Accepted for publication in Analytical Chemistry, August, 2011 SUPPORTING INFORMATION FOR. address:

Size: px
Start display at page:

Download "Accepted for publication in Analytical Chemistry, August, 2011 SUPPORTING INFORMATION FOR. address:"

Transcription

1 Accepted for publication in Analytical Chemistry, August, 0 SUPPORTING INFORMATION FOR Microfabricated Gas Chromatograph for the Selective Determination of Trichloroethylene Vapor at Sub-Parts-Per-Billion Concentrations in Complex Mixtures Sun Kyu Kim,, Hungwei Chang,, Edward T. Zellers,,3* Department of Environmental Health Sciences, University of Michigan, Ann Arbor, MI Center for Wireless Integrated Microsystems, University of Michigan, Ann Arbor, MI Department of Chemistry, University of Michigan, Ann Arbor, MI address: ezellers@umich.edu The information provided below includes the rationale for using 6 m of total microcolumn length, the table of 46 test VOCs, a representative F heating profile, Golay plots and a representative chromatogram from the dual 3-m microcolumn ensemble, descriptions of the printed circuit boards and DAQ boards used for instrument control, and methods used for data analysis and chemometrics. Rationale for Two 3-m Microcolumns. The use of two microcolumns in series, each 3 m in length, can be rationalized on the basis of the required number of theoretical plates, N req, which is defined as follows for the peaks of a critical pair of compounds, and : S N k req 6Rs (eq. S) k where R s is the resolution, α is the separation factor, k is the retention factor of the later eluting compound (TCE in this case), and each of these variables is defined as follows: ( tr t ) R Rs.8 (eq. S) ( fwhm fwhm ) where t Ri is the retention time of compound i, fwhm i is the full width at half maximum the peak; and k k (eq. S3)

2 Accepted for publication in Analytical Chemistry, August, 0 k t t ' R M (eq. S4) where t R is the adjusted retention time and t M is the hold-up time (evaluated as the retention time of methane). For TCE, a k value of 3.3 and a fwhm value of.0 s were assumed on the basis of experimental data, S and values of =. and R s =.5 (i.e., baseline separation) were imposed. This yields an N req value of 7,400 plates. For operation at. ml/min, it is estimated that each microcolumn generates ~,400 plates per meter (derived by extrapolation of the Golay plot presented in Figure S; see below). Therefore, a total microcolumn length 5.3 meters would be required. Table S. List of 46 test compounds and their vapor pressures (p v ). ID Compound p v (torr) a ID Compound p v (torr) a Pentane 54 4 n-octane 4.,-Dichloroethane n-butyl Acetate 5 3 Methylene Chloride Chlorobenzene.8 4 Acetonitrile 73 7 Ethylbenzene Propanol (Isopropyl Alcohol) 44 8 m,p-xylenes Acrylonitrile 97 9 Bromoform 5 7 -Butanone (MEK) o-xylene n-hexane 50 3 Styrene Ethanol 60 3 n-nonane Tetrahydrofuran (THF) 6 33 Cumene 3.5 Chloroform alpha-pinene 4.75 Acetone 3 35 n-propylbenzene 3.4 3,,-Trichloroethane 00 36,,4-Trimethylbenzene. 4 Ethyl Acetate Ethyltoluene 3 5 Carbon Tetrachloride 3 38,3,5-Trimethylbenzene.48 6 Benzene d-limonene.98 7 Cyclohexane 98 40,,4-Trichlorobenzene Trichloroethylene (TCE) 69 4,4-dichlorobenzene Methyl--pentanone 0 4 Naphthalene ,,-Trichloroethane 7 43 n-decane.4 Toluene n-undecane Hexanone 45 n-dodecane 0. 3 Tetrachloroethylene (PCE) n-tridecane 0.08 a from reference S3.

