MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20
|
|
- Alaina Sparks
- 6 years ago
- Views:
Transcription
1 MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20 Introduction Protein aggregation is recognised as a major issue in the biopharmaceutical industry. Proteins have a tendency to aggregate over time and the risk for biologic drugs is that the presence of aggregates will stimulate an immune response. Size-exclusion chromatography (SEC) is a powerful tool that is commonly used to look at the aggregation of proteins. SEC separates proteins by size. It is commonly used to measure their molecular weight and characterise aggregation. While most SEC systems use a single concentration detector such as ultraviolet (UV), the addition of light scattering allows the measurement of the molecular weight of the protein independent of its retention volume. The molecular weight of the protein monomer, oligomers and aggregates in a sample can then be measured to understand its behaviour. Multi angle light scattering (MALS) can also be used to measure the radius of gyration (Rg) of molecules that scatter light anisotropically (unevenly with respect to angle). The Viscotek SEC-MALS 20 system (figure 1) is a 20 angle light scattering device capable of measuring the molecular weight of protein and protein aggregates independent of elution volume. The addition of further detectors such as concentration detectors (RI, UV), intrinsic viscosity (IV) and dynamic light scattering (DLS) can greatly increase the amount of information available from a single SEC measurement. In this application note, a selection of proteins are separated using SEC. The molecular weights of their oligomers and aggregates are measured by SEC-MALS and the results discussed. Figure1: The Viscotek SEC-MALS 20 system 2013 Malvern Instruments Limited MRK
2 Materials and methods The SEC-MALS 20 system was connected to a Viscotek TDAmax system using the TDA RI detector to measure concentration. Samples were separated using 2 x Viscotek protein columns. The mobile phase was phosphate-buffered saline and all the samples were dissolved in the mobile phase. The SEC-MALS system was calibrated using bovine serum albumin, a protein with a well characterised molecular weight. The detectors and columns were all held at 30 C to ensure a good separation and maximise baseline stability of the detectors. Results Samples of BSA are typically a mixture of monomer, oligomers and larger aggregates. Here, monomer, dimer and trimer peaks are clearly visible and their molecular weights have accurately been measured by MALS. The sample also contains some more polydisperse higher order aggregates that are not individually resolved. The BSA chromatogram can be seen in figure 2 and the results from the BSA measurement can be seen in table 1. Figure 1: Chromatogram of BSA showing the refractive index (red) and SEC-MALS (90 ) (orange) detector signals. The molecular weight measured by SEC-MALS is overlaid in black. Table 1: Molecular weights of the different components of the BSA sample. Aggregates Trimer Dimer Monomer Peak RV (ml) Mn (kda) Mw (kda) Mw / Mn Rg(w) (nm) N/C N/C N/C N/C Wt Fr (Peak) The molecular weights across the monomer, dimer and trimer peaks are clearly seen to be very stable. This is expected with proteins which individually have very tightly controlled molecular weights and are considered to be monodisperse. While these molecular weights are stable, indicating that these are oligomers, the molecular weight trace is more variable in the earliest eluting material that is made up from a more polydisperse population. This is more typical of denatured aggregates which are likely to no longer be functional. The fraction of the total sample in each peak is identified in the Wt Fr row. SEC-MALS 20 application note - MRK
3 The second protein studied was thyroglobulin and its chromatogram can be seen in figure 2. The MALS signals from the different angles can be seen in the MALS plot in figure 3 and the calculated results can be seen in table 2. Figure 2: Chromatogram of thyroglobulin showing the refractive index (red) and SEC-MALS (90 ) (orange) detector signals. The molecular weight measured by SEC-MALS is overlaid in black. Figure 3: Chromatogram of SEC-MALS detector signals for thyroglobulin. SEC-MALS 20 application note - MRK
4 Table 2: Measured molecular weights of the different peaks of the thyroglobulin sample. Aggregates Monomer Peak RV (ml) Mn (kda) Mw (kda) Mw / Mn Rg(w) (nm) N/C N/C Wt Fr (Peak) The thyroglobulin sample has been separated into 2 peaks according to the RI detector but 3 according to the SEC-MALS detector. The SEC-MALS detector is identifying a very small amount of very large aggregates that are only just visible to the RI detector. The molecular weight of the monomer has been measured at 686 kda. This is within 3% of the 669 kda molecular weight of thyroglobulin. This is stable across the peak indicating this population is monodisperse. The aggregate peaks, which have been analysed as a single component, show a higher and more variable (polydisperse) molecular weight clearly identifying them as some