1 SOP Code: AP09 Issue Date: Revision No.: 04 Page 1 of 6 Crop and Environmental Sciences Division Standard Operating Procedure ph MEASUREMENT -AP09 This document is issued under the authority of MS. LILIA R. MOLINA Assistant Manager II THIS IS A CONTROLLED DOCUMENT

2 SOP Code: AP09 Issue Date: Revision No.: 04 Page 2 of 6 SIGNATURE PAGE The signatures below signify approval and concurrence of the signatories with the contents of this SOP coupled with the commitment to provide the resources necessary to ensure proper quality operation of the ASL laboratory. Issuing Authority: Name & Title Signature Date (yyyy-mm-dd) Lilia R. Molina Manager/QM Christy S. Daniel Deputy Quality Manager

3 SOP Code: AP09 Issue Date: Revision No.: 04 Page 3 of 6 -AP09 ph MEASUREMENT BY THE ELECTROMETRIC METHOD 1.0 Scope and application This SOP is designed primarily for the determination of soil ph, although the procedure applies equally well for the ph determination of any aqueous solutions or waters. ASL s default method is to use deionized water as the suspending medium in a 1:1 (soil:suspending medium) ratio unless specifically requested otherwise by the customer. Soil ph is measured to assess soil chemical properties, crop suitability, lime needs and relative nutrient availability. 2.0 Summary. Soil ph is measured electrometrically in a saturated paste or a supernatant liquid that is in equilibrium with a soil suspension of a 1:1 soil-to-liquid mixture. The most popular ph method is to use water or 0.01 M calcium chloride and a soil to liquid ratio of 1:2 w/v. ASL s default method is to use a 1:1 ratio and deionized water as the suspending liquid without stirring, unless requested otherwise to use an electrolyte (1 M KCl or 0.01 M CaCl 2 ) by the customer. Soils containing greater than 15% organic matter may require a 1:5 or 1:10 soil:water ratio for the determination of ph. The method is generally reproducible within ± 0.07 ph units (Kalra, 1995). 3.0 Apparatus and materials 3.1 ph meter preferably with automatic temperature compensation (Beckman Φ 310 ph meter or equivalent) 3.2 Numbered 125 ml polyethylene bottles, with a good fitting top 3.3 Dispenser bottle 3.4 Reciprocating shaker (Everbach Model Cat. No or equivalent) 3.5 Analytical balance, with a readability of ±0.01 g

4 SOP Code: AP09 Issue Date: Revision No.: 04 Page 4 of Reagents and standards 4.1 Standard buffer solutions (ph 4.0 ± 0.01, ph 7.00 ± 0.01 and ph ± 0.01) 4.2 Verification standard buffer of known ph obtained from a different manufacturer than M Potassium chloride solution Dissolve g potassium chloride (AR Grade) in ~200 ml deionized water in a 1000 ml volumetric flask. When dissolved, make to volume with deionized water and shake well to mix M Calcium chloride solution Dissolve 1.11 g of calcium chloride (AR Grade) in about 400 ml deionized water in a 1000 ml volumetric flask. When dissolved, make to volume with deionized water and shake well to mix M KCl solution (storage and fill solution for the electrode). 5.0 Sample preparation, preservation and storage 5.1 ASL does not prepare samples. 5.2 The customers are the one responsible for preparation of their samples and may use either the procedures prescribed by ASL in QM-AD036-MP24, or their own procedures. But the customer is required to indicate the method of sample preparation when they submit the sample charge slip to ASL. The soil sample must be air dried and sieved to 2 mm mesh size. 5.3 Samples submitted by customers in brown envelope are kept in the soils storage room until taken for analysis (no preservative is added). 6.0 Procedure 6.1 Weigh out 10 g of sample (air dried, passed in a 2 mm mesh sieve) and place in a clean 125 ml polyethylene bottle. 6.2 Add 10 g or 10 ml of the suspending liquid as specified by the customer (whether deionized water, 1M KCl or 0.01 M CaCl 2 ).

