Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Size: px
Start display at page:

Download "Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System"

Transcription

1 Fast Analysis of USP 7 Residual Solvents using the Agilent 789A GC and Low Thermal Mass (LTM) System Application Note Pharmaceutical Author Roger L Firor Agilent Technologies, Inc. 8 Centerville Road Wilmington, DE 988 USA Abstract A dual column residual solvent analysis according to USP 7 (8-9 revision) was performed using the Low Thermal Mass (LTM) system installed on an Agilent 789A GC system. The G888 Automated Headspace sampler connected to the volatiles interface was used for sample introduction. A Capillary Flow Technology (CFT) two way splitter was used to split the sample equally to a inch 7 M x. mm x. µm Agilent J&W DB- column and a inch 7 M x. mm x. µm Agilent J&W HP-INNOWax column. Each column module was connected to its own FID by retention gaps. Aqueous solutions of Class, Class A, and Class B solvents were analyzed. Sensitivity, linearity, and resolution met the requirements of USP 7. Overall cycle times for the analysis of all specified Class and Class A and B solvents were reduced to min.

2 Introduction Residual solvents in pharmaceuticals may remain from the manufacturing process of the active pharmaceutical ingredients (API) or final product. The level of residual solvents are monitored and controlled for a number of reasons that include safety, effect on crystalline form, solubility, bioavailability, and stability. All drug substances, excipients, and products are included under USP 7. The LTM (Low Thermal Mass) chromatographic system is combined with static headspace sampling for the analysis of residual solvents in pharmaceuticals according to USP 7 revised general chapter 8. [] This chapter follows guidelines set by the International Conference for Harmonization (ICH) QC. [] Residual solvents are divided into three classes based on possible toxicity. Class solvents are considered the most toxic and should be avoided in manufacture. These solvents may also pose an environmental risk. Class solvents (A, B, and C) are less toxic with limited use. Class C solvents have higher boiling points and some of them require analysis by non-headspace methods. Class are least toxic and should be used as solvents where practical. Headspace GC is used for determination of Class and Class solvents, while most Class solvents are analyzed by a nonspecific method such as loss on drying. Each Class solvent has a "permitted daily exposure" (PDE) limit. If a given solvent tests below the PDE limit then further testing is not required (daily dose below grams). Option of the general chapter, which looks at the total solvent added for all components of the drug product, is used for daily amounts above g. This work follows the guidelines of the method with the exception of column dimensions and GC oven programs. Column dimensions and program rates were optimized to gain a significant reduction in analysis time and overall cycle time. Alternate methodologies such as those described here can be used, however, validation and comparison to the original USP monograph may be required. The FDA also requires that any new ANDA provide the data necessary to prove control of residual solvents prior to a generic drug approval. USP 7 specifies three procedures as follows for Class and Class residual solvents:. Procedure A: Identification and limit test. Procedure B: Confirmatory test. Procedure C: Quantitative test Procedure A uses a G phase (Agilent J&W DB- column in this work) and Procedure B uses a G phase (Agilent J&W HP-INNOWax column in this work). In general a particular co-elution that occurs on one of these phases will not occur on the other. Experimental The water soluble procedures were used for standard sample preparation to demonstrate system performance. Insoluble articles require use of DMSO, DMF, or other suitable nonaqueous solvent. The methodology used is very similar for both solvent systems. A diagram of the dual column system used is shown in Figure. The setup splits the effluent from the headspace equally to the J&W DB- and J&W HP-INNOWax columns for a simultaneous dual channel analysis. Previous work has been described using conventional oven heating and two-hole ferrules with the split/splitless inlet for dual column residual solvents. [] Configuration and parameter settings for the LTM system are given in Table. Agilent G888 Headspace Sampler Volatiles Interface LTM : 7 M. mm. µm Agilent J&W DB- FID CFT Splitter Agilent 789A GC System LTM : 7 M. mm. µm Agilent J&W HP-INNOWax FID Figure. System diagram showing CFT splitter use in dual LTM column configuration.