disordered aggregates. It is unlikely that any of the proteins in this peak will have any activity and if it were a drug sample, these aggregates would be very likely to be immunogenic. The aggregates account for approximately 17% of the sample. It is worth noting that despite the high molecular weight of this protein, the SEC-MALS display clearly shows that the scattering from the monomer and the aggregates is isotropic. As such, it is not possible to measure the Rg of any of the peaks in this sample. The final sample measured was pepsin and its chromatogram can be seen in figure 4. The SEC-MALS signals from the different angles can be seen in the MALS plot in figure 5 and the calculated results can be seen in table 3. Figure 4: Chromatogram of pepsin showing the refractive index (red) and SEC-MALS (90 ) (orange) detector signals. The molecular weight measured by SEC-MALS is overlaid in black. SEC-MALS 20 application note - MRK
5 Figure 5: Chromatogram of SEC-MALS detector signals for pepsin. Table 3: Measured molecular weights of the different peaks of the pepsin sample. Aggregates Monomer Degradation products Peak RV (ml) Mn (kda) Mw (kda) Mw / Mn Rg(w) (nm) 69.9 N/C N/C Wt Fr (Peak) The pepsin sample has been separated into 2 main peaks. The molecular weight of the second peak is measured at 34.7 kda which is very close to known molecular weight of pepsin of 35 kda. The larger peak has a much higher and more variable molecular weight clearly identifying it as some disordered aggregates which are unlikely to be active. Despite being such a large light scattering peak, the aggregates account for less than 1% of the total material in the sample. In this case, the MALS plot shows that these aggregates are strongly anisotropic scatterers. This means that their Rg can actually be measured, and is shown in the table to be nearly 70 nm. Pepsin has a nonglobular structure which means that it is larger than generally expected for a protein of that molecular weight. The size of the aggregate shows that the structure actually persists in the aggregates, which then become very large. This compares with the thyroglobulin aggregates which were much smaller (<15 nm) despite being over half the molecular weight of the pepsin aggregates. The pepsin monomer is too small for the measurement of Rg. There is a third peak in the sample which is quite poorly defined but it is possible to measure its molecular weight as approximately 6 kda. As a digestive enzyme, it is quite likely that this pepsin sample is digesting itself and that this represents some of the digestion products. Discussion This application note has shown the successful measurement of the molecular weight and aggregate content of a number of proteins using the Viscotek SEC-MALS 20 system. This instrument can measure protein molecular weight independent of a protein s elution volume or structure. The molecular weight of BSA oligomers and thyroglobulin and pepsin aggregates were successfully measured and their amounts quantified. Where the aggregates are large enough, their size (Rg) can also be measured. SEC-MALS 20 application note - MRK
6 Malvern Instruments Ltd Enigma Business Park Grovewood Road Malvern Worcestershire UK WR14 1XZ Tel: +44 (0) Fax: +44 (0) Malvern Instruments Worldwide Sales and service centers in over 50 countries for details visit Malvern Instruments Ltd 2013 more information at SEC-MALS 20 application note - MRK
The power of multi-detector SEC in the analysis of antibodies
The power of multi-detector SEC in the analysis of antibodies How OMNISEC can measure antibody absolute MW independent of size and structural changes MOLECULAR SIZE MOLECULAR STRUCTURE MOLECULAR WEIGHT
More informationMeasurements of protein electrophoretic mobility using the Zetasizer Nano ZSP
Measurements of protein electrophoretic mobility using the Zetasizer Nano ZSP Introduction Zetasizer Nano is the market leader in dynamic and electrophoretic light scattering technology for measurements
More informationDifferentiation of polymer branching and composition using the Mark Houwink plot
Differentiation of polymer branching and composition using the Mark Houwink plot MOLECULAR SIZE MOLECULAR STRUCTURE MOLECULAR WEIGHT Introduction The manipulation of polymer properties through changes
More informationGPC/SEC Analysis of Polymer Degradation
MOLECULAR WEIGHT Introduction MOLECULAR Understanding the effects of processing a polymer on its molecular weight, STRUCTURE molecular weight distribution, and solution viscosity are important to MOLECULAR
More informationSample measurements to demonstrate Zetasizer specifications (Zetasizer Nano, Zetasizer APS, Zetasizer μv)
Sample measurements to demonstrate Zetasizer specifications (Zetasizer Nano, Zetasizer APS, Zetasizer μv) Contents Size using Non-Invasive back-scatter (NIBS) optics Zetasizer Nano S/ZS...2 Minimum size
More informationGPC/SEC Analysis of Polymer Solutions Used in Inkjet Printing
GPC/SEC Analysis of Polymer Solutions Used in Inkjet Printing Introduction Inkjet printing of polymer solutions is an area of interest for depositing precise, high-throughput patterns of functional polymer.
More informationMALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE
MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE MALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD Setting the Standard WHAT CAN ADVANCED GPC/SEC DETECTION OFFER YOU? A modern research environment
More informationThe Diffusion Barrier Technique, Practical Aspects and Data interpretation
The Diffusion Barrier Technique, Practical Aspects and Data interpretation Introduction The diffusion barrier technique is a method for reliably measuring the electrophoretic mobility of proteins by minimising
More informationIntroduction to the calculators in the Zetasizer software
Introduction to the calculators in the Zetasizer software PARTICLE SIZE ZETA POTENTIAL MOLECULAR WEIGHT MOLECULAR SIZE Introduction The calculators are a series of tools in the Zetasizer software that
More informationOptimization of PLA/PLGA molecular weight and structure to suit final application requirements
Optimization of PLA/PLGA molecular weight and structure to suit final application requirements MOLECULAR SIZE MOLECULAR STRUCTURE MOLECULAR WEIGHT Introduction The biodegradable and biocompatible polymers,
More informationHow to use GPC/SEC for compositional analysis
How to use GPC/SEC for compositional analysis Determining the relative concentration of two components in a polymer sample MOLECULAR SIZE MOLECULAR STRUCTURE MOLECULAR WEIGHT Introduction Over the last
More informationUnderstanding the colloidal stability of protein therapeutics using dynamic light scattering
Understanding the colloidal stability of protein therapeutics using dynamic light scattering A Malvern Instruments' Bioscience Development Initiative Executive summary The combination of dynamic light
More informationUse of SEC-MALS. (Size Exclusion Chromatography - Multi Angle. Light Scattering) for protein quality and characterization
Use of SEC-MALS (Size Exclusion Chromatography - Multi Angle Light Scattering) for protein quality and characterization Methods for protein characterization Analytical SEC is a common method to characterize
More informationHow switching mobile phases can improve your OMNISEC results
How switching mobile phases can improve your OMNISEC results Analysis of ester- and acid-capped PLGA samples in mobile phases of DCM and THF MOLECULAR SIZE Introduction MOLECULAR Biodegradable, thermoplastic
More informationBenefit of light scattering technologies (RALS/LALS/MALS) and multidetection characterization in life science research?
Benefit of light scattering technologies (RALS/LALS/MALS) and multidetection characterization in life science research? Bert Postma Business Support Separations and MicroCal Size Exclusion Chromatography
More informationAdvantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis
Advantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis Comparing Columns from Different Vendors to Improve Data Quality Technical Overview Introduction Size exclusion chromatography
More informationMalvern Macromolecular Solutions. Innovative Solutions in Material Characterization
Malvern Macromolecular Solutions Innovative Solutions in Material Characterization MALVERN HEADQUARTERS Malvern Instruments is a Spectris company. Innovative Solutions in Material Characterization Particle
More informationDEVELOPING A BIOFORMULATION STABILITY PROFILE
DEVELOPING A BIOFORMULATION STABILITY PROFILE Light Scattering and Micro-Capillary Viscometry A Malvern Instruments Bioscience Development Initiative Executive Summary: When used in combination, light
More informationUse of DLS/Raman to study the thermal unfolding process of lysozyme
Use of DLS/Raman to study the thermal unfolding process of lysozyme A Malvern Instruments' Bioscience Development Initiative Executive summary The combination of dynamic light scattering (DLS) and Raman
More information10 REASONS THAT THE ZETASIZER NANO IS CHOSEN FOR SCIENTIFIC SUCCESS
10 REASONS THAT THE ZETASIZER NANO IS CHOSEN FOR SCIENTIFIC SUCCESS Abstract: New components and measurement techniques continue to be introduced for the Zetasizer Nano. New to the system are a Surface
More informationInvestigating the Relationship Between the Rheological Properties of Hyaluronic Acid and its Molecular Weight and Structure using Multidetector
Investigating the Relationship Between the Rheological Properties of Hyaluronic Acid and its Molecular Weight and Structure using Multidetector SEC and SEC-MALS Presented by Bassem Sabagh, PhD Technical
More informationLatest Developments in GPC Analysis of Adhesive and Sealant Polymers Mark Pothecary PhD Americas Product Manager Malvern Instruments
Latest Developments in GPC Analysis of Adhesive and Sealant Polymers Mark Pothecary PhD Americas Product Manager Malvern Instruments Molecular weight The most fundamental molecular property that controls
More informationThe ultimate in desktop particle characterization
Particle size Zeta potential Molecular weight ZS The ultimate in desktop particle characterization detailed specification from www.malvern.com/zetasizernano The Zetasizer Nano ZS brings you the practicality
More informationElectrophoretic Light Scattering Overview
Electrophoretic Light Scattering Overview When an electric field is applied across an electrolytic solution, charged particles suspended in the electrolyte are attracted towards the electrode of opposite
More informationREVERSE ENGINEERING PHARMACEUTICAL FORMULATION USING THE MORPHOLOGI G3-ID.