3 Temperature ( C) Accepted for publication in Analytical Chemistry, August, s Time (s) Figure S. Representative (experimental) heating profile for the F during desorption/ injection. Application of 36 V leads to an increase 5 to 5 C in 0.45 s (440 C/s). Subsequent application of 6 V maintains the F between 5 and 50 C. Microcolumn Characterization. In order to compare the chromatographic efficiency of the current microcolumn design, with chamfered corners, to that of the previous design, with right-angle corners, the relationship between the linear velocity and the plate height was determined using n-octane as the test compound. For these experiments two series coupled 3-m microcolumns of a given design were connected between the injector and FID of a conventional GC (7890, Agilent Technologies, Pal Alto, CA) and mounted in the GC oven held at 30 C. Serial injections (000: split) of methane and n-octane (headspace above liquid, k = 3.7) were performed over a range of N carrier gas inlet pressures that produced a range of average linear velocities, u, estimated from the methane hold-up times. Retention times and peak widths were used to calculate the number of theoretical plates, N, from which the theoretical plate height, H, was determined (i.e., H = N/L, where L is the microcolumn length; N = 5.54(t R /fwhm), where t R is the adjusted retention time and fwhm is the full-width at half maximum of the peak). These data were used to create the Golay plots for n-octane shown in Figure S. The value of the optimal velocity, u opt, was ~ 0 cm/s (0. ml min - ) for both types of microcolumns, but those with the chamfered corners yielded a minimum value of H = 0.0 cm, which was 0% smaller than those with sharp corners (minimum H = 0.07 cm). This results in a commensurate increase in N produced by the dual 3-m microcolumn ensemble from,000 to 7,300 plates (i.e., 4,550 plates m - ). 3

4 H (cm) Accepted for publication in Analytical Chemistry, August, 0 a) b) c) u (cm/s) Figure S. SEM images of sub-sections of the etched-si channels used in the 3-m-long microcolumns of the GC prototype prior to sealing and coating with PDMS stationary phase: a) previous design with right-angle corners; b) current design with chamfered corners. The Golay plots in c) were generated with n-octane (000: split, k = 3.7) using N as the carrier gas by connecting the dual 3-m microcolumn ensemble between the injector and FID of a bench-scale GC: previous design (filled symbols) and current design (unfilled symbols). The chromatogram in Figure S3 shows the.-min separation of a 0-component mixture, including TCE, using the dual 3-m microcolumns of the current design, configured as described above for generating the data in Figure S. The separation conditions, including the temperature program used with the integrated microcolumn heaters, are given in the caption of Figure S3. All compounds were baseline separated, TCE eluted in about 45 s, and the entire mixture eluted in <. min. The fwhm values ranged from 0.6 to.9 s. On the basis of 6 replicate injections, the retention time variations ranged from 0.35 to 0.89% of the average values. 4

5 Accepted for publication in Analytical Chemistry, August, 0 TCE Time (min) Figure S3. TCE separation from 0 VOC interferences using a conventional (bench scale) GC inlet/injection port and FID, and the dual 3-m microcolumns of the current design. (0. L injection of the neat mixture; inlet pressure: 4 psi; inlet temperature: 50 o C; split ratio: 00:; Temperature program of st microcolumn: hold at 5 C for 60 s, heat to 60 C at 70 C/min, heat to 00 C at 80 C/min, hold at 00 C for 30 s. Temperature program of nd microcolumn: hold at 5 C for 60 s, heat to 60 C at 70 C/min, heat to 0 C at 0 C/min, hold at 0 C for 30 s. Compounds:, n-hexane;, benzene; 3, TCE; 4, toluene; 5. -hexanone; 6, PCE; 7, ethylbenzene; 8, o-xylene; 9, nonane; 0, cumene;, n-propylbenzene. Microcolumn Temperature Programming. For applied dc biases of 5-5 V (0.6-5 W), the corresponding steady-state microcolumn temperature ranged from C, respectively. The maximum heating rate, taken from the initial stages of the 5-W curve is 7.6 C s -. Higher heating rates may be possible but were not explored. Actuation and Control Circuitry. A custom pneumatic control circuit board and associated digital I/O card (USB-650, National Instruments, Austin, TX) for actuating the valves, pumps, and the heaters on the pre-trap and sampler, were located beneath the manifold on the chassis floor. A second printed circuit board and associated 6-bit multi-functional DAQ card (USB-68, National Instrument, Austin, TX) for monitoring and controlling the devices in the analytical subsystem (i.e., the µf and microcolumn heaters and temperature sensors, and the sensors in the CR array ) were located beneath the analytical subsystem components. A USB hub permitted connections to a laptop computer running a control program written in LabView (Ver. 8.5, National Instruments, Austin, TX). 5

