Forensic and toxicological drug screening using liquid chromatography-tandem mass spectrometry with MS/MS library based identification

Similar documents
Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

Application News. No. C154. Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Benzodiazepines and Barbiturates

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

Semi-quantitative Determination of Concentrations in Systematic Toxicological Analysis by LC-QTOF-MS

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

Liquid Chromatography Mass Spectrometry. Shimadzu Corporation, UK; 2 Phytocontrol, France; 3 Shimadzu France, France. Trimesium S + CH 3 ETU

N-(1-carbamoyl-2-methyl-propyl)-1-(5-chloropentyl)indazole-3- carboxamide

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

WADA Technical Document TD2003IDCR

5Cl-AKB48 and 5Br-AKB48

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

TargetScreener. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Methyl 1-[(4-fluorophenyl)methyl]indazole-3-carboxylate

Multi-residue analysis of pesticides by GC-HRMS

Food & Environmental. Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou SCIEX, Concord, Ontario, Canada

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS

Liquid Chromatography Mass Spectrometry. Expanding Capabilities in Multi-Residue Pesticide Analysis Using The LCMS-8060

Clinical, Forensic and Pharmaceutical Applications

Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou AB SCIEX, Concord, Ontario, Canada

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Methyl 2-[[1-(5-fluoropentyl)indole-3-carbonyl]amino]-3,3- dimethyl-butanoate

MassHunter TOF/QTOF Users Meeting

Overview. Introduction

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Liquid Chromatograph Mass Spectrometer LCMS-8040 C146-E188A

4-methyl-1-phenyl-2-pyrrolidin-1-yl-pentan-1-one

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Liquid Chromatograph Mass Spectrometer LCMS-8040 C146-E188

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

TOX. Scheme Description. Toxicology Proficiency Testing Scheme

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Quantitative Analysis and Identification of Migrants in Food Packaging Using LC-MS/MS

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Introduction. Abstract

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

Authors. Introduction. Abstract. Forensics

Theory and Practical Consequences.

isobaric Barbiturates in Serum using LDTD- MS/MS combined with differential mobility

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

2-[[2-(4-ethyl-2,5-dimethoxy-phenyl)ethylamino]methyl]phenol

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry

Screening of pesticides in water using SPE on line

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

A Case of Pesticide Poisoning: The Use of a Broad-Scope Tof Screening Approach in Wildlife Protection

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Journal of Forensic Toxicology & Pharmacology

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

AppNote 2/2013. A High Throughput Automated Sample Preparation and Analysis Workfl ow for Comprehensive Toxicology Urine Screenings using LC/MS/MS

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

1-(1,3-benzodioxol-5-yl)-2-(ethylamino)hexan-1-one

4-Methyldiethcathinone (4-MDEC)

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Application Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis

Agilent 6400 Series Triple Quadrupole LC/MS/MS Users Session

SUSPECT AND NON-TARGET SCREENING OF ORGANIC MICROPOLLUTANTS IN WASTEWATER THROUGH THE DEVELOPMENT OF A LC-HRMS BASED WORKFLOW

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Guide to Peptide Quantitation. Agilent clinical research

+ butanal + 2-picolineborane. + acetic acid. + butanal + 2-picolineborane. + acetic acid

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry

Journal of Analytical Toxicology Advance Access published April 24, 2012

Definitive EtG/EtS LC-MS/MS Analysis:

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1

Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid. Chromatography-Tandem Mass Spectrometry

2010 Waters Corporation. Xevo TQ-S for Bioanalysis

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

Transcription:

POCON696E Forensic and toxicological drug screening using liquid chromatographytandem mass spectrometry with MS/MS library based identification MSACL 6 EU Alan Barnes, Tiphaine Robin, Neil Loftus, Pierre Marquet, Sylvain Dulaurent, Franck SaintMarcoux Shimadzu, Manchester, UK, CHU Limoges, Limoges, France