REVERSE ENGINEERING PHARMACEUTICAL FORMULATION USING THE MORPHOLOGI G3-ID. Introduction According to the definition established by the FDA, a generic drug is a drug product which is comparable to a reference
More informationThe ultimate in desktop particle characterization
Particle size Zeta potential Molecular weight ZS The ultimate in desktop particle characterization detailed specification from www.malvern.com/zetasizernano The Zetasizer Nano ZS brings you the practicality
More informationComponent specific particle characterisation of the active components in a pharmaceutical topical formulations
Component specific particle characterisation of the active components in a pharmaceutical topical formulations PARTICLE SIZE PARTICLE SHAPE CHEMICAL IDENTIFICATION Introduction Most pharmaceutical products
More informationDetermining cure profile and post-cure shrinkage of photopolymers using UV accessory on a rotational rheometer
Determining cure profile and post-cure shrinkage of photopolymers using UV accessory on a rotational rheometer RHEOLOGY AND VISCOSITY Introduction A photopolymer usually consists of monomers, oligomers
More informationNanobubble Applications and Characterization by Nanoparticle Tracking Analysis
Nanobubble Applications and Characterization by Nanoparticle Tracking Analysis Review of ultra-fine bubble (nanobubble) applications and characterization examples PARTICLE CONCENTRATION PARTICLE SIZE Introduction
More informationStudying the Effect of Steric Layer Thickness on Emulsion Stability
Studying the Effect of Steric Layer Thickness on Emulsion Stability Dr Tim J. Wooster, Professor Mary A. Augustin, Food Science Australia (CSIRO) of Chemistry, Monash University, Victoria 3800, Australia
More informationUsing High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Light Scattering Detection
Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Light Scattering Detection Jennifer Gough 1, Michael Jones 1, Damian Morrison 1, and John Stenson 2
More informationSample preparation and characterization around SAXS
Sample preparation and characterization around SAXS Experimental verification and validation? Rob Meijers EMBL Hamburg Garbage in? The right stuff Molecular weight Oligomerization state Monodispersity
More informationZetasizer Nano ZSP: A Perfect Tool For Life Science Applications
Zetasizer Nano ZSP: A Perfect Tool For Life Science Applications Dr Mike Kaszuba Technical Support Manager E-mail: michael.kaszuba@malvern.com Contents Zetasizer Nano ZSP Software Enhancements Protein
More informationTips & Tricks GPC/SEC: From a Chromatogram to the Molar Mass Distribution
Tips & Tricks GPC/SEC: From a Chromatogram to the Molar Mass Distribution Peter Kilz and Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany. Molar masses cannot be measured directly by gel
More informationOvercoming and quantifying Wall Slip in measurements made on a rotational rheometer
Overcoming and quantifying Wall Slip in measurements made on a rotational rheometer RHEOLOGY AND VISCOSITY Introduction When making shear rheological measurements on structured liquids, in particular suspensions,
More informationThe Illumination of Structure using Light Scattering. Michael Caves Product Technical Specialist for Biophysical Characterisation
The Illumination of Structure using Light Scattering Michael Caves Product Technical Specialist for Biophysical Characterisation Light Scattering Laser Scattered light Detector Scattered light intensity
More informationZS90. The ultimate in desktop particle characterization. Particle size. Zeta potential. Molecular weight
Particle size Zeta potential Molecular weight ZS90 The ultimate in desktop particle characterization detailed specifications from www.malvern.com/zetasizernano A routine analysis system for particle characterization
More informationClearing the Confusion: GPC, SEC, GFC What, When, Why, and How?
