Ion Selective Electrode Probe

Similar documents
Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION

Operating Instructions for Ammonium ISE Specifications

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

Fig. Electrochemical Cell/ Potentiometric Titration

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

1) A ph/mv meter or an ion meter, either line operated or portable.

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Ion Selective Electrodes

What is a WQSensor? Software Installation. Uninstalling WQSensors Software. NexSens Technology, Inc. TABLE OF CONTENTS

Neutralization Titration Package (SCS-200) ph Neutralization Titration experiment

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement

Experiment 7 Buffer Capacity & Buffer Preparation

POTASSIUM ION SELECTIVE ELECTRODE

Guide to ph Analysis.

NITRATE ION SELECTIVE ELECTRODE

This sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode.

F- Cl- Br- I- CN- S2- NO3-

EXPERIMENT 6. Properties of Buffers INTRODUCTION

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL

CALCIUM SELECTIVE ELECTRODE

WEEK 2 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY

phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL 1. A ph/mv meter or an ion meter, either line operated or portable.

READING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3.

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

Chemistry MultiMeasure Sensor PS-2170

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA. Background

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth

perfection Guidebook perfection Combination Cyanide Electrode Successful Ion Measurement

NH 3 -selective electrodes

Buffer Preparation. Learning Objectives:

EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY

COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL

Ion Selective Electrodes

Determination of the K a of a Weak Acid and the K b of a Weak Base from ph Measurements

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

symphony Multisensor Probes

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

DOC /2013, Edition 1. User Manual

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

GENERAL INSTRUCTIONS GENERAL PREPARATION

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL

symphony Reference Electrodes

10 samples per second

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014

perfection Guidebook perfection Combination Cupric Electrode Successful Ion Measurement

EXPERIMENT 16 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion

Experiment C-10 Titration of a Strong Acid and a Strong Base

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

#13 Electrochemical Cells

User Guide. Calcium Ion Selective Electrode

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC

Experiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University

Lab #14: Electrochemical Cells

phoenix Electrode Company Ammonia Gas-Sensing Electrode Instruction Manual

ELECTROCHEMICAL CELLS

Experimental Procedure. Lab 406

Use quantitative and qualitative analyses to identify the powder in Mr. Orlow s car.

Diffusion through Membranes. Evaluation copy. dialysis tubing, 2.5 cm 12 cm

The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions.

Part I: Solubility!!!

ELECTROCHEMICAL CELLS

User Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode

phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual 1. A ph/mv meter or an ion meter, either line operated or portable.

Solubility of CaSO 4 Experiment 8

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement

User Guide. High Performance Ammonia Ion Selective Electrode

ORP Probe. Mini ORP. Reads. +/-2000mV. Range SMA. Connector. 95% in 1s. Response time 100 PSI. Max pressure. 60m (197 ft) Max depth 1 80 C

ELECTROCHEMICAL CELLS

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

Chemistry 1B Experiment 14 65

ph 5 Meter Manual The LaMotte 5 Series

Volumetric Measurement Techniques. Technique #1 Use of a Burette. Technique #2 Use of a Pipette. Technique #3 Use of a Volumetric Flask

ph Probes Lab vs. Process Care, Calibrations and Comparison Presented by: Michael Blackburn Technical Sales Specialist at Hach Company

NO 2. Orion Nitrite Electrode INSTRUCTION MANUAL. Orion 93-46, Orion 97-46, ionplus. Analyze Detect Measure Control

PRACTICAL 3 ph AND BUFFERS

MEASUREMENT OF ph: INTRODUCTION TO BUFFER AND BASICS OF ph Meter

User Guide. High Performance Ammonia Ion Selective Electrode

1.22 Concentration of Solutions

Na+ K+ Ag+ Ca2+ NH + 4. F- Cl- Br- I- CN- S2- NO3-

Manual ph Probe & Oxidation Reduction Potential Probe Models 6528A and 6528B

Transcription:

Rev. 1 Ion Selective Electrode Probe KDS-1064 Calcium KDS-1065 Ammonium KDS-1066 Nitrate KDS-1067 Chloride Type : Ion selective electrode Range : 1 M to 5x10-7 M(40,000 ~ 0.01 ppm) Sensor description The Ion Selective Electrode is a hand crafted PVC membrane ion-selective electrode which measures nitrate ions in aqueous solutions simply, quickly, economically, and accurately. They are used to conduct water quality studies. The concentration range of the Nitrate Electrode is 1 M to 7 x 10-7 M or 0.01 ppm to 40,000 ppm. The ISE(Ion Selective Electrode) set consists of the FastFil Ion selective combination electrode, combining the Ion Selective electrode and the reference electrode, and ISE Probe amplifier. This product is to be used for educational purpose only. It is not appropriate for industrial, medical, research, commercial use.

