Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Similar documents
Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Ed George and Anaïs Viven Varian, Inc.

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Solid Phase Micro Extraction of Flavor Compounds in Beer

Methanol Extraction of high level soil samples by USEPA Method 8260C

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Optimizing. Abstract: is standardd. procedures. altered to

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Maximizing Sample Throughput In Purge And Trap Analysis

Maximizing Production While Minimizing Costs

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

2017 EPA Method Update Rule and EPA Method 624.1

Solid Phase Micro Extraction of Flavor Compounds in Beer

Optimizing Standard Preparation

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Analytical Trap Comparison for USEPA Method 8260C

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

EPA TO-17 Volatile Organic Compounds

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Determination of Volatile Organic Compounds in Air

Target Compound Results Summary

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

STANDARD OPERATING PROCEDURES

CALA Directory of Laboratories

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

Associates. Project No.: October 25, 2017

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Thermal Desorption Technical Support

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Volatile Organic Compounds in Water PBM

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Copies: Dave Favero, RACER File

In-Tube Extraction Sample Preparation for Gas Chromatography

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

Study of Residual Solvents in Various Matrices by Static Headspace

Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc

Date Issued: April 01, 2013 Expiration Date: June 30, 2013

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

Volatile Organic Compounds in Every Day Food

/J~ GOLDER. Golder Associates Inc. golder.com. Project No April 25, 2018

317 Elm Street Milford, NH (603) Fax (603)

Table 1 Soil and Groundwater Sampling Summary Table

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

January 19, Dear Mr. Nightingale:

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

GOLDER. Golder Associates Inc. Al-1!1f;s. golder.com. June 22, Project No

Laboratory Report Number(s) Lone Tree Road (Deep) 10/5/15, 10/19/15, 6/13/16. Address Sample Date(s) Analyses Requested

Method 1624 Revision C Volatile Organic Compounds by Isotope Dilution GCMS

Transcription:

Application Note US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Thermo Scientific TRACE 1310 GC and ISQ LT MS System Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract US EPA Method 524.4 is used primarily by environmental labs for the analysis of volatile organic compounds (VOCs) in drinking water. While this method is effective at concentrating the trace levels of VOCs sometimes found in finished drinking water, it also tends to transfer significant quantities of water vapor to the gas chromatography-mass spectrometry instruments. Teledyne Tekmar s Lumin purge and trap (P&T) concentrator incorporates a moisture control system (MCS) specifically designed to reduce the amount of water transferred during desorb, in comparison to current purge and trap technologies. This application note will demonstrate the ability of the Lumin to reduce the amount of water transferred to the Thermo Scientific GC-MS system when using this method. Introduction Gas chromatography (GC) inlets, columns and mass spectrometers (MS) are sensitive to water. However, to detect toxic VOC compounds, the VOCs must be purged and trapped from water. This places water vapor on the GC-MS system, which contributes to column and inlet issues and frequent MS cleaning to remove the oxidation from the MS source. US EPA Method 524.4 places a large amount of water vapor on the GC-MS system. This method allows flexibility in the dry purge and desorb time of the analytical trap to reduce the large volumes of water that inundate the GC-MS system and lead to poor chromatography for early eluting gases. It also allows the use of nitrogen for the purge gas. Traditionally, purge and trap instruments reduced the amount of water transferred to GC-MS instrument through numerous water management techniques. The Teledyne Tekmar Lumin significantly reduces the amount of water transferred to the GC-MS system over current P&T instrumentation, through the design of its MCS. Additionally, improvements in analytical trap cooling have further reduced sample cycle times, permitting more samples to be processed within a 12-hour period. Sample Preparation Calibration standards were prepared from Restek 524.3 VOA MegaMix and 524.3 Gas Calibration Mix standards in deionized water containing the required amount of maleic acid and ascorbic acid. A calibration curve was prepared from 0.2 ppb to 50 ppb for all of the compounds. The relative response factor (RF) was calculated for each VOC using either 1,4-difluorobenzene, chlorobenzene-d5 or 1,4-dichlorobenzene-d4 as an internal standard. Surrogate standards consisted of: methyl-t-butyl ether-d3, 4-bromofluorobenzene and 1,2-dichlorobenzene-d4. A linear calibration curve with 1/X weighting was used for all compounds. Seven 0.5 ppb standards were prepared to calculate the method detection limit (MDL), accuracy and precision data. All calibration and MDL samples were analyzed with the Lumin and AQUATek 100 conditions in Table I and the GC-MS conditions in Table II. The correlation coefficient of the calibration curve (r 2 ), MDL, accuracy and precision data are shown in Table III. A 0.2 ppb standard was analyzed to indicate the lowest detectable peak for the initial seven gases unencumbered by excessive water in Figure 1. Figure 2 displays a 5 ppb standard, indicating excellent peak resolution for all of the VOCs. Sales/Support: 800-874-2004 Main: 513-229-7000

