SBC FLATS: PRISM P-FLATS and IMAGING L-FLATS

Size: px
Start display at page:

Download "SBC FLATS: PRISM P-FLATS and IMAGING L-FLATS"

Transcription

1 Instrument Science Report ACS SBC FLATS: PRISM P-FLATS and IMAGING L-FLATS R. C. Bohlin & J. Mack December 2006 ABSTRACT The internal deuterium lamp was used to illuminate the SBC detector through the PR110L and PR130L prisms for 12.2 hours each to produce a total of ~12,000 counts/pixel. This illumination does not simulate the OTA optics and, therefore, is not suitable for the production of a low frequency L-flat. However, the pixel-to-pixel P-flat is an improvement over the laboratory SBC P-flat currently used in the ACS pipeline for the two dispersing modes. In addition, short exposure internal lamp flats were obtained in the standard imaging filters. These flats have sufficient signal to define the low frequency L-flat field for five filters relative to the high signal F125LP flat, assuming that the relative lamp illumination does not vary with wavelength. These five ratio L-flats are smoother than the ratios of the current pipeline L-flats; but there is evidence for variation of the internal lamp illumination with wavelength. Thus, the current SBC L-flats may have some errors of a few percent due to local inappropriate lumpiness; but the alternative flats defined by the internal illumination may also have errors. 1. Introduction Because uniform flat field illumination in the Å region is available neither in the lab nor in space, the low-frequency L-flat field correction for the SBC filters on ACS is problematic. However, pipeline processing does not require an L-flat correction for prism data. Instead, Larsen (2006) derives sensitivities that depend on location on the detector. The internal deuterium lamp does produce a smoothly varying illumination 1

2 pattern that can be used to define a P-flat. Section 2 of this ISR addresses the creation of a new P-flat for the SBC prism modes, while section 3 discusses the relative L-flat correction for five imaging filters. 2. Prism P-flat The prism P-flat differs from the imaging filters, because the illumination has a different angle of incidence Bohlin, Hartig, and Meurer (1999, BHM). Because the P-flat for the prisms has probably changed by an amount that is comparable to the change found by Bohlin & Mack (2005) from ground to space for the imaging filters, an additional set of internal lamp exposures are analyzed to define a P-flat appropriate for on-orbit use for the prism modes. Figure 1 - Relative count rate vs. total time of illumination of the SBC detector by the internal deuterium flat field lamp. The 32 points after ~21 hours are 16 PR130L and 16 PR110L points, which are normalized to the earlier F125LP data. 2

3 The internal lamp exposure times are all 2750s for the 16 PR110L and 16 PR130L images that were obtained in program over the time period of 2005Oct4 to 2005Nov5. The total exposure time is 24.4 hours, which resulted in ~12,000 total counts/pixel near the center of the SBC MAMA detector. Figure 1 shows the degradation of the lamp output vs. exposure time. Based on pre-launch testing, the declining lamp output is probably caused by polymerization of contaminants on the lamp window. In order to verify the claim by BHM that the SBC flats are independent of wavelength, the 16 individual exposures for each prism are co-added and fit following the procedures summarized in Bohlin et al. (1999) The IDL routine flatall along with its main subroutines flats and fitflat are used to produce the P-flat by fitting the overall vignetting pattern with low order splines. This internal vignetting of the lamp illumination is about 10% from image center to the edges, while the P-flat is equal to the original co-added image divided by the fitted pattern. The data quality flags in the third extension of the flat field file are all zero, except for values of 512 around the edges, in the un-illuminated region beyond column 850, in rows 599:604 with dead MAMA diodes, and for 295 pixels listed in the bad pixel table lch1502jj_bpx.fits. The new PR110L and new PR130L P-flats are typically identical to % of excess residual scatter over the differences expected from the Poisson counting statistics. With an expected Poisson uncertainty per pixel of 0.9% in the combined 32 prism exposures, the intrinsic differences between PR110L and PR130L are small enough so that the best P-flat is made from the co-addition of all 32 internal prism exposures. The ratio of this new prism flat to the original lab P-flat is illustrated in Figure 2. In the dashed box of Figure 2, the expected Poisson scatter of 1.13% is comprised of 0.7% from the original lab flat and 0.9% from the new in-flight P-flat. The error in using the original flat is 1.08% from the combined (in quadrature) of its 0.70% Poisson statistics and the 0.82% systematic error. Assuming 0.3% systematic error caused by averaging the PR110L and PR130L images and combining with the 0.9% statistical uncertainty, the new P-flat uncertainty of 0.95% per pixel is slightly better than the 1.08% error expected in the old P-flat. The intrinsic differences at the lower right in Figure 2 increase from 0.82% to over 1% due to a slight shift in the Moire interference pattern between the micro-channel plate and the anode array of the SBC MAMA detector. Thus, the advantage of the new prism P-flat is greater in the lower right corner of the SBC. 3

