Proper Quality Assurance Means for Radon Monitoring Devices - A Convenient and Accurate Method for Routine Calibration Traceable to NIST, USA

Size: px
Start display at page:

Download "Proper Quality Assurance Means for Radon Monitoring Devices - A Convenient and Accurate Method for Routine Calibration Traceable to NIST, USA"

Transcription

1 Proper Quality ssurance Means for Radon Monitoring Devices - Convenient and ccurate Method for Routine Calibration Traceable to NIST, US Walter Bürkin Genitron Instruments GmbH, Frankfurt, Germany wbuerkin@genitron.de bstract. convenient method for routine calibration of radon devices traceable to the US National Institute of Standards (NIST) is presented. s an appropriate Q measure and with a minimum of auxiliary equipment the user himself can perform periodic calibration checks in order to maintain accuracy and reliability of applied radon measurement devices on a high level. 1. Introduction Performing reliable measurements of radon in homes and workplaces which are in compliance with given regulations require the application of measuring devices disposing of stable and proven accuracy. To ensure this accuracy on a permanent basis periodic verification of the applied calibration constants is required. Consecutively a radon calibration method is described which serves for periodic testing of active radon devices in order to trace their calibration back to a radon standard provided by the United States National Institute of Standards and Technology (NIST). The method consists in a relatively simple procedure which can be carried out by the user himself without a lot of sophisticated laboratory instrumentation. 1. Hardware Requirements Following equipment is required: (1) NIST Standard Reference Material stored in an Emanation Bulb (see FIG. 1) radon source provided by NIST, type SRM 4968 consisting of a small polyethylene-capsulated radium-226 solution with known (certified) activity (~ 400 Bq each) and emanation of radon-222 gas (~ 90 %) [1]. This standard reference material serves as primary reference for calibrating the radon device (reference unit) traceable to NIST, US. glass bulb is used for storing the capsule and for undisturbed radon accumulation maintaining inside the vessel a 100 % humidity atmosphere. FIG. 1 (2) Emanation & Calibration Container (see FIG. 2) gas tight container made of stainless steal serves for retaining the reference monitor to be calibrated. Two leak tested valves allow the exchange of gas e.g. injecting the required radon concentration into the container. FIG. 2 1

2 (3) Gas tight electronic pump (see FIG. 3) gas proof electronic pump is required to transfer the accumulated radon gas from the glass bulb into the container without gas loss and dilution respectively. FIG. 3 (4) Radon Monitor (see FIG. 4) s reference for calibrating other active and passive radon measurement detectors a reliable radon monitor with accurate calibration is needed. Here a device of type lphagurd which disposes of a long-time stable calibration factor and of a highly sophisticated inbuilt Q-System is applied [2]. FIG Calibration Procedure One or more radon devices are exposed within the calibration container. fter reducing and determining the radon background the radon which has been released from the NIST standard material and accumulated within the glass bulb is transferred into the vessel 1) (see FIG. 5). FIG. 5. Setup for transferring the radon activity released by a standard source to the inner of the container in order to calibrate an active radon measurement device traceable to NIST. 1) Recommendation: To avoid the need of additional mathematical corrections the period of undisturbed built-up process of radon should last 3 weeks minimum. 2

3 4. Evaluation of Measurement For data evaluation following steps have to be carried out: (1) Data of the calibration experiment have to be read out and displayed using DataEXPERT software (see FIG. 6). (2) The background concentration has to be defined by averaging the radon concentration over the period of low-level measurement (see FIG. 6). Background / Mean radon concentration FIG. 6. Data series of radon concentration covering background and calibration measurement. 3

4 (3) synthetic straight line is fitted in the data curve representing the radon decay function (see FIG. 7). FIG. 7. Radon data of calibration experiment with fit in of decay function (4) The initial radon concentration (C Rn m ) at the time the radon was separated from the standard source and injected into the container can now be defined. This is obtained by extrapolation using previously created decay function (see FIG. 8). C Rn m FIG. 8. Radon data of calibration experiment with zoom in and determination of initial radon concentration C Rn m 4

