DAILY) "Nova Chemistry Controls + Creat : 2 levels (Multipack) liquid control; stored frozen, shipped with Ice Packs. (Open Vial Life: 2 weeks refrigerated) ** 3rd QC Level (high QC; sold separately) ** "Best" control: Stable TCO2 & Creat balance for CRT electrodes. Hct Controls: 2 levels (STP version; aqueous) 4. "Performance Check Solution" (Glucose/BUN/Creatinine): (DAILY) New material: Stored frozen/shipped w/ Ice Packs; in ampules for single use. a.) Notes: As membrane ages, very high end results may appear 'normal' due to lower activity (Gluc). BUN can show falsely higher OR normal results with high-end samples as membrane ages so BUN has "range value" to verify. Creatine added to solution to verify Creat Membrane can differentiate between Creatine and Creatinine. b.) Testing Results: (must check results on "Sensor Data" Screens) Gluc: > 700 mg/dl (> 38.9 mmol/l) for acceptable membrane (other CRT models: > 700 mg/dl) BUN: > 100 (> 71.4 mmol/l) for acceptable membrane (other CRT models: > 100 mg/dl) Creat: < 1.6 mg/dl (< 142 umol/l) for acceptable membrane 5. Linearity Standards: (AS NEEDED) "Linearity Stds 'SET F'": (4 levels including ALL parameters plus Creat) Used exclusively on CRT 16. Stored frozen; shipped w/ Ice Packs. Run to verify linearity &/or for troubleshooting. NOT ROUTINE. 6. Probe/Air Dectector Cleaning/Flushing. (WEEKLY) a.) Press FLOWPATH; Remove ADET from Flowpath b.) Clamp R & H tubings (going into electrodes) with hemostat/clamp. c.) Flush from ADET down through S-Line/Probe w/ Cleaning Soln (PN 11272). d.) Let sit 2 minutes. Then Flush with water (x2) and air (x1). (Continued)

8 CRT 16 Maintenance Summary General Maintenance/QC (Continued): Page 2 of 6 Probe/Air Dectector Cleaning/Flushing (continued): e.) Attach syringe/tubing to TCO2 Flowcell port, and flush downward with water through flowpath. f.) Flush the flowpath with air, then reconnect the W-line and unclasp R/H lines. Exit Flowpath and calibrate. Flowpath Cleaning. (AS NEEDED) a.) Non-automated cleaning of entire flowpath. b.) Build-up of protein/wastes common at Na, Mixer, Reference, TCO2 areas c.) Procedure (notes) 1) Go to 'Flowpath Maintenance' (Probe is slightly outside Septum). 2.) Clamp R & H tubings (going into electrodes) with hemostat/clamp. 3.) Disconnect W-line from top of TCO2 Flowcell. 4.) Place cup/beaker under Septum/Probe. 5.) Attach syringe/tubing to TCO2 Flowcell port, and flush downward with water through flowpath. 6.) Flush with air, then reconnect W-line and unclasp R/H lines. Exit Flowpath and calibrate. d.) Done AS NEEDED (officially) for routine. "Best" if done ONCE/WEEK routinely combined with Na+ Conditioning. e.) Done as first troubleshooting step for flow/clog problems. f.) Combine with Na Conditioning on monthly schedule. 7. Mixer Cleaning: (AS NEEDED) a.) Semi-automated cycle with instructions on use of fluids/tubing connections. b.) Frequency: Done AS NEEDED (officially). "Best" if done ONCE/MONTH for serum users; EVERY 2 WKS for Plasma/WB users. c.) Cleans Mixer and TCO2. (manual cleaning of Reference is recommended). 8. Sodium Conditioning: (TWICE/WEEK) a.) Condition (with Na/pH Cond Solution) TWICE/WEEK. b.) Manual removal, soak 2 minutes in Na/pH Conditioning Soln (PN 00018). Access "Flowpath" for fluid purge before electrode removal. Do in conjunction with Flowpath Cleaning c.) Condition NEW electrode before use.

