Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR

Similar documents
Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR

03.1 Experimental Error

Source: Chapter 5: Errors in Chemical Analyses

Ch 3. EXPERIMENTAL ERROR

Chapter 5. Errors in Chemical Analysis 熊同銘.

Statistics: Error (Chpt. 5)

Lecture 3. - all digits that are certain plus one which contains some uncertainty are said to be significant figures

Chapter 3 Math Toolkit

Measurements, Sig Figs and Graphing

ANALYTICAL CHEMISTRY 1 LECTURE NOTES

Topic 2 Measurement and Calculations in Chemistry

Chapter 3 Experimental Error

Experiment 1 - Mass, Volume and Graphing

Why the fuss about measurements and precision?

Chem 222 #3 Ch3 Aug 31, 2004

Measurements Chapter 3

Jan 18, 2005 #3. Average (Ch. 4) Standard deviation Q-test Significant Figures (Ch 3) Error

Data Analysis I. CU- Boulder CHEM-4181 Instrumental Analysis Laboratory. Prof. Jose-Luis Jimenez Spring 2007

Chapters 0, 1, 3. Read Chapter 0, pages 1 8. Know definitions of terms in bold, glossary in back.

Chemistry Calibration of a Pipet and Acid Titration

Accuracy and Precision of Laboratory Glassware: Determining the Density of Water

What is measurement uncertainty?

ADVANCED ANALYTICAL LAB TECH (Lecture) CHM

Topic 11: Measurement and Data Processing and Analysis. Topic Uncertainties and Errors in Measurement and Results

Chem 321 Lecture 4 - Experimental Errors and Statistics 9/5/13

CHEM 334 Quantitative Analysis Laboratory

Source slideplayer.com/fundamentals of Analytical Chemistry, F.J. Holler, S.R.Crouch. Chapter 6: Random Errors in Chemical Analysis

Error Analysis. Table 1. Tolerances of Class A Pipets and Volumetric Flasks

Identification Of The Common Laboratory Glassware, Pipettes And Equipment. BCH 312 [Practical]

Uncertainties in Measurement

Measurements and Errors

NaOH (aq) + HCl (aq) NaCl (aq) + H 2 O (l)

precision accuracy both neither

Practice Lab. Balances and calibration of volumetric tools

Assessment of Accuracy and Precision

Fundamentals of data, graphical, and error analysis

Chemistry Lab: Introduction to Measurement

Introduction to Measurements & Error Analysis

University of Massachusetts Boston - Chemistry Department Physical Chemistry Laboratory Introduction to Maximum Probable Error

Part 01 - Notes: Identifying Significant Figures

TOPIC 3: READING AND REPORTING NUMERICAL DATA

Measurement Uncertainties

Uncertainty Analysis of Experimental Data and Dimensional Measurements

Density of Aqueous Sodium Chloride Solutions

ECE 102 Engineering Computation

Instrumentation & Measurement AAiT. Chapter 2. Measurement Error Analysis

Experimental Uncertainty (Error) and Data Analysis

Uncertainty, Error, and Precision in Quantitative Measurements an Introduction 4.4 cm Experimental error

SIGNIFICANT FIGURES. x 100%

ents & Uncertainties Significant Figures 1.005, Round best to the experimental want to meters and use 300 m 2. significant figures because of

Measurements and Data Analysis

1 Measurement Uncertainties

EXPERIMENT 5 THE ASSAY OF ASPIRIN

EXPERIMENT 30A1: MEASUREMENTS. Learning Outcomes. Introduction. Experimental Value - True Value. 100 True Value

Chapter 2. Theory of Errors and Basic Adjustment Principles

Propagation of Uncertainty

11.1 Uncertainty and error in measurement (1 Hour) 11.2 Uncertainties in calculated results (0.5 Hour) 11.3 Graphical techniques (0.

