Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

Size: px
Start display at page:

Download "Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements"

Transcription

1 Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements Experimental Design Doc. Martti Heinonen

2 Outline 1. Establishing traceability in measurements 2. Estimating uncertainty 2.1 Objectives and level of modelling 2.2 What should be covered by the analysis 3. Physical measurements - example 4. Chemical and biological measurements 4.1 Sampling 4.2 Example 2

3 1 Establishing traceability

4 Traceability Tower D U Measurement D U D U Calibration standard D U Reference standard D U Secondary standard D U Primary standard K C K C Measuring instrument SI System of Units K C K C K C K C Measurement uncertainty The tower is collapsed if any part of it is missing or incomplete i.e. there is no traceability unless all levels include all the characteristics of traceability At any level the measurement uncertainty can t be smaller than levels below. 4

5 Characteristics of Unbroken Traceability For each calibration of the chain: Uncertainty estimation Documented and generally acknowledged procedures, documented results Competence Calibration is valid for the application. (interval of calibrations, conditions etc.) P U V C SI System of Units 5

6 2 Estimating uncertainty 2.1 Objectives and level of modeling 2.2 What should be covered by the analysis? 2.3 Benefits from uncertainty analysis

7 6 steps to evaluating uncertainty 1) Measurement model: List essential input quantities (i.e. parameters x i having a significant effect on the result) and build up a mathematical model (function) showing how they are related to the final result: y = f(x 1,,, x i ) 2) Standard uncertainty: Estimate the standard uncertainty of each input quantity (x i ) 3) Use the model in uncertainty calculations: Determine the uncertainty due to standard uncertainty of each input quantity (x i ): u i (y) = c i u(x i ) 4) Correlation: Determine correlation between the input quantities (if relevant). 5) Calculate the combined standard uncertainty 6) Calculate the expanded uncertainty. 7

8 2.1 Objectives and level of modelling

9 Objectives of modelling to describe how the measurement result is calculated from input data (incl. measurement values, the data often include information from earlier measurements, specifications, calibration certificates etc. ) to show how various factors affect the result to provide a tool for calculating the estimate and the uncertainty 0.30 t x = tind + δ tcal + δtd + δtresol + δtg t x = (21,6 ± 0,3) C Ref - Ind ( C) Temperature ( C) 9

10 Level of modelling The model is never complete; approximations are needed: high accuracy more details in the model low accuracy rough approximations in the model A measurement model is never identical with a presentation of a physical phenomenon: All data used in the calculation of the estimate and the uncertainty should have corresponding input quantities in the model All input quantities of the model should be connected to data actually used in the calculations 10

11 Source of data: t x = t ind + δ t + δt + δt + δt Cal D resol G display calibration certificate long-term instability monitoring display earlier measurements (validated specifications) 11

12 Completeness of the model: RH ( p c ce w dind dc dd dr wg = 100 % rh ( p + δp d + δp ) e de ( t ) e w ( t Ind + δt + δt C + δt + δt D + δt + δt R ) + δe + δt G ) + δrh Calc S1 ews ( t) = 1 Pa exp[ + S S (/ t C, ) S (/ t C, ) t / C, S ln( t/ C+ 273, 15)] 5 2 RH = RH ind + δ RH + δrh C ni 12

13 2.2 What should be covered by the analysis?

14 Source of data: t x = t ind + δ t + δt + δt + δt Cal D resol G display calibration certificate long-term instability monitoring display earlier measurements (validated specifications) 14

15 Issues to be considered Measurement target/object Measurement method Measurement device / equipment Measurement conditions / environment Measurer 15

16 Issues to be considered Measurement target/object examples: gradients homogeneity disturbance due to the measurement temporal variations 16

17 Issues to be considered Measurement method examples: reproducibility variations affecting the target and/or the equipment representativity (number of measurements, measurement time, sampling frequency etc.) errors in sampling thermoelectric effects, non-ideal contacts, trace gases in tubing, contaminations, etc. 17

18 Issues to be considered Measurement device / equipment examples: calibration & reference standard drift (zero & full scale) resolution sensitivity non-linearity interaction of different parameters 18

19 Issues to be considered Measurement conditions / environment examples: ambient temperature ambient humidity ambient air velocity vibrations electric noise lighting magnetic field 19

20 Issues to be considered Measurer examples: skills: ability to obtain repeatable & reproducible & comparable results heat and moisture contamination 20

21 2.3 Benefits from uncertainty analysis

22 Improving quality Modelling = analysis of measurement Mathematical model shows explicitly how different factors affect the result and measurement process Uncertainty analysis (including the modelling) brings out weak points factors in which efforts should be focused to get most effective improvement in quality Benefitting in risk analysis 22

23 Innovating and planning By modelling you can test different methods before setting up an actual measurement system efficient and low-cost You can study the effects of both methods and devices on the measurement quality on the basis of existing data. You can judge if a new approach has potential to achieve the required accuracy level. 23

24 3 Physical measurements - Example

25 Example 3.1: Fluid temperature in a pipe Ω Measurement set-up: -Fluid: water -Pt-100 thermometer -immersed 5 cm -in a thermometer well -DMM for resistance measurements 25