9 CRT 16 Maintenance Summary (continued) Electrode Maintenance Information: Page 3 of 6 Sodium (Na): { STP Electrode (same). { Condition (with Na/pH Cond Solution) TWICE/WEEK (combine with ADET cleaning). { Condition NEW electrode before use. { Use PN Na/pH Cond Soln (EL/CRT/NUC version). Potassium (K): { Mid-life electrode (Range: Life = 3-9 months typical). { STP Electrode (same). { Flowpath conditioning (blood) will improve life/slope; done as needed. { Has Flowcell Washer (STP type). Chloride (Cl): { Electrode and Cap; STP same. { Has Flowcell Washer (STP type). { Cap = Short Life (Range: Life = 2-8 weeks). Change only AS-NEEDED (slope/results). { Cap Replacement: Fill cap to very TOP of threads with Cl IFS. Total CO2 (TCO2): { Electrode and Cap: New Design.; Electrode = Long Life (Range = 1-4 years) { Cap = Short Life (Range: Life =2-12 weeks). Change only AS NEEDED (slope/results). { Cap Replacement: Push-On type (similar to pco2 STP); must degas/shake-down & Flowpath Condition before calibration. Fill cap with TCO2 IFS to slightly below line above Vent Hole. { Conditioning with pco2/tco2 Conditioning Solution upon install of NEW electrode and/or for troubleshooting ONLY. (Soak for 15 minutes outside unit as required) { TCO2 Flowcell: Separate part from TCO2 electrode/membrane. { Has Flowcell Washer (new design). BUN/Urea: { Electrode (new version; longer wire than STP); Membrane (new design). { Push-On Membrane Cap: Replace AS-NEEDED based on slope/results/overrange soln. { Electrode: Range: Life = 1-2 years; Membrane: Range: Life = 2-8 weeks. { Run Overrange Solution daily to verify membrane/results. { Polishing NEVER done. { Membranes stored in refrigerator; shipped non-refrigerated.

10 CRT 16 Maintenance Summary (continued) Electrode Maintenance Information (continued): Page 4 of 6 Glucose (Gluc): { Electrode (STP same); Membrane (CRT 12 same). { Push-On Membrane Cap: Replace AS-NEEDED based on slope/results/overrange soln. { Electrode: Range: Life = 1-3 years; Membrane: Range: Life = 2-8 weeks. { Run Overrange Solution daily to verify membrane/results. { Auto-Polishing done daily or on-demand on CRT 16; time selectable. { Manual Electrode Polishing: for Troubleshooting Only. a.) Polishing 'Conditions': ** Consistently cannot recover > 700 mg/dl on Overrange Solution. ** Chronic slope and/or results problems not resolved by membrane changing. b.) Polishing Procedure: ** Polish using only new YELLOW (1 u) Polishing Paper only. ** Polish for 10 seconds (use H20 and Polishing Paper) ** Condition for 2 minutes in CRT Cleaning Solution. ** Rinse with H20 and 'blot' dry. { Membranes stored in refrigerator; shipped non-refrigerated. Creatinine: { New electrode/membrane (can use STP glucose electrode to troubleshoot if required). { Push-On Membrane Cap: Replace AS-NEEDED based on slope/results/overrange soln. { Electrode: Range: Life = 9-24 months; Membrane: Range: Life = 5-14 days. { Run Performance Check Solution daily to verify membrane/results. (Perf Check Solution: Heat sensitive for Creat: Stored FROZEN; ship ICE) { Auto-Polishing done daily or on-demand on CRT 16; time selectable. { Manual Electrode Polishing: NEVER TO BE DONE ** Change electrode for Overrange Soln results problems and/or slope/qc problems not resolved through membrane change. { "Chloridizing" of Electrode: (Done with new electrode install and EACH time membrane is replaced.) a.) Immerse in "Electrode Cleaning Solution" (PN 11272) for 2 minutes min. b.) Rinse with water; blot dry only, do NOT wipe. c.) Metallic ring should be dull; not bright. { Membranes heat sensitive: Stored FROZEN; shipped w/ Ice Packs &/or Dry Ice. { Storage (Membranes): Refrigerate (2-4 C) { Shipment to Dealers: Dry-Ice/Ice-Packs; To Customers Should use only ice. { Can store large quantities of membranes frozen. { Membrane Warranty = 3 days or 200 samples. (CONTINUED)

11 Electrode Maintenance Information (continued): Page 5 of 6 Creatinine (Continued): { To replace Creat Membrane: * Replace membrane in "Flowpath Maintenance" cycle. After completion, exit Flowpath wait 10 minutes. * After 10 minute wait, perform 3 calibrations minimum. If after 4th cal calibration, slope is still unacceptable, replace membrane again (must Chloridize electrode also). HCT/Air Detector: { Same as CRT version (Hct Detector). { Has heater circuit; only used for Hct, non-electrode dependent. { CRT 16 does not have automated "Probe (ADET) Cleaning Cycle" (must do manual ADET cleaning; see "Cleaning Maintenance" notes). { Clean ADET (manually) AS NEEDED (officially). "Best" if done ONCE/MONTH. Reference Electrode: { New version (CRT type with connector wire). { In-line between 'Main Flowcell' and Mixer. { Clean (manually) in conjunction with 'Mixer Cleaning' maintenance (more for plasma/wb users). Mixer Assembly: { New design; non-stp or CRT. { Similar function of CRT: mixes H-solution (acid) with sample for CO2 gas liberation. { 'Mixer Cleaning' cycle same as CRT models { Frequency: Done AS NEEDED (officially). "Best" if done ONCE/MONTH for serum users; EVERY 2 WKS for Plasma/WB users.

