COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

Size: px
Start display at page:

Download "COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL"

Transcription

1 COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Industrial Nitrate Ion Electrode is used to quickly, simply, accurately, and economically measure nitrate in in-line batch processing, control processing, or water treatment applications. Required Equipment 1. Cole-Parmer Ion Concentration Controller, Cat. No Cord with U.S.standard plug, Cat. No or cord with European plug, Cat. No is necessary. 2. Cole-Parmer Industrial Nitrate Ion Electrode, Cat. No It is a sealed, double junction electrode which is easily connected to the Ion Concentration Controller with stripped wire ends. 3. Cole-Parmer Industrial Housings to protect the Nitrate Ion Electrode. See page 1185 of the Cole-Parmer Catalog to choose an Industrial Housing. Required Solutions 1. Deionized or distilled water for solution and standard preparation. 2. Cole-Parmer Nitrate Standard, 1000 ppm NO 3-1, Cat. No To prepare this solution from your own laboratory stock, add 1.37 grams of reagent-grade sodium nitrate to a one liter volumetric flask about half full with distilled water. Swirl the flask gently to dissolve the solid. Fill to the mark with distilled water, cap and upend several times to mix the solution. 3. Cole-Parmer Ionic Strength Adjuster (ISA), 2M (NH 4 ) 2 SO 4, Cat. No (OPTIONAL). To prepare this solution 1

2 from your own laboratory stock, half fill a 1000 ml volumetric flask with distilled water and add 264 grams of reagent-grade ammonium sulfate, (NH 4 ) 2 SO 4. Swirl the flask gentle to dissolve the solid. Fill the flask to the mark with distilled water, cap, and upend several times to mix the contents. ISA is added at the rate of 2 ml of ISA to each 100 ml of standard or sample to adjust the ionic strength to about 0.12M. ELECTRODE PREPARATION Remove the rubber cap covering the electrode tip. Gently shake the electrode downward in the same manner as a clinical thermometer to remove any air bubbles which may be trapped behind the nitrate membrane. Prior to first usage, or after long-term storage, immerse the nitrate membrane in nitrate standard for thirty minutes. The electrode is now ready for use. CALIBRATION PROCEDURE This quick start procedure is designed to get the unit operational. To set alarms and relays, consult the Operator's manual. Standards must be made in the range of 0.01 to 1000 ppm, as the controller will not accept standards outside this range. 1. Connect the electrode with the stripped wire ends to the controller by inserting the center wire into slot #22 and inserting the outer wire braid into slot # Press the [Mode] key 9 times until the display reads [Configuration] 3. Press the [Cal] key. 4. The display should read [KB UNLOCKED]. Press [ENT] 5. The display should now read: [LCD Contrast: 4]. Press [ENT]. 6. The display should now read: [Access Nr.: 0]. Press [ENT]. 7. The display should now read [Valence: #]. Pressing the [/\] and [\/] keys, choose the proper valence of the ion you are testing for. Press [ENT] (If the factory setting is already the correct #, press [ENT]). E.G. For Silver/sulfide, choose -1; For Sodium, choose +1; for Calcium, choose If you chose a valence which is different from the factory setting, the display will read [Reset ppm Cal?] Press [ENT]. 9. The display will now read [Scale: ppm]. Press [Mode]. 10.The display will now read [Configuration]. 11.Press [Mode] key 6 times. Make sure the display reads: [Termoc.: OFF]. If so, go to step 14. If the display reads [Termoc.: ON], go to step 12. 2

3 12.Press the [CAL] key. Press the [\/] key so the display reads [CAL TC: OFF]. The display will flash [UPDATE] and then read: [CAL TC:0.198%/oC]. 13.Press the [Mode] key. The display will now read: [Termoc.: OFF]. 14.Press the [Mode] Key 8 times. The display should read [ppm 5 point cal].(note: It is not necessary to do a 5 point calibration. You can calibrate from 1 to 5 points). Prepare 1 to 5 standard ISE solutions whose concentrations vary by tenfold. Use the 1000 ppm ISE solution as the stock solution. Use the serial dilution method for this preparation and add the ISA when required to each standard 15.To start the calibration sequence, press the [cal] button. The display will now read [Cal Point Nr 1]. 16.Insert the ISE electrode in the standard with the lowest ppm value.(note: for best results, ppm standards should be used in the following order: {0.1, 1, 10, 100, 1000} 17.Press the [cal] button. 18.When a stable millivolt reading is obtained, press the [ENT] button. 19.The display should now read [P1 Decade: 100.0] 20.Choose the ppm range within which the standard lies using the [/\] or [\/] buttons. - select if the standard solution value is < ppm - select if the standard solution value is > ppm - and < ppm - select 1000 if the standard solution is > ppm 21.Press the [ENT] button. 22.The display should now read [P1 Value: (# selected in step 20)]. 23.Choose the actual ppm value of the standard solution by pressing the [/\] or [\/] buttons until the exact ppm value is displayed. Press the [ENT] key to confirm the ppm value. 24.The screen should read [CAL POINT Nr. 2]. If no further calibration points are required, go to step 25. If you wish to enter more calibration points, go to step If no further calibration points are desired, press the [ENT] button. The display will now read [End Point Cal?]. Press the [ENT] button once more. The screen will now display [Zr Pt updated]. Press [ENT]. The display will now read [ppm 5 point cal]. To display Concentration value, press the [Mode] key 10 times. To display millivolt signal supplied by electrode, press the [Mode] key once. 26.If further calibration points are required, repeat steps 16 through 25. After the 5th calibration point is confirmed at step 11, the meter will flash [chk cal pt] and then read [VALUE UPDATED]. Press the [ENT] key. The display will now 3