The AP Chemistry Summer assignment is meant to help prepare you for the first few weeks of class

that relative errors are dimensionless. When reporting relative errors it is usual to multiply the fractional error by 100 and report it as a percenta

Density of Aqueous Sodium Chloride Solutions

1 Measurement Uncertainties

CHM Accuracy, Precision, and Significant Figures (r14) C. Taylor 1/10

SPH3U1 Lesson 03 Introduction. 6.1 Expressing Error in Measurement

A.0 SF s-uncertainty-accuracy-precision

Measurement: The Basics

PURPOSE: To determine the Rate Law for the following chemical reaction:

REVIEW OF LAB TECHNIQUES

REVIEW OF LAB TECHNIQUES

Decimal Scientific Decimal Scientific

Appendix B: Skills Handbook

Titration of an Unknown Acid

Metric Prefixes UNITS & MEASUREMENT 10/6/2015 WHY DO UNITS AND MEASUREMENT MATTER?

Liquid-in-glass thermometer

Determination of the K a Value and Molar Mass of an Unknown Weak Acid

Notes Errors and Noise PHYS 3600, Northeastern University, Don Heiman, 6/9/ Accuracy versus Precision. 2. Errors

Dr. Kevin Moore CHM 111

Uncertainty in numbers

Spectrophotometric Determination of Ferrocyanide in Effluents

MEASUREMENTS AND ERRORS

The periodic table currently lists 116 different atoms. New atoms are being discovered.

Glossary of Common Laboratory Terms

CEM 333 Instrumental Analysis

CHEM 334 Quantitative Analysis Laboratory

Since the publication of the ISO Guide to the Expression

5. Statistical Evaluation of Acid-Base Indicators 1

Measurement. New Topics accuracy vs. precision rounding in chemistry significant figures determining uncertainty of a measurement % error moles - 1 -

Experimental Uncertainty (Error) and Data Analysis

Guide to the Expression of Uncertainty in Measurement (GUM)- An Overview

And how to do them. Denise L Seman City of Youngstown

Engineering Fundamentals and Problem Solving, 6e. Chapter 6 Engineering Measurements

For this lab, you will determine the purity of the aspirin by titration and by spectrophotometric analysis.

FUNDAMENTAL CONCEPTS IN MEASUREMENT & EXPERIMENTATION (continued) Measurement Errors and Uncertainty:

Measurement And Uncertainty

3. Measurement Error and Precision

PART I: MEASURING MASS

Experiment 20-Acid-Base Titration: Standardization of KOH and Determination of the Molarity and/or Percent Composition of an Acid Solution

Cyanide, colorimetric, pyridine-pyrazolone

Calibration of Volumetric Glassware. Prepared by Allan Fraser May 2016 APPLICATION Note 1

Statistical Analysis of Chemical Data Chapter 4

Volumetric Analysis: Analysis of antacid tablets Analysis of Cl - concentrations in IV solutions

Transcription:

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR

Chapter 3. Experimental Error -There is error associated with every measurement. -There is no way to measure the true value of anything. - The best way we can do in a chemical analysis is to apply a technique that experience tells us is reliable. - Repetition of one method of measurement several times tells us the precision (reproducibility) of the measurement - If the results of measuring the same quantity by different methods agree with one another, then we become confident that the results are accurate, which means they are near the true value.

Chapter 3. Experimental Error 3-1. Significant Figures The number of significant ifi tfigures is the minimum i number of fdigits it needed dto write a given value in scientific notation without loss of accuarcy. Fig. 3-1. Scale of a spectrophotometer. p Absorbance is a logarithmic scale. Absorbance = 0.34 by interpolation. and 3 are completely certain and the number 4 is an estimate. This number has three significant figures.

