Applications of Ion-Selective Electrodes
|
|
- Meagan Martin
- 6 years ago
- Views:
Transcription
1 Applications of Ion-Selective Electrodes Analyte Ammonia Carbon dioxide Chloride Chlorine residual Cyanide Fluoride Nitrate Nitrogen oxide/nitrite Oxygen, dissloved Sulfide Electrode type Gas sensing Gas sensing Solid State Liquid Membrane Combination Solid state Solid state Combination Solid state Liquid membrane Gas sensing Liquid membrane Solid state Ionic strength adjuster ph adjusting ISA CO 2 buffer 5 M NaNO Iodide reagent 5 M NaNO 3 TISAB M (NH 4 )SO 4 Acid buffer SAOB
2 Ion-Selective Electrodes (ISE) Analysis The principle of potentiometric analysis by ISE is based on the Nernst equation which can be written as: E= E 0 + S log C where: E = the measured voltage E 0 = reference potential S = the slope of the electrode C = the concentration of the ion in the solution Different types of sensor electrodes including solid state, liquid membrane, and gas-sensing types are available. The reference electrodes can either be a single junction or double junction type that contains the filling solution of KCl, AgCl, or KNO 3
3 The voltage difference between the sensing and the reference electrodes can be a measure of the concentration of the analyte ion, selective to the sensor electrode. This can be written as follows: C x = C s * 10 E/S Where: C x and C s = the concentrations of the unknown and the standard solutions, respectively. E = the difference of potentials between that of the standardizing solution and the sample solution. S = the slope of the electrode that measures the change in electrode potential per tenfold change in concentration.
4 Standard Addition Method The concentration of the standard can be calculated as follows: Conc. Of the standard to be spiked = Sample volume * Estimated sample conc. Volume of standard to be spiked
5 The concentration, Cx, of the analyte ion in the sample may be calculated from the following equation: Cx = ρcy [(1+ρ) * 10 E/S 1] Where: C y = the concentration of the standard ρ = the ratio of the volume of spiked standard to the volume of sample taken. S = the electrode slope = the millivolt difference, E 2 E 1
6 Example Five ml of 75 ppm chloride standard was added to 100 ml of a wasstewater sample for chloride analysis. The initial millivolt reading for the sample was 17.5 mv. After adding the standard, the reading was 35.3 mv. The slope was measured previously as 58.1 mv. Calculate the concentration of chloride in the sample.
7 Sample Addition Method Conc. Of the standard = Volume of sample to be spiked * Estimated sample conc. Volume of standard taken
8 Determine the concentration Cx of the analyte in the sample from the following equation: C x = [(1+ρ) * 10 E/S ρ] * C y Where: C y = the concentration of the standard ρ = the ratio of the volume of the standard taken to the volume of sample spiked E = the millivolt difference E 2 E 1 S = electrode slope
9 A wastewater sample was found to contain 3.8 mg/l cyanide. A 100 ml aliquot of this sample was spiked with 10 ml of 50 mg/l cyanide standard solution. The concentration of this spiked solution was measured to be 8.1 mg/l.
