Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS
|
|
- Beverley Linda Fisher
- 6 years ago
- Views:
Transcription
1 Application Note Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS Objective By: Glynda Smith In this paper, an evaluation of different methods of moisture control will be presented and the Tekmar-Dohrmann 3 Sample Concentrator will be optimized for the analysis of the USEPA compound list using GC/MS. This optimization of method parameters will focus on achieving the optimal chromatographic response by reducing the effect of water on the analysis. Introduction Purge and Trap coupled with gas chromatography offers a simple method for the extraction and concentration of volatile organic compounds (VOCs) from multiple matrices such as waters and soils. USEPA Method monitors a wide range of VOCs in drinking water. Due to the range and number of compounds analyzed in Method 524.2, there are several analytical problems that can arise. One of the most common problems involves the transfer of water to the analytical column. Excessive water transfer can cause problems with both the mass spectrometer and the chromatography. A mass spectrometer is especially sensitive to the presence of water. Various types of vacuum pumps may be incorporated (e.g., roughing pumps, turbomolecular pumps, diffusion pumps, etc.) and the pumping efficiencies can vary with time and with the level of maintenance. The pressure load that water imposes upon a mass spectrometer interferes with the efficiency of electron ionization, shortens the lifetime of the filaments, and results in higher maintenance costs. With respect to the chromatography, excess water often produces unacceptable broadening of early eluting peaks. Removal of water can be facilitated through the use of hydrophobic trapping materials and incorporation of a moisture control system (MCS) which condenses water from the desorb gas during the transfer to the gas chromatograph. The research described in this paper involves the evaluation of the moisture control system and two types of traps under varying conditions of sample mount temperatures, MCS temperatures, dry purge times, and desorb times. Experimental Method Rev. 4 standards were obtained from Restek Corp. (VOA Kit - Cat #3447). Duplicate data runs employed 5 ml samples containing ppb of each of the analytes. The parameters evaluated are as follows: I. Vocarb 3 Hydrophobic Trap A. Moisture Control System (MCS) Installed. Vary 2. Vary Dry Purge 3. Vary Mount Temperature 4. Vary Desorb B. MCS Uninstalled. Vary Dry Purge 2. Vary Mount Temperature 3. Vary Desorb II. #3 (Tenax/Charcoal/Silica Gel) Hydrophilic Trap A. Moisture Control System (MCS) Installed. Vary 2. Vary Mount Temperature 3. Vary Desorb B. MCS Uninstalled. Vary Mount Temperature 2. Vary Desorb 3-.doc; 9-Jun-3 Sales/Support: Main: Socialville Foster Rd., Mason, OH 454
2 The Tekmar-Dohrmann 3 Purge & Trap Sample Concentrator parameters that were held constant throughout all of the data runs are described below: Line Temperature 5 C Trap Pressure Control 4 psi Valve Temperature 5 C Bake 2 min. Purge Ready Temp. 35 C Bake Temperature 27 C (Vocarb);23 C () Purge min. Desorb Temperature 25 C (); 225 C () Desorb Preheat 245 C (); 22 C () Table. Conditions for HP 689GC/5973 MS Injector: Column: Temperature Program: Carrier: C, Split 2: with Vocarb trap; Split 5: with #3 trap Restek Corp. Rtx-VMS, 6m x.25mm x.4µm 45 C (hold min). Increase to 9 C at 2 C/min and hold 2 min. at 9 C. Increase to 225 C at 6 C/min and hold min. at 225 C. Helium,.2 ml/min MS Source Temperature: 23 C MS Quad Temperature: 5 C Electron Multiplier: Mass Range Scanned: Vacuum System: 62 volts amu Turbomolecular pump, x -5 Torr Splitting at the injection port is a good technique for water removal. A 5: split was needed for samples that were run on the hydrophilic #3 trap in order to fully resolve the light gases. When lower split ratios were examined, the higher baselines and broadened peak shapes of the early eluting components resulted in wide variability for the calculated response factors of the light gases. The Vocarb 3 trap, however, is very effective at removing water due to its hydrophobicity. As such, a lower 2: split ratio provided sufficient chromatographic resolution for samples prepared with the Vocarb 3 trap. Results and Discussion Tables 2 through 8 list the response factors obtained