In order to get the energy response of the CZT detector, the measurements were done using the mono-energetic X-ray beam sources from 10 to 40 kev and

Similar documents
Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector

(a) Mono-absorber. (b) 4-segmented absorbers. (c) 64-segmented absorbers

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers

RESPONSE CALCULATIONS OF THE CdZnTe DETECTOR USING EGS4. James C. Liu, W. R. Nelson and R. Seefred

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Pixels GaAs Detectors for Digital Radiography. M.E. Fantacci. and. Abstract

MEASURE SYNCHROTRON RADIATION AT PEP-II? y. W. R. Nelson. Stanford Linear Accelerator Center. Stanford University, Stanford, CA 94309

Since the beam from the JNC linac is a very high current, low energy beam, energy loss induced in the material irradiated by the beam becomes very lar

Efficiency and Attenuation in CdTe Detectors

Neutron and Gamma-ray Emission Double Dierential Cross Sections. *5 Energy Conversion Engineering, Kyushu University, Kasuga-koen, Kasuga-shi 816.

Z 14. 2, 3, 4, 7, 10, 13 Contaminated Soil (Photon source) 5, 6, 9, 11, 12 Gap 8 NaI Crystal

MEASUREMENT OF DIAGNOSTIC X-RAY SPECTRA USING CdZnTe DETECTOR

1.E Neutron Energy (MeV)

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

1 cm. Cu electrode. Ge crystal. end cap. Dead layer. crystal cup

THE spectroscopic performance of large volume CdZnTe

PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY

2 Improvements of Programs 2.1 L-X ray Energy dependent L subshell cross sections In 1996 version of L-X ray calculation [5], the ratio of L sub

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

Estimate of Large CZT Detector Absolute Efficiency

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

Oslo Cyclotron Laboratory

Dependence on neutron energy of neutron induced peaks in Ge detectors. E. Gete, D.F. Measday B.A. Moftah, M.A. Saliba, T.J. Stocki

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

DETECTORS. I. Charged Particle Detectors

Compton Camera. Compton Camera

Count. Count. 200mm. PMT scintillation. PMT flourescence. PMT fluorescence. PMT fluorescence. PMT fluorescence. 50mm. 42mm

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

New trends in CdTe detectors for X and γ-ray applications

EEE4106Z Radiation Interactions & Detection

Photon and primary electron arithmetics in photoconductors for digital mammography: Monte Carlo simulation studies

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions

Advantages / Disadvantages of semiconductor detectors

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

Semiconductor Detectors

Development and characterization of 3D semiconductor X-rays detectors for medical imaging

Spatial resolved efficiency determination of CdZnTe semiconductor detectors with a collimated gamma-ray source for the C0BRA experiment

SENSITIVITY ANALYSIS TO EVALUATE THE TRANSPORT PROPERTIES OF CdZnTe DETECTORS USING ALPHA PARTICLES AND LOW-ENERGY GAMMA-RAYS

SIMPLE USE OF EGS4 BY EXCEL MACRO FILE

Chemical Engineering 412

PHITS calculation of the radiation field in HIMAC BIO

Drift plane. substrate (20ÉIm polyimide) 200ÉIm. Back strip (180ÉIm width) Base (Ceramic) Anode strip (10ÉIm width) Cathode strip (100ÉIm width)

Radiation Detection and Measurement

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

CADMIUM telluride (CdTe) and cadmium zinc telluride

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

Gamma Spectroscopy. References: Objectives:

Results of a Si/CdTe Compton Telescope

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

Detecting high energy photons. Interactions of photons with matter Properties of detectors (with examples)

Position Sensitive Germanium Detectors for the Advanced Compton Telescope

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Chapter Seven (Nuclear Detectors)

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Stability and Characteristics of Large CZT Coplanar Electrode Detectors

Hard X- and g-ray measurements with a large volume coplanar grid CdZnTe detector

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

Gamma and X-Ray Detection

Radioactivity. Lecture 6 Detectors and Instrumentation

A Study On Radioactive Source Imaging By Using A Pixelated CdTe Radiation Detector

Application Note ANCZT-2 Rev. 3 Charge Trapping in XR-100T-CdTe and -CZT Detectors

