HPLC Method Development: Standard Practices and New Columns

Similar documents
HPLC Method Development with Eclipse Plus: Standard Practices and New Columns. Agilent Technologies

HPLC Method Development: Standard Practices and New Columns. Agilent Technologies February 2007

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

A Look at Column Choices. Ed Kim Application Engineer Agilent Technologies, Inc February 12, 2009

LC and LC/MS Column Selection Flow Chart

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

Choosing Columns and Conditions for the Best Peak Shape

William E. Barber, Ph.D. Applications Chemist October

Does one-size-fit-all?

High Resolution Fast LC

Too Polar for Reversed Phase What Do You Do?

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

New To HPLC Avoiding Beginner Pitfalls. Rita Steed LC Columns Application Engineer April 24, 2013

Fast Separation of Vastly Different Compounds by Isocratic HPLC

Efficiently analyze your most challenging compounds. Deliver high-quality results faster. Agilent ZORBAX Column Selection Guide for HPLC

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Shodex TM ODP2 HP series columns

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think!

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

penta-hilic UHPLC COLUMNS

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

The new Agilent 1290 Infinity Quaternary LC

Advanced Topic: Tips and Tricks of HPLC Separations and Troubleshooting. Rita Steed LC Advanced Topics Seminar Series August 25, 2010.

Tips and Tricks of HPLC System Troubleshooting

Validated HPLC Methods

HPLC Separation Fundamentals. Ed Kim Application Engineer Agilent Technologies, Inc January 14, 2009

How proteins separate on reverse-phase HPLC

A Look at Agilent HPLC Columns Choices

Reversed Phase Solvents

penta-hilic UHPLC COLUMNS

HICHROM. Chromatography Columns and Supplies. LC COLUMNS Zorbax. Catalogue 9. Hichrom Limited

State-of-the-art C18 HPLC Columns

Fast Screening Methods for Steroids by HPLC with Agilent Poroshell 120 Columns

(HILIC) Bill Champion Agilent Technologies, Inc opt 3/opt3/opt 2 HILIC - Agilent Restricted

Hamilton. Reversed-Phase Anion Exchange Cation Exchange Ion Exclusion

ACE. For increased polar retention and alternative selectivity. Alternative selectivity for method development

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

HPLC Column Troubleshooting

New Product Update. HPLC Columns

Method Development Kits

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Getting More Resolution from Your LC Michael Woodman Columns and Consumables September 10, 2008

HICHROM. Chromatography Columns and Supplies. LC COLUMNS HALO and HALO-5. Catalogue 9. Hichrom Limited

Phenyl-Hexyl. UHPLC Columns. Alternate, complementary selectivity to C18 and C8 bonded phases

An HPLC column offering moderate retentivity with superb peak shape Inertsil ODS Sprint

Packings for HPLC. Packings for HPLC

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Part 2. Overview of HPLC Media

8. Methods in Developing Mobile Phase Condition for C18 Column

InertSustainSwift C8

LC Column Troubleshooting. Do you have an equivalent column for my LC column?

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think!

Automated LC Method Development: Fundamentals and Columns, Hardware, and Software

Inertsil ODS-EP Technical Information

ACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes

Synthesis and Use of a New Covalently Bonded C18 Modified Carbon Clad Microporous Zirconia for Fast High Temperature Separations

InertSustainSwift C8

Inertsil Technical Library

Reversed Phase Chromatography: Mobile phase Considerations

ACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

ACE Excel SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

Separation of Structurally Similar Steroids on HALO C18 and PFP

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

HPLC Separation Robustness and Ruggedness Assessing the Effects of Experimental Variables During Method Development

Fast Screening Methods for Analgesics and Non-Steroidal Anti-Inflammatory Drugs by HPLC

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

zorbax Packed by Phenomenex Material Characteristics High efficiency Guaranteed quality Phenomenex columns packed with Zorbax

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Your Reversed Phase Method

Eliminating Unwanted Variability in LC Methods: More Uptime, More Sample Throughput

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Practical Applications of Method Translation Using the Agilent Method Translation Tool Rita Steed Inside Application Engineer Agilent Technologies

Dilution(*) Chromatography

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns

P R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns

HPLC Column Troubleshooting: Is It Really The Column? LC Columns and Consumables. Robyn C. Woolley Online Application Engineer July 24, 2007.