3 Table. Residual Solvent System Parameters Standards Class : p/n 9-, equivalent to USP Mixture RS Class A: p/n 9-9, equivalent to USP Mixture A RS Class B: p/n 9-9, equivalent to USP Mixture B RS Software ChemStation: B.. Headspace: G9AA, A.. LTM: G8AA 789A Configuration and Method Parameters Inlet: Volatiles Interface, C Pressure program: psig ( min) to psig ( min) at. psi/min Split ratio: :, He carrier Detectors: Dual FID CFT: Two-way splitter, G8B 789A oven: Isothermal at C LTM Module : 7 M. mm. µm J&W DB- LTM Module : 7 M. mm. µm J&W HP-INNOWax Module connections to CFT splitter and FID's:. M. mm deactivated retention gap LTM module programs: See Table. G888A Headspace Parameters Oven: 8 C Loop: 9 C Transfer line: C Cycle time: LTM program dependent Vial Equilibration time: min Pressurize time:. min Loop fill:. min Loop equilibration: min Inject:. min Vials: ml, high shake Vial pressure:. psig for 789A Aux channel Standard solutions of the Class, Class A, and Class B solvents were prepared in pure water according to the USP 7 procedures shown in Table. These stock solutions can be stored for to months at room temperature in a well sealed volumetric. Two grams of sodium sulfate was added to each headspace vial to assist with headspace extraction. Table. Standard Preparation Class solvents.. ml stock solution vial plus 9 ml DMSO diluted to ml.. ml from step diluted to ml with water. ml from step diluted to oo ml with water.. ml step and ml water in ml HS vial Class A solvents.. ml stock solution vial, diluted to ml.. ml from step in ml water in ml HS vial Class B solvents.. ml stock solution vial, diluted to ml.. ml step in ml water in ml HS vial Headspace samples were also prepared for the Class A solvents at other concentrations ranging from about times below USP limit values to times above to demonstrate linearity. Results are shown in Figure. For example, according to USP Procedure A, the limit concentration (in prepared headspace vials) for, dioxane is.7. The concentrations used () for linearity were.9,.7,.9,.7, and 9. in water. Methanol R = Acetonitrile R =.9999 Dichloromethane R = trans-,-dichloroethene R =.9999 Figure. Calibration curves for Class A solvents from approximately times below limit values to times above. (Continued)

4 8 7 cis-,-dichloroethene R =.9999 THF R = Cyclohexane R =.999 Methylcyclohexane R = R =.9999 R =.9999,-Dioxane Chlorobenzene 7 R =.9999 R =.9998 Toluene Ethylbenzene m-xylene, p-xylene o-xylene R =.9999 R = Figure. Calibration curves for Class A solvents from approximately X below limit values to X above.

5 Discussion In temperature-programmed gas chromatography, which is required for residual solvent analysis, oven cool down time has a major impact on overall cycle time. LTM column modules cool down at considerably faster rates compared to air bath ovens due to their very low thermal mass and cooling fan configuration directly below the column assembly. LTM columns are also capable of much higher temperature programmed ramp rates, which shorten cycle time further. Maximum practical program rates will depend on a number of factors including column dimensions, phase ratio, carrier gas, and the separation required. When translating a conventional air bath method to the LTM format, Agilent Method Translation software can be used to calculate starting conditions. An example is shown in Figure for translating from a standard M column to a 7 M LTM module. LTM program rates ranging from C/min to C/min gave acceptable results in terms of meeting required resolution of specific solvent pairs. Previous work describing the use of LTM technology for a generic set of residual solvents employed a m.8 mm,. µm J&W DB- column. [] A comparison of various column dimensions and program rates are shown in Table. Air bath and LTM methods are included. The table includes entries for the same column (7 M) dimension in air bath and LTM configurations, which allows a valid comparison of overall cycle time. Note that the maximum air oven program rates possible for the 789A V and V GC systems, over the range needed for this application ( C to C), are C/min and C/min, respectively. When comparing against a V 789A GC, the LTM still achieves a % reduction in cycle time. Throughout this work, both LTM columns were controlled from the LTM ChemStation Software add-on module and operated with identical oven programs. However, the LTM columns can each have unique programs that assist with optimization. The only restraint is that both column programs start at the same time. Ending times may be different Figure. Method translation from standard M column to an Agilent J&W DB- 7 M LTM column. See to download this tool.