4 4

5 Figure 2 - Ratio of the on-orbit SBC P-flat to the laboratory P-flat for the prisms. The name of the lab flat, psbcex99marsm20prism, is the sum 20 images with external illumination obtained in March of 1999 and is also known as jref$m5t12222j_pfl in the ACS pipeline processing system. The grey scale calibration is indicated on the reference bar at the top right. The line of text "P-flat rms(%)= " indicates one-sigma values in the 101x101 pixel standard region at 452:552,462:562 (dashed box) for the total pixel-to-pixel scatter, the Poisson counting statistic, and the intrinsic rms variation of the ratio. The horizontal grey strip is the region of bad anodes, where the flats are set to unity. The vertical feature near column 578 is caused by the slightly different illumination in the region of the repeller wire. The uniform grey region beyond column 850 at right hand side is outside the field-of-view of the prisms. 3. Imaging L-flats Even though the P-flat is independent of wavelength, the L-flat is not. Mack, et al. (2005, MGVB) used observations of the NGC6681 star cluster to determine SBC L-flats on a lightly smoothed 16x16 grid. The S/N of the hot blue HB stars in NGC6681 is the highest at the shorter wavelengths. The formal uncertainties in the L-flat are the lowest for F125LP, which also has the most observations. An alternative method for defining the remaining five SBC imaging L-flats uses short exposure internal deuterium lamp flats. These flats have sufficient signal to define the low frequency L-flat field for five filters relative to the high signal stellar F125LP L-flat, assuming that the relative lamp illumination does not vary with wavelength. These five ratios of internal lamp images are smoother than the corresponding ratios of the current MGVB pipeline L-flats. For example in the worst case with the lowest S/N for F165LP, the ratio F165LP/F125LP is shown in Figure 3. If the deuterium lamp illumination of the detector-filter combination is the same for all six filters, the internal flat field lamp can be used to find differences in the L-flat between filters. Single exposures of 2500s obtained in program in each of F115LP, F122M, F140LP, F150LP, and F165LP are sufficient to define the delta L-flats relative to the F125LP L-flat from MGVB. The ratio of the MGVB stellar L-flats in panel c) of Figure 3 is lumpier than the corresponding ratio of the L-flat structure in the internal lamp images illustrated in panel d) by up to 7% in a hot spot (white) in the lower left quadrant of panel c). MGVB estimate a systematic uncertainty of ~4% for F165LP in their table 3. Alternative L-flats, L', for the five filters in Table 1 are formed by multiplying the F125LP stellar L-flat by the appropriate internal lamp ratio, eg. multiplying panel b) by the F165LP/F125LP ratio of panel d) from the internal lamp. 5

6 Figure 3 - Top panels: L-flats from MGVB a) F165LP and b) F125LP scaled to unity in the central 101x101 pixel box. Lower panels: c) Ratio of panel a) to panel b) for the stellar L-flats from MGVB. d) Corresponding ratio of L-flat response to the internal deuterium lamp for F165LP/F125LP. 6

7 Table 1 summarizes the large-scale and small-scale differences between the five L-flats derived from the deuterium lamp data and the MGVB pipeline flats. The two "64 Bin" columns are the large-scale minimum and maximum values of the MGVB/internal L/L' ratio averaged over 8x8=64 boxes of 128x128 pixels per box, while the last two columns are the local, small-scale minimum and maximum of the L-flat ratios within the central 512x512 pixels of the 1024x1024 pixel SBC image format. The total range of 64-bin differences exceeds 10% only for F165LP. Both the min and max of the range of these large-scale differences are usually at an edge of image, where any change in the internal lamp illumination pattern with wavelength would contribute the most. The min and max of the local, smaller-scale lumpiness in the central 512x512 pixel region minimize any effect of changing illumination with wavelength and demonstrate agreement between L and L' to 4%, except for F165LP. TABLE 1. Ratio of Stellar to Internal-lamp L-flats Large scale Smaller scale Filter 64 Bin Central 512 px min max min max F115LP F122M F140LP F150LP F165LP Complete reduction, including time dependent corrections, is done for all the NGC6681 F165LP data; and stellar photometry is derived using the MGVB pipeline L-flats. This standard photometry, L, can be corrected to L' photometry for the alternative internal lamp L'-flat. If the lamp L'-flat is superior to the stellar L-flat, then the rms scatter in the dithered observations of NGC6681 should be lower in the L' photometry. The biggest differences between L and L' photometry should be for stars lying nearest to the hotspot, where L/L' reaches a maximum of 1.07 in panel c) of Figure 3. There are seven stars where at least one observation lies in box centered on the hotspot at (360,320) with x-size of 64 pixels and y-size of 74 pixels. Each of these seven stars has 33 observations, so that the seven rms measures are robust. Five stars show an improved rms scatter with the L' lamp flat, where the biggest improvement is from to mag. However, the rms scatter for two of the seven stars increases; and most of the outliers for the L' lamp photometry tend to lie near the top of the image. Thus, the L'-flat may be good locally; but there may be wavelength dependent errors that slowly vary across the full field. 7

8 To test for local improvements, the only photometry considered is still for the same seven stars; but every point is restricted to lie within a box of 200x200 pixels, centered on the hotspot. For this restricted case, the number of repeat observations is 7-8 for three stars and for four stars; and the rms scatter is always better for the L' photometry by as much as a factor of two. In summary, the relative L'-flats from the deuterium internal lamp suggest that the flats derived from the stellar observations by MGVB are somewhat too lumpy; and the final L-flats should be restricted to have stiffer fits to the stellar data. Unfortunately, the relative L'-flats from the internal lamp data may have significant low order errors across the full field that could be attributed to a gradient in the relative lamp illumination that varies with wavelength. Ongoing SBC observations of NGC6681 will be combined with existing observations; and revised L-flats will be made that should have ratios similar to the internal lamp ratios, eg. Figure 3d. REFERENCES Bohlin, R., Hartig, G., Lindler, D., Meurer, G., & Cox, C. 1999, Instrument Science Report, ACS 99-01, (Baltimore:STScI) Bohlin, R., Hartig, G., & Meurer, G. 1999, Instrument Science Report, ACS 99-02, (Baltimore:STScI), BHM Bohlin, R. C., & Mack, J. 2005, Instrument Science Report, ACS , (Baltimore:STScI) Larsen, S. 2006, Instrument Science Report, ACS 06-02, (Baltimore:STScI) Mack, J., Gilliland, R., van der Marel, R., & Bohlin, R. 2005, Instrument Science Report, ACS , (Baltimore:STScI), MGVB 8

Earth Flats. 1. Introduction. Instrument Science Report ACS R. C. Bohlin, J. Mack, G. Hartig, & M. Sirianni October 25, 2005

Earth Flats. 1. Introduction. Instrument Science Report ACS R. C. Bohlin, J. Mack, G. Hartig, & M. Sirianni October 25, 2005 Instrument Science Report ACS 2005-12 Earth Flats R. C. Bohlin, J. Mack, G. Hartig, & M. Sirianni October 25, 2005 ABSTRACT Since the last ISR 2003-02 on the use of Earth observations for a source of flat

More information

SBC L-Flat Corrections and Time-Dependent Sensitivity

SBC L-Flat Corrections and Time-Dependent Sensitivity SBC L-Flat Corrections and Time-Dependent Sensitivity J. Mack, R. Gilliland, R. van der Marel, and R. Bohlin November 17, 2005 ABSTRACT The uniformity of the SBC detector response has been assessed using

