1 OPEN SUBJECT AREAS: ANALYTICAL CHEMISTRY CHEMISTRY Received 5 June 014 Accepted 9 September 014 Published 16 October 014 On-site Rapid Detection of Trace Non-volatile Inorganic Explosives by Stand-alone Ion Mobility Spectrometry via Acid-enhanced Evaporization Liying Peng 1,, Lei Hua 1, Weiguo Wang 1, Qinghua Zhou 1, & Haiyang Li 1 1 Key Laboratory of Separation Science for Analytical Chemistry, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian, 11603, People s Republic of China, University of Chinese Academy of Sciences, Beijing, , People s Republic of China. Correspondence and requests for materials should be addressed to H.L. (hli@dicp.ac.cn) New techniques for the field detection of inorganic improvised explosive devices (IEDs) are urgently developed. Although ion mobility spectrometry (IMS) has been proved to be the most effective method for screening organic explosives, it still faces a major challenge to detect inorganic explosives owing to their low volatilities. Herein, we proposed a strategy for detecting trace inorganic explosives by thermal desorption ion mobility spectrometry (TD-IMS) with sample-to-sample analysis time less than 5 s based on in-situ acidification on the sampling swabs. The responses for typical oxidizers in inorganic explosives, such as KNO 3, KClO 3 and KClO 4 were at least enhanced by a factor of 3000 and their limits of detection were found to be subnanogram. The common organic explosives and their mixtures with inorganic oxidizers were detected, indicating that the acidification process did not affect the detection of organic explosives. Moreover, the typical inorganic explosives such as black powders, firecrackers and match head could be sensitively detected as well. These results demonstrated that this method could be easily employed in the current deployed IMS for on-site sensitive detection of either inorganic explosives or organic ones. I n recent years, rapid and sensitive detection of improvised explosive devices (IEDs) has become the most pressing issues with the increase of the globalization of terrorist acts during the last decades, such as those in Oklahoma, U.S.A. (1995), Bali, Indonesia (00, 005), Pune, India (010), Oslo, Norway (011), Boston, U.S.A. (013) and Abuja, Nigeria (014) 1 3. On-site screening is an effective way to reduce the risk of explosion, which demands the analytical methods with high requirements in terms of speed, sensitivity, and robustness The instrumental methods such as X-rays, neutron analysis, nuclear quadrupole resonance, and colorimetric detection are always adopted for the bulk detection of organic and inorganic explosives, while the thermal desorption ion mobility spectrometry (TD-IMS) is proven to be a practical technique for trace detection (from ng to pg) of organic explosives such as,4,6-trinitrotoluene (TNT) and cyclo-1,3,5-trimethylene-,4,6-trinitramine (RDX) et al IMS is a gas-phase ion separation and detection technique in a uniform electric field based on the mobility difference of gaseous ions More than 0,000 stand-alone IMS have been deployed at airports and subway stations worldwide for security applications, due to the advantages of fast speed, low cost, analytical flexibility, portability and commercial availability 0 3. However, it still remains a great challenge for TD-IMS to detect trace inorganic explosives except for ammonium nitrate-fuel oil (ANFO) and sulphur (S) in black powder, owing to their ultra-low vapour pressure even at the typical maximum desorber temperature (#80uC) 13,0,4,5. Inorganic explosives generally consist of an inorganic oxidizer such as potassium nitrate (KNO 3 ), potassium chlorate (KClO 3 ), or potassium perchlorate (KClO 4 ) and a fuel such as carbon source, sulphur (S), sugar or powdered metals 6,7. Inorganic explosives are extensively used in terrorist attacks owing to the readily available, low cost and legally purchased components. Ion chromatography (IC) 8,9, capillary electrophoresis (CE) 30,31, electrospray ionization mass spectrometry (ESI-MS) 3 34 and electrospray ionization ion mobility spectrometry (ESI-IMS) 35 have been performed to identify their characteristic ions, such as nitrate (NO 3 ), chlorate (ClO 3 ), sulphate (SO 4 ) and perchlorate (ClO 4 ) etc. contained within inorganic explosives 36,37. However, the fielddeployment of these techniques is hampered by the tedious and time-consuming procedures for handling the SCIENTIFIC REPORTS 4 : 6631 DOI: /srep
2 Figure 1 Schematic of thermal desorption ion mobility spectrometer and the analysis process of inorganic explosives. (a) Sampling: sampling swabs were swiped with samples; (b) Acidification: samples on the swab were sprayed with minimal regime 3% H 3 PO 4 ; (c) Analysis: the swab was inserted into the thermal desorber and analysed by IMS. aqueous samples. Additionally, the relatively long analysis time (.5 minutes) of IC and CE, and the relatively large bulk, high expensive cost of mass spectrometer make them not suitable for on-site screening as well. Obviously, an analytical method which could rapidly and sensitively detect both inorganic and organic explosives is highly demanded. Herein, we firstly reported a simple method to detect inorganic explosives involving KNO 3, KClO 3 and KClO 4 using TD-IMS via acid-enhanced evaporization for on-site detection. Samples on the swab were acidized by spraying H 3 PO 4, and enhancements of signal for KNO 3, KClO 3 and KClO 4 by a factor of over 3000 were achieved, realizing the detection of inorganic explosives at ng even pg level. Moreover, the common nitro-explosives and its mixture with the inorganic oxidizers and the frequently-used inorganic explosives such as black powder, firecracker and match head were sensitively detected. The influence of acidification process on the detection of organic explosives was also investigated. Results and Discussion Design of TD-IMS for analysis. A swiped-based TD-IMS apparatus operated in negative mode is schematically shown in Fig. 1. The key structure of IMS is similar to those reported in detail previously 5,38 expect that an extra inlet named inlet used for the sample inlet was punched in the reaction region at the vicinity of BN ion gate. Thus, only the purge gas flowed through the 63 Ni source, so the pollution and acid corrosion of 63 Ni source could be effectively avoided. In the test, samples were