6 Accepted for publication in Analytical Chemistry, August, 0 Table S. Timetable for SPIRON µgc operation. Component Sampling (5-6 min) Focusing (3 min) Stabilization (3 min) Separation & Detection (3.5 min) Manifold valve On On Off Off Off Off Off Manifold valve Off Off Off Off Off Off Off Manifold valve 3 Off On Off Off Off Off Off Manifold valve 4 Off Off Off Off Off Off Off Manifold valve 5 Off Off Off Off Off Off Off Manifold valve 6 Off Off On On On On On Analysis pump Off Off On On On On On Sampling pump On On Off Off Off Off Off Pretrap Off Off Off Off Off Off Off Sampler Off On Off Off Off Off Off Focuser Off Off Off On Off Off Off Column Off Off On On On On On Sensor Off Off On On On On On Data Analysis and Chemometrics. Peak heights and peak areas were determined after importing the raw response data into GRAMS AI/3 (Ver. 6.00, Thermo Scientific Inc., West Palm Beach, FL), and linear regressions of calibration data were performed using Excel. The performance of the CR array in differentiating among TCE and several potential interfering VOCs eluting nearby was assessed using Monte Carlo simulations coupled with extended disjoint principal components regression (EDPCR) classification models. Details of this approach to array assessment have been published elsewhere (see, for example, refs 4d, 4g, and 4h in the main article) and are summarized in the following paragraph. Using the experimental sensitivity values, synthetic MPN-CR responses to each vapor were generated by randomly selecting a vapor concentration within the range of 5-0 LOD, where the LOD was dictated by the least sensitive sensor in the array to ensure that all sensors contributed to the response patterns. The response was calculated from the calibration-curve regression equation for each sensor. Then, error was introduced by adding to the response a value obtained by multiplying that response value by a factor derived from randomly sampling a Gaussian distribution with a mean of zero and a standard deviation corresponding to the random sensitivity errors derived from the calibration data (Figure 3, main body) for each sensor for TCE (i.e., C8, 8.%; DPA,.7%; OPH,.%; HME, 9.5%). The error enhanced responses from all sensors were combined and the location of the resulting response vector was projected onto the principal component corresponding to the original calibrations for each vapor via EDPCR. The identity of the vapor assigned to this synthetic response vector was determined by the shortest Euclidean distance. This procedure was performed iteratively (i.e., 500 samples) to yield a statistical estimate of recognition rate (RR) for each vapor. 6

7 Accepted for publication in Analytical Chemistry, August, 0 Table S3. Confusion matrix for single-vapor discrimination. a Compound HEX BEN TCE MIBK TOL HEX BEN TCE MIBK TOL Recognition rate (%) a Based on Monte Carlo simulations and EDPCR classification models (see text); HEX = n-hexane, BEN = benzene, TCE = trichloroethylene, MIBK = 4-methyl--pentanone, TOL = toluene; actual identities are listed in the top row and assigned identities are listed in the first column; n = 500 iterations for each vapor. References S. Jennings, W.; Mittlefehldt, E.; Stremple, P. Analytical Gas Chromatography, nd Ed., Academic Press, 997. S. Chang, H.; Kim, S. K.; Sukaew, T.; Bohrer, F.; Zellers, E. T. Procedia Engineering 00, 5, S3. Department of Chemistry at the University of Akron, The Chemical Database, available at 7

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley MATRIX Reference Code Analyte Code Description ASTM C471M- 14 2014 ASTM D5504-12 2012 3964 Orthorhombic Cyclooctasulfur 4842 1-Propanethiol 6113 2,5-Dimethylthiophene 4544 2-Ethylthiophene 4843 2-Propanethiol

More information

Headspace Technology for GC and GC/MS: Features, benefits & applications

Headspace Technology for GC and GC/MS: Features, benefits & applications Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over

More information

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)