Typical Experiments Ion Selective electrodes are used in a wide variety of applications for determining the concentrations of various ions in aqueous solutions. The following is a list of some of the main areas in which Ion Selective electrodes have been used:. Conducting water quality studies. Determining the concentration of ammonium ions in aqueous solutions. Pollution Monitoring: CN, F, S, Cl, NO3, etc., in effluents, and natural waters. Agriculture: NO3, Cl, NH4, K, Ca, I, and CN in soils, plant material, fertilizers and feedstuffs. Food Processing: NO3, NO2 in meat preservatives. Salt content of meat, fish, dairy products, fruit juices, brewing solutions. F in drinking water and other drinks. Ca in dairy products and beer. K in fruit juices and wine making effect of NO3 in canned foods. Detergent Manufacture: Ca, Ba, F for studying effects on water quality. Paper Manufacture: S and Cl in pulping and recovery-cycle liquors. Explosives: F, Cl, NO3 in explosive materials and combustion products. Electroplating: F and Cl in etching baths; S in anodizing baths. Biomedical Laboratories: Ca, K, Cl in body fluids (blood, plasma, serum, sweat). F in skeletal and dental studies. Education and Research: Wide range of applications

Figure 1: NO3 - Ion Selective Electrode How it works The Ion Selective Electrode (ISE) uses ion selective membrane to allow ions to penetrate to the electrode. A potential drop is developed between the two sides of the sensing membrane. This potential is proportional to the logarithm of the concentration of the Nitrate ion according with the Nernst equation: E = E0 + S ln(a)

Where E is the measured voltage, 0 E the reference potential, S the slope and a is the Nitrate activity. The slope is given by: S = RT/ nf Where R is the gas constant, T the temperature in Kelvin, n charge of the ion and F is Faraday constant. If the ionic strength is high and constant, Nernst equation can be written as: E = E0 + S ln(c) Where C is the Ionic concentration To adjust the background ionic strength to a high and constant value, ionic strength adjuster (ISA) must be added to all samples and standards. The potential develops due to the formation of a double layer consisting of a charged layer on the surface of the membrane of the ions sensed by the electrode and an opposite charged layer of counter ions from the sample (ions of opposite charge to the ones sensed by the electrode). As with any measurement of potentials, all values are relative to the built in reference electrode whose potential is constant. The reference solution aids in making electrical contact between the reference electrode (which is not in physical contact with the sample) and the sample. It consists of a solution of a salt that is able to conduct electricity but does not interfere with the measurement of the ion of interest. Items included Calcium ISE Probe Combination Ca++ Electrode Calcium ISE Probe amplifier 1 oz Ca++ Reference Filling Solution (RF0005) 1 oz Ca++ Ionic Strength Adjuster (ISA) (AJ0004) 1 oz Ca++ 10ppm as Ca Standard (SD2054) 1 oz Ca++ 1000ppm as Ca Standard (SD2008)

Ammonium ISE Probe Combination NH4+ Electrode Ammonium ISE Probe amplifier 1 oz. NH4+ Reference Filling Solution (RF0012) 1 oz. NH4+ Ionic Strength Adjuster (ISA) (AJ0015) 1 oz. NH4+ 10ppm as N Standard (SD2052) 1 oz. NH4+ 1000ppm as N Standard (SD2002) Nitrate ISE Probe Combination NO3- Electrode(AC017) Nitrate ISE Probe amplifier(kds-1066) 1 oz NO3- Reference Filling Solution (RF0011) 1 oz NO3- Ionic Strength Adjuster (ISA) (AJ0011) 1 oz NO3-10ppm as N Standard (SD2051) 1 oz NO3-1000ppm as N Standard (SD2030) Chloride ISE Probe Combination Cl- Electrode Chloride ISE Probe amplifier(kds-1067) 1 oz Cl- Reference Filling Solution (RF0007) 1 oz Cl- Ionic Strength Adjuster (ISA) (AJ0013) 1 oz Cl- 10ppm as Ca Standard (SD2053) 1 oz Cl- 1000ppm as Ca Standard (SD2012) Required Equipment ScienceCube Lite II, Pro or Max Series Wash bottle with distilled or de-ionized water Several clean beakers 1 ml, 10 ml and 100 ml pipettes Electrode Preparation 1. The sensing element comes pre-mounted on the end of the electrode with a protective bottle, but can be removed by unscrewing the electrode end.

