Scintillation Detector

Similar documents
hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Gamma Spectroscopy. References: Objectives:

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Analysis of γ spectrum

GAMMA RAY SPECTROSCOPY

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics

Detection and measurement of gamma-radiation by gammaspectroscopy

EEE4106Z Radiation Interactions & Detection

Gamma Ray Spectroscopy

Nuclear Physics and Astrophysics

SCINTILLATION DETECTORS AND PM TUBES

Mass of the electron m 0

MEASURING THE LIFETIME OF THE MUON

Experiment 6 1. The Compton Effect Physics 2150 Experiment No. 6 University of Colorado

PHY 192 Compton Effect Spring

Radionuclide Imaging MII Detection of Nuclear Emission

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

DETECTORS. I. Charged Particle Detectors

Chapter 4 Scintillation Detectors

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Advanced lab course for Bachelor s students

Multi Channel Analyzer (MCA) Analyzing a Gamma spectrum

Radiation Detection and Measurement

THE COMPTON EFFECT Last Revised: January 5, 2007

Quality Assurance. Purity control. Polycrystalline Ingots

Absolute activity measurement

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

Radiation (Particle) Detection and Measurement

Nuclear Lifetimes. = (Eq. 1) (Eq. 2)

Radioactivity. Lecture 6 Detectors and Instrumentation

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

ABSORPTION OF BETA AND GAMMA RADIATION

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

Radiation Detection. Laboratory & Computational Physics 2. Last compiled March 2, Constructed from legacy notes with amendments by Innes Bigaran

AN34 Experiments in Nuclear Science Laboratory Manual Fourth Edition

COMPTON SCATTERING OF GAMMA RAYS

Radiation Detection. Laboratory & Comp. Physics 2

The Mössbauer Effect

PHYS 3446 Lecture #12

Alpha-Energies of different sources with Multi Channel Analyzer

Gamma Ray Spectroscopy, Using NaI(Tl) Detectors. Department of Physics The University of Hong Kong

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University

Experiments 30a and 30b INTERACTION OF GAMMA RAYS WITH MATTER

BETA-RAY SPECTROMETER

Foundations of Modern Physics by Tipler, Theory: The dierential equation which describes the population N(t) is. dn(t) dt.

Overview: In this experiment we will study the decay of a radioactive nucleus, Cesium. Figure 1: The Decay Modes of Cesium 137

Overview: In this experiment we study the decay of a radioactive nucleus, Cesium 137. Figure 1: The Decay Modes of Cesium 137

Particle Energy Loss in Matter

Lecture 3 - Compton Scattering

Nuclear Spectroscopy (1) Precautions

Particle Energy Loss in Matter

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P )

III. Energy Deposition in the Detector and Spectrum Formation

PARTICLES REVELATION THROUGH SCINTILLATION COUNTER

Chapter Seven (Nuclear Detectors)

Radioactivity and Ionizing Radiation

Basic physics Questions

II. 5. Study for NaI(Tl) and Scintillation Fiber with 80 MeV Proton Beam Toward ESPRI Experiment at NIRS-HIMAC, RIKEN-RIBF

Neutrino Helicity Measurement

dn(t) dt where λ is the constant decay probability per unit time. The solution is N(t) = N 0 exp( λt)

arxiv: v1 [physics.ins-det] 3 Feb 2011

Interaction of charged particles and photons with matter

Gamma ray coincidence and angular correlation

Pulse height non-linearity in LaBr 3 :Ce crystal for gamma ray spectrometry and imaging

At the conclusion of this lesson the trainee will be able to: a) Write a typical equation for the production of each type of radiation.

David B. Cassidy. Department of Physics and Astronomy, University of California, Riverside, USA. Varenna, July 09

Contents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung

Total probability for reaction Yield

Scintillation Detectors

arxiv:physics/ v1 3 Aug 2006

4.1. The Positron Sources

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

Measurement of Mean μ-lifetime

Interaction of Electron and Photons with Matter

Researchers at the University of Missouri-Columbia have designed a triple crystal

Rutherford Scattering

LAB 4: Gamma-ray coincidence spectrometry (2018)

Beta Spectroscopy. Glenn F. Knoll Radiation Detection and Measurements, John Wiley & Sons, Inc. 2000

The interaction of radiation with matter

1.4 The Tools of the Trade!

The Compton Effect. Martha Buckley MIT Department of Physics, Cambridge, MA (Dated: November 26, 2002)

THE GEIGER-MULLER TUBE AND THE STATISTICS OF RADIOACTIVITY

Scintillators General Characteristics

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

1.5. The Tools of the Trade!

1 of :32

Direct WIMP Detection in Double-Phase Xenon TPCs

General Overview of Gas Filled Detectors

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Photon Instrumentation. First Mexican Particle Accelerator School Guanajuato Oct 6, 2011

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER

Modern Physics Laboratory (Physics 6180/7180)

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

Alpha-particle Stopping Powers in Air and Argon

Radioactivity. The Nobel Prize in Physics 1903 for their work on radioactivity. Henri Becquerel Pierre Curie Marie Curie