deposited on the swabs firstly (Fig. 1(a)) and then sprayed with minimal regime 3% H 3 PO 4 (Fig. 1(b)). Afterwards, the swab was inserted into the thermal desorber with temperature of 180uC for IMS analysis (Fig. 1 (c)). Enhancement of acidification on the sensitivity. Ion mobility spectra for typical inorganic oxidizers of KNO 3, KClO 3 and KClO 4 obtained without and with 3% H 3 PO 4 acidfication were depicted in Fig., respectively. Clearly, in lack of H 3 PO 4 acidfication on the swab, only two weak product ion peaks with K 0 of.18 and.05 cm V 1 s 1 were observed for 0 mg KNO 3 and the signal intensity of characteristic peak (.05 cm V 1 s 1 ) was only about 83 mv. Whereas, its response for 0.0 mg KNO 3 instantly increased to 80 mv in the existence of H 3 PO 4. It was over 3 times higher than that of 0 mg KNO 3 obtained without acidification and the sensitivity were at least enhanced by a factor of Similarly, no responses for KClO 3 and KClO 4 even with quantity of 50 mg and 600 mg, respectively were obtained in the absence of H 3 PO 4, while two strong ion peaks of 0.05 mg KClO 3 were appeared at.16 and 1.88 cm V 1 s 1 and peaks with K 0 of.15 and 1.77 cm V 1 s 1 were observed for 0.60 mg KClO 4 in the assistance of H 3 PO 4. And the signal intensities for ion peaks of KClO 3 (1.88 cm V 1 s 1 ), KClO 4 (1.77 cm V 1 s 1 ) were 184 and 179 mv, respectively. Furthermore, we also adopted other weak acids such as acetic acid (CH 3 COOH) and boric acid (H 3 BO 3 )to perform the acidification process. The results demonstrated that H 3 PO 4 was the most suitable for the analysis due to the perfect sensitivity and its simple background (see Supplementary Fig. S1 ). It is noteworthy that the time taken for sample-to-sample analysis is less than 5 s, and the analysis speed is fast enough for rapid on-site screening. Mechanism for the acid-enhancement. The product ions of inorganic oxidizers in 63 Ni source were assigned by an atmospheric pressure negative ionization time-of-flight mass spectrometer (TOF- MS) (see Supplementary Fig. S3). The typical mass spectra were demonstrated in Fig. 3; the reduced ion mobilities (K 0 ), mass-tocharge ratios (m/z) and corresponding product ions were summarized in Table 1. Accordingly, two product ions of KNO 3 at m/z 6 and m/z 15 were assigned to NO 3 and HNO 3?NO 3, and SCIENTIFIC REPORTS 4 : 6631 DOI: /srep06631
3 Figure Ion mobility spectra of 0 mg KNO 3,50mg KClO 3, and 600 mg KClO 4 obtained without 3% H 3 PO 4 and ion mobility spectra of 0.0 mg KNO 3, 0.05 mg KClO 3, and 0.6 mg KClO 4 obtained with 3% H 3 PO 4. two ions of KClO 3 at m/z 83, 85 and m/z 167, 169, 171 were attributed to ClO 3 and HClO 3?ClO 3, while the ions assigned to ClO 4 (m/z 5 99, 101) and HClO 4?ClO 4 (m/z 5 199, 01, 03) were observed for KClO 4. The product ions observed for KNO 3 were same as those in previous experiments of ammonium nitrate (AN, the main component of ANFO) where NO 3 and complex HNO 3?NO 3 were the dominant ions, likely formed by HNO 3 (the decomposition product of AN) 3. According to the observed product ions, gaseous inorganic acids, such as HNO 3, HClO 3 and HClO 4 were believed to be released in the thermal desorber after acidification through the metathesis reaction beween the solid oxidizers and H 3 PO 4 at high temperature of 180uC, and then ionized by the reactant ions O in the ion source. Hence, inorganic oxidizers could be indirectly detected by identifying their gaseous inorganic acids in the existence of H 3 PO 4. It was further verified by ion mobility spectra of trace HNO 3 and HClO 4 (see Supplementary Fig. S4) where the ion peak positions of HNO 3 and HCIO 4 were same as those of KNO 3 and KClO 4 with acidification in Figure 3 TOF-mass spectra of 0 ng KNO 3, 50 ng KClO 3 and 600 ng KClO 4 obtained with 3% H 3 PO 4. Fig.. The mechanism of acid-enhanced detection of inorganic salt oxidizers are summarized as equation (1) to (3) and X presents N or Cl atom. 3KXO n zh 3 PO 4?3HXO n :zk 3 PO 4 ð1þ HXO n zo {?XO n { zho XO n { zhxo n?hxo n : XOn { Quantitative analysis. To further investigate the sensitivities of inorganic explosives, we used ion peaks at K 0 of.05, 1.88 and 1.77 cm V 1 s 1 for quantitative analysis of KNO 3, KClO 3 and KClO 4. The results were summarized in Table 1; their detectable linear range were 60 ng, ng and ng, respectively and the RSD (n 5 9) were 17.10%, 16.7%, and 9.99%, respectively. The limits of detection (LOD, S/N 5 3) for KNO 3, KClO 3 and KClO 4 were 0.10 ng, 0.84 ng and 6.30 ng, respectively. These results exhibit the good quantitative capability of the current method for detection of inorganic explosive. Detection performances of actual samples. The acidification process did not affect the detection of organic explosives and typical nitro-explosives such as TNT and RDX could be sensitively detected by current method, and their ion mobility spectra were displayed in Fig. S5 in the Supplementary Information. Clearly, product ion with K 0 of 1.54 cm V 1 s 1 was observed for TNT, and the product ion peaks at K 0 of 1.65, 1.53 and 1.44 cm V 1 s 1 were observed for RDX. Additionally, the mixtures of TNT and inorganic oxidizers could also be sensitively measured as well, and Fig. 4 shows the ion mobility spectra of 5 ng TNT mixed with 0 ng KNO 3 and 50 ng KClO 3, respectively. From the spectra, the characteristic ion peaks of TNT, KNO 3 and KClO 3 are clearly observed simultaneously for the mixtures. Moreover, the oxidizers in typical inorganic explosives were also detected. Figure 5 shows the ion mobility spectra of black powder, firecracker and match head obtained in the existence of H 3 PO 4. From the spectra, the characteristic ion peaks of KNO 3, KClO 4 and KClO 3 were distinctly observed for black powder, firecracker and match head, respectively. It indicates that inorganic oxidizers in black powder, firecracker and match head are KNO 3, KClO 4 and KClO 3, respectively and other substances in the explosives such as powdered metals and charcoals had little impact on the identification. In particular, only S was detected in the firecracker and black powder without H 3 PO 4 acidification (see Supplementary Fig. S6), so the detection of both S and inorganic oxidizers all could help identification the materials used in IEDs. Finally, as shown in Supplementary Fig. S7, KNO 3, KClO 3, KClO 4 in sugar (0 ng KNO 3 /sugar (451), ðþ ð3þ SCIENTIFIC REPORTS 4 : 6631 DOI: /srep