Caution: Do not touch the PVC membrane with your fingers or over tighten the sensing element 2. The reference chamber must be filled with Reference Fill Solution and remain open during testing and measuring: a) Slide the sleeve of the electrode cap down to uncover the fill hole (see Figure 2) b) Fill the reference chamber with the Reference Fill Solution provided above the reference fill line on the electrode (see Figure 1) Figure 2: Reference Fill Cap 3. Shake the electrode downward like a thermometer to remove any air bubbles trapped inside. 4. Rinse the electrode with Dl water, blot dry. Do not rub dry. 5. Condition the electrode in a 10ppm solution for 30 minutes. 6. After the conditioning period, rinse the tip of the electrode with DI water. Defining the sensor with ScienceCube Interface 1. Turn on the Interface 2. Connect the Interface to the computer 3. Open ScienceCube program

4. Click Excel on the [Experiment] - [Setting], then click [Zero] to open a dialog: MultiLab will update the defined sensor in your data logger. 5. High Standard Calibration Point: The ISE should still be soaking in the High Standard. When the displayed value reading for Reading 1 stabilizes(~ 1 minute), click. 6. Low Standard Calibration Point: Remove the ISE from the High Standard, rinse well with distilled water from a wash bottle, and gently blot dry with a paper towel or lab wipe. Place the electrode into the Low Standard (included with your ISE). Make sure the ISE is not resting on the bottom of the container, and that the small white reference contacts are immersed. Make sure no air bubbles are trapped below the ISE. Checking Electrode Operation (Slope) Check the electrode every day when measurements will be conducted 1. Connect the electrode to the Electrode Amplifier(KDS-1043), then connect the Probe to the Interface. 2. Place 100 ml Dl water into a 150 ml beaker. Add 2 ml ISA to the Dl water and stir horoughly 3. Begin recording 4. Rinse the electrode with Dl water, blot dry and place in the solution prepared in step 2 5. Pipette 1 ml of 1000 ppm Nitrate Standard into the beaker. Stir thoroughly and then record he potential (E1) in mvs when a stable reading is displayed. 6. Pipette 10 ml of the same standard into the same beaker. Stir thoroughly. When a stable eading is displayed, record the potential (E2) in mvs. 7. The difference between the second and the first potential readings (E1-E2) is defined as the electrode slope. 8. The normal range for the slope is : Calcium : 27 mv at 25 C Ammonium : 56 mv at 25 C Nitrate : -56 mv at 25 C Chloride : -56 mv at 25 C Troubleshooting If the electrode slope is not within the normal range, the following procedure may restore the electrode. 1. Soak the electrode in the 10 ppm standard solution for 2 hours before use

2. Repeat "Checking Electrode Operation" procedure again Note: All standard solutions should be prepared fresh. Use ISA in all solutions. Periodically check the Reference Fill Solution level in the reference chamber. The solution level must be above than the reference fill line (see Figure 1). If the electrode slope is still outside the normal range after this procedure, replace the sensing module. Figure 4: replacing the Sensing Module Reading a Sample with the Electrode Various procedures may be used to determine the concentration of a sample. The most common is the Direct Calibration method, which is described below. In Direct Calibration a series of standard solutions of differing concentrations are used to calibrate the electrode. Then each sample requires only a single reading, which is compared with the calibration readings to obtain the sample concentration. ISA is added to all solutions to ensure the samples and the standards have the same ionic strength. Calibrate once a day before measurements. The filling hole must remain open during measurements (see Figure 2). Set up: 1. Prepare the electrode as described in "Electrode Preparation" 2. Connect the electrode to the ISE amplifier, then connect the amplifier to the data logger 3. Prepare two standard solutions that differ in concentration by a factor of ten. The standards should be at the same temperature as the sample

Measurement: 1. Place 100 ml of the more dilute standard into a 150 ml beaker. Add 2 ml of ISA. Stir thoroughly 2. Rinse electrode with Dl water, blot dry and place in the beaker. Wait for a stable reading, and then record the voltage reading 3. Measure 100 ml of the more concentrated standard into a second 150 ml beaker. Add 2 ml of ISA and stir 4. Rinse electrode with Dl water, blot dry and place in the second beaker. Wait for a stable reading, and then record the voltage reading of the second standard 5. On a semi-logaithmic graph paper, plot the voltage readings (linear axis) against the concentration (logarithmic axis). See Figure 5 for a typical calibration curve