P a g e 2 Experimental Instrument Conditions Table I Lumin and AQUATek 100 Conditions Standby Variable Bake Variable Valve Oven Temp 150 C Bake Time 2.00 min Transfer Line Temp 160 C Bake Temp 260 C Sample Mount Temp 90 C MCS Bake Temp 180 C Purge Ready Temp 35 C Bake Flow 200 ml/min MCS Purge Temp 20 C AQUATek 100 Variable Standby Flow 20 ml/min Sample Loop Time 0.35 min Purge Variable Sample Transfer Time 0.35 min Purge Time 11.00 min Rinse Loop Time 0.30 min Purge Flow 40 ml/min Sweep Needle Time 0.30 min Dry Purge Temp 20 C Presweep Time 0.25 min Dry Purge Time 0 min Water Temp 90 C Dry Purge Flow 0 ml/min Bake Rinse Drain Cycles 3 Desorb Variable Bake Rinse Drain Time 0.35 min Desorb Preheat Temp 245 C Desorb Time 1.20 min Trap 9 Desorb Temp 250 C Chiller Tray On Drain Flow 300 ml/min Purge Gas Nitrogen Table II Thermo Scientific TRACE 1310 GC and ISQ LT MS System Conditions Thermo Scientific TRACE 1310 GC with Split/Splitless Instant Connect Injector Conditions Column Oven Profile Inlet Rtx -VMS, 20 m x 0.18 mm ID, 1 µm Film; Helium 0.8 ml/min 35 C, 2 min, 10 C/min to 100 C, 25 C/min to 225 C, 2 min hold; Run Time 15.5 min 200 C, 65:1 Split, P&T Adapter, Gas Saver Flow - 20 ml/min after 1 min Thermo Scientific ISQ LT MS Conditions Temp Transfer Line 230 C; Ion Source 300 C Scan Range 35 amu to 260 amu, Time 0.2 sec; Delay 0.8 min Current Emission Current 25 µa, Gain 3 e 5