5 5. Calculation of Deviation In order to obtain the deviation of the calibration factor of the tested device the measured initial radon concentration (C Rn m ) has to be referred to the actually released activity of the standard material. This has to be done in three steps (1) Calculation of actually released activity from the NIST material S ( SRM ) = SRM f (1) S(SRM) : Emanated activity by NIST SRM [Bq] SRM : Certified activity of NIST SRM [Bq] f : Emanation fraction (2) Converting measured radon concentration to activity Rn Cal ( C C ) ( V nv ) = 0 (2) Rn m Rn Con lgu Rn Cal : Measured net activity [Bq] C Rn m : Initial radon concentration [Bq/m³] at the time point the radon gas was injected into the container C Rn 0 : Measured background radon concentration [Bq/m³] V Con : Volume of container [m³] n : Number of measurement devices exposed within the container V lgu : Volume of one measurement device [m³] (here lphagurd : V=0,00255m³) (3) Defining Deviation in ctivity = : Deviation [%] Rn Cal S ( SRM ) S ( SRM ) 100% (3) The realized result of above calculations complies with the variation in calibration of the tested radon device to the NIST standard reference material SRM References 1. Collé R., Hutchinson J.M.R., Unterweger M.P., The NIST Primary Radon-222 Measurement System, in J. Res. Nat. Inst. Stds. Tech. 95, (1990) 2. Genrich V., lphagurd PQ2000 / MC50, Multiparameter Radon Monitor, Characterization of its physical properties under normal climatic and severe environmental conditions, (Genitron Instruments 1994) 5

NIST TRACEABLE RADON CALIBRATION SYSTEM FOR CALIBRATING TRUE INTEGRATING RADON MONITORS -

NIST TRACEABLE RADON CALIBRATION SYSTEM FOR CALIBRATING TRUE INTEGRATING RADON MONITORS - NIST TRACEABLE RADON CALIBRATION SYSTEM FOR CALIBRATING TRUE INTEGRATING RADON MONITORS - E-PERM@* P. Kotrappa and L.R. S tieff Rad Elec Inc. Frederick, MD ABSTRACT The NIST (National Institute of Standards

More information

W T RADON EMANATION STANDARD - SRM 4968 THEORETICAL AND PRACTICAL ASPECTS OF IT'S APPLICATIONS

W T RADON EMANATION STANDARD - SRM 4968 THEORETICAL AND PRACTICAL ASPECTS OF IT'S APPLICATIONS W T RADON EMANATION STANDARD - SRM 4968 THEORETICAL AND PRACTICAL ASPECTS OF IT'S APPLICATIONS Paul Kotrappa Rod Bee,Inc. Frederick, MD ABSTRACT Several years of research in developing a radon emanation

More information

Radon-Thoron mixed atmosphere: realization, characterization, monitoring and use for detector calibration.

Radon-Thoron mixed atmosphere: realization, characterization, monitoring and use for detector calibration. Radon-Thoron mixed atmosphere: realization, characterization, monitoring and use for detector calibration. Raffaele Buompane Dipartimento di Matematica e Fisica, Università degli Studi della Campania Luigi

More information

White Paper. Perform Conductivity Measurements In Compliance with USP <645>

White Paper. Perform Conductivity Measurements In Compliance with USP <645> Perform Conductivity Measurements In Compliance with USP Water is the most widely used substance, raw material, or ingredient in the production, processing and formulation of compendial articles.

More information

Natural Rock Sample Manual

Natural Rock Sample Manual Natural Rock Sample Manual Revision 2014-06-12 DURRIDGE Company Inc. 524 Boston Road Billerica, MA 01821 Tel: (978) 667-9556 Fax: (978) 667-9557 service@durridge.com www.durridge.com 2014, DURRIDGE Company

More information

RESPONSE OF A RADON CHARCOAL CANNISTER TO CLIMATIC AND RADON VARIATIONS IN THE INTE RADON CHAMBER. A. Vargas, X. Ortega, I.