10 C y = 75 ppm ρ= 5 ml = ml E = 35.3 mv 17.5 mv= 17.8 mv S = 58.1 mv C x = 0.05 * 75 ppm [(1+0.05) * mv/58.1 mv 1] = 3.75 ppm [(1.05 * ) 1] = 3.75 ppm [(1.05 * 2.02) 1] = 3.75 ppm = 3.34 ppm
11 Relative Percent Deviation (RPD) RPD = ( a 1 a2) or( a2 ( a1 + a2) 2 a 1 ) *100% Where a 1 and a 2 are the results of duplicate analysis of a sample
12 RPD = (11.1mg (11.1mg / / L L + 9.7mg 9.7mg 2 / / L) L) *100% = (1.4mg / L) 10.4mg / L *100% = 13.5
13 s = x 2 n ( 1 n x ) 2 Where Σx 2 = sum of the squares of individual measurements Σx = sum of individual measurements, and n = number of individual measurements RSD = s x *100 %
14 The sum of x = 30.5 The sum of x 2 = Σx = 30.5 Σx 2 = (Σx) 2 = (30.5) 2 = ; n = 6 s = mg/l = 2.86 mg/l = = 0.29mg = 0.29 mg
15 % Recovery = 100(X s X u ) / K Where X s = measured value for the spiked sample X u = measured value for the unspiked sample adjusted for dilution of the spike, and K = known value of the spike in the sample
16 % Recovery = Measured concentration Theoretical concentration * 100% Theoretical concentration can be calculated as: (C u *V u ) (C s * V s ) + (V u + V s ) (V u + V s ) Where C u = measured concentration of the unknown sample, and C s = concentration of the standard While V u and V s are the volumes of the unknown sample and standard, respectively
CHEMICALS & REAGENTS
BUFFER SOLUTION, ph 1.68 B-41 3.8 14.45 21.60 64.90 BUFFER SOLUTION, ph 2.00 ± 0.02 B-42 3.8 BUFFER SOLUTION, ph 4.00 ± 0.02 Color coded red BUFFER SOLUTION, COLORLESS, ph 4.00 (ph 4.00 ± 0.02 - Colorless
More informationMETHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE
METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking
More informationISO-TECH SYSTEM, VARANASI Analysis by ph and ion selective electrodes
ISO-TECH SYSTE, VARANASI-221010 Related link : ph Combination Electrodes Analysis by ph and ion selective electrodes A comparative view of analysis by ph and Ion Selective Electrodes. Types of ISE electrodes.
More informationMethod Abstract. Flow Solution Fluoride, USGS by ISE and FIA
Scope Summary Interferences Fluoride, USGS by ISE and FIA This method is used for the determination of fluoride in drinking water, surface water, and domestic and industrial wastes according to USGS Method
More informationWater & Wastewater. Laboratory Math
Water & Wastewater Laboratory Math VWEA Good Lab Practices Meeting Presented by: Wayne Staples Virginia DEQ July 24, 2017 Required to: Math in the Laboratory Determine a test result Alkalinity calculation
More informationChem 321 Lecture 17 - Potentiometry 10/24/13
Student Learning Objectives Chem 321 Lecture 17 - Potentiometry 10/24/13 Electrodes The cell described in the potentiometric chloride titration (see 10/22/13 posting) consists of a Ag/AgCl reference electrode
More informationThermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant
Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant Approved ASTM Method for Fluoride in Drinking Water and Wastewater Analyze free fluoride ions
More informationF- Cl- Br- I- CN- S2- NO3-
Middle of the 60 s First ISE for fluoride Na+ K+ Ag+ Ca2+ NH + 4 Cu2+ Pb2+ F- Cl- Br- I- CN- S2- NO3- Environment Food industry and agriculture Medecine, pharmaceutical, cosmetics Power plants Fluoride
More informationThermo Scientific Orion Solutions, Standards and Buffers. The high-quality solutions you need for electrochemistry measurement
Thermo Scientific Orion Solutions, Standards and Buffers The high-quality solutions you need for electrochemistry measurement High-quality solutions deliver correct measurements every time Whether in the
More informationIon Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015
Ion Selective Electrodes for the Laboratory Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015 Outline Review ISE measurement technology How to properly calibrate
More informationDirect Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution
, drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific
More informationCOLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL
COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Industrial Nitrate Ion Electrode is used to quickly, simply, accurately, and economically
More informationCOLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL
COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com
More informationCOLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL
COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com
More informationUnit 3: Chemical Equilibrium Chemistry Write balanced chemical equations for each of the following. Pay close attention to the physical states!