Figure 1. Decay Scheme for 60Co

Transcription:

Scintillation Detector Introduction The detection of ionizing radiation by the scintillation light produced in certain materials is one of the oldest techniques on record. In Geiger and Marsden s famous scattering experiment of α- particles off Gold nuclei, the scattered α s were observed via the scintillation light they produced when they hit a ZnS- screen. ZnS produced light flashes, called scintillation light, when hit by an α particle. The scintillation process remains one of the most useful methods available for the detection and spectroscopy of a wide assortment of radiation. The scintillation material used in this experiment is NaI(Tl) where (Tl) means that there are small mounts of Tl (Thallium) mixed into the NaI. Scintillation light is produced when a fast, charged particle excites the electrons in the NaI crystal from their valence band into the conduction band and leaving a hole in the valence band. The Thallium impurity in NaI produces so- called activator centers, which allow the electrons and the holes to recombine quickly with the emission of visible light. This light is detected by a photo- multiplier (see below) that produces an electrical current pulse proportional to the number of photons observed. A good scintillator should convert the kinetic energy of the charged particle in to visible light efficiently and linearly. The decay time of the light should be fast so that fast pulses can be generated and high particle rates can be measured. Scintillator Photo Multiplier (PMT) photo cathode dynodes anode +HV NaI(Tl) scintillation photon current pulse charged particle photo electron +HV

Photo Multiplier The photomultiplier is an instrument that can detect photons. An incoming photon (e.g. a scintillation photon) hits the photo- cathode and via photo effect releases an electron. The probability for this process is about 20%, so you need on the average 5 photons to produce a single photo- electron. The released electron is then accelerated toward a metal plate called a dynode where upon impact it releases more electrons, typically 2. These in turn are accelerated to the next dynode where each electron knocks out 2 more electrons. In this matter one obtains an exponentially growing number of electrons. At the end all the released electrons are collected at the Anode where they produce a short, measurable electric current pulse. Electronics and Data Acquisition The output pulses of the PMT are amplified and then recorded. The size of a pulse is proportional to the number of photons hitting the photocathode. These in turn are proportional to the amount of energy the charged particle deposited in the scintillator. Therefore the output pulse is proportional to the energy deposited in the scintillator. The proportionality constant has to be determined experimentally. This process is called calibrating the detection system. The amplified pulses are analyzed by an Analog to Digital Converter (ADC) which determines the pulse height and converts it into a number (typically between 0 and 1023, or 0 to 2047). These (ADC) numbers are then histogrammed i.e. each possible ADC value has a counter associated with it and each time a certain value occurs the corresponding counter is incremented. So if your ADC has a range from 0 to 1023, you have 1024 counters associated. Each ADC value is called a channel and the counter value is called the channel contents. If you take data for a while you will see that some values occur more frequently than others. The total set of counter values and the corresponding ADC values is called a spectrum. When you plot a histogram the x- axis corresponds to the ADC value and the y- axis represents the corresponding counter values or channel contents. The process described above performed by a Multi Channel Analyzer (MCA). The hardware part is installed in a PC and the associated software (Maestro) is running on the computer. ADC MCA/Computer PMT Amplifier

Detection of Photons A scintillator is primarily sensitive to the passage of charged particles. High energy photons, like all photons, are not charged and can only be observed if they produce a fast moving charged particle which subsequently interacts with the scintillator material. The energy that is measured in a scintillator is therefore the energy deposited by charged particles, mostly electrons, that interacted with the incoming high energy photon. The main reactions between a photon and the electrons in the scintillator are: Photo effect: a photon is absorbed by an electron which then interacts with the crystal. K e = E! " B e, where K e is the kinetic energy of the electron, E! is the photon energy and B e is the electron s binding energy. Compton scattering: a photon scatters off an atomic electron, the electron recoils and has kinetic energy, the original photon changes direction and loses energy corresponding to the recoil energy of the electron. The recoiling electron interacts with the crystal. Its energy is given by: $ & K e = E &! & & % M e E! ' ) 1" cos(#) ), where! is the photon scattering angle ) + 1" cos(#) ) ( Pair production: a photon (with a an energy > 1.2 MeV) converts into an electron positron (anti- electron) pair. The electron is stopped in the crystal, the positron eventually annihilates with an atomic electron and produces 2 0.511 MeV photons which may or may not be detected. Rayleigh scattering: a photon scatters off an atom without excitation. This does not produce a measurable effect in the scintillator. The probability of each of these process depends on the photon energy and the composition of the scintillator material. From all the processes which produce signals in the detector (those highlighted above) the most important one for the source we are using is Compton scattering. This can be seen from studying the cross sections for these processes for Carbon and Lead in the following figure. Note that this is a log- log plot! For Carbon, Compton scattering is dominant for photon energies between about 100KeV and 10 MeV while for Lead the photo electric process dominates up to about 700KeV when Compton scattering starts to dominate.

