DEVELOPMENT OF A SENSITIVE THERMAL DESORPTION GC-MS METHOD USING SELECTIVE ION MONITORING FOR A WIDE RANGE OF VOCS

Size: px
Start display at page:

Download "DEVELOPMENT OF A SENSITIVE THERMAL DESORPTION GC-MS METHOD USING SELECTIVE ION MONITORING FOR A WIDE RANGE OF VOCS"

Transcription

1 DEVELOPMENT OF A SENSITIVE THERMAL DESORPTION GC-MS METHOD USING SELECTIVE ION MONITORING FOR A WIDE RANGE OF VOCS C Jia *, S Batterman and S Chernyak Department of Environmental Health Sciences, University of Michigan 1420 Washington Heights, Ann Arbor, MI ABSTRACT Thermal desorption (TD) of volatile organic compounds (VOCs) collected on adsorbent tubes is increasingly used in IAQ applications. This paper develops and evaluates the use of a TD-GC-MS method using selective ion monitoring () that can significantly improve method detection limits (MDLs). Two protocols were developed for 99 target compounds: a full scan method with ions ranging from 26 to 250 amu, and a method using 16 windows over the min GC run time. In the laboratory, calibration curves and MDLs were determined for the target compounds using both modes. In the field, duplicate samples were taken inside and outside of 38 residences, and analyzed using both scan and modes. Calibration curves for scan and modes were very similar, allowing a single calibration for both modes. MDLs in mode were 1.5 to 6 times lower than those in the scan mode and often well below 0.01 μg m -3. In the field, many more VOCs were detected and quantified using mode. The new method offers greater sensitivity and maintains the high selectivity of the traditional GC-MS method. INDEX TERMS Adsorbents, Mass spectrometry, Sampling, Volatile organic compounds, VOCs. INTRODUCTION Adsorbent sampling followed by thermal desorption, gas chromatography and mass spectrometry (TD-GC-MS) has been extensively used for VOC analysis (Peng and Batterman 2000, Woolfenden 1997, Heavner et al. 1992). The MS detector is considered to provide definitive identification of compounds since the ion spectrum obtained by the MS can often resolve co-eluting pollutants. Co-elution is a significant problem in environmental sampling using less specific detectors, e.g., flame ionization detectors (FID) and electron capture detectors (ECD). Compounds with similar structures often co-elute, e.g., isomers such as p-xylene and m-xylene. Sometimes compounds with very different structures co-elute, e.g., benzene and carbon tetrachloride. Either case can result in misidentification and incorrect quantification using non-specific detectors. Non-specific detectors have the advantage that they can often provide much higher sensitivity than MS detectors, e.g., FID have high sensitivity to many hydrocarbons and a wide linear range. Thus, GC-MS is sometimes coupled to FID, with the FID used for quantification (Na et al. 2004, Mohamed et al. 2002, Jurvelin et al. 2001, Brickus et al. 1998, Cheng et al. 1997). Other configurations include dual-column tandem GC/ECD-FID (Begerow et al. 1996), or GC/FID-RIM (retention index monitoring), again with the purpose of confirming the identification of compounds and simultaneously obtaining high sensitivity. These methods have been used in a variety of settings. For example, Fernandez-Martinez et al. (1999) developed a method for 15 VOCs using GC-FID and GC-MS, and obtained MDLs for 4 L samples from to μg m -3 (GC-FID), to μg m -3 (GC-MS, mode), and to 0.08 μg m -3 (GC-MS, SCAN mode). However, multi-detector approaches require specialized equipment, are relatively costly, and calibrations and analyses can be time consuming. A mass spectrometer has two modes: scan and selective ion monitoring (). In scan mode, the MS scans every ion over a specified (and typically wide) range. In mode, only specific ions are measured. mode is routinely applied to identify compounds in environmental samples since many compounds may be present and the ion spectrum can be used to confirm the identification of compounds. However, scan mode may not provide sufficient sensitivity for many important environmental contaminants. MS sensitivity can be significantly increased using the selective ion monitoring () mode, potentially yielding * Corresponding author chunrong@umich.edu 2194