Practice Questions Section. The Equilibrium Constant 1. Write balanced chemical equations for each of the following. Pay close attention to the physical states! Also - you must include the charge when
More informationNanotechnology Sensors
ESILAB PACK : 1 Ion + ph Nanotechnology Sensors email : contact@water-kits.fr This revolutionary technology based on carbon nanotubes hugely simplifies the potentiometric measurements. Measurement & calibration
More informationIon Selective Electrodes
Ion Selective Electrodes A n a l y t i c a l S e n s o r s & I n s t r u m e n t s I O N S E L E C T I V E E L E C T R O D Ion Selective Electrodes (ISE) are membrane electrodes that respond selectively
More informationION SELECTIVE MEASUREMENT
Electrodes Solutions Instrumentation ION SELECTIVE MEASUREMENT catalog featuring our full line of Ion Selective Electrodes Introduction Hanna Instruments Ion Selective Measurement Catalog Authorized Distributor
More information1225 Lab # 11 Fluoride Speciation and Analysis by Flow Injection using Ion-Selective Electrode: Measurement of Total Fluoride in Water
1225 Lab # 11 Fluoride Speciation and Analysis by Flow Injection using Ion-Selective Electrode: Measurement of Total Fluoride in Water Introduction Fluoride is recognized world- wide to help build stronger
More information3.2.5 Group VII. Trends in Physical Processes. 70 minutes. 70 marks. Page 1 of 7
3.2.5 Group VII Trends in Physical Processes 70 minutes 70 marks Page 1 of 7 Q1. (a) State and explain the trend in electronegativity down Group VII from fluorine to iodine. Trend... Explanation... (i)
More informationCOLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL
GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Ammonium Ion Electrode is used to quickly, simply, accurately, and economically
More informationphoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual 1. A ph/mv meter or an ion meter, either line operated or portable.
phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The phoenix Electrode Company Silver/Sulfide Ion Electrodes are used to quickly, simply, accurately,
More informationCOLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL
COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com
More informationIon Selective Electrode Probe
Rev. 1 Ion Selective Electrode Probe KDS-1064 Calcium KDS-1065 Ammonium KDS-1066 Nitrate KDS-1067 Chloride Type : Ion selective electrode Range : 1 M to 5x10-7 M(40,000 ~ 0.01 ppm) Sensor description The
More informationExperiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University
Experiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University Adapted from "Experiment 3: Analysis of Phosphate in Water," Laboratory Experiments
More informationMeasurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste
Experiment ISE: Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste 67 You have been hired by the government to check the fluoride concentration labelling on some major
More informationCOLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL
COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com
More information2. Which of the following statements best describes the movement of electrons in an electrochemical cell?
Exam 2 Chem 311 Evans Fall 2009 112: 2 pts each 1. Consider the following unbalanced redox equation: Pb (s) + PbO 2 (s) + 2 HSO 4 (aq) 2 PbSO 4 (s) Which species is being oxidized? A. HSO 4 B. Pb(s) C.
More informationDetermination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)
Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with
More informationNomenclature Naming Ionic Compounds Worksheet #1
Naming Ionic Compounds Worksheet #1 In forming ionic compounds with non-metals, the transition metals often exhibit more than one valence. For example, in the reaction between iron and chlorine, two products
More informationWhat is a WQSensor? Software Installation. Uninstalling WQSensors Software. NexSens Technology, Inc. TABLE OF CONTENTS
TABLE OF CONTENTS 1 Sensor Operation...3 1.1 Ion Selective Electrode Sensors...3 1.1.1 WQ-FL Fluoride ISE Sensor... 3 1.1.2 Making ISE Sensor Measurements... 4 1.1.3 Maintenance and Care... 8 1.1.4 ISE
More information(a) Complete Figure 9 by placing one tick in each row to show whether the salt is soluble or insoluble. salt soluble insoluble.
1 The method used to prepare a salt depends on its solubility in water. (a) Complete Figure 9 by placing one tick in each row to show whether the salt is soluble or insoluble. ammonium chloride salt soluble
More informationNa+ K+ Ag+ Ca2+ NH + 4. F- Cl- Br- I- CN- S2- NO3-
1 1 HISTORY 2 Middle of the 60 s First ISE for fluoride Na+ K+ Ag+ Ca2+ NH + 4 Cu2+ Pb2+ F- Cl- Br- I- CN- S2- NO3- Until the 60 s the only ISE is in fact the ph electrode with a glass sensitive to protons.
More informationApplication Note #119 Alkalinity and Chloride by Titration
Application Note #119 Alkalinity and Chloride by Titration Updated July 2017 Introduction: The alkalinity of water may be defined as its capacity to neutralize acid. Alkali substances in water include
More informationCP Chapter 15/16 Solutions What Are Solutions?