4 Table 1 Reduced ion mobilities (K 0 ), mass-to-charge ratios (m/z), product ions and quantitative analysis results obtained with 3% H 3 PO 4 KNO 3 KClO 3 KClO 4 Peak 1 Peak Peak 1 Peak Peak 1 Peak K 0 (cm V 1 s 1 ) m/z , , 169, , , 01, 03 Product ions NO 3 HNO 3?NO 3 ClO 3 HClO 3?ClO 3 ClO 4 HClO 4?ClO 4 Linear equation y x 1.90 y x 1.18 y x Linear range (ng) LOD [a] (ng, S/N 5 3) RSD [b] (%, n 5 9) [a] LOD 5 limit of detection. [b] RSD 5 relative standard deviation. 50 ng KClO 3 /sugar (451) and 600 ng KClO 4 /sugar (35)) could also be sensitively measured, which means that the common fuel in inorganic explosives did not affect the detection. Conclusions We demonstrated a new rapid and sensitive method based on IMS via acid-enhanced detection of oxidizers including KNO 3, KClO 3 and KClO 4 to detect inorganic explosives. It owns the advantages of fast speed (about 5 s), low cost, analytical flexibility, portability, commercial availability, and suitable for detecting inorganic explosives and organic ones at the level of ng even pg. Moreover, it is envisioned to be easily deployed to the rapid and sensitive on-site screening of both inorganic explosives and organic ones at airports, subway stations, sports venues and even blast site, being an effective way to screen the substances in the detonation scene and reduce the risk of explosion. Methods Experiment. A 500 ng ml 1 KNO 3, KClO 3 and 600 ng ml 1 KClO 4 standard stock solutions and 3% H 3 PO 4 solution were prepared by weighing sample and dissolving in distilled water. The standard solutions and test samples were obtained by gradually diluting the relevant stock solutions with distilled water or other volatile solvents. 0 ng KNO 3 /sugar (451), 50 ng KClO 3 /sugar (451), 600 ng KClO 4 /sugar (35) were made up by mixing KNO 3, KClO 3, KClO 4 and sugar in corresponding proportion while match head, firecrackers, black powders,,4,6-trinitrotoluene (TNT) and cyclo-1,3,5-trimethylene-,4,6-trinitramine (RDX) were commercial grade. A known quantity sample solutions were added on or solid samples were rubbed on the swabs firstly and the solvent was evaporated to dryness on a hot plate of 50uC, and then the swabs with samples were sprayed minimal regime 3% H 3 PO 4 (several microlitres). After that the swab was inserted into the thermal desorber for analysis by IMS operated in negative mode. It is noted that enough acid content is essential for the sensitivity. Operating process and conditions. The sample on the swab ( cm 3 8 cm) made of Teflon was heated to 180uC in the thermal desorber and vaporized into a gas flow that passed into the drift tube for analysis. In the 63 Ni ionization source, the sample molecules were ionized, and then the product ions were injected into the drift region through the BN gate with an injection pulse width of 300 ms. In the drift region, ions were separated according to their mobility with an electric field of 391 V/cm and then detected by the Faraday plate and the ion current was amplified with a gain of 10 9 V/A. The drift tube temperature was kept at a constant of 90uC. Air dried by silica gel, activated carbon and 133 molecular sieve traps was divided into drift, carrier and purge gases via flow controllers which were optimized at 800, 400 and 400 ml/min, respectively. The operating parameters were summarized in Table. Calculation. The reduced ion mobility, K 0 for unknown peaks were calculated with a standard via the following equation, K 0a ~K 0s t ds ð4þ t da where K 0a is the reduced mobility of the analyte, t ds and t da is the drift time of the standard and the analyte, respectively. (TNT-H) was used as the standard substance, Figure 4 Ion mobility spectra of the mixture of 5 ng TNT and 0 ng KNO 3 and mixture of 5 ng TNT and 50 ng KClO 3 obtained with 3% H 3 PO 4. Figure 5 Ion mobility spectra of 00 ng match head, 600 ng firecracker and 0 ng black powder obtained with 3% H 3 PO 4. SCIENTIFIC REPORTS 4 : 6631 DOI: /srep
5 Table The IMS operating parameters used in this work Operating conditions Setting Desorber temperature 180uC Drift tube temperature 90uC Electric field 391 V cm 1 BN gate opening time 300 ms Carrier gas Purified Air Drift gas Purified Air Carrier gas flow rate 400 ml min 1 Drift gas flow rate 800 ml min 1 Purge gas flow rate 400 ml min 1 whose K 0 was 1.54 cm V 1 s 1 and its drift time was 11.8 ms under the same experimental conditions. 1. Stierstorfer, J. & Klapötke, T. M. High energy materials. propellants, explosives and pyrotechnics. Angew Chem Int Ed 49, (010).. Blanco, G. A. et al. Identification of inorganic improvised explosive devices using sequential injection capillary electrophoresis and contactless conductivity detection. Anal Chem 83, (011). 3. Cheng, S. et al. Fast switching of CO 3 (H O) n and O (H O) n reactant ions in dopant-assisted negative photoionization ion mobility spectrometry for explosives detection. Anal Chem 86, (014). 4. Nagarkar, S. S., Joarder, B., Chaudhari, A. K., Mukherjee, S. & Ghosh, S. K. Highly selective detection of nitro explosives by a luminescent metal-organic framework. Angew Chem Int Ed 5, (013). 5. Bhattacharjee, Y. New efforts to detect explosives require advances on many fronts. Science 13, (008). 6. Hutchinson, J. P. et al. Identification of inorganic improvised explosive devices by analysis of postblast residues using portable capillary electrophoresis instrumentation and indirect photometric detection with a light-emitting diode. Anal Chem 79, (007). 7. Dicinoski, G. W., Shellie, R. A. & Haddad, P. R. Forensic identification of inorganic explosives by ion chromatography. Anal Lett 39, (006). 8. Hazarika, P. & Russell, D. A. Advances in fingerprint analysis. Angew Chem Int Ed 51, (01). 9. Ewing, R. G., Clowers, B. H. & Atkinson, D. A. Direct real-time detection of vapors from explosive compounds. Anal Chem 85, (013). 10. Arndt, J. et al. Preliminary evaluation of the persistence of organic gunshot residue. Forensic Sci Int, (01). 11. Marshall, M. & Oxley, J. C. Aspects of explosives detection. Elsevier (009). 1. Caygill, J. S., Davis, F. & Higson, S. P. J. Current trends in explosive detection techniques. Talanta 88, 14 9 (01). 13. Najarro, M., Dávila Morris, M. E., Staymates, M. E., Fletcher, R. & Gillen, G. Optimized thermal desorption for improved sensitivity in trace explosives detection by ion mobility spectrometry. Analyst 137, (01). 