26 Example 3.1 continuing: Ω Step 1: Measurement model - How we get temperature from the resistance? Calibration equation - What are the input quantities? - What is the measurement model? 26

27 Example 3.1 Input quantities Measurement target/object Error due to temperature gradients in the pipe δt G Fluid: Water & liquid in the thermometer well temperature difference between the fluid and the tip of the thermometer is negligible error due to different time constants is negligible Measurement method Thermometer is immersed only partly error due to heat conduction along the probe δt cond (disturbance due to the thermometer is negligible) Error in resistance measurement (wires etc.) δr ni 27

28 Example 3.1 Input quantities Measurement device / equipment Errors in the calibration equation of the Pt-100: δt cal Drift of the Pt-100 since the last calibration: δt drift Reading of the DMM: I R Calibration correction of the DMM: δr cal Drift of the DMM since the last calibration: δr drift Error due to non-ideal resolution of the DMM: δr resol Error due to non-linearity of the DMM: δr nl 28

29 Example 3.1 Input quantities Measurement conditions / environment Effect environmental conditions on the DMM: negligible Effect environmental conditions on the PRT & wiring: included in δt cond and δr ni Measurer: permanent set-up & automatic measurement the effect of a measurer is negligible 29

30 Example 3.1 Measurement model t 4 = a i= 0 i ( I R + i δ Rcal + δrdrift + δrresol + δrnl + δrni ) + δtcal + δtcond + δtdrift + δt G 30

31 Example 3.1 continuing: Ω Step 2: Standard uncertainty of the input quantities - How we estimate? 31

32 Example 3.1 continuing: 4 i t = ai ( I R + δ Rcal + δrdrift + δrresol + δrnl + δrni ) + δtcal + δtcond + δtdrift + δt i= 0 G Quantity, X i Estimate, x i Standard unc., u (x i ) Probability Notes on the determination i Description Value unit Value unit distribution of the standard uncertainty reading of 1 DMM Ohm Ohm normal display calibration 2 equation 0 C C normal taken from the calibration certificate calibration of 3 the DMM Ohm Ohm normal taken from the calibration certificate drift of the 4 Pt100 0 C C rectangular comparing last two calibration results drift of the 5 DMM 0 ohm Ohm rectangular comparing last two calibration results resolution of 6 the DMM 0 ohm Ohm rectangular display temp. 7 gradients 0 C C rectangular from specification (or thermal modelling) earlier measurement data, e.g. with calibration heat flow along 8 the Pt C C rectangular recording 9 method 0 Ohm Ohm validation results non-linearity of the DMM 0 C Ohm rectangular 10 calculated from the calibration certificate 32

33 Example 3.1 continuing: Ω Step 3: Effect on the combined uncertainty 33

34 Example 3.1 continuing: 4 i t = ai ( I R + δ Rcal + δrdrift + δrresol + δrnl + δrni ) + δtcal + δtcond + δtdrift + δt i= 0 G Quantity, X i Estimate, x i Standard unc., u (x i ) Probability Sensitivity coeff., c i Notes on the determination i Description Value unit Value unit distribution Value unit of the standard uncertainty reading of 1 DMM Ohm Ohm normal C/Ohm display calibration 2 equation 0 C C normal 1 taken from the calibration certificate calibration of 3 the DMM Ohm Ohm normal C/Ohm taken from the calibration certificate drift of the 4 Pt100 0 C C rectangular 1 comparing last two calibration results drift of the 5 DMM 0 ohm Ohm rectangular C/Ohm comparing last two calibration results resolution of 6 the DMM 0 ohm Ohm rectangular C/Ohm display temp. 7 gradients 0 C C rectangular 1 from specification (or thermal modelling) earlier measurement data, e.g. with calibration heat flow along 8 the Pt C C rectangular 1 recording 9 method 0 Ohm Ohm C/Ohm validation results non-linearity of the DMM 0 C Ohm rectangular 1 10 calculated from the calibration certificate 34

35 Example 3.1 continuing: Ω Step 4: Correlations: - Input quantities can be considered independent on each other 35

36 Example 3.1 continuing: Ω Step 5: Combined standard uncertainty Step6: Expanded uncertainty 36

37 Example 3.1 continuing: 4 i t = ai ( I R + δ Rcal + δrdrift + δrresol + δrnl + δrni ) + δtcal + δtcond + δtdrift + δt i= 0 G Quantity, X i Estimate, x i Standard unc., u (x i ) Probability Sensitivity coeff., c i Uncertainty contribution, u i Notes on the determination i Description Value unit Value unit distribution Value unit Value unit of the standard uncertainty reading of 1 DMM Ohm Ohm normal C/Ohm C display calibration 2 equation 0 C C normal C taken from the calibration certificate calibration of 3 the DMM Ohm Ohm normal C/Ohm C taken from the calibration certificate drift of the 4 Pt100 0 C C rectangular C comparing last two calibration results drift of the 5 DMM 0 ohm Ohm rectangular C/Ohm C comparing last two calibration results resolution of 6 the DMM 0 ohm Ohm rectangular C/Ohm C display temp. 7 gradients 0 C C rectangular C from specification (or thermal modelling) heat flow along 8 the Pt C C rectangular C recording 9 method 0 Ohm Ohm C/Ohm C validation results non-linearity of the DMM 0 C Ohm rectangular C 10 combined standard uncertainty: C Estimate: C Expanded uncertainty: C earlier measurement data, e.g. with calibration calculated from the calibration certificate Measurement r 82.1 C ± 0.3 C 37