6 Table. Cycle Times for Various Column and Oven Type Configurations Heating Column Program Cool down Cycle time 789A (V) M. mm. µm C ( min) to C min sec with min 7 min Agilent J&W DB- ( min) at C/min oven equil. 789A () 7 M. mm. µm C ( min) to C 8 min sec with min min Agilent J&W DB- ( min) at C/min* oven equil. sec 789A () 7 M. mm. µm C ( min) to C 8 min sec with min min Agilent J&W DB- ( min) at C/min** oven equil..sec LTM (Fast) 7 M. mm. µm C ( min) to C min sec (one min Agilent J&W DB- ( min) at C/min module system) sec LTM (Faster) 7 M. mm. µm C ( min) to C min sec (one min Agilent J&W DB- ( min) at C/min module system) sec LTM (Fastest) 7 M. mm. µm C ( min) to C min sec (one min Agilent J&W DB- ( min) at C/min module system) sec LTM chromatograms of Class solvents are shown in Figures A and B on Agilent J&W DB- and J&W HP-INNOWax modules, respectively. Resolution between two Class A solvents (acetonitrile and methylene chloride on J&W DB- columns) meets method requirements as shown in Figure. Signal-to-noise ratio's for all Class solvents are greater than at specified limit concentrations. Chromatograms for Class A and B solvents on both J&W DB- and J&W HP-INNOWax phases are shown in Figures and 7. All Class, A, and B solvents combined at limit concentrations are shown in Figure 8. Peak identifications and limit concentrations in prepared headspace vials are shown in Table. Note that operating at C/min yields a cycle time of. minutes. Headspace vial equilibration times were kept at min in this work, following USP 7. However, it should be noted that equivalent results can be obtained with min heating times []. Additional benefits in sensitivity and repeatability are possible using electronic back pressure control of the headspace vial venting (loop fill) process. This is discussed at length in Application Note EN []. Norm....,-dichloroethene.,,-trichloroethane. Carbon tetrechloride. Benzene.,-dichloroethane. 7 Figure A. Class residual solvents at limit concentration on an Agilent J&W DB- column at C/min program rate.

7 Norm. 7, ,-dichloroethene.,,-trichloroethane. CCl. Benzene.,-dichloroethene min Figure B. Class residual solvents at limit concentration on an Agilent J&W HP-INNOWax column at C/min. Resolution, R =.9 USP requirements, R = or greater. Acetonitrile. Methylene chloride Figure. Acetonitrile/methlyene chloride resolution. 7

8 Norm Methanol. Acetonitrile. Methyl cyclohexane. trans,-dichloroethene/thf. cis,-dichloroethene. Tetrahydrofuran 7. Cyclohexane 8. Methylene chloride 9.,-dioxane. Toluene. Chlorobenzene. Ethylbenzene. m-xylene, p-xylene. o-xylene 9 DMSO Agilent J&W DB-, C/min, Class A Figure A. Class A solvents at limit concentration on an Agilent J&W DB- column, C/min. Norm Cyclohexane. Methyl cyclohexane. trans,-dichloroethene/thf. Methanol. Methylene chloride. cis,-dichloroethene 7. Acetonitrile 8. Toluene 9.,-dioxane. Ethylbenzene. p-xylene. m-xylene. o-xylene. Chlorobenzene DMSO Resolution, R =. for acetonitrile and cis,-dichloroethene Agilent J&W HP-INNOWAX, C/min, Class A Figure B. Class A solvents at limit concentration on an Agilent J&W HP-INNOWax column, C/min. 8

9 8 7. Hexane. Nitromethane. Chloroform.,-dimethoxyethane. Trichloroethene. Pyridine 7. -hexanone 8. Tetralin 7 DMSO Agilent J&W DB-, C/min, Class B Figure 7A. Class B solvents at limit concentration on Agilent J&W DB- column, C/min 7. Hexane.,-dimethoxyethane. Trichloroethene. Chloroform. -hexanone. Nitromethane 7. Pyridine 8. Tetralin 8 7 Agilent J&W HP-INNOWAX, C/min, Class B Figure 7B. Class B solvents at limit concentration on Agilent J&W HP-INNOWax column, C/min. 9