14. Staymates, M. E., Smith, W. J. & Windsor, E. Thermal desorption and vapor transport characteristics in an explosive trace detector. Analyst 136, (011). 15. Yinon, J. Field detection and monitoring of explosives. Trac-trend Anal Chem 1, (00). 16. Davis, E. J., Grows, K. F., Siems, W. F. & Hill, H. H. Improved ion mobility resolving power with increased buffer gas pressure. Anal Chem 84, (01). 17. Du, Y., Wang, W. & Li, H. Bradbury nielsen gate grid structure for further enhancing the resolution of ion mobility spectrometry. Anal Chem 84, (01). 18. Dietiker, R., Lena, F. D. & Chen, P. Fourier transform ion mobility measurement of chain branching in mass-selected, chemically trapped oligomers from methylalumoxane-activated, metallocene-catalyzed polymerization of ethylene. J Am Chem Soc 19 (10), (007). 19. Ewing, R. G. Ion mobility spectrometry, nd Edition. J Am Chem Soc 18 (16), (006). 0. Mäkinen, M., Nousiainen, M. & Sillanpää, M. Ion spectrometric detection technologies for ultra-traces of explosives: A review. Mass Spectrom Rev 30, (011). 1. Song, J., Grün, C. H., Heeren, R. M. A., Janssen, H. G. & van den Brink, O. F. Highresolution ion mobility spectrometry-mass spectrometry on poly (methyl methacrylate). Angew Chem Int Ed 49, (010).. Eiceman, G. A. & Stone, J. A. Ion mobility spectrometers in national defense. Anal Chem 76, 390A 397A (004). 3. Buryakov, I. A. Detection of explosives by ion mobility spectrometry. Anal Chem 66, (011). 4. Steinfeld, J. I. Explosives detection: a challenge for physical chemistry. Annu Rev Phys Chem 49, 03 3 (1998). 5. Liang, X. et al. Sensitive detection of black powder by a stand-alone ion mobility spectrometer with an embedded titration region. Anal Chem 85, (013). 6. Marshall, A. The nature of explosives. Nature 96, (1916). 7. Moretti, J. D., Sabatini, J. J. & Chen, G. Periodate salts as pyrotechnic oxidizers: development of barium- and perchlorate-free incendiary formulations. Angew Chem Int Ed 51, (01). 8. Johns, C. et al. Identification of homemade inorganic explosives by ion chromatographic analysis of post-blast residues. J Chromatogr A 118, (008). 9. Meng, H., Wang, T., Guo, B., Hashi, Y., Guo, C. & Lin, J. Simultaneous determination of inorganic anions and cations in explosive residues by ion chromatography. Talanta 76, (008). 30. Hutchinson, J. P. et al. Identification of inorganic improvised explosive devices by analysis of postblast residues using portable capillary electrophoresis instrumentation and indirect photometric detection with a light-emitting diode. Anal Chem 79, (007). 31. Doyle, J. M. & Mccord, B. R. Novel electrolyte for the analysis of cations in low explosive residue by capillary electrophoresis. J Chromatogr B 714, (1998). 3. Tamiri, T. Characterization of the improvised explosive urea nitrate using electrospray ionization and atmospheric pressure chemical ionization. Rapid Commun Mass Spectrom 19, (005). 33. Sigman, M. E. & Armstrong, P. Analysis of oxidizer salt mixtures by electrospray ionization mass spectrometry. Rapid Commun Mass Spectrom 0, (006). 34. Flanigan, P. M., Brady, J. J., Judge, E. J. & Levis, R. J. Determination of inorganic improvised explosive device signatures using laser electrospray mass spectrometry detection with offline classification. Anal Chem 83, (011). 35. Dwivedi, P., Matz, L. M., Atkinson, D. A. & Hill, J. H. H. Electrospray ionizationion mobility spectrometry: a rapid analytical method for aqueous nitrate and nitrite analysis. Analyst 19, (004). 36. Detata, D. A., Collins, P. A. & McKinley, A. J. A comparison of common swabbing materials for the recovery of organic and inorganic explosive residues. J Forensic Sci 58, (013). 37. Pavlov, J. & Attygalle, A. B. Direct Detection of inorganic nitrate salts by ambient pressure helium-plasma ionization mass spectrometry. Anal Chem 85, 78 8 (013). 38. Du, Y., Wang, W. & Li, H. Resolution enhancement of ion mobility spectrometry by improving the three-zone properties of the bradbury-nielsen gate. Anal Chem 84, (01). Acknowledgments This research is partially supported by NSF of China (117714, ) and the National Key Scientific Instrument and Equipment Development Project (013YQ ). Author contributions L.P. designed the study and wrote the manuscript. L.H. contributed significantly to the measurements on TOF-MS identification. W.W. contributed significantly to the discussion of results and manuscript refinement. Q.Z. contributed to the construction of IMS. H.L. contributed to propose of the project and the refinement of this paper. All the authors contributed to the writing of the manuscript. Additional information Supplementary information accompanies this paper at scientificreports Competing financial interests: The authors declare no competing financial interests. How to cite this article: Peng, L., Hua, L., Wang, W., Zhou, Q. & Li, H. On-site Rapid Detection of Trace Non-volatile Inorganic Explosives by Stand-alone Ion Mobility Spectrometry via Acid-enhanced Evaporization. Sci. Rep. 4, 6631; DOI: /srep06631 (014). This work is licensed under a Creative Commons Attribution-NonCommercial- NoDerivs 4.0 International License. The images or other third party material in this article are included in the article s Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder in order to reproduce the material. To view a copy of this license, visit creativecommons.org/licenses/by-nc-nd/4.0/ SCIENTIFIC REPORTS 4 : 6631 DOI: /srep
Rapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography
Rapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography L. Canella, C. Austin, A. Beavis, P. Maynard, M. Dawson, C Roux, and P. Doble Centre for Forensic Science, University of Technology,
More informationFundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics
Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass
More informationChemistry Instrumental Analysis Lecture 37. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at
More informationSupplementary Information
Supplementary Information Tuning the Luminescence of Metal-Organic Frameworks for Detection of Energetic Heterocyclic Compounds Yuexin Guo, Xiao Feng,*, Tianyu Han, Shan Wang, Zhengguo Lin, Yuping Dong,
More informationDetermination of Elemental Sulfur in Explosives and. Explosive Residues by Gas Chromatography-Mass Spectrometry. Kimberly S.