5 SOP Code: AP09 Issue Date: Revision No.: 04 Page 5 of Place all of the bottles in a reciprocating shaker and shake for 30 min. at high speed ( excursions per minute). 6.4 Remove from the shaker and allow soil particles to settle for 5 minutes but not more than 30 minutes before ph measurement. 6.5 Measure the ph of the sample in a previously calibrated ph meter (calibrate ph meter using standard buffers of ph 4, 7 and 10). Result is expressed as ph up to the first decimal point. 6.6 Rinse the electrodes thoroughly in-between measurements with deionized water making sure there are no adhering particles on the electrode bulb before making further measurements. 6.7 When all measurements are done, rinse the electrodes thoroughly before storing them in 3M KCl solution. 7.0 Calculations and reporting 7.1 The ph reading is reported in not more than one decimal point. 7.2 The extracting liquid is also indicated in the analysis report. 8.0 Quality control, acceptance criteria Each analytical batch of samples should include measurement of an independent buffer solution of known ph, in-house reference material, sample repeats and a buffer for continuing calibration verification. 9.0 References 9.1 Caroll, S.D., Herron, C.G., Lafex, D.C. Determination of Soil ph. University of Arkansas Soil Testing and Research. 9.2 FAO and ISRIC Guidelines for quality management in soil and plant laboratories. FAO Soils Bulleting Jackson, M.L. Soil Chemical Analysis Prentice Hall, Inc. Inglewood Cliffs. N.J. p 46.

6 SOP Code: AP09 Issue Date: Revision No.: 04 Page 6 of Kalra, Y.P.; Maynard, D.G Methods manual for forest soil and plant analysis. For. Can., Northwest Reg., North For. Cent., Edmonton, Alberta, Canada. Inf. Rep. NOR-X Thomas GW Soil ph and soil acidity. In: Sparks DL et al, editors. Methods of soil analysis: Part 3 Chemical methods. SSSA Book Series No. 5. Madison, Wisconsin (USA): Soil Science Society of America. pp (END OF DOCUMENT)

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge Standard Operating Procedure for GLNPO Specific Conductance: Marvin Palmer United States Environmental Protection Agency Great Lakes National Program Office Region 5 77 West Jackson 9th Floor Chicago,

More information

QAM-I-111 Operation and Calibration of the Conductivity Meter

QAM-I-111 Operation and Calibration of the Conductivity Meter 1.0 Applicability and Purpose i. This procedure applies to the operation and calibration of the YSI Model 3200 conductivity meter. This procedure is performed prior to any analysis using this meter. By

More information

SOP-C-120 Determination of ph

SOP-C-120 Determination of ph i. Identification of the method a. SM 4500-H + (approved 2011) and EPA 9045C (approved 1997) ii. Applicable matrix or matrices a. Waters or aqueous solutions and suspensions iii. Limits of detection and

More information

Procedure for the Determination of Permanganate Oxidizable Carbon

Procedure for the Determination of Permanganate Oxidizable Carbon Procedure for the Determination of Permanganate Oxidizable Carbon Overview: Steve Culman, Mark Freeman, Sieglinde Snapp Kellogg Biological Station, Michigan State University, Hickory Corners, MI, 49060

More information

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required

More information

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA , 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine

More information

ISO INTERNATIONAL STANDARD. Soil quality Determination of ph. Qualité du sol Détermination du ph. Second edition

ISO INTERNATIONAL STANDARD. Soil quality Determination of ph. Qualité du sol Détermination du ph. Second edition INTERNATIONAL STANDARD ISO 10390 Second edition 2005-02-15 Soil quality Determination of ph Qualité du sol Détermination du ph Reference number ISO 10390:2005(E) ISO 2005 ISO 10390:2005(E) Provläsningsexemplar

More information

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow 1. Application Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow Parameters and Codes: Nitrogen, ammonia, dissolved, I-2522-90 (mg/l as N): 00608 Nitrogen, ammonia, total-in-bottom-material,

More information

SOIL SURVEY STANDARD TEST METHOD PARTICLE SIZE ANALYSIS

SOIL SURVEY STANDARD TEST METHOD PARTICLE SIZE ANALYSIS Department of Sustainable Natural Resources SOIL SURVEY STANDARD TEST METHOD PARTICLE SIZE ANALYSIS ABBREVIATED NAME PSA TEST NUMBER P7 TEST METHOD TYPE B VERSION NUMBER 3 RECORD OF AMENDMENTS Version

