Therapeutic Substance Monitoring/Quantitative Toxicology Proficiency Test Program. Statistical Summary - August 2015 (Event 15-3)
|
|
- Arabella Carpenter
- 5 years ago
- Views:
Transcription
1 ANDREW M. CUOMO Governor HOWARD A. ZUCKER, M.D., J.D. Commissioner SALLY DRESLIN, M.S., R.N. Executive Deputy Commissioner Therapeutic Substance Monitoring/Quantitative Toxicology Proficiency Test Program Statistical Summary - August 2015 (Event 15-3) This statistical report summarizes participant data for the five Therapeutic Substance Monitoring proficiency survey specimens shipped 31 August Test samples (Vials T21, T22, T23, T24, T25) were prepared by the quantitative transfer of constituents to pooled human serum. The material was subsequently sterile filtered, dispensed into aliquots, stored at -80 o C, and distributed to participants for analysis. Results for individual instrument and reagent systems where the number of laboratories using those systems is three or greater are provided. Mean and Standard Deviation (± 1 SD) values are calculated by a robust statistical technique that does not assume a Gaussian distribution. Disclaimer: Note: The use of brand and/or trade names in this report does not constitute an endorsement of the products on the part of the Wadsworth Center or the New York State Department of Health Should you have any questions regarding this report, please contact the Therapeutic Substance Monitoring Section at (518) Empire State Plaza, Corning Tower, Albany, NY health.ny.gov
2 Acetaminophen (mg/l) ± ± ± ± ± 5.87 n = 211 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 0.92 n = 20 [ABJ] Abbott Architect c System ± ± ± ± ± 3.97 n = 21 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 2.86 n = 10 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± 0.42 n = 7 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 1.30 n = 7 [JJF] Ortho Vitros 5,1FS ± ± ± ± ± 0.98 n = 20 [JJG] Ortho Vitros ± ± ± ± ± 1.44 n = 20 [ROC] Roche cobas c ± ± ± ± ± 0.76 n = 6 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 3.37 n = 8 [ROD] Roche MODULAR D/P ± ± ± ± ± 1.02 n = 8 [BYE] Siemens ADVIA ± ± ± ± ± 1.26 n = 18 [DUE] Siemens Dimension EXL ± ± ± ± ± 1.43 n = 3 [DUR] Siemens Dimension RxL ± ± ± ± ± 1.60 n = 46 [DUT] Siemens Dimension Vista ± ± ± ± ± 2.08 n = 6 [DUX] Siemens Dimension Xpand ± ± ± ± ± 0.92 n = 20 [AB1] Abbott ± ± ± ± ± 2.59 n = 18 [BC1] Beckman Coulter ± ± ± ± ± 3.51 n = 15 [OL1] Beckman Coulter AU Series ± ± ± ± ± 0.90 n = 30 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± 1.46 n = 23 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 3.03 n = 6 [RO2] Roche Hitachi and Modular D/P ± ± ± ± ± 0.77 n = 5 [RO1] Roche Integra and MIRA ± ± ± ± ± 1.45 n = 10 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± 1.55 n = 73 [DA5] Siemens Dimension ± ± ± ± ± 5.00 n = 3 [SY2] Syva Emit ± ± ± ± ± 5.07 n = 4 [SY5] Syva Emit tox 1
3 Carbamazepine (mg/l) ± ± ± ± ± 0.71 n = 211 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 0.56 n = 9 [ABJ] Abbott Architect c System ± ± ± ± ± 0.23 n = 11 [ABH] Abbott Architect i System ± ± ± ± ± 0.46 n = 22 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 0.08 n = 10 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± 0.29 n = 8 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 0.39 n = 5 [JJF] Ortho Vitros 5,1FS ± ± ± ± ± 0.43 n = 20 [JJG] Ortho Vitros ± ± ± ± ± 0.92 n = 18 [ROC] Roche cobas c ± ± ± ± ± 0.50 n = 3 [ROG] Roche cobas c ± ± ± ± ± 0.17 n = 9 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 0.41 n = 8 [ROD] Roche MODULAR D/P ± ± ± ± ± 0.49 n = 6 [BYE] Siemens ADVIA ± ± ± ± ± 0.48 n = 10 [COB] Siemens ADVIA Centaur ± ± ± ± ± 0.35 n = 10 [DUE] Siemens Dimension EXL ± ± ± ± ± 0.56 n = 3 [DUR] Siemens Dimension RxL ± ± ± ± ± 0.35 n = 47 [DUT] Siemens Dimension Vista ± ± ± ± ± 0.10 n = 3 [DUX] Siemens Dimension Xpand ± ± ± ± ± 0.38 n = 20 [AB1] Abbott ± ± ± ± ± 0.23 n = 20 [BC1] Beckman Coulter ± ± ± ± ± 0.35 n = 10 [OL1] Beckman Coulter AU Series ± ± ± ± ± 0.68 n = 7 [MG1] Microgenics CEDIA ± ± ± ± ± 0.42 n = 25 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± 0.86 n = 21 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 0.18 n = 5 [RO2] Roche Hitachi and Modular D/P ± ± ± ± ± 0.20 n = 11 [RO1] Roche Integra and MIRA ± ± ± ± ± 0.48 n = 12 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± 0.26 n = 7 [BY5] Siemens ADVIA/Syva Emit ± ± ± ± ± 0.36 n = 63 [DA5] Siemens Dimension ± ± ± ± ± 0.30 n = 5 [SY4] Syva Emit
4 Digoxin (µg/l) ± ± ± ± ± n = 268 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± n = 8 [ABJ] Abbott Architect c System ± ± ± ± ± n = 15 [ABH] Abbott Architect i System ± ± ± ± ± n = 4 [SAA] Beckman Coulter ACCESS ± ± ± ± ± n = 32 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± n = 11 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± n = 8 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± n = 3 [BCU] Beckman Coulter UniCel DxI ± ± ± ± ± n = 7 [JJF] Ortho Vitros 5,1FS ± ± ± ± ± n = 21 [JJG] Ortho Vitros ± ± ± ± ± n = 3 [ROJ] Roche cobas c ± ± ± ± ± n = 20 [ROC] Roche cobas c ± ± ± ± ± n = 3 [ROG] Roche cobas c ± ± ± ± ± n = 3 [ROA] Roche cobas e ± ± ± ± ± n = 5 [ROT] Roche Cobas INTEGRA ± ± ± ± ± n = 8 [ROD] Roche MODULAR D/P ± ± ± ± ± n = 9 [BYE] Siemens ADVIA ± ± ± ± ± n = 16 [COB] Siemens ADVIA Centaur ± ± ± ± ± n = 21 [DUE] Siemens Dimension EXL ± ± ± ± ± n = 4 [DUR] Siemens Dimension RxL ± ± ± ± ± n = 48 [DUT] Siemens Dimension Vista ± ± ± ± ± n = 6 [DUX] Siemens Dimension Xpand ± ± ± ± ± n = 4 [DPD] Siemens Immulite ± ± ± ± ± n = 23 [AB1] Abbott ± ± ± ± ± n = 29 [BC1] Beckman Coulter ± ± ± ± ± n = 19 [OL1] Beckman Coulter AU Series ± ± ± ± ± n = 5 [MG2] Microgenics DRI ± ± ± ± ± n = 29 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± n = 25 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± n = 5 [RO3] Roche Elecsys/Modular E/e601/e ± ± ± ± ± n = 3 [RO2] Roche Hitachi and Modular D/P ± ± ± ± ± n = 6 [RO1] Roche Integra and MIRA ± ± ± ± ± n = 5 [RO5] Roche Tina-quant ± ± ± ± ± n = 26 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± n = 44 [DA5] Siemens Dimension ± ± ± ± ± n = 35 [DA6] Siemens Dimension LOCI ± ± ± ± ± n = 5 [DP5] Siemens Immulite ± ± ± ± ± n = 6 [SY4] Syva Emit