12 Fluid Pack Information: CRT 16 Maintenance Summary (continued) Page 6 of 6 Activation: { Stds A,B,C are BAG fluids with 2 pouches: one liquid section, and one powdered section. Clip dividing pouches must be removed before use and contents combined and thoroughly mixed. Pack Life/Use: { Pack (RMS tracking) good for 21 days from RMS pod install (activation). { Date pack "activated" (RMS pod installed) denoted on RMS (Fluid Pack Change) screen. { 2-3 week used life "estimated" at full pack usage with volume of almost 0 samples/day. Pack may be used after "expired" date: unit simply denotes pack expired on both Base Screen AND Fluid Pack Change screen. Use beyond RMS expiration based on: * Creat results usage dependent on control and overrange solution results. * Other parameters unaffected by Use Life issue and can be used until pack end. Other Information: { Stds A/B/C: Used for calibration of all analytes. { Stds A/B/C: Have bacterial/fungal preservative added. { Other Fluids: No preservatives added. Other Maintenance Items: Tubings: { W/R/H Pump Tubings: Change every 3 months. { Tubing Harness: Change every months. { Septum: Change with every fluid pack (Not recommended to run more than one (1) month. Increased septum wear with CRT 16 since Probe must move all the way out for air aspiration vs. position # 3 on all other CRT models.) { Probe: Change yearly or as needed.

13

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

ph electrode Instruction Manual For use with the RAH-210 & RPH-250 ph electrode Instruction Manual For use with the RAH-210 & RPH-250 The information in this manual was current at the time of printing. The most current versions of all Hydro Instruments manuals can be

More information

EASYLYTE Na/K/Cl/Ca/Li SOLUTIONS PACK REF mL

EASYLYTE Na/K/Cl/Ca/Li SOLUTIONS PACK REF mL EASYLYTE Na/K/Cl/Ca/Li SOLUTIONS PACK REF 2124 800mL INTENDED USE The EasyLyte Na/K/Cl/Ca/Li Solutions Pack is intended for the quantitative determination of sodium (Na + ), potassium (K + ), chloride

More information

Chemistry MultiMeasure Sensor PS-2170

Chemistry MultiMeasure Sensor PS-2170 ä 5 Instruction Manual 012-09523A Chemistry MultiMeasure Sensor PS-2170 ph/ise/orp PRESSURE CHEMISTRY TEMPERATURE SENSOR VOLTAGE 1 2 3 4 PS-2170 5 6 7 cc 8 10 15 20 25 30 35 40 45 50 55 60 Included Equipment

More information

STANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER

STANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER STANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER Estimation of Dietary Intake of Sodium (DIS) in 24h Urine on electrolyte

More information

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7 Troy, NY 12180 Date: 03/14/2012 Page 1 of 7 USGS New York Water Science Center Soil and Low Ionic Strength Water Quality Laboratory Electrometric Measurements and Titration Standard Operating Procedure

More information

Rapid RH Portable ph Meter. User's Manual

Rapid RH Portable ph Meter. User's Manual Rapid RH Portable ph Meter User's Manual This page intentionally left blank. 2 Contents 1. Introduction... 4 2. Included with the Rapid RH Portable ph Meter... 4 3. Understanding ph... 5 4. Calibrating

More information

User instructions. Gel filled ORP/Redox electrode: Model MTC10101, MTC10103, MTC10105, MTC10110, MTC10115 or MTC10130

User instructions. Gel filled ORP/Redox electrode: Model MTC10101, MTC10103, MTC10105, MTC10110, MTC10115 or MTC10130 User instructions Gel filled ORP/Redox electrode: Model MTC10101, MTC10103, MTC10105, MTC10110, MTC10115 or MTC10130 Safety Information DOC272.53.80033 Precautionary Labels Read all labels and tags attached

More information

Automated Clinical Analyzer

Automated Clinical Analyzer Automated Clinical Analyzer Improved ease of operation and better test efficiency, and provides the optimal solution for routine operations, STAT operation and a broad range of clinical analysis requirements.