More information

CHEMISTRY DEPARTMENT

CHEMISTRY DEPARTMENT POINT LEPREAU GENERATING STATION GADOLINIUM ARSENAZO III Issued By: Date: APPROVED BY: of 6 REVISION RECORD AUTHOR REV DESCRIPTION DATE 0 Initial Issue W. Mawhinney Not Tracked Oct./9 W. T. Underhill 2

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 9 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Tex-620-J, Determining Chloride and Sulfate Contents in Soil Contents in Soil Contents: Section 1 Overview...2 Section 2 Sample Preparation...3 Section 3 Ion Chromatography Method...5 Section 4 Wet Chemical Method...9 Section 5 Archived Versions...15 Texas Department

More information

DR/4000 PROCEDURE NITRATE. 2. The display will show:

DR/4000 PROCEDURE NITRATE. 2. The display will show: Method 8192 Powder Pillows DR/4000 PROCEDURE Cadmium Reduction Method LR (0 to 0.50 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number

More information

PRACTICAL 3 ph AND BUFFERS

PRACTICAL 3 ph AND BUFFERS PRACTICAL 3 ph AND BUFFERS ph and Buffers Structure 3.1 Introduction 3.2 ph and Buffers: Basic Concept 3.2.1 ph 3.2.2 Buffers and Buffer Solutions 3.3 Methods for Determining ph Experiment 1: Measurement

More information

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 , MR, 8171 DOC316.53.01069 Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 N) Powder Pillows or AccuVac Ampuls Scope and Application: For water, wastewater and seawater Test preparation

More information

Acid Deposition and Oxidant Research Center (ADORC)

Acid Deposition and Oxidant Research Center (ADORC) Standard Operating Procedures for H measurement Acid Deposition and Oxidant Research Center (ADORC) Introduction - What is SOPs? - Objectives of SOPs - Major Items to be included in the SOPs in wet deposition

More information

LACTIC ACID. The method is applicable to the determination of lactic acid and lactate salts (Note 2) in light or heavy steepwater.

LACTIC ACID. The method is applicable to the determination of lactic acid and lactate salts (Note 2) in light or heavy steepwater. LACTI.01-1 LACTIC ACID PRINCIPLE SCOPE Lactic acid in steepwater (Note 1) is oxidized to acetaldehyde following treatment with copper sulfate and calcium hydroxide to remove interfering substances. Acetaldehyde

More information

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University. Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared

More information

Mehlich and Modified Mehlich Buffer for Lime Requirement

Mehlich and Modified Mehlich Buffer for Lime Requirement Mehlich and Modified Mehlich Buffer for Lime Requirement David H. Hardy Application and Principle The development of the Mehlich buffer was to estimate acidity actually affecting plant growth in North

More information

For Research Use Only

For Research Use Only L-Ascorbate Assay Kit Kit-0873 Lot. No. (See product label) Specification Description Size Kit Components Materials Required but Not Supplied Preparation L-Ascorbic acid (Vitamin C), an anti-oxidant and

More information

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. Using Powder Pillows DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

SOIL ORGANIC CONTENT USING UV-VIS METHOD

SOIL ORGANIC CONTENT USING UV-VIS METHOD Test Procedure for SOIL ORGANIC CONTENT USING UV-VIS METHOD TxDOT Designation: Tex-148-E Effective Date: March 2016 1. SCOPE 1.1 This method determines the soil organic content based on the amount of humic

More information

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT METHOD 9040B ph ELECTROMETRIC MEASUREMENT 1.0 SCOPE AND APPLICATION 1.1 Method 9040 is used to measure the ph of aqueous wastes and those multiphase wastes where the aqueous phase constitutes at least

More information

Citric Acid Analysis L-6-1. MOISTURE (Karl Fischer)

Citric Acid Analysis L-6-1. MOISTURE (Karl Fischer) Citric Acid Analysis L-6-1 MOISTURE (Karl Fischer) PRINCIPLE SCOPE The sample is dissolved in a mixture of methanol and formamide (50:50 v/v) and then titrated with standardized Karl Fischer reagent. The

More information

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater

More information

DR/4000 PROCEDURE. IRON, Total

DR/4000 PROCEDURE. IRON, Total Method 8365 Powder Pillows DR/4000 PROCEDURE FerroMo Method* (0 to 1.800 mg/l) Scope and Application: For cooling water containing molybdate-based treatment; digestion is required for determining total