38 Example 3.1 continuing: Ω Measurement result: (82.0 ± 0.3) C 38

39 Example 3.1 continuing: Ω Conclusions on factors affecting the result: - temperature gradients dominate the uncertainty 39

40 4 Chemical and biological measurements 4.1 Sampling 4.2 Example

41 4.1 Sampling

42 Objectives of sampling The objective is to get reliable information on the whole target of interest Most significant error sources: taking samples + handling and analysis of the samples Often several samples are needed to be taken in steps Sampling is often the dominating uncertainty component 42

43 Methods of sampling Sampling methods: random samples systematic samples (periodicity) representative sample combined sample divided sample (sample is too large for the analysis) layered sample 43

44 Sampling plan The sampling plan should state: size and number of samples locations, times/dates, sampling method handling of samples sample container; cleaning, closing, storage labelling and records requirements for the analysis environmental conditions during sampling other requirements (e.g. authorized persons) 44

45 Sources of uncertainty in sampling inhomogeneity of the material shape of particles instability of the material in time layers in the material weighting error number of samples errors in handling the samples errors in the analysis 45

46 Uncertainty of sampling Deviation of samples When analysed n b pieces from total number N of pieces and from each piece we take n w samples and we carry out n a analysis with each samples, then the standard deviation of the mean is: 2 N 1 σ = σ b + n n n ( nb ) σ w σ a nb N nbnw b w a σ w = deviation of analysed samples σ b = deviation of pieces σ a = deviation of analysis n t = total number of analyses =n b n w n a for homogenious material: σ w = 0 if all samples are analysed: N = n b 46

47 Further information on sampling Minimum number of sampling σ s = standard deviation R = permitted maximum error t = student-t factor n = 2 t σ R 2 s 2 Literature EN ISO/IEC 17025:2005, General requirements for the competence of testing and calibration laboratories. Sections 5.2.5, 5.7 Measurement uncertainty arising from sampling: A guide to methods and approaches, EURACHEM / CITAC Guide (2007)1st edition (can be downloaded at ISO 15189/2003: Medical laboratories -- Particular requirements for quality and competence ISO ,2(1993): Statistics 47

48 4.2 Example

49 Preparation of a Cd calibration standard for AAS (Atomic absorption spectroscopy) 1) Description of the measurement: Metal oxide contamination is removed with an acid mixture treatment. A volumetric flask of 100 ml is weighed with and without the purified metal inside. 1 ml of nitric acid (65 %m/m) and 3 ml ion-free water are added to the flask to dissolve the cadmium. Afterwards the flask is filled with ionfree water up to the mark and mixed by inverting the flask at least thirty times. The concentration is: c Cd = m P V m = mass of metal P = purity of metal, i.e. mass fraction V = volume of the liquid of the calibration standard 1) Based on the Example A1 presented in [1] 49

50 Preparation of a Cd calibration standard - continuing Step 1: Measurement model Supplier s certification Repeatability (dissolve efficiency) 50

51 Preparation of a Cd calibration standard - continuing Step 1: Measurement model Supplier s certification Repeatability (dissolve efficiency) c Cd = m P V = [( m Igross m Itare ) + δm V cal lin + δv + δm temp rep + δv + δm filling resol ] ( P + δp ) certif eff 51

52 Preparation of a Cd calibration standard - continuing Step 1: Measurement model c Cd = m P V [( m = Igross m Itare ) + δm m Igross, m Itare = balance readings with and without the metal δm lin = correction due to non-linearity of the balance δm rep, δm resol = correction due to non-ideal repeatability and resolution of the balance P = purity of the metal according to the supplier s certificate δp = correction due to the non-ideal dissolving V cal = the inner volume of the flask according to its calibration certificate δv temp = thermal expansion of the measured volume cal δv tilling = error in filling water up to the mark V lin + δv + δm temp rep + δv + δm filling resol ] ( P + δp ) certif eff 52

53 Preparation of a Cd calibration standard - continuing Step 2: Standard uncertainty of the input quantities Quantity, X i Estimate, x i Standard unc., u (x i ) Probability Notes on the determination i Description Value unit Value unit distribution of the standard uncertainty balance readings with the standard deviation of recorded 1 metal mg mg normal readings balance readings without 2 the metal 0 mg mg normal standard deviation of recorded readings after taring non-linearity of the 3 balance 0 mg mg rectangular According to calibration, the nonlinearity is less than ±0.03 mg non-ideal repeatability of 4 the balance 0 mg mg rectangular According to repeated measurements in calibration of the balance, the repeatability is less than ±0.02 mg non-ideal resolution of the 5 balance 0 mg mg rectangular 6 purity of the metal rectangular Manufacturer's certificate states that the purity of the metal is ± non-ideal dissolving rectangular According to repeated preparations the efficiency is larger than % 8 inner volume of the flask 100 ml 0.05 ml normal 9 thermal expansion of the measured volume 0 ml ml rectangular 10 filling error 0 ml ml normal The flask was calibrated by weighing with distilled water. The result was (100 ± 0.1) ml with k =2 Because the thermal expansion of glass is much smaller than water, only the water (2.1 x 10^-4 1/ C) is taken into account. Temperature variation is ± 4 C at maximum. standard deviation from ten fillings and weighings 53

