1 The First International Proficiency Testing Conference Sinaia, România th th October, 7 SISLI ETFAL RESEARCH AND TRAINING HOSPITAL ENDOCRINOLOGY PROFICIENCY TESTING SCHEME: FIRST NATIONAL ENDOCRINOLOGY PROFICIENCY TESTING IN TURKEY: A PILOT PROJECT Berna Aslan, Nezaket Eren Nihal Yücel, Ayşen Yaral, Leyla Koç Öztürk Şişli Etfal Research and Training Hospital, Biochemistry laboratory, 46 ŞİŞLİ-İSTANBUL Marmara University, Faculty of Dentistry, Department of Biochemistry, NİŞANTAŞI-İSTANBUL aslan_berna@hotmail.com, nezaketeren@yahoo.co.uk, nihal.yucel@turk.net Abstract In our country there are few Proficiency Testing providers in the field of medical laboratory. There is not any national PT provider who offers endocrinology tests in Turkey. By establishing a PT scheme in the fields of endocrinology, we tried to fulfill the need in the field. In addition to answer the requisition in the market we also would like to use this scheme to give practical training about different aspects of PT s to our residents. They all will be prospective participants of the PT schemes and some of them will be providers. We compared the thyroid stimulating hormone, thyroxin, free thyroxin, tri-iodotyronin and free tri-iodotyronin results of the participant laboratories. A series of two samples constituted with pooled human serum and analytes at clinically relevant concentrations are prepared for each PT event. The PT specimens are subsequently verified, sterile-filtered (. µm pore size), and aliquot. Aliquots were stored frozen at -8 C until lyophilization (freeze drying) procedure. The variability of the test materials received by the participants should be small. For these purpose analytical outliers were checked and deleted from the data by using Cochran s test then one-way analysis of variance is carried out. The samples were sufficiently homogeneous. Consensus of Participants Results approach is used for determining the assigned value for a test material. Firstly, clearly spurious results are excluded to prevent the reporting errors. Then, the median/ robust standard deviation method is used to calculate the assigned value and robust standard deviation. Acceptable ranges are established using fixed criteria specified by CLIA '88 Minitab statistical software and MS Excel are used for statistical evaluation. 44

2 Key words Proficiency Tests, Interlaboratory Comparisons, External quality assessment schemes, Robust statistics. INTRODUCTION Quality improvements in the modern clinical laboratory require the continuous inspection of the laboratory s performances and refinement of the procedures being used so their services can meet the needs and expectations of the users. Laboratories which participate proficiency testing schemes obtain an objective tool for assessing and the demonstrating the reliability of the data they are producing []. Although there are various types of proficiency testing schemes, most of them compare the laboratory test results of two or more laboratory. In our country, there are few PT schemes in the field of clinical laboratory, but any of them coordinates PT schemes for endocrinology tests. Big laboratories supply their proficiency tests from foreign suppliers. This condition carry the some drawbacks: Firstly, PT schemes provided by the foreign suppliers are very expensive and they do not reach the small laboratories in the country. Secondly, they do not compare the results of our country, so they are very weak to show the regional problems. By establishing a PT scheme in the fields of endocrinology, we tried to meet the need in the field. In addition to answer the requisition in the market we also would like to use this scheme to give practical training about different aspects of PT s to our residents. They all will be prospective participants of the PT schemes and some of them will be providers. We compared the thyroid stimulating hormone (TSH), thyroxine (T4), free thyroxine (FT4), tri-iodotyronin (T) and free tri-iodotyronin (FT) results of the participant laboratories. EXPERIMENTAL DESIGN. Preparation of Samples A series of two samples constituted with pooled human serum and analytes at clinically relevant concentrations were prepared for the PT event: Sample A, and sample B. We prepared human serum pool by using patient samples accepted for HBsAg, Anti-HCV and HIV tests in our Hospitals Microbiology laboratory. Samples which are negative for HBsAg, Anti-HCV and HIV tests were used for preparing human serum pool. We mixed the pool gently, and then we divide it into two parts. In order to prepare high level sample, patients plasmas with high hormone level were used. After that two pools we obtained were processed separately. We filtered both pools by using. µm pore size filters to prevent the microorganisms in the samples that may deteriorate the components of the sample. After preparing the filtered pools, we separated them ml aliquots. Aliquots were stored frozen at -8 C until lyophilization (freeze drying) procedure. No preservatives or stabilizers are added to the PT samples to eliminate the potential interference from the additives. Samples were labeled after freeze drying. QC samples were stored in -8 o C after lyophilization. 45

3 We performed homogeneity test as explained in the ISO guide 8 []. Samples distributed to 4 clinical chemistry laboratories in İstanbul. Transportation took maximum -4 hours.. Participants This was the first round of the scheme and it is the closed one. Participants were invited by phone calls and were informed about the scheme. Than documents about the scheme and result return sheets were prepared and sent with samples to the participants who would like to attend. 4 laboratories attended to this first round of the scheme; of them were in University Hospitals, 7 were in Research and Training hospitals, 5 were in State Hospitals, were in private hospitals.. Schedule of PT Scheme Three PT events were scheduled annually: April, August and December. A series of two samples is sent for each round. Participants report their results in a week after they receive the samples.4 Statistical analysis We evaluated homogeneity of the PT samples by using Cocran s test and one way ANOVA. For evaluating test results of the participant laboratories we used robust mean and robust standard deviation. Which are statistical techniques recommended in The International Harmonized Protocol for the Proficiency Testing of (Chemical) Analytical Laboratories [, 4]. RESULTS. Checking samples for homogeneity Any conditions relating to the samples affect the outcome of the PT scheme. Even if efforts to ensure homogeneity, samples prepared for PT schemes are usually heterogeneous to some degree. Aliquots of the bulk material which is distributed to participant laboratories show slight variation in composition among themselves [6]. Procedure recommended in ISO 58 [] and IUPAC Harmonized Protocol Appendix [6] for checking samples homogeneity aims to show if this variation is sufficiently small for the purpose. Sufficient homogeneity in PT samples is that this variation in composition among the distributed units (sampling standard deviation, σ sam ) should be small enough to be meaningless in relation to the variation introduced by the measurements conducted by the participants in the proficiency test (the standard deviation for proficiency assessment σ p ) [,, 5]. The 99 Harmonized protocol [7] required that σ sam, should be smaller than, σ p. In our study we choose samples in random order and analyzed TSH levels twice in random order again. By using Cochran s test, we investigated analytical outliers. There was not any analytical outlier. We performed one-way analysis of variance (ANOVA) as explained in the protocol by using Excel software. Then we calculated the σ sam compared it with σ p. We founded that σ sam was, and σ p was, in Sample A and because,<,*, it was sufficiently homogeneous for our 46

