The Agilent 1290 Infinity II Multi Method Solution

Similar documents
Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Determination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infinity Analytical SFC System

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Application Note. David A. Weil Zoltan Timar Michael Zumwalt Edgar Naegele. Abstract

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Application Note. Authors. Abstract. Food Testing & Agriculture. Stephen E. Reichenbach University of Nebraska Lincoln, NE, USA

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Fast Screening Methods for Steroids by HPLC with Agilent Poroshell 120 Columns

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

The new Agilent 1290 Infinity Quaternary LC

Application Note. Food. Abstract. Author

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Fast Screening Methods for Analgesics and Non-Steroidal Anti-Inflammatory Drugs by HPLC

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Agilent LC & LC-MS Cannabis Testing

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

LC and LC/MS Column Selection Flow Chart

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Application Note. Abstract. Authors. Pharmaceutical. William J. Long Agilent Technologies, Inc Centerville Road Wilmington, DE USA

Method Development in Comprehensive 2D-LC Finding the Most Orthogonal Separation Systems for RPLC RPLC Using Column and Solvent Screening

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

Application Note. Edgar Naegele. Abstract

Fast Screening Methods for Beta Blockers by HPLC with Agilent Poroshell 120 Columns

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Automated LC Method Development: Fundamentals and Columns, Hardware, and Software

Analysis of Positional Isomers with Agilent Poroshell 120 PFP Columns

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Agilent ESI and APCI sources: for polar to non-polar compounds

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

Fast methods for the determination of ibuprofen in drug products

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Agilent 385-ELSD Evaporative Light Scattering Detector

Determination of Polymer Modifier in Asphalt

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Successfully Scaling and Transferring HPLC and UPLC Methods

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Dionex IonPac AS28-Fast-4µm column

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

COC Biotechnology Program

Agilent 1260 Infinity Evaporative Light Scattering Detector (ELSD)

Transcription:

The Agilent 129 Infinity II Multi Method Solution Switch between seven applications on a single LC system by using automatic column and solvent selection Application ote Food Testing & Agriculture Author Edgar aegele Agilent Technologies, Inc. Abstract This Application ote describes the use of the Agilent 129 Infinity II LC for the automated operation of seven different applications using different methods (different stationary and mobile phases) in food analysis. The exchange of columns is controlled by the method used. All columns are located in a single Agilent 129 Infinity II Multicolumn Thermostat. Accordingly, the Agilent 129 Infinity II Flexible Pump is attached to two external solvent-selection valves for automated solvent switching, according to the method.

Introduction A typical laboratory that analyzes food quality offers a large variety of measurements for different applications, for example, mycotoxins, parabens, and so on. These applications are usually measured on one UHPLC instrument, making it necessary to exchange columns and solvents to adapt the instrument to the desired application. If these adaptations are done manually, analyzing a larger variety of analytes is timeconsuming. If an application finishes after the working day, the instrument waits for hours or even days for manual interaction and a restart. This can be overcome by using an instrument that employs multiple columns with multiple solvents, such as the Agilent 129 Infinity II LC. This system can operate up to eight columns with up to 26 different solvents, enabling more than 1, different analytical conditions. The individual combination of solvent and column is chosen by a predefined method optimized for the analysis of the desired analytes, making it easy to automatically change the analysis conditions within a sequence for different analyses. The system we describe comprises two highly sophisticated modules from the Agilent 129 Infinity II LC, the Agilent 129 Infinity II Multicolumn Thermostat (MCT) and the Agilent 129 Infinity II Multisampler, which enhance the system so that it delivers the highest possible performance. The 129 Infinity II MCT enables precise column thermostatting over a broad temperature range. Heat exchange is done by heat exchangers with the lowest dead volume and highest efficiency, which enables excellent retention time stability for highly reproducible retention times 1. The 129 Infinity II Multisampler, with its advanced design elements such as sample hotel, needle handling routines, and carryover reduction, handles a large number of different samples and analyte types without interactions 2. This allows optimum performance for a multimethod system. We describe the use of the 129 Infinity II LC with an eight-column selection valve located in the 129 Infinity II MCT, and two solvent-selection valves attached to the 129 Infinity II Flexible Pump for multimethod analysis. Experimental Instrumentation The Agilent 129 Infinity II LC solution for multimethod analysis comprises: Agilent 129 Infinity II Flexible Pump (G714A) Agilent 129 Infinity II Multisampler (G7167B) Agilent 129 Infinity II Multicolumn Thermostat MCT (G7116B) with valve drive (G7116B#58) equipped with Agilent Quick Change eight column selector valve (G4239C) and capillary kit (G4239C#5, p/n 567-4248). This kit contains all Agilent Quick Connect heat exchangers (standard flow) and capillaries needed for the installation of up to eight columns. Agilent 129 Infinity II (G7117B) Agilent 129 Infinity External Valve Drive (2 x G117A) with Quick Change solvent selector valve (2 G4235A) Software Agilent penlab A.2.2 CDS, ChemStation Edition for LC and LC/MS Systems, Rev. C.1.7 Instrument setup The columns used for all applications were introduced in the column list in ChemStation (Figure 1). In the ChemStation column list, open a new line by clicking Append to add a new column. For the columns, a description, geometric data, particle size, and limitations such as ph, pressure, and temperature must be given. Installed analytical columns are indicated in the table by YES. Figure 1. The Agilent ChemStation column list that provides an overview of all columns available in the laboratory. 2