10 Norm DMSO 7 Class, A, B on an Agilent J&W DB- column at C/min Figure 8. Class, A, and B solvents at limit concentration on Agilent J&W DB- column, C/min. Peak IDs in Table. Table. Peak Numbering for Figure 8 and Actual Headspace Vial Concentrations Class Conc () Class A Conc () Class B Conc ().,-dichloroethene.7. Methanol.. Hexane.8 9.,,-trichloroethane 8.. Acetonitrile. 7. Nitromethane.8 9. Carbon tetrachloride.. Methylene chloride. 8. Chloroform..,-dichloroethane.7. trans-,-dichloroethene 7.8.,-dimethoxyethane.7. Benzene.7 7. cis-,-dichloroethene 7.8. Trichloroethene. 8. Tetrahydrofuran.. Pyridine. Coelutions on DB- 9. Cyclohexane. 7. -hexanone.8 cis-,-dichloroethene, nitromethane. Methylcyclohexane 9.8. Tetralin.7 THF, chloroform.,-dioxane.7 Cyclohexane, CCl,,,-trichloroethane. Toluene 7. Benzene,,-dichloroethane 8. Chlorobenzene. 9. Ethylbenzene.7. m, p-xylene.8. o-xylene.

11 Conclusions A X overall reduction in cycle time is possible when converting from the standard methodology to a LTM based system for residual solvent analysis. Capillary flow technology can be employed to conveniently analyze on two column phases (Agilent J&W DB- and Agilent J&W HP-INNOWax columns) simultaneously from a single headspace injection. LTM column dimensions of 7M x. mm provide a good compromise among speed, ease-of-use, and capacity while meeting the resolution requirements of USP 7. This general methodology using LTM technology should be particularly attractive to new drug development where variations to the USP procedures are appropriate and fast method optimization is desired. For More Information For more information on our products and services, visit our Web site at The Chemstation method integrates Agilent 789A GC/Agilent G888A Headspace, and LTM control through add-on software modules for ease of setup, operation, method integration, and compliance. References. USP -NF 7, General Chapter USP <7> Organic Volatile Impurities, United States Pharmacopeia. Pharmacopoeia Convention Inc., Rockville, MD, 8/9.. International Conference on Harmonization (ICH) of Technical Requirements for the Registration of Pharmaceuticals for Human Use, QC (R): Impurities guideline for residual solvents, Step, July Joseph M. Levy and Michael Kraft,"Simultaneous Dual Capillary Column Headspace GC With Flame Ionization Confirmation and Quantification According to <USP7>, Agilent Technologies publication EN, 8.. Frank David, Roman Szucs, Jay Makwans, and Pat Sandra, "Fast Capillary GC using a Low Thermal Mass Column Oven for the Determination of Residual Solvents in Pharmaceuticals," Pfizer Analytical Research Centre, Ghent University, Krijgslann, Ghent, Belgium, J. Sep. Sce., 9, 9-98,.. Roger L. Firor,"The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G888 Network Headspace Sampler," Agilent Technologies publication 989-EN,.. Albert E. Gudat and Roger L. Firor, "Improved Retention Time, Repeatability, and Sensitivity for Analysis of Residual Solvents," Agilent Technologies publication EN, 7.

12 Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., Printed in the USA January, 99-9EN

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System Application Note Residual Solvent Analysis of USP Method Residual Solvents on the Agilent 889 GC System Author Lukas Wieder, Jie Pan, and Rebecca Veeneman Agilent Technologies, Inc. 8 Centerville Road

More information

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC Applying the Agilent 5977A MSD to the Analysis of USP Residual Solvents with the 7697A Headspace Sampler and 7890B GC Application Note Pharmaceuticals Authors Roger L. Firor and Mike Szelewski Agilent

More information

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP Application Note Joseph M. Levy Michael Kraft Abstract Agilent Equipment 7890A

More information

USP<467> residual solvents

USP<467> residual solvents Application Note Pharmaceuticals USP residual solvents Applying the Agilent 977A MSD with the Agilent 797A headspace sampler and Agilent 7890B GC Authors Roger L. Firor and Mike Szelewski Agilent Technologies,

More information

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application Pharmaceuticals Author Roger L. Firor Agilent Technologies, Inc. 2850 Centerville

More information

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Residual solvents analysis using an Agilent Intuvo 9000 GC system Residual solvents analysis using an Intuvo GC system Technology advantage: simplified dual column analysis with Flow Chip modularity Introduction USP defines a method for analysis of residual solvents

More information

Study of Residual Solvents in Various Matrices by Static Headspace

Study of Residual Solvents in Various Matrices by Static Headspace Application Note Abstract United States Pharmacopeia (USP) chapter is a widely used method for identifying and quantifying Organic Volatile Impurities (OVI) used in the production of pharmaceuticals.