2 sensitivities that are comparable to non-specific detectors like FIDs (Message 1984, Chen 1979). Within a specified ion interval, the sensitivity can be increased significantly, for example, the Agilent 5973 inert GC/MS system has electronic impact (EI) scan sensitivity of 60:1 s/n (signal to noise) for 1 pg OFN (octofluoronaphthalene) scanning from amu at nominal m/z 272 ion, and EI sensitivity of 10:1 s/n for pg OFN at nominal m/z 272 ion (Agilent). Adjusted to the same s/n, the mode sensitivity increase is about 8 times. MS in mode has been occasionally used for VOC analyses. Almasi et al. (1993) detected 41 VOCs with concentration range from 0.8 to 80 μg m -3 using small sampling volumes, i.e., 0.06 L. Heavner et al. (1992) used a multisorbent TD-GC-MS method for 28 target VOCs and obtained LODs from 0.03 to 0.8 μg m -3 for 4 L samples. Otson et al. (1994) collected passive samples in 757 Canadian residences and analyzed 52 target VOCs using a method, but samples were desorbed using solvent extraction, which is less sensitive than thermal desorption. The goal of this study is to develop and evaluate a method for VOCs measured in indoor and ambient air. The method is evaluated using laboratory and field tests. EXPERIMENTAL Thermal desorption tubes (TDTs) The TDTs have been previously described by Peng and Batterman (2000). Each stainless steel tube is packed with two sorbents (Tenax GR followed by Carbosieve SIII) for active sampling, or with a single sorbent (Tenax GR) for passive sampling since VOCs cannot reach the second sorbent bed by diffusion. Target compounds Ninety-nine VOCs were selected based on their toxicity and occurrence in indoor and outdoor air. Many of these VOCs are listed in the US EPA TO-14 (US EPA 1999A) and TO-15 (US EPA 1997) methods, and include aromatics, halogenated compounds, terpenes, alkanes, etc. This paper describes results for six of these VOCs in two groups, selected on the basis of their occurrence and concentration in indoor and outdoor environments in a previous study (Jia et al. 2004). VOCs in the first group are frequently detected at relatively high concentrations: toluene, limonene and 1,4-dichlorobenzene. VOCs in the second group are found at low concentrations and are detected less frequently: chloroform, styrene and heptadecane. The six VOCs were also selected to represent aromatic, terpene, chlorinated and aliphatic compounds. Laboratory performance evaluation Seven point calibration curves were obtained in scan and modes using an Agilent 6890/5973 running Chemstation (G1701BA, Version B.01, Hewlett-Packard, Palo Alto, CA, USA) at concentrations from 0.15 to 50.0 ng μl -1, spanning the range of interest to determine instrument sensitivity and linearity. Calibrations used 2 μl aliquots of liquid standards and duplicate TDTs at each concentration. Method detection limits (MDLs) were determined using low concentration standard solutions that contained and ng μl -1 of each compound for scan and modes, respectively. Each tube was loaded with 2 μl of these standards, thus obtaining injection masses of 0.03 and 0.01 ng tube -1 for scan and modes, respectively. The MDL was calculated as 3.14 S where S = the standard deviation of 7 replicate analyses, and 3.14 = Student s t-value for 99% confidence level (US EPA 1999B). Field study The field study was conducted in Ann Arbor, MI, USA. Duplicate passive samplers were placed inside and outside of houses for 3-4 weekdays during August and September, The estimated sampling volume was ~2 L. Collected samples were analyzed using scan and modes. Overall, 38 pairs of indoor measurements and 38 pairs of outdoor measurements were collected. Of these, one indoor and one outdoor sample failed quality assurance/quality control requirements, so comparisons omit these two sample pairs. Sample analysis and quality control The GC-MS scan method and quality control procedure is described elsewhere (Peng and Batterman 2000). In mode, we applied the same GC conditions, but divided the target compounds into 16 groups, based on their retention times, and assigned from 4 to 19 ions to each time window in the min GC run. Several compounds were included in two or even three adjacent windows due to the possibility of shifting peaks during the run. A target ion (for both quantification and confirmation) and one or two qualifier ions (for confirmation) were selected for each compound. Totally 79 different ions (from 29 to 260 m/z) were monitored over the whole run. 2195