More information

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the Introduction Although analysis of VOCs by purge and trap is considered a mature technique, advances in

More information

Methanol Extraction of high level soil samples by USEPA Method 8260C

Methanol Extraction of high level soil samples by USEPA Method 8260C Methanol Extraction of high level soil samples by USEPA Method 8260C Abstract In order to determine the concentration of Volatile Organic Compounds (VOCs) in soil and waste samples the USEPA developed

More information

Target Compound Results Summary

Target Compound Results Summary Phone/ (86)88-4800 / (86)88-471 http://www.emsl.com to1lab@emsl.com Customer ID: EMSL0A K12312.D 28 cc 0 0/01/2017 K12327.D 21 cc 600 Target Compound Results Summary Propylene 11-07-1 42.08 ND.0 ND 8.6

More information

Thermal Desorption Technical Support

Thermal Desorption Technical Support Thermal Desorption Technical Support Note 86a: US EPA Method TO-17 for Monitoring Air Toxics in Ambient Air Using Sorbent Tubes and Automated, Cryogen-free Thermal Desorption Application Note Environmental,

More information

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Application Note US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Author Amy Nutter Applications Chemist, Teledyne Tekmar Abstract US EPA Method 8260 in conjunction

More information

EPA TO-17 Volatile Organic Compounds

EPA TO-17 Volatile Organic Compounds EPA TO-17 Volatile Organic Compounds Method TO-17 is used to analyze samples for volatile organic compounds collected on multi-bed sorbent tubes, which are thermally desorbed and cryo-focused on the capillary

More information

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd. 5/26/2017 Ms. Anna Kelley Hamilton Co. Environmental Services 250 Wm. Howard Taft Cincinnati OH 45219 Project Name: Sun Coke Project #: Workorder #: 1705303 Dear Ms. Anna Kelley The following report includes

More information

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions Validation of Volatile Organic Compound by USEPA Method 8260C Application Note Abstract In order to determine the concentration of volatile organic compounds (VOCs) in water and soil matrices the USEPA

More information

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Application Note Abstract Due to regional shortages and increasing costs of helium, the preferred carrier gas in gas chromatography, alternative

More information

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS Application Note By Tyler Trent Abstract In order to determine the concentration of volatile organic compounds

More information

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler Application Note By: Anne Jurek Abstract In order to determine the amount

More information

Maximizing Sample Throughput In Purge And Trap Analysis

Maximizing Sample Throughput In Purge And Trap Analysis Maximizing Sample Throughput In Purge And Trap Analysis LINDSEY PYRON ANNE JUREK INTRODUCTION There are several demands and requirements imposed on chemists performing volatile organic analysis (VOC) in

More information

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Validation of USEPA Method 524.2 Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Application Note By: Nathan Valentine Abstract The US EPA developed Method 524.2¹, Measurement

More information

Solid Phase Micro Extraction of Flavor Compounds in Beer

Solid Phase Micro Extraction of Flavor Compounds in Beer Solid Phase Micro Extraction of Flavor s in Beer ANNE JUREK USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler Application Note Environmental

More information

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Residual solvents analysis using an Agilent Intuvo 9000 GC system Residual solvents analysis using an Intuvo GC system Technology advantage: simplified dual column analysis with Flow Chip modularity Introduction USP defines a method for analysis of residual solvents

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC / 5977A MS Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Abstract US EPA Method

More information

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers Electronic Supplementary Material (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 2017 Electronic Supplementary Material Experimentally Validated Mathematical

More information

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator Application Note Introduction Environmental laboratories have utilized the Purge & Trap sample concentration technique for over

More information

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing

More information

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1806 P+T VOC ENV US EPA Method 524.2 with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS Amy Nutter, Applications Chemist; Teledyne

More information

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas A Comparison of Volatile Organic Compound When Using Nitrogen as a Gas Application Note By: Anne Jurek Abstract For many years Helium has been the gas of choice for purging Volatile Organic Compounds (VOCs).