RESPONSE OF A RADON CHARCOAL CANNISTER TO CLIMATIC AND RADON VARIATIONS IN THE INTE RADON CHAMBER. A. Vargas, X. Ortega, I. RESPONSE OF A RADON CHARCOAL CANNISTER TO CLIMATIC AND RADON VARIATIONS IN THE INTE RADON CHAMBER A. Vargas, X. Ortega, I. Serrano Institut de Tècniques Energètiques (INTE), Universitat Politècnica de

More information

CALIBRATION OF INSTRUMENTS MEASURING RADON OVER A LARGE ACTIVITY RANGE. N. Michielsen, V. Voisin

CALIBRATION OF INSTRUMENTS MEASURING RADON OVER A LARGE ACTIVITY RANGE. N. Michielsen, V. Voisin Radon in the Living Environment, 011 CALIBRATION OF INSTRUMENTS MEASURING RADON OVER A LARGE ACTIVITY RANGE N. Michielsen, V. Voisin Institut de Protection et de Sûreté Nucléaire, Département de Prévention

More information

NATURAL ROCK SAMPLE. Performance Monitoring Accessory for the RAD7 User Manual

NATURAL ROCK SAMPLE. Performance Monitoring Accessory for the RAD7 User Manual NATURAL ROCK SAMPLE Performance Monitoring Accessory for the RAD7 User Manual TABLE OF CONTENTS TABLE OF CONTENTS 2 1 INTRODUCTION 3 2 RAD7 RADON SENSITIVITY CHECK 4 2.1 Purge the RAD7 5 2.2 Hook up to

More information

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES The National Association of Testing Authorities (NATA) requirements for accreditation have undergone

More information

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP. NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE ENVIRONMENTAL LABORATORY CERTIFICATION PROGRAM ON-SITE LABORATORY EVALUATION RADIOCHEMISTRY PROCEDURES Gross Alpha-Gross Beta

More information

Ion Chamber. Radon Measurements. Theremino System Rev.1. Theremino System IonChamber_ENG Page 1

Ion Chamber. Radon Measurements. Theremino System Rev.1. Theremino System IonChamber_ENG Page 1 Ion Chamber Radon Measurements Theremino System Rev.1 Theremino System IonChamber_ENG Page 1 Table of Contents Misure con Camera a Ioni... 3 Theory... 3 Equipment... 3 Radon in Buildings - Rn 222... 4

More information

Traceable Mass Determination and Uncertainty Calculation

Traceable Mass Determination and Uncertainty Calculation Traceable Mass Determination and Uncertainty Calculation Falko Hilbrunner, Thomas Fehling, Siego Mühlich, Mario Schreiber Sartorius Lab Instruments GmbH Weender Landstrasse 94-108 Goettingen, Germany ABSTRACT

More information

Mercury Stack Monitor SM-4

Mercury Stack Monitor SM-4 Member of the envea Group Mercury Stack Monitor SM-4 Automatic continuous emission monitor (CEM) for mercury Continuous operation Sample dilution directly at stack - works with virtually any sample matrix

More information

Experiment Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado

Experiment Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado Experiment 10 1 Introduction Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado Some radioactive isotopes formed billions of years ago have half- lives so long

More information

WM2018 Conference, March 18-22, 2018, Phoenix, Arizona, USA. PVT and LaBr3(Ce)-based Radon Express Analyzers 18164

WM2018 Conference, March 18-22, 2018, Phoenix, Arizona, USA. PVT and LaBr3(Ce)-based Radon Express Analyzers 18164 PVT and LaBr3(Ce)-based Radon Express Analyzers 864 Vladislav Kondrashov *, Stephen Steranka* and Glenn Paulson** * RadComm Systems Corp. 293 Portland Dr, Oakville, Ontario L6H 5S4, CANADA ** Paulson and

More information

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple Reliable TOC measurements Stop throwing away your first rep. QbD1200 has virtually eliminated sample-to-sample carryover

More information

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES APPLICATION NOTE 43145 Analysis of trace elements in polymer s to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES Authors Introduction Sanja Asendorf, Application

More information

Quality assurance of analytical measurements of CO2 parameters. Andrew G. Dickson

Quality assurance of analytical measurements of CO2 parameters. Andrew G. Dickson Quality assurance of analytical measurements of CO2 parameters Andrew G. Dickson 1 Quality assurance constitutes the system by which an analytical laboratory can assure outside users that the analytical

More information

Product Description and Specification

Product Description and Specification Product Description and Specification Catalog No 7 Synthra [ 11 C]Choline Description Synthra [ 11 C]Choline is a completely automated synthesis system for routine production of [ 11 C]choline. Using in-target