Clearing the Confusion: GPC, SEC, GFC What, When, Why, and How? Jean Lane Applications Engineer LC Columns & Consumables Technical Support January 19, 2017 What we will cover What - Clarification of the
More informationTips & Tricks GPC/SEC: Protein Analysis with Size-Exclusion Chromatography
Tips & Tricks GPC/SEC: Protein Analysis with Size-Exclusion Chromatography Daniela Held and Thorsten Hofe, PSS Polymer Standards Service GmbH, Mainz, Germany Gel permeation chromatography/size-exclusion
More informationDiscriminating between polymorphs of acetaminophen using Morphologically-Directed Raman Spectroscopy (MDRS)
Discriminating between polymorphs of acetaminophen using Morphologically-Directed Raman Spectroscopy (MDRS) CHEMICAL IDENTIFICATION PARTICLE SIZE PARTICLE SHAPE Introduction Interest in studying mixed
More informationAn Introductions to Advanced GPC Solutions
An Introductions to Advanced GPC Solutions Alan Brookes Sales Manager GPC Instruments EMEAI 9 th April 2014 Agilent GPC/SEC Solutions 1 Introduction to Polymers Polymers are long chain molecules produced
More informationComparison of Polymer Separation by Size Exclusion Chromatography and Asymmetric Flow Field Flow Fractionation
Comparison of Polymer Separation by Size Exclusion Chromatography and Asymmetric Flow Field Flow Fractionation Stepan Podzimek, 1 Christoph Johann 2 1 SYNPO / University of Pardubice, Czech Republic, stepan.podzimek@synpo.cz
More informationGPC / SEC Theory and Understanding
Dr. Jason S. Davies, Smithers Rapra, UK Gel permeation chromatography (GPC), also known as size exclusion chromatography (SEC) is a branch of liquid chromatography specifically concerned with characterisation
More informationUnderstanding the conformational stability of protein therapeutics using Raman spectroscopy
Understanding the conformational stability of protein therapeutics using Raman spectroscopy A Malvern Instruments' Bioscience Development Initiative Executive summary The combination of Dynamic Light Scattering
More informationStatic and dynamic light scattering. Cy Jeffries EMBL Hamburg
Static and dynamic light scattering. Cy Jeffries EMBL Hamburg Introduction. The electromagnetic spectrum. visible 10-16 10-10 10-8 10-4 10-2 10 4 (l m) g-rays X-rays UV IR micro wave Long radio waves 400
More informationSize Exclusion Chromatography in the Presence of an Anionic Surfactant
Size Exclusion Chromatography in the Presence of an Anionic Surfactant Intact Protein Profiling Application Note Author Andy Coffey Agilent Technologies, Inc Abstract Sodium dodecyl sulfate (SDS, or SLS)
More informationProgress toward reliable NC molecular mass distribution by GPC
Progress toward reliable NC molecular mass distribution by GPC Dr E Stubbs Emma.Stubbs@awe.co.uk www.awe.co.uk British Crown Owned Copyright 2016/AWE Content Introduction to GPC New instrumentation Sample
More informationTips & Tricks GPC/SEC: Inter-Detector Delay
Tips & Tricks GPC/SEC: Inter-Detector Delay Daniela Held and Wolfgang Radke, PSS Polymer Standards Service GmbH, Mainz, Germany Combinations of detectors are often used in gel permeation chromatography/size-exclusion
More informationEvaluating the rheological properties of hyaluronic acid hydrogels for dermal filler applications
Evaluating the rheological properties of hyaluronic acid hydrogels for dermal filler applications RHEOLOGY AND VISCOSITY PARTICLE SIZE Introduction Hyaluronic acid (HA) is a naturally occurring polysaccharide
More informationProtein Aggregate Assessment of Ligand Binding Assay (LBA) Reagents Using SEC-MALS
Protein Aggregate Assessment of Ligand Binding Assay (LBA) Reagents Using SEC-MALS Justin Low, Mehraban Khosraviani, Sylvia (Kyung-Joo) Lee, and Jihong Yang, Genentech, Inc. Summary Ligand Binding Assays
More informationExecutive Summary. Introduction WHITEPAPER
The Characterization of Single Wall Carbon Nanotubes Using The Malvern Instruments Helix System: Integrating Dynamic Light Scattering (DLS) and Raman Spectroscopy Executive Summary The combination of DLS
More informationCharacterization of Single Wall Carbon Nanotubes (SWNTs) by combining Dynamic Light Scattering and Raman Spectroscopy
Characterization of Single Wall Carbon Nanotubes (SWNTs) by combining Dynamic Light Scattering and Raman Spectroscopy PARTICLE SIZE Executive Summary The combination of DLS and Raman spectroscopy offers
More informationCharacterization of protein aggregates in suspension and on a filter membrane by Morphologically-Directed Raman Spectroscopy