5 Ethanol (mg/dl) ± ± ± ± ± 5.71 n = 218 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 3.01 n = 18 [ABJ] Abbott Architect c System ± ± ± ± ± 5.98 n = 20 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 5.23 n = 11 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± 8.65 n = 7 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 4.85 n = 9 [GCC] Gas Chromatograph ± ± ± ± ± 6.59 n = 7 [JJF] Ortho Vitros 5,1FS ± ± ± ± ± 3.38 n = 20 [JJG] Ortho Vitros ± ± ± ± ± 3.82 n = 18 [ROC] Roche cobas c ± ± ± ± ± 3.08 n = 6 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 3.11 n = 9 [ROD] Roche MODULAR D/P ± ± ± ± ± 7.40 n = 8 [BYE] Siemens ADVIA ± ± ± ± ± 4.47 n = 18 [DUE] Siemens Dimension EXL ± ± ± ± ± 1.02 n = 3 [DUR] Siemens Dimension RxL ± ± ± ± ± 5.24 n = 45 [DUT] Siemens Dimension Vista ± ± ± ± ± 3.37 n = 6 [DUX] Siemens Dimension Xpand ± ± ± ± ± 2.84 n = 19 [AB1] Abbott ± ± ± ± ± 7.07 n = 19 [BC1] Beckman Coulter ± ± ± ± ± 5.75 n = 14 [OL1] Beckman Coulter AU Series ± ± ± ± ± 3.64 n = 8 [IH1] In-House ± ± ± ± ± 4.21 n = 30 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± 4.12 n = 22 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 3.11 n = 9 [RO2] Roche Hitachi and Modular D/P ± ± ± ± ± 2.87 n = 7 [RO1] Roche Integra and MIRA ± ± ± ± ± 6.49 n = 6 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± 5.65 n = 4 [BY5] Siemens ADVIA/Syva Emit ± ± ± ± ± 4.89 n = 72 [DA5] Siemens Dimension ± ± ± ± ± n = 3 [SY2] Syva Emit 4
6 Ethosuximide (mg/l) ± ± ± ± ± n = 7 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± n = 3 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± n = 6 [SY2] Syva Emit 5
7 Gentamicin (mg/l) 7.20 ± ± ± ± ± 1.43 n = 167 [---] All Methods & Instruments [---] Weigh-in value 7.17 ± ± ± ± ± 0.31 n = 6 [ABJ] Abbott Architect c System 7.33 ± ± ± ± ± 1.53 n = 11 [ABH] Abbott Architect i System 7.66 ± ± ± ± ± 1.03 n = 17 [OLC] Beckman Coulter AU Chemistry System 7.87 ± ± ± ± ± 0.39 n = 6 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± 0.51 n = 7 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 0.88 n = 5 [JJF] Ortho Vitros 5,1FS 7.94 ± ± ± ± ± 0.99 n = 15 [JJG] Ortho Vitros ± ± ± ± ± 1.15 n = 12 [ROC] Roche cobas c ± ± ± ± ± 1.76 n = 3 [ROS] Roche Cobas INTEGRA ± ± ± ± ± 0.36 n = 9 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 0.79 n = 7 [ROD] Roche MODULAR D/P 7.33 ± ± ± ± ± 1.86 n = 11 [COB] Siemens ADVIA Centaur 6.96 ± ± ± ± ± 0.53 n = 6 [DUE] Siemens Dimension EXL 6.94 ± ± ± ± ± 0.61 n = 46 [DUT] Siemens Dimension Vista 7.25 ± ± ± ± ± 0.69 n = 17 [AB1] Abbott 7.88 ± ± ± ± ± 0.53 n = 14 [BC1] Beckman Coulter 7.58 ± ± ± ± ± 1.04 n = 9 [OL1] Beckman Coulter AU Series 7.80 ± ± ± ± ± 0.99 n = 20 [JJ1] Ortho Clinical Diagnostics 6.68 ± ± ± ± ± 1.74 n = 9 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 1.09 n = 5 [RO2] Roche Hitachi and Modular D/P 6.73 ± ± ± ± ± 0.51 n = 12 [RO1] Roche Integra and MIRA 7.33 ± ± ± ± ± 1.86 n = 11 [BY1] Siemens ADVIA/ADVIA Centaur 7.09 ± ± ± ± ± 0.82 n = 3 [BY5] Siemens ADVIA/Syva Emit ± ± ± ± ± 0.63 n = 53 [DA5] Siemens Dimension 7.67 ± ± ± ± ± 0.71 n = 7 [SY4] Syva Emit ± ± ± ± ± 0.53 n = 4 [TH1] Thermo Scientific 6
8 Lithium (mmol/l) ± ± ± ± ± n = 197 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± n = 16 [ABJ] Abbott Architect c System ± ± ± ± ± n = 21 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± n = 8 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± n = 9 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± n = 7 [JJF] Ortho Vitros 5,1FS ± ± ± ± ± n = 21 [JJG] Ortho Vitros ± ± ± ± ± n = 19 [ROC] Roche cobas c ± ± ± ± ± n = 3 [ROG] Roche cobas c ± ± ± ± ± n = 5 [ROT] Roche Cobas INTEGRA ± ± ± ± ± n = 7 [ROD] Roche MODULAR D/P ± ± ± ± ± n = 9 [BYE] Siemens ADVIA ± ± ± ± ± n = 7 [DUE] Siemens Dimension EXL ± ± ± ± ± n = 45 [DUT] Siemens Dimension Vista ± ± ± ± ± n = 3 [DUX] Siemens Dimension Xpand ± ± ± ± ± n = 15 [AB1] Abbott ± ± ± ± ± n = 16 [BC1] Beckman Coulter ± ± ± ± ± n = 16 [OL1] Beckman Coulter AU Series ± ± ± ± ± n = 3 [IH1] In-House ± ± ± ± ± n = 29 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± n = 23 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± n = 5 [RO1] Roche Integra and MIRA ± ± ± ± ± n = 11 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± n = 57 [DA5] Siemens Dimension ± ± ± ± ± n = 12 [TH1] Thermo Scientific 7
9 Phenobarbital (mg/l) ± ± ± ± ± 1.66 n = 185 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 0.66 n = 4 [ABJ] Abbott Architect c System ± ± ± ± ± 1.07 n = 15 [ABH] Abbott Architect i System ± ± ± ± ± 1.20 n = 21 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 1.10 n = 10 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± 1.48 n = 8 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 5.74 n = 3 [JJG] Ortho Vitros ± ± ± ± ± 1.24 n = 22 [ROC] Roche cobas c ± ± ± ± ± 0.77 n = 3 [ROG] Roche cobas c ± ± ± ± ± 0.68 n = 9 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 1.79 n = 8 [ROD] Roche MODULAR D/P ± ± ± ± ± 1.98 n = 10 [COB] Siemens ADVIA Centaur ± ± ± ± ± 1.12 n = 9 [DUE] Siemens Dimension EXL ± ± ± ± ± 3.13 n = 3 [DUR] Siemens Dimension RxL ± ± ± ± ± 1.41 n = 47 [DUT] Siemens Dimension Vista ± ± ± ± ± 0.94 n = 3 [DUX] Siemens Dimension Xpand ± ± ± ± ± 1.06 n = 19 [AB1] Abbott ± ± ± ± ± 1.35 n = 20 [BC1] Beckman Coulter ± ± ± ± ± 0.76 n = 11 [OL1] Beckman Coulter AU Series ± ± ± ± ± 0.18 n = 3 [MG1] Microgenics CEDIA ± ± ± ± ± 1.17 n = 26 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 0.84 n = 11 [RO1] Roche Integra and MIRA ± ± ± ± ± 0.80 n = 7 [RO6] Roche ONLINE ± ± ± ± ± 1.98 n = 10 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± 0.45 n = 5 [BY5] Siemens ADVIA/Syva Emit ± ± ± ± ± 1.55 n = 62 [DA5] Siemens Dimension ± ± ± ± ± 0.43 n = 5 [SY4] Syva Emit
