Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Size: px
Start display at page:

Download "Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)"

Transcription

1 Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with an analytical method called direct potentiometry. Students will use direct potentiometry to determine the content of fluorides in a sample of river water, black tea infusion and tap water. 1. Introduction Potentiometry is an electroanalytical method, in which the concentration (activity) of ionic species is measured using an electrochemical cell consisting of two special electrodes. One of these electrodes is called a reference electrode, the other is known as indicating or sensing electrode. When these two electrodes are in contact with the electrolyte (a sample to be tested) they form an electrochemical (galvanic) cell and each of electrodes develop a property called potential1. The most important feature of the cell is the difference of potentials (E) between the indicating (Eind) and reference (Eref) electrodes: (Eq 1) This difference of potentials is sometimes called a cell voltage or electromotive force (EMF). Illustration 1: Instrumental setup for determination of fluorides using; (A) two separate electrodes, (B) so called "combined" electrode. The potential of a single electrode depends on number of variables and can be described by Nernst's equation: 1 It is impossible to measure the potential of a single electrode, but we can measure the difference of potentials. Therefore the galvanic cell consist of two electrodes.

2 (Eq 2) where: Ex Eo R T z F ln aox, ared potential of the electrode standard potential of the electrode (a constant) universal gas constant absolute temperature (in Kelvin degrees) charge of the ionic species Faraday's constant natural logarithm (with base e= ) electrochemical activity of the ionic species The above equation (Eq 2) may be directly applied to the redox based electrodes, i.e. when a reductionoxidation reaction is the source of potential and two forms of ionic species, at different oxidation states, are present. The copper electrode may serve as an example: (Eq 3) In this case the Nernst's equation will look like this: (Eq 4) When working with membrane ion selective electrodes (ISE), no redox reactions involving ions of interest occur. ISE's potential develop due to the fact that ions to be measured have different concentrations at both sides of the membrane being the part of ISE. This membrane is the most important part of ISE. It is responsible for selectivity and sensitivity of ISE. The membrane may be manufactured in many forms and from different materials. It can be made of glass, a single crystal, many small crystals or a polymeric material. In case of fluoride ion selective electrodes, the membrane is commonly made of single crystal of lanthanum fluoride (LaF3) doped with some europium fluoride (EuF2). In the laboratory practice the potential of ISE's is described by somewhat simplified Nernst's equation: (Eq 5) where: Ex K S cx log potential of the electrode sensitive to ion x a constant slope of the electrode's characteristics concentration of ion x (mol dm3) logarithm with base 10 (common logarithm)

3 The potential of the indicating electrode is (due to its construction), directly related to the concentration (activity) of the ions to be measured, while the reference electrode is always chosen in a such way that its potential is constant and independent on sample composition. Therefore the resulting cell voltage is dependent only on the concentration of ionic species of interest (fluorides, in case of this laboratory exercise). 2. Practical considerations There are two aspects of potentiometric measurements that have to be taken into account in practice. First is that equation 5 is only valid for very dilute solutions or solutions of constant ionic strength. The second is that electrodes respond only to free ions. Therefore some measures have to be taken to make measurement meaningful and reliable. In case of fluorides (F) determination this is achieved with the help of Total Ionic Strength Adjustment Buffer (TISAB). TISAB is added to all samples to keep the ionic strength at a high and constant value. The TISAB also contains a buffer to maintain the ph of the samples at a fixed value (below ph of around 5, significant portion of F ions will be replaced by HF or HF2). Additionally, the TISAB contains a complexing agent (CDTA) which complexes multivalent cations (eg. Al3+) which otherwise can interfere with the fluorides determination. 3. Instructions 3.1. Labware potentiometer fluoride ISE (combined) 100 ml volumetric flasks 50 ml polyethylene containers pipette 10 ml pipette 25 ml magnetic stirrer with a stirring bar (6 pieces) (8 pieces) (8 pieces) 3.2. Reagents standard solution of F (1 mg/ml) TISAB solution Prepare calibration solutions: Use a dedicated pipette for each solution. Stock solution: add 20 ml of fluoride standard solution (1 mg/ml) to a 100 ml volumetric flask. Calibration solution No. 1: add 10 ml of stock solution to a 100 ml volumetric flask. Make up to the