More information

TO-15 Checklist Determination of VOCs in Air by GC-MS

TO-15 Checklist Determination of VOCs in Air by GC-MS LAB ID: DATE: LAB NAME: ASSESSOR NAME: Method Number: TO-15 Checklist Determination of VOCs in Air by GC-MS SOP Number: Revision Number: SOP Date: Personnel records observed: Data records observed: Revision

More information

1. Recycled mobile phase could contaminate my samples during the chromatographic run.

1. Recycled mobile phase could contaminate my samples during the chromatographic run. Page 1 of 5 1. Recycled mobile phase could contaate my samples during the chromatographic run. This is a reasonable concern especially in a regulated laboratory environment. At first this might seem like

More information

A 3 Vapor pressure of volatile liquids

A 3 Vapor pressure of volatile liquids Versuchsanleitungen zum Praktikum Physikalische Chemie für Anfänger 1 A 3 Vapor pressure of volatile liquids Purpose The purpose of this experiment is to determine the vaporization enthalpy, entropy and

More information

Traceable UV-Vis Reference Materials

Traceable UV-Vis Reference Materials UV-Vis Spectroscopy Traceable UV-Vis Reference Materials For Regulatory Compliance Introduction Many laboratories, particularly in the pharmaceutical industry, are under increasing pressure to show proof

More information

Application of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids. October 2006 Ottawa, Canada

Application of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids. October 2006 Ottawa, Canada Application of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids October 26 Ottawa, Canada Summary The introduction of a growing number of targeted protein-based drug formulations

More information

7.1 Atomic Theory and Radioactive Decay

7.1 Atomic Theory and Radioactive Decay 7.1 Atomic Theory and Radioactive Decay exists all around us. This radiation consists of high energy particles or waves being emitted from a variety of materials. is the release of high energy particles

More information

Demonstration of radon removal from SF 6. using molecular sieves. Journal of Instrumentation. Related content OPEN ACCESS

Demonstration of radon removal from SF 6. using molecular sieves. Journal of Instrumentation. Related content OPEN ACCESS Journal of Instrumentation OPEN ACCESS Demonstration of radon removal from SF 6 using molecular sieves To cite this article: A.C. Ezeribe et al View the article online for updates and enhancements. Related

More information

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length Yun-Kyung Bae and In-Jik Jeong Industrial & Physical Instrument Center, Korea

More information

Automation of the radiochemical procedures for the sequential separation of radionuclides

Automation of the radiochemical procedures for the sequential separation of radionuclides LSC2017 - An International Conference on Advances in Liquid Scintillation Spectrometry, Copenhagen Denmark, 1 5 May 2017 Automation of the radiochemical procedures for the sequential separation of radionuclides

More information

Fusion UV/Persulfate TOC Analyzer

Fusion UV/Persulfate TOC Analyzer www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Fusion UV/Persulfate TOC Analyzer The Fusion Total Organic Carbon (TOC) Analyzer

More information

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS ABSTRACT LEDs are used in many applications, ranging from signalling to ambient lighting to displays. Specifications and tolerances are used

More information

QSM 5.1 FAQs. For a test with a recommended maximum holding time measured in hours, the holding time shall be tracked by the hour.

QSM 5.1 FAQs. For a test with a recommended maximum holding time measured in hours, the holding time shall be tracked by the hour. QSM 5.1 FAQs DoD ELAP FAQ Holding Time - Hours For a test with a recommended maximum holding time measured in hours, the holding time shall be tracked by the hour. For example, an exceedance of holding

More information

Comparing the MICROLAB 600 to Volumetric Glassware and Air Displacement Pipettes. June 2011

Comparing the MICROLAB 600 to Volumetric Glassware and Air Displacement Pipettes. June 2011 Comparing the MICROLAB 600 to tric Glassware and Air Displacement Pipettes June 2011 tric Glassware and Air Displacement Pipettes Table of Contents ``Abstract...3 ``Introduction...4 ``Method & Results...5

More information

REPORT RESULTS OF RADON DIFFUSION COEFFICIENT (SAMPLES SYNTHESIA O-14015)