4 read [ppm 5 point cal]. To display Concentration value, press the [Mode] key 10 times. To display millivolt signal supplied by electrode, press the [Mode] key once. MEASUREMENT Measuring Hints Check electrode slope weekly if not daily. Always use fresh standards for calibration. All samples and standards should be at the same temperature for precise measurement and below 40 o C. A difference of 1 o C in temperature will result in a 2% measurement error. All samples and standards should be at the same pressure for precise measurement. The sensing membrane is normally subject to water uptake and might appear milky. This has no effect on performance. For samples with high ionic strength, prepare standards whose composition is similar to the sample. Always check to see that the membrane is free from air bubbles after immersion into standard or sample. A slow responding electrode may be caused by interferences to the electrode. To restore proper performance, soak the electrode in distilled water for about 5 minutes to clean the membrane, rinse, and soak in standard solution for about 5 minutes. Sample Requirements All samples must be aqueous and not contain organics which can dissolve in the membrane or extract out the liquid ion exchanger. Interferences should be absent. If they are present, use the procedures found in the Interferences section to remove them. The ph range for the nitrate ion electrode is Neutralize samples outside this range with acid or base to bring them in range. Units of Measurement Nitrate concentrations are measured in units of ppm as sodium nitrate, ppm as nitrate, or any other convenient concentration 4

5 unit. Table 1 indicates some of the concentration units. TABLE 1: Concentration Unit Conversion Factors ppm NaNO ppm NO ELECTRODE CHARACTERISTICS Reproducibility Electrode measurements reproducible to ±2% can be obtained if the electrode is calibrated every hour. Factors such as temperature fluctuations, drift, and noise limit reproducibility. Reproducibility is independent of concentration within the electrode's operating range. Interferences Certain anions are electrode interferences and will cause electrode malfunction, drift or measurement errors if present in high enough levels. The level of interfering common anions that will cause a 10% error at three levels of nitrate is given in Table 3. TABLE 3: Concentration of Possible Interferences Causing a 10% Error at Various Levels of Nitrate; Background Ionic Strength of 0.12M (NH 4 ) 2 SO 4. Interferences (ppm) 100 ppm N 10 ppm N 1 ppm N Cl NO Br CN ClO I

6 Interferences such as bromide, iodide and cyanide can be removed by precipitation with 0.5 grams of silver sulfate added to 100 ml of sample. Nitrite interference can be removed by adding 0.3 grams of sulfamic acid to 100 ml of sample. Carbonate and bicarbonate, which are weak interferences, can be removed by acidifying the sample to ph 4.5 with sulfuric acid. Organic (carboxylic) anions hinder the nitrate electrode response and can be removed by adding 10 grams of aluminum sulfate to 100 ml of sample. ClO 4-1 The above interference removal procedures require similar treatment of standards as well as samples. If the electrode is exposed to high levels of interfering ions which cannot be removed, the electrode reading may drift and the response may become sluggish. Restore performance by soaking in distilled water for 30 minutes followed by soaking in nitrate standard for 30 minutes. Temperature Influences Samples and standards should be at the same temperature, since electrode potentials are influenced by changes in temperature. A 1 o C difference in temperature results in a 2% error at the 10 ppm level. Because of the solubility equilibria on which the electrode depends, the absolute potential of the reference electrode changes slowly with temperature. The slope of the nitrate electrode, as indicated by the factor "S" in the Nernst equation, also varies with temperature. Table 4 gives values for the "S" factor in the Nernst equation for the nitrate ion. The operating range of the nitrate ion electrode is 0 o -40 o C, provided that temperature equilibrium has occurred. If the temperature varies substantially from room temperature, equilibrium times up to one hour are recommended. TABLE 4: Temperature vs. Values for the Electrode Slope Temp ( o C) "S"

7 Electrode Response Plotting the mv potential against the nitrate concentration on semi-logarithmic paper results in a straight line with a slope of about 56 mv per decade. (Refer to Figure 1.) 7