4 purpose. In sample B, σ sam was,89 and σ p was,55. Because,89<,*,55 it was sufficiently homogeneous for our scheme. The results obtained in Sample A and B homogeneity test are shown in figure. Homogeineity Data For TSH in Sample A TSH(U/L),8,8,78,76,74,7,7,68, Aliquot number σ an =,; σ sam =,; σ p =,; σ sam =,9 Critical value(c) =,4 c> σ sam Homogeneity Data for TSH in Sample B TSH (U/L),6,4,,8,6,4, Aliquot number σ an =,4; σ sam =,89; σ p =,55; σ sam =,8 Critical value(c) =,48 c> σ sam Figure -Results of homogeneity tests (σ an, analytic variation; σ sam sampling σ p, standard deviation; standard deviation for proficiency assessment). Evaluation of the participant Laboratories results We first checked the frequency distribution of the series. By using Minitab software we determine whether the distributions fit normal distribution by using Anderson Daring test for normality. All the distributions except the one of the sample B FT, showed normal distribution. Secondly, we calculate the arithmetic mean, standard deviation, uncertainty of arithmetic mean, robust, robust standard deviation, and uncertainty of robust mean by using Microsoft Excel. Consensus of Participants Results approach is used for determining the assigned value for a test material. The influences of the outliers are minimized by the use of robust statistical procedures [,, ]. Results are shown in table. 47

5 Table -. Summary of the results of the first round of the pilot study. Number of Arithmetic Participants Standard Deviation Uncertainty of Arithmetic Robust Robust Standard Deviation CV % Uncertainty of Robust Sample A TSH (U/L) 4,669,5,,655, 8,7 T4 (µg/dl) 9,7,4,8 9,,45 6, T (ng/ml),,4,48,8,85 6,4 FT4 (ng/dl) 4,94,5,5,95,7 7,48 FT (pg/ml) 4,957,6,,94,46 5,89 Sample B TSH 4,6,84,74,95,556 7,8 T4 (µg/dl) 8,64,9,9 8,79,877,8 T (ng/ml),,4,5,5,78 5,8 FT4 (ng/dl) 4,5,88,8,55,6 4, FT (pg/ml) 4,86,65,,79,45 9,5 Finally, z-scores were calculated by using robust standard deviation. The z-scores were estimated by dividing the difference between each laboratory result and target robust mean by robust standard deviation ((Laboratory result Target Robust )/robust standard deviation). The z- score values outside the ± standard deviation (SD) were evaluated as a procedure that needed to be investigated. These were considered unsatisfactory, while the values inside the ± SD limits were considered satisfactory, and z-scores outside the ± SD limit but inside the ± SD limits were considered questionable. The z-scores of the participant laboratories for TSH, FT4, FT, T4 and T obtained in Sample A and B are shown in Figure. For TSH only one laboratory got the z- scores lower than ± SD for both sample A and B. All the other laboratories got the z- scores in the range of ± SD and their results considered satisfactory. This laboratory s results were also outside the limits permitted by the CLIA 88 for TSH (Target ± SD). Laboratories got performance score of satisfactory for FT4 except two laboratories whose performance scores were questionable. One laboratory obtained z-scores higher than ± SD. The other had z-scores between ± ±and ± SD in sample B while the z- scores for the sample A were inside the ± SD limits. Results of all the laboratories were inside the limits stated by the CLIA 88 (Target ± SD) [7]. We found two laboratories whose T4 results were outside the ± SD limit, but inside the ± SD limits for sample B. z-scores of these laboratories in the sample A were inside the ± SD limits as all the other laboratories results were. These results in sample B were also outside of the CLIA 88 limits for T4 (Target mean ± % or. µg/dl) [7]. 48

6 TSH-A and TSH-B SCATTERPLOT Scatterplot of T4 -A vst4-b TSH-(SAMPLE A) z-score T4 (SAMPLE A) z-score TSH-(SAMPLE B) z-score T4 (SAMPLE B) z-score - Scatterplot of FT4 -A vs FT4-B Scatterplot of FT -A vs FT-B FT4 (SAMPLE B) z-score T (SAMPLE A) z-score FT4 (SAMPLE B) z-score -5, -,5,,5 FT (SAMPLE B) z-score 5, 7,5 Scatterplot of T -A vst-b T (SAMPLE A) z-score T (SA MPLE B) z-score 4 Figure - z-scores of the participant laboratories Frequency distributions of FT results obtained in sample B did not fit normal distribution (figure ). There was one laboratory finding the higher FT levels and having z-sores higher than ± SD in both samples. In addition, there was one laboratory whose z-score for sample B was lower than ± SD and between ± and ± SD for sample A. Other two laboratories found questionable results at least in one sample. of the laboratories sent results within the ± SD limits. When we compare the results with limits stated by the CLIA 88, we found all laboratories having questionable and unsatisfactory z-scores, also failed for CLIA 88 [7] criteria (Target ± 5% or.5 pg/ml). 49

7 laboratories sent results for T, of them were inside ± standard deviation limits, while one of the remaining laboratory had z-scores greater than SD for sample B and greater than SD for sample A, other one had z-score inside ± SD limit for sample B and outside the SD for Sample A. Only one laboratory sent a result failed according to CLIA criteria [7, 8].for T (Target ± SD) Summary for FT-A A nderson-darling Normality Test A-Squared,55 P-V alue,4,958 StDev,68 V ariance,6 Skew ness,995 Kurtosis,8689 N 4,,5,,5 4, 4,5 5, Minimum,9 st Q uartile,585 Median,94 rd Q uartile, Maximum 4,78 95% C onfidence Interv al for,697,45 95% C onfidence Interv al for Median,744,69 95% Confidence Intervals 95% C onfidence Interv al for StDev,4669,848 Median,7,8,9,,, Summary for FT-B A nderson-darling Normality Test A-Squared,98 P-V alue,,79 StDev,565 V ariance,77 Skew ness,665 Kurtosis 4,478 N 4,6,4, 4, Minimum,48 st Q uartile,5575 Median,79 rd Q uartile,94 Maximum 4,4 95% C onfidence Interv al for,57,57 95% C onfidence Interv al for Median,65,969 95% Confidence Intervals 95% C onfidence Interv al for StDev,49,786 Median,6,7,8,9, Figure - Graphical Summary of Basic Statistics of Sample A and Sample B 5