for a group of representative compounds under the conditions described in Table. All six of the gas standards are included (dichlorodifluoromethane, chloromethane, vinyl chloride, bromomethane, trichlorofluoromethane) along with three mid-boiling range compounds (methyl acrylate, toluene, pentachloroethane) and three high boiling range compounds (,2,4- trichlorobenzene, naphthalene,,2,3-trichlorobenzene). Vary Table 2.. Vary during Desorb Vocarb 3 Trap Analyte 4 C 5 C 6 C Analyte 4 C 5 C 6 C Dichlorodifluoromethane Dichlorodifluoromethane Chloromethane Chloromethane Vinyl Chloride Vinyl Chloride Bromomethane Bromomethane Chloroethane Chloroethane Trichlorofluoromethane Trichlorofluoromethane Methyl Acrylate Methyl Acrylate Toluene Toluene Bromoform Bromoform Pentachloroethane Pentachloroethane ,2,4-Trichlorobenzene ,2,4-Trichlorobenzene Naphthalene Naphthalene ,2,3-Trichlorobenzene ,2,3-Trichlorobenzene Table 2 shows the results obtained when the MCS temperature was increased during the desorb cycle. The MCS is typically cooled to ambient temperature during the desorb step in order to condense water prior to transfer of 3-.doc; 9-Jun-3 Page 2 of 7
3 the analytes from the trap to the GC. As the data indicates, the overall responses are higher with the Vocarb 3 trap relative to the #3 trap. Specifically, the average responses of the lighter components are about 8% better with the Vocarb trap. This is not surprising since the early eluting components are adversely affected by the presence of water. But for either trap, there is little variation in the analyte responses as a function of MCS temperature and there seems to be no analytical advantage in operating the system at elevated MCS temperatures. Graph provides a brief synopsis of this information showing the effect on two of the gas standards, a mid-boiling range compound and a high boiling range compound. Graph Vary - Dichlorodifluoromethane 5 Vary - Chloroethane Vary - Toluene Vary - Naphthalene Vary Dry Purge Hydrophobic traps such as the Vocarb 3 trap can be purged with dry gas prior to desorption of the sample. This step facilitates water removal. Hydrophilic traps, however, do not benefit from incorporation of a dry purge since water will preferentially remain in the silica gel. Thus, the data in tables 3 and 4 only refer to the Vocarb 3 trap. Table 3.. Vary Dry Purge (Applies to Vocarb 3 Trap Only) Vocarb 3 Trap Analyte min. 2 min. 4 min. 6 min. 8 min. Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane Methyl Acrylate Toluene Bromoform Pentachloroethane ,2,4-Trichlorobenzene Naphthalene ,2,3-Trichlorobenzene doc; 9-Jun-3 Page 3 of 7
4 Table 4.. Vary Dry Purge Vocarb 3 Trap Analyte min. 2 min. 4 min. 6 min. 8 min. Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane Methyl Acrylate Toluene Bromoform Pentachloroethane ,2,4-Trichlorobenzene Naphthalene ,2,3-Trichlorobenzene When a moisture control system is installed along with a hydrophobic trap such as the Vocarb 3, a dry purge step may or may not be beneficial and the length of time employed should be determined by the application. In the work described in this paper, the split ratios employed minimized the dry purge effect. When the MCS is bypassed inside the valve oven with a short piece of /6 Silcosteel tubing, however, a long dry purge step is needed in order to attain the same analyte responses as when the MCS is incorporated in the sample pathway (see Graph 2). Graph 2. Vary Dry Purge - Dichlorodifluoromethane Vary Dry Purge - Vinyl Chloride Vary Dry Purge - Bromoform Vary Dry Purge - Naphthalene doc; 9-Jun-3 Page 4 of 7
5 Vary Mount Temperature Tables 5 and 6 compare the results obtained at different mount temperatures when the MCS is installed and removed, respectively. Table 5.. Vary Mount Temperature Vocarb 3 Trap Analyte Ambient 75 C C Analyte Ambient 75 C C Dichlorodifluoromethane Dichlorodifluoromethane Chloromethane Chloromethane Vinyl Chloride Vinyl Chloride Bromomethane Bromomethane Chloroethane Chloroethane Trichlorofluoromethane Trichlorofluoromethane Methyl Acrylate Methyl Acrylate Toluene Toluene Bromoform Bromoform Pentachloroethane Pentachloroethane ,2,4-Trichlorobenzene ,2,4-Trichlorobenzene Naphthalene Naphthalene ,2,3-Trichlorobenzene ,2,3-Trichlorobenzene Table 6.. Vary Mount Temperature Vocarb 3 Trap Analyte Ambient 75 C C Analyte Ambient 75 C C Dichlorodifluoromethane Dichlorodifluoromethane