CP Chapter 15/16 Solutions What Are Solutions? What is a solution? A solution is uniform that may contain solids, liquids, or gases. Known as a mixture Solution = + o Solvent The substance in abundance
More informationCOLE-PARMER INDUSTRIAL SILVER/SULFIDE ION ELECTRODE INSTRUCTION MANUAL
GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL SILVER/SULFIDE ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Silver/sulfide Ion Electrode is used to quickly, simply, accurately,
More informationChapter 12 Tex-617, Determining Chloride in Concrete
Chapter 12 Tex-617, Determining Chloride in Contents: Section 1 Overview... 12-2 Section 2 Apparatus... 12-3 Section 3 Preparing Solutions... 12-4 Section 4 Procedures... 12-5 Section 5 Calculations...
More informationWhich of the following answers is correct and has the correct number of significant figures?
Avogadro s Number, N A = 6.022 10 23 1. [7 points] Carry out the following mathematical operation: 6.06 10 3 + 1.1 10 2 Which of the following answers is correct and has the correct number of significant
More informationWEEK 2 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY
University of Puget Sound Department of Chemistry Chem 230 WEEK 2 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY INTRODUCTION Nitrogen is an essential nutrient
More informationApplication Bulletin
Of interest to: General analytical laboratories A 1, 2, 4, 7, 10, 11, 12, 13 Chloride titrations with potentiometric indication Summary Besides acid/base titrations, the titrimetric determination of chloride
More informationUser Guide. High Performance Ammonia Ion Selective Electrode
User Guide High Performance Ammonia Ion Selective Electrode Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries. AQUAfast, AQUASensors, BOD AutoEZ, ionplus,
More informationThe Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions.
OAKTON Cadmium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous
More informationCyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0
Metrosep A Supp 10-100 /4.0 Branch Environment, Food, Beverages Keywords IC; 850; 858; Metrosep A Supp 10-100/4.0; Cyanide; Sulfide; 2.850.9110; DC Mode Summary The determination of sulfide and cyanide
More informationShow your work for all questions; answer all parts of all questions. No work = no credit.
Quiz: Ch 3 & 4 Name: Version M (32 pts) September 16, 2005 AP Chem Period: 1 2 3 4 Show your work for all questions; answer all parts of all questions. No work = no credit. 1. (10 pts) A compound contains
More informationAnswer Sheet for Sample Problems for Chemistry Level 1 Final Exam 2016 Study Guide
Answer Sheet for Sample Problems for Chemistry Level 1 Final Exam 2016 Study Guide Electrons in Atoms Chapter 5 1. What is the frequency of green light, which has a wavelength of 4.90 x 10-7 m? 8 c 3.00x10
More informationPart A Answer all questions in this part.
Part A Directions (1-24): For each statement or question, record on your separate answer sheet the number of the word or expression that, of those given, best completes the statement or answers the question.
More informationA New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores
Application Note 37890312 A New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores Keywords CNSolution 9310 On-line Cyanide Analyzer Cyanidation
More informationChem 321 Name Answer Key D. Miller
1. For a reversed-phase chromatography experiment, it is noted that the retention time of an analyte decreases as the percent of acetonitrile (CH 3 CN) increases in a CH 3 CN/H 2 O mobile phase. Explain
More informationMethod (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows
Chlorine Free, Indophenol, 10241 Indophenol 1 1 Patent pending. DOC316.53.01256 Method 10241 (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Scope and Application: For determining residual free chlorine levels
More informationMETHOD 9076 DETERMINATION OF INORGANIC ANIONS BY ION CHROMATOGRAPHY PRODUCTS BY OXIDATIVE COMBUSTION AND MICROCOULOMETRY
METHOD 9076 DETERMINATION OF INORGANIC ANIONS BY ION CHROMATOGRAPHY PRODUCTS BY OXIDATIVE COMBUSTION AND MICROCOULOMETRY 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total
More informationHow did JJ Thomson conclude that the mobile charged particle in the atom had a ( ) charge.