More information

Absolute Flux Calibration for STIS First-Order, Low-Resolution Modes

Absolute Flux Calibration for STIS First-Order, Low-Resolution Modes Instrument Science Report STIS 97-14 Absolute Flux Calibration for STIS First-Order, Low-Resolution Modes Ralph Bohlin, Space Telescope Science Institute Nicholas Collins, Hughes STX/LASP/GSFC Anne Gonnella,

More information

ACS after SM4: RELATIVE GAIN VALUES AMONG THE FOUR WFC AMPLIFIERS

ACS after SM4: RELATIVE GAIN VALUES AMONG THE FOUR WFC AMPLIFIERS Instrument Science Report ACS 2009-03 ACS after SM4: RELATIVE GAIN VALUES AMONG THE FOUR WFC AMPLIFIERS R. C. Bohlin, A. Maybhate, & J. Mack 2009 October 8 ABSTRACT For the default setting of gain=2, the

More information

FLAT FIELDS FROM THE MOONLIT EARTH

FLAT FIELDS FROM THE MOONLIT EARTH Instrument Science Report WFPC2 2008-01 FLAT FIELDS FROM THE MOONLIT EARTH R. C. Bohlin, J. Mack, and J. Biretta 2008 February 4 ABSTRACT The Earth illuminated by light from the full Moon was observed

More information

Calibration of ACS Prism Slitless Spectroscopy Modes

Calibration of ACS Prism Slitless Spectroscopy Modes The 2005 HST Calibration Workshop Space Telescope Science Institute, 2005 A. M. Koekemoer, P. Goudfrooij, and L. L. Dressel, eds. Calibration of ACS Prism Slitless Spectroscopy Modes S. S. Larsen, M. Kümmel

More information

An Algorithm for Correcting CTE Loss in Spectrophotometry of Point Sources with the STIS CCD

An Algorithm for Correcting CTE Loss in Spectrophotometry of Point Sources with the STIS CCD An Algorithm for Correcting CTE Loss in Spectrophotometry of Point Sources with the STIS CCD Ralph Bohlin and Paul Goudfrooij August 8, 2003 ABSTRACT The correction for the change in sensitivity with time

More information

First On-orbit Measurements of the WFC3-IR Count-rate Non-Linearity

First On-orbit Measurements of the WFC3-IR Count-rate Non-Linearity First On-orbit Measurements of the WFC3-IR Count-rate Non-Linearity A. G. Riess March 10, 2010 ABSTRACT Previous HgCdTe detectors on HST have suffered from a count-rate dependent non-linearity, motivating

More information

The in-orbit wavelength calibration of the WFC G800L grism

The in-orbit wavelength calibration of the WFC G800L grism The in-orbit wavelength calibration of the WFC G800L grism A. Pasquali, N. Pirzkal, J.R. Walsh March 5, 2003 ABSTRACT We present the G800L grism spectra of the Wolf-Rayet stars WR45 and WR96 acquired with

More information

Improved Photometry for G750L

Improved Photometry for G750L SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report STIS 2015-01(v1) Improved Photometry for G750L R.C. Bohlin 1, C.R. Proffitt 1 1 Space Telescope Science Institute,

More information

Study of the evolution of the ACS/WFC sensitivity loss

Study of the evolution of the ACS/WFC sensitivity loss Instrument Science Report ACS 2013 01 Study of the evolution of the ACS/WFC sensitivity loss Leonardo Úbeda and Jay Anderson Space Telescope Science Institute January 28, 2013 ABSTRACT We present a study

More information

Delivery of a new ACS SBC throughput curve for Synphot

Delivery of a new ACS SBC throughput curve for Synphot Delivery of a new ACS SBC throughput curve for Synphot Francesca R. Boffi, M. Sirianni, R. A. Lucas, N. R. Walborn, C. R. Proffitt April 18, 2008 ABSTRACT On November 12th, 2007 a new ACS SBC throughput

More information

SMOV Absolute Flux Calibration of the COS FUV Modes

SMOV Absolute Flux Calibration of the COS FUV Modes Instrument Science Report COS 2010-02(v1) SMOV Absolute Flux Calibration of the COS FUV Modes Derck Massa 1, Charles Keyes 1, Steve Penton 2, Ralph Bohlin 1, and Cynthia Froning 2 1 Space Telescope Science

More information

Updated flux calibration and fringe modelling for the ACS/WFC G800L grism

Updated flux calibration and fringe modelling for the ACS/WFC G800L grism Updated flux calibration and fringe modelling for the ACS/WFC G800L grism H. Kuntschner, M. Kümmel, J. R. Walsh January 25, 2008 ABSTRACT A revised flux calibration is presented for the G800L grism with

More information

Persistence in the WFC3 IR Detector: Spatial Variations

Persistence in the WFC3 IR Detector: Spatial Variations SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA WFC3 Instrument Science Report 2015-16 Persistence in the WFC3 IR Detector: Spatial Variations Knox S. Long, Sylvia M. Baggett, & John W. MacKenty

More information

ACS CCDs UV and narrowband filters red leak check

ACS CCDs UV and narrowband filters red leak check Instrument Science Report ACS 2007-03 ACS CCDs UV and narrowband filters red leak check Marco Chiaberge and Marco Sirianni May 01, 2007 ABSTRACT We present results of the observations of the star 15 Mon,

More information

Fringe Correction for STIS Near-IR Long-Slit Spectra using Contemporaneous Tungsten Flat Fields

Fringe Correction for STIS Near-IR Long-Slit Spectra using Contemporaneous Tungsten Flat Fields 1997 HST Calibration Workshop Space Telescope Science Institute, 1997 S. Casertano, et al., eds. Fringe Correction for STIS Near-IR Long-Slit Spectra using Contemporaneous Tungsten Flat Fields Paul Goudfrooij

More information

WFC3 IR Blobs, IR Sky Flats and the measured IR background levels

WFC3 IR Blobs, IR Sky Flats and the measured IR background levels The 2010 STScI Calibration Workshop Space Telescope Science Institute, 2010 Susana Deustua and Cristina Oliveira, eds. WFC3 IR Blobs, IR Sky Flats and the measured IR background levels N. Pirzkal 1 Space