Chloromethane Chloromethane Vinyl Chloride Vinyl Chloride Bromomethane Bromomethane Chloroethane Chloroethane Trichlorofluoromethane Trichlorofluoromethane Methyl Acrylate..6. Methyl Acrylate Toluene Toluene Bromoform Bromoform Pentachloroethane Pentachloroethane ,2,4-Trichlorobenzene ,2,4-Trichlorobenzene Naphthalene Naphthalene ,2,3-Trichlorobenzene ,2,3-Trichlorobenzene Elevated sample mount temperature eliminates a cold spot at the top of the glassware. On average, increasing the mount temperature from ambient conditions to 75 C results in a 6% increase in the response factors for the heavier components without sacrificing the response of the early eluting gases. Furthermore, the data clearly indicates that removal of the MCS results in the loss of analyte response on the hydrophilic #3 trap. For example, with the MCS removed, the response factor for naphthalene drops by 5% as shown in Graph 3. Graph 3. Vary Mount Temperature - Naphthalene s Temperature ( C) - Vocarb - Vocarb doc; 9-Jun-3 Page 5 of 7
6 Vary Desorb Tables 7 and 8 compare the results obtained at different desorb times when the MCS is installed and removed, respectively. Table 7.. Vary Desorb Vocarb 3 Trap Analyte 2 min. 4 min. 6 min. 8 min. Analyte 3 sec. min. 2 min. 4 min. Dichlorodifluoromethane Dichlorodifluoromethane Chloromethane Chloromethane Vinyl Chloride Vinyl Chloride Bromomethane Bromomethane Chloroethane Chloroethane Trichlorofluoromethane Trichlorofluoromethane Methyl Acrylate Methyl Acrylate Toluene Toluene Bromoform Bromoform Pentachloroethane Pentachloroethane ,2,4-Trichlorobenzene ,2,4-Trichlorobenzene Naphthalene Naphthalene ,2,3-Trichlorobenzene ,2,3-Trichlorobenzene Table 8.. Vary Desorb Vocarb 3 Trap Analyte 2 min. 4 min. 6 min. 8 min. Analyte 3 sec. min. 2 min. Dichlorodifluoromethane Dichlorodifluoromethane Chloromethane Chloromethane Vinyl Chloride Vinyl Chloride Bromomethane Bromomethane Chloroethane Chloroethane Trichlorofluoromethane Trichlorofluoromethane Methyl Acrylate Methyl Acrylate Toluene Toluene Bromoform Bromoform Pentachloroethane Pentachloroethane ,2,4-Trichlorobenzene ,2,4-Trichlorobenzene Naphthalene Naphthalene ,2,3-Trichlorobenzene ,2,3-Trichlorobenzene When the hydrophobic Vocarb 3 trap is installed, variations in desorb time between 2 8 minutes do not significantly affect analyte response. The hydrophilic #3 trap, however, shows a better overall response with a shorter desorb time. Longer desorb times yield poor chromatography with the early eluting components, i.e., higher baselines with broadened peaks due to the excess water that is also desorbed onto the column. The result is wide variability in the calculated response factors. Thus, it was deemed unnecessary to examine desorb times longer than 4 minutes with the #3 trap. Also, a longer desorb time in conjunction with removal of the moisture control system further compounds the problem for the #3 trap because the responses of the later eluting analytes begin to show significant variability. Graph 4 provides a graphical summary of this information. 3-.doc; 9-Jun-3 Page 6 of 7
7 Graph 4. Vary Desorb - Vinyl Chloride Vary Desorb - Vinyl Chloride #3 - #3 Vary Desorb - Toluene Vary Desorb - Naphthalene In general, incorporation of a moisture control system (MCS) appears to help with water removal. When the MCS is installed and a Vocarb 3 trap is used, a dry purge step can be minimized. With the MCS removed, however, a dry purge is needed in order to obtain responses similar to those observed when it is installed. The MCS also provides better consistency in the data obtained with the #3 trap. When the MCS is removed, the responses tend to drop because this trap is holds water in the silica gel layer and allows the water to be desorbed to the GC. The data further indicates that a heated mount feature, such as that provided on the Tekmar 3 Sample Concentrator, provides excellent responses for the heavier components in the Compound List when it is maintained at an elevated temperature. Conclusions Excess water can pose difficulties in purge & trap analyses. But using the combination of a dry purgeable trap, a moisture control system, split injection, and a heated mount provides maximum reduction of water that gets transferred to the column. Figures and 2 provide quick references to the best conditions for water removal with a Vocarb 3 trap and a #3 trap, respectively. Figure. Purge & Trap Conditions for Vocarb 3 Trap Figure 2. Purge & Trap Conditions for 3-.doc; 9-Jun-3 Page 7 of 7
Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1
Application Note US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Thermo Scientific TRACE 1310 GC and ISQ LT MS System Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract
More informationExploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time
Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Introduction Purge and Trap (P&T) concentrators
More informationRoger Bardsley, Applications Chemist; Teledyne Tekmar Page 1
Application Note US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC / 5977A MS Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Abstract US EPA Method
More informationAnalytical Trap Comparison for USEPA Method 8260C
Analytical Trap Comparison for USEPA Method 8260C Application Note Abstract Purge and trap concentration is a technique that is used for the analysis of Volatile Organic Compounds (VOCs). The major component
More informationUS EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1806 P+T VOC ENV US EPA Method 524.2 with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS Amy Nutter, Applications Chemist; Teledyne
More informationSolid Phase Micro Extraction of Flavor Compounds in Beer
Solid Phase Micro Extraction of Flavor s in Beer ANNE JUREK USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler Application Note Environmental
More informationMaximizing Sample Throughput In Purge And Trap Analysis
Maximizing Sample Throughput In Purge And Trap Analysis LINDSEY PYRON ANNE JUREK INTRODUCTION There are several demands and requirements imposed on chemists performing volatile organic analysis (VOC) in
More informationUS EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS
Application Note US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Author Amy Nutter Applications Chemist, Teledyne Tekmar Abstract US EPA Method 8260 in conjunction
More informationOptimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator
Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator Application Note Introduction Environmental laboratories have utilized the Purge & Trap sample concentration technique for over
More informationValidation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS
Validation of USEPA Method 524.2 Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Application Note By: Nathan Valentine Abstract The US EPA developed Method 524.2¹, Measurement
More informationAnalysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler
Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler Application Note By: Teri Dattilio Introduction Purge and Trap concentration
More informationAnalysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler
Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the Introduction Although analysis of VOCs by purge and trap is considered a mature technique, advances in
More informationValidation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS
Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS Application Note By Tyler Trent Abstract In order to determine the concentration of volatile organic compounds
More informationUsing Hydrogen as An Alternative Carrier Gas for US EPA 8260
Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Application Note Abstract Due to regional shortages and increasing costs of helium, the preferred carrier gas in gas chromatography, alternative
More informationOptimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note Environmental Author Anne Jurek Applications
More informationMethanol Extraction of high level soil samples by USEPA Method 8260C
Methanol Extraction of high level soil samples by USEPA Method 8260C Abstract In order to determine the concentration of Volatile Organic Compounds (VOCs) in soil and waste samples the USEPA developed
More informationAnalysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler
Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler Application Note By: Anne Jurek Abstract In order to determine the amount
More informationA Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas
A Comparison of Volatile Organic Compound When Using Nitrogen as a Gas Application Note By: Anne Jurek Abstract For many years Helium has been the gas of choice for purging Volatile Organic Compounds (VOCs).