Chapter 3. Experimental Error 3-. Significant Figures in Arithmetic ( rule of thumb ) 1) Addition & Substraction - In the addition or subtraction of numbers expressed in scientific notation, all numbers should first be expressed with the same exponent. - If the numbers being added do not have the same number of significant figures, we are limited by the least certain one. Rounding off - In the special case where the number is exactly halfway, round to the nearest even digit. ex). 1.45 1.4, 43.550000 43.6, When rounding off, look at all the digits beyond the last place desired. ex). 11.79500001 11.80 11.80500001 11.81

Chapter 3. Experimental Error 3-. Significant Figures in Arithmetic ( rule of thumb ) ) Multiplication & Division We are normally limited to the number of digits contained in the number with the fewest significant figures. But use the rule of thumb with Caution!! A rule of thumb sometimes suggested for multiplication and division that the answer should be rounded so that it contains the same number of significant digits as the original number with the smallest number of significant digits. Unfortunately, this procedure can lead to incorrect rounding. For example, consider the two calculations 4 4.5 100.0 = 1.08 and 4 4.0 = 100.0 0.965 B h l j d ibd h fi ldb dd 11 d By the rule just described, the first answer would be rounded to 1.1 and the second to 0.96.

Use the rule of thumb with Caution!! However, if the last digit of each number making up the first quotient is uncertain by 1, the relative uncertainties associated with each of these numbers are 1/4, 1/45, and 1/1000. Because the first relative uncertainty is much larger than the other two, the relative uncertainty in the result is also 1/4; the absolute uncertainty is then 1.08 1/4 = 0.045 = 0.04 By the same argument, the absolute uncertainty of the second answer is given by 0.965 1/4 = 0.040040 = 004 0.04 Therefore, the first result should be rounded to three significant figures, or 1.08, but the second should be rounded to only two; that is, 0.96 * Rounding should only be done on the final answer (not intermediate results), g y ( ) to avoid accumulating round-off errors

3-. Significant Figures in Arithmetic 3) Logarithms and Antilogarithms * Number of digits in mantissa of log x = number of significant figures in x : log (5.403 10-8 ) =-7.674 4 digits character mantissa * Number of digits in the mantissa of x = Number of significant figures in the antilogarithm, antilog x (=10 x ) antilog 6.14 = 10 6.14 = 1.39 10 6 mantissa, 3 digits 3 digits

3.3. Significant Figures and Graphs - When the graph is used to read data points, better is a fine grid superimposed on the graph - Rulings on graph should be compatible with the number of significant figures of the coordinates. a) Divisions are fine enough (0.53, 0.63), (1.08, 1.47)

3.3. Significant Figures and Graphs - When the graph is used to read data points, better is a fine grid superimposed on the graph -Rulings on graph should be compatible with the number of significant figures of the coordinates. b) Divisions are not fine enough (0.53, 0.63), (1.08, 1.47)

3.3. Significant Figures and Graphs When you are not expected to be able to read coordinates accurately on the graph, Qualitative behavior of data is OK (i.e., fine grid is not required)

3-4. Types of Errors in Experimental Data 1) Systematic errors (= Determinate Errors) - It is always in the same direction (unidirectional), and could be discovered and corrected. -It causes the mean of a set of data to differ from the accepted value ( Next slide, analyst 3 ) ) Random Error ( = Indeterminate error) 3) Gross Errors

Systematic error (Analyst 3 & 4) Fig 5-3, p.94

3-4. Types of Errors in Experimental Data 1) Systematic errors (= Determinate Errors) i) Instrument t Errors - imperfections in measuring devices. pipets, burets and volumetric flasks frequently deliver or contain volumes slightly different from those indicate by their graduations. The reasons are as following: Use of glassware at a temp. that differs significantly from the calibration temp. Distortions in container walls due to heating while drying. Errors in the original calibration. Contaminants on the inner surfaces of the container. Most systematic errors of this type are readily eliminated by calibration ( see next slide, textbook Fig. 3-3) - instruments powered by electricity. decreased voltage of battery-operated power supplies increased resistance in circuits because of dirty electrical contacts temp. effects on resistor