More information

Lab 4 Radial Velocity Determination of Membership in Open Clusters

Lab 4 Radial Velocity Determination of Membership in Open Clusters Lab 4 Radial Velocity Determination of Membership in Open Clusters Sean Lockwood 1, Dipesh Bhattarai 2, Neil Lender 3 December 2, 2007 Abstract We used the Doppler velocity of 29 stars in the open clusters

More information

Observer Anomaly(?): Recent Jitter and PSF Variations

Observer Anomaly(?): Recent Jitter and PSF Variations Instrument Science Report TEL 2005-01 Observer Anomaly(?): Recent Jitter and PSF Variations R. L. Gilliland gillil@stsci.edu February 2005 Abstract An anomaly in the HST Pointing Control System (PCS) has

More information

Grism Sensitivities and Apparent Non-Linearity

Grism Sensitivities and Apparent Non-Linearity Instrument Science Report NICMOS 2005-002 Grism Sensitivities and Apparent Non-Linearity Ralph C. Bohlin, Don J. Lindler, & Adam Riess May 10, 2005 ABSTRACT Recent grism observations using the NICMOS instrument

More information

NIRSpec Performance Report NPR / ESA-JWST-RP Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5.

NIRSpec Performance Report NPR / ESA-JWST-RP Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5. NIRSpec Performance Report NPR-2013-011 / ESA-JWST-RP-19649 Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5.2013 Revision: 1 estec European Space Research and Technology Centre Keplerlaan

More information

NICMOS Status and Plans

NICMOS Status and Plans 1997 HST Calibration Workshop Space Telescope Science Institute, 1997 S. Casertano, et al., eds. NICMOS Status and Plans Rodger I. Thompson Steward Observatory, University of Arizona, Tucson, AZ 85721

More information

Predicted Countrates for the UV WFC3 Calibration Subsystem using Deuterium Lamps

Predicted Countrates for the UV WFC3 Calibration Subsystem using Deuterium Lamps Predicted Countrates for the UV WFC3 Calibration Subsystem using Deuterium Lamps S.Baggett, J. Sullivan, and M. Quijada May 17,24 ABSTRACT Predicted WFC3 calibration subsystem countrates have been computed

More information

Internal Flat Field Monitoring II. Stability of the Lamps, Flat Fields, and Gain Ratios

Internal Flat Field Monitoring II. Stability of the Lamps, Flat Fields, and Gain Ratios Instrument Science Report WFPC2 99-01 Internal Flat Field Monitoring II. Stability of the Lamps, Flat Fields, and Gain Ratios C. O Dea, M. Mutchler, and M. Wiggs April 27, 1999 ABSTRACT We have compared

More information

Addendum: GLIMPSE Validation Report

Addendum: GLIMPSE Validation Report August 18, 2004 Addendum: GLIMPSE Validation Report The GLIMPSE Team 1. Motivation In our Validation Report of Jan. 30, 2004, we produced reliability calculations and discussed photometric accuracy estimates

More information

11041,11042,11043: ACS CCD Daily Monitor

11041,11042,11043: ACS CCD Daily Monitor 11041,11042,11043: ACS CCD Daily Monitor Purpose. This program consists of a series of basic tests to measure the readnoise and dark current of the ACS CCDs and to track the growth of hot pixels. The images

More information

COS Cycle 17: 20 programs 149 external orbits 446 internals FUV Detector Sensitivity Monitor 36 0 Oct. 3, 2010

COS Cycle 17: 20 programs 149 external orbits 446 internals FUV Detector Sensitivity Monitor 36 0 Oct. 3, 2010 Prop. ID Title External Internal Status/End obs Cycle 17 11891 NUV MAMA Fold Distribution 0 2 last visit on hold 11894 NUV Detector Dark Monitor 0 114 Oct. 31, 2010 11896 NUV Spectroscopic Sensitivity

More information

SOLSTICE I and II: Ultraviolet Variability. Marty Snow, Bill McClintock, Tom Woods, and Gary Rottman Laboratory for Atmospheric and Space Physics

SOLSTICE I and II: Ultraviolet Variability. Marty Snow, Bill McClintock, Tom Woods, and Gary Rottman Laboratory for Atmospheric and Space Physics SOLSTICE I and II: Ultraviolet Variability Marty Snow, Bill McClintock, Tom Woods, and Gary Rottman Laboratory for Atmospheric and Space Physics Outline SOLSTICE I and II Measurement Equations Individual

More information

WFC3/UVIS and IR Multi-Wavelength Geometric Distortion

WFC3/UVIS and IR Multi-Wavelength Geometric Distortion SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report WFC 2012-07 WFC3/UVIS and IR Multi-Wavelength Geometric Distortion V. Kozhurina-Platais, M. Dulude, T. Dahlen, C. Cox

More information

Flat Fields and Flux Calibrations for the COS FUV Channel at Lifetime Position 4

Flat Fields and Flux Calibrations for the COS FUV Channel at Lifetime Position 4 Instrument Science Report COS 2018-20(v1) Flat Fields and Flux Calibrations for the COS FUV Channel at Lifetime Position 4 William J. Fischer 1, Marc Rafelski 1, and Gisella de Rosa 1 1 Space Telescope

More information

Updated Measurements of ACS/SBC Dark Rates

Updated Measurements of ACS/SBC Dark Rates Instrument Science Report ACS 2017-04 Updated Measurements of ACS/SBC Dark Rates R.J. Avila May 3, 2017 Abstract The results of dark rate monitoring programs for the ACS/SBC are presented here. The dark

More information

The Accuracy of WFPC2 Photometric Zeropoints

The Accuracy of WFPC2 Photometric Zeropoints The Accuracy of WFPC2 Photometric Zeropoints Inge Heyer, Marin Richardson, Brad Whitmore, Lori Lubin July 23, 2004 ABSTRACT The accuracy of WFPC2 photometric zeropoints is examined using two methods. The

More information

WFC3 Calibration Using Galactic Clusters

WFC3 Calibration Using Galactic Clusters Instrument Science Report WFC3 2009-006 WFC3 Calibration Using Galactic Clusters E. Sabbi, J. Kalirai, A. Martel, S. Deustua, S.M. Baggett, T. Borders, H. Bushouse, M. Dulude, B. Hilbert, J. Kim Quijano,