4 mark with distilled water, close with a stopper and mix well. Calculate molar concentration of this solution. Calibration solution No. 2: add 50 ml of calibration solution No. 1 to a 100 ml volumetric flask. Calibration solution No. 3: add 20 ml of calibration solution No. 2 to a 100 ml volumetric flask. Calibration solution No. 4: add 50 ml of calibration solution No. 3 to a 100 ml volumetric flask. Calibration solution No. 5: add 20 ml of calibration solution No. 4 to a 100 ml volumetric flask. Prepare calibration curve: Add exactly 20 ml of a calibration solution and 20 ml of TISAB solution to a clean and dry polyethylene container. Carefully put the stirring bar in the mixture and place the container on the stirrer. Start mixing the content of the container, then immerse the tip of fluoride ISE in the solution. Wait for readout to stabilise and write it down. Repeat the procedure for all calibration solutions. Start from the most dilute one. Each students group will perform full calibration of their ISE. Wash the stirring bar and the ISE with distilled water after each measurement. After washing remove excess of water from ISE and stirring bar by blotting with an absorbent tissue (e.g. Kleenex). Draw the calibration curve for your ISE, i.e. the graph illustrating the relationship between the measured signal and log10 of molar concentration of fluorides. Analyse your samples: Note (or weigh) the mass of tea leaves being used (most commonly it is between 1.5 and 2 g). Prepare the tea infusion in usual way, soaking tea leaves for 5 minutes. Cool the infusion to the room temperature before taking measurements. Analyse your samples starting with tap water. Add exactly 20 ml of tap water sample and 20 ml of TISAB solution to a clean and dry polyethylene container. Carefully put the stirring bar in the mixture and place the container on the stirrer. Start mixing the content of the container, then immerse the tip of fluoride ISE in the solution. Wait for readout to stabilise and write it down. Wash the stirring bar and the ISE with distilled water after each measurement. After washing remove excess of water from ISE and stirring bar by blotting with an absorbent tissue (e.g. Kleenex). Repeat the procedure for other samples (river water, tea infusion).

5 Illustration 2: River water origin, Gdańsk, Sobieszewo, Martwa Wisła (North: 54 20' 29.21", East: 18 49' 54.07") Report Shortly describe the laboratory experiment, present all measured data and graphs. Discuss the impact of the fluorides on the environment and human health. Determine slope of the electrode's characteristics. How much does it differ from the theoretical (Nernstian) value? Using calibration curve of your ISE, determine the fluoride concentration in each of the samples. Exchange your result with the results obtained by other subgroups and provide the mean and 95% confidence interval for fluorides content in each sample. Provide an example of detailed calculations for at least one sample. Compare your results with a data from scientific literature do not use popular web sites as they frequently publish unproven, poor quality data. Try to use e.g. Google Scholar, Elsevier's Science Direct or In most cases these search engines will provide you with full texts when used form within GUT network. Each member of laboratory group must compare its group's results with one unique literature source. No duplications of sources between and within subgroups will be allowed! The faster you prepare your report the bigger choice you have. Give the name of the person doing comparison. Quality of conclusions drawn by a given student will affect its grade. The nopenalty period for returning and acceptance of a report is one week. After this time each started week of delay will result with 0,25 grade malus. This means that a report accepted after more than 9 weeks will be considered as nonexistent. The only accepted form of report submission is a PDF file sent by to: wasia@pg.gda.pl