More information

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing

More information

The determination of residual solvents in pharmaceuticals using the Agilent G1888 headspace/6890n GC/5975 inert MSD system Application Note

The determination of residual solvents in pharmaceuticals using the Agilent G1888 headspace/6890n GC/5975 inert MSD system Application Note The determination of residual solvents in pharmaceuticals using the Agilent G1888 headspace/6890n GC/5975 inert MSD system Application Note Roger L. Firor Albert E. Gudat Abstract The Agilent G1888 network

More information

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument Application Note Abstract The US Pharmacopeia recently released USP as the current monograph for determining residual solvents in pharmaceutical products by static headspace. The USP classified these

More information

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key

More information

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS Application Note Pharmaceuticals Authors Karine Jacq, Frank David, and Pat Sandra Research Institute

More information

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology APPLICATION NOTE Gas Chromatography Author: David Scott PerkinElmer, Inc. Shelton, CT Residual Solvents in Pharmaceuticals by USP Chapter Methodology Introduction The synthesis of active pharmaceutical

More information

Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method

Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method APPLICATION NOTE 0676 Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP method Authors Giulia Riccardino, Paolo Magni, Stefano

More information

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Application Note Environmental Authors Peter Gautschi and Harry Prest Senior Application Scientist

More information

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application Static Headspace Blood Alcohol Analysis with the G Network Headspace Sampler Application Forensics Author Roger L. Firor and Chin-Kai Meng Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-0

More information

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace_Rev3 Introducing

More information

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing Automated Preparation of Simulated Distillation Standards and Samples for ASTM Methods D2887, D7213, D7398 and D6352 using the 7693A System with Easy SamplePrep Software Author Roger L. Firor Agilent Technologies,

More information

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Application Note Forensic Toxicology and Drug Testing Author Jared Bushey Agilent Technologies, Inc. 285 Centerville

More information

Application Note. Abstract. Authors. Introduction

Application Note. Abstract. Authors. Introduction Multiple Headspace Extraction for the Quantitative Determination of Residual Monomer and Solvents in Polystyrene Pellets Using the Agilent 7697A Headspace Sampler Application Note Authors Wenwen Shen and

More information

Headspace Technology for GC and GC/MS: Features, benefits & applications

Headspace Technology for GC and GC/MS: Features, benefits & applications Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over

More information

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products. Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products. CHRISTOPHER M. ENGLISH, CHRISTOPHER S. COX, FRANK

More information

Application. Gas Chromatography February Introduction

Application. Gas Chromatography February Introduction Ambient Headspace Analysis with the Agilent 7683 Automatic Liquid Sampler Application Gas Chromatography February 1998 Authors Matthew S. Klee and Chin Kai Meng Agilent Technologies, Inc. 2850 Centerville

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System Application Note Author Roger L. Firor Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 Abstract

More information

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography Fast Analysis of Aromatic Solvent with.8 mm ID GC column Application Gas Chromatography Authors Yun Zou Agilent Technologies (Shanghai) Co. Ltd. Ying Lun Road Waigaoqiao Free Trade Zone Shanghai 3 P.R.

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)

More information

Author. Abstract. Introduction

Author. Abstract. Introduction Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert Gas Chromatography/Mass Spectrometry System Application Author James D. McCurry Agilent

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

A Capillary Gas Chromatographic Procedure for the Analysis of Nine Common Residual Solvents in Water-Insoluble Bulk Pharmaceuticals

A Capillary Gas Chromatographic Procedure for the Analysis of Nine Common Residual Solvents in Water-Insoluble Bulk Pharmaceuticals A Capillary Gas Chromatographic Procedure for the Analysis of Nine Common Residual Solvents in Water-Insoluble Bulk Pharmaceuticals Q. Chan Li and Kenneth A. Cohen* Analytical Sciences Department, Boehringer

More information

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ Application Note Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ -3 Application Note Abstract This application note demonstrates the

More information

Solid Phase Micro Extraction of Flavor Compounds in Beer

Solid Phase Micro Extraction of Flavor Compounds in Beer Solid Phase Micro Extraction of Flavor s in Beer ANNE JUREK USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler Application Note Environmental