More informationOptimizing. Abstract: is standardd. procedures. altered to
Optimizing Standard Preparation ANNE JUREK Abstract: Standardd preparation can often be a time consuming, tedious process. The opportunity for human error or inconsistencies between individual preparation
More informationValidation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions
Validation of Volatile Organic Compound by USEPA Method 8260C Application Note Abstract In order to determine the concentration of volatile organic compounds (VOCs) in water and soil matrices the USEPA
More informationValidation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler
Validation of USEPA Method 524.2 Using a Stratum PTC and the New AQUATek 100 Autosampler Application Note Abstract Automation is the key to increasing laboratory productivity and minimizing costs. It is
More informationSystem and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.
ANNE JUREK Optimal Conditions for USEPA Method 8260B Analysis Using the EST Analyticall Sampling System and the Shimadzu GCMS-QP2010s Introduction: The USEPA developedd Method 8260B for the determination
More informationINNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical
INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in
More informationAnalysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.
Optimal Conditions for USEPA Method 8260 Analysis Anne Jurek Introduction: Over the past decade, the need for environmental laboratoriess using purge and trap systems to report at or below the Method Detection
More informationANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.
Determination of Volatile Petroleum Hydrocarbons ANNE JUREK Abstract: Due to current events, the importance of determining volatile petroleum hydrocarbons in both soils and waters has become an issue.
More informationValidation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters
Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Application Note Abstract Water quality laboratories across the
More informationTO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated
TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated Lee Marotta, Field Application Scientist, PerkinElmer Miles Snow, Research Scientist, PerkinElmer
More informationApplication Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek
Requirements of an Automated Sample Delivery System in Today s Realm of Ever Increasing Sensitivity Demands Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek Abstract
More informationMethod 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8
application Note Gas Chromatography/ Mass Spectrometry Authors Ruben Garnica Dawn May PerkinElmer, Inc. Shelton, CT USA Method by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ
More informationApplication Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek
Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek
More informationPerformance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices
Application Note Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices Abstract By: Ed Price In today s laboratories, increased efficiency and productivity are of
More informationAnalysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70
Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70 Application Note Abstract Geosmin and 2-Methylisoborneol are organic compounds that have a distinct scent. These compounds
More informationDetermination of Volatile Organic Compounds in Air
JSB is an authorised partner of Determination of Volatile Organic s in Air Anne Jurek #126 Introduction Air pollution is a growing problem due to the global economy and industrial development in many countries.
More informationEvaluation of a New Analytical Trap for Gasoline Range Organics Analysis
Abstract Purge and Trap (P&T) is a concentration technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible
More informationU.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS
ENVIRONMENTAL application note U.S. EPA VOLATILE ORGANICS METHOD 524.2 USING PURGE AND TRAP GC/MS Cheri Coody, Public Health Laboratory, Mississippi State Health Department, Jackson, MS USA Michael J.