6 User maintenance and troubleshooting Symptom Possible cause Remedy Drift Junction blocked Remove and clean sleeve Noisy Inaccurate readings or poor reproducibility Membrane unclean or contaminated Poor connection to meter Junction not immersed fully Insufficient electrolyte or bubble Contaminated standard(s) Drift since last calibration Improperly made standards Temperature differences between calibration and sample measurement No TISAB used Incorrect ph Variations in measurement procedure between calibration and measurement Remove sleeve. Gently polish membrane with fine polishing agent Check connection Lower electrode into solution below junction Refill electrolyte Make up fresh calibration standards and recalibrate Recalibrate using freshly prepared standards Recalibrate using freshly made standards Ensure calibration and sample measurements are done at the same temperature Add TISAB buffer Ensure ph is between 5 7 Ensure stirring rates, temperature, and other conditions are the same in calibration and measurement steps Warranty IJ electrodes have a warranty of 12 months from date of purchase. Any electrode found to be faulty due to manufacture within this time will be replaced. Ionode reserves the right to limit or modify product warranty if it is deemed that the electrode has been used in unsuitable applications. Electrodes with broken stems, connectors or cables are not covered by warranty. Electrode life may be reduced in chemically aggressive or abrasive samples and at high temperatures. Visit ionode.com for more details. All specifications and values are subject to change without notice Images are illustrative only. Parameter Concentration range Temperature range 0 60ºC ph range Reference type Sensor materials Body and sleeve Overall length Barrel diameter Cable length Operating Range 0.02 to 19,000 mg/l 5 7 ph Double Junction Ag/AgCl/refillable Solid state crystal Polypropylene/PMMA 150mm 12mm 1m standard, longer to order. Maximum 20m Operators Manual ShortForm Intermediate Junction Series IJ F FLUORIDE ION SELECTIVE ELECTRODE Connector BNC standard, others on request GOOD CHEMISTRY

7 Introduction This guide contains the basic information for proper use of your new Fluoride Ionselective electrode. Preparation IJ series electrodes are shipped without sleeve electrolyte, and must be filled prior to use. To fill, hold the electrode by the sleeve and gently ease off the rubber wetting cap. Prepare as follows: Invert the electrode. Hold the 2. Fill the annular space with gel electrode just below the sleeve or electrolyte to approximately and with careful rotation and half to three quarter full. pulling along the axis of the electrode, remove the sleeve. DO NOT BEND Slide the sleeve back onto the electrode ensuring the black Oring is well seated within the electrode body. Do not exert sideways force. Any excess electrolyte will be expelled from the end of the electrode through the ground junction. Ensure there are no air bubbles in the sleeve.wash off any excess electrolyte before use. Application The Ionode IJF Ion selective electrode responds reversibly to Fluoride ions. The limit of detection is 0.02 ppm F. The response is theoretical down to approximately 0.2 ppm Interfering Ions The major interference is OH. For best results, ensure measurement is done in the ph range of 5 7 ph (optimally pH). In common applications such as drinking water analysis, this is easily achieved with the use of Tisab (Total Ionic Strength Adjuster Buffer). Cleaning If the membrane is poisoned by interferences, the surface may be carefully polishing with a fine polishing agent such as 3µm diamond paste. Polishing the membrane gently with fluoridated toothpaste can also restore sensitivity. Organic contaminants can be removed with ethanol. DO NOT use the electrode in chlorinated hydrocarbons. Routinely remove the sleeve and replace the potassium chloride electrolyte. Electrolyte Replacement The IJF has an inbuilt double junction Ag/AgCl with a replaceable sleeve electrolyte. Saturated potassium chloride is suitable for most applications. Calibration Standards Standard solutions of Fluoride Ion should preferably bracket the expected measurement range. For example, to determine Fluoride ions in drinking water (expected value around 1 ppm), it is normally sufficient to calibrate with 0.2 and 2.0 ppm Fluoride standards. Calibration Procedure Make up F calibration standards in distilled water bracketing the expected measurement range. Starting with the lowest concentration first, measure out 20ml of standard and add 20ml Tisab. Insert the electrode after rinsing, stir, and note the potential in mv. Perform this (carefully rinse between measurements to eliminate carry over) on each of the standards. At room temperature, a slope should be obtained of around 55mV/decade. (from 0.2ppm upwards) Potential mv Typical Calibration Curve Concentration PPM Sample Measurement Follow the same basic procedures as calibration, substituting the calibration standard for your sample. It is important to use the same stirring conditions, temperature, etc for best results. Use the same ratio of sample/isa as used in the calibration step. Read off the concentration from the graph or directly from Ion meters. Visit ionode.com for more details.

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with

More information

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required

More information

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste Experiment ISE: Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste 67 You have been hired by the government to check the fluoride concentration labelling on some major

More information

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0) 150mV 300mV Mono/Reference Combination S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) 1376 340 456 Fax: +44 (0) 1376 340 453 Email:

More information

Operating Instructions for Ammonium ISE Specifications

Operating Instructions for Ammonium ISE Specifications Operating Instructions for Ammonium ISE Specifications Range: The Ammonium Electrode responds to uncomplexed ion activity over the range 1 x 10-1M to 1 x 10-6M. Linear detection limit is about 10-5M Interference's:

More information

1) A ph/mv meter or an ion meter, either line operated or portable.