More information

Time Dependence of ACS WFC CTE Corrections for Photometry and Future Predictions

Time Dependence of ACS WFC CTE Corrections for Photometry and Future Predictions Time Dependence of ACS WFC CTE Corrections for Photometry and Future Predictions Adam Riess, Jennifer Mack May 5, 2004 ABSTRACT We present measurements of photometric losses due to imperfect parallel and

More information

WFPC2 Cycle 7 Calibration Plan

WFPC2 Cycle 7 Calibration Plan Instrument Science Report WFPC2-97-06 WFPC2 Cycle 7 Calibration Plan S. Casertano and the WFPC2 group August 18, 1997 ABSTRACT This report describes in detail the WFPC2 observations planned to maintain

More information

Accuracy of VIMOS night-time and daytime wavelength calibrations

Accuracy of VIMOS night-time and daytime wavelength calibrations Accuracy of VIMOS night-time and daytime wavelength calibrations Burkhard Wolff, Steffen Mieske, Juan Carlos Muñoz-Mateos, Sabine Möhler, Michael Hilker 10 December 2014 Abstract Night-time calibrations

More information

Dark Rate of the STIS NUV Detector

Dark Rate of the STIS NUV Detector SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report STIS 2011-03 Dark Rate of the STIS NUV Detector Wei Zheng 1, Charles Proffitt 2, and David Sahnow 1 1 Department of

More information

Cross-Talk in the ACS WFC Detectors. I: Description of the Effect

Cross-Talk in the ACS WFC Detectors. I: Description of the Effect Cross-Talk in the ACS WFC Detectors. I: Description of the Effect Mauro Giavalisco August 10, 2004 ABSTRACT Images acquired with the Wide Field Channel (WFC) of the Advanced Camera for Surveys (ACS) are

More information

WFC3 TV2 Testing: IR Channel Read Noise

WFC3 TV2 Testing: IR Channel Read Noise Instrument Science Report WFC3 2008-04 WFC3 TV2 Testing: IR Channel Read B. Hilbert 12 Feb 2008 ABSTRACT Using data taken during WFC3's Thermal Vacuum 2 (TV2) testing campaign, we have characterized the

More information

Lab 4: Differential Photometry of an Extrasolar Planetary Transit

Lab 4: Differential Photometry of an Extrasolar Planetary Transit Lab 4: Differential Photometry of an Extrasolar Planetary Transit Neil Lender 1, Dipesh Bhattarai 2, Sean Lockwood 3 December 3, 2007 Abstract An upward change in brightness of 3.97 ± 0.29 millimags in

More information

NIRSpec Performance Report NPR / ESA-JWST-RP Author(s): Date: October 11, Stephan Birkmann, Catarina Alves de Oliveira

NIRSpec Performance Report NPR / ESA-JWST-RP Author(s): Date: October 11, Stephan Birkmann, Catarina Alves de Oliveira estec NIRSpec Performance Report NPR-2013-007 / ESA-JWST-RP-19653 Date of Issue: October 11, 2013 Issue: 1 European Space Research and Technology Centre Keplerlaan 1 2201 AZ Noordwijk The Netherlands Tel.

More information

Impacts of focus on aspects of STIS UV Spectroscopy

Impacts of focus on aspects of STIS UV Spectroscopy Instrument Science Report STIS 2018-06 Impacts of focus on aspects of STIS UV Spectroscopy Allyssa Riley 1, TalaWanda Monroe 1, & Sean Lockwood 1 1 Space Telescope Science Institute, Baltimore, MD November

More information

Part III: Circumstellar Properties of Intermediate-Age PMS Stars

Part III: Circumstellar Properties of Intermediate-Age PMS Stars 160 Part III: Circumstellar Properties of Intermediate-Age PMS Stars 161 Chapter 7 Spitzer Observations of 5 Myr-old Brown Dwarfs in Upper Scorpius 7.1 Introduction Ground-based infrared studies have found

More information

The WFC3 IR Blobs Monitoring

The WFC3 IR Blobs Monitoring The WFC3 IR Blobs Monitoring N. Pirzkal, B. Hilbert Nov 13, 2012 ABSTRACT We present new results on the WFC3 IR Blobs based on analysis of data acquired using the WFC3 IR channel from 2010 to 2012. In

More information

Point-Source CCD Photometry with STIS: Correcting for CTE loss

Point-Source CCD Photometry with STIS: Correcting for CTE loss SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Point-Source CCD Photometry with STIS: Correcting for CTE loss Space Telescope Science Institute Methods to measure CTE of STIS CCD (Visual)

More information

NICMOS Focus Field Variations (FFV) and Focus Centering

NICMOS Focus Field Variations (FFV) and Focus Centering Instrument Science Report NICMOS-98-005 NICMOS Focus Field Variations (FFV) and Focus Centering A.Suchkov & G.Galas March 16, 1998 ABSTRACT NICMOS foci are known to vary across detector s field of view.

More information

The TV3 ground calibrations of the WFC3 NIR grisms

The TV3 ground calibrations of the WFC3 NIR grisms The TV3 ground calibrations of the WFC3 NIR grisms H. Kuntschner, H. Bushouse, J. R. Walsh, M. Kümmel July 10, 2008 ABSTRACT Based on thermal vacuum tests (TV3; March/April 2008), the performance of the

More information

COS FUV Focus Determination for the G140L Grating

COS FUV Focus Determination for the G140L Grating Instrument Science Report COS 2012-01 COS FUV Focus Determination for the G140L Grating Parviz Ghavamian 1 Space Telescope Science Institute, Baltimore, MD October 03, 2012 ABSTRACT The procedures for

More information

High Signal-to-Noise, Differential NICMOS Spectrophotometry

High Signal-to-Noise, Differential NICMOS Spectrophotometry Instrument Science Report NICMOS 2003-001 High Signal-to-Noise, Differential NICMOS Spectrophotometry R.L. Gilliland, S. Arribas January 7, 2003 ABSTRACT We report analysis for NICMOS CAL/9642, High S/N