Name Veritas Class Period Chemistry: Final Exam Practice Problems The final exam will focus on material covered in the spring semester. However, note that much of the material learned early in the year
More informationAmmonia Gas Sensing Combination Electrode
51927-88 Ammonia Gas Sensing Combination Electrode Model 51927-00 Hach Company, 2001. All rights reserved. Printed in the U.S.A. te/dk 08/01 1ed 2 TABLE OF CONTENTS SPECIFICATIONS...5 SAFETY PRECAUTIONS...7
More informationExam 2 Chem 311 Fall 2002 You must do (1 or 2), (3 or 4), (5 or 6), (7 or 8), and one other problem!!! All problems are worth 20 points
Exam 2 Chem 311 Fall 2002 You must do (1 or 2), (3 or 4), (5 or 6), (7 or 8), and one other problem!!! All problems are worth 20 points 1) A 100.0 ml solution containing 0.0100 M (NH 4 ) 2 Ce(NO 3 ) 6,
More informationDetermination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)
Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with
More informationPotentiometry: Ion-Selective Electrodes
Potentiometry: Ion-Selective Electrodes Determination of the sodium monofluorophosphate content of a toothpaste sample using a fluoride ion-selective electrode. 1. Aims The aims of the experiment are to:
More informationFor Practice 4.1 Magnesium hydroxide, the active ingredient in milk of magnesia, neutralizes stomach acid, primarily HCl, according to the reaction:
Stoichiometry For Practice 4.1 Magnesium hydroxide, the active ingredient in milk of magnesia, neutralizes stomach acid, primarily HCl, according to the reaction: What mass of HCl, in grams, is neutralized
More informationDetermination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)
Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with
More informationwater proof water proof Tap water, pure water application Precious, trace amount sample 3-in-1 Plastic Electrode D
ELECTRODE LINE UP electrode just got Touer! Enhanced durability of the glass membrane makes it four times stronger than the conventional Tou electrode. Response glass NEW The patented silver ion trap technology
More informationAutomated Photometry. Presented by Scott Tucker. Instruments
Application Breakthroughs in Automated Photometry Presented by Scott Tucker Application Chemist, Lachat Instruments Introduction Advantages of automated photometry FIAversus other techniques Lachat Application
More informationForm A. Exam 1, Ch 1-4 September 23, Points
Chem 130 Name Exam 1, Ch 1-4 September 23, 2011 100 Points Please follow the instructions for each section of the exam. Show your work on all mathematical problems. Provide answers with the correct units
More informationNomenclature for ionic compounds
Name: Nomenclature for ionic compounds Nomenclature is a system of naming. This worksheet presents a widely used system of nomenclature for ionic compounds. There are two types of metal cations with different
More informationEXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY
University of Puget Sound Department of Chemistry Chem 230 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY INTRODUCTION Nitrogen is an essential nutrient
More informationUser Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode
User Manual Thermo Scientific Orion Fluoride Ion Selective Electrode 254792-001 Revision B September 2016 Table of Contents Introduction...1 Required Equipment...3 Electrode Setup...5 Electrode Preparation...5
More informationCHAPTER 14: ELECTRODES AND POTENTIOMETRY
CHAPTER 14: ELECTRODES AND POTENTIOMETRY Chapter 14 Electrodes and Potentiometry Potentiometry : The use of electrodes to measure voltages that provide chemical information. (The cell voltage tells us
More informationCOLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL
COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com
More informationAmmonia Gas Sensing Combination Electrode
51940-88 Ammonia Gas Sensing Combination Electrode Model 50250 Hach Company, 1999. All rights reserved. Printed in the U.S.A. hm/dk 4/99 1ed TRADEMARKS OF HACH COMPANY 2 TABLE OF CONTENTS SPECIFICATIONS...
More informationGENERAL INSTRUCTIONS GENERAL PREPARATION
GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Fluoroborate Ion Selective Electrode is used to quickly, simply, accurately, and economically measure Fluoroborate in aqueous solutions.