1) A ph/mv meter or an ion meter, either line operated or portable. COLE-PARMER Sodium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS The Cole-Parmer Sodium Ion Electrodes are used to quickly, simply, accurately, and economically measure sodium ion concentrations

More information

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions. Cole-Parmer Ammonium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium

More information

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Fluoroborate Ion Selective Electrode is used to quickly, simply, accurately, and economically measure Fluoroborate in aqueous solutions.

More information

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic. GENERAL INSTRUCTIONS Introduction The Van London Co. Cyanide Ion Selective Electrode is used to measure dissolved cyanide ion in aqueous solutions. *IMPORTANT NOTICE* Acidic cyanide solutions produce hydrogen

More information

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

POTASSIUM ION SELECTIVE ELECTRODE

POTASSIUM ION SELECTIVE ELECTRODE Instruction Manual and Experiment Guide for the PASCO scientific Model CI-6733 012-06614A 9/97 POTASSIUM ION SELECTIVE ELECTRODE 1997 PASCO scientific $7.50 10101 Foothills Blvd. P.O. Box 619011 Roseville,

More information

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049 imagine explore learn Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049 Overview Fluoride is often added to drinking water and to toothpaste to prevent dental decay. In high concentrations,

More information

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions. COLE-PARMER NITRATE ION ELECTRODES INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate

More information

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London Co. Nitrogen Oxide Gas-Sensing Electrode is used to measure nitrite in aqueous solutions. The measurement is not affected by sample color or turbidity.

More information

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Ammonium Ion Electrode is used to quickly, simply, accurately, and economically

More information

COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement Contents Table of Contents 1. Introduction 1 2. Required Equipment 2 3. Electrode and Measurement Setup 4 Electrode

More information

Chem 321 Lecture 17 - Potentiometry 10/24/13

Chem 321 Lecture 17 - Potentiometry 10/24/13 Student Learning Objectives Chem 321 Lecture 17 - Potentiometry 10/24/13 Electrodes The cell described in the potentiometric chloride titration (see 10/22/13 posting) consists of a Ag/AgCl reference electrode

More information

The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions.

The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions. OAKTON Cadmium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous

More information

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml) GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Silver/Sulfide Ion Electrode is used to measure silver or sulfide ions in aqueous solutions. The two ions are virtually never present in solution

More information

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

READING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3.

READING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3. CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE READING Skoog, Holler and Crouch: Chapter 23 and Appendix 3. A. INTRODUCTION Potentiometry is a static electroanalytical method in which the potential

More information

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Calcium Ion Electrodes are used to quickly, simply, accurately, and economically

More information

perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement

perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement Contents Table of Contents 1. Introduction 1 2. Required Equipment 3 3. Electrode and Measurement Setup 4 Electrode

More information

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution , drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific

More information

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking

More information

phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL 1. A ph/mv meter or an ion meter, either line operated or portable.

phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL 1. A ph/mv meter or an ion meter, either line operated or portable. phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The phoenix Electrode Company Carbon Dioxide Gas-Sensing Electrode is used to quickly,

More information

perfection Guidebook perfection Combination Cyanide Electrode Successful Ion Measurement

perfection Guidebook perfection Combination Cyanide Electrode Successful Ion Measurement perfection Guidebook perfection Combination Cyanide Electrode Successful Ion Measurement Contents Table of Contents 1. Introduction 1 2. Required Equipment 3 3. Electrode and Measurement Setup 4 Electrode

More information

CALCIUM SELECTIVE ELECTRODE

CALCIUM SELECTIVE ELECTRODE Instruction Manual and Experiment Guide for the PASCO scientific Model CI-6727 012-06601A 9/97 CALCIUM SELECTIVE ELECTRODE 1997 PASCO scientific $7.50 10101 Foothills Blvd. P.O. Box 619011 Roseville, CA

More information

User Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode

User Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode User Manual Thermo Scientific Orion Fluoride Ion Selective Electrode 254792-001 Revision B September 2016 Table of Contents Introduction...1 Required Equipment...3 Electrode Setup...5 Electrode Preparation...5