More information2017 EPA Method Update Rule and EPA Method 624.1
Introduction Method 624 is for the determination of volatile organic compounds in industrial discharges and other liquid environmental samples by gas chromatography combined with mass spectrometry (GC/MS).
More informationAnalysis of Low Level Volatile Organic Compounds in Air Anne Jurek
JSB is an authorised partner of Analysis of Low Level Volatile Organic s in Air Anne Jurek #111 Introduction The New Jersey Department of Environmental Protection (NJDEP) developed a new low level TO-15
More informationOptimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s
JSB is an authorised partner of Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s Anne Jurek #120 Introduction: The USEPA developed
More informationOptimizing of Volatile Organic Compound Determination by Static Headspace Sampling
JSB is an authorised partner of Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling Anne Jurek Introduction: #134 Static headspace sampling has always been a viable option
More informationRoger Bardsley, Applications Chemist; Teledyne Tekmar Page 1
Application Note Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform ( 염화비닐, 아크릴로니트릴, 브로모포름 ) with Static and Dynamic Headspace GC/MS Roger Bardsley, Applications Chemist;
More informationOptimizing Standard Preparation
JSB is an authorised partner of Optimizing Standard Preparation Anne Jurek #106 Abstract: Standard preparation can often be a time consuming, tedious process. The opportunity for human error or inconsistencies
More informationVolatile Organic Compounds in Every Day Food
Volatile Organic Compounds in Every Day Food Application Note Abstract Volatile Organic Compounds (VOCs) have been analyzed in drinking water and pharmaceuticals extensively in recent years. Now, VOCs
More informationCopyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.
Optimizing Sampling and Analytical Parameters for Soil Vapour Samples using Automated Thermal Desorption / Gas Chromatography / Mass Spectrometry (ATD/GC/MS) Stephen Varisco, Technical Manager, CARO Analytical
More informationUCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring
UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring Application Note Abstract The United States Environmental Agency (USEPA) has established the Unregulated
More informationValidation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System
Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method
More informationA Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260
JSB is an authorised partner of A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 Anne Jurek #101 Introduction: The United States Environmental Protection Agency (USEPA)
More informationMaximizing Production While Minimizing Costs
JSB is an authorised partner of Maximizing Production While Minimizing Costs Anne Jurek #123 Abstract: The need to increase productivity and reduce cost is becoming of greater concern to environmental
More informationA Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer
A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer Application Note By: Anne Jurek Abstract This application note presents
More informationA Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist
A Single Calibration for Waters and Soil Samples Performing EPA Method 8260 Anne Jurek Applications Chemist Abstract A Single Calibration Method for Water AND Soil Samples The United States Environmental
More informationEd George and Anaïs Viven Varian, Inc.
Application Note SI-02408 Evaluation of EPA Method 524.3: a New Draft Method for the Analysis of VOCs in Drinking Water using GC/MS and V:Results GC/MS Software Ed George and Anaïs Viven Varian, Inc. Introduction
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationApplication Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.