Determination of Elemental Sulfur in Explosives and Explosive Residues by Gas Chromatography-Mass Spectrometry Kimberly S. Bradley Illinois State Police, Springfield Forensic Science Laboratory 24 Hill
More informationAnalysis of explosives using electrospray ionization/ion mobility spectrometry (ESI/IMS)
Talanta 50 (2000) 1291 1298 www.elsevier.com/locate/talanta Analysis of explosives using electrospray ionization/ion mobility spectrometry (ESI/IMS) G. Reid Asbury, Jörg Klasmeier, Herbert H. Hill Jr.
More informationInstrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai
303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in
More informationQuick Review. - Chemical equations - Types of chemical reactions - Balancing chemical equations - Stoichiometry - Limiting reactant/reagent
Quick Review - Chemical equations - Types of chemical reactions - Balancing chemical equations - Stoichiometry - Limiting reactant/reagent Water H 2 O Is water an ionic or a covalent compound? Covalent,
More informationExplosive Screening by Vapour Detection: Operational Aspects
Explosive Screening by Vapour Detection: Operational Aspects G. Fernandez de la Mora SEDET, Boecillo, Spain, gfdelamora@sedet.com November 14 th, 2013 1 ACES OVERVIEW Product: Air Cargo Explosives Screener
More informationChemistry Instrumental Analysis Lecture 34. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry
More informationTechnical know-how in thermal analysis measurement
Technical articles Technical know-how in thermal analysis measurement Evolved gas analysis by thermogravimetry-differential thermal analysis-mass spectrometry (TG-DTA-MS) technique Kazuko Motomura* and
More informationGas Chromatography (Chapter 2 and 3 in The essence of chromatography)
Gas Chromatography 1. Introduction. Stationary phases 3. Retention in Gas-Liquid Chromatography 4. Capillary gas-chromatography 5. Sample preparation and injection 6. Detectors (Chapter and 3 in The essence
More informationSoil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006
Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006 Introduction: Capillary electrophoresis (CE) is a relatively new, but rapidly growing separation
More informationLIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC
LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC INTRODUCTION Principle: LC/MS is a technique that combines physical separation capabilities of liquid
More informationMiniaturized Sensor for the Detection of Environmental Pollutants Alexander Graf 1, Ronald Stübner 1, Christian Kunath 1, Sebastian Meyer 1, Harald Schenk 1 1 Fraunhofer Institute for Photonic Microsystems
More informationIntroduction to Pharmaceutical Chemical Analysis
Introduction to Pharmaceutical Chemical Analysis Hansen, Steen ISBN-13: 9780470661222 Table of Contents Preface xv 1 Introduction to Pharmaceutical Analysis 1 1.1 Applications and Definitions 1 1.2 The
More informationApplication Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry
Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry Introduction The oil industry requires detailed information on the composition of
More informationSupporting Information. Use of Mass Spectrometric Vapor Analysis to Improve Canine Explosive Detection Efficiency
Supporting Information Use of Mass Spectrometric Vapor Analysis to Improve Canine Explosive Detection Efficiency Ta-Hsuan Ong, Ted Mendum, Geoff Geurtsen, Jude Kelley, Alla Ostrinskaya, Roderick Kunz Chemical,
More informationFluorescent Chemosensor for Selective Detection of Ag + in an. Aqueous Medium
Electronic supplementary information For A Heptamethine cyanine -Based Colorimetric and Ratiometric Fluorescent Chemosensor for Selective Detection of Ag + in an Aqueous Medium Hong Zheng *, Min Yan, Xiao-Xing
More informationLight-Controlled Switching of a Non- Photoresponsive Molecular Shuttle
Supporting Information Light-Controlled Switching of a Non- Photoresponsive Molecular Shuttle Liu-Pan Yang, a,b Fei Jia, a Jie-Shun Cui, a Song-Bo Lu, a and Wei Jiang* a a Department of Chemistry, South
More informationCHROMATOGRAPHY AND MASS SPECTROMETER
22 CHROMATOGRAPHY AND MASS SPECTROMETER 22.1 INTRODUCTION We know that the biochemistry or biological chemistry deals with the study of molecules present in organisms. These molecules are called as biomolecules
More informationM M e M M H M M H. Ion Sources
Ion Sources Overview of Various Ion Sources After introducing samples into a mass spectrometer, the next important step is the conversion of neutral molecules or compounds to gas phase ions. The ions could
More informationTANDEM MASS SPECTROSCOPY
TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER
More informationSupporting Information for
Supporting Information for Room Temperature Palladium-Catalyzed Arylation of Indoles icholas R. Deprez, Dipannita Kalyani, Andrew Krause, and Melanie S. Sanford* University of Michigan Department of Chemistry,
More informationSupplementary Information
Supplementary Information C aryl -C alkyl bond formation from Cu(ClO 4 ) 2 -mediated oxidative cross coupling reaction between arenes and alkyllithium reagents through structurally well-defined Ar-Cu(III)
More informationAnalysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF
Analysis of Biomarkers in Crude Oil Using the Agilent 7 GC/Q-TOF Application Note Petrochemical and Environmental Authors Frank David Research Institute for Chromatography, Kennedypark 6, B-85 Kortrijk,
More informationDetection and Classification of Explosive Compounds in Presence of Interferents Utilizing Laser Ion Mobility Spectrometry
B6.3 Detection and Classification of Explosive Compounds in Presence of Interferents Utilizing Laser Ion Mobility Spectrometry A. Langmeier, C. Oberhüttinger, W. Heep, H. Oberpriller, S. Beer, M. Kessler,
More informationINTRODUCTION OVERVIEW