More information

JAPAN QUALITY. Automated Clinical Analyzer

JAPAN QUALITY. Automated Clinical Analyzer JAPAN QUALITY Automated Clinical Analyzer Higher test performance and operability with new functions The clinical analyzer in the next stage 1 The best to the requirements for functionality and usability

More information

A new chapter. medicacorp.com

A new chapter. medicacorp.com A new chapter. \EDIC@ medicacorp.com Same operational reliability. Reliable performance, test after test. The photometer s xenon fl ash lamp typically has a 5-year use life and is maintenance-free. No

More information

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051 imagine explore learn Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051 Overview Sodium is highly abundant in the lithosphere (Earth s outer shell) where it occurs as rock salt deposits as

More information

JAPAN QUALITY. Automated Clinical Analyzer

JAPAN QUALITY. Automated Clinical Analyzer JAPAN QUALITY Automated Clinical Analyzer Higher test performance and operability with new functions The clinical analyzer in the next stage 1 The best to the requirements for functionality and usability

More information

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A imagine explore learn Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A Overview Ammonium ions (NH 4+ ) can be found in fertilizer, cleaning products, processed meats and in the waste of

More information

The Clinical Chemistry System that offers more so you do less!

The Clinical Chemistry System that offers more so you do less! The Clinical Chemistry System that offers more so you do less! A unique level of automation in a compact system Safe and flexible sample management Fully automatic cuvette management Cuvette management

More information

RayBio Human IDUA ELISA Kit

RayBio Human IDUA ELISA Kit RayBio Human IDUA ELISA Kit Catalog #: ELH-IDUA User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross, GA

More information

hydrideicp Hydride Generation System

hydrideicp Hydride Generation System HG-MP2-4-A hydrideicp Generation System hydrideicp Hydride Generation System User Manual MP 2 hydrideicp Generation System Elemental Scientific Phone: 1.402.991.7800 Email: esi@icpms.com 1500 North 24th

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required

More information

Bante210 Benchtop ph Meter Instruction Manual

Bante210 Benchtop ph Meter Instruction Manual Bante210 Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante210 Benchtop ph Meter 1 Introduction Thank you for selecting the Bante210 benchtop ph meter. This manual provides a step-by-step

More information

epoc Blood Analysis System: Summary of Analytical Methods and Performance

epoc Blood Analysis System: Summary of Analytical Methods and Performance White Paper epoc Blood Analysis System epoc Blood Analysis System: Summary of Analytical Methods and Performance siemens.com/poc White Paper epoc Blood Analysis System epoc Blood Analysis System White

More information

cobas c 111 analyzer Big analyzer performance in a small package

cobas c 111 analyzer Big analyzer performance in a small package For more information, contact your Roche Diagnostics representative or call 1-800-852-8766 (option 7). cobas c 111 analyzer Big analyzer performance in a small package www.poc.roche.com COBAS, COBAS B,

More information

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

RayBio Human Thyroid Peroxidase ELISA Kit

RayBio Human Thyroid Peroxidase ELISA Kit RayBio Human Thyroid Peroxidase ELISA Kit Catalog #: ELH-TPX User Manual Last revised July 6, 2017 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100

More information

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049 imagine explore learn Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049 Overview Fluoride is often added to drinking water and to toothpaste to prevent dental decay. In high concentrations,

More information

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution , drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific

More information

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

PHS-25CW/3BW Benchtop ph Meter Instruction Manual PHS-25CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides

More information

user manual blotting Hoefer PR648 Slot blot manifold um PR648-IM/Rev.F0/08-12

user manual blotting Hoefer PR648 Slot blot manifold um PR648-IM/Rev.F0/08-12 user manual blotting Hoefer PR648 Slot blot manifold um PR648-IM/Rev.F0/08-12 Page finder Introduction...1 Specifications...2 Procedure for standard use of the Hoefer PR648...3 Setting up the Hoefer PR648...3

More information

QAM-I-111 Operation and Calibration of the Conductivity Meter

QAM-I-111 Operation and Calibration of the Conductivity Meter 1.0 Applicability and Purpose i. This procedure applies to the operation and calibration of the YSI Model 3200 conductivity meter. This procedure is performed prior to any analysis using this meter. By

More information

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction User s Guide Waterproof Palm ph Meter Model PH220 Introduction Congratulations on your purchase of the Extech PH220 waterproof ph/mv meter. This microprocessor-based device with tactile buttons is battery

More information

Technical Procedure for ELISA Drug Screen Version 3 Toxicology Unit Effective Date: 02/15/2013. ELISA Drug Screen

Technical Procedure for ELISA Drug Screen Version 3 Toxicology Unit Effective Date: 02/15/2013. ELISA Drug Screen ELISA Drug Screen 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Tecan Freedom EVO 75 Workstation for an ELISA Drug Screen. 2.0 Scope This procedure applies

More information

symphony Multisensor Probes

symphony Multisensor Probes DOC022.53.80319 symphony Multisensor Probes 07/2012, Edition 1 User Manual Table of contents Safety information on page 3 Calibration on page 6 Specifications on page 3 Sample requirements on page 8 Product

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Fluoroborate Ion Selective Electrode is used to quickly, simply, accurately, and economically measure Fluoroborate in aqueous solutions.