The column list is directly connected to the column assignment in the 129 Infinity II MCT (Figure 2). In the assignment, the position of the column on the right or left can be chosen and associated with a color code. The column data from the column list can be selected under Column Tag Information and associated to a column position and color code. The column used for the individual method can be selected in the Method screen of the 129 Infinity II MCT (Figure 3). The appropriate column is chosen either by the pull-down menu, which shows all assigned columns, or by just clicking the column with the right color code in the image of the 129 Infinity II MCT. The current valve position, which then connects automatically to the chosen column, is shown. For quick information, the valve position, the color code of the chosen column, and its product number are shown in the ChemStation user interface (Figure 4). Figure 2. Column assignment in the Agilent 129 Infinity II MCT. Each column is assigned to an unambiguous location in the MCT. Figure 3. Column selection in the Method screen of the Agilent 129 Infinity II MCT. 3

The assignment of the solvents is done in a similar way. The solvent selection valves are assigned to the pump channels in the instrument configuration (not shown). In the Pump Valve Cluster Configuration screen, the positions of the solvent selection valve are named by the connected solvent (Figure 5). The compressibility calibration is chosen by taking the correct solvent or type of solvent from the pull-down menu. Molarity and ph values can be added. In the Method of Pump Valve Cluster screen, the right solvents for channels A and B are chosen and the valve automatically switches to the correct positions (Figure 6). The accessible multiple solvents are sketched out in the ChemStation pump user interface (Figure 4). Figure 4. Instrument user interface in Agilent ChemStation, highlighting the active column. Figure 5. The Pump Valve Cluster Configuration screen. 4

Methods The detailed methods for the analysis of the individual compounds are in Results and Discussion. The main analysis conditions for each group of analytes are summarized in Table 1. Figure 6. The Method of Pump Valve Cluster screen. Table 1. Summary of main analytical conditions for each group of analytes. Compound Column Mobile phase A Mobile phase B Temperature Preservatives Benzoic acid Salicylic acid Citric acid Parabens Methyl paraben Ethyl paraben Propy paraben Butyl paraben Antiparasitic Malachite green Coloring additives Sudan Red I Sudan Red II Sudan Red III Sudan Red IV Antimicrobial Chloramphenicol Mycotoxin Patulin Antioxidants PG TBHQ BHA BHT Agilent ZRBAX RRHD Eclipse C8, 2.1 1 mm, 1.8 µm (p/n 959764-96) Agilent Poroshell 12 C18, 3. 1 mm, 1.8 µm (p/n 695975-32) Agilent ZRBAX RRHD SB C18, 2.1 5 mm, 1.8 µm (p/n 8277-92) Agilent ZRBAX RRHD Eclipse plus C18, 2.1 1 mm, 1.8 µm (p/n 959758-92) Agilent ZRBAX RRHD SB C8, 2.1 1 mm, 1.8 µm (p/n 8287-96) Agilent ZRBAX RRHD Extend C18, 2.1 1 mm, 1.8 µm (p/n 7287-92) Agilent ZRBAX RRHD Eclipse C18, 2.1 5 mm, 1.8 µm (p/n 959741-92) Water +.5% TFA ph 2.1 Water + 5 mm H 4 Ac ph 6.73 Water +.1% formic acid ph 2.7 Water +.1% formic acid ph 2.7 Acetonitrile:isopropanol (1:1) +.5% TFA 4 C Methanol 45 C Acetonitrile +.1% formic acid Acetonitrile +.1% formic acid 4 C 3 C Water Methanol 3 C Water +.1% H 4 H ph 1.3 Water +.5% acetic acid ph 3.43 Acetonitrile +.1% H 4 H 4 C Acetonitrile 4 C 5