3.3. Significant Figures and Graphs Most systematic errors of this type are readily eliminated by calibration! Ex.) Buret reading: 9.43 ml, Correction: - 003mL 0.03 Actual volume = 9.43-0.03 =9.40 ml Fig. 3-3 Calibration Curve for a 50 ml buret

3-4. Types of Errors in Experimental Data ii) Method Errors Nonideal chemical or physical behavior of the reagents and reactions upon which an analysis is based. incompleteness of the reaction (ex: decomposition of pyridine ring) slowness of the reaction possible occurrence of side reactions that interfere with the measurement process in volumetric titration, the small excess of reagent required to cause an indicator to undergo the color change that signals completion of the reaction Errors inherent in a method are frequently difficult to detect, and are thus the most serious of the three types of determinate error.

3-4. Types of Errors in Experimental Data iii) Personal Errors Many measurements require personal ljudgments. - Prejudice (bias) Most of us have a natural tendency to estimate scale readings in a direction that improves the precision in a set of results or causes the results to fall closer to a preconceived notion of the true value. level of a liquid in buret color at the end point in titration the position of a pointer between two scales

3-4. Types of Errors in Experimental Data iii) Personal Errors Many measurements require personal ljudgments. - Number bias prefer 0 or 5 even number over odd number. * Detection of determinate errors. 1. Analysis of standard reference materials. a second reliable analytical method 3. Blank determinations 4. Variation in sample size in case of a constant error

3-4. Types of Errors in Experimental Data * Determinate Error may be either constant or proportional Constant Errors : magnitude is independent of size of the quantity measured. Ex). 500 mg of precipitate 0.5 mg wash out 50 mg of precipitate 0.5 mg wash out 0.5 relative error = - 100 = - 0.1 % 500 0.5 relative error = - 100 = -1.0% 50 Proportional Errors : increase or decrease in proportion to the size of the sample taken for analysis. Ex). the presence of interfering contaminants in the sample.

Absolute and Relative Uncertainty - Uncertainty is usually expressed as the standard deviation. Absolute Uncertainty : margin of uncertainty associated with a measurement. Ex). a buret reading : 1.35±0.0 ml Relative Uncertainty : comparing the size of the absolute uncertainty to the size of its associated measurement. R.U = absolute uncertainty magnitude of measurement 0.0 ml ex). RU R.U = = 0.0000 = ppt = 0 0. % 1.35 ml

3-4. Types of Errors in Experimental Data ) Random Error ( = Indeterminate error) - It arises from natural limitation on our ability to make physical measurements. - It causes data to be scattered more or less systematically around a mean value. The precision of the data reflects the indeterminate errors in an analysis. - Sometimes positive, sometimes negative. - It is the ultimate limitation on the determination of a quantity. See next slide Ex). Electrical noise : small fluctuation resulting from electrical instability of the meter itself (voltameter)

3-4. Types of Errors in Experimental Data 3) Gross Errors - personal and arise from carelessness or ineptitude on the part of the experimenter. - Gross errors usually affect only a single result in a set of replicate data, causing it to differ significantly from the remaining results for that set. (outliers) Ex). Arithmetic mistakes Reading scale backward Reversing a sign Spilling a solution - They can be eliminated through self-discipline

3-4-1 Methods For Expressing Precision and Accuracy Precision : A measure of reproducibility of a result defined as the agreement between the numerical values of two or more measurements that have been made in an identical method. Accuracy : How close a measured value is to the true one Accuracy : How close a measured value is to the true one - the nearness of a measurement to its accepted value - Accuracy is expressed in terms of error.