Measurement of 40 MeV Deuteron Induced Reaction on Fe and Ta for Neutron Emission Spectrum and Activation Cross Section

Chapter 4 Scintillation Detectors

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Boron-based semiconductor solids as thermal neutron detectors

Detection and measurement of gamma-radiation by gammaspectroscopy

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

POSITRON AND POSITRONIUM INTERACTIONS WITH CONDENSED MATTER. Paul Coleman University of Bath

1.4 The Tools of the Trade!

Chapter 1 Overview of Semiconductor Materials and Physics

Radiation Dose, Biology & Risk

The Best Gamma-Ray Detector

Introduction to Radiological Sciences Neutron Detectors. Theory of operation. Types of detectors Source calibration Survey for Dose

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

PROJECT STATUS AND PERSPECTIVES

A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate

CHARGED PARTICLE INTERACTIONS

Detection of X-Rays. Solid state detectors Proportional counters Microcalorimeters Detector characteristics

Quality Assurance. Purity control. Polycrystalline Ingots

Single Photon detectors

Compton suppression spectrometry

SECTION 8 Part I Typical Questions

Chem 481 Lecture Material 3/20/09

IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS

Georgia Institute of Technology. Radiation Detection & Protection (Day 3)

Analysis of γ spectrum

Characterizations and Diagnostics of Compton Light Source

Slides by: Prof. Abeer Alharbi

III. Energy Deposition in the Detector and Spectrum Formation

Attenuation of Radiation in Matter. Attenuation of gamma particles

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

Radiation (Particle) Detection and Measurement

Investigation of the Asymmetric Characteristics and Temperature Effects of CdZnTe Detectors

Introduction. X-Ray Production and Quality. Fluorescence Yield. Fluorescence X-Rays. Initiating event. Initiating event 3/18/2011

Transcription:

Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.144-151 Development of Gamma Ray Monitor Using CdZnTe Semiconductor Detector A. H. D. Rasolonjatovo, T. Shiomi, T. Nakamura, H. Nishizawa 1, Y. Tsudaka 1, H. Fujiwara 1, H. Araki 1, and K. Matsuo 1 Department of Quantum Science and Energy Engineering, Tohoku University Aza-aoba, Amaraki, Aoba-ku, Sendai, Miyagi 980-8578, JAPAN 1 Mitsubishi Electric Corporation 8-1-1, Tsukaguchi-Honmachi, Amagasaki, Hyogo 661, JAPAN Abstract In this study, we aimed to develop a new X-ray and gamma ray monitor using the CdZnTe semiconductor detector, whichhave high sensitivity at room temperature. The pulse height spectra and the detection eciencies of 10x10 mm 2 by 2 mm thick CdZnTe detector were measured in the energy range of 10 kev to 1.8 MeV by using monoenergetic X-ray and gamma ray sources. The measured results showed very good agreement with the results calculated using the EGS4 Monte Carlo code taking into account the charge collection eciency in the detector. By using two CZT detectors of 10x10x2 mm 3 and 3x3x2 mm 3 coupled with a lter, the weighted sum of a few energy channels with dierent cut-o energies was nally found out to realize a at energy response to equivalent dose (counts per msv) within 30% or 10% deviation. 1 Intoroduction In recent years the CdZnTe (CZT) semiconductor detector is preferred for X-ray and low energy gamma ray detection in many situations, because of its high eciency, low bias voltage, good energy resolution and its capability of operation at room and higher temperatures[1, 2, 3]. The CZT detector has a great advantage to have the high density ( =5:86 gcm ;3 ) and the high atomic number (Z=48 for Cd, 30 for Zn and 52 for Te) compared with Si (=2.4 gcm ;3, Z=14) and Ge (=5.36 gcm ;3, Z=32) semiconductor detectors. Nowadays a larger crystal size of CZT becomes available and the CZT detector can be used for higher energy gamma ray detection. In this study, we aimed to develop a new gamma ray monitor using a Cd0:5Zn0:5Te semiconductor detector. The present gamma ray monitor, therefore, possesses higher sensitivity than the commercially available Si semiconductor gamma ray monitor and at the same time has a at energy response to dose-equivalent in the wide photon energy region from 10 kev to 7 MeV. This gamma ray monitor could be used to operate at higher temperature such as in the power reactor core. 2 Materials and Methods As a gamma ray sensor, a Cd0:5Zn0:5Te detector having 10 mm x 10 mm size by 2 mm thickness (fabricated by ev Product, USA) was used. Table 1 shows the physical characteristics of the CZT detector used in this study. The detector was directly coupled to the low noise preamplier (Model 850, fabricated by Clear-Pulse Co. Ltd, Japan). The bias voltage of +24 V was supplied to the detector from the regulated power supply (PR 36-1.2 A fabricated by Kenwood, Japan). The preamplier was connected to a linear amplier ORTEC 571 with a 2 s shaping time. The detector and the preamplier were covered with an Al foil of 0.015 mm thickness to decrease the electromagnetic noise. 1