More information

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA , 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine

More information

Drug Testing Solutions INDIKO DRUG TESTING ANALYZER. 1 P F

Drug Testing Solutions INDIKO DRUG TESTING ANALYZER. 1  P F Drug Testing Solutions INDIKO DRUG TESTING ANALYZER 1 www.dtpm.ws P. 256-845-1261 F. 888-317-4178 sales@dtpm.ws Overview Indiko Sample oriented random access system for clinical and specialty testing,

More information

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Ammonium Ion Electrode is used to quickly, simply, accurately, and economically

More information

ph Probe ReallyEasyData com

ph Probe ReallyEasyData com ph Probe 9200001 Uses Whether you re monitoring ph for chemistry, physical science, life science, or earth science activity, this ph meter offers accurate readings in a convenient format. Use the RED ph

More information

Operating Instructions for Ammonium ISE Specifications

Operating Instructions for Ammonium ISE Specifications Operating Instructions for Ammonium ISE Specifications Range: The Ammonium Electrode responds to uncomplexed ion activity over the range 1 x 10-1M to 1 x 10-6M. Linear detection limit is about 10-5M Interference's:

More information

cobas pro integrated solutions Simplicity meets Excellence

cobas pro integrated solutions Simplicity meets Excellence cobas pro integrated solutions Simplicity meets Excellence Introducing cobas pro integrated solutions Simplicity meets Excellence 01 02 03 2 01 Sample supply unit Up to 1,000 samples per hour processing

More information

General. Power requirements: 110/220 Volts 50/60 VAC. Dimensions: Width 80 cm x Height 45 cm x Depth 58 cm. Weight: 57 Kg.

General. Power requirements: 110/220 Volts 50/60 VAC. Dimensions: Width 80 cm x Height 45 cm x Depth 58 cm. Weight: 57 Kg. General Power requirements: 110/220 Volts 50/60 VAC Dimensions: Width 80 cm x Height 45 cm x Depth 58 cm Weight: 57 Kg. Features Reagents cooling Cuvettes washer Continuous input of samples Auto re-run

More information

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7 Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes

More information

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-25CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides

More information

Clinical Chemistry Quality Control Policies, Processes, & Procedures

Clinical Chemistry Quality Control Policies, Processes, & Procedures Clinical Chemistry Quality Control Policies, Processes, & Procedures Prepared By: Brendan Jones Clinical Chemistry Quality Supervisor Laverty Pathology What is Quality Control? The statistical process

More information

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

RayBio Rat TNF-alpha ELISA Kit (For Lysates)

RayBio Rat TNF-alpha ELISA Kit (For Lysates) RayBio Rat TNF-alpha ELISA Kit (For Lysates) Catalog #: ELR-TNFa-CL User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite

More information

ELISA Technical Manual and Troubleshooting Guide

ELISA Technical Manual and Troubleshooting Guide ELISA Technical Manual and Troubleshooting Guide 2014 Contents: Page Kit Storage Guidelines 3 Materials and Equipment Required 4 Performing the ELISA Assay 5-8 Sample Handling and Storage 9 Troubleshooting

More information

CX4, CX5, CX7, CX9 PRO

CX4, CX5, CX7, CX9 PRO Beckman CX3, CX4CE, CX5CE, CX7, Synchron Beckman CX3, CX4, CX5, CX7 Delta, CX9 ALX Beckman CX3, CX4, CX5, CX7, CX9 PRO BK-A17652D A17652 Electrolyte Buffer Reagent 4.7 CX 3,5,7,9 2x2000mL BK-A17657D A17657

More information

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0) 150mV 300mV Mono/Reference Combination S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) 1376 340 456 Fax: +44 (0) 1376 340 453 Email:

More information

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C V 2.8 Revised 1/16/18 ph Probe Silver / silver chloride Reads ph Range 0 14 Resolution Response time Max pressure Max depth Temperature range C Cable length Internal temperature sensor Time before recalibration

More information

Warnings. This device is not suitable for use in a potentially explosive environment and cannot be used in an oxygen tent.

Warnings. This device is not suitable for use in a potentially explosive environment and cannot be used in an oxygen tent. ! " # # $ % Warnings This device is not suitable for use in a potentially explosive environment and cannot be used in an oxygen tent. Dispose of this device properly. It should not be disposed of in a

More information

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832 , TNTplus 832, 10205 DOC316.53.01083 Salicylate Method Method 10205 HR (2 to 47 mg/l NH 3 N) TNTplus 832 Scope and Application: For surface waters, municipal and industrial wastewaters. Test preparation

More information

H900 Electrolyte Analyzer. Operating/Technical Manual

H900 Electrolyte Analyzer. Operating/Technical Manual H900 Electrolyte Analyzer Operating/Technical Manual Shenzhen Xilaiheng Medical Electronics Co., Ltd. Product information Product model: H900 Product name: Electrolyte Analyzer Manufacturing site: No.