Cross section (barns /atom) 1 Mb 1 kb 1 b " p.e. " Rayleigh (a) Carbon (Z = 6)! experimental " tot photo electric effect! p.e.! Rayleigh! Compton! nuc Coherent scattering (Rayleigh scattering atom neither ionized nor excited) Incoherent scattering (Compton scattering off an electron) Pair production, nuclear field 10 mb " Compton # nuc # e! e Pair production, electron field 1 Mb " p.e. (b) Lead (Z = 82)! experimental " tot Cross section (barns /atom) 1 kb " Rayleigh # nuc 1 b " Compton # e 10 mb 10 ev 1 kev 1 MeV 1 GeV 100 GeV Photon Energy The goal of this experiment is to understand the spectrum produced by a scintillator when a high energy photon interacts with it. To understand it you should always remember that several of the processes listed above can occur sequentially with a single initial photon and that it is the total energy of the recoiling electrons that is measured in this process. If mono energetic photons enter the scintillator and transfer all their energy to the electrons which all get stopped in the scintillator then the signals at the end of the PMT should all have basically the same value. This means that the histogram should show a peak. If a Compton scattered photon leaves the scintillator without further interaction then the maximum energy deposited in Compton scattering would correspond to a photon scattering angle of 180 degrees. A schematic spectrum of a scintillator measuring photons with the same energy and where E! " 1MeV would look like:

Photo peak: entire photon energy deposited in scintillator Counts Compton edge: 180 o scattered photon energy deposited by recoiling electrons from compton scattering Energy The measured response of the scintillator depends on the composition, the size and the geometrical shape of the detector material. Testing the MCA Connect a signal generator (BK 2MHz function generator 3011B) with a BNC T to the oscilloscope (you might want to ask for help at this point to get familiar with the oscilloscope) and to the input of the ADC at the back of the PC. Set the signal size to 1V and the rate to 100Hz. On the PC start the Maestro program and click on the start symbol to start taking data. You should see the channel contents growing. When you click on the stop symbol the data taking stops. Take about 1000 events. Change the signal size by 1 V and take data without clearing the spectrum. Continue to do this until you reach 5 V in steps of 1 V. You should see now a total of five sharp peaks, each corresponding to the various signal sizes. This allows you to calibrate the MCA in terms of signal size / channel. It also allows you to check the linearity of the ADC. If you fit a quadratic function to your data how much does the quadratic term contribute and how big is its uncertainty? Now you are ready to take data with the scintillator and the PMT. PMT Operation The photo multiplier requires high voltage (HV) in order to accelerate the secondary electrons to the anode. It is supplied by the Ortec 556 module. Before you turn it on, make sure the knobs are set to 0. Turn the unit on and then set it to 700-800V. The output of the PMT should be connected to the input of the Ortec 590A amplifier. Connect the output of the amplifier with a BNC T to an Oscilloscope and continue the connection to the input of the ADC at the back of the PC. Place a 137 Cs source on the scintillator and check if you see pulses on the oscilloscope. Make sure the signals are in the range of 2 5 V. What you see is the histogram (spectrum) of pulses described previously.

You should save this spectrum using File! Save from the menu. Make sure that you select to save an ASCII spectrum with the extension.spe. You can load this file back into Maestro but you can also use it in LabTools for further analysis (look in the documentation). Identify the photo peak and the Compton edge. Replace the 137 Cs source with a 60 Co source. And take data for about 2 minutes. Decays schemes of 137 Cs and 60 Co. 55 Cs 137 30.17 a 7/2+ max 0.5120 MeV!- max 1.174 MeV!- 94.6% 5.4% 85.1% 56 Ba 137m 2.55m 0.6617 MeV! 56 Ba 137 stable 0.66 17 11/2-0 3/2+ 5.272 a 5+ 0.31 MeV! 99.88% 2.505 4+ 1.48 MeV! 0.12% 2.158 2+ 1.1732 MeV! 1.332 2+ 1.3325 MeV! 0+ The red arrows indicate the photons to be detected by the NaI(Tl) scintillator. The photons are emitted when the excited daughter nuclei transition into their ground states. The exited daughter nuclei are the result of a beta decay of the parent nuclei. Using the photo peak of these two sources calibrate the spectra such that instead of channels (or bin numbers) you have KeV ot MeV. Locate the various Compton edges and compare the theoretical values.

Back Scattering Peak It can happen that a photon does a 180 scattering outside the detector (e.g. in the thick glass of the PMT) and this backscattered photon is then detected. Find its energy and locate the corresponding peak. The recoiling electron from the backscattering process is not detected. NaI(Tl) high enery! backscattered! detected Background Measurement and Subtraction Remove all the sources and take data for 5 minutes. The spectrum you obtained in such a way is a background spectrum and is present in all the spectra you took so far. You can subtract the background from the spectra you took if you know how long each spectrum accumulated. You need to adjust your background spectrum to correspond to the same acquisition time as the one you took with the source. You can perform this subtraction either using the Maestro program (see the manual), or using the LabTools (see Working with MCA spectra).