8 4 DISCUSSION Proficiency testing (PT) schemes are integral parts of the laboratory quality assessments. These schemes provide very useful information about the technical performance of the laboratories including validity of analytical measurements. A PT scheme coordinates interlaboratory comparisons to evaluate the laboratories performances externally and objectively. By doing this, it supplies a very convenient tool for the laboratory to see their performances and the results of their corrective actions. Because accreditation bodies recommend that clinical laboratories that seek accreditation according to ISO 589 [8] to participate an appropriate PT schemes, growing number of laboratories would like to attend PT schemes. This pilot project aims to create a PT scheme that the participants around the country can easily reach. PT organizers distribute portions of a homogeneous material to their participants, collect the test results reports from them, perform statistical analysis and produce reports to inform the participants about their laboratory performances. Data produced by PT scheme can be crucial for many users and lead to misinterpretation. So the PT provider has to create the report formats which are easily interpreted by the users. For this purpose we designed reports containing graphical summary of the all results and showed the places of the user on it. Thus, the participants can see their places among the others visually. We also give number of participants, arithmetic mean, standard deviation, robust mean, robust standard deviation and performance scores of the laboratories calculated by using robust standard deviation in their own languages. We prepared PT specimens in our laboratory, we performed homogeneity tests, and we found that they were sufficiently homogeneous for our purpose, the cost were lower than that of the foreign suppliers. Although, all the distributions obtained in the scheme fit normal distribution except FT in sample B, we calculated performance scores of the participants by using robust mean to be able to eliminate the effects of the outliers. We also evaluated the participant results according to the criteria stated by CLIA 88. Because CLIA criteria for the TSH, FT4 and T is Target ± SD, results evaluated as questionable according to z-scores became satisfactory when CLIA rules were applied. For FT, we found all laboratories having questionable and unsatisfactory z-scores, also failed for CLIA 88 criteria (Target ± 5% or.5 pg/mol). We observed that two laboratories T4 results which were outside the ± SD limit, but inside the ± SD limits for sample B, were also outside of the CLIA 88 limits for T4 (Target mean ± % or. µg/do). These observations showed that performance scores produced in our schemes could catch the results that were outside the CLIA [7] limits. 5 CONCLUSIONS As a result of this pilot project, we coordinate endocrinology proficiency test first time in our country. We produced proficiency test samples in our laboratory with a very affordable cost. Participants of the first round were the medium and big laboratories in İstanbul, because the first round was a closed one. But the scheme was also affordable for small laboratories that perform thyroid tests. 5

9 REFERENCES [] ISO Guide 4-:997 Proficiency testing by interlaboratory comparisons - Part : Development and operation of proficiency testing schemes [] ISO/FDIS 58:5. Statistical methods for use in proficiency testing by interlaboratory comparisons M. Thompson, S.L.R. Ellison, R. Wood. The International Harmonized Protocol For The Proficiency Testing Of Analytical Chemistry Laboratories (IUPAC Technical Report). (Revised 6) [] IUPAC 6 (M. Thompson and R. Wood). The International Harmonised Protocol for the Proficiency Testing of (Chemical) Analytical Laboratories. Pure Appl. Chem. 78, [4] IUPAC 99 (M. Thompson and R. Wood). The International Harmonised Protocol for the Proficiency Testing of (Chemical) Analytical Laboratories. Pure Appl. Chem. 65, -44 [5] Robust statistics: a method of coping with outliers, Analytical Methods Committee, Royal Society of Chemistry, No:6, April [6] ISO Guide 4-:997 Proficiency testing by interlaboratory comparisons - Part : Selection and use of proficiency testing schemes by laboratory accreditation bodies [7] Center for Disease Control and Prevention, Current CLIA Regulations (including all changes through //4, Part 49 (Laboratory Requirements), subpart I (Proficiency Testing Programs for Non-waived Testing), section 49.9 Endocrinology, page 95, ( [8] ISO 589: specifies requirements for quality and competence particular to medical laboratories 5

Homogeneity of EQA samples requirements according to ISO/IEC 17043

Homogeneity of EQA samples requirements according to ISO/IEC 17043 Homogeneity of EQA samples requirements according to ISO/IEC 17043 Dr.-Ing. Frank Baumeister TGZ AQS-BW at Institute for Sanitary Engineering, Water Quality and Solid Waste Management University of Stuttgart

More information

Practical Statistics for the Analytical Scientist Table of Contents

Practical Statistics for the Analytical Scientist Table of Contents Practical Statistics for the Analytical Scientist Table of Contents Chapter 1 Introduction - Choosing the Correct Statistics 1.1 Introduction 1.2 Choosing the Right Statistical Procedures 1.2.1 Planning

More information

Sample Homogeneity Testing

Sample Homogeneity Testing Sample Homogeneity Testing Following the Protocol Outlined in: The International Harmonized Protocol for the Proficiency Testing of Analytical Laboratories, 006 (IHP), MICHAEL THOMPSON, STEPHEN L. R. ELLISON

More information

Schedule for a proficiency test

Schedule for a proficiency test Schedule for a proficiency test Once a laboratory haegistered the year s programme the interaction between the laboratory and Eurofins Miljø A/S will be according to the schedule described below. 1 Reminder

More information

Minitab Project Report - Practical 2. Time Series Plot of Electricity. Index

Minitab Project Report - Practical 2. Time Series Plot of Electricity. Index Problem : Australian Electricity Production Minitab Project Report - Practical Plot of the Data Australian Electricity Production Data Time Series Plot of Electricity Electricity Since these are monthly

More information

A critical view on ISO standard 13528

A critical view on ISO standard 13528 A critical view on ISO standard 13528 Wim Coucke EQALM symposium, Dublin, 20th of October 2017 WIV-ISP Rue Juliette Wytsmanstraat 14 1050 Brussels Belgium T+32 2 642 55 23 F +32 2 642 56 45 wim.coucke@wiv-isp.be

More information

Results of Proficiency Test Rubber/Compounds May 2005

Results of Proficiency Test Rubber/Compounds May 2005 Results of Proficiency Test Rubber/Compounds May 25 Organised by: Spijkenisse, the Netherlands Authors: ing. R.J. Starink Correctors: R. Agten & dr. R.G. Visser Report: iis5p2 June 25 CONTENTS 1 INTRODUCTION...