More information

Hepatitis B virus IgM ELISA Kit

Hepatitis B virus IgM ELISA Kit Hepatitis B virus IgM ELISA Kit Catalog Number KA0289 96 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay...

More information

STANDARD OPERATING PROCEDURE

STANDARD OPERATING PROCEDURE Ref.. Table of Contents Page. Scope 2 2. Objective 2 3. Definitions 2 4. Responsibility 3 5. Procedure 3 5.. Diagram of Display and Control Panel 3 5.2. Installation 4 5.3. Operation 4 5.4. Calibration

More information

IonPac Trace Cation Concentrator

IonPac Trace Cation Concentrator for the IonPac Trace Cation Concentrator (TCC-LP1, TCC-ULP1 and TCC-XLP1) Page 1 of 17 PRODUCT MANUAL FOR TRACE CATION CONCENTRATOR (TCC) TCC-LP1 Column Low Pressure, 4 x 35 mm (P/N 046027) TCC -ULP1 Column

More information

PC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual

PC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual PC100 Handheld ph/cond./tds/sal. Meter Instruction Manual Table of Contents 1. Introduction ------------------------------------------------------------------------------------------------------------------------

More information

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Water and Lab Products, Thermo Fisher Scientific Technical Note 505 Key Words Thermo Scientific Orion

More information

Manual ph Probe & Oxidation Reduction Potential Probe Models 6528A and 6528B

Manual ph Probe & Oxidation Reduction Potential Probe Models 6528A and 6528B Manual ph Probe & Oxidation Reduction Potential Probe Models 6528A and 6528B Revision History File name / Revision Date Authors Previous version BX 2004 RS/ JH Unidata Manual - 6528 ph and ORP Sensors

More information

This sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode.

This sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode. ph sensor with electrode ENPH-A016 ph sensor with flat electrode ENPH-F052 The ph sensor replaces the traditional ph meter adding powerful capabilities such as collecting the ph data and displaying the

More information

MassWWP Standard Operating Procedure L-6 ph and Alkalinity Rev.0 Page 1 of 12

MassWWP Standard Operating Procedure L-6 ph and Alkalinity Rev.0 Page 1 of 12 MassWWP Standard Operating Procedure L-6 ph and Alkalinity Rev.0 Page 1 of 12 MassWWP Standard Operating Procedure L-6 ph and Alkalinity Rev.0 Page 2 of 12 Table of Contents Field Equipment List...3 Sampling

More information

Prepared By: R. W. Waggoner, Jr. Effective Date: August 18, 2008

Prepared By: R. W. Waggoner, Jr. Effective Date: August 18, 2008 Name of Procedure: Use of the Dade Behring Viva Jr. EMIT Analyzer as a Drug Screen Suggested Uses: This procedure does not cover every aspect of the instrument used. The operator of the instrument should

More information

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure Troy, NY 12180 Date: 12/12/2017 Page 1 of 7 USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure 1. Scope and Application 1.1 Analytes Dissolved inorganic

More information

NITROGEN, TOTAL, HR, Test N Tube

NITROGEN, TOTAL, HR, Test N Tube Method 10072 NITROGEN, TOTAL, HR, Test N Tube 10 to 150 mg/l N TNT Persulfate Digestion Method For water and wastewater Digestion is required for determining total nitrogen. The digestion procedure is

More information

Ion Selective Electrode Probe

Ion Selective Electrode Probe Rev. 1 Ion Selective Electrode Probe KDS-1064 Calcium KDS-1065 Ammonium KDS-1066 Nitrate KDS-1067 Chloride Type : Ion selective electrode Range : 1 M to 5x10-7 M(40,000 ~ 0.01 ppm) Sensor description The

More information

SCIEX Cation Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

SCIEX Cation Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49109AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including

More information

INE-PHS-3E ph Meter Operating Manual

INE-PHS-3E ph Meter Operating Manual INE-PHS-3E ph Meter Operating Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.12.16 ADVICE: Read this instruction

More information

RayBio Human CA-125 ELISA Kit

RayBio Human CA-125 ELISA Kit RayBio Human CA-125 ELISA Kit Catalog #: ELH-CA125 User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross,

More information

IgG (Cat) ELISA Kit. Catalog Number KA assay Version: 01. Intend for research use only.