54 Preparation of a Cd calibration standard - continuing Step 3: Effect on the combined uncertainty Sensitivity coefficients: For m Igross, m Itare δm lin, δm rep and δm resol : c 1 = c m Cd = c m Cd For P and δp : For V cal, δv temp and δv tilling : c c 3 2 = = c P c V Cd = Cd = c P Cd c V Cd 54

55 Preparation of a Cd calibration standard - continuing Step 3: Effect on the combined uncertainty Quantity, X i Estimate, x i Standard unc., u (x i ) Probability Sensitivity coeff., c i Notes on the determination i Description Value unit Value unit distribution Value unit of the standard uncertainty balance readings with the standard deviation of recorded 1 metal mg mg normal /l readings balance readings without 2 the metal 0 mg mg normal /l standard deviation of recorded readings after taring non-linearity of the 3 balance 0 mg mg rectangular /l According to calibration, the nonlinearity is less than ±0.03 mg non-ideal repeatability of 4 the balance 0 mg mg rectangular /l According to repeated measurements in calibration of the balance, the repeatability is less than ±0.02 mg non-ideal resolution of the 5 balance 0 mg mg rectangular /l 6 purity of the metal rectangular mg/l Manufacturer's certificate states that the purity of the metal is ± non-ideal dissolving rectangular mg/l According to repeated preparations the efficiency is larger than % 8 inner volume of the flask 100 ml 0.05 ml normal g/l^2 9 thermal expansion of the measured volume 0 ml ml rectangular g/l^2 10 filling error 0 ml ml normal g/l^2 The flask was calibrated by weighing with distilled water. The result was (100 ± 0.1) ml with k =2 Because the thermal expansion of glass is much smaller than water, only the water (2.1 x 10^-4 1/ C) is taken into account. Temperature variation is ± 4 C at maximum. standard deviation from ten fillings and weighings 55

56 Preparation of a Cd calibration standard - continuing Step 4: Correlations: Input quantities can be considered independent on each other c Cd = m P V [( m = Igross m Itare ) + δm V cal lin + δv + δm temp rep + δv + δm filling resol ] ( P + δp ) certif eff 56

57 Preparation of a Cd calibration standard - continuing Step 5 & 6: Combined standard and expanded uncertainty Quantity, X i Estimate, x i Standard unc., u (x i ) Probability Sensitivity coeff., c i Uncertainty contribution, u i Notes on the determination i Description Value unit Value unit distribution Value unit Value unit of the standard uncertainty balance readings with the standard deviation of recorded 1 metal mg mg normal /l mg/l readings balance readings without 2 the metal 0 mg mg normal /l mg/l standard deviation of recorded readings after taring non-linearity of the 3 balance 0 mg mg rectangular /l mg/l According to calibration, the nonlinearity is less than ±0.03 mg non-ideal repeatability of 4 the balance 0 mg mg rectangular /l mg/l According to repeated measurements in calibration of the balance, the repeatability is less than ±0.02 mg non-ideal resolution of the 5 balance 0 mg mg rectangular /l mg/l 6 purity of the metal rectangular mg/l mg/l Manufacturer's certificate states that the purity of the metal is ± non-ideal dissolving rectangular mg/l mg/l According to repeated preparations the efficiency is larger than % 8 inner volume of the flask 100 ml 0.05 ml normal g/l^ mg/l 9 thermal expansion of the measured volume 0 ml ml rectangular g/l^ mg/l 10 filling error 0 ml ml normal g/l^ mg/l combined standard uncertainty: 0.83 mg/l Estimate: mg/l Expanded uncertainty: 1.66 mg/l The flask was calibrated by weighing with distilled water. The result was (100 ± 0.1) ml with k =2 Because the thermal expansion of glass is much smaller than water, only the water (2.1 x 10^-4 1/ C) is taken into account. Temperature variation is ± 4 C at maximum. standard deviation from ten fillings and weighings 57

58 Preparation of a Cd calibration standard - continuing Thus, the Cd content of the calibration standard prepared for an AAS was: (1003 ± 2) mg/l 58

59 References and literature [1] EURACHEM/CITAC Guide CG 4, Quantifying Uncertainty in Analytical measurement ( LITERATURE ISO/IEC Guide 99-12:2007, International Vocabulary of Metrology Basic and General Concepts and Associated Terms, VIM International vocabulary of metrology Basic and general concepts and associated terms (VIM), 3rd ed., JCGM 200:2008 (can be downloaded at Metrology - in short, 3rd edition, EURAMET 2008, 84 p. ( JCGM 100:2008, Evaluation of measurement data Guide to the expression of uncertainty in measurement, First edition, JCGM 2008 ( European cooperation for Accreditation, EA-4/02 Expression of the Uncertainty of Measurement in Calibration, December ( UKAS M3003, The Expression of Uncertainty and Confidence in Measurement S. A. Bell, A beginner's guide to uncertainty in measurement, Measurement Good Practice Guide No. 11, (Issue 2), National Physical Laboratory 2001, 41 p. ( MIKES has published several guides in Finnish on the uncertainty estimations in different fields. 59

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC 17025 Table of contents 1 GENERAL... 2 2 THE STANDARD SIST EN ISO/IEC 17025... 2 3 SA'S POLICY IN IMPLEMENTING

More information

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012 Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability

More information

What is measurement uncertainty?