More information

Objectives. Concepts in Quality Control Data Management. At the conclusion, the participant should be able to:

Objectives. Concepts in Quality Control Data Management. At the conclusion, the participant should be able to: Concepts in Quality Control Data Management Bio-Rad Laboratories Quality System Specialist Thomasine Newby Objectives At the conclusion, the participant should be able to: Describe how laboratory quality

More information

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) INTRODUCTION After a number of meetings and workshops, a group of representatives from 27 organizations

More information

The AAFCO Proficiency Testing Program Statistics and Reporting

The AAFCO Proficiency Testing Program Statistics and Reporting The AAFCO Proficiency Testing Program Statistics and Reporting Program Chair: Dr. Victoria Siegel Statistics and Reports: Dr. Andrew Crawford Contents Program Model Data Prescreening Calculating Robust

More information

Philipp Koskarti

Philipp Koskarti Österreichisches Forschungsinstitut für Chemie und Technik Austrian Research Institute for Chemistry and Technology ISSS round robin 2010 BIOENERGIE Philipp Koskarti 21.10.2010 content Introduction Details

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Use of a ISO template for quantitative schemes : the issues. Annette Thomas

Use of a ISO template for quantitative schemes : the issues. Annette Thomas Use of a ISO 13528 template for quantitative schemes : the issues Annette Thomas ISO 17043 and homogeneity and Stability ISO 17043 requires that an EQA provider demonstrates sufficient homogeneity and

More information

Validation and Standardization of (Bio)Analytical Methods

Validation and Standardization of (Bio)Analytical Methods Validation and Standardization of (Bio)Analytical Methods Prof.dr.eng. Gabriel-Lucian RADU 21 March, Bucharest Why method validation is important? The purpose of analytical measurement is to get consistent,

More information

Analysis of interlaboratory comparison when the measurements are not normally distributed

Analysis of interlaboratory comparison when the measurements are not normally distributed Analysis of interlaboratory comparison when the measurements are not normally distributed Alexandre Allard 1,*, Soraya Amarouche 1* 1 Laboratoire National de métrologie et d Essais, 1 rue Gaston Boissier

More information

Histamine Research ELISA

Histamine Research ELISA Instructions for use Histamine Research ELISA BA E-5800 Histamine Research ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of Histamine in different

More information

Instructions for use. Chromogranin A ELISA TM E-9000

Instructions for use. Chromogranin A ELISA TM E-9000 Instructions for use Chromogranin A ELISA TM E-9000 Chromogranin A ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of Chromogranin A in serum and plasma.

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Results of Proficiency Test OPP, PCP and TeCP in textile December 2016

Results of Proficiency Test OPP, PCP and TeCP in textile December 2016 Results of Proficiency Test OPP, PCP and TeCP in textile December 2016 Organised by: Author: Corrector: Report: (iis) Spijkenisse, the Netherlands ing. A.S. Noordman de Neef dr. R.G. Visser & ing. R. J.

More information

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC 17025 Table of contents 1 GENERAL... 2 2 THE STANDARD SIST EN ISO/IEC 17025... 2 3 SA'S POLICY IN IMPLEMENTING

More information

Analytische Qualitätssicherung Baden-Württemberg

Analytische Qualitätssicherung Baden-Württemberg Analytische Qualitätssicherung Baden-Württemberg Proficiency Test UKWIR SS 17 PAH in surface water with suspended solids Anthracene, Fluoranthene, Naphtalene, Benzo[a]pyrene, Benzo[b]fluoranthene, Benzo[k]fluoranthene,

More information

Results of Proficiency Test Benzene & Toluene March 2017

Results of Proficiency Test Benzene & Toluene March 2017 Results of Proficiency Test Benzene & Toluene March 217 Organised by: Authors: Correctors: Report: Institute for Interlaboratory Studies (iis) Spijkenisse Netherlands ing. G.A. Oosterlaken-Buijs dr. R.G.

More information

VitalStim Therapy Electrodes Compared to Generic Electrodes: Evaluating Impedance

VitalStim Therapy Electrodes Compared to Generic Electrodes: Evaluating Impedance VitalStim Therapy Electrodes Compared to Generic Electrodes: Evaluating Impedance VitalStim Therapy Electrodes Compared to Generic Electrodes: Evaluating Impedance The purpose of this document is to evaluate

More information

Analysis For quality control the reference material is analysed at the same time and in the same manner as other samples.

Analysis For quality control the reference material is analysed at the same time and in the same manner as other samples. Smedeskovvej 38 DK-8464 Galten CERTIFICATE FOR QC SW4.1B NUTRIENTS IN NATURAL SEAWATER BATCH: VKI-33-5-0217 INSTRUCTIONS FOR USE OF THE REFERENCE MATERIAL Description This reference material consists of

More information

Participants in the Proficiency Test THM 02/2016

Participants in the Proficiency Test THM 02/2016 2 February 2016 Participants in the Proficiency Test THM 02/2016 Reference: Letter 5th November 2015 Enclosed we will distribute the samples for the Proficiency Test THM 02/2016. In total, 6 laboratories

More information

Reference Materials and Proficiency Testing. CropLife International Meeting October 2015 Gina M. Clapper

Reference Materials and Proficiency Testing. CropLife International Meeting October 2015 Gina M. Clapper Reference Materials and Proficiency Testing CropLife International Meeting 15-16 October 2015 Gina M. Clapper Approaches to Improving and Demonstrating Method and Laboratory Performance Analytical quality

More information

Analysis For quality control the reference material is analysed at the same time and in the same manner as other samples.

Analysis For quality control the reference material is analysed at the same time and in the same manner as other samples. Eurofins Miljø A/S Smedeskovvej 38 DK-8464 Galten CERTIFICATE FOR QC SW3.2B NUTRIENTS IN NATURAL COASTAL/ESTUARINE WATER BATCH: VKI-32-4-0910 INSTRUCTIONS FOR USE OF THE REFERENCE MATERIAL Description

More information

Proficiency test evaluation Performance data evaluation Chapter 8.2.2

Proficiency test evaluation Performance data evaluation Chapter 8.2.2 Proficiency test evaluation Performance data evaluation Chapter 8.2.2 Thursday 1st September 2005 7th ISTA seminar on statistics Hohenheim Germany Sylvain Gregoire Statistical evaluations Sylvain GREGOIRE

More information

Uncertainty of Measurement

Uncertainty of Measurement Uncertainty of Measurement Contents Uncertainty of Measurement.... 3 Uncertainty of Measurement and Measurement Error.... 4 Measuring Uncertainty.... 5 Coverage factor k...6 Factors Affecting Uncertainty....