3 RESULTS Laboratory performance For each mode, the precisions for duplicate tubes across the concentrations used in the calibration were mostly below 10%, and were below 20% at the lowest concentrations (below 0.5 ng μl -1 ). Calibration curves were based on linear regressions with intercepts set to 0 in order to increase the sensitivity. R 2 values were high, showing good linearity over the concentration range tested. and mode calibrations showed good agreement. Based on a sample volume of 4 L used in routine analysis (US EPA 1999B), MDLs for target compounds were below μg m -3 in the scan mode and below μg m -3 in mode (Table 1). -mode sensitivity was 1.5 to 5.7 times better than that for scan mode. This increase was lower than expected, in part because we analyzed 100 compounds in each run, and most window included 10 compounds. With fewer target compounds, sensitivity will increase significantly. However, the increased sensitivity obtained here was still valuable as the mode detected ions that were not found by scan mode in the low concentration samples, as seen in the field study described next. Table 1. Calibration curves and MDLs measured in scan and modes. Compound MDL Slope R 2 MDL (μg ( 礸 m -3 ) Slope R 2 MDL (μg ( 礸 m -3 ) / Toluene ,4-Dichlorobenzene d-limonene Chloroform Styrene n-hexadecane Field study The six VOCs were detected in over 90% of the homes sampled. In ambient air, toluene, d-limonene and chloroform were frequently detected. 1,4-dichlorobenzene and styrene were less frequently detected. n-hexadecane was not detected in ambient air. In the first group of VOCs, indoor concentrations of toluene and d-limonene frequently exceeded 10 μg m -3 (Table 2). The large standard deviations reflected the differences between homes. Outdoor concentrations were much lower. Concentrations of VOCs in the second group were low indoors (<0.5 μg m -3 ) and even lower outdoors (<0.1 μg m -3 ). As expected, standard deviations of outdoor concentrations for all VOCs were small, reflecting the limited range of concentrations found. Detection frequencies of VOCs in scan and modes were computed after pooling the indoor and outdoor data (Table 2). Four VOCs were detected more frequently using mode, e.g., chloroform was detected in 19% of the samples in the scan mode, but 97% in the mode. Detection frequencies for toluene and n-hexadecane were unchanged. Table 2. Concentrations and detection probabilities for 6 compounds. Indoor Outdoor Sim Compound Mean Median St. Dev. Detected Mean St. Dev. Median Detected Detected Detected (µg m -3 ) (µg m -3 ) (µg m -3 ) (%) (µg m -3 ) (µg m -3 ) (µg m -3 ) (%) (%) (%) Toluene d-limonene ,4-Dichlorobenzene Chloroform Styrene n-hexadecane Figure 1 shows typical scan and chromatograms for ion 68 representing d-limonene. The mode peak is higher and smoother than that in the scan mode. The increased sensitivity is derived from the abundance of the 2196

4 quantification ions. The abundances of ion 68 are 5063 and 7342 in scan and modes, respectively, an increase of ~1.5 times. Since abundances of internal standards increase proportionally, one calibration still works. : Abundance Retention time : Abundance Retention time Figure 1. Target ions of d-limonene in scan and modes. Concentrations in both modes were calculated using the same calibration curves, plotted and fitted using linear regression (Figure 2). Concentrations were within 10% for the two modes for toluene, d-limonene, 1,4-dichlorobenzene and styrene. The small differences might result from sampling, analysis and calibration errors. For chloroform, mode measurements were 19% lower than scan mode determinations, though the correlation was high (R 2 =0.97). Most chloroform concentrations were <1 μg m -3. and scan mode measurements for n-hexadecane showed good agreement, though lower correlation (R 2 =0.51). Again, this might result from the low concentrations (<0.5 μg m -3 ) encountered n = 72 y = x R 2 = Toluene n = 41 y = x R 2 = d-limonene n = 23 y = x R 2 = ,4-Dichlorobenzene 2.00 n = y = x R 2 = Chloroform Styrene n = 36 y = x R 2 = n = 33 y = x R 2 = n-hexadecane Figure 2. Concentrations of 6 compounds measured by scan and modes. 2197