More information

4 What Happens When Compounds Are Added to Water?

4 What Happens When Compounds Are Added to Water? 4 What Happens When Compounds Are Added to Water? Name: Date: Lab Day/Time: Lab Partner: Objectives Understand the difference between electrolyte and nonelectrolyte solutions Learn to identify a compound

More information

maect PRODUCTION SOP M/1 JUNE 2007 version page: 1 Preparation of 2 liters of 10x concentrated PBS stock solution (phosphate buffered saline)

maect PRODUCTION SOP M/1 JUNE 2007 version page: 1 Preparation of 2 liters of 10x concentrated PBS stock solution (phosphate buffered saline) maect PRODUCTION SOP M/1 JUNE 2007 version page: 1 Preparation of 2 liters of 10x concentrated PBS stock solution (phosphate buffered saline) Intended use: Storage: Safety: After 1:10 dilution of the PBS

More information

Experiment 2: Analysis of Commercial Bleach Solutions

Experiment 2: Analysis of Commercial Bleach Solutions Experiment 2: Analysis of Commercial Bleach Solutions I. Introduction The ability of household bleach to remove stains is related to the amount of oxidizing agent in it. The oxidizing agent in bleach is

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Fluoroborate Ion Selective Electrode is used to quickly, simply, accurately, and economically measure Fluoroborate in aqueous solutions.

More information

EXPERIMENT 6. Properties of Buffers INTRODUCTION

EXPERIMENT 6. Properties of Buffers INTRODUCTION EXPERIMENT 6 Properties of Buffers INTRODUCTION A chemical buffer is any substance in a solution that tends to stabilize the hydronium ion concentration by neutralizing any added acid or base. Buffers

More information

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

Corn Sugar (crude & refined) Analysis F MOISTURE (Karl Fischer)

Corn Sugar (crude & refined) Analysis F MOISTURE (Karl Fischer) Corn Sugar (crude & refined) Analysis F-32-1 MOISTURE (Karl Fischer) PRINCIPLE SCOPE The sugar sample is dissolved in a mixture of methanol and formamide (50:50 v/v) and then titrated with standardized

More information

STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST

STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST Questa Rock Pile Stability Study SOP 52v6 Page 1 STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST REVISION LOG Revision Number Description Date 52v0 Original SOP by STM 6/9/2004 52v1

More information

Potentiometric Determination of the pka and the Equivalent Weight of a Weak Acid

Potentiometric Determination of the pka and the Equivalent Weight of a Weak Acid Potentiometric Determination of the pka and the Equivalent Weight of a Weak Acid Introduction In this experiment you will use a ph meter to follow changes in H + concentration during a titration. The laboratory

More information

EXPERIMENT C3: SOLUBILITY PRODUCT & COMMON ION EFFECT. Learning Outcomes. Introduction. Upon completion of this lab, the student will be able to:

EXPERIMENT C3: SOLUBILITY PRODUCT & COMMON ION EFFECT. Learning Outcomes. Introduction. Upon completion of this lab, the student will be able to: 1 EXPERIMENT C3: SOLUBILITY PRODUCT & COMMON ION EFFECT Learning Outcomes Upon completion of this lab, the student will be able to: 1) Measure the solubility product constant for a sparingly soluble salt.

More information

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7 Troy, NY 12180 Date: 03/14/2012 Page 1 of 7 USGS New York Water Science Center Soil and Low Ionic Strength Water Quality Laboratory Electrometric Measurements and Titration Standard Operating Procedure

More information

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Ammonium Ion Electrode is used to quickly, simply, accurately, and economically

More information

(b) Write the chemical equation for the dissolution of NaOH in water.

(b) Write the chemical equation for the dissolution of NaOH in water. Making a Solution and Measuring ph Prelab Assignment Read the entire lab. Write an objective and any hazards associated with this lab in your laboratory notebook. Answer the following 6 questions in your

More information

H 3 O + (aq) + P 2- (aq)

H 3 O + (aq) + P 2- (aq) PURPOSE: To standardize a solution of sodium hydroxide by titration with a primary standard, (KHC 8 H 4 O 4 ), potassium hydrogen phthalate (KHC 8 H 4 O 4 ) PRINCIPLES: Most shelf reagents, such as 0.10