8 The time needed to reach 99% of the stable electrode potential reading, the electrode response time, varies from one minute or less in highly concentrated solutions to several minutes near the detection limit. (Refer to Figure 2.) Limits of Detection The upper limit of detection in pure sodium nitrate solutions is 1M. In the presence of other ions, the upper limit of detection is above 10-1 M nitrate, but two factors influence this upper limit. Both the possibility of a liquid junction potential developing at the reference electrode and the salt extraction effect influence this upper limit. Some salts may infuse into the electrode membrane at high salt concentrations, causing deviation from the theoretical response. Either dilute samples between 1M and 10-1 M or calibrate the electrode at 4 or 5 intermediate points. The lower limit of detection is influenced by the slight water solubility of the ion exchanger used in the sensing portion of the electrode. Refer to Figure 1 for a comparison of the theoretical response to the actual response at low levels of nitrate. Nitrate measurements below 10-5 M NO 3-1 (0.6 ppm as NO 3-1 ) should employ low level procedures. 8

9 ph Effects The operating range of the nitrate electrode is from ph 2.5 to ph 11. Electrode Life The nitrate electrode will last six months in normal laboratory use. On-line measurement might shorten operational lifetime to several months or less. In time, the response time will increase and the calibration slope will decrease to the point calibration is difficult and electrode replacement is required. Electrode Storage The nitrate electrodes may be stored for short periods of time in 10-2 M nitrate solution. For longer storage (longer than two weeks), rinse and dry the nitrate membrane and cover the tip with any protective cap shipped with the electrodes. ELECTRODE THEORY Electrode Operation The nitrate electrode consists of an electrode body containing a liquid internal filling solution in contact with a gelled organophilic membrane containing a nitrate ion exchanger. When the membrane is in contact with a solution containing free nitrate ions, an electrode potential develops across the membrane. This electrode potential is measured against a constant reference potential, using a standard ph/mv meter or an ion meter. The level of nitrate ions, corresponding to the measured potential, is described by the Nernst equation: E = E - S log X where: E = measured electrode potential E = reference potential (a constant) S = electrode slope (~56 mv/decade) X = level of nitrate ions in solution The activity, X, represents the effective concentration of the ions in solution. The total nitrate ion concentration, C t, is the sum of free nitrate ion, C f, and complexed or bound perchlorate ion, C b. The electrode is able to respond to only the free ions, whose concentration is : 9

10 C f = C t - C b Since nitrate ions form very few stable complexes, the free ion concentration may be equated to the total ion concentration. The activity is related to the free ion concentration, C f, by the activity coefficient, γ, by: X = γ C f Activity coefficients vary, depending on total ions strength, I, defined as: I = 1/2 ΣC x Z x ² where: C x = concentration of ion X Z x = charge of ion X Σ = sum of all of the types of ions in the solution In the case of high and constant ionic strength relative to the sensed ion concentration, the activity coefficient, γ, is constant and the activity, X, is directly proportional to the concentration. To adjust the background ionic strength to a high and constant value, ionic strength adjuster (ISA) is added to samples and standards. The recommended ISA for nitrate is (NH 4 ) 2 SO 4. Solutions other than this may be used as ionic strength adjusters as long as ions that they contain do not interfere with the electrode's response to nitrate ions. The reference electrode must also be considered. When two solutions of different composition are brought into contact with one another, liquid junction potentials arise. Millivolt potentials occur from the inter-diffusion of ions in the two solutions. Electrode charge will be carried unequally across the solution boundary resulting in a potential difference between the two solutions, since ions diffuse at different rates. When making measurements, it is important to remember that this potential be the same when the reference is in the standardizing solution as well as in the sample solution or the change in liquid junction potential will appear as an error in the measured electrode potential. The composition of the liquid junction filling solution in the reference electrode is most important. The speed with which the 10

11 positive and negative ions in the filling solutions diffuse into the samples should be as nearly equal as possible, that is, the filling solution should be equitransferent. No junction potential can result if the rate at which positive and negative charge carried into the sample is equal. Strongly acidic (ph = 0-2) and strongly basic (ph = 12-14) solutions are particularly troublesome to measure. The high mobility of hydrogen and hydroxide ions in samples make it impossible to mask their effect on the junction potential with any concentration of equitransferent salt. One must either calibrate the electrodes in the same ph range as the sample or use a known incremental method for ion measurement. COMMON MEASUREMENT PROBLEMS...SIMPLE CAUSES? -NOT REPRODUCIBLE sample carryover? sample interferences or complexing agents present? contaminated reference electrode junction -SLOW RESPONSE (READINGS SLOWLY CHANGING) electrode stored in wrong solution? electrode poisoned by sample? -OUT OF RANGE READING electrode(s) not plugged into controller properly? no reference electrode? not enough fill solution left in reference? electrode not in sample solution? -LOW SLOPE OR NO SLOPE standards are old? contaminated? made wrong? sample ph has not been adjusted properly to operating range of electrode? air bubble on electrode surface? controller okay? not enough fill solution left in reference? -NOISY RESPONSE (READINGS RANDOMLY CHANGING) controller not grounded? air bubble on electrode surface? controller okay? not enough fill solution left in reference? -DRIFTY RESPONSE (READINGS CONTINOUSLY CHANGING) excessive leaking at reference electrode junction? clogged reference electrode junction? sensing membrane poisoned by sample? 11