More information

Report on the 2011 Proficiency Test for the determination of T-2 and HT-2 toxins in wheat flour

Report on the 2011 Proficiency Test for the determination of T-2 and HT-2 toxins in wheat flour CODA CERVA Belgian National Reference Laboratory for Mycotoxins in Food and Feed Report on the 2011 Proficiency Test for the determination of T-2 and HT-2 toxins in wheat flour May 2012 Ph. Debongnie Table

More information

Method of Estimating Uncertainty of Measurement Values in Case of Measurement Methods Involving Non-Linear Calibration

Method of Estimating Uncertainty of Measurement Values in Case of Measurement Methods Involving Non-Linear Calibration Method of Estimating Uncertainty of Measurement Values in Case of Measurement Methods Involving Non-Linear Calibration Shigemi Hosogaya *1, Masato Ota *1, Yukio Ozaki *1, Katsuhiko Kuwa *2, Naotaka Hamasaki

More information

Analysis For quality control the reference material is analysed at the same time and in the same manner as other samples.

Analysis For quality control the reference material is analysed at the same time and in the same manner as other samples. Eurofins Miljø A/S Smedeskovvej 38 DK-8464 Galten CERTIFICATE FOR QC SW3.1B NUTRIENTS IN NATURAL COASTAL/ESTUARINE WATER BATCH: VKI-32-3-0910 INSTRUCTIONS FOR USE OF THE REFERENCE MATERIAL Description

More information

CHEK Proficiency study 664. Solvents in adhesive. Date 3 July 2017 Version 1. Report number CHEK Final CHEK proficiency study July 2017

CHEK Proficiency study 664. Solvents in adhesive. Date 3 July 2017 Version 1. Report number CHEK Final CHEK proficiency study July 2017 CHEK Proficiency study 664 Solvents in adhesive Date 3 July 2017 Version 1 Status Final Report number CHEK-17-664 page 0 Colophon Number 664 Name Solvents in adhesive Contact Author Authorization Netherlands

More information

CHGA (Human) ELISA Kit

CHGA (Human) ELISA Kit CHGA (Human) ELISA Kit Catalog Number KA1884 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3 General

More information

Analysis For quality control the reference material is analysed at the same time and in the same manner as other samples.

Analysis For quality control the reference material is analysed at the same time and in the same manner as other samples. Eurofins Miljø A/S Smedeskovvej 38 DK-8464 Galten CERTIFICATE FOR QC SW4.2B NUTRIENTS IN NATURAL SEAWATER BATCH: VKI-33-4-0910 INSTRUCTIONS FOR USE OF THE REFERENCE MATERIAL Description This reference

More information

Ivo Leito University of Tartu

Ivo Leito University of Tartu Using method validation and performance data for estimating measurement uncertainty Ivo Leito University of Tartu ivo.leito@ut.ee.05.01 1 Overview The main question of uncertainty evaluation The different

More information

CHEK Proficiency study 644. Solvents in adhesive. Date 15 October 2016 Version 1 Status Final Report number CHEK

CHEK Proficiency study 644. Solvents in adhesive. Date 15 October 2016 Version 1 Status Final Report number CHEK CHEK Proficiency study 644 Solvents in adhesive Date 15 October 2016 Version 1 Status Final Report number CHEK-16-644 page 0 Colophon Number 644 Name Solvents in adhesive Contact Author Netherlands Food

More information

Methanephrine (Plasma) EIA

Methanephrine (Plasma) EIA Methanephrine (Plasma) EIA For the quantitative determination of free methanephrine in EDTA or citrate plasma For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 17-METHU-E01-FST

More information

Scoring systems for quantitative schemes what are the different principles?

Scoring systems for quantitative schemes what are the different principles? Scoring systems for quantitative schemes what are the different principles? Dr.-Ing. Frank Baumeister TGZ AQS-BW at Institute for Sanitary Engineering, Water Quality and Solid Waste Management of the University

More information

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012 Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability

More information

SHBG-ELISA

SHBG-ELISA For Research Use Only. Not for invitro diagnostic use. SHBG-ELISA 40-056-205043 SHBG ELISA 40-056-205043 RESEARCH USE ONLY 1 CLINICAL BACKGROUND Sex-hormone-binding globulin (SHBG) is a B-globulin that

More information

ON USING BOOTSTRAP APPROACH FOR UNCERTAINTY ESTIMATION

ON USING BOOTSTRAP APPROACH FOR UNCERTAINTY ESTIMATION The First International Proficiency Testing Conference Sinaia, România 11 th 13 th October, 2007 ON USING BOOTSTRAP APPROACH FOR UNCERTAINTY ESTIMATION Grigore Albeanu University of Oradea, UNESCO Chair

More information

Statistical Reports in the Magruder Program

Statistical Reports in the Magruder Program Statistical Reports in the Magruder Program Statistical reports are available for each sample through LAB PORTAL with laboratory log in and on the Magruder web site. Review the instruction steps 11 through

More information

Development of a harmonised method for specific migration into the new simulant for dry foods established in Regulation 10/2011

Development of a harmonised method for specific migration into the new simulant for dry foods established in Regulation 10/2011 Development of a harmonised method for specific migration into the new simulant for dry foods established in Regulation 1/11 stablishment of precision criteria from an U interlaboratory comparison organised

More information

PILOT OF AN EXTERNAL QUALITY ASSURANCE PROGRAMME FOR CRYSTALLINE SILICA

PILOT OF AN EXTERNAL QUALITY ASSURANCE PROGRAMME FOR CRYSTALLINE SILICA PILOT OF AN EXTERNAL QUALITY ASSURANCE PROGRAMME FOR CRYSTALLINE SILICA MJ Shai, National Institute for Occupational Health ABSTRACT The mining industry has committed itself to reducing work-related silicosis.