In order to get the energy response of the CZT detector, the measurements were done using the mono-energetic X-ray beam sources from 10 to 40 kev and the radioactive photon sources of 241 Am, 57 Co, 137 Cs, 60 Co and 88 Yhaving photon energies of 59.54 kev to 1836 kev. The experiments using X-ray beam sources were performed at a beam line of the Photon Factory of High Energy Accelerator Research Organization (KEK), Tsukuba, Japan. Because a direct photon beam intensity from the beam line of synchrotron radiation is too high, a beam scattered at 90 was used. In order to investigate the variation of the energy response with the lters, we also measured the pulse height distributions of the CZT detector covered with Al, Cu, Cd and Pb lters of various thicknesses. The measurements using point sources were carried out at the Hot Laboratory of the Cyclotron and Radioisotope Center (CYRIC), Tohoku University, Japan. The point sources of about 2 mm diameter are encapsulated within thin vinyl cover. The source was placed at 10 cm from the detector surface. The fraction of backscattered photons was about 2%. In order to get the angular dependence of the detection eciency, the CZT detector was rotated to the source-to-detector axis in the directions of 0 to 90. The 137 Cs source was covered with a 1.5 mm thick Al foil to absorb the 661.7 kev conversion electrons from the electron capture reaction and the rays of 511 kev. The 60 Co was also covered with a 1 mm thick Al foil to absorb the rays of 318 kev. 3 s The pulse height spectra of the detector were also calculated using the electron-photon cascade Monte Carlo code EGS4[4] taking into account the carrier trapping eect[5]. The charge collection depends on the drift distance of the hole and the electron, which vary with the depth of interaction in the detector. For the CZT, the product of the mobility and lifetime, (), is very low, especially for holes and then most of the holes are trapped before reaching the cathode. The mean free path is given by = F = V d (1) where F is the electric eld, d is the thickness of the detector and V is the applied voltage. In the present case, we calculated the deposited energy E in the CZT detector for the geometry that a plane parallel photon beam is incident normal to the detector surface using the EGS4-PRESTA code[4], and taking into account the carrier trapping eect. According to the work by Nishizawa et al.[5], three cases were considered: energy absorption only, constant drifting distance of the carrier and exponential distribution of the drifting distance of the carrier. For the rst case where energy absorption only is considered, the induced charges are independent to the interaction depth, then Q(x) =Q0 = E " e (2) " is the energy needed to create an electron-hole pair and is equal to 4.43 ev[6], E is the deposited energy and e is the charge of the electron (e =1:6 10 ;19 C). For the second case, the induced charges can be written as the rst-order approximation of the Fetch's equation[7] as follows[5] If e >d; x and h >x ;! Q(x) = E " e If e >d; x and h <x ;! Q(x) = E " e d ; x If e <d; x and h <x ;! Q(x) = E " e e d + h d d + h d 2