Results P a g e 3 Table III Method 524.4 Calibration, Accuracy and Precision Data Compound Calibration Linearity (r 2 ) MDL MRL Confirmation (0.5 ppb) LPIR (> 50%) UPIR (<150%) Avg. Conc. Accuracy and Precision (n=7, 0.5 ppb) Accuracy (± 20%) Precision ( 20%) Dichlorodifluoromethane 0.993 0.11 60.1 114.3 0.44 87 7.8 Chlorodifluoromethane 0.997 0.12 66.6 127.9 0.49 97 8.0 Chloromethane 0.998 0.13 64.7 132.0 0.49 98 8.6 Vinyl Chloride 0.997 0.12 66.6 129.3 0.49 98 8.1 1,3-Butadiene 0.999 0.08 82.1 123.3 0.51 103 5.1 Bromomethane 1 0.998 0.06 79.7 109.4 0.47 95 3.9 Trichlorofluoromethane 0.998 0.08 78.9 120.1 0.50 99 5.2 Diethyl Ether 1.000 0.09 79.6 125.5 0.51 103 5.7 1,1-Dichloroethene 0.998 0.09 73.7 118.7 0.48 96 5.9 Carbon Disulfide 0.997 0.08 77.1 118.8 0.49 98 5.4 Iodomethane 0.999 0.07 80.6 117.6 0.50 99 4.7 Allyl Chloride 0.999 0.10 73.6 122.5 0.49 98 6.3 Methylene Chloride 0.997 0.03 81.4 97.9 0.45 90 2.3 trans-1,2-dichloroethene 0.998 0.08 71.1 109.0 0.45 90 5.3 Methyl Acetate 1.000 0.09 83.4 129.9 0.53 107 5.5 Methyl-t-Butyl Ether-d3 (Surr) 2.7 2 12.73 102 1.9 3 Methyl-t-Butyl Ether (MtBE) 1.000 0.04 94.1 116.0 0.53 105 2.6 t-butyl Alcohol (TBA) 0.999 0.18 51.2 140.7 0.48 96 11.8 Diisopropyl Ether (DIPE) 0.999 0.04 90.1 110.3 0.50 100 2.6 1,1-Dichloroethane 0.999 0.06 78.9 110.8 0.47 95 4.2 t-butyl Ethyl Ether (EtBE) 0.999 0.04 89.9 108.9 0.50 99 2.4 cis-1,2-dichloroethene 0.999 0.09 77.2 120.6 0.49 99 5.6 Bromochloromethane 4 1.000 0.05 90.3 114.9 0.51 103 3.0 Chloroform 0.998 0.09 69.4 112.9 0.46 91 6.1 Carbon Tetrachloride 0.999 0.13 64.7 128.7 0.48 97 8.4 Tetrahydrofuran 0.999 0.12 62.7 122.9 0.46 93 8.2 1,1,1-Trichloroethane 0.999 0.05 78.0 105.0 0.46 92 3.8 1,1-Dichloropropene 0.999 0.08 80.9 119.9 0.50 100 4.9 1-Chlorobutane 1.000 0.07 79.1 112.2 0.48 96 4.4 Benzene 1.000 0.06 81.0 111.2 0.48 96 4.0 t-amyl Methyl Ether (TAME) 0.999 0.06 85.8 116.1 0.50 101 3.8 1,2-Dichloroethane 0.999 0.07 82.2 118.2 0.50 100 4.5 Trichloroethene 1.000 0.06 73.3 105.7 0.45 89 4.6 1,4-Difluorobenzene (IS 1) 8.1 2 4.9 3 t-amyl Ethyl Ether (TAEE) 0.999 0.05 81.5 108.9 0.48 95 3.6

P a g e 4 Table III Method 524.4 Calibration, Accuracy and Precision Data Compound Calibration Linearity (r 2 ) MDL MRL Confirmation (0.5 ppb) LPIR (> 50%) UPIR (<150%) Avg. Conc. Accuracy and Precision (n=7, 0.5 ppb) Accuracy (± 20%) Precision ( 20%) Dibromomethane 1.000 0.09 78.3 125.4 0.51 102 5.8 1,2-Dichloropropane 1.000 0.05 86.7 109.7 0.49 98 3.0 Bromodichloromethane 1.000 0.05 82.3 108.0 0.48 95 3.4 cis-1,3-dichloropropene 0.999 0.03 97.2 112.9 0.53 105 1.9 Toluene 0.999 0.06 82.8 112.0 0.49 97 3.7 Tetrachloroethene 0.999 0.07 68.9 103.1 0.43 86 5.1 trans-1,3-dichloropropene 0.998 0.04 92.2 112.2 0.51 102 2.5 1,1,2-Trichloroethane 1.000 0.07 77.9 112.9 0.48 95 4.6 Ethyl Methacrylate 0.997 0.09 77.9 122.7 0.50 100 5.7 Dibromochloromethane 0.999 0.10 63.2 115.0 0.45 89 7.3 1,3-Dichloropropane 1.000 0.04 93.9 112.2 0.52 103 2.3 1,2-Dibromoethane (EDB) 1.000 0.06 84.0 115.1 0.50 100 3.9 Chlorobenzene-d5 (IS 2) 5.6 2 5.8 3 Chlorobenzene 1.000 0.09 74.9 118.9 0.48 97 5.8 Ethylbenzene 0.999 0.05 80.6 107.3 0.47 94 3.6 1,1,1,2-Tetrachloroethane 0.999 0.12 56.0 117.0 0.43 87 8.9 m-, p-xylene 1.000 0.12 78.1 107.1 0.93 185 4.0 o-xylene 0.999 0.09 71.5 116.7 0.47 94 6.1 Styrene 0.998 0.04 87.4 108.9 0.49 98 2.8 Bromoform 0.998 0.06 80.6 111.4 0.48 96 4.1 Isopropylbenzene 0.999 0.09 70.5 114.7 0.46 93 6.0 4-Bromofluorobenzene (Surr) 2.2 2 12.31 98 2.2 3 Bromobenzene 0.999 0.08 76.3 114.9 0.48 96 5.1 Propylbenzene 0.999 0.05 78.8 105.8 0.46 92 3.7 1,1,2,2-Tetrachloroethane 0.998 0.08 75.5 114.1 0.47 95 5.2 2-Chlorotoluene 0.999 0.10 63.9 115.3 0.45 90 7.2 1,3,5-Trimethylbenzene 1.000 0.12 70.4 129.8 0.50 100 7.5 1,2,3-Trichloropropane 0.999 0.05 77.6 102.8 0.45 90 3.5 4-Chlorotoluene 0.999 0.04 84.5 103.4 0.47 94 2.5 tert-butylbenzene 0.999 0.19 54.3 147.5 0.50 101 11.6 Pentachloroethane 1 0.998 0.09 71.9 115.3 0.47 94 5.8 1,2,4-Trimethylbenzene 0.999 0.05 76.5 101.7 0.45 89 3.5 sec-butylbenzene 0.999 0.04 79.4 100.1 0.45 90 3.0 4-Isopropyltoluene 0.999 0.07 70.9 107.0 0.44 89 5.2 1,3-Dichlorobenzene 1.000 0.06 76.6 104.3 0.45 90 3.9 1,4-Dichlorobenzene-d4 (IS3) 5.7 2 4.6 3