Fig 5-3, p.94 Precision and Accuracy

Accuracy (= nearness to the truth ) Methods for expressing accuracy 1) Absolute Error = x i x t x i : observed value x t : accepted value (It may itself lfbe subject to considerable uncertainty) ) Relative Error = x x x i t t 100 %

Precision (= reproducibility) Methods for expressing precision x i i Mean or Average = x = N Geometric Mean = n i x i 1) Standard Deviation For a very large set of data σ = N i= 1 ( x μ ) i N σ: population S. D. μ: population p mean ( = true value) σ : variance

Methods for expressing precision For a small number of replicate measurements, S = N i= 1 ( x x) i N -1 S : sample standard deviation x : measured mean for the small set ) Relative Standard d Deviation (RSD) S RSD = x 1,000 (ppt) S CV (coefficient of variation) = 100 % x

Methods for expressing precision 3) Variance (S ) S = N i= 1 ( x x) i N -1 4) Spread or Range (w) = highest lowest

3-5. Propagation of Uncertainty from Random Error - We can easily estimate or measure the random error associated with a measurement. - Uncertainty is usually expressed as the standard deviation of calculated results or as a confidence interval. These parameters are based upon a series of replicate measurements. -In most experiments, it is necessary to perform successive arithmetic operations on several numbers, each of which has its associated random error. - The most likely uncertainty in the result is not the sum of individual errors, because some of these are likely positive and some negative. We expect some cancellation of errors. - The first uncertain figure should be the last significant figure : ex). 0.0946 (±0.000)

3-5. Propagation of Uncertainty from Random Error 1) Addition and Subtraction 176( 1.76 (±0.03) 03) e 1 Uncertainty in addition + 1.89 (±0.0) e and subtraction: - 0.59 (±0.0) e 3 3.06 (± e 4 ) e 4 = e1 + e + e3 = 0.041 Absolute uncertainty e 4 = ( 0.03) + (0.0) + (0.0) = 0. 041 0.041 Percent relative uncertainty = 100 = 1. 3% 3.06 Two expressions of final result : 3.06 (±0.04) (absolute uncertainty) 3.06 (±1%) (relative uncertainty)

3-5. Propagation of Uncertainty from Random Error ) Multiplication and Division Uncertainty in multiplication and division: For example, 1.76( ± 0.03) 1.89( ± 0.0) = 0.59( ± 0.0) 0) % e + e 5.64 ± e 4 = (% e1 ) + (% e ) (% 3 ) First, convert absolute uncertainties to percent relative uncertainties. 4 1.76( ± 1. 7 %) 1.89( ± 1. 0.59( ± 3. 4 %) 1 %) = 5.64 ± e 4 % e4 = (1. 7 ) + (1. 1) + (3. 4 ) = 4. 0% 4. = 0 % 5.6 4 = 0.0404 0 5.6 4 0. 3 Two expressions of final result : 56(±0)( 5.6 (±0.) (absolute uncertainty) 5.6 (±4%) (relative uncertainty)

tip) Propagation of Uncertainty in the Product x x Table 3-1 says that the uncertainty in the function y = x a is % e y = a (%e x ). If y = X, then %e y = (%e x ). A 3% uncertainty in x leads to a ()(3%) = 6% uncertainty in y. But what if we just apply the multiplication li formula 3-6 to the product x x? ( x ± e ) x( ± e ) = x ( ± ) ( 1 e3 % e + e 3 = (% e 1 ) (% ) = ( 3%) + (3%) = 4. % Which uncertainty is correct, 6 % or 4. %?

tip) Propagation of Uncertainty in the Product x x

3-6. Propagation of Uncertainty from Systematic Error Systematic error occurs in some common situations and is treated differently from random error in arithmetic operation 1) The Rectangular Distribution : uncertainty in atomic mass - atomic mass of oxygen in the periodic table : 15.9994 ± 0.0003 ( Next slide) - The uncertainty t isnot mainly from random error in measuring atomic mass, but it is predominantly from isotopic variation in samples of oxygen from different sources. Therefore, the atomic mass of oxygen in a particular lot of reagent has a systematic uncertainty - There is approximately equal probability of finding any atomic mass between 15.9991 and 15.9997 and negligible probability of finding an atomic mass outside of the range. - Mean: 15.9994, range : 0.0003 - The standard uncertainty (standard deviation) = ± a (range)/ 3

End Papers: Periodic Table

3-6. Propagation of Uncertainty from Systematic Error -Oxygen atomic mass from different sources approximates a rectangular distribution. - There is approximately equal probability bili of finding any atomic mass between 15.9991 and 15.9997 and negligible probability of finding any atomic mass outside of the range (a).