If e <d; x and h >x ;! Q(x) = E " e e d + x d (3) where x is the interaction depth, e, h are the mean free paths of the electron and hole, respectively. Similarly for the third case, the equation of the induced charges is written as follows is a random number between 0 and 1. E X " Pe + P Q(x) h =e d i=1 (4) (5) = (; ln ) (6) If e >d; x and h >x ;! P e = d ; x P h = x If e >d; x and h <x ;! P e = d ; x P h = h If e <d; x and h >x ;! P e = e P h = x If e <d; x and h <x ;! P e = e P h = h (7) The pulse height spectra from these calculations were then compared with the measured pulse height spectra. Fig.1 is an example of the 40 kev photon beams. It can be seen from this gure that the third case gives the best t to the 40 kev measured spectrum. After comparison for all photons, the third case of an exponential distribution of the drifting distance of the carrier was adopted. In the calculation of the angular dependence of the detector eciency, the monodirectional parallel beam was assumed to be incident to the surface of CZT at the incident angle. In the slant incidence, the following correction was applied to the sensitive surface of the detector S = S t sin + S l cos (8) where S is the total surface of the CZT detector faced to the monodirectional parallel beam, S t is the front surface and S l is the lateral surface of the detector. 4 Results and Discussions 4.1 Pulse height spectra The output pulse height spectra were measured with the multi-channel analyzer. The noise level of the detector system was about 6 kev. Figures 1 to 4 show the comparison of measured and calculated pulse height spectra of 40 kev X-ray beams, 57 Co (14.4, 122.1 and 136.5 kev), 137 Cs (661.7 kev) and 88 Y(898and 1836 kev) sources in counts per unit uence incident to the CZT detector. In order to compare with the measured spectra, the energy deposited in the detector calculated by the EGS was smeared out using a Gaussian distribution having an energy resolution determined from the measured photopeak spectra. For 40 kev X-ray source, there can be seen a single sharp photopeak in Fig.1 (Case 3). The fullwidth at half-maximum (FWHM) of the 40 kev photopeaks is 4.3 kev. The measured pulse height distribution of the 40 kev X-ray presents two escape peaks at 10 kev from K X-ray ofte and at 14 kev from K X-ray ofcdinfig.1. The calculated spectra for these three cases show good agreement with the measured spectra above the noise level of 6 kev in absolute values, although the calculation does not consider the two escape peaks of 10 and 14 kev for 40 kev X-rays. For 57 Co, the pulse height spectra for the above three photon energies were calculated separately and summed up considering their branching ratios. The photopeaks of 122 and 136 kev are much 3

broader than that of 14 kev. This is because the charge collection eciency in the detector decreases for low applied voltage of only +24V and also the energy loss through the Compton scattering increases with increase of photon energy. The energy responses of two photon energies of 1836 kev and 898 kev for 88 Y, were calculated separately and added taking into account the branching ratio of each energy. For higher energy photons from 137 Cs and 88 Y, the photopeak cannot clearly be seen because of energy loss dominated by Compton scattering and a small fraction of energy deposition in 2 mm thick CZT detector. Then, the measured spectra have monotonically decreasing shapes. The agreement between the measured and calculated pulse height spectra for these three photon sources is very good. In this study, we used low biasvoltage of +24V sacricing better energy resolution obtained with higher bias voltage. Since our aim is to develop a gamma ray monitor of high sensitivity in this work, the low bias voltage is preferred for reasons of portability. Generally speaking, the agreement between the measured and the calculated pulse height spectra is good in absolute values over the wide energy range of 10 kev (not shown here) up to 1836 kev. 4.2 Energy dependence of the detection eciency To get the detection eciency, the measured and calculated pulse height spectra were rst summed up above the cut-o energy xed at 6 kev. But for 241 Am the subsidiary peaks from K X-rays and the 26.4 kev gamma ray were not considered and the two peaks of 14 and 136 kev gamma rays of 57 Co were also excluded from the measured results in the eciency estimation. The counts thus-obtained were normalized to unit photon uence incident on the detector to get counts per cm 2. This value was converted to counts perunitequivalent dose in msv by using the uence-to-equivalent dose conversion factor (msv per cm ;2 ) for the A-P (anterior-posterior) geometry given by ICRP Publ.74[8]. We nally obtained the detection eciency in cps (count per second) per msv/h from 10 kev to 1836 kev as shown in Fig. 5. For 60 Co gamma rays the measured result is 20% higher than the calculated result, while for other photon energies the agreement between experiment and calculation is quite good within 3%. The eciency is high for 10 kev and rather at from 20 kev to 120 kev, and rapidly decreases with energy. The high eciency at 10 kev is due to the small value of the uence-to-equivalent dose conversion factor. 4.3 Angular dependence of the detection eciency The angular dependence of the detection eciency with 241 Am, 57 Co, 137 Cs, 60 Co and 88 Y photon sources were also studied and we found a good agreement between experimental and calculated results except for 60 Co with a standard deviation of about 20%. The angular dependence of the eciency is almost constant for 137 Cs, 60 Co and 88 Y in the directions of 0 to 90, but for lower energy photons of 241 Am and 57 Co, the eciency decreases beyond 60 because of the self-absorption through the thicker layer for slant incidence. These results revealed that this CZT detector has an almost isotropic eciency within 60. 4.4 Eect of various lters Figure 6 shows the variation of the eciencies for 10, 20 and 40 kev X-rays with various thicknesses (mm) of Al lter. Solid lines are the measured results and the dotted lines the calculated results. The cut-o energy was also xed at 6 kev. The variation of the eciency with Al lter thicknesses is signicant at 10 kev but is negligible at 40 kev. The agreement between measurement and calculation is good. 4