10 Phenytoin (mg/l) ± ± ± ± ± 2.33 n = 253 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 0.67 n = 5 [ABJ] Abbott Architect c System ± ± ± ± ± 0.59 n = 17 [ABH] Abbott Architect i System ± ± ± ± ± 1.77 n = 34 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 1.05 n = 11 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± 0.66 n = 9 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 0.95 n = 7 [JJF] Ortho Vitros 5,1FS ± ± ± ± ± 0.78 n = 21 [JJG] Ortho Vitros ± ± ± ± ± 1.39 n = 3 [ROJ] Roche cobas c ± ± ± ± ± 1.81 n = 22 [ROC] Roche cobas c ± ± ± ± ± 1.37 n = 4 [ROG] Roche cobas c ± ± ± ± ± 1.14 n = 10 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 0.63 n = 8 [ROD] Roche MODULAR D/P ± ± ± ± ± 3.82 n = 8 [BYE] Siemens ADVIA ± ± ± ± ± 3.30 n = 11 [COB] Siemens ADVIA Centaur ± ± ± ± ± 1.54 n = 18 [DUE] Siemens Dimension EXL ± ± ± ± ± 1.45 n = 3 [DUR] Siemens Dimension RxL ± ± ± ± ± 1.11 n = 48 [DUT] Siemens Dimension Vista ± ± ± ± ± 1.95 n = 5 [DUX] Siemens Dimension Xpand ± ± ± ± ± 0.77 n = 22 [AB1] Abbott ± ± ± ± ± 0.92 n = 22 [BC1] Beckman Coulter ± ± ± ± ± 2.13 n = 19 [OL1] Beckman Coulter AU Series ± ± ± ± ± 0.76 n = 4 [MG1] Microgenics CEDIA ± ± ± ± ± 0.84 n = 29 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± 1.70 n = 29 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 1.29 n = 12 [RO1] Roche Integra and MIRA ± ± ± ± ± 1.31 n = 7 [RO6] Roche ONLINE ± ± ± ± ± 3.14 n = 13 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± 2.90 n = 8 [BY5] Siemens ADVIA/Syva Emit ± ± ± ± ± 1.28 n = 74 [DA5] Siemens Dimension ± ± ± ± ± 1.22 n = 7 [SY4] Syva Emit
11 Free Phenytoin (mg/l) 1.49 ± ± ± ± ± 0.38 n = 20 [---] All Methods & Instruments 1.36 ± ± ± ± ± 0.42 n = 6 [OLC] Beckman Coulter AU Chemistry System 1.58 ± ± ± ± ± 0.16 n = 5 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 0.31 n = 3 [OL1] Beckman Coulter AU Series 1.58 ± ± ± ± ± 0.25 n = 7 [RO1] Roche Integra and MIRA 1.47 ± ± ± ± ± 0.37 n = 3 [SY4] Syva Emit
12 Primidone (mg/l) 7.77 ± ± ± ± ± 0.74 n = 13 [---] All Methods & Instruments [---] Weigh-in value 7.77 ± ± ± ± ± 0.47 n = 5 [OLC] Beckman Coulter AU Chemistry System 7.70 ± ± ± ± ± 0.54 n = 3 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 0.54 n = 3 [RO1] Roche Integra and MIRA 7.86 ± ± ± ± ± 0.48 n = 7 [SY2] Syva Emit 11
13 Procainamide (mg/l) ± ± ± ± ± 0.68 n = 10 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 0.56 n = 3 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 1.26 n = 3 [RO4] Roche cobas c311/c501/c502/c701/c702 12
14 N-Acetyl-Procainamide (mg/l) ± ± ± ± ± 0.47 n = 10 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 0.83 n = 3 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 0.09 n = 3 [RO4] Roche cobas c311/c501/c502/c701/c702 13
15 Quinidine (mg/l) 6.18 ± ± ± ± ± 0.46 n = 9 [---] All Methods & Instruments [---] Weigh-in value 6.28 ± ± ± ± ± 0.32 n = 3 [RO4] Roche cobas c311/c501/c502/c701/c702 14
16 Salicylate (mg/l) ± ± ± ± ± 2.32 n = 209 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 1.50 n = 20 [ABJ] Abbott Architect c System ± ± ± ± ± 4.12 n = 19 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 1.89 n = 10 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± 1.52 n = 8 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 3.69 n = 7 [JJF] Ortho Vitros 5,1FS ± ± ± ± ± 2.10 n = 20 [JJG] Ortho Vitros ± ± ± ± ± 1.49 n = 20 [ROC] Roche cobas c ± ± ± ± ± 0.56 n = 5 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 0.94 n = 7 [ROD] Roche MODULAR D/P ± ± ± ± ± 0.73 n = 8 [BYE] Siemens ADVIA ± ± ± ± ± 0.37 n = 18 [DUE] Siemens Dimension EXL ± ± ± ± ± 0.19 n = 3 [DUR] Siemens Dimension RxL ± ± ± ± ± 0.70 n = 46 [DUT] Siemens Dimension Vista ± ± ± ± ± 0.32 n = 5 [DUX] Siemens Dimension Xpand ± ± ± ± ± 1.50 n = 20 [AB1] Abbott ± ± ± ± ± 1.82 n = 18 [BC1] Beckman Coulter ± ± ± ± ± 4.21 n = 15 [OL1] Beckman Coulter AU Series ± ± ± ± ± 2.36 n = 30 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± 1.79 n = 24 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 0.94 n = 7 [RO2] Roche Hitachi and Modular D/P ± ± ± ± ± 0.78 n = 6 [RO1] Roche Integra and MIRA ± ± ± ± ± 0.96 n = 11 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± 0.58 n = 72 [DA5] Siemens Dimension ± ± ± ± ± 1.06 n = 3 [SY5] Syva Emit tox 15
17 Theophylline (mg/l) ± ± ± ± ± 2.91 n = 181 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 0.76 n = 3 [ABJ] Abbott Architect c System ± ± ± ± ± 1.68 n = 9 [ABH] Abbott Architect i System ± ± ± ± ± 4.03 n = 18 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 0.82 n = 9 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± 0.75 n = 8 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 3.13 n = 5 [JJF] Ortho Vitros 5,1FS ± ± ± ± ± 1.29 n = 18 [JJG] Ortho Vitros ± ± ± ± ± 0.73 n = 17 [ROC] Roche cobas c ± ± ± ± ± 1.35 n = 3 [ROG] Roche cobas c ± ± ± ± ± 0.57 n = 5 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 1.15 n = 9 [ROD] Roche MODULAR D/P ± ± ± ± ± 2.74 n = 10 [COB] Siemens ADVIA Centaur ± ± ± ± ± 2.00 n = 7 [DUE] Siemens Dimension EXL ± ± ± ± ± 1.36 n = 46 [DUT] Siemens Dimension Vista ± ± ± ± ± 1.45 n = 3 [DPD] Siemens Immulite ± ± ± ± ± 1.42 n = 12 [AB1] Abbott ± ± ± ± ± 0.78 n = 17 [BC1] Beckman Coulter ± ± ± ± ± 4.38 n = 12 [OL1] Beckman Coulter AU Series ± ± ± ± ± 1.72 n = 23 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± 0.86 n = 19 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 0.84 n = 4 [RO2] Roche Hitachi and Modular D/P ± ± ± ± ± 1.32 n = 7 [RO1] Roche Integra and MIRA ± ± ± ± ± 1.69 n = 6 [RO6] Roche ONLINE ± ± ± ± ± 2.74 n = 10 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± 0.59 n = 4 [BY5] Siemens ADVIA/Syva Emit ± ± ± ± ± 1.49 n = 57 [DA5] Siemens Dimension ± ± ± ± ± 1.45 n = 3 [DP5] Siemens Immulite ± ± ± ± ± 3.32 n = 6 [SY4] Syva Emit