COMPARISON OF CCS(N 2 ) MEASUREMENTS OBTAINED FROM TWO DIFFERENT T-WAVE ION S WITH DIRECT MEASUREMENTS USING A DRIFT TUBE ION Kevin Giles, Martin Palmer, Keith Richardson, Nick Tomczyk Waters MS Technologies
More informationElectronic Supplementary Information
Electronic Supplementary Information Uniform and Rich Wrinkled Electrophoretic Deposited Graphene Film: A Robust Electrochemical Platform for TNT Sensing Longhua Tang, Hongbin Feng, Jinsheng Cheng and
More informationApplying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation
Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Introduction Fire investigation involves many different types of analyses from crime scene
More informationApplication of Ion Mobility Spectrometry for Determination of Morphine in Human Urine
Caspian J. Chem. 2(2013) 59-64 Application of Ion Mobility Spectrometry for Determination of Morphine in Human Urine Ali Sheibani*, M. Reza Shishehbore, Shohrah Vafaie-Shahi Department of Chemistry, Yazd
More informationTargeting an Achilles Heel in Olefin Metathesis: A Strategy for High-Yield Synthesis of Second-Generation Grubbs Methylidene Catalysts
Supplementary Information for: Targeting an Achilles Heel in Olefin Metathesis: A Strategy for High-Yield Synthesis of Second-Generation Grubbs Methylidene Catalysts Justin A.M. Lummiss, a Nicholas J.
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?
More informationMass Spectrometry. What is Mass Spectrometry?
Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance
More informationThis method describes the identification of the following prohibited colorants in cosmetic products:
A. IDENTIFICATION BY TLC 1. SCOPE AND FIELD OF APPLICATION This method describes the identification of the following prohibited colorants in cosmetic products: Names C I number Pigment Orange 5 12075 Metanil
More informationMass spectrometric determination of the surface compositions of ethanol water mixtures
International Journal of Mass Spectrometry 212 (2001) 267 271 www.elsevier.com/locate/ijms Cluster/kinetic method Mass spectrometric determination of the surface compositions of ethanol water mixtures
More informationAnion recognition in water by a rotaxane containing a secondary rim functionalised cyclodextrin stoppered axle
Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2015 Supplementary Information: Anion recognition in water by a rotaxane containing a secondary rim
More informationChapter 4. Reactions in Aqueous Solution
Chapter 4 Reactions in Aqueous Solution Topics General properties of aqueous solutions Precipitation reactions Acid base reactions Oxidation reduction reactions Concentration of solutions Aqueous reactions
More informationInnovation for Excellence in Detection Technology
[Type text] Excellims Electrospray Ionization High Performance Ion Mobility Spectrometry for Rapid On-site Cleaning Validation in Pharmaceutical Manufacturing Abstract A novel high performance ion mobility
More informationYujuan Zhou, Kecheng Jie and Feihe Huang*
Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2017 A redox-responsive selenium-containing pillar[5]arene-based macrocyclic amphiphile: synthesis,
More information--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.
Translated English of Chinese Standard: GB5009.17-2014 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA National Food Safety Standard-Determination
More informationPRACTICE EXAMINATION QUESTIONS FOR 1.2 AMOUNT OF SUBSTANCE
PRACTICE EXAMINATION QUESTIONS FOR 1.2 AMOUNT OF SUBSTANCE 1. Nitroglycerine, C 3 H 5 N 3 O 9, is an explosive which, on detonation, decomposes rapidly to form a large number of gaseous molecules. The
More informationIon Chromatography Validation for the Analysis of Anions in Gunshot Residue
Ion Chromatography Validation for the Analysis of Anions in Gunshot Residue Student Report Written for: Dr. Samide CH 424 November 22, 2010 Abstract The purpose of this study is to determine whether or
More informationQuantitative and Qualitative Determination of Organics in Water Using Electrospray Ionization Coupled with Ion Mobility Spectrometry
Quantitative and Qualitative Determination of Organics in Water Using Electrospray Ionization Coupled with Ion Mobility Spectrometry Final Progress Report 15 Mar 98 14 Sep 02 Agreement No. DAAG55-98-1-0107
More informationElectronic supplementary information
Electronic supplementary information Surface plasmon resonance enhanced upconversion luminescence in aqueous media for TNT selective detection Nina Tu and Leyu Wang* State Key Laboratory of Chemical Resource
More informationLiquid chromatography/mass spectrometric analysis of explosives: RDX adduct ions
RAPID COMMUNICATIONS IN MASS SPECTROMETRY Rapid Commun. Mass Spectrom. 2003; 17: 943 948 Published online in Wiley InterScience (www.interscience.wiley.com). DOI: 10.1002/rcm.1006 Liquid chromatography/mass
More informationAnalysis of Pressure Sensitive Adhesives by GC/MS and GC/AED with Temperature Programmable Pyrolyzer
2000 The Japan Society for Analytical Chemistry 627 Analysis of Pressure Sensitive Adhesives by GC/MS and GC/AED with Temperature Programmable Pyrolyzer Sadao NAKAMURA,* Masahiko TAKINO,* and Shigeki DAISHIMA**
More informationChem Hughbanks Exam 1 - Spring 2013
Chem 107 - Hughbanks Exam 1 - Spring 2013 UIN # Section 502 Exam 1, Version # A On the last page of this exam, you ve been given a periodic table and some physical constants. You ll probably want to tear
More informationLecture 15: Introduction to mass spectrometry-i
Lecture 15: Introduction to mass spectrometry-i Mass spectrometry (MS) is an analytical technique that measures the mass/charge ratio of charged particles in vacuum. Mass spectrometry can determine masse/charge
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?