REPORT RESULTS OF RADON DIFFUSION COEFFICIENT (SAMPLES SYNTHESIA O-14015) REPORT RESULTS OF RADON DIFFUSION COEFFICIENT (SAMPLES SYNTHESIA O-14015) Prof. Dr. Carlos Sainz Fernandez Technical Director REPORT RESULTS RADON DIFFUSION COEFFICIENT - SYNTHESIA CONTENTS 1. INTRODUCTION

More information

Radon Emanation Testing for DRIFT

Radon Emanation Testing for DRIFT Radon Emanation Testing for DRIFT DRIFT-IIa @ Boulby Direct & independent measurement of Rn emanation from detector components. Sean Paling - Sheffield. 1 CYGNUS mtg - July 2007 Boulby Cathode crossers

More information

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013 Comparing the Microlab 600 to tric Glassware and Air Displacement Pipettes January 2013 White Paper: Comparing the Microlab 600 to tric Table of Contents Abstract... 3 Introduction... 4 Methods & Results...

More information

MATHEMATICAL MODEL OF RADON ACTIVITY MEASUREMENTS

MATHEMATICAL MODEL OF RADON ACTIVITY MEASUREMENTS 2015 International Nuclear Atlantic Conference - INAC 2015 São Paulo, SP, Brazil, October 4-9, 2015 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-06-9 MATHEMATICAL MODEL OF RADON ACTIVITY

More information

STATIC GAS MONITOR Type SGM/DEW Revision A of 20 aprile 2016

STATIC GAS MONITOR Type SGM/DEW Revision A of 20 aprile 2016 APPLICATIONS Moisture monitoring of air or gas (SF6) Suitable for indoor or outdoor Industrial, medical or aerospace fields HIGHLIGHTS High voltage circuit breakers commonly used for distribution and transmission

More information

TRACERLAB. Natural Radiation, the Determination of Radon, Thoron and it s Progenies. Quality and Technology - Made in Germany - TRACERLAB GmbH

TRACERLAB. Natural Radiation, the Determination of Radon, Thoron and it s Progenies. Quality and Technology - Made in Germany - TRACERLAB GmbH Quality and Technology - Made in Germany - TRACERLAB TRACERLAB offers a wide range of development, engineering and production of instruments, such as sampling-, sample preparation- and measuring systems,

More information

Radon Determination by Activated Charcoal Adsorption and Liquid Scintillation Measurement. Canoba, A.C.; López, F.O. and Oliveira, A.A.

Radon Determination by Activated Charcoal Adsorption and Liquid Scintillation Measurement. Canoba, A.C.; López, F.O. and Oliveira, A.A. Radon Determination by Activated Charcoal Adsorption and Liquid Scintillation Measurement Canoba, A.C.; López, F.O. and Oliveira, A.A. Publicado en: Journal of Radioanalytical and Nuclear Chemistry, vol.

More information

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment

More information

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM IC Application Work AW CH6-1048-012011 Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM Branch: Chemical industry; Water, wastewater, environmental protection,

More information

In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL

In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL J-B. Pruvost 1, P. Berkvens 2, F. Justine 1, C. Mage 1 1: Synchrotron SOLEIL BP.48, 91192

More information

S8 TIGER Series 2 for ASTM D 6443

S8 TIGER Series 2 for ASTM D 6443 Lab Report XRF 139 S8 TIGER Series 2 for ASTM D 6443 Standard Test Method for Determination of Ca, Cl, Cu, Mg, P, S and Zn in Unused Lubricating Oils and Additives Introduction Lubricating oils are generally

More information

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Introduction Humidity generators are mechanical devices that provide stable temperature and humidity

More information

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies AUTOMATED QA/QC CHECK

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies AUTOMATED QA/QC CHECK AUTOMATED QA/QC CHECK FOR β- COINCIDENCE DETECTOR Matthew W. Cooper, James C. Hayes, Tom R. Heimbigner, Charles W. Hubbard, Justin I. McIntyre, Michael D. Ripplinger, and Brian T. Schrom Pacific Northwest

More information

Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado

Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado Experiment 10 1 Introduction Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado Some radioactive isotopes formed billions of years ago have half-lives so long

More information

Draft PS 18 APPENDIX A Dynamic Spiking Procedure ( ) 1.1 This appendix to Performance Specification 18

Draft PS 18 APPENDIX A Dynamic Spiking Procedure ( ) 1.1 This appendix to Performance Specification 18 Draft PS 18 APPENDIX A Dynamic Spiking Procedure (4-4-2013) A1. Scope and Application 1.1 This appendix to Performance Specification 18 describes the procedure and performance requirements for dynamic

More information

881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise.