More information

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS Application Note Multi-detector Splitter Authors Charlie Spanjers and Andrew Jones Activated Research Company 7561

More information

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Sensitive Detection of 2-MIB and Geosmin in Drinking Water Sensitive Detection of -MIB and Geosmin in Drinking Water Application Note Environmental Author Yean-Woong You Agilent Technologies, Inc. Seoul, Korea Abstract An automated SPME extraction method for easy

More information

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT Acta Poloniae Pharmaceutica ñ Drug Research, Vol. 66 No. 6 pp. 611ñ615, 2009 ISSN 0001-6837 Polish Pharmaceutical Society DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE

More information

Modernizing the USP Monograph for Acetaminophen

Modernizing the USP Monograph for Acetaminophen Modernizing the USP Monograph for cetaophen pplication Note Small Molecule Pharmaceuticals and Generics uthor Rongjie Fu gilent Technologies (Shanghai) Co. Ltd bstract new HPLC method was developed for

More information

GC METHOD FOR QUANTITATIVE DETERMINATION OF RESIDUAL 2-(2-CHLOROETHOXY)ETHANOL (CEE) AND N-METHYL-2-PYRROLIDINONE (NMP) IN QUETIAPINE

GC METHOD FOR QUANTITATIVE DETERMINATION OF RESIDUAL 2-(2-CHLOROETHOXY)ETHANOL (CEE) AND N-METHYL-2-PYRROLIDINONE (NMP) IN QUETIAPINE Acta Poloniae Pharmaceutica ñ Drug Research, Vol. 64 No. 2 pp. 187ñ189, 2007 ISSN 0001-6837 Polish Pharmaceutical Society GC METHOD FOR QUANTITATIVE DETERMINATION OF RESIDUAL 2-(2-CHLOROETHOXY)ETHANOL

More information

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns Application Food, Flavors and Fragrances Authors Mark Sinnott and Simon Jones Agilent Technologies, Inc. 9 Blue Raving Road

More information

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes Application Note Environmental Authors Ken Lynam and Doris Smith Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Validation of USEPA Method 524.2 Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Application Note By: Nathan Valentine Abstract The US EPA developed Method 524.2¹, Measurement

More information

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-exyl Columns to Separate Estrogens Application Note Pharmaceutical, Environmental Authors John W. enderson Jr. and William J. Long Agilent

More information

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application Pharmaceuticals Author Cliff Woodward and Ronald Majors Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808

More information

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column Analyze Hydrocarbon Impurities in,3-butadiene with an Agilent J&W GS-Alua PT Column Application Note Energy and Chemicals Authors Yun Zou and Chunxiao Wang Agilent Technologies Shanghai Ltd Abstract The

More information

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers Electronic Supplementary Material (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 2017 Electronic Supplementary Material Experimentally Validated Mathematical

More information

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract. Application Note 8 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-5 Application Note Environmental, Canister, TO-5, Air Monitoring, Air Toxics Abstract This application

More information

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Automated Sample Preparation in Quality Control of Eye-Drop Formulation Automated Sample Preparation in Quality Control of Eye-Drop Formulation Application Note Authors Bart Tienpont, Maria Rambla Alegre, Frank David, and Pat Sandra Research Institute for Chromatography Kennedypark

More information

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Application Note Energy and Fuels - Biofuels Authors Ron Honnold, Ph.D. and Robert Kubas Agilent Technologies, Inc. Santa Clara, CA

More information

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection Trade Science Inc. September 2009 Volume 8 Issue 3 ACAIJ, 8(3) 2009 [346-349] Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization

More information

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler Validation of USEPA Method 524.2 Using a Stratum PTC and the New AQUATek 100 Autosampler Application Note Abstract Automation is the key to increasing laboratory productivity and minimizing costs. It is

More information

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors Trace-Level Analysis of Melamine in Milk Products on Agilent 789A/5975C GC/MSD Using a ew Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges Application ote Gas Chromatography/Mass Spectrometry/Food

More information

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE Authors: DANI Instruments SpA viale Brianza, 87 Cologno Monzese Milano Italy Key

More information

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Application Note Pesticides Authors Charlie Spanjers and Paul Dauenhauer University of Minnesota, Twin Cities