More information

NITRATE ION SELECTIVE ELECTRODE

NITRATE ION SELECTIVE ELECTRODE Instruction Manual and Experiment Guide for the PASCO scientific Model CI-6735 012-06616A 9/97 NITRATE ION SELECTIVE ELECTRODE 1997 PASCO scientific $7.50 10101 Foothills Blvd. P.O. Box 619011 Roseville,

More information

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015 Ion Selective Electrodes for the Laboratory Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015 Outline Review ISE measurement technology How to properly calibrate

More information

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA , 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine

More information

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Industrial Nitrate Ion Electrode is used to quickly, simply, accurately, and economically

More information

perfection Guidebook perfection Combination Cupric Electrode Successful Ion Measurement

perfection Guidebook perfection Combination Cupric Electrode Successful Ion Measurement perfection Guidebook perfection Combination Cupric Electrode Successful Ion Measurement Contents Table of Contents 1. Introduction 1 2. Required Equipment 3 Introduction 3. Electrode and Measurement Setup

More information

phoenix Electrode Company Ammonia Gas-Sensing Electrode Instruction Manual

phoenix Electrode Company Ammonia Gas-Sensing Electrode Instruction Manual phoenix Electrode Company Ammonia Gas-Sensing Electrode Instruction Manual GENERAL INSTRUCTIONS Introduction The phoenix Electrode Company Ammonia Gas-Sensing Electrode is used to quickly, simply, accurately,

More information

Potentiometry: Ion-Selective Electrodes

Potentiometry: Ion-Selective Electrodes Potentiometry: Ion-Selective Electrodes Determination of the sodium monofluorophosphate content of a toothpaste sample using a fluoride ion-selective electrode. 1. Aims The aims of the experiment are to:

More information

symphony Multisensor Probes

symphony Multisensor Probes DOC022.53.80319 symphony Multisensor Probes 07/2012, Edition 1 User Manual Table of contents Safety information on page 3 Calibration on page 6 Specifications on page 3 Sample requirements on page 8 Product

More information

Guide to ph Analysis.

Guide to ph Analysis. Guide to ph Analysis. XX Why Contents ph Matters Table of Contents Why ph Matters pg 4 What You Need pg 6 How to Measure pg 14 Slope and offset pg 18 Checklist pg 20 Best Practices pg 21 2 Guide to ph

More information

Electrochemistry LEC Potentiometric ph titration (phosphoric acid in a soft drink) What you need: What you can learn about

Electrochemistry LEC Potentiometric ph titration (phosphoric acid in a soft drink) What you need: What you can learn about Electrochemistry LEC 06 What you can learn about Galvanic cell Types of electrodes Nernst equation Potentiometry Principle and tasks The cell voltage and the Galvani voltage of the electrodes of an galvanic

More information

The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in aqueous solutions.

The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in aqueous solutions. OAKTON Lead Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in

More information

phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual 1. A ph/mv meter or an ion meter, either line operated or portable.

phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual 1. A ph/mv meter or an ion meter, either line operated or portable. phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The phoenix Electrode Company Silver/Sulfide Ion Electrodes are used to quickly, simply, accurately,

More information

Analysis of cations and anions by Ion- Selective Electrodes (ISEs)

Analysis of cations and anions by Ion- Selective Electrodes (ISEs) Analysis of cations and anions by Ion- Selective Electrodes (ISEs) Purpose: The purpose of this assignment is to introduce potentiometric measurements of ionic species by ion selective electrodes (ISEs)

More information

EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION

EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION I. INTRODUCTION This experiment introduces the direct-measurement approach to potentiometry. Principal purposes of the study are a) to understand quantitative

More information

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Water and Lab Products, Thermo Fisher Scientific Technical Note 505 Key Words Thermo Scientific Orion

More information

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051 imagine explore learn Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051 Overview Sodium is highly abundant in the lithosphere (Earth s outer shell) where it occurs as rock salt deposits as

More information

Understanding ph Troubleshooting and Diagnostic information

Understanding ph Troubleshooting and Diagnostic information Understanding ph Troubleshooting and Diagnostic information The desire is to achieve an accurate, reliable measurement with a reasonable electrode life expectancy while minimizing frequency or complexity