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

LABORATORY EXERCISE: USING SPECTROPHOTOMETRY FOR QUALITY CONTROL: NIACIN

LABORATORY EXERCISE: USING SPECTROPHOTOMETRY FOR QUALITY CONTROL: NIACIN SURVEY OF QUALITY REGULATIONS AND STANDARDS LABORATORY EXERCISE: USING SPECTROPHOTOMETRY FOR QUALITY CONTROL: NIACIN Submitted by Madison Area Technical College Contact Person: Lisa Seidman, Lseidman@matcmadison.edu

More information

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

GETTING THE END POINT TO APPROXIMATE. Two hours

GETTING THE END POINT TO APPROXIMATE. Two hours Chem 1312 Handout Experiment ONE Laboratory Time Required Special Equipment and Supplies Objective Safety First Aid GETTING THE END POINT TO APPROXIMATE THE EQUIVALENCE POINT Two hours Balance Potassium

More information

National standard of People s Republic of China

National standard of People s Republic of China National standard of People s Republic of China GB5413.24-2010 Determination of chlorine in foods for infants and young children, raw milk and dairy products Issued at 2010-03-2 Implemented at:2010-06-01

More information

STANDARD OPERATING PROCEDURE

STANDARD OPERATING PROCEDURE Ref.. Table of Contents Page. Scope 2 2. Objective 2 3. Definitions 2 4. Responsibility 3 5. Procedure 3 5.. Diagram of Display and Control Panel 3 5.2. Installation 4 5.3. Operation 4 5.4. Calibration

More information

Chapter 9. Volumetric Analysis

Chapter 9. Volumetric Analysis Chapter 9 Volumetric Analysis The terms volumetric analysis, titrimetry and titration are used interchangeably to describe a procedure which analyses chemicals in solution by accurate volume measurement.

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Total Mercury Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-106-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94.

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94. Ethanol Project Chemical Analysis and Testing Task Laboratory Analytical Procedure Procedure Title: LAP-013 HPLC Analysis of Liquid Fractions of Process Samples for Monomeric Sugars and Cellobiose Author:

More information

P E R O X Y G E N S A sector group of Cefic

P E R O X Y G E N S A sector group of Cefic P E R O X Y G E N S A sector group of Cefic CEFIC PEROXYGENS H 2 O 2 AM-7160 Hydrogen Peroxide for industrial use Determination of apparent ph (pha) Potentiometric method March 2003 Notice This document

More information

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description WHO/SIT/19.R4 TEMEPHOS TECHNICAL TECHNICAL TEMEPHOS 1. Specification 1.1 Description Full specification WHO/SIT/19.R4 Revised 10 December 1999 The material shall consist of temephos together with related

More information

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,: sop-c-102 Determination of Chemical Oxygen Demand Revision 6 Approval: Laborato ry Manager/LQAO/RS O Date Z/zs Date Z-z* I A Renewat date: 7lt:,: Effective date: 3, tnitiats: wu\ Texas lnstitute for Applied

More information

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution , drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific

More information

Determination of Orthophosphate Ion

Determination of Orthophosphate Ion Determination of Orthophosphate Ion Introduction Phosphorous, in the form of phosphate, is one of several important elements in the growth of plants. Excessive algae growth in water is stimulated by the

More information

Chemistry Determination of Mixed Acids

Chemistry Determination of Mixed Acids Chemistry 3200 Acid-base titration is one of the most common operations in analytical chemistry. A solution containing an unknown amount of ionizable hydrogen can be titrated with a solution of standard

More information

Chapter 12 Tex-617, Determining Chloride in Concrete

Chapter 12 Tex-617, Determining Chloride in Concrete Chapter 12 Tex-617, Determining Chloride in Contents: Section 1 Overview... 12-2 Section 2 Apparatus... 12-3 Section 3 Preparing Solutions... 12-4 Section 4 Procedures... 12-5 Section 5 Calculations...