18 Tobramycin (mg/l) ± ± ± ± ± 0.34 n = 70 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 0.10 n = 8 [ABJ] Abbott Architect c System ± ± ± ± ± 0.14 n = 8 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 0.27 n = 4 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 0.11 n = 5 [JJG] Ortho Vitros ± ± ± ± ± 0.17 n = 6 [ROC] Roche cobas c ± ± ± ± ± 0.04 n = 4 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 0.05 n = 4 [ROD] Roche MODULAR D/P ± ± ± ± ± 0.14 n = 5 [COB] Siemens ADVIA Centaur ± ± ± ± ± 0.09 n = 3 [DUE] Siemens Dimension EXL ± ± ± ± ± 0.14 n = 17 [DUT] Siemens Dimension Vista ± ± ± ± ± 0.10 n = 8 [AB1] Abbott ± ± ± ± ± 0.25 n = 6 [BC1] Beckman Coulter ± ± ± ± ± 0.13 n = 4 [OL1] Beckman Coulter AU Series ± ± ± ± ± 0.18 n = 6 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± 0.18 n = 6 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 0.06 n = 6 [RO1] Roche Integra and MIRA ± ± ± ± ± 0.05 n = 5 [RO6] Roche ONLINE ± ± ± ± ± 0.14 n = 5 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± 0.13 n = 20 [DA5] Siemens Dimension ± ± ± ± ± 0.11 n = 4 [SY4] Syva Emit
19 Valproic Acid (mg/l) ± ± ± ± ± n = 242 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 6.29 n = 5 [ABJ] Abbott Architect c System ± ± ± ± ± 8.79 n = 17 [ABH] Abbott Architect i System ± ± ± ± ± 9.91 n = 32 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± n = 10 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± 9.98 n = 8 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 6.74 n = 6 [JJF] Ortho Vitros 5,1FS ± ± ± ± ± n = 20 [JJG] Ortho Vitros ± ± ± ± ± n = 22 [ROC] Roche cobas c ± ± ± ± ± 6.40 n = 3 [ROG] Roche cobas c ± ± ± ± ± 7.00 n = 7 [ROT] Roche Cobas INTEGRA ± ± ± ± ± n = 9 [ROD] Roche MODULAR D/P ± ± ± ± ± 7.79 n = 8 [BYE] Siemens ADVIA ± ± ± ± ± n = 16 [COB] Siemens ADVIA Centaur ± ± ± ± ± 5.49 n = 14 [DUE] Siemens Dimension EXL ± ± ± ± ± 6.54 n = 4 [DUR] Siemens Dimension RxL ± ± ± ± ± 5.10 n = 48 [DUT] Siemens Dimension Vista ± ± ± ± ± n = 4 [DUX] Siemens Dimension Xpand ± ± ± ± ± n = 22 [AB1] Abbott ± ± ± ± ± n = 20 [BC1] Beckman Coulter ± ± ± ± ± 9.10 n = 20 [OL1] Beckman Coulter AU Series ± ± ± ± ± n = 27 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± n = 25 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 9.06 n = 9 [RO1] Roche Integra and MIRA ± ± ± ± ± n = 7 [RO6] Roche ONLINE ± ± ± ± ± n = 18 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± 8.59 n = 9 [BY5] Siemens ADVIA/Syva Emit ± ± ± ± ± 6.19 n = 70 [DA5] Siemens Dimension ± ± ± ± ± n = 10 [SY4] Syva Emit
20 Vancomycin (mg/l) ± ± ± ± ± 1.10 n = 219 [---] All Methods & Instruments [---] Weigh-in value ± ± ± ± ± 0.39 n = 5 [ABJ] Abbott Architect c System ± ± ± ± ± 0.33 n = 15 [ABH] Abbott Architect i System ± ± ± ± ± 0.40 n = 25 [OLC] Beckman Coulter AU Chemistry System ± ± ± ± ± 0.61 n = 8 [BCG] Beckman Coulter UniCel DxC ± ± ± ± ± 0.47 n = 9 [BCH] Beckman Coulter UniCel DxC ± ± ± ± ± 0.31 n = 6 [JJF] Ortho Vitros 5,1FS ± ± ± ± ± 0.35 n = 18 [JJG] Ortho Vitros ± ± ± ± ± 0.87 n = 20 [ROC] Roche cobas c ± ± ± ± ± 0.64 n = 3 [ROG] Roche cobas c ± ± ± ± ± 0.33 n = 7 [ROT] Roche Cobas INTEGRA ± ± ± ± ± 0.94 n = 7 [ROD] Roche MODULAR D/P ± ± ± ± ± 1.13 n = 5 [BYE] Siemens ADVIA ± ± ± ± ± 0.60 n = 16 [COB] Siemens ADVIA Centaur ± ± ± ± ± 0.44 n = 15 [DUE] Siemens Dimension EXL ± ± ± ± ± 0.53 n = 47 [DUT] Siemens Dimension Vista ± ± ± ± ± 0.48 n = 4 [DUX] Siemens Dimension Xpand ± ± ± ± ± 0.49 n = 20 [AB1] Abbott ± ± ± ± ± 0.63 n = 19 [BC1] Beckman Coulter ± ± ± ± ± 0.33 n = 14 [OL1] Beckman Coulter AU Series ± ± ± ± ± 0.33 n = 25 [JJ1] Ortho Clinical Diagnostics ± ± ± ± ± 0.81 n = 23 [RO4] Roche cobas c311/c501/c502/c701/c ± ± ± ± ± 0.46 n = 3 [RO2] Roche Hitachi and Modular D/P ± ± ± ± ± 0.45 n = 9 [RO1] Roche Integra and MIRA ± ± ± ± ± 1.53 n = 5 [RO6] Roche ONLINE ± ± ± ± ± 0.59 n = 18 [BY1] Siemens ADVIA/ADVIA Centaur ± ± ± ± ± 0.39 n = 6 [BY5] Siemens ADVIA/Syva Emit ± ± ± ± ± 0.55 n = 68 [DA5] Siemens Dimension ± ± ± ± ± 0.44 n = 9 [SY4] Syva Emit
Therapeutic Substance Monitoring/Quantitative Toxicology Proficiency Test Program. Statistical Summary - May 2016 (Event 16-2)
ANDREW M. CUOMO Governor HOWARD A. ZUCKER, M.D., J.D. Commissioner SALLY DRESLIN, M.S., R.N. Executive Deputy Commissioner Therapeutic Substance Monitoring/Quantitative Toxicology Proficiency Test Program
More informationTherapeutic Substance Monitoring/Quantitative Toxicology Proficiency Test Program. Statistical Summary - August 2016 (Event 16-3)
ANDREW M. CUOMO Governor HOWARD A. ZUCKER, M.D., J.D. Commissioner SALLY DRESLIN, M.S., R.N. Executive Deputy Commissioner Therapeutic Substance Monitoring/Quantitative Toxicology Proficiency Test Program
More informationStatistical Report: Quantitative Urine Clinical Chemistry Mail out date: October 26, 2015
ANDREW M. CUOMO Governor HOWARD A. ZUCKER, M.D., J.D. Commissioner SALLY DRESLIN, M.S., R.N. Executive Deputy Commissioner Statistical Report: Quantitative Urine Clinical Chemistry Mail out date: October
More informationBlood ph and Gases Proficiency Test Program. Statistical Summary November 2015 (Event 15-3)
ANDREW M. CUOMO Governor HOWARD A. ZUCKER, M.D., J.D. Commissioner SALLY DRESLIN, M.S., R.N. Executive Deputy Commissioner Blood ph and Gases Proficiency Test Program Statistical Summary November 2015
More informationState of the Art for Measurement of Urine Albumin: Comparison of Routine. Measurement Procedures to Isotope Dilution Tandem Mass Spectrometry
1 of 27 State of the Art for Measurement of Urine Albumin: Comparison of Routine Measurement Procedures to Isotope Dilution Tandem Mass Spectrometry Supplemental Information, Tables and Data 2 of 27 Supplemental
More informationBLOOD GAS ELECTROLYTE (ISE) CHEMISTRY (ISE)
PRODUCT LIST 01 02 03 04 05 BLOOD GAS ELECTROLYTE (ISE) CHEMISTRY (ISE) QUALITY CONTROLS ELECTRODES TABLE OF CONTENTS BLOOD GAS / ISE / CO-OXIMETRY ANALYZERS Siemens-Bayer (Ciba-Corning / Chiron) 238,248,348
More informationStatistical Summary Blood ph and Gas/Chemistry Proficiency Testing Event July 23, 2012
Statistical Summary Blood ph and Gas/Chemistry Proficiency Testing Event 12-2 - July 23, 2012 Attached is a summary of participant performance (mean and standard deviation) for the Blood ph and Gas/Chemistry
More informationStatistical Summary Blood ph and Gas/Chemistry Proficiency Testing Event 13-2 July 22, 2013