5 Conguration of Flat Energy Response By utilizing the results obtained, we tried to obtain at energy response to eective dose of the CZT semiconductor detector in the energy range of 10 kev to 7 MeV. This is an indispensable characteristic for use as a gamma-ray monitor. As very good agreement between experiment and calculation was obtained, we extend to obtain at energy response for higher energies up to 7 MeV by calculation. Three parameters were considered for this realization: lters, cut-o energy and detector size. The energy responses of the detector with dierent kinds of lter materials such as Al, Cu, Cd, Fe and Pb with various thicknesses were calculated in the energy range from 10 kev to 7 MeV. Low energy photons are very sensitive to lter material while high energy photons are almost insensitive to any kind of lters. We therefore found that it is rather dicult to get a at response only using lters. At the same time, we varied the cut-o energy of the pulse height spectra from 6 to 1500 kev. For low energy below 100 kev, a at energy response of the detection eciency can be obtained with a cut-o energy around 6 kev but for higher energy, a cut-o energy higher than hundreds of kev is necessary. By combing these above-mentioned results parametrically, we tried to get the at energy dependence of the detection eciency to equivalent dose in units of cps per msv/h within 30% and 10% dierences over a wide energy region of 10 kev to 7 MeV. After many parametric surveys on lters, cut-o energies and detector sizes, we nally optimized to achieve the at detection eciency to equivalent dose, ", by using two sizes of the CZT detectors 3x3 mm 2 by 2 mm thick CZT, and 10x10 mm 2 by 2mm thick CZT as follows, "(cps per Sv=h) = X i a i "1 + X i b i "2 (9) where "1 and "2 are the detection eciencies of 3x3x2 mm 3 and 10x10x2 mm 3 CZT, respectively, and a i and b i are multiplication factors. Fig. 7 and 8 show the results obtained by theabove process using Eq.(9). 6 Conclusion In this study, the pulse height spectra and the detection eciencies of the CZT detector of 10x10 mm 2 by 2 mm thickness are presented for 10 kev to 1.8 MeV photon energies from both experiment and calculation using EGS4 code. The agreement between the experiment and the calculation is very good in absolute values. The variation of the detection eciency with the cut-of energy and with various lters was also obtained by calculation in the photon energy of 10 kev to 7 MeV. By using two CZT detectors, 10x10 mm 2 by 2 mm thick CZT with 1.2 mm Pb lter and 3x3 mm 2 by2mmthick CZT without lter, we nally obtained the at energy response to eective dose within 30% and 10% deviations in the photon energy 15 kev to 7 MeV by taking the weighted sum of counts integrated above a few dierent cut-o energies for the two CZT detectors. From these results, it is evident that it is possible to realize a gamma ray monitor of good quality having a at energy response to equivalent dose in a wide photon energy range. Field measurements will be further needed to compare with other commercially available survey meters. Acknowledgement We are grateful to Professors H. Hirayama and S. Ban, Drs. T. Kurosawa and Y. Namito for their fruitful collaboration and advices during the calculation and experiment at KEK. 5