What is measurement uncertainty? What is measurement uncertainty? What is measurement uncertainty? Introduction Whenever a measurement is made, the result obtained is only an estimate of the true value of the property being measured.

More information

Introduction to the evaluation of uncertainty

Introduction to the evaluation of uncertainty Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials SECTION 1 Introduction to the evaluation of uncertainty F A Kandil National Physical Laboratory

More information

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014 Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

Document No: TR 12 Issue No: 1

Document No: TR 12 Issue No: 1 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Prepared by: SADCAS Technical Manage Approved

More information

Calibration of Temperature Block Calibrators

Calibration of Temperature Block Calibrators European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)

More information

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014) SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February

More information

Since the publication of the ISO Guide to the Expression

Since the publication of the ISO Guide to the Expression DE SILVA: JOURNAL OF AOAC INTERNATIONAL VOL. 86, NO. 5, 003 1077 SPECIAL GUEST EDITOR SECTION Uncertainty of Analytical Determinations GALAPPATTI M.S. DE SILVA 4, PB Alwis Perera Mawatha, Katubedda, Sri

More information

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results David MILDE Traceability Validation Measurement Uncerainty Outline Introduction to 3 pillars Recommended

More information

CALIBRATION REPORT FOR THERMOHYDROMETER

CALIBRATION REPORT FOR THERMOHYDROMETER CALIBRATION REPORT FOR THERMOHYDROMETER SAMPLE CUSTOMER S/N: XXXXX Report Number: Y205188 Page 1 of 5 Report No. Y205188 ICL CALIBRATION LABORATORIES, INC. ISO/IEC 17025 and ANSI/NCSL Z540-1 accredited

More information

Measurement Uncertainty, March 2009, F. Cordeiro 1

Measurement Uncertainty, March 2009, F. Cordeiro 1 Measurement Uncertainty, March 2009, F. Cordeiro 1 Uncertainty Estimation on the Quantification of Major Proteins in Milk by Liquid Chromatography Contribution to a Chemical Reference Measurement System

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION Page 1 of 5 CERTIFICATE OF CALIBRATION 1K Thermistor CERTIFICATE NO.: MANUFACTURER: Measurement Specialties DATE OF CALIBRATION: 26-Apr-17 TC15249 CUSTOMER: AMBIENT TEMPERATURE: (21 ± 3) C JOB NUMBER:

More information

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

The Uncertainty of Reference Standards

The Uncertainty of Reference Standards 2009 Cerilliant Corporation The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations and Vendor Certifications Authors: Ning Chang PhD, Isil

More information

Example A1: Preparation of a Calibration Standard

Example A1: Preparation of a Calibration Standard Suary Goal A calibration standard is prepared fro a high purity etal (cadiu) with a concentration of ca.1000 g l -1. Measureent procedure The surface of the high purity etal is cleaned to reove any etal-oxide

More information

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany Guidelines on the Calibration of Temperature Block Calibrators Authorship and Imprint This document was developed by the EURAMET e.v., Technical Committee for Thermometry. Authors: Yves Hermier (LNE-INM,

More information

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015 Uncertainty of Measurement (Analytical) Maré Linsky 14 October 015 Introduction Uncertainty of Measurement GUM (Guide to the Expression of Uncertainty in Measurement ) (Analytical) Method Validation and

More information

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors 1. Scope... 2 2. Introduction... 2 3. Selection of thermocouples

More information

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

Vocabulary of Metrology

Vocabulary of Metrology Vocabulary of Metrology For Understating Uncertainty and Traceability Prepared by: Kim, Sang Ho Engineering Team Leader Modal Shop, Inc A PCB Group Company SI National Metrology Institute Secondary Calibration

More information

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

R SANAS Page 1 of 7

R SANAS Page 1 of 7 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Approved By: Chief Executive Officer:

More information

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. ESTIMATION OF UNCERTAINTY OF MEASUREMENT POLICY...

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

Comparability and Traceability Point of view from a metrological institute

Comparability and Traceability Point of view from a metrological institute Comparability and Traceability Point of view from a metrological institute Ed W.B. de Leer NMi Van Swinden Laboratory Delft, The Netherlands 1 My Discussions in Metrologyland 2 The CCQM in 1996 3 Traceability

More information

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. EVALUATION OF MEASUREMENT UNCERTAINTY POLICY... 5

More information

Certificate of Accreditation to ISO/IEC 17025:2005

Certificate of Accreditation to ISO/IEC 17025:2005 United States Department of Commerce National Institute of Standards and Technology Certificate of Accreditation to ISO/IEC 17025:2005 NVLAP LAB CODE: 200508-0 Insco Metrology, Inc. Miami, FL is accredited

More information

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number:

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION. Peter B. Crisp. Fluke Precision Measurement Ltd, 52 Hurricane Way, Norwich, UK

UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION. Peter B. Crisp. Fluke Precision Measurement Ltd, 52 Hurricane Way, Norwich, UK UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION Peter B. Crisp Fluke Precision Measurement Ltd, 5 Hurricane Way, Norwich, UK Introduction NAMAS Accredited laboratories are required to follow recommended

More information

Uncertainty of Measurement A Concept

Uncertainty of Measurement A Concept Uncertainty of Measurement A Concept Shweta V. Matey 1, Dr. Nitin K. Mandavgade 2, and Dr. R.R. Lakhe 3 1 (Asst. Professor, Dept. of Mechanical Engg, Lokmanya Tilak College of Engineering, Mumbai University,

More information

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT# 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement Publication Reference EA-10/11 EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical PURPOSE This document has been produced by EA as a means of giving advice for calibrating

More information

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium, 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018 Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018 Guidelines on the Calibration of Automatic Instruments for

More information

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands Calibration, traceability, uncertainty and comparing measurement results Workshop 16 March 2015 VSL, Delft The Netherlands 1 Calibration Calibrating an instrument involves comparing the measurements of

More information

Determination of Cadmium Release from Ceramic Ware by Atomic Absorption Spectrometry

Determination of Cadmium Release from Ceramic Ware by Atomic Absorption Spectrometry This is the example A5 of the EURACHEM / CITAC Guide "Quantifying in Analytical Measurement", Second Edition. The amount of released cadmium from ceramic ware is determined using atomic absorption spectrometry.

More information

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE Michael Bair and Pierre Delajoud 2002 DEC 10 Revised 2004 MAR 26 FORWARD FPG8601 is a pressure standard designed

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Total Mercury Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-106-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY 15th EFLM Continuing Postgraduate Course in Clinical Chemistry and Laboratory Medicine How to assess the quality of your method? Zagreb, 24-25 October 2015 HOW TO ASSESS THE MEASUREMENT UNCERTAINTY Centre

More information

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST DEUTSCHER KALIBRIERDIENST Guideline DKD-R 5-7 Calibration of Climatic Chambers Edition 07/2004 English translation 02/2009 Page 1 of 31 Published by the Accreditation Body of the Deutscher Kalibrierdienst

More information

Uncertainty sources of reference measurement procedures for enzymes

Uncertainty sources of reference measurement procedures for enzymes Medical School Hannover Institute for Clinical Chemistry Director: Prof. Dr. K. Brand Uncertainty sources of reference measurement procedures for enzymes R. Klauke and G. Schumann Renate Strache Rainer

More information

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,

More information

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) Dr. Steven D. Phillips Precision Engineering Division National Institute

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

CHEM 334 Quantitative Analysis Laboratory

CHEM 334 Quantitative Analysis Laboratory Calibration of Volumetric Glassware Introduction Volumetric glassware is a class of glass vessels that are calibrated to contain or deliver certain volumes of substances. Graduated cylinders, pipettes

More information

Reproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process

Reproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process Flowchart for Nordtest Method (a) Specify measurand Quantify components for within lab reproducibility A control samples B possible steps, not covered by the control sample Quantify bias components Convert

More information

Practical and Valid Guidelines for Realistic Estimation of Measurement Uncertainty in Pesticide Multi-Residue Analysis*

Practical and Valid Guidelines for Realistic Estimation of Measurement Uncertainty in Pesticide Multi-Residue Analysis* Practical and Valid Guidelines for Realistic Estimation of Measurement Uncertainty in Pesticide Multi-Residue Analysis* Antonio Valverde Pesticide Residue Research Group University of Almería * This presentation

More information

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com CERTIFICATE OF ANALYSIS tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 ACCREDITATION / REGISTRATION

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

CERTIFICATE OF. Single Analyte Custom Grade Solution g/ml (measured at 20 ± 4 ºC) Graviametric NIST SRM Lot Number: See Sec. 4.

CERTIFICATE OF. Single Analyte Custom Grade Solution g/ml (measured at 20 ± 4 ºC) Graviametric NIST SRM Lot Number: See Sec. 4. CERTIFICATE OF ANALYSIS 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 ACCREDITATION / REGISTRATION

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS. A.Gnanavelu

CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS. A.Gnanavelu CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS A.Gnanavelu UNCERTAINTY (U) Every measurement has error which is unknown and unknowable. This unknown error is called measurement uncertainty. Types of Error

More information

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR Davor Zvizdic, Tomislav Stasic, Lovorka

More information

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,

More information

The uncertainty evaluation for the hydrometer calibration at NMISA. Presenter: Bongani Ndlovu Date: 09/10/2018

The uncertainty evaluation for the hydrometer calibration at NMISA. Presenter: Bongani Ndlovu Date: 09/10/2018 The uncertainty evaluation for the hydrometer calibration at NMISA Presenter: Bongani Ndlovu Date: 09/10/2018 A HYDROMETER Thin graduated hollow tube Weighted at bottom end to make it float upright USED

More information

ANALYSIS CERTIFICATE OF

ANALYSIS CERTIFICATE OF CERTIFICATE OF ANALYSIS 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 INORGANIC VENTURES is an