More informationNovel fluorescent cationic benzothiazole dye response to G-quadruplex aptamer as a novel K + sensor
Electronic Supplementary Material (ESI) for Analyst. This journal is The Royal Society of Chemistry 2017 Novel fluorescent cationic benzothiazole dye response to G-quadruplex aptamer as a novel K + sensor
More informationed. Brad Collins Aqueous Chemistry Chapter 5 Some images copyright The McGraw-Hill Companies, Inc. Sunday, August 18, 13
ed. Brad Collins Aqueous Chemistry Chapter 5 Some images copyright The McGraw-Hill Companies, Inc. A solution is a homogenous mixture of 2 or more substances at the molecular level The solute(s) is(are)
More informationIDENTIFICATION AND DETERMINATION OF HYDROQUINONE IN COSMETIC PRODUCTS 2 14/11/17 ACM 003 BY TLC AND HPLC
A. IDENTIFICATION BY TLC 1. SCOPE AND FIELD OF APPLICATION The method describes the identification of hydroquinone in cosmetic products. 2. PRINCIPLE Hydroquinone is identified by thin layer chromatography
More informationRSC Advances.
This is an Accepted Manuscript, which has been through the Royal Society of Chemistry peer review process and has been accepted for publication. Accepted Manuscripts are published online shortly after
More informationGas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5
Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH 5730-6730 Analysis Topic 5 Chromatography! Group of separation techniques based on partitioning (mobile phase/stationary phase). Two immiscible
More informationAccurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS
Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS Application Note Authors E. Michael Thurman and Imma Ferrer Center for Environmental
More informationCluster measurements at CLOUD using a high resolution ion mobility spectrometermass spectrometer combination
Cluster measurements at CLOUD using a high resolution ion mobility spectrometermass spectrometer combination Alexey Adamov, Heikki Junninen, Jonathan Duplissy, Mikko Sipilä, Markku Kulmala, and CLOUD Collaboration
More information2401 Gas (liquid) Chromatography
2401 Gas (liquid) Chromatography Chromatography Scheme Gas chromatography - specifically gas-liquid chromatography - involves a sample being vaporized and injected onto the head of the chromatographic
More informationSupporting Information
Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 2017 Supporting Information Lithium Triethylborohydride-Promoted Generation of α,α-difluoroenolates
More informationPhysical and Chemical Changes
Objectives Introduction Physical and Chemical Changes Gezahegn Chaka, Ph.D. Collin College Department of Chemistry To observe physical and chemical changes. To identify and characterize physical and chemical
More informationThe precursor (TBA) 3 [H 3 V 10 O 28 ] was synthesised according to the literature procedure. 1 (TBA = n tetrabutylammonium).
An unprecedented silver decavandate dimer investigated using Ion Mobility Mass Spectrometry Thomas McGlone, Johannes Thiel, Carsten Streb, De Liang Long and Leroy Cronin* Supporting Information Experimental
More informationI. Properties of Aqueous Solutions A) Electrolytes and Non-Electrolytes B) Predicting Solubility* II. Reactions of Ionic Compounds in Solution*
Chapter 5 Reactions in Aqueous Solutions Titrations Kick Acid!!! 1 I. Properties of Aqueous Solutions A) Electrolytes and Non-Electrolytes B) Predicting Solubility* II. Reactions of Ionic Compounds in
More informationThe Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)
The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) Basically all compounds which are less volatile than the mobile phase can be detected. Detection is based on a Universal property of
More informationElectronic Supplementary Information
Electronic Supplementary Information Tailored TiO 2 layers for the photocatalytic ozonation of cumylphenol, a refractory pollutant exerting hormonal activity S. Ardizzone, G. Cappelletti, D. Meroni and
More informationChromatography & instrumentation in Organic Chemistry
Chromatography & instrumentation in Organic Chemistry What is Chromatography? Chromatography is a technique for separating mixtures into their components in order to analyze, identify, purify, and/or quantify
More informationISSN X CODEN (USA): PCHHAX. Determination of captopril with potassium permanganate chemiluminescence system*
Available online at www.derpharmachemica.com ISSN 975-413X CODEN (USA): PCHHAX Der Pharma Chemica, 16, 8(1):289-293 (http://derpharmachemica.com/archive.html) Determination of captopril with potassium
More informationSupporting Information
ne-pot synthesis of pyrrolidino- and piperidinoquinolinones by three-component aza-diels Alder reactions of -arylimines with in situ generated cyclic enamides. Wenxue Zhang, Yisi Dai, Xuerui Wang, Wei
More informationThe First Asymmetric Total Syntheses and. Determination of Absolute Configurations of. Xestodecalactones B and C
Supporting Information The First Asymmetric Total Syntheses and Determination of Absolute Configurations of Xestodecalactones B and C Qiren Liang, Jiyong Zhang, Weiguo Quan, Yongquan Sun, Xuegong She*,,
More informationNational standard of People s Republic of China
National standard of People s Republic of China GB 5413.22 2010 National food safety standard Determination of phosphorus in foods for infants and young children, milk and milk products Issued on 2010-03-26
More informationElectronic Supplementary Information (12 pages)
Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2014 A C 2 -responsive pillar[5]arene: synthesis and self-assembly in water Kecheng Jie, Yong Yao, Xiaodong
More informationDetermination of Volatile Substances Proof of Food Adulteration
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of Volatile Substances Proof of Food Adulteration (method: gas chromatography with mass spectrometric detection) Exercise guarantor:
More informationSelective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation
Electronic Supplementary Material (ESI) for Chemical Communications. This journal is The Royal Society of Chemistry 2015 Content: Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of
More informationSimultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID
Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.