After an exchange of a method in a sequence of samples, which includes the exchange of the column and the solvents, it is necessary to flush the system with the new solvents and equilibrate the column on the starting conditions of the method. Generation of solvent exchange methods The exchange of the solvent can be done by taking the analytical method for the desired application. The solvent composition was set to 5:5 and the column selection valve switched to the bypass position (a capillary was installed on position eight instead of a column). Under these conditions, the solvents were run through the system at 5 ml/min for 4 minutes isocratically and without injection. The created method can be stored separately and used in sequence. Generation of column equilibration methods For the equilibration of the column, the analytical method was taken and run without injection under isocratic conditions at the starting solvent composition. The run time was the same as used for re-equilibration between sample runs. The created method can be stored separately and used in sequence. Chemicals All chemicals were purchased from Sigma Aldrich, Corp., Germany. All solvents were purchased from Merck, Germany. Fresh ultrapure water was obtained from a Milli-Q Integral system equipped with LC-Pak Polisher and a.22-μm membrane point-of-use cartridge (Millipak). Results and Discussion The following analyses can be done in a single sequence by exchanging the column and the solvents with the method. After each exchange of solvents, a solvent wash method must be run. In addition, after an exchange of the column, the chosen column must be conditioned using the following starting conditions with the respective method. The analysis methods and results belong to preservatives, parabens, antiparasitic drugs, coloring additives, antimicrobial drugs, mycotoxins, and antioxidants (Figures 7 to 13). These compounds can be found in food and beverages. Analysis of preservatives Preservatives are used to prevent the growth of fungi and bacteria in food and beverages. Typical compounds are sodium benzoate (E211), salicylic acid, and sorbic acid (E2). Injection volume 1 µl Sorbic acid, benzoic acid, salicylic acid, 3 mg/1 ml methanol each, 1:1 dilution 5 to 5% B in 1 minutes 1 minutes 5 minutes Column temperature 45 C 25 2 15 1 5 H 3 C.4 ml/min 25/4 nm, Ref. 36/1 nm, data rate 1 Hz, path length 1 mm, slit 8 nm H Sorbic acid Benzoic acid 5.516 5.399 6.142 Salicylic acid 2 3 4 5 6 7 8 9 H H H Time (min) Figure 7. Separation of the food preservatives sorbic acid, benzoic acid, and salicylic acid. 6

Parabens Parabens are typically used as preservatives in cosmetics and pharmaceuticals. Widely used parabens are methyl paraben (E218), ethyl paraben (E214), propyl paraben (E216), and butyl paraben. Injection volume Column temperature Methyl, ethyl, propyl, butyl paraben, 1 mg each, dissolved in 1 ml acetonitrile, diluted 1:1 with acetonitrile 4 to 55% B in 1 minutes 1 minutes 5 minutes.5 ml/min 1 µl 45 C 254/4 nm, Ref. 36/1 nm, data rate 1 Hz, path length 1 mm, slit 8 nm 5 4 3 2 1 Methyl paraben 2.682 Ethyl paraben 4.26 Propyl paraben 5.621 Butyl paraben 7.162 1 2 3 4 5 6 7 8 9 H R R = Methyl, ethyl, propyl, or butyl Figure 8. Separation of the preservatives methyl, ethyl, propyl, and butyl paraben, widely used in cosmetics and pharmaceuticals. Time (min) Antiparasitic drugs Malachite green is a typical antiseptic or antiparasitic drug used against fungal and bacterial infections, especially in fish and fish eggs. It is banned from use in breeding fish for human consumption because of health concerns. Injection volume Column temperature Malachite green, 2 mg, dissolved in methanol, 1 ml, diluted 1:1 with methanol 5 to 95% B in 5 minutes, 2 minutes at 95% B 7 minutes 3 minutes.5 ml/min 1 µl 4 C 55 nm/1 nm, Ref. off, data rate 2 Hz, path length 1 mm, slit 8 nm 45 4 35 3 25 3.952 + Cl 2 H 3 C CH 3 15 CH 3 CH 3 1 5 2 3 4 5 6 Time (min) Figure 9. Determination of the antiparasitic drug malachite green, banned from use in fish farming. 7