More information

NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC

NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC OPERATING MANUAL ba15362e03 10/2014 NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC NITRATE SENSITIVE ELECTRODE NO 500 NO 800 Note The latest version of the present operating manual can be found on the Internet

More information

1225 Lab # 11 Fluoride Speciation and Analysis by Flow Injection using Ion-Selective Electrode: Measurement of Total Fluoride in Water

1225 Lab # 11 Fluoride Speciation and Analysis by Flow Injection using Ion-Selective Electrode: Measurement of Total Fluoride in Water 1225 Lab # 11 Fluoride Speciation and Analysis by Flow Injection using Ion-Selective Electrode: Measurement of Total Fluoride in Water Introduction Fluoride is recognized world- wide to help build stronger

More information

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014 All about ph! Is it as simple as it seems? The ph Workshop Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014 Common Questions: Measuring ph What is ph? Potential Hydrogen

More information

INSTRUCTION MANUAL. Type: ORP500 C, ORP500 CD, ORP222 CD, ORP223 CD EN Introduction Helpful Operating Tips. 2

INSTRUCTION MANUAL. Type: ORP500 C, ORP500 CD, ORP222 CD, ORP223 CD EN Introduction Helpful Operating Tips. 2 ORP/Reference Electrodes INSTRUCTION MANUAL Type: ORP500 C, ORP500 CD, ORP222 CD, ORP223 CD EN 02-05 Table of Contents 1. Introduction. 2 2. Helpful Operating Tips. 2 3. Calibration Procedure 3 3.1. Laboratory

More information

STANDARD OPERATING PROCEDURE

STANDARD OPERATING PROCEDURE Ref.. Table of Contents Page. Scope 2 2. Objective 2 3. Definitions 2 4. Responsibility 3 5. Procedure 3 5.. Diagram of Display and Control Panel 3 5.2. Installation 4 5.3. Operation 4 5.4. Calibration

More information

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C V 2.8 Revised 1/16/18 ph Probe Silver / silver chloride Reads ph Range 0 14 Resolution Response time Max pressure Max depth Temperature range C Cable length Internal temperature sensor Time before recalibration

More information

Ion Selective Electrode Probe

Ion Selective Electrode Probe Rev. 1 Ion Selective Electrode Probe KDS-1064 Calcium KDS-1065 Ammonium KDS-1066 Nitrate KDS-1067 Chloride Type : Ion selective electrode Range : 1 M to 5x10-7 M(40,000 ~ 0.01 ppm) Sensor description The

More information

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION INDUSTRIAL COATINGS CS19 PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION is a chromium-free passivating rinse. It is formulated to provide improved adhesion

More information

A) QUICK GUIDE TO BASIC OPERATION

A) QUICK GUIDE TO BASIC OPERATION Operating Instructions and Notes on Measuring Procedures for all ELIT Ion-Selective Electrodes (Last updated 11 May 2015 -- Chris C Rundle, BSc, PhD -- Nico2000 Ltd.) For more information, visit: www.nico2000.net

More information

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A imagine explore learn Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A Overview Ammonium ions (NH 4+ ) can be found in fertilizer, cleaning products, processed meats and in the waste of

More information

Bante210 Benchtop ph Meter Instruction Manual

Bante210 Benchtop ph Meter Instruction Manual Bante210 Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante210 Benchtop ph Meter 1 Introduction Thank you for selecting the Bante210 benchtop ph meter. This manual provides a step-by-step

More information

What is a WQSensor? Software Installation. Uninstalling WQSensors Software. NexSens Technology, Inc. TABLE OF CONTENTS

What is a WQSensor? Software Installation. Uninstalling WQSensors Software. NexSens Technology, Inc. TABLE OF CONTENTS TABLE OF CONTENTS 1 Sensor Operation...3 1.1 Ion Selective Electrode Sensors...3 1.1.1 WQ-FL Fluoride ISE Sensor... 3 1.1.2 Making ISE Sensor Measurements... 4 1.1.3 Maintenance and Care... 8 1.1.4 ISE

More information

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan APPENDIX 7 PH MEASURING 1. General information PH probe is a unit, measuring the solution acidity or alkalinity degree. It is measured on scale of 0 to 14. The term ph is derived from "p", the mathematical