Characterization of protein aggregates in suspension and on a filter membrane by Morphologically-Directed Raman Spectroscopy Morphologi G3-ID CHEMICAL IDENTIFICATION PARTICLE SHAPE PARTICLE SIZE PROTEIN
More informationBSA pegylation PG Journal of chromatographic A, 1147 (2007)
BSA pegylation BSA 5mg/mL, pi= 4.7, Sigma A796, 65-7% free sulfyhydryl, cysteine 34 of BSA is not paired with another cysteine in the protein structure 1. PEG 1kD, kd, 3 kd 1mg/mL PEG:BSA=1.2:1 (molar
More informationGel Permeation Chromatography (GPC) or Size Exclusion Chromatography (SEC)
Gel Permeation Chromatography (GPC) or Size Exclusion Chromatography (SEC) Size Exclusion Chromatography (SEC) is a non-interaction based separation mechanism in which compounds are retained for different
More informationUsing runt Disposable Cuvettes
Tech Note Guide Table of Contents Ease of Use... 3 Sensitivity... 4 Resolution... 5 Accuracy... 6 Faster Sample Prep... 7 Measuring Molecular Weight... 8 Observing Protein Aggregation... 9 Characterize
More informationInteraction of Gold Nanoparticle with Proteins
Chapter 7 Interaction of Gold Nanoparticle with Proteins 7.1. Introduction The interfacing of nanoparticle with biomolecules such as protein is useful for applications ranging from nano-biotechnology (molecular
More informationClemens Stilling Characterisation, Analytical Sciences for Biologicals, UCB Pharma S.A. Alun, living with Parkinson s disease
Clemens Stilling Characterisation, Analytical Sciences for Biologicals, UCB Pharma S.A. Alun, living with Parkinson s disease Outline Short intro Aggregate characterisation during candidate selection and
More informationUsing rheology to study the hardness and spreadability of butter
Using rheology to study the hardness and spreadability of butter RHEOLOGY AND VISCOSITY Introduction Butter is a multiphase emulsion consisting of fat globules, crystalline fat and an aqueous phase dispersed
More informationOVERVIEW INTRODUCTION
Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Dynamic and Multi-Angle Light Scattering Detection Jennifer Gough 1, Robert Birdsall 1, Damian Morrison
More informationHow Molecular Weight and Branching of Polymers Influences Laser Sintering Techniques
How Molecular Weight and Branching of Polymers Influences Laser Sintering Techniques Dr. Bernd Tartsch Malvern Instruments GmbH Rigipsstr. 19, D-71083 Herrenberg Tel: +49-703-97 770, Fax: +49-703-97 854
More informationTips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages
Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany Gel permeation chromatography/size-exclusion chromatography (GPC/SEC)
More informationInform is a series of white papers designed to provide advice on material characterization issues. Mie theory The rst 100 years
Inform is a series of white papers designed to provide advice on material characterization issues Mie theory The rst 100 years Mie theory the rst 100 years One hundred years ago Gustav Mie developed an
More informationDr. Christoph Johann Wyatt Technology Europe GmbH Copyright Wyatt Technology Europe GmbH All Rights reserved 1
Dr. Christoph Johann Wyatt Technology Europe GmbH 2010 Copyright Wyatt Technology Europe GmbH All Rights reserved 1 Introduction Overview The Nature of Scattered Light: Intensity of scattered light Angular
More informationUsing the Zetasizer Nano to optimize formulation stability
Using the Zetasizer Nano to optimize formulation stability PARTICLE SIZE ZETA POTENTIAL The top five ways you can use the Zetasizer Nano to help improve the stability of your formulation. Formulation development
More informationHighly automated protein formulation development: a case study
Application Note Highly automated protein formulation development: a case study Introduction and background Therapeutic proteins can be inherently prone to degradation and instability, and they often pose
More informationPrecision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System
Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System Keywords: Hydrodynamic radius, diffusion coefficient, sizing, Taylor dispersion analysis, protein, antibodies,
More informationSample characterization: Quality control and sample handling prior to data collection
Sample characterization: Quality control and sample handling prior to data collection Marc JAMIN UMI 3265 UJF-EMBL-CNRS Unit of Virus Host Cell interactions Grenoble, France jamin@embl.fr Take home message