More information

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR M. Araya 1, D. Almeida 2 1 National Laboratory of Temperature of Chile, LCPNT, Santiago, Chile 2 Ecuadorian Standardization

More information

Practical Interpretation of Metrological Traceability on Gauge Block as to VIM 3

Practical Interpretation of Metrological Traceability on Gauge Block as to VIM 3 MAPAN - Journal Practical of Metrology Interpretation Society of Metrological India, Vol. 26, Traceability No. 1, 2011; on pp. Gauge 79-86Block as to VIM 3 ORIGINAL ARTICLE Practical Interpretation of

More information

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Hards Laboratories cc, t/a Technology Solutions

More information

A basic introduction to reference materials. POPs Strategy

A basic introduction to reference materials. POPs Strategy A basic introduction to reference materials POPs Strategy 2009-2010+ A tutorial 16 September 2009 Angelique Botha R&D metrologist Contents Why do we need reference materials? comparability of results metrological

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION Issued By Transmille Ltd. Date of Issue 20 January 2016 0324 Approved Signatory Transmille Ltd. Unit 4, Select Business Centre Lodge Road Staplehurst, Kent. TN12 0QW. TEL 01580 890700 FAX 01580 890711

More information

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB)

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB) JAB NOTE4 ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) nd edition: January 5 01 1 st edition: March 5 00 Japan Accreditation Board (JAB) Initial edition: 00-0-5 1/ nd

More information

AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS

AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS WHITE PAPER AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS The Energy Conservatory Minneapolis, MN Introduction: The purpose of this document is to explain the details of the accuracy specifications

More information

Examples of Method Validation Studies Conducted in Different Economies

Examples of Method Validation Studies Conducted in Different Economies Examples of Method Validation Studies Conducted in Different Economies Template for group discussion National Metrology Laboratory, Malaysia 8th APMP TCQM-DEC MiC Workshop, Kobe, Japan December 2011 Description

More information

1551A Ex/1552A Ex Intrinsically Safe Stik Thermometer. The new gold standard of industrial temperature calibration

1551A Ex/1552A Ex Intrinsically Safe Stik Thermometer. The new gold standard of industrial temperature calibration 1551A Ex/155A Ex Intrinsically Safe Stik Thermometer The new gold standard of industrial temperature calibration Designed to go where you work Finally, a digital substitute for your mercury-in-glass thermometers!

More information

CERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC)

CERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC) CERTIFICATE OF ANALYSIS 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 ACCREDITATION / REGISTRATION

More information

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES XVII IMEKO World Congress Metrology in the 3rd Millennium June 7, 003, Dubrovnik, Croatia MODEL OR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION O THERMOCOUPLES Georges Bonnier, Eliane Renaot, DavorZvizdic,

More information

EURAMET Project 858: Hydrostatic weighing exchange of experiences. Project outline

EURAMET Project 858: Hydrostatic weighing exchange of experiences. Project outline EURAMET Project 858: Hydrostatic weighing exchange of experiences Project outline A bilateral comparison has been carried out between the national measurement institutes INRIM, Italy and IPQ, Portugal,

More information

Introduction to Blackbody Sources

Introduction to Blackbody Sources Introduction to s This section contains dedicated blackbody sources for low uncertainty calibration of infrared thermometers. A range of portable primary blackbody sources combine high emissivity with

More information

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3103a Lot Number:

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3103a Lot Number: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Traceable UV-Vis Reference Materials

Traceable UV-Vis Reference Materials UV-Vis Spectroscopy Traceable UV-Vis Reference Materials For Regulatory Compliance Introduction Many laboratories, particularly in the pharmaceutical industry, are under increasing pressure to show proof

More information

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES Proceedings of the ASME 2014 Power Conference POWER2014 July 28-31, 2014, Baltimore, Maryland, USA POWER2014-32205 UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES Michael S. Hering Rosemount DP

More information

CERTIFICATE OF. Single Analyte Custom Grade Solution. ICP Assay NIST SRM 3120a Lot Number: Calculated NIST SRM Lot Number: See Sec. 4.

CERTIFICATE OF. Single Analyte Custom Grade Solution. ICP Assay NIST SRM 3120a Lot Number: Calculated NIST SRM Lot Number: See Sec. 4. 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com CERTIFICATE OF ANALYSIS tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 ACCREDITATION / REGISTRATION

More information

The stimation of Uncertainty by the Utilization of Validation and Quality Control Data; Lead in Gasoline by AAS

The stimation of Uncertainty by the Utilization of Validation and Quality Control Data; Lead in Gasoline by AAS The stimation of ncertainty by the tilization of Validation and Quality Control Data; Lead in Gasoline by AAS Stefica Celap, Ina Rafinerija nafte Rijeka, Laboratorij, rinj bb, e-mail:stefica.celap@ina.hr

More information

Traceable Mass Determination and Uncertainty Calculation

Traceable Mass Determination and Uncertainty Calculation Traceable Mass Determination and Uncertainty Calculation Falko Hilbrunner, Thomas Fehling, Siego Mühlich, Mario Schreiber Sartorius Lab Instruments GmbH Weender Landstrasse 94-108 Goettingen, Germany ABSTRACT