881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise. Ion analysis 881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise. 881 Compact IC pro and 882 Compact IC plus. Intelligent. Reliable. Precise. Metrohm s top ion

More information

The Elzone II Particle Count and Size Analyzer

The Elzone II Particle Count and Size Analyzer The Elzone II 5390 Particle Count and Size Analyzer Elzone II Overview of Basic Theory Operates using the Electrical Sensing Zone (ESZ) Principle, also known as the Coulter Principle. First commercially

More information

INFLUENCE OF EXPOSURE GEOMETRY ON THE RESPONSE OF CR39 SSNT RADON DETECTORS *

INFLUENCE OF EXPOSURE GEOMETRY ON THE RESPONSE OF CR39 SSNT RADON DETECTORS * Romanian Reports in Physics, Vol. 63, No. 2, P. 376 382, 11 INFLUENCE OF EXPOSURE GEOMETRY ON THE RESPONSE OF CR39 SSNT RADON DETECTORS * ELENA ROBU 1, FRANZ JOSEF MARINGER 2, MASSIMO GARAVALIA 3, LUCA

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

Method for Nitrite determination on Low-Range Samples

Method for Nitrite determination on Low-Range Samples Method for Nitrite determination on Low-Range Samples 1. PRINCIPLE This method is designed for the determination of nitrite in soil, water and other forms of samples. The sample is prepared offline if

More information

The integration of NeSSI with Continuous Flow Reactors and PAT for Process Optimization

The integration of NeSSI with Continuous Flow Reactors and PAT for Process Optimization The integration of NeSSI with Continuous Flow Reactors and PAT for Process Optimization Michael F. Roberto, Thomas I. Dearing and Brian J. Marquardt U.S. Food and Drug Administration 1 FDA Sponsored CFR

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium, 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin Elemental Scientific prepfast IC: Inline Autodilution Method to Eliminate Species Interconversion for LC-ICPMS Based Applications Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan

More information

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine pplication Note Small Molecule Pharmaceuticals, Generics ccurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine igh-resolution sampling D-LC uthors Rongjie Fu and Xinlei

More information

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW Paula Hong, John Van Antwerp, and Patricia McConville Waters Corporation, Milford, MA, USA Historically UV detection has been

More information

by A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock +

by A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock + Bilateral Comparison of 1.018 V and 10 V Standards between the INTI (Argentina) and the BIPM, August to October 2009 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b) by A.Tonina*, R.Iuzzolino*,

More information

Samples compliant with the WATER

Samples compliant with the WATER ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions

More information

NSG 4070C APPLICATION FOR MIL-STD 461G CS 114

NSG 4070C APPLICATION FOR MIL-STD 461G CS 114 out < +0 meter ch. < +0 ch. < +0 ch. < +7 M nd NSG 0C APPLICATION FOR MIL-STD G CS Test parameter Equipment Standard: MIL-STD G CS Frequency range: to 00 M Curve to : see diagrams Modulation: pulse modulation,

More information

2. QUALITY CONTROL IN MASS SPECTROMETRY

2. QUALITY CONTROL IN MASS SPECTROMETRY IAEA-SM-367/5/03 Evaluation of Uncertainties for Pu and U Measurements Achieved in the On-Site Laboratory by Thermal Ionisation Mass Spectrometry during Two Years of Operation E. Zuleger, K. Mayer, L.