More information

Optimizing. Abstract: is standardd. procedures. altered to

Optimizing. Abstract: is standardd. procedures. altered to Optimizing Standard Preparation ANNE JUREK Abstract: Standardd preparation can often be a time consuming, tedious process. The opportunity for human error or inconsistencies between individual preparation

More information

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ Application Note Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ -3 Application Note Abstract This application note demonstrates the

More information

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler Application Note By: Teri Dattilio Introduction Purge and Trap concentration

More information

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System Application Note Residual Solvent Analysis of USP Method Residual Solvents on the Agilent 889 GC System Author Lukas Wieder, Jie Pan, and Rebecca Veeneman Agilent Technologies, Inc. 8 Centerville Road

More information

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in

More information

Analytical Trap Comparison for USEPA Method 8260C

Analytical Trap Comparison for USEPA Method 8260C Analytical Trap Comparison for USEPA Method 8260C Application Note Abstract Purge and trap concentration is a technique that is used for the analysis of Volatile Organic Compounds (VOCs). The major component

More information

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Application Note Abstract Water quality laboratories across the

More information

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C APPLICATION NOTE 10441 Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C Authors Andrea Caruso, 1 Tommaso Albertini, 1 Jacob A. Rebholz 2 ; 1 Thermo Fisher Scientific,

More information

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products. Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products. CHRISTOPHER M. ENGLISH, CHRISTOPHER S. COX, FRANK

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Application Note US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Thermo Scientific TRACE 1310 GC and ISQ LT MS System Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract

More information

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography Fast Analysis of Aromatic Solvent with.8 mm ID GC column Application Gas Chromatography Authors Yun Zou Agilent Technologies (Shanghai) Co. Ltd. Ying Lun Road Waigaoqiao Free Trade Zone Shanghai 3 P.R.

More information

Date Issued: April 01, 2013 Expiration Date: June 30, 2013

Date Issued: April 01, 2013 Expiration Date: June 30, 2013 State of Florida Department of Health~Bureau of Public Health Laboratories I his is to certify that ALS ENVIRONMENTAL - SIMI VALLEY 2655 PARK CENTER DRIVE, SUITE A SIMI VALLEY, CA 93065 has complied with

More information

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP Application Note Joseph M. Levy Michael Kraft Abstract Agilent Equipment 7890A

More information

Study of Residual Solvents in Various Matrices by Static Headspace

Study of Residual Solvents in Various Matrices by Static Headspace Application Note Abstract United States Pharmacopeia (USP) chapter is a widely used method for identifying and quantifying Organic Volatile Impurities (OVI) used in the production of pharmaceuticals.

More information

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method

More information

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key

More information

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek JSB is an authorised partner of Analysis of Low Level Volatile Organic s in Air Anne Jurek #111 Introduction The New Jersey Department of Environmental Protection (NJDEP) developed a new low level TO-15

More information

Determination of Volatile Organic Compounds in Air

Determination of Volatile Organic Compounds in Air JSB is an authorised partner of Determination of Volatile Organic s in Air Anne Jurek #126 Introduction Air pollution is a growing problem due to the global economy and industrial development in many countries.

More information

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to. Optimal Conditions for USEPA Method 8260 Analysis Anne Jurek Introduction: Over the past decade, the need for environmental laboratoriess using purge and trap systems to report at or below the Method Detection

More information

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler Validation of USEPA Method 524.2 Using a Stratum PTC and the New AQUATek 100 Autosampler Application Note Abstract Automation is the key to increasing laboratory productivity and minimizing costs. It is

More information

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument Application Note Abstract The US Pharmacopeia recently released USP as the current monograph for determining residual solvents in pharmaceutical products by static headspace. The USP classified these

More information

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Introduction Purge and Trap (P&T) concentrators

More information

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds Permit Guidance 12 Final National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds Statutory references: ORC 6111.01, 6111.03, 6111.04 Rule references: OAC 3745-1-05,

More information

Maximizing Production While Minimizing Costs

Maximizing Production While Minimizing Costs JSB is an authorised partner of Maximizing Production While Minimizing Costs Anne Jurek #123 Abstract: The need to increase productivity and reduce cost is becoming of greater concern to environmental