More information

NICMOS Cycles 13 and 14 Calibration Plans

NICMOS Cycles 13 and 14 Calibration Plans Instrument Science Report NICMOS 2005-005 NICMOS Cycles 13 and 14 Calibration s Santiago Arribas, Eddie Bergeron, Roelof de Jong, Sangeeta Malhotra, Bahram Mobasher, Keith Noll, Al Schultz, Tommy Wiklind,

More information

GALEX GR1 Instrument Performance and Calibration Review

GALEX GR1 Instrument Performance and Calibration Review GALEX GR1 Instrument Performance and Calibration Review Patrick Morrissey Caltech Operations Overview GALEX is operating quite smoothly in 2004. Active space weather has mostly dissipated. FUV is shut

More information

Performance of the NICMOS ETC Against Archived Data

Performance of the NICMOS ETC Against Archived Data Performance of the NICMOS ETC Against Archived Data M. Sosey June 19, 2001 ABSTRACT A robust test of the newest version of the NICMOS Exposure Time Calculator (ETC) was conducted in order to assess its

More information

Statistics of Accumulating Signal

Statistics of Accumulating Signal Instrument Science Report NICMOS 98-008 Statistics of Accumulating Signal W.B. Sparks, STScI May 1998 ABSTRACT NICMOS detectors accumulate charge which can be non-destructively. The charge accumulation

More information

Astrometric Performance of STIS CCD CTI Corrections on Omega Cen Images

Astrometric Performance of STIS CCD CTI Corrections on Omega Cen Images Instrument Science Report STIS 2015-05 Astrometric Performance of STIS CCD CTI Corrections on Omega Cen Images John Biretta, Sean Lockwood, and John Debes September 28, 2015 ABSTRACT We are in the process

More information

On the calibration of WFCAM data from 2MASS

On the calibration of WFCAM data from 2MASS On the calibration of WFCAM data from 2MASS Authors: Simon Hodgkin, Mike Irwin Draft: September 28 th 2006 Modifications: ID: VDF-TRE-IOA-00011-0000* 1 Introduction The requirement on VDFS is to photometrically

More information

WFC3/IR Persistence as Measured in Cycle 17 using Tungsten Lamp Exposures

WFC3/IR Persistence as Measured in Cycle 17 using Tungsten Lamp Exposures WFC3/IR Persistence as Measured in Cycle 17 using Tungsten Lamp Exposures Knox S. Long, Sylvia Baggett, Susana Deustua & Adam Riess November 17, 2010 ABSTRACT Like most IR arrays, the IR detector incorporated

More information

estec NIRSpec Technical Note NTN Spectral intensity of the MSA glow sources Abstract:

estec NIRSpec Technical Note NTN Spectral intensity of the MSA glow sources Abstract: NIRSpec Technical Note NTN-2013-014 Authors: P. Ferruit & G. Giardino Date of Issue: 29.11.2013 Version: 1 estec European Space Research and Technology Centre Keplerlaan 1 2201 AZ Noordwijk The Netherlands

More information

Data Processing in DES

Data Processing in DES Data Processing in DES Brian Yanny Oct 28, 2016 http://data.darkenergysurvey.org/fnalmisc/talk/detrend.p Basic Signal-to-Noise calculation in astronomy: Assuming a perfect atmosphere (fixed PSF of p arcsec

More information

Relative Astrometry Within ACS Visits

Relative Astrometry Within ACS Visits Instrument Science Report ACS 2006-005 Relative Astrometry Within ACS Visits Richard L. White August 07, 2006 ABSTRACT The log files from APSIS, the ACS science team s image processing pipeline, have been

More information

XMM-Newton Optical-UV Monitor: introduction and calibration status OM instrument and calibration

XMM-Newton Optical-UV Monitor: introduction and calibration status OM instrument and calibration XMM-Newton Optical-UV Monitor: introduction and calibration status OM instrument and calibration Antonio Talavera XMM-Newton Science Operation Centre, ESAC, ESA OM: Instrument Description 30 cm Ritchey-Chretien

More information

Atmospheric Extinction

Atmospheric Extinction Atmospheric Extinction Calibrating stellar photometry requires correction for loss of light passing through the atmosphere. Atmospheric Rayleigh and aerosol scattering preferentially redirects blue light

More information

A Recalibration of Optical Photometry Based on STIS Spectrophotometry

A Recalibration of Optical Photometry Based on STIS Spectrophotometry The 2005 HST Calibration Workshop Space Telescope Science Institute, 2005 A. M. Koekemoer, P. Goudfrooij, and L. L. Dressel, eds. A Recalibration of Optical Photometry Based on STIS Spectrophotometry J.

More information

On-orbit Calibration of ACS CTE Corrections for Photometry

On-orbit Calibration of ACS CTE Corrections for Photometry On-orbit Calibration of ACS CTE Corrections for Photometry Adam Riess August 15, 2003 ABSTRACT We present the first on-orbit calibration of the photometric losses due to imperfect CTE on ACS HRC and WFC.

More information

Lab 7: The H-R Diagram of an Open Cluster

Lab 7: The H-R Diagram of an Open Cluster Lab 7: The H-R Diagram of an Open Cluster Due Date: 2007 Nov 27 (after thanksgiving) 1 Introduction: The HR Diagram In this two week project you will do absolute (not differential) photometry with a CCD

More information

1 The Preliminary Processing

1 The Preliminary Processing AY 257 Modern Observational Techniques...23 1 The Preliminary Processing Frames must be corrected for a bias level and quantum efficiency variations on all scales. For a minority of CCDs and most near-ir

More information

EM Waves in Media. What happens when an EM wave travels through our model material?