12 temperature problems? sample too concentrated? sensing membrane needs conditioning? -INNACURATE (BUT CALIBRATION IS OK) standards are incorrect? sample is not ph adjusted properly? sample carryover? sample interferences or complexing agents present? TROUBLESHOOTING DIFFICULT MEASUREMENT PROBLEMS -COMPONENTS OF THE ELECTRODE SYSTEM -CONTROLLER -ION-SELECTIVE ELECTRODE -PROPER CALIBRATION -STANDARD(S) -ph ADJUSTMENT -SAMPLE VARIABLES -CONCENTRATION RANGE -PRESSURE -TEMPERATURE -ph -INTERFERENCES -COMPLEXATION perform checkout procedure in instruction manual perform electrode slope check, inspect for physical damage. Polish, soak, or rebuild sensing membrane. monitor leak rate on reference side prepare fresh standards by serial dilution adjust ph to operational range of electrode sample out of range? wrong sample pressure? wrong sample temperature? wrong sample ph? interferences in the sample? complexing agents in sample? SOME QUESTIONS ASKED ABOUT ION-SELECTIVE ELECTRODES 1.How often do you need to calibrate? Recalibrate at least once a week, or daily if in doubt of your results being accurate or reproducible. 2.How long after being opened or made are standards good? The stock standard will last at least six months before discarding, whereas diluted standards treated with ISA/pH buffer should be prepared weekly. 3.How do you store the electrodes? It is best to store them dry when they are not to be used in the next week or so. Empty out the filling solution in the gas-sensing electrode. For shorter periods, store in dilute standard approximating the sample concentration, and ISA/pH 12

13 buffer added when required. 4.Can you do temperature compensation with ISE? Yes, it is possible, but somewhat difficult. First, you have to know the isopotential point for a given electrode system. Second, the concentration of the sample has to be similar in concentration to the isopotential point or else the temperature correction will be very inaccurate. Third, the temperature of the sample can not exceed the operational temperature range of the ISE. Fourth, very little isopotential point data for ISE's is available at this time. It is best to standardize and measure samples at the same temperature without using temperature compensation. 5.Can you do in-line continous ISE measurement without treating the sample? Yes, direct measurement is possible in many cases without ISA/buffer addition to the sample stream. However, Fluoride, Sulfide, Ammonia, and Sodium Electrodes do require ph adjustment and must have ISA/ buffer added to the sample stream. 6.Which standards should be used with the ISE? The most obvious choice will be determined by what concentration units are desired (e.g. ppm as what?). Also, if an electrode is being used to measure another ion (e.g. sulfate with a lead electrode), use a stock standard of the ion to be measured (e.g. sulfate). 7.Why buy a combination electrode instead of seperate ones? Advantages: no external reference electrode needed more economical than price of both one less electrode to fit in the process NITRATE ELECTRODE SPECIFICATIONS (MODEL # ) -SLOPE 55 +/- 3 mv/decade -REPRODUCIBILITY +/- 2% -INTERFERENCES ClO4-1, I -1,CN -1,BF4-1 -TEMPERATURE RANGE -PRESSURE RANGE -RESPONSE TIME -STORAGE -CONCENTRATION RANGE 0 to 40 degrees C 0 to 30 psi 95% response in 30 seconds store in dilute nitrate standard for short-term, and dry long-term 7 x 10-6 M to 1M 13

14 -ph RANGE -RESISTANCE -CARE, MAINTENANCE -TEMPERATURE COMPENSATION -ISOPOTENTIAL POINT -ON-LINE CAPABILITY -APPLICATIONS: -SIZE: 0.5 ppm to 62,000 ppm 2.5 to 11 ph 100 megaohms treat with distilled water followed by nitrate standard Yes 10 ppm nitrate Yes Surface Waters and Sewage Effluent Drinking Waters and Soil Extracts 80 mm in length 12 mm in diameter 3 m cable length ORDERING INFORMATION P/N DESCRIPTION Nitrate Industrial Electrode, combination, sealed, double junction, epoxy body with stripped wire ends Nitrate Standard,1000 ppm NO Nitrate ISA (Ionic Strength Adjustor), 2 M(NH 4 ) 2 SO Ion Concentration Controller Cord with U.S.standard plug Cord with European plug Cole-Parmer Industrial Housings to protect the Nitrate Ion Electrode. See page 1185 of the Cole-Parmer Catalog to choose an Industrial Housing. 14

15 TABLE OF CONTENTS General Instructions...1 introduction...1 required equipment...1 required solutions...1 General Preparation...2 electrode preparation...2 Calibration...2 Measurement...4 measuring hints...4 sample requirements...4 units of measurement...4 Electrode Characteristics...5 reproducibility...5 interferences...5 temperature influences...6 electrode response...7 limits of detection...8 ph effects...9 electrode life...9 electrode storage...9 Electrode Theory...9 electrode operation...9 Common Measurement Problems, Simple Causes...11 Troubleshooting Difficult Measurement Problems...12 Some Questions Asked About Ion-Selective Electrodes...12 Nitrate Electrode Specifications...13 Ordering Information...14 Table of Contents

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Ammonium Ion Electrode is used to quickly, simply, accurately, and economically

More information

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

COLE-PARMER INDUSTRIAL SILVER/SULFIDE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL SILVER/SULFIDE ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL SILVER/SULFIDE ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Silver/sulfide Ion Electrode is used to quickly, simply, accurately,

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Fluoroborate Ion Selective Electrode is used to quickly, simply, accurately, and economically measure Fluoroborate in aqueous solutions.