5 DISCUSSION While the laboratory-based determination of -mode MDLs showed modest (1.5 to 5.7-fold) improvements compared to scan mode, -mode analyses significantly increased the detection frequency of VOCs in the field study. A scan-mode chromatogram reflects the many substances found in complex environmental samples. mode reduces noise and achieves better signal-to-noise ratios, resulting in clearer peaks that are easier to identify and quantify. Thus, in practice, mode detects about 50% more compounds than the traditional scan mode, including compounds that are rarely detected in scan mode analyses, e.g., 1,1,1-trichloroethane, trichloroethylene and phenol. and scan mode calibration curves did not change significantly, e.g., calibration curves were within 15%, thus allowing a single and convenient calibration for both modes. It is important to measure even low concentration VOCs. Some VOCs may affect human health even at low levels. Also, more complete detection provides information that can be used to identify sources. The mode may also facilitate the detection of compounds that are masked by the solvent from the internal standard solution, e.g., 1,3-butadiene, a very reactive and toxic compound, as well as methanol and other VOCs. The method offers greater sensitivity while maintaining the high selectivity of traditional MS detector. The new method appears to provide a viable option for analyzing environmental samples. ACKNOWLEDGEMENTS Portions of the study were financially supported by the American Chemistry Council (Grant 2401) and the Michigan Education and Research Center which is funded by the National Institute of Occupational Safety and Health (Grant T42 CC ). REFERENCES Agilent 5973 Inert GC/MS Specifications, Almasi E., Kirshen N. and Kern H The determination of sub part-per-billion levels of volatile organic-compounds in air by preconcentration from small sample volumes, International Journal of Environmental Analytical Chemistry. 52 (1-4): Begerow J., Jermann E., Keles T., Koch T. and Dunemann L Screening method for the determination of 28 volatile organic compounds in indoor and outdoor air at environmental concentrations using dual-column capillary gas chromatography with tandem electron-capture-flame ionization detection, Journal of Chromatography A. 749 (1): Chen ECM Selective ion monitoring, Practical Mass Spectrometry. Edited by Middleditch B.S. Plenum Press, New York. Cheng L., Fu L., Angle RP. and Sandhu HS Seasonal variations of volatile organic compounds in Edmonton, Alberta, Atmospheric Environment. 31 (2): Fernandez-Martinez G.; Lopez-Mahia P.; Muniategi-Lorenzo S. et al., Development of a method for determination of volatile organic compounds (C6-C9) by thermal desorption-gas chromatography: application to urban and rural atmospheres, Analytical Letters. 32 (14): Heavner DL., Ogden MW. and Nelson PR Multisorbent thermal desorption/gas chromatography/mass selective detection method for the determination of target volatile organic compounds in indoor air, Environ Science & Techn. 26 (9): Jia C., Godwin C., Batterman S. and Alfred F Sources and significance of VOC exposures in three microenvironments: schools, residences and vehicles, Air & Waste Management Association Conference, Indianapolis, IN, USA, June Jurvelin J., Edwards R., Saarela K., Laine-Ylijoki J., et al Evaluation of VOC measurements in the EXPOLIS study, Journal of Environmental Monitoring. 3(1): Jurvelin J., Edwards R., Saarela K., Laine-Ylijoki J., et al Distributions of indoor and outdoor air pollutants in Rio de Janeiro, Brazil: implications to indoor air quality in Bayside offices, Environmental Science and Technology. 32 (22): Message GM Practical Aspects of Gas Chromatography/Mass Spectrometry. John Wiley & Sons, Inc. pp Mohamed M., Kang D. and Aneja V Volatile organic compounds in some urban locations in United States, Chemosphere. 47 (8): Na K., Kim YP., Moon I. and Moon KC Chemical composition of major VOC emission sources in the Seoul atmosphere, Chemosphere. 55 (4): Otson R., Fellin P. and Tran Q VOCs in representative Canadian residences, Atmospheric Environment, 28 (22):

6 Peng C. and Batterman S Performance evaluation of a sorbent tube sampling method using short path thermal desorption for volatile organic compounds, Journal of Environmental Monitoring. 2 (4): US EPA Compendium of Methods for the Determination of Toxic Organic Compounds in Ambient Air, TO-15, U. S. Environmental Protection Agency, Washington, DC, 2nd ed., Report No. EPA/625/R-96/010b. US EPA. 1999A. Compendium Method TO-14A. Determination of volatile organic compounds in ambient air using specially prepared canisters with subsequent analysis by gas chromatograpy. January US EPA. 1999B. Compendium Method TO-17. Determination of volatile organic compounds in ambient air using active sampling onto sorbent tubes. January Woolfenden E Monitoring VOCs in air using sorbent tubes followed by thermal desorption capillary GC analysis: Summary of data and practical guidelines, Journal of the Air & Waste Management Association. 47 (1):

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ Application Note Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ -3 Application Note Abstract This application note demonstrates the

More information

AA-ANSIBIFMA-Jun2408, Page 1 of 16

AA-ANSIBIFMA-Jun2408, Page 1 of 16 356-001-01AA-ANSIBIFMA-Jun2408, Page 1 of 16 BERKELEY ANALYTICAL ASSOCIATES, LLC 815 Harbour Way South, Suite 6 Richmond, CA 94804-3612 Ph. 510-236-2325; Fax 510-236-2335 E-mail baalab@berkeleyanalytical.com

More information

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)

More information

Application Note 032

Application Note 032 Application Note 3 Analysis of Sulfur Compounds Using On-line and Off line TD GC Application Note Abstract This application note shows how Markes TD technology is compatible with trace level sulfur compounds,

More information

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS ENVIRONMENTAL application note U.S. EPA VOLATILE ORGANICS METHOD 524.2 USING PURGE AND TRAP GC/MS Cheri Coody, Public Health Laboratory, Mississippi State Health Department, Jackson, MS USA Michael J.