Application Note 8 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-5 Application Note Environmental, Canister, TO-5, Air Monitoring, Air Toxics Abstract This application
More informationAUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL
AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL Challenges of Water Monitoring Volatile Organic Compounds (VOCs) can have negative health impacts even at ppb levels VOC concentrations can
More informationMeasuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS
application Note Gas Chromatography Author Heidi Grecsek PerkinElmer, Inc. Shelton, CT 06484 USA Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS Introduction
More informationReactions of Vinyl Chloride and Methanol in a Quadrupole Ion Trap Mass Spectrometer during VOC Analysis
Environ. Sci. Technol. 1999, 33, 1528-1533 Reactions of Vinyl Chloride and Methanol in a Quadrupole Ion Trap Mass Spectrometer during VOC Analysis LIANGQIAO BIAN, EARL G. ALLEY,, AND BERT C. LYNN, JR.*,,
More informationChemistry Instrumental Analysis Lecture 27. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)
More informationThermal Desorption Technical Support
Thermal Desorption Technical Support Note 86a: US EPA Method TO-17 for Monitoring Air Toxics in Ambient Air Using Sorbent Tubes and Automated, Cryogen-free Thermal Desorption Application Note Environmental,
More informationSolid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS
Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS LECO Corporation; Saint Joseph, Michigan USA Key Words: GC-TOFMS,
More informationAmy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1
Application Note Analysis of Epichlorohydrin in Drinking Water Using the Lumin P & T Concentrator Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract Epichlorohydrin (ECH) is a versatile
More informationOptimizing Conditions for Volatiles Analysis
July 2016 SPEXCertiPrep.com/ECS Optimizing Conditions for Volatiles Analysis by Mark A. Ferry GC/MS Technical Consultant MFerry@SPEX.com A publication by Environmental Consulting & Supplies a Division
More informationDetection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS
Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)
More informationSolid Phase Micro Extraction of Flavor Compounds in Beer
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Automed Sampling of Methanol Extractions Applicion Note Environmental Author Anne Jurek Applicions Chemist EST Analytical Cincinni, OH
More informationApplication News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS
HS-20 & GCMS-QP2010Ultra Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS Introduction Volatile Organic Compounds (VOCs) refer to a group
More informationApplying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation
Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Introduction Fire investigation involves many different types of analyses from crime scene
More informationHelium conservation in volatile organic compound analysis using U.S. EPA Method 8260C
APPLICATION NOTE 10441 Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C Authors Andrea Caruso, 1 Tommaso Albertini, 1 Jacob A. Rebholz 2 ; 1 Thermo Fisher Scientific,
More informationHeadspace Technology for GC and GC/MS: Features, benefits & applications
Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over
More informationDetermination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS
PO-CON1609E Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS Pittcon 2016 1110-1 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole
More informationReduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System
Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System Application Note 19080203 Keywords Dual Purge-and-Trap Eclipse Sample Concentrator Gas Chromatography GC GC/MS Model 4560 Sample
More informationRapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air
Environmental Applications Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air By Jason S. Herrington Abstract The following study evaluated the efficacy of using Nutech s 8900DS preconcentrator
More informationPractical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software
Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software GC Columns and Consumables Mark Sinnott Application Engineer January 8 th, 2008 Page 1 Questions to Ask
More informationChromatography. Gas Chromatography
Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture
More informationMeeting NJ Low Level TO-15 Air Testing Method Requirements
Meeting NJ Low Level TO-15 Air Testing Method Requirements By Jason S. Herrington, PhD Abstract The following study evaluated the efficacy of pairing a Markes Unity with CIA Advantage preconcentrator with
More informationImproved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source
Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Application Note Environmental Authors Peter Gautschi and Harry Prest Senior Application Scientist
More informationMETHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS
METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS 1.0 SCOPE AND APPLICATION 1.1 Method 5021 is a general purpose method for the preparation of
More informationAN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION
AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION Paper # 164-8P Pittsburgh Conference 24 T. Wampler, C. Zawodny, L. Mancini CDS Analytical, Inc 465 Limestone Road, Oxford,
More information1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction
1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)
More informationApplication Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ
Application Note Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ -3 Application Note Abstract This application note demonstrates the
More informationAnalysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler
No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing
More informationVOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography
VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography Introduction ASTM International publishes a number of different methods for the determination of volatile compounds in water-based coatings.
More informationActivity in the FID Detection Port: A Big Problem if Underestimated
Activity in the FID Detection Port: A Big Problem if Underestimated Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands It is commonly known in gas chromatography, that many problems can be
More informationEPA TO-17 Volatile Organic Compounds
EPA TO-17 Volatile Organic Compounds Method TO-17 is used to analyze samples for volatile organic compounds collected on multi-bed sorbent tubes, which are thermally desorbed and cryo-focused on the capillary
More informationEnhanced Method Detection Limits for irm-gc/ms of Volatile Organic Compounds Using Purge and Trap Extraction
Luc Zwank, Michael Berg, Swiss Federal Institute for Environmental Science and Technology (EAWAG), Switzerland Key Words Compound Specific Isotope Analysis, Isotope Ratio MS, Purge and Trap, Volatile Organic
More informationDetermination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC
Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,
More informationCALA Directory of Laboratories
CALA Directory of Laboratories Membership Number: 2628 Laboratory Name: Caduceon Environmental Laboratories (Richmond Hill) Parent Institution: Caduceon Enterprises Inc. Address: 110 West Beavercreek Rd.