More information

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com CERTIFICATE OF ANALYSIS tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 ACCREDITATION / REGISTRATION

More information

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000 Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000 Welcome to the Isotech Journal of Thermometry Fundamentals of Thermometry Part 1 Temperature Scales Practical Calibration of Thermometers

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Automated volume measurement for weihts using acoustic volumeter

Automated volume measurement for weihts using acoustic volumeter IMEKO 20th TC3, 3rd TC16 and 1st TC22 International Conference Cultivating metrological knowledge 27th to 30th November, 2007. Merida, Mexico. Automated volume measurement for weihts using acoustic volumeter

More information

Precision estimated by series of analysis ISO and Approach Duplicate Approach

Precision estimated by series of analysis ISO and Approach Duplicate Approach Agenda Item 9 JOINT FAO/WHO FOOD STANDARDS PROGRAMME CODEX COMMITTEE ON METHODS OF ANALYSIS SAMPLING Thirty-seventh th Session Budapest, Hungary, 6 February 016 (Comments prepared by the German Delegation)

More information

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE Publication Reference EA-10/13 EA Guidelines on the Calibration of Temperature Block Calibrators PURPOSE This document been produced by EA to improve the harmonisation in the calibration of temperature

More information

Designed to go where you work

Designed to go where you work Designed to go where you work Finally, a digital substitute for your mercury-in-glass thermometers! Accurate and repeatable to ± 0.05 C over its full range, the 55A/55A Stik Thermometer is the new gold

More information

Guide for volume determination within the scope of reference measurement procedures in medical reference measurement laboratories, Part 1

Guide for volume determination within the scope of reference measurement procedures in medical reference measurement laboratories, Part 1 Page 1 Translation of eitfaden für die Volumenbestimmung bei Referenzmessprozeduren in medizinischen Referenzlaboratorien, Teil 1: Kalibrierflüssigkeit Wasser, PTB-Mitteilungen 112 (2002) No. 2, pp.139-149

More information

Infrared Thermometer Calibration A Complete Guide

Infrared Thermometer Calibration A Complete Guide Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate

More information

A Statistical Approach to Reporting Uncertainty

A Statistical Approach to Reporting Uncertainty A Statistical Approach to Reporting Uncertainty on Certified Values of Chemical Reference Materials for Trace Metal Analysis Nimi Kocherlakota, Ralph Obenauf, and Robert Thomas This article discusses an

More information

Calibration and verification: Two procedures having comparable objectives and results

Calibration and verification: Two procedures having comparable objectives and results Calibration and verification: Two procedures having comparable objectives and results KLAUS-DIETER SOMMER, Landesamt für Messund Eichwesen Thüringen (LMET), Germany SAMUEL E. CHAPPELL, Consultant, Formerly

More information

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment

More information

Essentials of expressing measurement uncertainty

Essentials of expressing measurement uncertainty Essentials of expressing measurement uncertainty This is a brief summary of the method of evaluating and expressing uncertainty in measurement adopted widely by U.S. industry, companies in other countries,

More information

Humidity Calibration Solutions

Humidity Calibration Solutions Instruments Humidity Solutions www.michell.com Instruments Humidity Solutions The Importance of Regular The reliable operation of a hygrometer and indeed any measuring instrument, can only be verified

More information

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L Supplementary comparison SIM.M.FF-S12 Final Report for Volume of Liquids at 20 L Pilot: M. Maldonado (CENAM, Mexico) mmaldona@cenam.mx km 4.5 Carretera a los Cués, El Marqués, Qro 76241, Mexico Participants:

More information

Guidelines on the Calibration of Non-Automatic Weighing Instruments

Guidelines on the Calibration of Non-Automatic Weighing Instruments European Association of National Metrology Institutes Guidelines on the Calibration of Non-Automatic Weighing Instruments EURAMET/cg-18/v.01 Previously EA-10/18 July 007 Calibration Guide EURAMET/cg-18/v.01

More information

Hiroshi Sawai TKG Laboratory Center International Standardisation Office

Hiroshi Sawai TKG Laboratory Center International Standardisation Office Achieving ISO17025 Accreditation For Au Assay by ICP-OES Using Au-Reference Materials Hiroshi Sawai TKG Laboratory Center International Standardisation Office TANAKA KIKINZOKU KOGYO K.K. 7-8 March 2011

More information

Draft EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement. Second Edition. Draft: June 1999

Draft EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement. Second Edition. Draft: June 1999 Draft EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition Draft: June 1999 Prepared by the EURACHEM Measurement Uncertainty Working Group in collaboration with members

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Palenque 1 A, Bello Horizonte,. (Hereinafter called

More information

PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS

PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS XIX IMEKO World Congress Fundamental and Applied Metrology September 611, 009, Lisbon, Portugal PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS Adriana Vâlcu 1, Sterică Baicu 1 National Institute

More information

Guide to the Expression of Uncertainty in Measurement (GUM)- An Overview

Guide to the Expression of Uncertainty in Measurement (GUM)- An Overview Estimation of Uncertainties in Chemical Measurement Guide to the Expression of Uncertainty in Measurement (GUM)- An Overview Angelique Botha Method of evaluation: Analytical measurement Step 1: Specification

More information