More information

Rapid RH Portable ph Meter. User's Manual

Rapid RH Portable ph Meter. User's Manual Rapid RH Portable ph Meter User's Manual This page intentionally left blank. 2 Contents 1. Introduction... 4 2. Included with the Rapid RH Portable ph Meter... 4 3. Understanding ph... 5 4. Calibrating

More information

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

PHS-25CW/3BW Benchtop ph Meter Instruction Manual PHS-25CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides

More information

Advanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX:

Advanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX: IMPORTANT NOTES FOR FLUORIDE ISE MEASUREMENT SYSTEM WITH 56 ANALYZER IMPORTANT NOTE # 1: This addendum ONLY applies to the 2 nd generation Model 56 ISE Analyzers with the 2.19 software loaded. This addendum

More information

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT METHOD 9040B ph ELECTROMETRIC MEASUREMENT 1.0 SCOPE AND APPLICATION 1.1 Method 9040 is used to measure the ph of aqueous wastes and those multiphase wastes where the aqueous phase constitutes at least

More information

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction User s Guide Waterproof Palm ph Meter Model PH220 Introduction Congratulations on your purchase of the Extech PH220 waterproof ph/mv meter. This microprocessor-based device with tactile buttons is battery

More information

Instrumental Chemical Analysis. Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013

Instrumental Chemical Analysis. Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013 0510212 Instrumental Chemical Analysis Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013 Chapter 1 Electroanalytical Methods Electroanalytical Chemistry Electroanalytical

More information

CHEM 334 Quantitative Analysis Laboratory

CHEM 334 Quantitative Analysis Laboratory Calibration of Volumetric Glassware Introduction Volumetric glassware is a class of glass vessels that are calibrated to contain or deliver certain volumes of substances. Graduated cylinders, pipettes

More information

Electrodes for ph measurements. - when you need to be sure...

Electrodes for ph measurements. - when you need to be sure... Electrodes for ph measurements - when you need to be sure... Different types Combined Reference Glass Electrode working principle Glass electrode Reference electrode Electrode body Internal reference element

More information

LEAD ION SELECTIVE ELECTRODE

LEAD ION SELECTIVE ELECTRODE Instruction Manual and Experiment Guide for the PASCO scientific Model CI-6736 012-06617A 8/97 LEAD ION SELECTIVE ELECTRODE 1997 PASCO scientific $7.50 10101 Foothills Blvd. P.O. Box 619011 Roseville,

More information

User Guide. High Performance Ammonia Ion Selective Electrode

User Guide. High Performance Ammonia Ion Selective Electrode User Guide High Performance Ammonia Ion Selective Electrode Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries. AQUAfast, AQUASensors, BOD AutoEZ, ionplus,

More information

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth V 2.4 Revised 3/1/18 Spear Tip ph Probe Silver / silver chloride Reads ph Range 0 14 Resolution Response time Max pressure Max depth Temperature range C Cable length Internal temperature sensor Time before

More information

COLE-PARMER INDUSTRIAL SILVER/SULFIDE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL SILVER/SULFIDE ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL SILVER/SULFIDE ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Silver/sulfide Ion Electrode is used to quickly, simply, accurately,

More information

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION INDUSTRIAL COATINGS CS100 PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION is a chromium-free organic passivating rinse. It is formulated to provide improved

More information

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-25CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides

More information

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION INDUSTRIAL COATINGS CS59 PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION is a chromium-free concentrate for use as a final rinse after phosphate with CHEMFOS iron or zinc

More information

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide:

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide: Weak Acid Titration v010516 You are encouraged to carefully read the following sections in Tro (3 rd ed.) to prepare for this experiment: Sec 4.8, pp 168-174 (Acid/Base Titrations), Sec 16.4, pp 769-783

More information

Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant

Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant Approved ASTM Method for Fluoride in Drinking Water and Wastewater Analyze free fluoride ions

More information

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge Standard Operating Procedure for GLNPO Specific Conductance: Marvin Palmer United States Environmental Protection Agency Great Lakes National Program Office Region 5 77 West Jackson 9th Floor Chicago,

More information

Intermediate Junction Electrode Instruction Manual Models

Intermediate Junction Electrode Instruction Manual Models Intermediate Junction Electrode Instruction Manual Models IJ44A ph electrode ph 0-12 IJ44C ph electrode ph 0-14 IJ64 ORP/Redox electrode IJ14 Reference electrode IJAg Silver Billet electrode IJAu Gold