IgG (Cat) ELISA Kit. Catalog Number KA assay Version: 01. Intend for research use only. - IgG (Cat) ELISA Kit Catalog Number KA2048 96 assay Version: 01 Intend for research use only www.abnova.com Introduction and Background A. Test principle The principle of the double antibody sandwich

More information

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth V 2.4 Revised 3/1/18 Spear Tip ph Probe Silver / silver chloride Reads ph Range 0 14 Resolution Response time Max pressure Max depth Temperature range C Cable length Internal temperature sensor Time before

More information

рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically USER MANUAL

рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically USER MANUAL рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically Complete Set ph PRO Controller - 1 pcs. ph electrode - 1 pcs. Calibration

More information

RayBio Human VEGFR1 ELISA Kit

RayBio Human VEGFR1 ELISA Kit RayBio Human VEGFR1 ELISA Kit Catalog #: ELH-VEGFR1 User Manual Last revised July 6, 2017 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross,

More information

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT METHOD 9040B ph ELECTROMETRIC MEASUREMENT 1.0 SCOPE AND APPLICATION 1.1 Method 9040 is used to measure the ph of aqueous wastes and those multiphase wastes where the aqueous phase constitutes at least

More information

Human anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit

Human anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit Human anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit Catalog Number. MBS700087 For the qualitative determination of human anti-myelin associated glycoprotein antibody (MAG) concentrations

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

Mouse IgG2B ELISA. Cat. No. KT-405 K-ASSAY. For the quantitative determination of IgG2B in mouse biological samples. For Research Use Only.

Mouse IgG2B ELISA. Cat. No. KT-405 K-ASSAY. For the quantitative determination of IgG2B in mouse biological samples. For Research Use Only. K-ASSAY Mouse IgG2B ELISA For the quantitative determination of IgG2B in mouse biological samples Cat. No. KT-405 For Research Use Only. 1 Rev. 020808 K-ASSAY PRODUCT INFORMATION Mouse IgG2B ELISA Cat.

More information

ABX Pentra 400. Clinical Chemistry System. Up to 420 tests/hour 55 Assays Validation Station

ABX Pentra 400. Clinical Chemistry System. Up to 420 tests/hour 55 Assays Validation Station ABX Pentra 400 Clinical Chemistry System Up to 420 tests/hour 55 Assays Validation Station ABX Pentra400 From routine to esoteric testing ABX Pentra 400 is an extremely compact Clinical Chemistry benchtop

More information

TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX

TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX LANCER reserves the right to modify, constantly its documentation for its improvement. The values of adjustments indicated in the displays of this manual are

More information

RayBio Human LBP ELISA Kit

RayBio Human LBP ELISA Kit RayBio Human LBP ELISA Kit Catalog #: ELH-LBP User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross, GA

More information

technical bulletin Installation Instructions: Wall-Mount Dewpoint Probe ( )

technical bulletin Installation Instructions: Wall-Mount Dewpoint Probe ( ) Installation Instructions: Wall-Mount Dewpoint Probe (809-2011) The HMW90 Wall-Mount Dewpoint Probe (P/N 809-2011) HMW90 measures dewpoint for indoor applications such as anti-sweat heater control and

More information

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple Reliable TOC measurements Stop throwing away your first rep. QbD1200 has virtually eliminated sample-to-sample carryover

More information

Jumbo Cross Beam Ionizer

Jumbo Cross Beam Ionizer Jumbo Cross Beam Ionizer Version 1.0 September 30, 2008 Page 1 of 19 Table of Contents Cross Beam Ionizer... 1 Table of Contents... 2 1.0 Packing List... 3 1.1 Packing List for Cross Beam Ionizer... 3

More information

Aflatoxin M1 (AFM1) ELISA Kit

Aflatoxin M1 (AFM1) ELISA Kit Aflatoxin M1 (AFM1) ELISA Kit Catalog Number. CSB-EL027236 This immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power. This package insert

More information

Providing Vital Solutions for Your Laboratory

Providing Vital Solutions for Your Laboratory Test Menu General Chemistry Assays Albumin Bilirubin, Direct Bilirubin, Total Calcium Creatinine Glucose Iron, Total Magnesium Phosphorus Protein, Total Urea Nitrogen (BUN) Uric Acid Enzyme Assays Alanine

More information

Human rheumatoid factor (RF) antibody (IgM) ELISA Kit

Human rheumatoid factor (RF) antibody (IgM) ELISA Kit Human rheumatoid factor (RF) antibody (IgM) ELISA Kit Catalog Number. CSB-EQ027654HU For the qualitative determination of human rheumatoid factor (RF) antibody (IgM) concentrations in serum, plasma. This

More information

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Catalog No: E0785Mo 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

HIPPURIC ACID, o-m-p-methylhippuric ACIDS and CREATININE IN URINE by UV CODE Z07510