Figure 5: Typical calibration curve Electrode Storage Short Term: Rinse the electrode thoroughly with Dl water and place the tip in a diluted standard solution (10ppm is satisfatory) between measurements. Slide the sleeve up to close refill hole. Make sure that the reference electrolyte does not exhaust the solution that the electrode is stored in does not dry. We re-command cleaning with DI water (see long term storage) at least once a week for solid results. Long Term: Empty reference chamber of Reference Fill Solution. Flush reference chamber with Dl water several times. Empty Dl water from the reference chamber and store the electrode dry. Replace the storage bottle and hand tighten the storage bottle cap. Follow procedures in "Electrode Preparation" and "Checking Electrode Operation" when using the electrode again. Specifications: Calcium ISE Probe Range: 1 M to 5 x 10-7 M (40,000 0.02 ppm) 12-bit Resolution: 0.5mV

ph Range: 2.5 to 11 ph Temperature Range: 0 to 40 C Electrode Resistance: 1 to 4MΩ Reproducibility: ±4% Minimum Sample Size: 3 ml in a 50 ml beaker Interfering Ions: Pb2+, Hg2+, Si2+, Fe2+, Cu2+, Ni2+, NH3,Na+, Li+, Tris+, K+, Ba+, Zn2+, Mg2+ Ammonium ISE Probe Range: 1 M to 5 x 10-6 M (18,000 0.1 ppm) 12-bit Resolution: 0.5mV ph Range: 4 to 10 ph Temperature Range: 0 to 50 C Electrode Resistance: 1 to 4MΩ Reproducibility: ±4% Minimum Sample Size: 3 ml in a 50 ml beaker Interfering Ions: K+, Na+ Nitrate ISE Probe Range: 1 M to 5 x 10-7 M (14,000 0.1 ppm) 12-bit Resolution : 0.5mV ph Range: 2.5 to 11 ph Temperature Range: 0 to 40 C Electrode Resistance: 1 to 4MΩ Reproducibility: ±4% Minimum Sample Size: 3 ml in a 50 ml beaker Interfering Ions: Pb2+, Hg2+, Si2+, Fe2+, Cu2+, Ni2+, NH3,Na+, Li+, Tris+, K+, Ba+, Zn2+, Mg2+ Chloride ISE Probe Range: 1 M to 5 x 10-6 M (35,000 1.8 ppm) 12-bit Resolution: 0.5mV ph Range: 2.0 to 12 ph Temperature Range: 0 to 80 C Electrode Resistance: less than 1 megohms Reproducibility: ±2% Minimum Sample Size: 3 ml in a 50 ml beaker

Interfering Ions: CN-, Br-, I-, OH-, S2- Solutions Calcium ISE Probe 1000ppm Ca (0.249 M Ca): Dissolve 3.668g CaCl2.2H2O in DI water and dilute to 1000mL ISA 4M KCl: 300g KCl in 1000mL DI water Reference (1M KNO3): 100g KNO3 in 1000mL DI water Ammonium ISE Probe 1000ppm NH4+ (0.0554 M NH4+): Dissolve 2.965g NH4Cl in DI water and dilute to 1000mL ISA 1M NaCl: 58.443g NaCl in 1000mL DI water Reference (0.1M NaCl): 29.22g NaCl in 1000mL DI water Nitrate ISE Probe 1000 ppm : 3 NO3- (0.0161 M NO3- ) Dissolve 1.631 g KNO3 in DI water and dilute to 1000 ml ISA 2 M (NH4)2 SO4: 264.3 g (NH4)2 SO4 in 1000 ml DI water Reference 1 M (NH4)2 SO4: 132.1 g (NH4)2 SO4 in 1000 ml DI water Chloride ISE Probe 1000ppm Cl (0.0282 M Cl): Dissolve 1.648g NaCl in DI water and dilute to 1000mL ISA 5M NaNO3: 425g NaNO3 in 1000mL DI water Reference (1M KNO3): 100g KNO3 in 1000mL DI water

This product is to be used for educational purpose only. It is not appropriate for industrial, medical, research, commercial use. Our products and the contents are subject to change without any notice. In consequence we cannot assume responsibility for any consequential or other damage resulting from the use of this instrument. and Science Cube are registered trademarks or trademarks of Co., Ltd. ScienceCube Lite, ScienceCube Lite II ScienceCube Pro and ScienceCube Max are trademarks of Co., Ltd. Windows and Excel are registered trademarks of Microsoft Corporation. Other product and brand names may be trademarks or registered trademarks of their respective organizations. KoreaDigital Co. Ltd. #804 Ace Twin Tower 2, Guro, Seoul, 152-779 South Korea TEL (+82) 2-2109-8880 FAX (+82) 2-2109-8881 www.sciencecube.com www.koreadigital.com www.sensecube.com mbl@sciencecube.com info@koreadigital.com