3-6. Propagation of Uncertainty from Systematic Error - Systematic error occurs in some common situations, and is treated differently from random error in arithmetic operations 1) The Rectangular Distribution : uncertainty in atomic mass - The periodic table gives the atomic mass of oxygen as 15.9994 ± 0.0003g/mol. The uncertainty is not mainly from random error in measuring the atomic mass, but it is predominantly from isotopic variation in sampled oxygen from different sources. - The atomic mass of oxygen in a particular lot of reagent has a systematic uncertainty.

3-6. Propagation of Uncertainty from Systematic Error - Propagation of systematic uncertainty : uncertainty of n identical atoms = n x (standard uncertainty in atomic mass) - The sum of atomic masses of different elements : Use the rules for propagation of random uncertainty Example: what is the standard uncertainty in molecular mass of C H 4? From periodic table, Atomic mass of C = 1.0107 ± 0.0008/ 3 =1.0107 ± 0.00046 Atomic mass of H = 1.00794 ± 0.00007/ =1.00794 ± 0.000040 3

3-6. Propagation of Uncertainty from Systematic Error -For the uncertainty in the sum of the masses of C +4H, we use equations for random error, because the uncertainties in the mass of C and H are independent of each other. One might be positive and one might be negative. 8.053 16 ± 0.0009 + 0.00016 = 8.053 16 ± 0.000000 93 = 8.053 ± 0.000 9

3-6. Propagation of Uncertainty from Systematic Error ) The Triangular Distribution : multiple deliveries from one pipet - A 5 ml pipet is certified to deliver 5.00 ± 0.03 ml - The volume delivered by a pipet is reproducible, but can be in the range of 4.97 ~ 5.03 ml, depending on a pipet. p - However, the manufacturer works hard to make the volume close to 5.00mL. In such case, we approximate the volumes of a large number of pipets by the triangular distribution.

3-6. Propagation of Uncertainty from Systematic Error - The probability falls off approximately in a linear manner as the volume deviates from 5.00 ml because the manufacturer works hard to make the volume close to 5.00mL. - There is negligible probability that a volume outside of 5.00 ± 0.03 ml will be delibered. - The standard uncertainty (standard deviation) = ± a (range)/ 6 3

3-6. Propagation of Uncertainty from Systematic Error Example: 1) What is the uncertainty in 100 ml if you use an uncalibrated 5 ml pipet four times to deliver of 100 ml? The uncertainty is a systematic error. So the standard uncertainty = ± 4x 0.03/ 003/ 6 = ± 4 x 0.01 = ± 0.048 ml So uncalibrated pipet volume = 100.00 ± 005 0.05 ml - The difference between 5.00 ml and the actual volume delibered by a particular pipet is a systematic error. In other words, it is always the same, within a small random error - You could calibrate the pipet by weighing the water it delivers to eliminate a systematic error. In such case, the pipet always delivers, say, 4.991 ± 0.006 ml. The uncertainty (± 0.006 ml) is random error.

3-6. Propagation of Uncertainty from Systematic Error Example: 1) What is the uncertainty in 100 ml if you use a calibrated pipet which delivers a mean volume of 4.991 ml with a standard uncertainty of ± 0.006 ml? You deliver four times, so the volume is 4 x 4.991 ml = 99.964 964 ml But the uncertainty = ± 0.006 + 0.006 + 0.006 + 0.006 = ± 0.01 ml We use equations for random error, because the uncertainties in four delivered aliquots are independent of each other. One might be positive and one might be negative because they are random error. So calibrated pipet volume = 99. 964 ± 0.01 ml