More information

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates Abstract Foods, Flavors & Fragrances Applications A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates By Corby Hilliard; Amanda Rigdon; William Schroeder*,

More information

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System Analysis of Terpenes in Cannabis Using the Agilent 7697A/789B/977B Headspace GC-MSD System Faster Analysis Time = Greater Productivity Application Note Cannabis, Food Authors Ronald Honnold, Robert Kubas,

More information

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD) Introduction There has been an increase in concern for the detection and removal of organic chloride species from crude aromatic, naphtha, and other hydrocarbon streams. One process called catalytic reforming

More information

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL Challenges of Water Monitoring Volatile Organic Compounds (VOCs) can have negative health impacts even at ppb levels VOC concentrations can

More information

Application. Gas Chromatography March 1998

Application. Gas Chromatography March 1998 Micro Liquid-Liquid Extraction with the Agilent 7683 Automatic Liquid Sampler Application Gas Chromatography March 1998 Authors Matthew S. Klee Agilent Technologies, Inc. 285 Centerville Road Wilmington,

More information

C146-E209. Headspace Samplers. HS-20 Series

C146-E209. Headspace Samplers. HS-20 Series C146-E209 Headspace Samplers HS-20 Series HS-20 Series of Headspace Samplers A Revolutionary System Aimed at Performance and Ease of Use The HS-20 Series is the optimal solution for volatile component

More information

Maximizing Sample Throughput In Purge And Trap Analysis

Maximizing Sample Throughput In Purge And Trap Analysis Maximizing Sample Throughput In Purge And Trap Analysis LINDSEY PYRON ANNE JUREK INTRODUCTION There are several demands and requirements imposed on chemists performing volatile organic analysis (VOC) in

More information

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software High Efficiency GC Columns Page 1 Variables for Shortening Run Times Stationary Phase Shorten Column Length

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform ( 염화비닐, 아크릴로니트릴, 브로모포름 ) with Static and Dynamic Headspace GC/MS Roger Bardsley, Applications Chemist;

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 12 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Optimizing. Abstract: is standardd. procedures. altered to

Optimizing. Abstract: is standardd. procedures. altered to Optimizing Standard Preparation ANNE JUREK Abstract: Standardd preparation can often be a time consuming, tedious process. The opportunity for human error or inconsistencies between individual preparation

More information

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application Food Safety, Environmental Authors Chin-Kai Meng and Bruce Quimby Agilent Technologies, Inc. 2850 Centerville

More information

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions Validation of Volatile Organic Compound by USEPA Method 8260C Application Note Abstract In order to determine the concentration of volatile organic compounds (VOCs) in water and soil matrices the USEPA

More information

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Lee Marotta Tom Kwoka David Scott Miles Snow Toby Astill PerkinElmer, Inc. Shelton, CT Fast, Quantitative Analysis of Residual Solvents in

More information

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application Pharmaceuticals Authors John W. Henderson Jr., Nona Martone, and Cliff Woodward

More information

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Application Note Abstract Due to regional shortages and increasing costs of helium, the preferred carrier gas in gas chromatography, alternative

More information

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates Abstract Due to rapid growth in the medical cannabis industry, demand is increasing for analysis of residual solvents in cannabis concentrates in order to protect consumer safety. This application note

More information

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas A Comparison of Volatile Organic Compound When Using Nitrogen as a Gas Application Note By: Anne Jurek Abstract For many years Helium has been the gas of choice for purging Volatile Organic Compounds (VOCs).

More information

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX Technical Note Agilent s SampliQ WAX provides Applications for strongly acidic, acidic and neutral compounds Excellent

More information

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler Application Note By: Anne Jurek Abstract In order to determine the amount

More information

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Application Note Abstract Water quality laboratories across the

More information

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Application Note Environmental Author Suli Zhao, Andy Zhai Agilent Technologies Co., Ltd. 412 Yinglun Road

More information

Methanol Extraction of high level soil samples by USEPA Method 8260C

Methanol Extraction of high level soil samples by USEPA Method 8260C Methanol Extraction of high level soil samples by USEPA Method 8260C Abstract In order to determine the concentration of Volatile Organic Compounds (VOCs) in soil and waste samples the USEPA developed

More information

Introduction to Capillary GC

Introduction to Capillary GC Introduction to Capillary GC LC Columns and Consumables Simon Jones Chromatography Applications Engineer February 20, 2008 Page 1 Introduction to Capillary GC t r K c?? Kβ k = - tr t m? t m R s Page 2

More information

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC / 5977A MS Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Abstract US EPA Method

More information

Determination of Volatile Organic Compounds in Air

Determination of Volatile Organic Compounds in Air JSB is an authorised partner of Determination of Volatile Organic s in Air Anne Jurek #126 Introduction Air pollution is a growing problem due to the global economy and industrial development in many countries.