More information

Manual ph Probe & Oxidation Reduction Potential Probe Models 6528A and 6528B

Manual ph Probe & Oxidation Reduction Potential Probe Models 6528A and 6528B Manual ph Probe & Oxidation Reduction Potential Probe Models 6528A and 6528B Revision History File name / Revision Date Authors Previous version BX 2004 RS/ JH Unidata Manual - 6528 ph and ORP Sensors

More information

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82 ALLOWAY Method Outline Standard Laboratory Method: SM4500-F - B and SM4500-F-C Parameter: Total Fluoride via Distillation Method: Distillation then Specific Ion Date Issued: 01/04/82 Date Revised: 04/09/02,

More information

Ka Acid Dissociation Constant Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.1.16

Ka Acid Dissociation Constant Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.1.16 Ka Acid Dissociation Constant Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.1.16 I. Introduction Monoprotic acetic acid, CH 3 COOH is sometimes written as HCH 3 COO, HC

More information

Electro Analytical Methods

Electro Analytical Methods CH 2252 Instrumental Methods of Analysis Unit II Electro Analytical Methods Dr. M. Subramanian Associate Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam

More information

User Guide. Perchlorate Ion Selective Electrode

User Guide. Perchlorate Ion Selective Electrode User Guide Perchlorate Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information

Chemistry MultiMeasure Sensor PS-2170

Chemistry MultiMeasure Sensor PS-2170 ä 5 Instruction Manual 012-09523A Chemistry MultiMeasure Sensor PS-2170 ph/ise/orp PRESSURE CHEMISTRY TEMPERATURE SENSOR VOLTAGE 1 2 3 4 PS-2170 5 6 7 cc 8 10 15 20 25 30 35 40 45 50 55 60 Included Equipment

More information

Chapter 3 Electrochemical methods of Analysis-Potentiometry

Chapter 3 Electrochemical methods of Analysis-Potentiometry Chapter 3 Electrochemical methods of Analysis-Potentiometry Electroanalytical chemistry Contents Introduction Galvanic and electrolytic cells Salt bridge Electrode potential and cell potential Indicator

More information

User Guide. High Performance Ammonia Ion Selective Electrode

User Guide. High Performance Ammonia Ion Selective Electrode User Guide High Performance Ammonia Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT,

More information

DOC /2013, Edition 1. User Manual

DOC /2013, Edition 1. User Manual DOC022.52.80407 50 49 08/2013, Edition 1 User Manual Table of Contents Specifications...3 General information...3 Safety information...3 Use of hazard information...3 Precautionary labels...4 Product

More information

GETTING THE END POINT TO APPROXIMATE. Two hours

GETTING THE END POINT TO APPROXIMATE. Two hours Chem 1312 Handout Experiment ONE Laboratory Time Required Special Equipment and Supplies Objective Safety First Aid GETTING THE END POINT TO APPROXIMATE THE EQUIVALENCE POINT Two hours Balance Potassium

More information

The Decomposition of Hydrogen Peroxide. Evaluation copy

The Decomposition of Hydrogen Peroxide. Evaluation copy The Decomposition of Hydrogen Peroxide Computer 12 The decomposition of hydrogen peroxide in aqueous solution proceeds very slowly. A bottle of 3% hydrogen peroxide sitting on a grocery store shelf is

More information

General Medicine 2016/17

General Medicine 2016/17 ÚSTAV LÉKAŘSKÉ BIOCHEMIE A LABORATORNÍ DIAGNOSTIKY 1. LF UK Buffers, buffer capacity. Oxidoreduction, electrode processes Practical lesson on medical biochemistry General Medicine Martin Vejražka, Tomáš

More information

NH 3 -selective electrodes

NH 3 -selective electrodes NH 3 -selective electrodes Manual 8.109.8031EN Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com NH 3 -selective electrodes Manual 8.109.8031EN

More information

EXPERIMENT 16 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion

EXPERIMENT 16 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion EXPERIMENT 16 Electrochemical Cells: A Discovery Exercise 1 Introduction This lab is designed for you to discover the properties of electrochemical cells. It requires little previous knowledge of electrochemical

More information

User Guide. Calcium Ion Selective Electrode

User Guide. Calcium Ion Selective Electrode User Guide Calcium Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information