More information

A New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores

A New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores Application Note 37890312 A New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores Keywords CNSolution 9310 On-line Cyanide Analyzer Cyanidation

More information

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur Combustion Ion Chromatography Fast and reliable determination of halogens and sulfur Combustion digestion and ion chromatography combined in one system 02 Combustion Ion Chromatography (CIC) extends the

More information

VOC measurements in ambient air using Selected Ion Flow Tube Mass Spectrometry-automation and calibration considerations

VOC measurements in ambient air using Selected Ion Flow Tube Mass Spectrometry-automation and calibration considerations VOC measurements in ambient air using Selected Ion Flow Tube Mass Spectrometry-automation and calibration considerations Environmental Chemistry group, water Science Forum and the Separation Science Group

More information

Noble Gas Control Room Accident Filtration System for Severe Accident Conditions (N-CRAFT)

Noble Gas Control Room Accident Filtration System for Severe Accident Conditions (N-CRAFT) E-Journal of Advanced Maintenance Vol.7-1 (2015) 34-42 Japan Society of Maintenology Noble Gas Control Room Accident Filtration System for Severe Accident Conditions (N-CRAFT) Axel HILL *1, Dr. Cristoph

More information

The Claviature of Gas Analysis

The Claviature of Gas Analysis The Claviature of Gas Analysis Mass spectrometric gas analysis for quality assurance of gases and gas mixtures High purity and special gases are becoming increasingly important in the development of modern

More information

User Manual. Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio. Model with NIST-Traceable Calibration

User Manual. Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio. Model with NIST-Traceable Calibration User Manual Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio with NIST-Traceable Calibration Model 20250-09 THE STANDARD IN PRECISION MEASUREMENT Introduction The Digi-Sense

More information

Minimum (-31) every 10 s -35 (-31)

Minimum (-31) every 10 s -35 (-31) Humidity, and Dew Point Data Logger with LCD ORDERING INFORMATION Standard Data Logger (Data Logger, Software on CD, Battery) Replacement Battery EL-USB-2-LCD BAT 3V6 1/2AA FEATURES 0 to 100 measurement

More information

Purification of N 2 and Ar. Hardy Simgen, MPIK Heidelberg

Purification of N 2 and Ar. Hardy Simgen, MPIK Heidelberg Purification of N 2 and Ar Hardy Simgen, MPIK Heidelberg 1 Outline Requirements Gas purification by adsorption N 2 purification Ar purification Techniques for analysis Summary Page 2 Production N 2, Ar

More information

Fusion UV/Persulfate TOC Analyzer

Fusion UV/Persulfate TOC Analyzer Fusion UV/Persulfate TOC Analyzer The Fusion Total Organic Carbon (TOC) Analyzer utilizes powerful Ultra Violet (UV) Persulfate oxidation allowing superior carbon liberation from even the most challenging

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

Spa & Wellness for Pipettes

Spa & Wellness for Pipettes epservices for premium performance Spa & Wellness for Pipettes Pipette Performance Plans Treat your pipettes well with premium services from the experts! Total approach philosophy For over 60 years Eppendorf

More information

SSi Super Systems Inc.

SSi Super Systems Inc. Model DPC2530 Continuous Digital Dew Point Analyzer For measurement of: Endothermic Atmosphere Exothermic Atmosphere Nitrogen / Hydrogen Atmosphere Plant Air Systems SPECIFICATIONS Measurement Range: -50

More information

Performance Characteristics of Deterministic Leak Detection Systems

Performance Characteristics of Deterministic Leak Detection Systems Performance Characteristics of Deterministic Leak Detection Systems Michael Kasch 3S Consult München GmbH Germany Abstract The aim of leak detection and leak location is the limitation of losses and possible

More information

ANALYSIS CERTIFICATE OF

ANALYSIS CERTIFICATE OF CERTIFICATE OF ANALYSIS 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 INORGANIC VENTURES is an

More information

Finnigan LCQ Advantage MAX

Finnigan LCQ Advantage MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer

More information

High Precision Mass Measurement in Automation Tatjana Ivanova 1, Jānis Rudzītis 2

High Precision Mass Measurement in Automation Tatjana Ivanova 1, Jānis Rudzītis 2 Solid State Phenomena Vol. 164 (2010) pp 19-24 Online available since 2010/Jun/30 at www.scientific.net (2010) Trans Tech Publications, Switzerland doi:10.4028/www.scientific.net/ssp.164.19 High Precision

More information

The LSC Approach to Radon Counting in Air and Water

The LSC Approach to Radon Counting in Air and Water CHAPTER 32 The LSC Approach to Radon Counting in Air and Water Charles J. Passo, Jr. and James M. Floeckher INTRODUCTION Various methods exist to monitor 222Rn in air. There are seven commonly used types