More informationUltra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry
PO-CON18E Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) ASMS 21 TP 3 Jing Dong, Satoshi Yamaki, Yuki Hashi, Naoki Hamada Shimadzu China,
More informationIntroduction. Chapter 1. Learning Objectives
Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography
More informationSupporting Information
Electronic Supplementary Material (ESI) for Organic & Biomolecular Chemistry. This journal is The Royal Society of Chemistry 2016 Supporting Information Merging visible-light photoredox and copper catalysis
More informationIntroduction to Instrumental Detection Technology: IMS and DMS
Patricia Guerra-Diaz, Ph.D. Introduction to Instrumental Detection : IMS and DMS Outline Trace vs. bulk detection Instrumental detection technologies Ion Mobility Spectrometry (IMS) Morpho Detection Itemiser
More informationDirect Swabbing and Surface Recovery with Ion Trap Mobility Spectrometry
Direct Swabbing and Surface Recovery with Ion Trap Mobility Spectrometry By Derek Brand, Mei Guo, Dr. Ralf Wottrich and Tim Wortley Abstract Ion Trap Mobility Spectrometry has been discussed as a fast
More informationCharacterization of petroleum samples via thermal analysis coupled to APCI FTMS
Characterization of petroleum samples via thermal analysis coupled to APCI FTMS Abstract Thermal analysis, by means of a thermo balance (TG), was coupled to an atmospheric pressure chemical ionization
More informationatomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive
Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific
More informationAn EZ-DART spectrum was taken for both fifty (SI figure 2a) and twenty (figure 3 and
Electronic Supplementary Material (ESI) for Analytical Methods. This journal is The Royal Society of Chemistry 2017 Electronic Supplemental Information Maleic Acid Data SI Figure 1 Spectrum of maleic acid
More informationChemical Aspects of Mass Spectrometry
Chemical Aspects of Mass Spectrometry Abraham Badu Chemistry and Biochemistry, OSU July 12, 217 Spread of MS by Discipline https://masspec.scripps.edu/ mass spectrometry 2 1 Current Challenges in Mass
More informationChoosing the metabolomics platform
GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation
More informationAmmonium-Bearing Dinuclear Copper(II) Complex: A Highly Selective and Sensitive Colorimetric Probe for Pyrophosphate
Ammonium-Bearing Dinuclear Copper(II) Complex: A Highly Selective and Sensitive Colorimetric Probe for Pyrophosphate Wenxiang Yu, Jian Qiang, Jun Yin, Srinivasulu Kambam, Fang Wang, Yong Wang, and Xiaoqiang
More informationA rapid and highly selective colorimetric method for direct detection of tryptophan in proteins via DMSO acceleration
A rapid and highly selective colorimetric method for direct detection of tryptophan in proteins via DMSO acceleration Yanyan Huang, Shaoxiang Xiong, Guoquan Liu, Rui Zhao Beijing National Laboratory for
More informationSupporting information for A simple copper-catalyzed two-step one-pot synthesis of indolo[1,2-a]quinazoline
Supporting information for A simple copper-catalyzed two-step one-pot synthesis of indolo[1,2-a]quinazoline Chunpu Li 1,2, Lei Zhang 2, Shuangjie Shu 2 and Hong Liu* 1,2 Address: 1 Department of Medicinal
More informationChemistry Instrumental Analysis Lecture 27. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)
More informationIDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)
Document A. THIN LAYER CHROMATOGRAPHY (TLC) 1. SCOPE AND FIELD OF APPLICATION The method describes the identification of hydrocortisone acetate, dexamethasone, betamethasone, betamethasone 17-valerate
More informationMOF-76: From Luminescent Probe to Highly Efficient U VI Sorption Material
MOF-76: From Luminescent Probe to Highly Efficient U VI Sorption Material Weiting Yang, a Zhi-Qiang Bai, b Wei-Qun Shi*, b Li-Yong Yuan, b Tao Tian, a Zhi-Fang Chai*, c Hao Wang, a and Zhong-Ming Sun*
More informationOrganoselenium-Catalyzed Mild Dehydration of Aldoximes: An Unexpected Practical Method for Organonitrile Synthesis
Supporting Information for Organoselenium-Catalyzed Mild Dehydration of Aldoximes: An Unexpected Practical Method for Organonitrile Synthesis Lei Yu,*,,, Hongyan Li, Xu Zhang,, Jianqing Ye, Jianping Liu,
More informationTrace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection
Trade Science Inc. September 2009 Volume 8 Issue 3 ACAIJ, 8(3) 2009 [346-349] Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization
More informationRapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry
Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,
More informationElectronic Supplementary Information
Electronic Supplementary Information Novel [2]pseudorotaxanes constructed by self-assembly of bis-urea-functionalized pillar[5]arene and linear alkyl dicarboxylates Qunpeng Duan, Wei Xia, Xiaoyu Hu, Mengfei
More informationHighly Ordered Binary Assembly of Silica Mesochannels. and Surfactant Micelles for Extraction and Electrochemical
Supporting Information to Highly Ordered Binary Assembly of Silica Mesochannels and Surfactant Micelles for Extraction and Electrochemical Analysis of Trace Nitroaromatic Explosives and Pesticides Fei
More informationOVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA
Use of High Speed/High Resolution Size-Based Chromatographic Separation of Surfactants and Oligomeric Materials with Single Quadrupole Mass Spectrometry Michael O Leary, Jennifer Gough, Tanya Tollifson
More informationCHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.
CHROMATOGRAPHY The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments. THEORY OF CHROMATOGRAPHY: Separation of two sample components in
More informationFigure S1 - Enzymatic titration of HNE and GS-HNE.
Figure S1 - Enzymatic titration of HNE and GS-HNE. Solutions of HNE and GS-HNE were titrated through their reduction to the corresponding alchools catalyzed by AR, monitoring the decrease in absorbance
More informationDEVELOPMENT AND OPTIMIZATION OF THE UV REDUCER FOR THE DETERMINATION OF EXPLOSIVES
DEVELOPMENT AND OPTIMIZATION OF THE UV REDUCER FOR THE DETERMINATION OF EXPLOSIVES Alena Bumbová, Zbyněk Večeřa, Pavel Mikuška, Josef Kellner and Alena Langerová Abstract: In the study the procedure of
More informationChemistry 283g- Experiment 4
EXPEIMENT 4: Alkenes: Preparations and eactions elevant sections in the text: Fox & Whitesell, 3 rd Ed. Elimination eactions of Alcohols: pg. 426-428, 431-432 Electrophilic Addition to Alkenes: pg. 484-488,
More informationMethod for Characterization of Gum Rosin by Capillary Gas Chromatography
PCA TEST METHOD PCTM 27 Published 1/1 2004 1. Scope Method for Characterization of Gum Rosin by Capillary Gas Chromatography 1.1 These test methods cover the determination of the amounts of certain rosin
More information