References [1] J. Franc, P. Haschl, E. Belas, R. Grill, P. Hledek, P. Moravec, J. Bok, \CdTe and CdZnTe crystals for room temperature gamma-ray detectors", Nucl. Instrum. Methods A 434(1999)146-151. [2] A. S. Zerrai, K. Cherkaoui, G. Marrakchi, G. Bremond, P. Fougeres, M. Hage-Ali, J. M. Koebel, P. Siert, \Inuence of deep levels on CdZnTe nuclear detectors", Journal of Crystal Growth 197(1999)646-649. [3] S. E. Pfantiel, K. J. Hofstetter, T. A. DeVol, \Comparison of four types of gamma and X-ray detectors for environmental applications in the 10-450 kev energy range", Journal of Radioanalytical and Nuclear Chemistry 223 1-2(1997)89-98. for air, nitrogen at 30 kev", Appl. Radiat. Isot. 44(1993)769-772. [4] H. Hirayama, \Overview of the EGS4 Code System", KEK Internal 97-10, High Energy Accelerator Research Organization, 1997, pp. 6-14. [5] H. Nishizawa, K. Ikegami, K. Takashima, T. Usami, T. Hayakawa, T. Yamamoto, \Development of Multi-layered CdTe Semiconductor Detectors", Ionizing Radiation Journal Vol.22, No.3(1996)27-36. [6] Y. X. Dardenne, T. F. Wang, A. D. Lavietes, G. J. Mauger, W. D. Ruhter, S. A. Kreek, \Cadmium Zinc Telluride Spectral modeling", Nucl. Instrum. Methods A 422(1999)159-163. [7] Y. Eisen, A. Shor, \CdTe and CdZnTe materials for room temperature X-ray and gamma ray detectors", Journal of Crystal Growth 184/185(1998)1302-1312. [8] International Commission on Radiological Protection, \ Conversion Coecients for use in Radiological Protection against External Radiation", ICRP Publication 74, Annals of the ICRP Vol.26/3, 1995. 6

Table 1. Physical characteristics of the CZT detector Size 10x10x2mm 3 Density 5.86 gcm ;3 Resistivity 3x10 10 cm Band gap 1.56 ev Mobility and lifetime product 3x10 ;3 cm 2 V ;1 for electron and 2x10 ;5 cm 2 V ;1 for hole Depletion layer 2mm Counts per unit incident fluence 0.016 0.012 Case 1 Case2 Case3 0.008 0.004 0 0 200 400 600 800 1000 1200 Channel Count per unit incident fluence 1.E-02 1.E-03 1.E-04 1.E-05 0 50 100 150 200 Fig.1 Comparison of measured pulse height spectra for Fig.2 Pulse height spectra of 57 Co photon source 40 kev X-rays with the calculated spectra for three models of charge collection. Case 1: Energy absorption only; case 2: Constant drift distance of carriers; case 3: The drift distance of carriers has an exponential distribution Counts per unit incident fluence 1.E-03 1.E-04 1.E-05 1.E-06 0 200 400 600 800 1000 Count per unit incident fluence 1.E-02 1.E-04 1.E-06 0 500 1000 1500 2000 Fig.3 Pulse height spectra of 137 Cs photon source Fig.4 Pulse height spectra of 88 Y photon source 7

Efficiency per equivalent dose [cps/( Sv/h)] 10000 1000 100 10 1 1 10 100 1000 10000 Efficiency per equivalent dose [cps/( Sv/h)] 10000 1000 100 0.1mm 0.2mm 0.5mm No filter 10 0.8mm 1.0mm 1 0 5 10 15 20 25 30 35 40 45 Fig.5 Measured and calculated detection efficiencies of the Fig.6 Variation of the detection efficiency with CdZnTe semiconductor detector as a function of photon energy aluminum filter for 10 to 40 kev photons 1000 +30% Efficiency per equivalent dose [cps/(µsv/h)] 100 10 (1) CZT (3x3x2) mm 3 No filter (2) CZT (10x10x2) mm 3 Pb filter -30% 1 10 100 1000 10000 Fig.7 Energy dependence of detection efficiency to effective dose within 30% difference. 1000 Efficiency per equivalent dose [cps/(µsv/h)] 100 (1) CZT (3x3x2) mm 3 No filter (2) CZT (10x10x2) mm 3 Pb filter +10% -10% 10 10 100 1000 10000 Fig.8 Energy dependence of detection efficiency to effective dose within 10% difference. 8