Statistical Summary Blood ph and Gas/Chemistry Proficiency Testing Event 13-2 July 22, 2013 Attached is a summary of participant performance (mean and standard deviation) for the Blood ph and Gas/Chemistry
More informationStatistical Summary Blood ph and Gas/Chemistry Proficiency Testing Event 12-3 November 26, 2012
Statistical Summary Blood ph and Gas/Chemistry Proficiency Testing Event 12-3 November 26, 2012 Attached is a summary of participant performance (mean and standard deviation) for the Blood ph and Gas/Chemistry
More informationDRI OXYCODONE APPLICATION BECKMAN COULTER DxC 700 AU
Microgenics Corporation Part of Thermo Fisher Scientific DRI OXYCODONE APPLICATION BECKMAN COULTER DxC 700 AU Beckman Coulter Reagent REF 100248 Intended for the qualitative and semiquantitative determination
More informationepoc Blood Analysis System: Summary of Analytical Methods and Performance
White Paper epoc Blood Analysis System epoc Blood Analysis System: Summary of Analytical Methods and Performance siemens.com/poc White Paper epoc Blood Analysis System epoc Blood Analysis System White
More informationSEQC EXTERNAL QUALITY ASSURANCE PROGRAMS
SPANISH SOCIETY OF CLINICAL BIOCHEMISTRY AND MOLECULAR PATHOLOGY SEQC EXTERNAL QUALITY ASSURANCE PROGRAMS Commutable Serum with Reference Values (SCR) Impact on standardization Beatriz Boned (1) / Emma
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Biochemistry Department Macclesfield District General Hospital Victoria Road Macclesfield SK10 3BL United Kingdom Contact: Jane E Clarke Tel:
More informationHematology Proficiency Test Program. Statistical Summary February 2016 (Event 16-1)
ANDREW M. CUOMO Governor HOWARD A. ZUCKER, M.D., J.D. Commissioner SALLY DRESLIN, M.S., R.N. Executive Deputy Commissioner Hematology Proficiency Test Program Statistical Summary February 2016 (Event 16-1)
More informationClinical Chemistry and Immunodiagnostics Markets
MEDICAL DEVICES PHARMACEUTICALS CHEMICALS FOOD & BEVERAGE ELECTRONICS Clinical Chemistry and Immunodiagnostics Markets NEW PRODUCT DEVELOPMENT OPPORTUNITES AND BUSINESS EXPANSION STRATEGIES FOR INSTRUMENT
More informationStatistical validation of reagent lot change in the clinical chemistry laboratory can confer insights on good clinical laboratory practice
Original Article Annals of Clinical Biochemistry 2014, Vol. 51(6) 688 694! The Author(s) 2014 Reprints and permissions: sagepub.co.uk/journalspermissions.nav DOI: 10.1177/0004563214520749 acb.sagepub.com
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2014 Beckman Coulter, Inc. All rights reserved. SALY Salicylate 378194 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date
More informationCEDIA CYCLOSPORINE PLUS (LOW RANGE) APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800
CEDIA CYCLOSPORINE PLUS (LOW RANGE) APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 Beckman Coulter Reagent REF A31849 The CEDIA Cyclosporine PLUS assay is for the in vitro quantitative determination
More informationTake your lab s performance to the peak with cobas 8000 modular analyzer series
Take your lab s performance to the peak with cobas 8000 modular analyzer series Take your lab s performance to the peak with cobas 8000 modular analyzer series While your lab is facing rising complexity
More informationRoles and responsibilities in verification of traceability of in vitro medical diagnostics (IVD)
Roles and responsibilities in verification of traceability of in vitro medical diagnostics (IVD) Federica Braga University of Milan Medical School Centre for Metrological Traceability in Laboratory Medicine
More informationCEDIA COCAINE OFT ASSAY APPLICATION (UK) BECKMAN COULTER AU680
Microgenics Corporation Part of Thermo Fisher Scientific CEDIA COCAINE OFT ASSAY APPLICATION (UK) BECKMAN COULTER AU680 Catalog No. 10014734, 10014740 The Thermo Scientific CEDIA Cocaine OFT Assay is intended
More informationCEDIA Cyclosporine PLUS High Range Application Ortho Clinical Diagnostics VITROS 5600 Integrated System, VITROS 5,1 FS and 4600 Chemistry Systems
Microgenics Corporation Part of Thermo Fisher Scientific CEDIA Cyclosporine PLUS High Range Application Ortho Clinical Diagnostics VITROS 5600 Integrated System, VITROS 5,1 FS and 4600 Chemistry Systems
More informationCEDIA OPIATE OFT ASSAY APPLICATION (UK) - Beckman Coulter AU680
Microgenics Corporation Part of Thermo Fisher Scientific CEDIA OPIATE OFT ASSAY APPLICATION (UK) - Beckman Coulter AU680 Catalog No. 10010612, 10010659 The Thermo Scientific CEDIA Opiate OFT Assay is intended
More informationCEDIA Cyclosporine PLUS Low Range Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 5,1 FS and 4600 Chemistry Systems
Microgenics Corporation Part of Thermo Fisher Scientific CEDIA Cyclosporine PLUS Low Range Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 5,1 FS and 4600 Chemistry Systems
More informationCUVETTES. Item # Unit Cuvette for Horiba ABX Mira Plus and Mira S Plus, 500/case
CUVETTES 20 2 23 Cuvette for Cobas Mira, Individually Wrapped, Bulk, /cs Cuvette for Cobas Mira, Individually Wrapped, Bulk, 300/cs Cuvette for Cobas Mira Changer, /Rack, 40/cs 300 40 60 Cuvette for Horiba
More informationBLOOD GAS ELECTROLYTE (ISE) CHEMISTRY (ISE)
PRODUCT LIST 01 02 03 04 05 BLOOD GAS ELECTROLYTE (ISE) CHEMISTRY (ISE) QUALITY CONTROLS ELECTRODES Our Company Phoenix Diagnostics is a premier alternative manufacturer of consumables for Blood Gas, Electrolyte,
More informationJAPANESE CHEMISTRY AND IMMUNODIAGNOSTICS MARKETS. Table of Contents
JAPANESE CHEMISTRY AND IMMUNODIAGNOSTICS MARKETS Introduction Table of Contents Worldwide Market and Technology Overview A. Major Routine Chemistry Tests 1. Albumin 2. Alkaline Phosphatase 3. ALT/SGPT
More informationCX4, CX5, CX7, CX9 PRO
Beckman CX3, CX4CE, CX5CE, CX7, Synchron Beckman CX3, CX4, CX5, CX7 Delta, CX9 ALX Beckman CX3, CX4, CX5, CX7, CX9 PRO BK-A17652D A17652 Electrolyte Buffer Reagent 4.7 CX 3,5,7,9 2x2000mL BK-A17657D A17657
More informationCEDIA Tacrolimus Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 5,1 FS and 4600 Chemistry Systems