EM Waves in Media. What happens when an EM wave travels through our model material? EM Waves in Media We can model a material as made of atoms which have a charged electron bound to a nucleus by a spring. We model the nuclei as being fixed to a grid (because they are heavy, they don t

More information

C. Watson, E. Churchwell, R. Indebetouw, M. Meade, B. Babler, B. Whitney

C. Watson, E. Churchwell, R. Indebetouw, M. Meade, B. Babler, B. Whitney Reliability and Completeness for the GLIMPSE Survey C. Watson, E. Churchwell, R. Indebetouw, M. Meade, B. Babler, B. Whitney Abstract This document examines the GLIMPSE observing strategy and criteria

More information

Validation of GOMOS version 6.01 ozone profiles using ground-based lidar observations. October 2012

Validation of GOMOS version 6.01 ozone profiles using ground-based lidar observations. October 2012 Validation of GOMOS version 6.01 ozone profiles using ground-based lidar observations October 2012 Anne.van.Gijsel@knmi.nl Introduction In this study we have compared GOMOS IPF version 6.01 ozone profiles

More information

WFC3 TV2 Testing: UVIS-2 Dark Frames and Rates

WFC3 TV2 Testing: UVIS-2 Dark Frames and Rates Instrument Science Report WFC3 007-06 WFC3 TV Testing: UVIS- Dark Frames and Rates A.R. Martel November 0, 007 ABSTRACT WFC3 underwent its second thermal vacuum testing in the Space Environment Simulation

More information

Observing with the Infrared Spectrograph

Observing with the Infrared Spectrograph Observing with the Infrared Spectrograph C. Grillmair, L. Armus GO Workshop 21-22 November 2002 Outline 1) Meet the IRS IST 2) Basic IRS capabilities 3) Observing and readout modes 4) Data products and

More information

Breathing, Position Drift, and PSF Variations on the UVIS Detector

Breathing, Position Drift, and PSF Variations on the UVIS Detector SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report WFC3 1-1 Breathing, Position Drift, and PSF Variations on the UVIS Detector L. Dressel July 13, 1 ABSTRACT This study

More information

The Effective Spectral Resolution of the WFC and HRC Grism

The Effective Spectral Resolution of the WFC and HRC Grism The Effective Spectral Resolution of the WFC and HRC Grism A. Pasquali, N. Pirzkal, J.R. Walsh, R.N. Hook, W. Freudling, R. Albrecht, R.A.E. Fosbury March 7, 2001 ABSTRACT We present SLIM simulations of

More information

COS Monthly Status Review. August 19, Ball

COS Monthly Status Review. August 19, Ball COS Ball Page 1 Agenda Management Changes Results from Initial Calibration BATC Presentation Financial Splinter J. Green E. Wilkinson R. Higgins GSFC/Ball/CU Page 2 Management Changes Jon Morse has moved

More information

Monitoring The HRC-S UV Rate: Observations of Vega

Monitoring The HRC-S UV Rate: Observations of Vega Monitoring The HRC-S UV Rate: Observations of Vega Deron Pease, Vinay Kashyap, Jeremy Drake and Michael Juda 10 May 2005 Abstract We present an interim report on Chandra HRC-S calibration observations

More information

Verification of COS/FUV Bright Object Aperture (BOA) Operations at Lifetime Position 3

Verification of COS/FUV Bright Object Aperture (BOA) Operations at Lifetime Position 3 Instrument Science Report COS 2015-05(v1) Verification of COS/FUV Bright Object Aperture (BOA) Operations at Lifetime Position 3 Andrew Fox 1, John Debes 1, & Julia Roman-Duval 1 1 Space Telescope Science

More information

UNIVERSITY COLLEGE LONDON. Measuring a Stellar Spectrum

UNIVERSITY COLLEGE LONDON. Measuring a Stellar Spectrum University Of London Observatory 1 Introduction UNIVERSITY COLLEGE LONDON Measuring a Stellar Spectrum 1B30 Practical Astronomy 1B13 Frontiers of Astronomy The primary way astronomers find out about physical

More information

Mimir NIR Spectroscopy Data Processing Cookbook V2.0 DPC

Mimir NIR Spectroscopy Data Processing Cookbook V2.0 DPC Mimir NIR Spectroscopy Data Processing Cookbook V2.0 DPC - 20111130 1. Fetch and install the software packages needed a. Get the MSP_WCT, MSP_CCS, MSP_SXC packages from the Mimir/Software web site: http://people.bu.edu/clemens/mimir/software.html

More information

Impressions: First Light Images from UVIT in Orbit

Impressions: First Light Images from UVIT in Orbit Impressions: First Light Images from UVIT in Orbit Drafted by S N Tandon on behalf of UVIT team. December 4, 2015; V1.0 1. Introduction: Ultraviolet Imaging Telescope (UVIT) is the long wavelength eye

More information

Exploring Data. Keck LRIS spectra. Handbook of CCD Astronomy by Steve Howell Chap. 4, parts of 6

Exploring Data. Keck LRIS spectra. Handbook of CCD Astronomy by Steve Howell Chap. 4, parts of 6 Exploring Data Keck LRIS spectra Handbook of CCD Astronomy by Steve Howell Chap. 4, parts of 6 FITS: Flexible Image Transport System Digital file format used to store astronomical images, data, and catalogs.

More information

Calibration Goals and Plans

Calibration Goals and Plans CHAPTER 13 Calibration Goals and Plans In This Chapter... Expected Calibration Accuracies / 195 Calibration Plans / 197 This chapter describes the expected accuracies which should be reached in the calibration

More information

Extraction of Point Source Spectra from STIS Long Slit Data

Extraction of Point Source Spectra from STIS Long Slit Data 1997 HST Calibration Workshop Space Telescope Science Institute, 1997 S. Casertano, et al., eds. Extraction of Point Source Spectra from STIS Long Slit Data J. R. Walsh Spect Telescope European Coordinating

More information

WFPC2 Dark Current vs. Time

WFPC2 Dark Current vs. Time WFPC2 Dark Current vs. Time J. Mack, J. Biretta, S. Baggett, C. Proffitt June 7, 2001 ABSTRACT On-going measurements of the dark current in the WFPC2 detectors indicate that the average level of dark current

More information

70µm Warm Campaign 10K Sensitivity Tests D. Frayer, A. Noriega-Crespo, S. Wachter, & D. Fadda (MIPS-IST) 2005-Jun-16, Version 1.