More information

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to. ANNE JUREK Optimal Conditions for USEPA Method 8260B Analysis Using the EST Analyticall Sampling System and the Shimadzu GCMS-QP2010s Introduction: The USEPA developedd Method 8260B for the determination

More information

Optimizing Standard Preparation

Optimizing Standard Preparation JSB is an authorised partner of Optimizing Standard Preparation Anne Jurek #106 Abstract: Standard preparation can often be a time consuming, tedious process. The opportunity for human error or inconsistencies

More information

Date: March 6, 2008 RWDI Reference #: W B Pages (Including Cover):

Date: March 6, 2008 RWDI Reference #: W B Pages (Including Cover): RWDI AIR Inc. 650 W oodlawn R oa d W est, Guelph, Ontari o, Ca nad a N1K 1B8 Tel: (51 9) 823-1 31 1 E-mail: info@rwdi.com Fax: (51 9) 823-1 31 6 W eb: www. r wdi.com Date: March 6, 2008 RWDI Reference

More information

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Meeting NJ Low Level TO-15 Air Testing Method Requirements Meeting NJ Low Level TO-15 Air Testing Method Requirements By Jason S. Herrington, PhD Abstract The following study evaluated the efficacy of pairing a Markes Unity with CIA Advantage preconcentrator with

More information

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 JSB is an authorised partner of A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 Anne Jurek #101 Introduction: The United States Environmental Protection Agency (USEPA)

More information

CALA Directory of Laboratories

CALA Directory of Laboratories CALA Directory of Laboratories Membership Number: 2628 Laboratory Name: Caduceon Environmental Laboratories (Richmond Hill) Parent Institution: Caduceon Enterprises Inc. Address: 110 West Beavercreek Rd.

More information

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Application Note Environmental Authors Peter Gautschi and Harry Prest Senior Application Scientist

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds INSTRUMENTS IN. Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic ompounds Application Note 00 Preconcentrator G/MS Analysis Authors Thomas X. Robinson, Daniel B. ardin, Entech

More information

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System Fast Analysis of USP 7 Residual Solvents using the Agilent 789A GC and Low Thermal Mass (LTM) System Application Note Pharmaceutical Author Roger L Firor Agilent Technologies, Inc. 8 Centerville Road Wilmington,

More information

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air Environmental Applications Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air By Jason S. Herrington Abstract The following study evaluated the efficacy of using Nutech s 8900DS preconcentrator

More information

Texas Commission on Environmental Quality INTEROFFICE MEMORANDUM

Texas Commission on Environmental Quality INTEROFFICE MEMORANDUM Texas Commission on Environmental Quality INTEROFFICE MEMORANDUM To: Lorinda Gardner, Director, R15 Date: Carlos Rubinstein, Texas Border Area Director From: Valerie E. Meyers, Ph.D. Toxicology Section,

More information

COMPLIANCE EMISSIONS TEST California Dept. of Public Health Standard Method Version 1.1 and FloorScore Flooring Evaluation

COMPLIANCE EMISSIONS TEST California Dept. of Public Health Standard Method Version 1.1 and FloorScore Flooring Evaluation COMPLIANCE EMISSIONS TEST California Dept. of Public Health Standard Method Version 1.1 and FloorScore Flooring Evaluation SAMPLE DESCRIPTION & TESTING PARAMETERS Tesoro Woods submitted exemplars of their

More information

CHEM 429 / 529 Chemical Separation Techniques

CHEM 429 / 529 Chemical Separation Techniques CHEM 429 / 529 Chemical Separation Techniques Robert E. Synovec, Professor Department of Chemistry University of Washington Lecture 1 Course Introduction Goal Chromatography and Related Techniques Obtain

More information

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS Application Note Pharmaceuticals Authors Karine Jacq, Frank David, and Pat Sandra Research Institute

More information

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract. Application Note 8 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-5 Application Note Environmental, Canister, TO-5, Air Monitoring, Air Toxics Abstract This application

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Application Note Forensic Toxicology and Drug Testing Author Jared Bushey Agilent Technologies, Inc. 285 Centerville