More informationNovel instrumentation to gain insight into the aggregation state of proteins by Taylor Dispersion Analysis: a comparison
Novel instrumentation to gain insight into the aggregation state of proteins by Taylor Dispersion Analysis: a comparison with Dynamic Light Scattering Wendy Louise Hulse *, Rob Forbes, David Goodall, Alex
More informationIntroduction to Biological Small Angle Scattering
Introduction to Biological Small Angle Scattering Tom Grant, Ph.D. Staff Scientist BioXFEL Science and Technology Center Hauptman-Woodward Institute Buffalo, New York, USA tgrant@hwi.buffalo.edu SAXS Literature
More informationImpurities In Biomolecules Institute for International Research. Monitoring & Predicting Biomolecular Aggregation Using Light Scattering
Impurities In Biomolecules Institute for International Research Monitoring & Predicting Biomolecular Aggregation Using Light Scattering Kevin Mattison Malvern Instruments Ewa Folta-Stogniew Yale University
More informationBiophysics Service at the MPIB Biochemistry Core Facility Stephan Uebel, Biochemistry Core Facility
Biophysics Service at the MPIB Biochemistry Core Facility 30.11.2015 Stephan Uebel, Biochemistry Core Facility uebel@biochem.mpg.de Overview Peptide Chemistry - Peptide synthesis -Amino acid analysis -
More informationSetting the Standard for GPC. Complete Guide for GPC / SEC / GFC Instrumentation and Detection Technologies. The Right Instrument for Your Application
TM Setting the Standard for GPC Complete Guide for GPC / SEC / GFC Instrumentation and Detection Technologies The Right Instrument for Your Application Viscotek is the global leader in Gel Permeation /
More informationSem /2007. Fisika Polimer Ariadne L. Juwono
Chapter 8. Measurement of molecular weight and size 8.. End-group analysis 8.. Colligative property measurement 8.3. Osmometry 8.4. Gel-permeation chromatography 8.5. Ultracentrifugation 8.6. Light-scattering
More informationSepax SRT -C, Zenix -C SEC Phases
Sepax SRT -C, Zenix -C SEC Phases Complimentary Phases to SRT for Derivatized Monoclonal Antibodies General Description Stationary Phase Structure SRT, Zenix, SRT-C and Zenix-C SEC phases developed based
More informationStrategies to compare far-uv Circular Dichroism spectra of similar proteins using Chirascan -plus ACD
Chirascan Series Application note Strategies to compare far-uv Circular Dichroism spectra of similar proteins using Chirascan -plus ACD Abstract Circular Dichroism (CD) spectroscopy is an extremely sensitive
More informationGel Permeation Chromatography
Gel Permeation Chromatography Polymers and Coatings Laboratory California Polytechnic State University Gel permeation chromatography (GPC) has become the most widely used technique for determination of
More informationGPC - Gel Permeation Chromatography. aka Size Exclusion Chromatography- SEC
GPC - Gel Permeation Chromatography aka Size Exclusion Chromatography- SEC Wendy Gavin Biomolecular Characterization Laboratory Version 1 May 2016 1 Table of Contents 1. GPC Introduction. Page 3 2. How
More informationHigh Performance Liquid Chromatography
High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).
More informationSupporting Information
Supporting Information Self-cleaning membranes from comb-shaped copolymers with photoresponsive side groups Papatya Kaner 1, Xiaoran Hu 2, Samuel W. Thomas III 2, Ayse Asatekin 1,* 1 Department of Chemical
More informationMeet Our Uncle: 12 Stability Applications on One Platform
Tech Note Meet Our Uncle: 12 Stability Applications on One Platform Uncle is an all-in-one stability platform that enables twelve different applications with one instrument. Fluorescence, static light
More informationCharacterization of polyphenylene sulphide using the Agilent PL-GPC 220 High Temperature GPC System with triple detection
materials analysis Characterization of polyphenylene sulphide using the Agilent PL-GPC 220 High Temperature GPC System with triple detection Solutions for Your Analytical Business Markets and Applications
More informationNo.106. Aqueous SEC Columns for Analysis of Cationic Polymers: TSKgel PWXL-CP Series. Contents. Page
No.106 Aqueous SEC Columns for Analysis of Cationic Polymers: TSKgel PWXL-CP Series Contents Page 1. Introduction 2. Features 3. Basic characteristics 3-1. Pore characteristics 3-2. Sample injection volume
More informationParticle Characterization Laboratories, Inc.