More informationUse of Ion Selective Probes
World Bank & Government of The Netherlands funded Training module # WQ - 33 Use of Ion Selective Probes New Delhi, February 2000 CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas, New Delhi 11 00 16
More informationMolecule smallest particle of a substance having its chemical properties Atoms connected via covalent bonds Examples:
Ionic equations, calculations involving concentrations, stoichiometry MUDr. Jan Pláteník, PhD Molecule smallest particle of a substance having its chemical properties Atoms connected via covalent bonds
More informationElectro Analytical Methods
CH 2252 Instrumental Methods of Analysis Unit II Electro Analytical Methods Dr. M. Subramanian Associate Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam
More informationComplex equilibria and calculations of formation constants from potentiometric data
Experiment 8: /Ag 2 Based Ion elective Electrode Determination of Lead/Hydroxide Equilibria While lead hydroxide does not give a good quantitative measure of lead, particularly in the presence of competing
More informationName: Unit 9- Stoichiometry Day Page # Description IC/HW
Name: Unit 9- Stoichiometry Day Page # Description IC/HW Due Date Completed ALL 2 Warm-up IC 1 3 Stoichiometry Notes IC 1 4 Mole Map IC X 1 5 Mole to Mole Practice IC 1 6 Mass to Mole Practice IC 1/2 X
More informationAuthorized Distributor Clarkson Laboratory & Supply Inc. Phone Fax
Table of Contents Part A. General Information Preface should be read first Who is Weiss Research? Company Mission Consultant Dr. K. L. Chang Sales Terms & Conditions How to Order Technical Support Promises
More informationPreparation and Characterization of Home-Made Silver Sulphide Based Chloride Selective Electrode
Armila Rajbhandari et al Available online at www.pharmaresearchlibrary.com/ijcps ISSN: 2321-3132 Research Article International Journal of Chemistry and Pharmaceutical Sciences www.pharmaresearchlibrary.com/ijcps
More informationAP Chemistry Summer Assignment
AP Chemistry Summer Assignment Due Date: Thursday, September 1 st, 2011 Directions: Show all of your work for full credit. Include units and labels. Record answers to the correct number of significant
More informationC = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)
GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Silver/Sulfide Ion Electrode is used to measure silver or sulfide ions in aqueous solutions. The two ions are virtually never present in solution
More information1. What is the mass percent of sulfur in Al 2 (SO 4 ) 3? A % C % B % D %
1. What is the mass percent of sulfur in Al 2 (SO 4 ) 3? A. 9.372 % C. 28.12 % B. 21.38 % D. 42.73 % 2. How many grams of phosphorus are in 35.70 g of P 2 O 5? A. 6.359 g C. 15.58 g B. 23.37 g D. 31.16
More informationIon-selective Measurement (ISE)
Ion-selective Measurement (ISE) Page 1 of 39 Table of contents 1 Basic information 1.1 Introduction to ion-selective measurement 1.2 Reference electrode design and principle of operation 1.3 Concentration
More information3. Monoatomic anions of an element are identified by the element s name. 4. A binary compound is made of two or more different elements.
CHAPTER 7 HOMEWORK 7-1 (pp. 203 206) Write true or false for each statement. 1. A chemical formula indicates the relative number of molecules of each kind in a chemical compound. 2. Monoatomic ions are
More informationALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82
ALLOWAY Method Outline Standard Laboratory Method: SM4500-F - B and SM4500-F-C Parameter: Total Fluoride via Distillation Method: Distillation then Specific Ion Date Issued: 01/04/82 Date Revised: 04/09/02,
More informationUnit-8 Equilibrium. Rate of reaction: Consider the following chemical reactions:
Unit-8 Equilibrium Rate of reaction: Consider the following chemical reactions: 1. The solution of sodium chloride and silver nitrate when mixed, there is an instantaneous formation of a precipitate of
More informationUnit 6 Problem Set Which pair of formulas represents the empirical formula and the molecular formula of a compound?
Unit 6 Problem Set - 2015 Name: ate: 1. Which pair of formulas represents the empirical formula and the molecular formula of a compound? 1.. H 2 O, 4 H 6 O 4. HO, 6 H 12 O 6. H 4, 5 H 12. H 2, 3 H 6 2.