More information

CHEK Proficiency study 642. Migration of formaldehyde from melamine kitchenware. Date 7 July 2016 Version 1 Status Final Report number CHEK

CHEK Proficiency study 642. Migration of formaldehyde from melamine kitchenware. Date 7 July 2016 Version 1 Status Final Report number CHEK CHEK Proficiency study 642 Migration of formaldehyde from melamine kitchenware Date 7 July 2016 Version 1 Status Final Report number CHEK-16-642 page 1 Colophon Number 642 Name Migration of formaldehyde

More information

COLA Mass Spec Criteria

COLA Mass Spec Criteria Wednesday April 5, 2017 2pm COLA Mass Spec Criteria Kathy Nucifora, MPH, MT(ASCP) COLA, Accreditation Division Director Columbia, MD DESCRIPTION: Kathy Nucifora, COLA Accreditation Division Director, will

More information

FRANKLIN UNIVERSITY PROFICIENCY EXAM (FUPE) STUDY GUIDE

FRANKLIN UNIVERSITY PROFICIENCY EXAM (FUPE) STUDY GUIDE FRANKLIN UNIVERSITY PROFICIENCY EXAM (FUPE) STUDY GUIDE Course Title: Probability and Statistics (MATH 80) Recommended Textbook(s): Number & Type of Questions: Probability and Statistics for Engineers

More information

Results of Proficiency Test Phthalates in Plastics May 2015

Results of Proficiency Test Phthalates in Plastics May 2015 Results of Proficiency Test Phthalates in Plastics May 2015 Organised by: Author: Corrector: Report: Institute for Interlaboratory Studies Spijkenisse, the Netherlands ing. C.M. Nijssen-Wester dr. R.G.

More information

Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400

Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400 v02-jan2014 Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400 Intended Use The canine CRP immunoassay on Horiba Medical's ABX Pentra 400 is an in vitro diagnostic test for quantitative

More information

BATCH: VKI least 97%. Use. Analysis For quality ples. Production. in accordance

BATCH: VKI least 97%. Use. Analysis For quality ples. Production. in accordance Smedeskovvej 39 DK-8464 Galten CERTIFICATE FOR QC ENDOCRINE DISRUPTORS IN MUNICIPAL SLUDGE S ORGANIC MATERIAL BATCH: VKI-27-1-0297 INSTRUCTION FOR USE OF THE REFERENCE MATERIAL L Description, Quantity

More information

Instructions for use. Glycine ELISA BA E-2100

Instructions for use. Glycine ELISA BA E-2100 Instructions for use Glycine ELISA BA E-2100 Glycine Urine ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of Glycine in urine. After derivatization

More information

Exercises from Chapter 3, Section 1

Exercises from Chapter 3, Section 1 Exercises from Chapter 3, Section 1 1. Consider the following sample consisting of 20 numbers. (a) Find the mode of the data 21 23 24 24 25 26 29 30 32 34 39 41 41 41 42 43 48 51 53 53 (b) Find the median

More information

A basic introduction to reference materials. POPs Strategy

A basic introduction to reference materials. POPs Strategy A basic introduction to reference materials POPs Strategy 2009-2010+ A tutorial 16 September 2009 Angelique Botha R&D metrologist Contents Why do we need reference materials? comparability of results metrological

More information

i-stat ctni Performance Verification vs. ABBOTT AxSYM

i-stat ctni Performance Verification vs. ABBOTT AxSYM Report #1 i-stat ctni Performance Verification vs. ABBOTT AxSYM i-stat Corporation - 1 Summary of Evaluation Study Results Correlation Slope Note 1 with comparative assay for [ctni]

More information

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant SWGDRUG GLOSSARY These definitions were developed and adopted by the SWGDRUG Core Committee from a variety of sources including The United Nations Glossary of Terms for Quality Assurance and Good Laboratory

More information

Method Validation and Accreditation

Method Validation and Accreditation SELAMAT Mycotoxins Workshop China, 11-15th December 2006 Method Validation and Accreditation Dr Hamide Z Şenyuva Senior Research Scientist TÜBİTAK-ATAL, TURKEY hamide.senyuva@tubitak.gov.tr SELAMAT Mycotoxins

More information

Chapter 2 Exercise Solutions. Chapter 2

Chapter 2 Exercise Solutions. Chapter 2 Chapter 2 Exercise Solutions Chapter 2 2.3.1. (a) Cumulative Class Cumulative Relative relative interval Frequency frequency frequency frequency 0-0.49 3 3 3.33 3.33.5-0.99 3 6 3.33 6.67 1.0-1.49 15 21

More information

Calibration Verification Linearity in the Clinical Lab. Did I Pass or Fail? 2017 ASCLS New Jersey

Calibration Verification Linearity in the Clinical Lab. Did I Pass or Fail? 2017 ASCLS New Jersey Calibration Verification Linearity in the Clinical Lab. Did I Pass or Fail? 2017 ASCLS New Jersey Presentation Topics & Objectives Calibration Verification Key Definitions Why do I need to perform CV?

More information

EN-TRAP. The European Network of Testing Facilities for the Quality Checking of Radioactive Waste Packages

EN-TRAP. The European Network of Testing Facilities for the Quality Checking of Radioactive Waste Packages EN-TRAP The European Network of Testing Facilities for the Quality Checking of Radioactive Waste Packages Pierre Van Iseghem SCK CEN 1 March 2008 Why / when characterize RW? Independent Verification Regulator

More information

i-stat ctni Performance Verification

i-stat ctni Performance Verification Report #1 i-stat ctni Performance Verification vs. IMMULITE TURBO DPC i-stat Corporation - 1 Summary of Evaluation Study Results Correlation Slope Note 1 with comparative assay for all ctni data 0.32 Critical

More information

Instructions for use. Tryptophan ELISA BA E-2700

Instructions for use. Tryptophan ELISA BA E-2700 Instructions for use ELISA BA E-2700 ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of in urine, serum, and plasma samples. After extraction and derivatization

More information

Review Inter-laboratory studies in analytical chemistry. Edelgard Hund, D.Luc Massart, Johanna Smeyers-Verbeke

Review Inter-laboratory studies in analytical chemistry. Edelgard Hund, D.Luc Massart, Johanna Smeyers-Verbeke Analytica Chimica Acta 423 (2000) 145 165 Review Inter-laboratory studies in analytical chemistry Edelgard Hund, D.Luc Massart, Johanna Smeyers-Verbeke Chemo AC, Farmaceutisch Instituut, Vrije Universiteit

More information

i-stat ctni Performance Verification vs. J&J Vitros ECi

i-stat ctni Performance Verification vs. J&J Vitros ECi Report #1 i-stat ctni Performance Verification vs. J&J Vitros ECi i-stat Corporation - 1 - Summary of Evaluation Study Results Correlation Slope Note 1 with comparative assay for [ctni]