More information

Supplementary Methods: Determination of the spectral

Supplementary Methods: Determination of the spectral Supplementary Methods: Determination of the spectral responsivity of fluorescence instruments. The spectral responsivity s( em ) of a fluorescence instrument like a fluorescence spectrometer can be best

More information

Measurement Uncertainty, March 2009, F. Cordeiro 1

Measurement Uncertainty, March 2009, F. Cordeiro 1 Measurement Uncertainty, March 2009, F. Cordeiro 1 Uncertainty Estimation on the Quantification of Major Proteins in Milk by Liquid Chromatography Contribution to a Chemical Reference Measurement System

More information

881 Compact IC pro 882 Compact IC plus. Compact ion chromatography for intelligent, reliable and precise analysis

881 Compact IC pro 882 Compact IC plus. Compact ion chromatography for intelligent, reliable and precise analysis 881 Compact IC pro 882 Compact IC plus Compact ion chromatography for intelligent, reliable and precise analysis 881 Compact IC pro and 882 Compact IC plus the compact solution for routine analysis 02

More information

Low Background Counting At SNOLAB

Low Background Counting At SNOLAB Low Background Counting At SNOLAB Ian Lawson Collaboration Meeting Minneapolis, Minnesota, June 22-23, 212 1 Outline SNOLAB and description of the SNOLAB Low Background Gamma Counting System Other material

More information

Exploitation of ground based GPS for Climate and Numerical Weather Prediction applications COST action 716

Exploitation of ground based GPS for Climate and Numerical Weather Prediction applications COST action 716 Exploitation of ground based GPS for Climate and Numerical Weather Prediction applications COST action 716 COST Objectives and status of COST 716 Overview of work packages / projects Near real-time demonstration

More information

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software V5.2 Plutonium and Uranium Isotopic Analysis Software Advanced Isotopic Ratio Analysis Software for HPGe Gamma-Ray Spectra Analyzes Pu, and a wide variety of heterogeneous samples containing Pu, Am, U,

More information

Mehlich and Modified Mehlich Buffer for Lime Requirement

Mehlich and Modified Mehlich Buffer for Lime Requirement Mehlich and Modified Mehlich Buffer for Lime Requirement David H. Hardy Application and Principle The development of the Mehlich buffer was to estimate acidity actually affecting plant growth in North

More information

Certificate: / 29 April 2014

Certificate: / 29 April 2014 Test report: 936/21218896/A of 20 September 2013 Initial certification: 01 April 2014 Date of expiry: 31 March 2019 Publication: BAnz AT 01 April 2014 B12, chapter IV, No. 5.1 Approved application The

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for:

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for: A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for: generating monodisperse aerosol by mobility classification with automatic concentration detection

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

The Plus for your lab!

The Plus for your lab! BioPhotometer plus NEW! analysis system The Plus for your lab! The easy, rapid and reliable BioPhotometer plus analysis system See it clearly! UVette Precision in plastics The patented* 1 UVette is the

More information

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD APPLICATION NOTE OPS-001 The Optical Particle Sizer (OPS) is calibrated for size with Polystyrene Latex (PSL) spheres (per ISO 12501-1/4) at TSI. This method

More information

ɣ-radiochromatography

ɣ-radiochromatography nuclear instruments Kapitel_y_Graphy.indd 1 19.08.09 10:27 n u c l e a r i n s t r u m e n t s Index HPLC-detectors TLC-detectors GABI* analog output GABI* digital output to PC minigita*, 1 trace scanner

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Multiphase Oscillatory Flow Strategy for in Situ Measurement and Screening of Partition Coefficients

Multiphase Oscillatory Flow Strategy for in Situ Measurement and Screening of Partition Coefficients Supporting Information Multiphase Oscillatory Flow Strategy for in Situ Measurement and Screening of Partition Coefficients Milad Abolhasani, Connor W. Coley, and Klavs F. Jensen * Department of Chemical

More information

Determination of Polymer Modifier in Asphalt

Determination of Polymer Modifier in Asphalt Standard Method of Test for Determination of Polymer Modifier in Asphalt AASHTO Designation: T xxx-xx (2005) 1. SCOPE 1.1. This method of test is used to determine the polymer content of an asphalt sample.

More information

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information