More informationSelection of a Capillary
Selection of a Capillary GC Column - Series 3 Mark Sinnott Application Engineer March 19, 2009 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector
More informationDetermination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD
Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Application Note Environmental Author Suli Zhao, Andy Zhai Agilent Technologies Co., Ltd. 412 Yinglun Road
More informationVUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0.
System Description The VUV PIONA + Analyzer is a VGA-100 Gas Chromatography (GC) detector configured for PIONA compound analysis by the VUV PIONA + method. The VUV PIONA + method uses relatively simple
More informationLow-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)
Application Note 16181100 Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD) Keywords Beer PFPD Purge and Trap Sulfur Introduction The presence or absence
More informationStudy of Residual Solvents in Various Matrices by Static Headspace
Application Note Abstract United States Pharmacopeia (USP) chapter is a widely used method for identifying and quantifying Organic Volatile Impurities (OVI) used in the production of pharmaceuticals.
More informationANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5
ANALYTICAL REPORT TestAmerica Laboratories, Inc. TestAmerica Savannah 5102 LaRoche Avenue Savannah, GA 31404 Tel: (912)354-7858 Client For: Imperial Paints LLC PO BOX 489 Fairforest, South Carolina 29336
More informationSelection of a Capillary GC Column
Selection of a Capillary GC Column Mark Sinnott Application Engineer March 13, 2008 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer
More informationAn Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.
An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. Jaap de Zeeuw, Ron Stricek and Gary Stidsen Restek Corp Bellefonte, USA To analyze basic compounds at
More informationELECTRONICALLY REPRINTED FROM. Volume 34 Number 3 MARCH
ELECTRONICALLY REPRINTED FROM Volume 34 Number 3 MARCH 2016 2016 www.chromatographyonline.com Extending the Hydrocarbon Range for the Analysis of Soil Gas Samples Using Automated Thermal Desorption Coupled
More informationTHE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS
International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp
More informationThermal Desorption Technical Support
Thermal Desorption Technical Support Note 47: The Analysis of Landfill Gas Compounds using Thermal Desorption GC/MS and a Retention Time Locked Database Key Words: VOCs, directive, compliance, identifi
More informationANALYTICAL REPORT. Job Number: Job Description: Transform Complete
ANALYTICAL REPORT Job Number: 640-4848-1 Job Description: Transform Complete For: American Master*Tech Scientific, Inc. 1330 Thurman Street PO BOX 2539 Lodi, CA 95240 Attention: Mr. Adam Kowalczyk Chad
More informationUSP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument
Application Note Abstract The US Pharmacopeia recently released USP as the current monograph for determining residual solvents in pharmaceutical products by static headspace. The USP classified these
More informationIdentifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application
Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application Food Safety, Environmental Authors Chin-Kai Meng and Bruce Quimby Agilent Technologies, Inc. 2850 Centerville
More informationHighly Sensitive and Rugged GC/MS/MS Tool
Highly Sensitive and Rugged GC/MS/MS Tool For Pesticide Multiresidue Analysis in Food Samples Agilent 7 Series Triple Quadrupole GC/MS. The world s first MS/MS designed specifically for GC Analysis Introduction
More informationVolatile organic compounds (VOCs):
Volatile organic compounds (VOCs): Organic chemicals with a high vapour pressure at room temperature. High vapour pressure results from a low boiling point. The World Health Organization (WHO) defined
More information