More information

Precision estimated by series of analysis ISO and Approach Duplicate Approach

Precision estimated by series of analysis ISO and Approach Duplicate Approach Agenda Item 9 JOINT FAO/WHO FOOD STANDARDS PROGRAMME CODEX COMMITTEE ON METHODS OF ANALYSIS SAMPLING Thirty-seventh th Session Budapest, Hungary, 6 February 016 (Comments prepared by the German Delegation)

More information

A case study on how to maintain confidence of thermal properties test: Thermal conductivity of building insulation materials

A case study on how to maintain confidence of thermal properties test: Thermal conductivity of building insulation materials Thermochimica Acta 455 (2007) 90 94 A case study on how to maintain confidence of thermal properties test: Thermal conductivity of building insulation materials Young-Sun Jeong a,b,, Gyoung-Soek Choi a,

More information

Instructions for use 5-HIAA ELISA BA E-1900

Instructions for use 5-HIAA ELISA BA E-1900 Instructions for use ELISA BA E-1900 ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of 5-Hydroxy-3-Indole Acetic Acid () in urine. First, is quantitatively

More information

The Levey-Jennings Chart. How to get the most out of your measurement system

The Levey-Jennings Chart. How to get the most out of your measurement system Quality Digest Daily, February 1, 2016 Manuscript 290 How to get the most out of your measurement system Donald J. Wheeler The Levey-Jennings chart was created in the 1950s to answer questions about the

More information

CHEK Proficiency study 671. CMI, MI and free formaldehyde in body lotion. Date 30 November 2017 Version 1. Report number CHEK

CHEK Proficiency study 671. CMI, MI and free formaldehyde in body lotion. Date 30 November 2017 Version 1. Report number CHEK CHEK Proficiency study 671 CMI, MI and free formaldehyde in body lotion Date 30 November 2017 Version 1 Status Final Report number CHEK-17-671 page 0 Colophon Number 671 Name CMI, MI and free formaldehyde

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis ISO GUIDE 34 ANAB Cert# AR-1470 ISO/IEC 17025 ANAB Cert# AT-1467 ATROPINE SULFATE CERTIFIED REFERENCE MATERIAL CERTIFIED PURITY: 99.9%, U crm = ±0.2% k = 2 (Mass Balance/anhydrous

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Measurement Uncertainty - How to Calculate It In The Medical Laboratory

Measurement Uncertainty - How to Calculate It In The Medical Laboratory Measurement Uncertainty - How to Calculate It In The Medical Laboratory Godfrey C. Moses, PhD, FCACB National Director, Gamma-Dynacare Medical Laboratories 11/18/2009 Moses, GC; GDML 1 Introduction Disclaimers

More information

Inferences About the Difference Between Two Means

Inferences About the Difference Between Two Means 7 Inferences About the Difference Between Two Means Chapter Outline 7.1 New Concepts 7.1.1 Independent Versus Dependent Samples 7.1. Hypotheses 7. Inferences About Two Independent Means 7..1 Independent

More information

centro tecnológico Development of reference materials: a case study on packaging material as a source of taints in foods

centro tecnológico Development of reference materials: a case study on packaging material as a source of taints in foods centro tecnológico ESN CONFERENCE: SENSORY ANALYSIS. MORE THAN JUST FOOD Session 5: Proficiency testing in sensory analysis. Development of reference materials: a case study on packaging material as a

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis ISO GUIDE 34 ACLASS Cert# AR-1470 ISO/IEC 17025 ACLASS Cert# AT-1467 AMOXICILLIN TRIHYDRATE CERTIFIED REFERENCE MATERIAL HO O H H N H S CH 3 3H 2 O NH 2 N CH 3 COOH CERTIFIED PURITY:

More information

Proficiency testing: Aqueous ethanol. Test and measurement Workshop Marcellé Archer. 20 September 2011

Proficiency testing: Aqueous ethanol. Test and measurement Workshop Marcellé Archer. 20 September 2011 Proficiency testing: Aqueous ethanol Test and measurement Workshop Marcellé Archer 20 September 2011 NMISA Ethanol PT schemes Purpose of a PT scheme Consistent good performance is the goal One bad result

More information

Volume 11 Issue 6 Version 1.0 November 2011 Type: Double Blind Peer Reviewed International Research Journal Publisher: Global Journals Inc.

Volume 11 Issue 6 Version 1.0 November 2011 Type: Double Blind Peer Reviewed International Research Journal Publisher: Global Journals Inc. Volume 11 Issue 6 Version 1.0 2011 Type: Double Blind Peer Reviewed International Research Journal Publisher: Global Journals Inc. (USA) Online ISSN: & Print ISSN: Abstract - Time series analysis and forecasting

More information

Bovine FSH(Follicle Stimulating Hormone) ELISA Kit

Bovine FSH(Follicle Stimulating Hormone) ELISA Kit Bovine FSH(Follicle Stimulating Hormone) ELISA Kit Catalogue No.: EB0141 Size: 48T/96T Reactivity: Bovine Detection Range: 1.563-100mIU/ml Sensitivity:

More information

Quality assurance of analytical measurements of CO2 parameters. Andrew G. Dickson

Quality assurance of analytical measurements of CO2 parameters. Andrew G. Dickson Quality assurance of analytical measurements of CO2 parameters Andrew G. Dickson 1 Quality assurance constitutes the system by which an analytical laboratory can assure outside users that the analytical

More information

Inference for Regression Inference about the Regression Model and Using the Regression Line, with Details. Section 10.1, 2, 3

Inference for Regression Inference about the Regression Model and Using the Regression Line, with Details. Section 10.1, 2, 3 Inference for Regression Inference about the Regression Model and Using the Regression Line, with Details Section 10.1, 2, 3 Basic components of regression setup Target of inference: linear dependency

More information

Analytische Qualitätssicherung Baden-Württemberg

Analytische Qualitätssicherung Baden-Württemberg Analytische Qualitätssicherung Baden-Württemberg Proficiency Test 7/16 - Ions in waste water - ammonium-nitrogen, nitrate-nitrogen, nitrite-nitrogen, total phosphorous, total cyanide, cyanide (week acid

More information

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015 Uncertainty of Measurement (Analytical) Maré Linsky 14 October 015 Introduction Uncertainty of Measurement GUM (Guide to the Expression of Uncertainty in Measurement ) (Analytical) Method Validation and