Microgenics Corporation Part of Thermo Fisher Scientific CEDIA Tacrolimus Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 5,1 FS and 4600 Chemistry Systems Catalog No. 10008656
More informationCEDIA COCAINE OFT ASSAY APPLICATION (EU) BECKMAN COULTER AU680
Microgenics Corporation Part of Thermo Fisher Scientific CEDIA COCAINE OFT ASSAY APPLICATION (EU) BECKMAN COULTER AU680 Catalog No. 10014734, 10014740 The Thermo Scientific CEDIA Cocaine OFT Assay is intended
More informationProduct Information. IVD Catalogue 2016 COMING MEDICAL. Hematology Analyzer Reagent. Biochemistry Analyzer Reagent & ISE Solution
IVD Catalogue 2016 Product Information Hematology Analyzer Reagent Biochemistry Analyzer Reagent & ISE Solution Chemiluminescence Immunoassay System Reagent Coagulation Analyzer Reagent COMING MEDICAL
More informationDRI Ecstasy Assay Application BECKMAN COULTER UniCel DxC and Synchron System(s)
DRI Ecstasy Assay Application BECKMAN COULTER UniCel DxC and Synchron System(s) Beckman Coulter Reorder Number A39935 Homogeneous enzyme immunoassay for the qualitative or semi-quantitative determination
More informationProduct Information. IVD Catalogue 2014 COMING MEDICAL. Hematology Analyzer Reagent. Biochemistry Analyzer Reagent & ISE Solution
IVD Catalogue 2014 Product Information Hematology Analyzer Reagent Biochemistry Analyzer Reagent & ISE Solution Chemiluminescence Immunoassay System Reagent Coagulation Analyzer Reagent COMING MEDICAL
More informationFuture US Clinical Chemistry and Immunodiagnostics Market: Growth Opportunities and Business Expansion Strategies.
Future US Clinical Chemistry and Immunodiagnostics Market: Growth Opportunities and Business Expansion Strategies Introduction Table of Contents Worldwide Market and Technology Overview A. Major Routine
More informationGentian Canine CRP Immunoassay Application Note for Siemens ADVIA * 1800/1650
v03-jan2014 Gentian Canine CRP Immunoassay Application Note for Siemens ADVIA * 1800/1650 Intended Use The canine CRP immunoassay on Siemens ADVIA 1800/1650 is an in vitro diagnostic test for quantitative
More informationEvaluation of Survey results for regulatory grading
Important information regarding grading of specimen C -02 Specim en C-02 w as a carefully prepared fresh frozen hum an serum sam ple that w as sent to all participants in the Survey. Your laboratory s
More informationguide [ tech ] Chemistry Analyzers Abbott Park, Ill (847) Brea, Calif (800)
Chemistry Analyzers Abbott Beckman Coulter Beckman Coulter Abbott Park, Ill (847) 937-6100 www.abbottdiagnositcs.com Name of product Abbott ARCHITECT immunochemistry analyzers AU480 AU680 1) What does
More informationApplication Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010
Page of 6 Application Note for Gentian Cystatin C Immunoassay on For in vitro diagnostic use only Cobas c50 (Optimised Volumes II), Roche Diagnostics v06-september 200 General Information Intended Use
More informationTOX. Scheme Description. Toxicology Proficiency Testing Scheme
TOX Toxicology Proficiency Testing Scheme Scheme Description LGC Standards Proficiency Testing 1 Chamberhall Business Park Chamberhall Green Bury Lancashire BL9 0AP United Kingdom Telephone: +44 (0) 161
More informationHematology Proficiency Test Program. Statistical Summary June 2016 (Event 16-2)
ANDREW M. CUOMO Governor HOWARD A. ZUCKER, M.D., J.D. Commissioner SALLY DRESLIN, M.S., R.N. Executive Deputy Commissioner Hematology Proficiency Test Program Statistical Summary June 2016 (Event 16-2)
More informationFor In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Potassium REF & 467915 467935 REF & A28945 A28937 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date Refer to coversheet in front of
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2017 Beckman Coulter, Inc. All rights reserved. TPm Total Protein 465986 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date
More informationFor In Vitro Diagnostic Use. Rx Only. ANNUAL REVIEW Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2017 Beckman Coulter, Inc. All rights reserved. IBCT Total Iron Binding Capacity 465970 465971 465976 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed
More informationPotassium Enzymatic Assay Kit (Liquid Stable) Catalog Number: BQ 010-EAEL
Note: Serum specimens and all materials coming in contact with them should be handled and disposed as if capable of transmitting infection. Avoid contact with skin by wearing gloves and proper laboratory
More informationAnalysis of Reagent Lot-to-Lot Comparability Tests in Five Immunoassay Items
Available online at www.annclinlabsci.org Analysis of Reagent Lot-to-Lot Comparability Tests in Five Immunoassay Items 165 Hyun Soo Kim 1, Hee Jung Kang 1, Dong Hee Whang 2, Seong Gyu Lee 3, Min Jeong
More informationCALC PRINCIPLE REF & REF A28937 & A ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Calcium REF 467935 & 467915 REF A28937 & A28945 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed
More informationCME/SAM. An Examination of the Usefulness of Repeat Testing Practices in a Large Hospital Clinical Chemistry Laboratory
Clinical Chemistry / Repeat Testing Practices in Clinical Chemistry An Examination of the Usefulness of Repeat Testing Practices in a Large Hospital Clinical Chemistry Laboratory Carl O. Deetz, MD, PhD,
More informationcobas pro integrated solutions Simplicity meets Excellence
cobas pro integrated solutions Simplicity meets Excellence Introducing cobas pro integrated solutions Simplicity meets Excellence 01 02 03 2 01 Sample supply unit Up to 1,000 samples per hour processing
More informationActivity. This procedure is to be used in conjunction with the package insert for PLAC Test for Lp-PLA 2
Test for Lp-PLA 2 Abbott ARCHITECT c8000 and ci8200 Chemistry Analyzer Application Sheet This procedure is to be used in conjunction with the package insert for Test for Lp-PLA 2. Read and follow the package
More informationDate. Urea nitrogen measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.