Introduction MultiTargeted Screening (MTS) was applied to systemic toxicological analysis to reduce false positive and negative reporting using MS/MS spectral library based identification. The MTS method used threshold triggered multiple reaction monitoring (MRM) and MS/MS product ion scans at three collision energies to confirm the compound identification based on mass spectral library searching. The MS/MS library was created using certified reference materials and included electrospray spectral data from over compounds relevant to clinical and forensic toxicology in both positive and negative ion modes. The MTS approach was applied to screening whole blood samples at three concentration levels to evaluate screening at therapeutic, overdose and toxic concentrations. Methods MTS methods were developed to screen whole blood spiked with a range of commonly observed compounds including antidepressant compounds, anxiolytic drugs, analgesics and antipsychotic agents. Samples were prepared by QuEChERS method with inclusion of ten internal standard compounds to normalise sample matrix effects and quality control assessment. Data acquisition parameters were set to a single MRM per compound with threshold triggered MS/MS at 3 collision energies (, 35, 55V) enabling confirmation of parent ion (low) and fragment ions at medium and high CE voltages. Library searching was performed on all CE spectral data in addition to a mergedce spectrum. Internal standards deuterated compounds 3 antiepileptics, antipsychotics, barbiturates 36 benzodiazepines 5 cannabinoids. 3. 4. 5. 6. 7. 8. 9... Figure. MRM chromatograms for a panel of drugs extracted from whole blood using a QuEChERS method corresponding to a concentration of ug/l.

Table. LCMS/MS data acquisition conditions. The method included full scan and MRM data acquisition in both positive and negative ion mode. Liquid chromatography UHPLC : Nexera LC system Analytical column : Restek Raptor Biphenyl.7um x.mm Column temp. : 5 C Injection cycle : 5 µl injection volume Flow rate :.3 ml/min Solvent A : Water mm ammonium formate.% formic acid Solvent B : Methanol mm ammonium formate.% formic acid Mass spectrometry LCMS/MS : LCMS86 Ionisation mode : Heated ESI Scan speed : 3, u/sec Polarity switching time : 5 msec MRM Dwell time : 5 msec Pause time : 3 msec LCMS/MS method set up for simultaneous full scan and MRM data acquisition with polarity switching Type Event Polarity Name m/z Time (3mins) MRM 5 Target 7aminonitrazepam 5.>. 6 > CE:, 3.. 7 > CE: 35, 3.. 8 > CE: 5, 3.. MRM 9 Target 7aminoclonazepam 86.5>. > CE:, 3.. > CE: 35, 3.. Spectral Library > compounds The Forensic & Toxicological spectral library was created using authentic standard reference material at collision energies 6V. Compounds that ionised efficiently with more than one adduct state were saved resulting in 58 Library entries from 5 compounds (94 positive mode, 4 negative mode). Spectral Library information was registered for CE, 35 and 55V. Optimised MRM transitions were determined for all compounds with chromatographic retention time and peak area measured to enable reference ionratio calculation. RT analysis included internal standard compounds for relative RT calculation. Compound information was populated including: CAS number, formula, synonyms, compound class/properties, ChemSpider URL and ID number, mol file, InChI and InChIKey. 3

Results Using a MRM method followed by three product ion scans at different collision energies resulted in linear calibration curves over the concentration range of 55 ug/l (r >.996 for all compounds). With regard to accuracy and precision; accuracy was between 8% and precision <% throughout the calibration range. As a result of fast data scanning, the peak sampling rate resulted in more than data points across a peak. Area Ratio (x).5. Flurazepam Quantitation MRM Rt 7.37 mins 388.5>35. Area Ratio 3..5..5 Temazepam Quantitation MRM Rt 8.365 mins 3.5>55.5 Area Ratio..5. Quantitation MRM Rt 9.47mins 85.>93.5.5 Linear regression analysis R² =.9993 Matrix spiked calibration..5 Linear regression analysis R² =.9999 Matrix spiked calibration.5 Linear regression analysis R² =.9999 Matrix spiked calibration. 5 Concentration (ug/l) 5. 5 Concentration (ug/l) 5. 5 Concentration (ug/l) 5 Mean Accuracy Repeatability Rt (mins) MRM Calibration standards (n=3) for each calibration level ug/l replicate (n=5) ug/l replicate (n=5) 5ug/L ug/l ug/l 5ug/L Mean Conc (ug/l) %RSD Mean Conc (ug/l) %RSD Flurazepam 7.37 388.5>35. 9.5 95.5 95.3. 8. 3.. 7.7 Temazepam 8.365 3.5>55.5 3.4 97.8 99.. 8.8.9.5 6. 9.47 85.>93.5. 99.5 98.6.3 8.6.5. 5.6 The scope of the method was to ensure robust quantitation and a high level of confidence in the reported result. Using a MRM method followed by three product ion scans at different collision energies resulted in linear calibration curves over the concentration range of 55 ug/l (r >.996 for all compounds). With regard to accuracy and precision; accuracy was between 8% and precision <% throughout the calibration range. Using a pause time of 3msecs and a dwell time of 5msec, the scan time was set to 5msecs (3,u/sec scan speed, from 3u). As a result of fast data scanning, the peak sampling rate resulted in more than data points across a peak. 4