More information

EURL Food Contact Material. ILC 2009/02 BPA in 50% ethanol

EURL Food Contact Material. ILC 2009/02 BPA in 50% ethanol EURL Food Contact Material. ILC 9/ BPA in 5% ethanol Report of an interlaboratory comparison organised by the European Union Reference Laboratory for Food Contact Materials ILC1 1 - Photoinitiators in

More information

Chapter 2. Mean and Standard Deviation

Chapter 2. Mean and Standard Deviation Chapter 2. Mean and Standard Deviation The median is known as a measure of location; that is, it tells us where the data are. As stated in, we do not need to know all the exact values to calculate the

More information

Inter-laboratory study for tetracyclines in poultry muscle

Inter-laboratory study for tetracyclines in poultry muscle Projectnr.: 7.036.01 NRL tasks, residues in animal products Project manager: J.A. van Rhijn This study was conducted with financial support of the Ministry of Agriculture, Nature and Food Quality of The

More information

Application Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010

Application Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010 Page of 6 Application Note for Gentian Cystatin C Immunoassay on For in vitro diagnostic use only Cobas c50 (Optimised Volumes II), Roche Diagnostics v06-september 200 General Information Intended Use

More information

TC2 EXPERIENCES IN COLLABORATIVE STUDIES, METHOD VALIDATION AND PROFICIENCY TESTING

TC2 EXPERIENCES IN COLLABORATIVE STUDIES, METHOD VALIDATION AND PROFICIENCY TESTING Distributed in TC2 as document No ICG/TC-2/07-1444 TC2 EXPERIENCES IN COLLABORATIVE STUDIES, METHOD VALIDATION AND PROFICIENCY TESTING E. Guadagnino Stazione Sperimentale del Vetro, Via Briati 10, 30141

More information

Scanning Electron Microscopy Scheme

Scanning Electron Microscopy Scheme Group Report Round 7A November 2017 Scanning Electron Microscopy Scheme BACKGROUND This report covers Round 7 of the SEMS asbestos fibre counting PT scheme. The scheme is operated by HSL, in collaboration

More information

Instrumentation (cont.) Statistics vs. Parameters. Descriptive Statistics. Types of Numerical Data

Instrumentation (cont.) Statistics vs. Parameters. Descriptive Statistics. Types of Numerical Data Norm-Referenced vs. Criterion- Referenced Instruments Instrumentation (cont.) October 1, 2007 Note: Measurement Plan Due Next Week All derived scores give meaning to individual scores by comparing them

More information

Continuous random variables

Continuous random variables Continuous random variables A continuous random variable X takes all values in an interval of numbers. The probability distribution of X is described by a density curve. The total area under a density

More information

Assessing Test Quality from EQA/PT Surveys. James O. Westgard University of Wisconsin Medical School Sten A. Westgard Westgard QC, Inc.

Assessing Test Quality from EQA/PT Surveys. James O. Westgard University of Wisconsin Medical School Sten A. Westgard Westgard QC, Inc. Assessing Test Quality from EQA/PT Surveys James O. Westgard University of Wisconsin Medical School Sten A. Westgard Westgard QC, Inc. Madison, WI Issues to consider Is analytical quality better than needed

More information

Minidex Trial Specific Sample Collection, Handling, Processing, Analysis and Storage Guidance

Minidex Trial Specific Sample Collection, Handling, Processing, Analysis and Storage Guidance 16A: Microsampling protocol Minidex Trial Specific Sample Collection, Handling, Processing, Analysis and Storage Guidance EudraCT No: 2015-005342-63, REC Ref: 16/NW/0396 Guidance Sheet 16A Version 2.0

More information

CHAPTER 4 VARIABILITY ANALYSES. Chapter 3 introduced the mode, median, and mean as tools for summarizing the

CHAPTER 4 VARIABILITY ANALYSES. Chapter 3 introduced the mode, median, and mean as tools for summarizing the CHAPTER 4 VARIABILITY ANALYSES Chapter 3 introduced the mode, median, and mean as tools for summarizing the information provided in an distribution of data. Measures of central tendency are often useful

More information

Technical Guide 4 A Guide on Measurement Uncertainty in Medical Testing

Technical Guide 4 A Guide on Measurement Uncertainty in Medical Testing ACCREDITATION SCHEME FOR LABORATORIES Technical Guide 4 A Guide on Measurement Uncertainty in Medical Testing Technical Guide 4, Feb 13 Technical Guide 4, First Edition, February 013 Page 1 of 51 The SAC

More information

Zusammenfassung der Ergebnisse SEMS 7 Scanning Electron Microscopy Scheme - Round 7

Zusammenfassung der Ergebnisse SEMS 7 Scanning Electron Microscopy Scheme - Round 7 Zusammenfassung der Ergebnisse SEMS 7 Scanning Electron Microscopy Scheme - Round 7 Ergebnisse CRB Ergebnisse Ringversuch Sample 1-7SEM1 [ fibres mm -2 ] Amphibole Chrysotile Other inorg. fibres Total

More information

WELCOME!! LABORATORY MATH PERCENT CONCENTRATION. Things to do ASAP: Concepts to deal with:

WELCOME!! LABORATORY MATH PERCENT CONCENTRATION. Things to do ASAP: Concepts to deal with: WELCOME!! Things to do ASAP: Read the course syllabus; information regarding testing, homework, lecture schedules, expectations and course objectives are all there Read the weekly overview; lecture objectives

More information

Descriptive Statistics-I. Dr Mahmoud Alhussami

Descriptive Statistics-I. Dr Mahmoud Alhussami Descriptive Statistics-I Dr Mahmoud Alhussami Biostatistics What is the biostatistics? A branch of applied math. that deals with collecting, organizing and interpreting data using well-defined procedures.

More information

Lactate Dehydrogenase Assay Kit

Lactate Dehydrogenase Assay Kit Lactate Dehydrogenase Assay Kit Catalog Number KA0878 500 assays Version: 10 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials

More information

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 NPL REPORT CMAM 73 Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 Peter Hanes October 2001 The National Physical Laboratory is operated on behalf

More information

Part A: Salmonella prevalence estimates. (Question N EFSA-Q ) Adopted by The Task Force on 28 March 2007

Part A: Salmonella prevalence estimates. (Question N EFSA-Q ) Adopted by The Task Force on 28 March 2007 The EFSA Journal (2007) 98, 1-85 Report of the Task Force on Zoonoses Data Collection on the Analysis of the baseline survey on the prevalence of Salmonella in broiler flocks of Gallus gallus, in the EU,

More information