More information

CHAPTER 6 GAS CHROMATOGRAPHY

CHAPTER 6 GAS CHROMATOGRAPHY CHAPTER 6 GAS CHROMATOGRAPHY Expected Outcomes Explain the principles of gas chromatography Able to state the function of each components of GC instrumentation Able to state the applications of GC 6.1

More information

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling JSB is an authorised partner of Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling Anne Jurek Introduction: #134 Static headspace sampling has always been a viable option

More information

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8 application Note Gas Chromatography/ Mass Spectrometry Authors Ruben Garnica Dawn May PerkinElmer, Inc. Shelton, CT USA Method by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ

More information

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application Pharmaceuticals Author Roger L. Firor Agilent Technologies, Inc. 2850 Centerville

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

J. Li et al. Correspondence to: S. D. Xie

J. Li et al. Correspondence to: S. D. Xie Supplement of Atmos. Chem. Phys., 15, 7945 7959, 2015 http://www.atmos-chem-phys.net/15/7945/2015/ doi:10.5194/acp-15-7945-2015-supplement Author(s) 2015. CC Attribution 3.0 License. Supplement of Characterization

More information

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR: SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD 21044 PREPARED FOR: HOWARD COUNTY PUBLIC SCHOOL SYSTEM 10910 ROUTE 108 ELLICOTT

More information

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s JSB is an authorised partner of Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s Anne Jurek #120 Introduction: The USEPA developed

More information

systems Dibyadeep Paul

systems Dibyadeep Paul Designing Thermal modulators for portable GC GC systems by Dibyadeep Paul A dissertation submitted in partial fulfillment of the requirements for the degree of Doctor of Philosophy (Mechanical Engineering)

More information

2017 EPA Method Update Rule and EPA Method 624.1

2017 EPA Method Update Rule and EPA Method 624.1 Introduction Method 624 is for the determination of volatile organic compounds in industrial discharges and other liquid environmental samples by gas chromatography combined with mass spectrometry (GC/MS).

More information

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL Challenges of Water Monitoring Volatile Organic Compounds (VOCs) can have negative health impacts even at ppb levels VOC concentrations can

More information

CASE STUDY : airmozone : 88 VOC from PAMS and TO14 Ozone precursors by FID from WATER or AIR 0 / 325 µg/m 3 or higher

CASE STUDY : airmozone : 88 VOC from PAMS and TO14 Ozone precursors by FID from WATER or AIR 0 / 325 µg/m 3 or higher CASE STUDY : VOC analysis in ambient air by FID dectection - Complete system airmozone Page 1 of 13 E X P E R T S I N G A S A N A L Y S I S CASE STUDY : airmozone : 88 VOC from PAMS and TO14 Ozone precursors

More information

Ed George and Anaïs Viven Varian, Inc.

Ed George and Anaïs Viven Varian, Inc. Application Note SI-02408 Evaluation of EPA Method 524.3: a New Draft Method for the Analysis of VOCs in Drinking Water using GC/MS and V:Results GC/MS Software Ed George and Anaïs Viven Varian, Inc. Introduction

More information

Solid Phase Micro Extraction of Flavor Compounds in Beer

Solid Phase Micro Extraction of Flavor Compounds in Beer Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Automed Sampling of Methanol Extractions Applicion Note Environmental Author Anne Jurek Applicions Chemist EST Analytical Cincinni, OH

More information

Improve lives IT S ABOUT IMPROVING PEOPLE S LIVES IT S ABOUT INTENT

Improve lives IT S ABOUT IMPROVING PEOPLE S LIVES IT S ABOUT INTENT THE DESIGNERS ROLE Different roles Improve lives IT S ABOUT IMPROVING PEOPLE S LIVES IT S ABOUT INTENT INTENT 1) DEFINE THE INTENT The road to hell THE ROAD TO HELL IS PAVED WITH GOOD INTENTIONS Toothbrush

More information

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS LECO Corporation; Saint Joseph, Michigan USA Key Words: GC-TOFMS,

More information

Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010

Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010 Secrets of GC Column Dimensions GC Columns and Consumables Mark Sinnott Application Engineer Folsom California March 12, 2010 Page 1 Secrets of GC Column Dimensions Do I have the right column phase? Resolution

More information

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS ENVIRONMENTAL application note U.S. EPA VOLATILE ORGANICS METHOD 524.2 USING PURGE AND TRAP GC/MS Cheri Coody, Public Health Laboratory, Mississippi State Health Department, Jackson, MS USA Michael J.