More information

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions. COLE-PARMER NITRATE ION ELECTRODES INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate

More information

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required

More information

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions. Cole-Parmer Ammonium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium

More information

1) A ph/mv meter or an ion meter, either line operated or portable.

1) A ph/mv meter or an ion meter, either line operated or portable. COLE-PARMER Sodium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS The Cole-Parmer Sodium Ion Electrodes are used to quickly, simply, accurately, and economically measure sodium ion concentrations

More information

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic. GENERAL INSTRUCTIONS Introduction The Van London Co. Cyanide Ion Selective Electrode is used to measure dissolved cyanide ion in aqueous solutions. *IMPORTANT NOTICE* Acidic cyanide solutions produce hydrogen

More information

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Calcium Ion Electrodes are used to quickly, simply, accurately, and economically

More information

The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions.

The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions. OAKTON Cadmium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous

More information

COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml) GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Silver/Sulfide Ion Electrode is used to measure silver or sulfide ions in aqueous solutions. The two ions are virtually never present in solution

More information

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

NITRATE ION SELECTIVE ELECTRODE

NITRATE ION SELECTIVE ELECTRODE Instruction Manual and Experiment Guide for the PASCO scientific Model CI-6735 012-06616A 9/97 NITRATE ION SELECTIVE ELECTRODE 1997 PASCO scientific $7.50 10101 Foothills Blvd. P.O. Box 619011 Roseville,

More information

POTASSIUM ION SELECTIVE ELECTRODE

POTASSIUM ION SELECTIVE ELECTRODE Instruction Manual and Experiment Guide for the PASCO scientific Model CI-6733 012-06614A 9/97 POTASSIUM ION SELECTIVE ELECTRODE 1997 PASCO scientific $7.50 10101 Foothills Blvd. P.O. Box 619011 Roseville,

More information

The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in aqueous solutions.

The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in aqueous solutions. OAKTON Lead Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in

More information

phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual 1. A ph/mv meter or an ion meter, either line operated or portable.

phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual 1. A ph/mv meter or an ion meter, either line operated or portable. phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The phoenix Electrode Company Silver/Sulfide Ion Electrodes are used to quickly, simply, accurately,

More information

phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL 1. A ph/mv meter or an ion meter, either line operated or portable.

phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL 1. A ph/mv meter or an ion meter, either line operated or portable. phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The phoenix Electrode Company Carbon Dioxide Gas-Sensing Electrode is used to quickly,

More information

Operating Instructions for Ammonium ISE Specifications

Operating Instructions for Ammonium ISE Specifications Operating Instructions for Ammonium ISE Specifications Range: The Ammonium Electrode responds to uncomplexed ion activity over the range 1 x 10-1M to 1 x 10-6M. Linear detection limit is about 10-5M Interference's:

More information

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0) 150mV 300mV Mono/Reference Combination S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) 1376 340 456 Fax: +44 (0) 1376 340 453 Email:

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London Co. Nitrogen Oxide Gas-Sensing Electrode is used to measure nitrite in aqueous solutions. The measurement is not affected by sample color or turbidity.

More information

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015 Ion Selective Electrodes for the Laboratory Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015 Outline Review ISE measurement technology How to properly calibrate

More information

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking

More information

perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement

perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement Contents Table of Contents 1. Introduction 1 2. Required Equipment 3 3. Electrode and Measurement Setup 4 Electrode

More information

perfection Guidebook perfection Combination Cyanide Electrode Successful Ion Measurement

perfection Guidebook perfection Combination Cyanide Electrode Successful Ion Measurement perfection Guidebook perfection Combination Cyanide Electrode Successful Ion Measurement Contents Table of Contents 1. Introduction 1 2. Required Equipment 3 3. Electrode and Measurement Setup 4 Electrode

More information

CALCIUM SELECTIVE ELECTRODE

CALCIUM SELECTIVE ELECTRODE Instruction Manual and Experiment Guide for the PASCO scientific Model CI-6727 012-06601A 9/97 CALCIUM SELECTIVE ELECTRODE 1997 PASCO scientific $7.50 10101 Foothills Blvd. P.O. Box 619011 Roseville, CA

More information

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution , drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific

More information

phoenix Electrode Company Ammonia Gas-Sensing Electrode Instruction Manual

phoenix Electrode Company Ammonia Gas-Sensing Electrode Instruction Manual phoenix Electrode Company Ammonia Gas-Sensing Electrode Instruction Manual GENERAL INSTRUCTIONS Introduction The phoenix Electrode Company Ammonia Gas-Sensing Electrode is used to quickly, simply, accurately,

More information

User Guide. Perchlorate Ion Selective Electrode

User Guide. Perchlorate Ion Selective Electrode User Guide Perchlorate Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C V 2.8 Revised 1/16/18 ph Probe Silver / silver chloride Reads ph Range 0 14 Resolution Response time Max pressure Max depth Temperature range C Cable length Internal temperature sensor Time before recalibration

More information

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement Contents Table of Contents 1. Introduction 1 2. Required Equipment 2 3. Electrode and Measurement Setup 4 Electrode

More information

Advanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX:

Advanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX: IMPORTANT NOTES FOR FLUORIDE ISE MEASUREMENT SYSTEM WITH 56 ANALYZER IMPORTANT NOTE # 1: This addendum ONLY applies to the 2 nd generation Model 56 ISE Analyzers with the 2.19 software loaded. This addendum

More information

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan APPENDIX 7 PH MEASURING 1. General information PH probe is a unit, measuring the solution acidity or alkalinity degree. It is measured on scale of 0 to 14. The term ph is derived from "p", the mathematical

More information

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049 imagine explore learn Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049 Overview Fluoride is often added to drinking water and to toothpaste to prevent dental decay. In high concentrations,

More information

LEAD ION SELECTIVE ELECTRODE

LEAD ION SELECTIVE ELECTRODE Instruction Manual and Experiment Guide for the PASCO scientific Model CI-6736 012-06617A 8/97 LEAD ION SELECTIVE ELECTRODE 1997 PASCO scientific $7.50 10101 Foothills Blvd. P.O. Box 619011 Roseville,

More information

User Guide. Fluoroborate Ion Selective Electrode

User Guide. Fluoroborate Ion Selective Electrode User Guide Fluoroborate Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051 imagine explore learn Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051 Overview Sodium is highly abundant in the lithosphere (Earth s outer shell) where it occurs as rock salt deposits as

More information

perfection Guidebook perfection Combination Cupric Electrode Successful Ion Measurement

perfection Guidebook perfection Combination Cupric Electrode Successful Ion Measurement perfection Guidebook perfection Combination Cupric Electrode Successful Ion Measurement Contents Table of Contents 1. Introduction 1 2. Required Equipment 3 Introduction 3. Electrode and Measurement Setup

More information

Understanding ph Troubleshooting and Diagnostic information

Understanding ph Troubleshooting and Diagnostic information Understanding ph Troubleshooting and Diagnostic information The desire is to achieve an accurate, reliable measurement with a reasonable electrode life expectancy while minimizing frequency or complexity

More information

Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant

Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant Approved ASTM Method for Fluoride in Drinking Water and Wastewater Analyze free fluoride ions

More information

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth V 2.4 Revised 3/1/18 Spear Tip ph Probe Silver / silver chloride Reads ph Range 0 14 Resolution Response time Max pressure Max depth Temperature range C Cable length Internal temperature sensor Time before

More information

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A imagine explore learn Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A Overview Ammonium ions (NH 4+ ) can be found in fertilizer, cleaning products, processed meats and in the waste of

More information

Ion Selective Electrode Probe

Ion Selective Electrode Probe Rev. 1 Ion Selective Electrode Probe KDS-1064 Calcium KDS-1065 Ammonium KDS-1066 Nitrate KDS-1067 Chloride Type : Ion selective electrode Range : 1 M to 5x10-7 M(40,000 ~ 0.01 ppm) Sensor description The

More information

[Agilent Technologies] [Electrochemistry Products] Quick Reference

[Agilent Technologies] [Electrochemistry Products] Quick Reference Overview Using I9121 Ammonia Gas Sensing Probe on 3200I/M This document demonstrates how to use I 9121 Ammonia Gas Sensing Probe on 3200I or 3200 M to measure pn or the concentration of [NH 4 + ]. Applied

More information

User Guide. Cyanide Ion Selective Electrode

User Guide. Cyanide Ion Selective Electrode User Guide Cyanide Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information

NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC

NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC OPERATING MANUAL ba15362e03 10/2014 NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC NITRATE SENSITIVE ELECTRODE NO 500 NO 800 Note The latest version of the present operating manual can be found on the Internet

More information

User Guide. Bromide Ion Selective Electrode

User Guide. Bromide Ion Selective Electrode User Guide Bromide Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information

Rapid RH Portable ph Meter. User's Manual

Rapid RH Portable ph Meter. User's Manual Rapid RH Portable ph Meter User's Manual This page intentionally left blank. 2 Contents 1. Introduction... 4 2. Included with the Rapid RH Portable ph Meter... 4 3. Understanding ph... 5 4. Calibrating