Analytical services Particle size analysis Dynamic Light Scattering Static Light Scattering Sedimentation Diffraction Zeta Potential Analysis Single Point Titration Isoelectric point determination Aqueous
More informationPARTICLE SIZE CHARACTERISATION OF AGRICULTURAL SPRAYS USING LASER DIFFRACTION
ILASS-Europe 2002 Zaragoza 9 11 September 2002 PARTICLE SIZE CHARACTERISATION OF AGRICULTURAL SPRAYS USING LASER DIFFRACTION Paul Kippax *, Steve Parkin ** and Clive Tuck ** Paul.Kippax@malvern.co.uk *
More informationGel Permeation Chromatography
Gel Permeation Chromatography Polymers and Coatings Laboratory California Polytechnic State University San Luis Obispo, CA Gel permeation chromatography (GPC) has become the most widely used technique
More informationMACROIMS MACROION MOBILITY SPECTROMETER MOLECULAR WEIGHT AND PARTICLE SIZE DETERMINATION OF POLYVINYLPYRROLIDONE (PVP)
MACROIMS MACROION MOBILITY SPECTROMETER MOLECULAR WEIGHT AND PARTICLE SIZE DETERMINATION OF POLYVINYLPYRROLIDONE (PVP) APPLICATION NOTE CHEMC-009 (A4) Polyvinylpyrrolidone (PVP) Polymer Polyvinylpyrrolidone
More informationnext generation Welcome to the Dispersion options Smarter particle sizing Your sample, expertly processed and prepared Particle size
Particle size Welcome to the next generation Your sample, expertly processed and prepared Dispersion options 3000 detailed specification sheets from Getting your sample preparation right The US Pharmacopeia
More informationSize Exclusion Chromatography
Springer Laboratory Size Exclusion Chromatography Bearbeitet von Sadao Mori, Howard G Barth 1. Auflage 1999. Buch. xiv, 234 S. Hardcover ISBN 978 3 540 65635 7 Format (B x L): 15,5 x 23,5 cm Gewicht: 1170
More informationEXCLUSION HIGH PERFORMANCE LIQUID CHROMATOGRAPHY ANALYSIS OF A CELLULOSE ACETATE-AMMONIUM NITRATE GAS GENERATOR PROPELLANT
I SIZE EXCLUSION HIGH PERFORMANCE LIQUID CHROMATOGRAPHY ANALYSIS OF A CELLULOSE ACETATE-AMMONIUM NITRATE GAS GENERATOR PROPELLANT G. D. KNOWLTON and D. V. HAUN Talley Industries of Arizona, Mesa, Arizona
More informationFFF-MALS method development and measurements of size and molecular weight
Project: FFF-MALS method development and measurements of size and molecular weight Measurement of particle size distribution of protein binding, of mean molecular weight of polymeric NP components, study
More informationCOURSE MATERIAL: Unit 3 (Part 1) Polymer Science LT8501 (Click the link Detail to download)
COURSE MATERIAL: Unit 3 (Part 1) Polymer Science LT8501 (Click the link Detail to download) Dr. Debasis Samanta Senior Scientist & AcSIR Assistant Professor Polymer Science & Technology Department., CSIR-CLRI,
More informationField-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized by Conventional GPC/SEC Techniques
The Field-Flow Fractionation Platform Field-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized by Conventional GPC/SEC Techniques Trevor Havard, Evelin Moldenhaur, Soheyl
More informationVrije Universiteit Amsterdam
Vrije Universiteit Amsterdam MSc Chemistry Analytical Sciences Literature thesis Separation techniques for the quantification of protein aggregates by Debbie van der Burg October 2017 12 ECT Supervisor:
More informationSRT -10/10C (10 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC. Sepax Technologies, Inc. Preparative Size Exclusion Chromatography
Sepax Technologies, Inc. Preparative Size Exclusion Chromatography SRT -0/0C (0 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC Better Surface Chemistry for Better Separation SRT -0/0C Size Exclusion
More informationAnalysis of PEGylated Proteins with Agilent AdvanceBio SEC Columns
Analysis of PEGylated Proteins with Agilent AdvanceBio SEC Columns Application Note Bio-Pharmaceutical Author M.Sundaram Palaniswamy Agilent Technologies, Ltd India Abstract PEGylation of therapeutic proteins
More informationThe ph-responsive behaviour of aqueous solutions of poly(acrylic acid) is dependent on molar mass
Electronic Supplementary Material (ESI) for Soft Matter. This journal is The Royal Society of Chemistry 2016 The ph-responsive behaviour of aqueous solutions of poly(acrylic acid) is dependent on molar
More informationAnalytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals
Analytical Methods to Qualify and Quantify and ylated Biopharmaceuticals Christopher Crafts, Bruce Bailey, Marc Plante, Ian Acworth Thermo Fisher Scientific, Chelmsford, MA Application Note 1015 Key Words
More informationOVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA
Use of High Speed/High Resolution Size-Based Chromatographic Separation of Surfactants and Oligomeric Materials with Single Quadrupole Mass Spectrometry Michael O Leary, Jennifer Gough, Tanya Tollifson
More informationQuick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE
Quick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE Introduction Gel permeation chromatography (GPC) and size exclusion chromatography
More information