More information

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method

More information

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software NEMC, Washington D.C. August 6 th 2012 Nicola Watson Environmental Specialist Markes International

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

Thermal Desorption Technical Support

Thermal Desorption Technical Support Thermal Desorption Technical Support Note 86a: US EPA Method TO-17 for Monitoring Air Toxics in Ambient Air Using Sorbent Tubes and Automated, Cryogen-free Thermal Desorption Application Note Environmental,

More information

COMPLIANCE EMISSIONS TEST California Dept. of Public Health Standard Method Version 1.1 and FloorScore Flooring Evaluation

COMPLIANCE EMISSIONS TEST California Dept. of Public Health Standard Method Version 1.1 and FloorScore Flooring Evaluation COMPLIANCE EMISSIONS TEST California Dept. of Public Health Standard Method Version 1.1 and FloorScore Flooring Evaluation SAMPLE DESCRIPTION & TESTING PARAMETERS Tesoro Woods submitted exemplars of their

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION. APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Author: Kira. Yang PerkinElmer, Inc. Shanghai, China Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS Introduction

More information

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract. Application Note 8 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-5 Application Note Environmental, Canister, TO-5, Air Monitoring, Air Toxics Abstract This application

More information

Vapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs

Vapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs Vapor Intrusion Sampling Options: Performance Data for s, Badges, and Sorbent Tubes for VOCs Linda S. Coyne SKC Inc., 863 Valley View Road, Eighty Four, PA 1533 George Havalias, Maria C. Echarte American

More information

Determination of Volatile Organic Compounds in Air

Determination of Volatile Organic Compounds in Air JSB is an authorised partner of Determination of Volatile Organic s in Air Anne Jurek #126 Introduction Air pollution is a growing problem due to the global economy and industrial development in many countries.

More information

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System Application Note Residual Solvent Analysis of USP Method Residual Solvents on the Agilent 889 GC System Author Lukas Wieder, Jie Pan, and Rebecca Veeneman Agilent Technologies, Inc. 8 Centerville Road

More information

Analytical Trap Comparison for USEPA Method 8260C

Analytical Trap Comparison for USEPA Method 8260C Analytical Trap Comparison for USEPA Method 8260C Application Note Abstract Purge and trap concentration is a technique that is used for the analysis of Volatile Organic Compounds (VOCs). The major component

More information

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES J Kouzaki 1*, S Sato 1, S Nakai 1, Y Shirasuna 2, K Hirano 2 1 Graduate School of Environmental and Information

More information

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Lee Marotta Tom Kwoka David Scott Miles Snow Toby Astill PerkinElmer, Inc. Shelton, CT Fast, Quantitative Analysis of Residual Solvents in

More information

Thermal Desorption Technical Support

Thermal Desorption Technical Support Thermal Desorption Technical Support Note 47: The Analysis of Landfill Gas Compounds using Thermal Desorption GC/MS and a Retention Time Locked Database Key Words: VOCs, directive, compliance, identifi

More information

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA 94804-3614 Ph. 510-236-2325; Fax 510-236-2335 E-mail info@berkeleyanalytical.com VOC Emissions from Building Products Customer & Building

More information

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in

More information

Test Report in accordance CDPH-IAQ

Test Report in accordance CDPH-IAQ Scan Underlay ApS Smedeskovvej 38 Dalgas Avenue 50 8464 Galten 8000 Aarhus C Denmark Denmark voc@eurofins.com www.eurofins.com/voc-testing Test Report in accordance CDPH-IAQ Date 6 January 2015 1 Sample

More information

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC Aysel T. ATIMTAY Middle East Technical University, Environmental Engineering Department, 06531 Ankara Turkey (aatimtay@metu.edu.tr)

More information

GC/MS: A Valid Tool for Soil Gas Hydrocarbons Speciation

GC/MS: A Valid Tool for Soil Gas Hydrocarbons Speciation International Journal of Analytical Mass Spectrometry and Chromatography, 2015, 3, 54-62 Published Online September 2015 in SciRes. http://www.scirp.org/journal/ijamsc http://dx.doi.org/10.4236/ijamsc.2015.33007

More information

U.S. EPA Method 8270 for multicomponent analyte determination

U.S. EPA Method 8270 for multicomponent analyte determination ENVIRONMENTAL application note U.S. EPA Method 8270 for multicomponent analyte determination Elaine A. LeMoine and Herman Hoberecht Introduction Multicomponent analytes are compounds that yield several

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Introduction Purge and Trap (P&T) concentrators

More information

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers Electronic Supplementary Material (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 2017 Electronic Supplementary Material Experimentally Validated Mathematical

More information

Test Report in accordance CDPH-IAQ

Test Report in accordance CDPH-IAQ Hilti Entwicklungsgesellschaft GmbH Hiltistraße 6 86916 Kaufering Germany Test Report in accordance CDPH-IAQ Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing

More information

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA 94804-3614 Ph. 510-236-2325; Fax 510-236-2335 E-mail info@berkeleyanalytical.com VOC Emissions from Building Products Customer & Building

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe Protox Fabriksvej 19 6000 Kolding DENMARK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 13 August 2015 Test Report- VOC emission regulations

More information

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No. G12871B_02

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No. G12871B_02 Polyseam A/S Ravneveien 7 Linnestad Næringsområde 3174 Revetal Norway Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 09 October 2014

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis Abstract Purge and Trap (P&T) is a concentration technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible

More information

Author. Abstract. Introduction

Author. Abstract. Introduction Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert Gas Chromatography/Mass Spectrometry System Application Author James D. McCurry Agilent

More information

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Validation of USEPA Method 524.2 Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Application Note By: Nathan Valentine Abstract The US EPA developed Method 524.2¹, Measurement

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe Protox Aps Fabriksvej 19 6000 Kolding Denmark Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 14 November 2014 Test Report- VOC emission

More information

The Application of Method TO-15 to Naphthalene Measurements in Indoor Air

The Application of Method TO-15 to Naphthalene Measurements in Indoor Air The Application of Method TO-15 to Naphthalene Measurements in Indoor Air Extended Abstract #13 Heidi C. Hayes and Diane J. Benton Air Toxics Ltd. 180 Blue Ravine Rd. Ste. B Folsom, CA 95630 INTRODUCTION

More information

Real Time On-Site Odor and VOC Emission Measurements Using a znose

Real Time On-Site Odor and VOC Emission Measurements Using a znose Real Time On-Site Odor and VOC Emission Measurements Using a znose Edward J. Staples, Electronic Sensor Technology, EST@ESTCAL.COM Remediation Site Description Remediation of contaminated soil from where

More information

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology Gas Chromatography/Mass Spectrometry Author Harry Prest 1601 California Avenue

More information

Odor Detection and Analysis using GC/SAW znose

Odor Detection and Analysis using GC/SAW znose Complimentary Copy Odor Detection and Analysis using GC/SAW znose Edward J. Staples Electronic Sensor Technology 1077 Business Center Circle Newbury Park, California 91320 Ph. 1-805-480-1994 FAX 1-805-480-1984

More information

DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES

DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES Mihriban Yılmaz Civan, Öznur Kuntasal and Gürdal Tuncel 1 1 OUTLINE Introduction Importance and

More information

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp

More information

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing

More information

Methanol Extraction of high level soil samples by USEPA Method 8260C

Methanol Extraction of high level soil samples by USEPA Method 8260C Methanol Extraction of high level soil samples by USEPA Method 8260C Abstract In order to determine the concentration of Volatile Organic Compounds (VOCs) in soil and waste samples the USEPA developed

More information

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS Agilent Technologies/Wasson ECE Monomer Analyzer Application Authors Fred Feyerherm 119 Forest Cove Dr. Kingwood, TX

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 12 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring Application Note Abstract The United States Environmental Agency (USEPA) has established the Unregulated

More information

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP Application Note Joseph M. Levy Michael Kraft Abstract Agilent Equipment 7890A

More information

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Introduction Fire investigation involves many different types of analyses from crime scene

More information

Characterising Volatile Organic Compounds from Sewer Emissions by Thermal Desorption coupled with Gas- Chromatography-Mass Spectrometry

Characterising Volatile Organic Compounds from Sewer Emissions by Thermal Desorption coupled with Gas- Chromatography-Mass Spectrometry A publication of CHEMICAL ENGINEERING TRANSACTIONS VOL. 30, 2012 Guest Editor: Renato Del Rosso Copyright 2012, AIDIC Servizi S.r.l., ISBN 978-88-95608-21-1; ISSN 1974-9791 The Italian Association of Chemical

More information

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek JSB is an authorised partner of Analysis of Low Level Volatile Organic s in Air Anne Jurek #111 Introduction The New Jersey Department of Environmental Protection (NJDEP) developed a new low level TO-15

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe CONICA AG Industriestr. 26 CH-8207 Schauffhausen Switzerland Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 31 July 2014 Test Report-

More information

VOC EMISSION TEST REPORT ISO 16000

VOC EMISSION TEST REPORT ISO 16000 Proxy A/S Gothersgade 12, 1.th 1123 København K DENMARK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark CustomerSupport@eurofins.com www.eurofins.com/voc-testing 1 Sample Information VOC

More information

CHEM 429 / 529 Chemical Separation Techniques

CHEM 429 / 529 Chemical Separation Techniques CHEM 429 / 529 Chemical Separation Techniques Robert E. Synovec, Professor Department of Chemistry University of Washington Lecture 1 Course Introduction Goal Chromatography and Related Techniques Obtain

More information

VOC TEST REPORT M1. 16 August Regulation or protocol Conclusion Version of regulation or protocol

VOC TEST REPORT M1. 16 August Regulation or protocol Conclusion Version of regulation or protocol SGS-Doors A/S Thrigesvej 8-16 7430 Ikast DENMARK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT M1 16 August 2016 1 Sample

More information

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No A_04

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No A_04 Sika Technology AG Tüffenwies 16 8048 Zürich SWITZERLAND EMICODE Test Report 1 Sample Information Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing

More information

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc. Optimizing Sampling and Analytical Parameters for Soil Vapour Samples using Automated Thermal Desorption / Gas Chromatography / Mass Spectrometry (ATD/GC/MS) Stephen Varisco, Technical Manager, CARO Analytical

More information

VOC TEST REPORT M1. 6 April Regulation or protocol Conclusion Version of regulation or protocol

VOC TEST REPORT M1. 6 April Regulation or protocol Conclusion Version of regulation or protocol Dana Lim A/S Københavnsvej 220 4600 Køge DENMARK Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT M1 6 April 2017 1 Sample

More information

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling JSB is an authorised partner of Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling Anne Jurek Introduction: #134 Static headspace sampling has always been a viable option

More information

VOC EMISSION TEST REPORT BREEAM NOR

VOC EMISSION TEST REPORT BREEAM NOR Hey di AS Tretjerndalsveien 68 Pb 13 2017 Frogner NORWAY Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark CustomerSupport@eurofins.com www.eurofins.com/voc-testing 1 Sample Information

More information

Solid Phase Micro Extraction of Flavor Compounds in Beer

Solid Phase Micro Extraction of Flavor Compounds in Beer Solid Phase Micro Extraction of Flavor s in Beer ANNE JUREK USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler Application Note Environmental

More information

VOC TEST REPORT CDPH

VOC TEST REPORT CDPH Baux AB Östermalmsgatan 26A 114 26 Stockholm SWEDEN Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT CDPH 29 September 2017

More information

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.)

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.) LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.) SOP-316 391 S. Vidal St., Sarnia, Ontario, N7T 7L1 Phone: (519) 344-4747 Fax: (519) 344-2350 E-Mail: info@lambtonscientific.com Method

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS PO-CON1609E Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS Pittcon 2016 1110-1 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole

More information

EVALUATION OF A PID-BASED PORTABLE DETECTION SYSTEM FOR TVOC AS AN INDOOR POLLUTION ANALYZER

EVALUATION OF A PID-BASED PORTABLE DETECTION SYSTEM FOR TVOC AS AN INDOOR POLLUTION ANALYZER EVALUATION OF A PID-BASED PORTABLE DETECTION SYSTEM FOR TVOC AS AN INDOOR POLLUTION ANALYZER KH Kim *, JD Lee 1, YJ Hong, YJ Choi Department of Earth & Environmental Sciences, Sejong University, Goon Ja

More information

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS PAGE: 1 of 18 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Emission measurements according to, M1 (3 appendices)

Emission measurements according to, M1 (3 appendices) issued by an Accredited Testing Laboratory Contact person Ulrika Johansson 2017-01-30 6F024225-1 1 (6) Chemistry, Materials and Surfaces +46 10 516 53 22 ulrika.johansson@sp.se Accred. No. 1002 Testing

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe L Isolante K-flex Srl Via Leonardi da Vinci 36 20877 Roncello (MB) Italy Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 29 August 2013

More information

Ed George and Anaïs Viven Varian, Inc.

Ed George and Anaïs Viven Varian, Inc. Application Note SI-02408 Evaluation of EPA Method 524.3: a New Draft Method for the Analysis of VOCs in Drinking Water using GC/MS and V:Results GC/MS Software Ed George and Anaïs Viven Varian, Inc. Introduction

More information

EPA TO-17 Volatile Organic Compounds

EPA TO-17 Volatile Organic Compounds EPA TO-17 Volatile Organic Compounds Method TO-17 is used to analyze samples for volatile organic compounds collected on multi-bed sorbent tubes, which are thermally desorbed and cryo-focused on the capillary

More information

METHOD 3600B CLEANUP

METHOD 3600B CLEANUP METHOD 3600B CLEANUP 1.0 SCOPE AND APPLICATION 1.1 Method 3600 provides general guidance on selection of cleanup methods that are appropriate for the target analytes of interest. Cleanup methods are applied

More information

TPH Methods and Measurements and Petroleum Vapor Intrusion (PVI) Risks

TPH Methods and Measurements and Petroleum Vapor Intrusion (PVI) Risks TPH Methods and Measurements and Petroleum Vapor Intrusion (PVI) Risks Robert Truesdale, RTI International Heidi Hayes, Eurofins Air Toxics Ian Hers, Golder Associates 1 RTI International is a trade name

More information

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler Validation of USEPA Method 524.2 Using a Stratum PTC and the New AQUATek 100 Autosampler Application Note Abstract Automation is the key to increasing laboratory productivity and minimizing costs. It is

More information

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Chapter 23. Gas Chromatography What did they eat in the year 1,000? GC of Cholesterol and other lipids extracted from