More information

DETERMINATION OF ORGANIC VOLATILE IMPURITIES IN NEPAFENAC BY GC METHOD

DETERMINATION OF ORGANIC VOLATILE IMPURITIES IN NEPAFENAC BY GC METHOD Acta Poloniae Pharmaceutica ñ Drug Research, Vol. 71 No. 6 pp. 1013ñ1022, 2014 ISSN 0001-6837 Polish Pharmaceutical Society DETERMINATION OF ORGANIC VOLATILE IMPURITIES IN NEPAFENAC BY GC METHOD MARIOLA

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India Effective use of pharmacopeia guidelines to reduce cost of chromatographic analysis Optimized, cost-effective HPLC analysis of atorvastatin by varying column dimensions within the USP allowed limts

More information

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract. Separation of Explosives in EPA 833: Column Choices Optimize Speed, Resolution, and Solvent Use Application Environmental Authors John W. Henderson Jr. and William J. Long Agilent Technologies, Inc. 28

More information

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF Analysis of Biomarkers in Crude Oil Using the Agilent 7 GC/Q-TOF Application Note Petrochemical and Environmental Authors Frank David Research Institute for Chromatography, Kennedypark 6, B-85 Kortrijk,

More information

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7B Triple Quadrupole GC/MS Application Note Authors Dale R. Walker and Fred Feyerherm Agilent Technologies, Inc.

More information

GAS CHROMATOGRAPHY (GC)

GAS CHROMATOGRAPHY (GC) GAS CHROMATOGRAPHY (GC) Pre-Lab Questions Questions are to be answered before the beginning of the laboratory. The answers are due at the beginning of each experiment (the questions are for credit and

More information

Thermal Desorption Technical Support

Thermal Desorption Technical Support Thermal Desorption Technical Support Note 86a: US EPA Method TO-17 for Monitoring Air Toxics in Ambient Air Using Sorbent Tubes and Automated, Cryogen-free Thermal Desorption Application Note Environmental,

More information

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System Margaret Maziarz, Sean M. McCarthy, Michael D. Jones, and Warren B. Potts Waters Corporation, Milford, MA, USA A P

More information

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS LECO Corporation; Saint Joseph, Michigan USA Key Words: GC-TOFMS,

More information

Analysis of BTEX in Natural Water with SPME

Analysis of BTEX in Natural Water with SPME Analysis of BTEX in Natural Water with SPME Application Note Environmental Author Giordano Vassalli Sezione Protezione Aria Introduction Benzene, toluene, ethylbenzene and xylene (BTEX) isomers are monocyclic

More information

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS Application Note By Tyler Trent Abstract In order to determine the concentration of volatile organic compounds

More information

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer Application Note By: Anne Jurek Abstract This application note presents

More information

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Application Note Author Harry Prest, Ph.D. Senior Chemist in GC-MS Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara,

More information

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Environmental Analysis Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville

More information

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology Gas Chromatography/Mass Spectrometry Author Harry Prest 1601 California Avenue

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column Competitive Comparison Authors Ngoc-A Dang and Allen K. Vickers Agilent

More information

Strategies to minimize the impact of presence of residual solvents in APIs

Strategies to minimize the impact of presence of residual solvents in APIs CBC-PROCOS S.p.A. Strategies to minimize the impact of presence of residual solvents in APIs DamianaGentili Qualified Person, Quality Unit Director CBC ProcosS.p.A. Pavia, 02 nd October 2015 slide# 1 Agenda

More information

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek JSB is an authorised partner of Analysis of Low Level Volatile Organic s in Air Anne Jurek #111 Introduction The New Jersey Department of Environmental Protection (NJDEP) developed a new low level TO-15

More information