More informationChapter 12. Answers to Questions. 1. (a) nitrogen dioxide (b) hydrogen sulfide (c) hydrogen chloride
Chapter 12 Answers to Questions 1. (a) nitrogen dioxide (b) hydrogen sulfide (c) hydrogen chloride 2. (a) hydrogen cyanide (b) dinitrogen oxide (c) ammonia 3. X = sulfur, Y = oxygen, Z = sulfur dioxide
More informationISEmax CAM40/CAS40. Technical Information
Technical Information Online measurement of nutrient parameters Ion-selective electrode system for the continuous measurement of ammonium and nitrate Application The ion-selective electrode system works
More informationOperating Instructions for Ammonium ISE Specifications
Operating Instructions for Ammonium ISE Specifications Range: The Ammonium Electrode responds to uncomplexed ion activity over the range 1 x 10-1M to 1 x 10-6M. Linear detection limit is about 10-5M Interference's:
More informationMethod to 0.50 mg/l NH 3 N Powder Pillows
, 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows
More informationS ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)
150mV 300mV Mono/Reference Combination S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) 1376 340 456 Fax: +44 (0) 1376 340 453 Email:
More informationUser Guide. High Performance Ammonia Ion Selective Electrode
User Guide High Performance Ammonia Ion Selective Electrode ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT,
More informationThe Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.
COLE-PARMER NITRATE ION ELECTRODES INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate
More informationChapter Six Chemical Names and Formulas WS C U1C6
Chapter Six Chemical Names and Formulas WS C U1C6 Name Period Section 6.1 Part I: Matching. Match the definition with the term that best correlates to it. No definition will be used more than once. 1.
More informationIon-selective Measurements
Ion-selective Measurements Ion-selective measurement is a method for determining the concentration of dissolved ions. Potassium ions, sodium ions, fluoride or chloride are examples of such cations and
More informationTimberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis
Timberline Ammonia-001 Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis Prepared By: Dr. Edward F. Askew Date: June 24, 2011 Table of Contents 1. SCOPE
More informationUnit 2 Electrochemical methods of Analysis
Unit 2 Electrochemical methods of Analysis Recall from Freshman Chemistry: Oxidation: Loss of electrons or increase in the oxidation number Fe 2 e - Fe 3 Reduction: Gain of electrons or decreases in the
More informationChemistry 51 Chapter 8 TYPES OF SOLUTIONS. Some Examples of Solutions. Type Example Solute Solvent Gas in gas Air Oxygen (gas) Nitrogen (gas)
TYPES OF SOLUTIONS A solution is a homogeneous mixture of two substances: a solute and a solvent. Solute: substance being dissolved; present in lesser amount. Solvent: substance doing the dissolving; present
More informationIntroduction. Terminology
Use of ASTM D 7284-08 Standard Test Method for Total Cyanide in Water by Micro Distillation followed by Flow Injection Analysis with Gas Diffusion Separation and Amperometric Detection Introduction A new
More informationPhysical Properties of Solutions
Physical Properties of Solutions Physical Properties of Solutions Types of Solutions (13.1) A Molecular View of the Solution Process (13.2) Concentration Units (13.3) Effect of Temperature on Solubility
More informationUnit 9 Stoichiometry Notes
Unit 9 Stoichiometry Notes Stoichiometry is a big word for a process that chemist s use to calculate amounts in reactions. It makes use of the coefficient ratio set up by balanced reaction equations to
More informationSummer Assignment Part 2
Summer Assignment Part 2 Name: 1. Metric Conversions. Remember 1 cm 3 = 1 ml 1 L = 1 dm 3 ITEM GIVEN METRIC UNIT DESIRED METRIC UNIT A 8.43 cm mm B 2.41 x 10 2 cm m C 294.5 nm cm D 1.445 x 10 4 m km E
More informationMethod to 0.50 mg/l NH 3 N Powder Pillows
, 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows
More informationFunsheet 3.0 [WRITING & BALANCING EQUATIONS] Gu/R. 2017
Funsheet 3.0 [WRITING & BALANCING EQUATIONS] Gu/R. 2017 Balance the following chemical equations. Remember, it is not necessary to write "1" if the coefficient is one. 1. N 2 + H 2 NH 3 2. KClO 3 KCl +
More information