More information

Solubility of KHT and Common ion Effect

Solubility of KHT and Common ion Effect Solubility of KHT and Common ion Effect v010516 You are encouraged to carefully read the following sections in Tro (3 rd ed.) to prepare for this experiment: Sec 16.5, pp 783-788 (Solubility Equilibria

More information

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy EXPERIMENT 7 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium

More information

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) 683 METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) NOTE: This method does not include all of the specifications (e.g., equipment and

More information

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan APPENDIX 7 PH MEASURING 1. General information PH probe is a unit, measuring the solution acidity or alkalinity degree. It is measured on scale of 0 to 14. The term ph is derived from "p", the mathematical

More information

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions. Cole-Parmer Ammonium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium

More information

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Method 8171 NITRATE, Mid Range (0 to 5.0 mg/l NO 3- -N) For water, wastewater and seawater* Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Using Powder Pillows 1. Enter the stored program

More information

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows FILL LINE Method 8155 Salicylate Method 1 Powder Pillows Scope and Application: For water, wastewater, and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) (0.01 to 0.50 mg/l NH 3 N) Test Preparation

More information

NITRITE, Low Range (0 to mg/l NO 2- -N)

NITRITE, Low Range (0 to mg/l NO 2- -N) NITRITE, Low Range (0 to 0.350 mg/l NO 2- -N) Method 8507 For water, wastewater, seawater Diazotization Method * (Powder Pillows or AccuVac Ampuls); USEPA approved for reporting wastewater and drinking

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 7 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Textiles - Determination of ph of Aqueous Extract

Textiles - Determination of ph of Aqueous Extract Translated English of Chinese Standard: GB/T7573-2009 www.chinesestandard.net Sales@ChineseStandard.net ICS 59. 080. 01 W 04 NATIONAL STANDARD GB OF THE PEOPLE'S REPUBLIC OF CHINA GB/T 7573-2009 Replacing

More information

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml) GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Silver/Sulfide Ion Electrode is used to measure silver or sulfide ions in aqueous solutions. The two ions are virtually never present in solution

More information

Graphene oxide was synthesized from graphite using the MH (modified Hummer s method) 30 and

Graphene oxide was synthesized from graphite using the MH (modified Hummer s method) 30 and Supplemental Information Synthesis of Graphene Oxide from Graphite Graphene oxide was synthesized from graphite using the MH (modified Hummer s method) 30 and the Tour methods 31. For the MH method, SP-1

More information

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater. Nitrogen, Ammonia DOC316.53.01078 USEPA 1 Nessler Method 2 Method 8038 0.02 to 2.50 mg/l NH 3 N Reagent Solution Scope and application: For water, wastewater and seawater. Distillation is required for

More information

Experiment 8 Introduction to Volumetric Techniques I. Objectives

Experiment 8 Introduction to Volumetric Techniques I. Objectives Experiment 8 Introduction to Volumetric Techniques I Objectives 1. To learn the proper technique to use a volumetric pipette. 2. To learn the proper technique to use a volumetric flask. 3. To prepare a

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Fluorides Former numbering: ECSS/CN 311-1982 & ESPA/CN-E-110-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7 Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes

More information

Application Determination of isocyanate (NCO-) content

Application Determination of isocyanate (NCO-) content Determination of isocyanate (NCO-) content date: 20.06.2013 page 1 from 10 Use This method is applicable for material containing reactive isocyanate groups.the isocyanate is reacted with di-n-butyl amine

More information

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

Experiment 8 and 9 Weak Acids and Bases: Exploring the Nature of Buffers

Experiment 8 and 9 Weak Acids and Bases: Exploring the Nature of Buffers Experiment 8 and 9 Weak Acids and Bases: Exploring the Nature of Buffers Pre-Laboratory Assignments Reading: Textbook Chapter 16 Chapter 17:1-3 This Laboratory Handout Pre-Laboratory Assignments: Complete

More information

Buffer Preparation. Learning Objectives:

Buffer Preparation. Learning Objectives: Proteomics Buffer Preparation Buffer Preparation Maintaining the optimum ph during the biological sample processing is to maintain the proper functional and structural aspects of the sample. It is important

More information

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Calcium Ion Electrodes are used to quickly, simply, accurately, and economically

More information

Density of Aqueous Sodium Chloride Solutions

Density of Aqueous Sodium Chloride Solutions Experiment 3 Density of Aqueous Sodium Chloride Solutions Prepared by Ross S. Nord and Stephen E. Schullery, Eastern Michigan University PURPOSE Determine the concentration of an unknown sodium chloride

More information

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

PHS-25CW/3BW Benchtop ph Meter Instruction Manual PHS-25CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides

More information

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water. Method 8337 Powder Pillows DR/4000 TT PROCEDURE QUATERNARY AMMONIUM COMPOUNDS Direct Binary Complex Method (0 to 5.00 mg/l as CTAB) Scope and Application: For cooling tower and pool/spa water. The estimated

More information

Carbonate content. SCAN-N 32:98 Revised White, green and black liquors and burnt lime sludge

Carbonate content. SCAN-N 32:98 Revised White, green and black liquors and burnt lime sludge Revised 1998 White, green and black liquors and burnt lime sludge Carbonate content 0 Introduction This SCAN-test Method replaces SCAN-N 32:88 from which it differs in that it, in addition to white and

More information

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049 imagine explore learn Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049 Overview Fluoride is often added to drinking water and to toothpaste to prevent dental decay. In high concentrations,

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Arsenic Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-105-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

[Agilent Technologies] [Electrochemistry Products] Quick Reference

[Agilent Technologies] [Electrochemistry Products] Quick Reference Overview Using I9121 Ammonia Gas Sensing Probe on 3200I/M This document demonstrates how to use I 9121 Ammonia Gas Sensing Probe on 3200I or 3200 M to measure pn or the concentration of [NH 4 + ]. Applied

More information

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis Technical Procedure for Solid Phase Extraction of THC and THC-COOH 1.0 Purpose - This procedure specifies the required elements for the extraction of THC and THC-COOH using United Technologies Styre Screen

More information

Experiment 7 Buffer Capacity & Buffer Preparation

Experiment 7 Buffer Capacity & Buffer Preparation Chem 1B Dr. White 57 Experiment 7 Buffer Capacity & Buffer Preparation Objectives To learn how to choose a suitable conjugate acid- base pair for making a buffer of a given ph To gain experience in using

More information

EXPERIMENT 22 SOLUBILITY OF A SLIGHTLY SOLUBLE ELECTROLYTE

EXPERIMENT 22 SOLUBILITY OF A SLIGHTLY SOLUBLE ELECTROLYTE EXPERIMENT 22 SOLUBILITY OF A SLIGHTLY SOLUBLE ELECTROLYTE INTRODUCTION Electrolytes are compounds that are present in solution as ions. They are more likely to be soluble in water than in most other liquids

More information

Supernatant: The liquid layer lying above the solid layer after a precipitation reaction occurs.

Supernatant: The liquid layer lying above the solid layer after a precipitation reaction occurs. Limiting Reagent Introduction The quantities of substances involved in a chemical reaction represented by a balanced equation are often referred to as stoichiometric amounts. Solution stoichiometry is

More information

Determination of Orthophosphate Ion

Determination of Orthophosphate Ion Determination of Orthophosphate Ion Introduction Phosphorous, in the form of phosphate, is one of several important elements in the growth of plants. Excessive algae growth in water is stimulated by the

More information

Quaternary Ammonium Compounds

Quaternary Ammonium Compounds , 8337 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB Powder Pillows Scope and Application: For cooling tower water and pool/spa water Test

More information

Determination of the Equivalent Weight and Ionization Constant of a Weak Acid

Determination of the Equivalent Weight and Ionization Constant of a Weak Acid Determination of the Equivalent Weight and Ionization Constant of a Weak Acid Introduction: The object of this experiment will be to determine the ionization constant, K a, and the equivalent weight of

More information

SUMMARY OF ASL TEST METHODS

SUMMARY OF ASL TEST METHODS SUMMARY OF ASL TEST METHODS 1. Determination of Macro and Micro Elements in Plant Tissue by Dual View Inductively Coupled Plasma Optical Emission Spectrometer This method describes multi-elemental determinations

More information

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8032 is used to determine trace amounts of acrylamide monomer (CAS No. 79-06-1) in aqueous matrices. This method may be

More information

INTERNATIONAL CHEMISTRY TESTING. 258 Main Street- Suite 311- Milford, MA Tel: Fax:

INTERNATIONAL CHEMISTRY TESTING. 258 Main Street- Suite 311- Milford, MA Tel: Fax: SOP TITLE: DETERMINATION OF TOTAL PROANTHOCYANIDINS BY REACTION WITH DIMETHYLAMINOCINNAMALDEHYDE (DMAC) SOP NUMBER: 803-02 EFFECTIVE DATE: 03/01/14 SUPERSEDES: 803-01 NEW ISSUE DATE: 3/24/20 INTRODUCTION

More information

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856 Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information