SYNCHRON CX 3 DELTA System Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Urea REF 443350 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED
More informationANALYZER CONSUMABLES DISPOSABLE LABORATORY PLASTICWARE
DISPOSABLE LABORATORY PLASTICWARE ANALYZER CONSUMABLES BIOHAZARD SPECIMEN BAGS BLOOD COLLECTION PRODUCTS CENTRIFUGE TUBES CONTAINERS CYTOLOGY PRODUCTS DRUG TEST CONTAINERS FUNNELS HISTOLOGY PRODUCTS MICROBIOLOGY
More information2014 Consumables. Blood Gas Electrolyte Chemistry Hematology Coagulation
2014 Consumables Blood Gas Electrolyte Chemistry Hematology Coagulation Instrumentation Diamond Diagnostics Inc. is a worldwide supplier of remanufactured and new Clinical Laboratory Instrumentation and
More informationSummary of Analytical Methods and Performance The epoc Blood Analysis System. Data Compilation provided by Alere
Summary of Analytical Methods and Performance The epoc Blood Analysis System Data Compilation provided by Alere Table of Contents Page Introduction.... 4 Glossary... 5 Methodologies.... 6 ph.... 8 pco
More informationGentian Canine CRP Immunoassay Application Note for ABX Pentra * 400
v02-jan2014 Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400 Intended Use The canine CRP immunoassay on Horiba Medical's ABX Pentra 400 is an in vitro diagnostic test for quantitative
More informationFor In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Iron REF 467910 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date Refer to coversheet in front of method PRINCIPLE INTENDED USE reagent,
More informationReporting Drug Concentrations: Mass or Molar Units?
Reporting Drug Concentrations: Mass or Molar Units? A Recommendation from the AACB, ASCEPT, RCPA, RACP Common Units for Reporting Drug Concentrations Working Party Submitted to the RCPA, AACB, ASCEPT and
More informationProduct Catalogue 2016
Product Catalogue 2016 CLINICHEM Ltd H-1117 Budafoki street 111-113 Budapest, Hungary Content Immunochemistry reagents Standards, controls Concentrated reagents (NEW) Supplying quality for clinical laboratories
More informationClinical Chemistry 62: (2016) Evidence-Based Medicine and Test Utilization
Clinical Chemistry 62:9 1255 1263 (2016) Evidence-Based Medicine and Test Utilization Analytical Bias Exceeding Desirable Quality Goal in 4 out of 5 Common Immunoassays: Results of a Native Single Serum
More information2009 Consumables. Blood Gas Electrolyte Chemistry Hematology Coagulation
2009 Consumables Blood Gas Electrolyte Chemistry Hematology Coagulation Instrumentation Diamond Diagnostics Inc., parent company of Mission Diagnostics, is a worldwide supplier of remanufactured and new
More informationNew York State Department of Health Wadsworth Center Blood ph and Gas/Chemistry Educational Proficiency Testing February 7, 2011
New York State Department of Health Wadsworth Center Blood ph and Gas/Chemistry Educational Proficiency Testing February 7, 2011 Glucose (mg/dl) 87.7 ± 3.63 160.2 ± 4.78 230.5 ± 7.04 266.0 ± 8.81 52.5
More informationCarbon Dioxide 2015 Beckman Coulter, Inc. All rights reserved.
SYNCHRON System(s) CO 2 Chemistry Information Sheet Carbon Dioxide 2015 Beckman Coulter, Inc. All rights reserved. A28937 & A28945 472515 & 472481 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed
More informationFor in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date
AU US Instructions For Use 2017 Beckman Coulter, Inc. All rights reserved. FER Ferritin OSR61203 4 x 24 ml R1, 4 x 12 ml R2 For in vitro diagnostic use only. ANNUAL REVIEW For Rx use only Reviewed by Date
More informationINSTRUCTIONS FOR USE
INSTRUCTIONS FOR USE VITROS Chemistry Products RF Reagent RF 680 1729 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products RF Reagent is used on the VITROS 5,1 FS Chemistry
More informationAdaptation o f EMIT D rug Assays to a Random-Access Automated Clinical Analyzer*
AN N A LS O F CLINICAL A N D LABORATORY SC IE N C E, Vol. 15, No. 1, Copyright 1985, Institute for Clinical Science, Inc. Adaptation o f EMIT D rug Assays to a Random-Access Automated Clinical Analyzer*
More informationPRINCIPLE. Kit Reorder # Kit Reorder # Kit Reorder # A17652 Kit Reorder # A ANNUAL REVIEW Reviewed by: Date.
SYNCHRON CX 3 DELTA System Chemistry Information Sheet Copyright 2007 Beckman Coulter, Inc. For In Vitro Diagnostic Use Chloride Kit Reorder # 450214 Kit Reorder # 472095 Kit Reorder # A17652 Kit Reorder
More informationauto-creatinine liquicolor
Design Verification auto-creatinine liquicolor Contents 1 Introduction...2 2 Imprecision...2 3 Linearity and Detection Limit...2 3.1 Linearity...2 3.2 Sensitivity...4 4 Comparison of Methods...4 5 Stability...5
More information2008 CM-A PARTICIPANT SUMMARY
2008 CM-A PARTICIPANT SUMMARY Evaluation Criteria QUALITATIVE For qualitative procedures in urinalysis, evaluation is based on participant consensus by method and instrument. For each analyte, a minimum
More informationReporting Drug Concentrations: Mass or Molar Units?