More information

Chemistry Gas Chromatography: Separation of Volatile Organics

Chemistry Gas Chromatography: Separation of Volatile Organics Chemistry 3200 Gas chromatography (GC) is an instrumental method for separating volatile compounds in a mixture. A small sample of the mixture is injected onto one end of a column housed in an oven. The

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions Validation of Volatile Organic Compound by USEPA Method 8260C Application Note Abstract In order to determine the concentration of volatile organic compounds (VOCs) in water and soil matrices the USEPA

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

A Single-Method Approach for the Analysis of Volatile and Semivolatile Organic Compounds in Air Using Thermal Desorption Coupled with GC MS

A Single-Method Approach for the Analysis of Volatile and Semivolatile Organic Compounds in Air Using Thermal Desorption Coupled with GC MS Electronically reprinted from Volume 32 Number 10 OCTOBER 2014 www.chromatographyonline.com A Single-Method Approach for the Analysis of Volatile and Semivolatile Organic Compounds in Air Using Thermal

More information

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3) Schedule Assignment Due Date Draft Section of Lab Report Monday 6pm (Jan 27) Quiz for Lab 2 Peer Review of Draft Complete Lab Report 1 Tuesday 9:30am Wednesday 6pm Friday 6pm Summary of Paper 2 Monday

More information

Experimental Determination of Pyrolysis Products from Carbon/Resin Ablative Materials

Experimental Determination of Pyrolysis Products from Carbon/Resin Ablative Materials Experimental Determination of Pyrolysis Products from Carbon/Resin Ablative Materials Hsi-Wu Wong, Jay Peck, and Robin Edwards Aerodyne Research, Inc., Billerica, MA Guillaume Reinisch University of Texas

More information

Application Note 059 Direct Desorption of Car Trim Materials for VOC and SVOC Analysis in Accordance with VDA Method 278

Application Note 059 Direct Desorption of Car Trim Materials for VOC and SVOC Analysis in Accordance with VDA Method 278 Application Note Direct Desorption of Car Trim Materials for VOC and SVOC Analysis in Accordance with VDA Method Application Note Abstract In this application note, we demonstrate the suitability of Markes

More information

Intertek Environmental VOC Emissions

Intertek Environmental VOC Emissions 4700 Broadmoor SE, Suite 200 Kentwood, MI 49512 Jesse.ondersma@intertek.com Telephone: 616-656-7401 Facsimile: 616-656-2022 www.intertek.com ELIZABETH FATIMA DO NASCIMENTO AIME P.O.: N.A. Page 1 of 8 Intertek

More information

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS HS-20 & GCMS-QP2010Ultra Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS Introduction Volatile Organic Compounds (VOCs) refer to a group

More information

Test Report- VOC emission regulations in Europe

Test Report- VOC emission regulations in Europe Scan Underlay ApS Dalgas Avenue 50 8000 Aarhus C Denmark Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voctesting Date 6 January 2015 Test Report VOC

More information

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES JPACSM 127 AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES Trace Analytical Inc. Menlo Park, CA ABSTRACT GC based gas analyzers with Reduction Gas Detector (RGD) and Flame Ionization

More information

TDTS. Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination

TDTS.   Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination Thermal Desorption Technical Support Note 75: Liquid standard injection, tube impedance and other factors which may cause discrimination during the calibration of thermal desorption methods Keywords TD

More information

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA 94804 Ph. 510-236-2325; Fax 510-236-2335 E-mail info@berkeleyanalytical.com VOC Emissions from Building Products Customer & Building Product

More information

ASTM D5116 Emissions Test Report with CDPH SM V Loading Scenarios per Cradle to Cradle Standard v3.0 Sect. 5.8

ASTM D5116 Emissions Test Report with CDPH SM V Loading Scenarios per Cradle to Cradle Standard v3.0 Sect. 5.8 ROMA USA LLC 554-B North Ave. NW Atlanta, GA 30318 Eurofins Product Testing NA 180 Blue Ravine Road, Suite B Folsom, CA 95630 voc@eurofinsus.com www.eurofinsuspt.com ASTM D5116 Emissions Test Report with

More information

Study of Residual Solvents in Various Matrices by Static Headspace

Study of Residual Solvents in Various Matrices by Static Headspace Application Note Abstract United States Pharmacopeia (USP) chapter is a widely used method for identifying and quantifying Organic Volatile Impurities (OVI) used in the production of pharmaceuticals.

More information

TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated

TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated Lee Marotta, Field Application Scientist, PerkinElmer Miles Snow, Research Scientist, PerkinElmer

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

Gas chromatography. Advantages of GC. Disadvantages of GC

Gas chromatography. Advantages of GC. Disadvantages of GC Advantages of GC Gas chromatography Fast analysis, typically minutes Effi cient, providing high resolution Sensitive, easily detecting ppm and often ppb Nondestructive, making possible on - line coupling;

More information