70µm Warm Campaign 10K Sensitivity Tests D. Frayer, A. Noriega-Crespo, S. Wachter, & D. Fadda (MIPS-IST) 2005-Jun-16, Version 1. 1 70µm Warm Campaign 10K Sensitivity Tests D. Frayer, A. Noriega-Crespo, S. Wachter, & D. Fadda (MIPS-IST) 2005-Jun-16, Version 1. Abstract Data were recently taken to quantify the 70µm sensitivity and

More information

SCIAMACHY IN-FLIGHT CALIBRATION

SCIAMACHY IN-FLIGHT CALIBRATION SCIAMACHY IN-FLIGHT CALIBRATION Ralph Snel SRON Netherlands Institute for Space Research Sorbonnelaan 2, 3584 CA Utrecht, the Netherlands Email: R.Snel@sron.nl ABSTRACT The options for SCIAMACHY in-flight

More information

SPACE TELESCOPE SCIENCE INSTITUTE. ACS Polarization Calibration: Introduction and Progress Report

SPACE TELESCOPE SCIENCE INSTITUTE. ACS Polarization Calibration: Introduction and Progress Report 1 ACS Polarization Calibration: Introduction and Progress Report J. Biretta, V. Platais, F. Boffi, W. Sparks, J. Walsh Introduction: Theory / ACS Polarizers / Supported Modes Potential Issues for ACS Polarization

More information

TECHNICAL REPORT. Doc #: Date: Rev: JWST-STScI , SM-12 August 31, Authors: Karl Gordon, Ralph Bohlin. Phone:

TECHNICAL REPORT. Doc #: Date: Rev: JWST-STScI , SM-12 August 31, Authors: Karl Gordon, Ralph Bohlin. Phone: When there is a discrepancy between the information in this technical report and information in JDox, assume JDox is correct. TECHNICAL REPORT Title: Title: JWST Absolute Flux Calibration II: Expanded

More information

INAF-Osservatorio astrofisico di Torino Technical Report nr. 167

INAF-Osservatorio astrofisico di Torino Technical Report nr. 167 INAF-Osservatorio astrofisico di Torino Technical Report nr. 167 Simulation of Visible Light and UV images for the METIS coronagraph Alessandro Bemporad Pino Torinese, 9 ottobre 2014 Simulation of Visible

More information

Report on the new EFOSC2 VPH grisms

Report on the new EFOSC2 VPH grisms Report on the new EFOSC2 VPH grisms Ivo Saviane Lorenzo Monaco v 1.0 March 01, 2008 1 Introduction In January 2008 the ULTRASPEC project delivered two volume-phased holographic grisms (VPHG) to be used

More information

New On-Orbit Sensitivity Calibration for All STIS Echelle Modes

New On-Orbit Sensitivity Calibration for All STIS Echelle Modes New On-Orbit Sensitivity Calibration for All STIS Echelle Modes Alessandra Aloisi, Ralph Bohlin, and Jessica Kim Quijano January 18, 2007 ABSTRACT On-orbit sensitivities for the 32 medium- and high-resolution

More information

PACS Spectrometer Wavelength Calibration

PACS Spectrometer Wavelength Calibration PACS Spectrometer Wavelength Calibration H. Feuchtgruber Max-Planck-Institut für extraterrestrische Physik, Garching PACS WaveCal 1 Description of the Calibration Problem - Blue array: 400 pixels - Red

More information

Sub-percent Accuracy for a Factor of a Million in Flux

Sub-percent Accuracy for a Factor of a Million in Flux WFC3 Instrument Science Report 2019-01 Calibration of the WFC3-IR Count-rate Nonlinearity, Sub-percent Accuracy for a Factor of a Million in Flux A. G. Riess, Gautham Narayan, Annalisa Calamida January

More information

The IRS Flats. Spitzer Science Center

The IRS Flats. Spitzer Science Center Spitzer Science Center Table of Contents The IRS Flats 1 Chapter 1. The Purpose of this Document... 3 Chapter 2.... 4 2.1 Make a finely-spaced map of a star....4 2.2 Simulate an extended source...6 2.3

More information

Transiting Exoplanet in the Near Infra-red for the XO-3 System

Transiting Exoplanet in the Near Infra-red for the XO-3 System Transiting Exoplanet in the Near Infra-red for the XO-3 System Nathaniel Rodriguez August 26, 2009 Abstract Our research this summer focused on determining if sufficient precision could be gained from

More information

Errors in the Data Measurements

Errors in the Data Measurements Voyager PLS Error Analysis Errors in the Data Measurements For all purposes in this Error Analysis memo, it will be assumed that currents are measured and displayed in femtoamps. This is the natural range

More information

The Impact of x-cte in the WFC3/UVIS detector on Astrometry

The Impact of x-cte in the WFC3/UVIS detector on Astrometry Instrument Science Report WFC3 2014-02 The Impact of x-cte in the WFC3/UVIS detector on Astrometry Jay Anderson April 4, 2014 ABSTRACT Recent observations of the center of globular cluster Omega Centauri

More information

Lecture 9. November 1, 2018 Lab 5 Analysis

Lecture 9. November 1, 2018 Lab 5 Analysis Lecture 9 November 1, 2018 Lab 5 Analysis News Lab 2 Handed back with solution; mean: 92.1, std dev: 5.5 Lab 3 Handed back next week (I hope). Lab 4 Due November 1 (today) News Lab 5 (Transiting Exoplanets)

More information

Measurement of the stellar irradiance

Measurement of the stellar irradiance Measurement of the stellar irradiance Definitions Specific Intensity : (monochromatic) per unit area normal to the direction of radiation per unit solid angle per unit wavelength unit (or frequency) per

More information

COS FUV Dispersion Solution Verification at the New Lifetime Position

COS FUV Dispersion Solution Verification at the New Lifetime Position SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report COS 2013-06(v1) COS FUV Dispersion Solution Verification at the New Lifetime Position Paule Sonnentrucker 1, Julia

More information

WHAT PHOTOMETRIC PRECISION CAN I ACHIEVE? DAVID BOYD

WHAT PHOTOMETRIC PRECISION CAN I ACHIEVE? DAVID BOYD WHAT PHOTOMETRIC PRECISION CAN I ACHIEVE? DAVID BOYD If you start using a CCD camera to carry out photometry on variable stars, this is a question that sooner or later you will ask yourself. Prompted by

More information