P a g e 5 Table III Method 524.4 Calibration, Accuracy and Precision Data Compound Calibration Linearity (r 2 ) MDL MRL Confirmation (0.5 ppb) LPIR (> 50%) UPIR (<150%) Avg. Conc. Accuracy and Precision (n=7, 0.5 ppb) Accuracy (± 20%) Precision ( 20%) 1,4-Dichlorobenzene 1.000 0.06 78.3 105.7 0.46 92 3.8 Butylbenzene 0.997 0.09 71.0 115.5 0.47 93 6.0 Hexachloroethane 1.000 0.16 51.8 133.8 0.46 93 11.1 1,2-Dichlorobenzene 1.000 0.08 74.1 112.4 0.47 93 5.1 1,2-Dichlorobenzene-d4 (Surr) 3.2 2 12.75 102 2.3 3 1,2-Dibromo-3-chloropropane 0.998 0.05 85.4 112.4 0.49 99 3.5 Hexachlorobutadiene 0.998 0.07 83.6 120.1 0.51 102 4.5 1,2,4-Trichlorobenzene 0.998 0.10 70.0 119.3 0.47 95 6.6 Naphthalene 0.997 0.05 77.5 104.4 0.45 91 3.8 1,2,3-Trichlorobenzene 0.998 0.07 78.1 114.1 0.48 96 4.8 1. Quadratic curve with 1/X weighting. 2. %RSD of the peak areas for the internal standard or the calculated amount for the surrogate standards from the calibration curve samples. 3. %RSD of the peak areas for the internal standard or the calculated amount for the surrogate standards from the seven MDL samples. 4. Mass ion 49 used. Figure 1 Primary Characteristic Ions for the First Seven Gases of a 0.2 ppb Standard Indicating Excellent Detection Limits with Minimal Interference from Water.

P a g e 6 Figure 2 Total Ion Chromatogram of a 5 ppb VOC Standard Indicating Consistent Peak Shapes for all Compounds with No Water Interference. Conclusion The Teledyne Tekmar Lumin purge and trap concentrator and AQUATek 100 was used to process water samples containing VOCs following US EPA Method 524.4 with detection by a Thermo Scientific TRACE 1310 GC and ISQ LT MS system. The linearity of the calibration curve from 0.2 ppb to 50 ppb passed all method requirements with no interference from excessive water. The MDL, precision and accuracy for seven 0.5 ppb standards also indicated no interference from excessive water. References 1. Method 524.4 - Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry Using Nitrogen Purge Gas; US EPA, Office of Water, May 2013.