Reporting Drug Concentrations: Mass or Molar Units? Draft for Comment Invitation to respond The following document has been produced by a Working Party with members from the Royal College of Pathologists
More informationLinearity Program Users Guide
U Linearity Program Users Guide New for 2016 Two New CVL Programs LN43 Lamellar Body Count CVL LN44 Fibrinogen CVL s r me e C ea e Pro e o s e L C ea ti e rotein CVL ur ey i no an in trument e i rogram
More informationIDEXX VetTest Chemistry Analyzer. Lawrence A. Dwyer, Ph.D. January 2000 Senior Scientist
Performance of IDEXX VetTest Chemistry Analyzer Lawrence A. Dwyer, Ph.D. January 2 Senior Scientist IDEXX Laboratories, Inc. The Evaluation Objective: To correlate accuracy and precision of the IDEXX VetTest
More informationVERSION FOR SIEMENS DIMENSION VISTA 1500 AUTOMATED ANALYZER
VERSION FOR SIEMENS DIMENSION VISTA 1500 AUTOMATED ANALYZER NON SUPPLIED ADDITIONAL EQUIPMENT: Preparation of the reagents requires the use of micropipettes or systems suitable for dispensing the specified
More informationBeta-2 M Beta-2-Microglobulin
AU US Instructions For Use 2015 Beckman Coulter, Inc. All rights reserved. Beta-2 M Beta-2-Microglobulin OSR6151 4 x 10 ml R1, 4 x 8 ml R2 ANNUAL REVIEW For in vitro diagnostic use only. Reviewed by Date
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. NA Sodium A28937 & A28945 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date
More informationINSTRUCTIONS FOR USE
INSTRUCTIONS FOR USE VITROS Chemistry Products PHOS Slides PHOS 151 3209 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products PHOS Slides quantitatively measure phosphorus (PHOS)
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. K Potassium A28937 & A28945 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by
More informationVMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual
Page 1 / 11 VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual ZV-3030-0200-20 200 2-8 C Page 2 / 11 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE... 3
More informationVERSION FOR WERFEN ILAB 600 AND 650 AUTOMATED ANALYZER
VERSION FOR WERFEN ILAB 600 AND 650 AUTOMATED ANALYZER NON SUPPLIED ADDITIONAL EQUIPMENT: Preparation of the reagents requires the use of micropipettes or systems suitable for dispensing the specified
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2017 Beckman Coulter, Inc. All rights reserved. CREm Creatinine 472525 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2017 Beckman Coulter, Inc. All rights reserved. CR-S Creatinine A40920 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2017 Beckman Coulter, Inc. All rights reserved. BUN Urea Nitrogen 442750 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date
More informationA Rapid Toxicology Screen for Emergency and Routine Care of Patients
CLIN. CHEM. 34/1, 158-162 (1988) A Rapid Toxicology Screen for Emergency and Routine Care of Patients dive H. HamlIn I have adapted, for use with a centrifugal analyzer, reagents from 10 different EMIT
More informationKartell SpA LABWARE Division
KARTELL 1/7 Noviglio, 22/03/2011 Reference: PRODUCTS "CE" MARK Descrizione: IN VITRO DIAGNOSTIC MEDICAL DEVICES VESSELS DISPOSABLES TIPS Products List: see attached list 1 We undersigned, Kartell S.P.A.,
More informationBEIPH Report. QCMD 2010 Mycobacterium tuberculosis DNA (MTBDNA10) EQA Programme
QUALITY CONTROL for MOLECULAR DIAGNOSTICS The Altum Building, Todd Campus, West of Scotland Science Park, Glasgow, G20 0XA Scotland Tel: +44 (0) 141 945 6474 Fax: +44 (0) 141 945 5795 www.qcmd.org info@qcmd.org
More informationClinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for
Clinical Toxicology Page of 0 Materials and Methods Biomass Component Extraction: The uneaten cooked plant specimen was prepared for chemical analysis as follows. The sample was frozen, diced, pulverized
More informationMicro Volume QuEChERS kit
225-37872 Sep. 2018 Small Capacity Pretreatment Kit Micro Volume QuEChERS kit Instruction Manual Read this manual thoroughly before you use the product. Keep this manual for future reference. This page
More informationIntercept i2he Cannabinoids Oral Fluid Assay
FOR FORENSIC USE ONLY For Use with Oral Fluid Collection Device 1001-0390 (65 ml Kit) INTENDED USE The is intended for use in the qualitative determination of Cannabinoids in human oral fluid at a cutoff
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. CR-S Creatinine A40920 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date
More informationTHC PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Cannabinoid 100 ng REF 445990 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE
More informationORG1 Syntheses of Acetaminophen and Aspirin
RG1 Syntheses of Acetaminophen and Aspirin Estimated Time Required: 60 minutes Introduction Ethanoylation (better known as acetylation) is the introduction of an acetyl functional group onto a suitable
More informationThe Fluorometric Determination of Acetylsalicylic Acid in an Aspirin Tablet
The Fluorometric Determination of Acetylsalicylic Acid in an Aspirin Tablet Introduction: Fluorescence is the emission of radiation from an atom or polyatomic species after the substance has been exposed
More informationProfessional Product List 2018 PARTS AND CONSUMABLES
Trade diagnostical products for clinical laboratories products for life. Professional Product List 2018 PARTS AND CONSUMABLES SPARE PARTS AND SERVICES immunology chemistry ISE modules electrolyte ISE coagulation
More informationScience. Reagent Carryover Studies: Preventing Analytical Error with Open Clinical Chemistry Systems
Reagent Carryover Studies: Preventing Analytical Error with Open Clinical Chemistry Systems Joseph Boneno, MS, SC(ASCP), 1 Michelle Fokakis, MT(ASCP), 2 Dave Armbruster, PhD, DABCC, FACB, C(ASCP) 3 ( 1
More information01 English - Ref.: 95. Intended use. End point reaction system for determination of calcium in blood and urine.
Intended use. End point reaction system for determination of calcium in blood and urine. Test principle. Calcium reacts with arsenazo III in a solution a little acid yielding the blue colored complex calcium-arsenazo
More informationFor In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Barbiturates REF 475012 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date Refer to coversheet in front of method PRINCIPLE INTENDED
More informationINSTRUCTIONS FOR USE
VITROS Chemistry Products Slides 145 0261 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products Slides quantitatively measure calcium () concentration in serum, plasma, and urine
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. AMM Ammonia 439770 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date PRINCIPLE
More informationResults of the June 3, 2013 Hematology Proficiency Test
To: From: Laboratory Directors and Laboratory Staff Robert Rej, Ph.D. Date: June 28, 213 Subject: Results of the June 3, 213 Hematology Proficiency Test Enclosed are results from the hematology proficiency
More informationWhat clinical and forensic toxicologists expect
Evaluation of an Improved General Unknown Screening Procedure Using Liquid- Chromatography-Electrospray-Mass Spectrometry by Comparison with Gas Chromatography and High-Performance Liquid-Chromatography
More informationHematology Proficiency Test Program. Statistical Summary June 2014 (Event 14-2)
Hematology Proficiency Test Program Statistical Summary June 2014 (Event 14-2) This statistical report summarizes participant data for the Hematology proficiency survey shipped 9 June 2014. Five test samples
More informationPooja Diagnostic Laboratory Mulund, Mumbai, India
Pooja Diagnostic Laboratory Mulund, Mumbai, India Modern-day India is unrecognizable from the country that existed when the Pooja Laboratory was founded in 1987 and the two have developed in parallel.
More information