More information

User Guide. High Performance Ammonia Ion Selective Electrode

User Guide. High Performance Ammonia Ion Selective Electrode User Guide High Performance Ammonia Ion Selective Electrode Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries. AQUAfast, AQUASensors, BOD AutoEZ, ionplus,

More information

Complex equilibria and calculations of formation constants from potentiometric data

Complex equilibria and calculations of formation constants from potentiometric data Experiment 8: /Ag 2 Based Ion elective Electrode Determination of Lead/Hydroxide Equilibria While lead hydroxide does not give a good quantitative measure of lead, particularly in the presence of competing

More information

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste Experiment ISE: Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste 67 You have been hired by the government to check the fluoride concentration labelling on some major

More information

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA , 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine

More information

User Guide. Iodide Ion Selective Electrode

User Guide. Iodide Ion Selective Electrode User Guide Iodide Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information

Guide to ph Analysis.

Guide to ph Analysis. Guide to ph Analysis. XX Why Contents ph Matters Table of Contents Why ph Matters pg 4 What You Need pg 6 How to Measure pg 14 Slope and offset pg 18 Checklist pg 20 Best Practices pg 21 2 Guide to ph

More information

NH 3 -selective electrodes

NH 3 -selective electrodes NH 3 -selective electrodes Manual 8.109.8031EN Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com NH 3 -selective electrodes Manual 8.109.8031EN

More information

User Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode

User Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode User Manual Thermo Scientific Orion Fluoride Ion Selective Electrode 254792-001 Revision B September 2016 Table of Contents Introduction...1 Required Equipment...3 Electrode Setup...5 Electrode Preparation...5

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS Model IQ120 and IQ125 Professional ph Meter with Silicon Chip Sensor (IQ125 additional instructions printed in RED) IQ Scientific Instruments 2075-E Corte del Nogal Carlsbad, CA

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with

More information

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction User s Guide Waterproof Palm ph Meter Model PH220 Introduction Congratulations on your purchase of the Extech PH220 waterproof ph/mv meter. This microprocessor-based device with tactile buttons is battery

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

CHEMICALS & REAGENTS

CHEMICALS & REAGENTS BUFFER SOLUTION, ph 1.68 B-41 3.8 14.45 21.60 64.90 BUFFER SOLUTION, ph 2.00 ± 0.02 B-42 3.8 BUFFER SOLUTION, ph 4.00 ± 0.02 Color coded red BUFFER SOLUTION, COLORLESS, ph 4.00 (ph 4.00 ± 0.02 - Colorless

More information

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Water and Lab Products, Thermo Fisher Scientific Technical Note 505 Key Words Thermo Scientific Orion

More information

NO 2. Orion Nitrite Electrode INSTRUCTION MANUAL. Orion 93-46, Orion 97-46, ionplus. Analyze Detect Measure Control

NO 2. Orion Nitrite Electrode INSTRUCTION MANUAL. Orion 93-46, Orion 97-46, ionplus. Analyze Detect Measure Control NO 2 Orion 93-46, Orion 97-46, ionplus Orion Nitrite Electrode INSTRUCTION MANUAL Analyze Detect Measure Control AQUAfast, Cahn, EZ Flash, Ionalyzer, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT,

More information

PRACTICAL 3 ph AND BUFFERS

PRACTICAL 3 ph AND BUFFERS PRACTICAL 3 ph AND BUFFERS ph and Buffers Structure 3.1 Introduction 3.2 ph and Buffers: Basic Concept 3.2.1 ph 3.2.2 Buffers and Buffer Solutions 3.3 Methods for Determining ph Experiment 1: Measurement

More information

Ammonia Gas Sensing Combination Electrode

Ammonia Gas Sensing Combination Electrode 51927-88 Ammonia Gas Sensing Combination Electrode Model 51927-00 Hach Company, 2001. All rights reserved. Printed in the U.S.A. te/dk 08/01 1ed 2 TABLE OF CONTENTS SPECIFICATIONS...5 SAFETY PRECAUTIONS...7

More information

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

PHS-25CW/3BW Benchtop ph Meter Instruction Manual PHS-25CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides

More information

F- Cl- Br- I- CN- S2- NO3-

F- Cl- Br- I- CN- S2- NO3- Middle of the 60 s First ISE for fluoride Na+ K+ Ag+ Ca2+ NH + 4 Cu2+ Pb2+ F- Cl- Br- I- CN- S2- NO3- Environment Food industry and agriculture Medecine, pharmaceutical, cosmetics Power plants Fluoride

More information

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

ph electrode Instruction Manual For use with the RAH-210 & RPH-250 ph electrode Instruction Manual For use with the RAH-210 & RPH-250 The information in this manual was current at the time of printing. The most current versions of all Hydro Instruments manuals can be

More information

STANDARD OPERATING PROCEDURE

STANDARD OPERATING PROCEDURE Ref.. Table of Contents Page. Scope 2 2. Objective 2 3. Definitions 2 4. Responsibility 3 5. Procedure 3 5.. Diagram of Display and Control Panel 3 5.2. Installation 4 5.3. Operation 4 5.4. Calibration

More information

Ammonia Gas Sensing Combination Electrode

Ammonia Gas Sensing Combination Electrode 51940-88 Ammonia Gas Sensing Combination Electrode Model 50250 Hach Company, 1999. All rights reserved. Printed in the U.S.A. hm/dk 4/99 1ed TRADEMARKS OF HACH COMPANY 2 TABLE OF CONTENTS SPECIFICATIONS...