Coloring additives From the class of azo dyes, the Sudan Red colors have been used to colorize food products such as paprika, curry, and chili powder. However, the use of these Sudan Red colors is now banned as use as food colorizers. In particular, Sudan Red I, Sudan Red II, and Sudan Red IV are classified carcinogens by the International Agency for Research on Cancer. Injection volume Column temperature Sudan Red I, Sudan Red II, Sudan Red III, Sudan Red IV, 2 mg each, dissolved in 2 ml methanol, diluted 1:1 in methanol 8 to 95% B in 1 minutes 1 minutes 5 minutes.4 ml/min 1 µl 3 C 5 nm/1 nm, Ref. off, data rate 1 Hz, path length 1 mm, slit 8 nm 6 5 4 3 2 1 Sudan Red I 2.151 H H 3 C Sudan Red II 3.844 Sudan Red III 5.37 Sudan Red IV 1 2 3 4 5 6 7 8 9 Time (min) Figure 1. Separation of Sudan Red I, II, III, and IV, formerly used as coloring additives in spices such as chili and paprika, but now banned. CH 3 H H CH 3 8.568 CH 3 H Antimicrobial drugs Chloramphenicol is an antibacterial and antimicrobial drug. Due to its broad-spectrum antibiotic behavior, it is used to protect livestock and can be found in food. Injection volume Column temperature Chloramphenicol, 2 mg, dissolved in 2 ml methanol, diluted 1:1 with methanol 1 to 9% B in 1 minutes 1 minutes 5 minutes.3 ml/min 1 µl 45 C 254/4 nm, Ref. 36/1 nm, data rate 1 Hz, path length 1 mm, slit 8 nm 9 H H 6.629 8 Cl + 7 H 6 Cl 5 4 3 2 1 1 2 3 4 5 6 7 8 9 Figure 11. Determination of the broad-spectrum antibiotic chloramphenicol. Time (min) 8

Mycotoxins The mycotoxin patulin, which is suspected to be genotoxic, is found in rotting apples. Therefore, apple products must be analyzed for this mycotoxin. The World Health rganization recommends limits below maximum concentration of 5 ng/l in, for example, apple juice. 2.2 H Injection volume Column temperature Patulin, 5 mg, dissolved in 1 ml acetonitrile, diluted 1:2 with acetonitrile 5% B for 3 minutes, at 3 minutes to 9% B 6 minutes 3 minutes.5 ml/min 1 µl 3 C 276/4 nm, Ref. 36/1 nm, data rate 1 Hz, path length 1 mm, slit 8 nm 4 Patulin 3 2 1 1 2 3 4 5 Figure 12. Determination of the mycotoxin patulin. Time (min) Antioxidants Antioxidants such as butylated hydroxyanisole (BHA, E32) and butylated hydroxytoluene (BHT) are primarily used as food additives. They act as scavengers and prevent free radical reactions. Propyl gallate (PG, E31) is used as an additive in oily and fatty food to reduce oxidation. For unsaturated vegetable oils, tert-butylhydroquinone (TBHQ, E319) is a very effective antioxidant that enhances storage lifetime. Injection volume Column temperature PG, BHT, BHA, TBHQ, 1 mg each, dissolved in 1 ml acetonitrile 1 to 95% B in 1 minutes 1 minutes 5 minutes.5 ml/min 1 µl 4 C 24/4 nm, Ref. 36/1 nm, data rate 1 Hz, path length 1 mm, slit 8 nm 14 12 1 8 6 4 2 H H H PG 2.895 H BHA 3.842 5.62 H TBHQ H BHT 8.77 H 1 2 3 4 5 6 7 8 9 Time (min) Figure 13. Separation of PG, BHA, BHT, and TBHQ, used as antioxidation additives in foods and cosmetics. 9

Conclusions This Application ote demonstrates the capability of the Agilent 129 Infinity II Multi-Method Solution to support multiple LC methods on one system (multimethod analysis) as needed, for example, in food analysis laboratories. Equipped with the Agilent 129 Infinity II Multicolumn Thermostat, this configuration provides access to up to eight different columns. In addition, with two solvent selection valves attached to the Agilent 129 Infinity II Flexible Pump, the configuration runs up to 26 different solvents as mobile phase. Hence, it is possible to create more than 1, different separation methods. All LC methods can be run automatically in one sequence by an automated switch of columns and solvents. The large number of samples is handled by the Agilent 129 Infinity II Multisampler. References 1. Anon. Performance Characteristics of the Agilent 129 Infinity II Multicolumn Thermostat; Technical overview, Agilent Technologies, Inc. Publication number 5991-5533E, 215. 2. Anon. Performance Characteristics of the Agilent 129 Infinity II Multisampler; Technical overview, Agilent Technologies, Inc. Publication number 5991-5348E, 214. 1

11

www.agilent.com/chem This information is subject to change without notice. Agilent Technologies, Inc., 215 Published in the USA, April 1, 215 5991-5679E