MRM triggered product ion spectrum Compound ID Molecular Formula C8H3ClFN3 CAS 5946778 Compound ID Molecular Formula C6H3ClNO CAS 43945 85 CE V CE 35V 3 9 49 9 9 87 58 36 63 89 65 36 9 44 9 Library Reference Spectrum 3 49 CE 35V 9 CE 35V 93 54 4 5 CE 55V 65 89 4 77 658 5 5 75 5 5 75 5 5 75 3 35 m/z Merged CE 55V 93 4 65 77 658 5 4 65 77 9 65 8 4 54 6 93 85 Library Reference Spectrum CE 55V 85 5 75 5 5 75 5 5 75 m/z Figure 3. MRM triggered product ion spectrum data for midazolam and diazepam. The library included spectra for each collision energy and a separate library for merged spectra enabling match criteria to be set for a specific fragmentation voltage (as shown for midazolam) or to use a broad band fragmentation and merged spectra (as in the case for diazepam). An MTS procedure for clinical and forensic toxicology screening was developed for a single LC/MS/MS method following a QuEChERS extraction of whole blood. This approach results in robust quantitation using MRM data and enables a higher degree of confidence in compound identification as shown in Table. 5

Table. Library search results for a panel of drugs spiked into whole blood and extracted by QuEChERS from three QC levels (low, medium and high QC s correspond to,, 5 ug/l). Most compounds can be identified as the first hit in a spectral based library match (5 compounds are identified as the second candidate in the library; for 4 compounds the hit was not identified as either the first or second candidate). Quality control level Quality control level Compound RT (min) Low Medium Library Hit High Compound RT (min) Low Medium Library Hit High Merged CE spectrum Merged CE spectrum Paracetamol 3. (amino5bromobenzoyl)pyridine 7.43 Levetiracetam 3.49 Dextropropoxyphene 7.44 Theophylline 3.67 Desalkylflurazepam 7.6 Scopolamine 4.5 Zolpidem 7.75 Felbamate 4.5 Hydroxyzine 7.8 Lamotrigine 4.8 Hydroxyalprazolam 7.8 Tramadol 5.8 4hydroxymidazolam 7.87 hydroxycarbamazepine 5.4 Chlordiazepoxide 7.89 7aminonitrazepam 5.33 hydroxymidazolam 8.7 7aminoclonazepam 5.36 Nordiazepam 8.7 Ketamine 5.67 Clobazam 8.9 Niaprazine 5.77 Flunitrazepam 8. Norbuprenorphine 6.8 Lormetazepam 8. 3Hydroxybromazepam 6.5 Estazolam 8.3 Doxylamine 6.36 Temazepam 8.37 LSD 6.45 Triazolam 8.38 Diphenhydramine 6.78 Ethyl loflazepate 8.44 Carbamazepine 6.9 Alprazolam 8.58 Zopiclone 6.99 9. Desmethylflunitrazepam 7.9 9.5 Ndesmethylclobazam 7.9 OHTHC 9.3 Lorazepam 7. THC 9.37 3hydroxyflunitrazepam 7.36 Clotiazepam 9.49 Oxazepam 7.37 THCCOOH 9.5 Flurazepam 7.37 Tetrazepam 9.6 Clonazepam 7.4 Cannabinol 9.94 Nitrazepam 7.4 Loprazolam.9 6

Conclusions A spectral based library of more than compounds has been created using certified reference materials acquired at three collision energies on a triple quadrupole mass spectrometry platform. A MRM triggered product ion spectra method to quantify and identify a panel of compounds commonly found in clinical and forensic toxicology was successfully applied to whole blood samples spiked with a panel of compounds. All compounds were detected at highest concentration and positively identified using product ion scan MS/MS library based searching generating higher data quality for compound identification. First Edition: October, 6