HIPPURIC ACID, o-m-p-methylhippuric ACIDS and CREATININE IN URINE by UV CODE Z07510 HIPPURIC ACID, o-m-p-methylhippuric ACIDS and CREATININE IN URINE by UV CODE Z07510 Biochemistry The determination of Hippuric Acid ( HPA ) in urine is important to determine the risk due to the exposition

More information

Human Superoxide Dismutase 2 (SOD2) ELISA

Human Superoxide Dismutase 2 (SOD2) ELISA Human Superoxide Dismutase 2 (SOD2) ELISA For the quantitative determination of human SOD2 in biological samples Cat. No. KT-50848 For Research Use Only. Not for use in diagnostic procedures. Page 1 of

More information

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information

More information

Richter Optica. Instructions for Model: S850 Stereo Zoom Microscope

Richter Optica. Instructions for Model: S850 Stereo Zoom Microscope Richter Optica info@richter-optica.com Instructions for Model: S850 Stereo Zoom Microscope Microscope Head Focus Knob Stage Clips Zoom Knob C-Mount Adapter Eyepiece Focusing Holder Focusing Holder Set

More information

Human Lipocalin-2/NGAL ELISA

Human Lipocalin-2/NGAL ELISA K-ASSAY Human Lipocalin-2/NGAL ELISA For the quantitative determination of Lipocalin-2/NGAL in human biological samples Cat. No. KT-542 For Research Use Only. 1 Rev. 13453542 K-ASSAY PRODUCT INFORMATION

More information

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions. Cole-Parmer Ammonium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium

More information

What is a WQSensor? Software Installation. Uninstalling WQSensors Software. NexSens Technology, Inc. TABLE OF CONTENTS

What is a WQSensor? Software Installation. Uninstalling WQSensors Software. NexSens Technology, Inc. TABLE OF CONTENTS TABLE OF CONTENTS 1 Sensor Operation...3 1.1 Ion Selective Electrode Sensors...3 1.1.1 WQ-FL Fluoride ISE Sensor... 3 1.1.2 Making ISE Sensor Measurements... 4 1.1.3 Maintenance and Care... 8 1.1.4 ISE

More information

Adenosine Deaminase Assay (ADA) Kit Instructions

Adenosine Deaminase Assay (ADA) Kit Instructions V.2 Crystal Chem Adenosine Deaminase Assay (ADA) Kit Instructions For the quantitative determination of adenosine deaminase in serum or plasma Catalog #80249 120 Assays For research use only. Not for use

More information

Procine GHb ELISA Kit

Procine GHb ELISA Kit Procine GHb ELISA Kit For the quantitative in vitro determination of Procine Glycated hemoglobin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH

More information

Detection limit: grain, feed 500 ppb; milk 50 ppb; cream, cheese 5 ppb

Detection limit: grain, feed 500 ppb; milk 50 ppb; cream, cheese 5 ppb Product information Background Deoxynivalenol (DON) Deoxynivalenol, called vomitoxin, is a toxic metabolite mainly produced by Fusarium graminearum. It is mainly found in wheat, barley, corn and feed.

More information

1) A ph/mv meter or an ion meter, either line operated or portable.

1) A ph/mv meter or an ion meter, either line operated or portable. COLE-PARMER Sodium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS The Cole-Parmer Sodium Ion Electrodes are used to quickly, simply, accurately, and economically measure sodium ion concentrations

More information

Common Acids and Bases

Common Acids and Bases Common Acids and Bases Objective To determine the ph values of common household substances. Modules and Sensors Computer with NeuLog TM Software Data Cable, 10cm S98242-48ND USB Module - S98242-45ND ph

More information

Providing Vital Solutions for Your Laboratory

Providing Vital Solutions for Your Laboratory Test Menu General Chemistry Assays Albumin Bilirubin, Direct Bilirubin, Total Calcium Creatinine Glucose Iron, Total Magnesium Phosphorus Protein, Total Urea Nitrogen (BUN) Uric Acid Enzyme Assays Alanine

More information

RUO. DRG Amyloid beta Auto-Ab (serum) As of 14 Dec rm (Vers. 1.1)

RUO. DRG Amyloid beta Auto-Ab (serum) As of 14 Dec rm (Vers. 1.1) 1 INTRODUCTION Intended Use The DRG Amyloid β Auto-Ab ELISA is an enzyme immunoassay for measurement of autoantibodies against Amyloid Beta (Aβ) in human serum For research use only. Not for use in diagnostic

More information

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

cobas e 411 analyzer Confidence in every result.

cobas e 411 analyzer Confidence in every result. 17591_Brochure.qxd 11/1/06 10:46 AM Page 2 The cobas e 411 analyzer offers... Accurate Results ECL technology delivers excellent low-end sensitivity and broad dynamic ranges Quality Results Assay-specific

More information