More information

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS Application Note Optimization of Method Parameters for the Evaluation of USEPA Method 524.2 Using a Purge and Trap with GC/MS Objective By: Glynda Smith In this paper, an evaluation of different methods

More information

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS HS-20 & GCMS-QP2010Ultra Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS Introduction Volatile Organic Compounds (VOCs) refer to a group

More information

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application Static Headspace Blood Alcohol Analysis with the G Network Headspace Sampler Application Forensics Author Roger L. Firor and Chin-Kai Meng Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-0

More information

USP<467> residual solvents

USP<467> residual solvents Application Note Pharmaceuticals USP residual solvents Applying the Agilent 977A MSD with the Agilent 797A headspace sampler and Agilent 7890B GC Authors Roger L. Firor and Mike Szelewski Agilent Technologies,

More information

Secrets of GC Column Dimensions

Secrets of GC Column Dimensions Secrets of GC Column Dimensions GC Columns and Consumables Simon Jones Application Engineer May 20, 2008 Slide 1 Secrets of GC Column Dimensions Do I have the right column phase? Resolution Equation Changes

More information

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas Agilent Porous Polymer PLOT Columns: New Products, Expanded Uses, Prices Cut in Half! Product Brief Need improved resolution of small volatile compounds? Didn't try a PLOT column due to high price, short

More information

In-Tube Extraction Sample Preparation for Gas Chromatography

In-Tube Extraction Sample Preparation for Gas Chromatography In-Tube Extraction Sample Preparation for Gas Chromatography CTC Analytics AG Industriestrasse 20 CH-4222 Zwingen Switzerland Tel: +41 61 7658100 Fax: +41 61 7658199 E-mail: support@ctc.ch Web: www.ctc.ch

More information

Selection of a Capillary

Selection of a Capillary Selection of a Capillary GC Column - Series 3 Mark Sinnott Application Engineer March 19, 2009 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector

More information

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist A Single Calibration for Waters and Soil Samples Performing EPA Method 8260 Anne Jurek Applications Chemist Abstract A Single Calibration Method for Water AND Soil Samples The United States Environmental

More information

Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method

Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method APPLICATION NOTE 0676 Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP method Authors Giulia Riccardino, Paolo Magni, Stefano

More information

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA Comparison of VOC : s, Target Lists, MAI Acenaphthene 83-32-9 Acetalaldehyde (Ethanal) 75-07-0 Acetone 67-64-1 Acetonitrile 75-05-8 Acetophenone 96-86-2 Acrolein (Propenal) 107-02-8 Acrylamide 79-06-1

More information

Application. Gas Chromatography February Introduction

Application. Gas Chromatography February Introduction Ambient Headspace Analysis with the Agilent 7683 Automatic Liquid Sampler Application Gas Chromatography February 1998 Authors Matthew S. Klee and Chin Kai Meng Agilent Technologies, Inc. 2850 Centerville

More information

Title Experiment 7: Gas Chromatography and Mass Spectrometry: Fuel Analysis

Title Experiment 7: Gas Chromatography and Mass Spectrometry: Fuel Analysis Title Experiment 7: Gas Chromatography and Mass Spectrometry: Fuel Analysis Name Manraj Gill (Partner: Tanner Adams, Lab Section: 102) Introduction In this experiment, we use chromatography and mass spectrometry

More information

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS application Note Gas Chromatography Author Heidi Grecsek PerkinElmer, Inc. Shelton, CT 06484 USA Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS Introduction

More information

January 19, Dear Mr. Nightingale:

January 19, Dear Mr. Nightingale: January 19, 2011 Mr. Steven F. Nightingale, P.E. Manager, Permit Section Illinois Environmental Protection Agency Bureau of Land 1021 North Grand Avenue East Springfield, Illinois 62794 Subject: Supplemental

More information

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Application Note Environmental Author Suli Zhao, Andy Zhai Agilent Technologies Co., Ltd. 412 Yinglun Road

More information