More information

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows FILL LINE Method 8155 Salicylate Method 1 Powder Pillows Scope and Application: For water, wastewater, and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) (0.01 to 0.50 mg/l NH 3 N) Test Preparation

More information

User Guide. High Performance Ammonia Ion Selective Electrode

User Guide. High Performance Ammonia Ion Selective Electrode User Guide High Performance Ammonia Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT,

More information

User Guide. ROSS Sodium Ion Selective Electrode

User Guide. ROSS Sodium Ion Selective Electrode User Guide ROSS Sodium Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-25CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides

More information

User Guide. Calcium Ion Selective Electrode

User Guide. Calcium Ion Selective Electrode User Guide Calcium Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information

ISO-TECH SYSTEM, VARANASI Analysis by ph and ion selective electrodes

ISO-TECH SYSTEM, VARANASI Analysis by ph and ion selective electrodes ISO-TECH SYSTE, VARANASI-221010 Related link : ph Combination Electrodes Analysis by ph and ion selective electrodes A comparative view of analysis by ph and Ion Selective Electrodes. Types of ISE electrodes.

More information

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014 All about ph! Is it as simple as it seems? The ph Workshop Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014 Common Questions: Measuring ph What is ph? Potential Hydrogen

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater

More information

Chapter 12 Tex-617, Determining Chloride in Concrete

Chapter 12 Tex-617, Determining Chloride in Concrete Chapter 12 Tex-617, Determining Chloride in Contents: Section 1 Overview... 12-2 Section 2 Apparatus... 12-3 Section 3 Preparing Solutions... 12-4 Section 4 Procedures... 12-5 Section 5 Calculations...

More information

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 , MR, 8171 DOC316.53.01069 Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 N) Powder Pillows or AccuVac Ampuls Scope and Application: For water, wastewater and seawater Test preparation

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Quaternary Ammonium Compounds

Quaternary Ammonium Compounds , 8337 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB Powder Pillows Scope and Application: For cooling tower water and pool/spa water Test

More information

QAM-I-111 Operation and Calibration of the Conductivity Meter

QAM-I-111 Operation and Calibration of the Conductivity Meter 1.0 Applicability and Purpose i. This procedure applies to the operation and calibration of the YSI Model 3200 conductivity meter. This procedure is performed prior to any analysis using this meter. By

More information

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow 1. Application Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow Parameters and Codes: Nitrogen, ammonia, dissolved, I-2522-90 (mg/l as N): 00608 Nitrogen, ammonia, total-in-bottom-material,

More information

User Guide. Cupric Ion Selective Electrode

User Guide. Cupric Ion Selective Electrode User Guide Cupric Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,

More information

Applications of Ion-Selective Electrodes

Applications of Ion-Selective Electrodes Applications of Ion-Selective Electrodes Analyte Ammonia Carbon dioxide Chloride Chlorine residual Cyanide Fluoride Nitrate Nitrogen oxide/nitrite Oxygen, dissloved Sulfide Electrode type Gas sensing Gas

More information

Cyanide, colorimetric, pyridine-pyrazolone

Cyanide, colorimetric, pyridine-pyrazolone Cyanide, colorimetric, pyridine-pyrazolone Parameters and Codes: Cyanide, dissolved, I-1300-85 mg/l as CN): 00723 Cyanide, total, I-3300-85 (mgll as CN): 00720 Cyanide, total-in-bottom-material, dry wt,

More information

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV) METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV) 1.0 SCOPE AND APPLICATION 1.1 Method 9012 is used to determine the concentration of inorganic cyanide in an aqueous waste or leachate.

More information

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge Standard Operating Procedure for GLNPO Specific Conductance: Marvin Palmer United States Environmental Protection Agency Great Lakes National Program Office Region 5 77 West Jackson 9th Floor Chicago,

More information

Bante210 Benchtop ph Meter Instruction Manual

Bante210 Benchtop ph Meter Instruction Manual Bante210 Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante210 Benchtop ph Meter 1 Introduction Thank you for selecting the Bante210 benchtop ph meter. This manual provides a step-by-step

More information

NITRITE, Low Range (0 to mg/l NO 2- -N)

NITRITE, Low Range (0 to mg/l NO 2- -N) NITRITE, Low Range (0 to 0.350 mg/l NO 2- -N) Method 8507 For water, wastewater, seawater Diazotization Method * (Powder Pillows or AccuVac Ampuls); USEPA approved for reporting wastewater and drinking

More information

User Guide. Lead Ion Selective Electrode

User Guide. Lead Ion Selective Electrode User Guide Lead Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion, phisa,

More information