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Zirconia-Based Phases as a Powerful Complement to Silica-Based Phases for LC and LC-MS under Non-extreme Mobile Phase Conditions

ACD/AutoChrom Assisted Method Development for Challenging Separations. Vera Leshchinskaya February 7, 2018

LC Technical Information

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

for Acclaim Mixed-Mode HILIC-1 Column

Acclaim Mixed-Mode WCX-1

State - of - the -art C18 HPLC Columns

Uptisphere CS Evolution Core Shell columns for fast & highly efficient identification & quantification of small molecules

Do you need trouble-free speed? The NEW Chromolith HR

HICHROM. Chromatography Columns and Supplies. LC COLUMNS SIELC Primesep. Catalogue 9. Hichrom Limited

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

ProntoSIL C18-EPS Reversed-Phase HPLC Columns

UPLC Method Development and Validation

Do you need trouble-free speed? The NEW Chromolith HR

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Transcription:

HPLC Method Development: Standard Practices and New Columns Edward Kim Application Engineer Agilent Technologies, Inc. March 8, 29

What are Some Standard Method Development Practices?. Follow preferred method development scheme and do hands-on method development based on selectivity-changing parameters e.g., ph, column, or mobile phase types 2. Use method development software run a few predictive runs and get prediction for best method 3. Evaluate multiple columns/multiple mobile phases in a manual or automated fashion, determine best results (e.g., using valve switching to change between columns, software to track peaks etc.) Practice # is used by many and incorporates hands on learning. It requires nothing special to make it work, so we will work through a method development scheme for this practice.

What Makes a Good Starting Point for Hands-on Method Development?. Choose a C8 column many choices, well known technologies a) A short column minimizes method development time b) Smaller particle sizes provide the resolution needed d in a short time c) Newer C8 column choices can improve results based on efficiency, and improved peak shape 2. Choose a simple mobile phase reliable, works with many samples a) Phosphate buffer ph 3, TFA, or formic acid in aqueous portion b) Acetonitrile or methanol as organic modifier 3. Adjust mobile phase to get the desired retention and resolution a) Adequate resolution of all peaks, R s 2. requires highest efficiency b) Retention of first peak preferred to be at least k = c) Analysis time below 2 minutes shorter d) Newer column choices, short columns with small particle sizes can provide more efficiency and resolution in a very short time, thereby speeding up method development

Typical Method Development Parameters - Effects of Selectivity, Efficiency, and Retention on Resolution 7. Selectivity Impacts Resolution Most 6 6. 5. Change bonded phase Change mobile phase Plates are easiest to increase Typical Method Development Parameters Resolution 4. 3. Increase N Increase Alpha Increase k' 2... R =N ½ s /4 (α-)/α k /(k +) k/(k+) 2 3 4 5 6 7 8 9 2 3 4 5 6 7 8 9 2 2 Plates: 5 5 2 25 Alpha:..35.6.85 2. k : 2. 4.5 7. 9.5 2.

Small Differences in α Selectivity - Can Provide Large Differences in Resolution Eclipse Plus C8 Greatest Resolution for this sample α Resolution peaks,2.68 6.2 peaks 2,3.4 4.97 caffeine e sulfametho oxazole barbital Remember to assess T f when comparing results! 2 3 4 Eclipse Plus C8 min α Resolution peaks,2.33 3.55 peaks 2,3.3 3.82 Cols: 4.6 x 5, 5 um m.p.: 4% MeOH, 6% water 2 3 4 Eclipse XDB-C8 min α Resolution peaks,2.43 4.5 peaks 2,3.29 3.55 2 3 4 min

What Do We Want to Base Our Complete Method Development Scheme On? (Beyond the Start) Changing selectivity to improve resolution with the two key parameters mobile phase and bonded phase Mobile phase ( st choice to change because it s easy) Mobile phase organic modifier (ACN, MeOH, etc.) Mobile phase ph over a wide ph range ph -2 if needed Bonded phase (provides a lot of optimization potential) Many choices other than C8 C8, Phenyl, y, CN (most orthogonal!) AQ for highly aqueous mobile phases Choose new columns with best peak shape performance

New Column Choices To Use for Standard Method Development. Shorter columns with smaller particle sizes Rapid Resolution High Throughput Columns (RRHT) with a.8 μm particle size 5 mm or shorter for fastest method development Longer columns for maximizing efficiency Many bonded phase choices for a complete method development scheme 2. Improved bonded phases for superior performance Better peak shape and efficiency result in more resolution Critical to the best methods New Eclipse Plus columns can deliver improved performance

. Start with Short RRHT C8 Columns with Highest Efficiency for Fastest Method Development Column: 4.6 x 5mm,.8um Mobile Phase: A=.% Formic Acid, B=ACN +.%Formic Acid (95:5) Flow rate:.5 ml/min Inj. Vol: 2 ul Sample: Xanthines:. -methylxanthine, 2.,3-dimethyluric acid, 3. 3,7-dimethylxanthine, 4.,7-dimethylxanthine 2 3 4 RRHT SB-C8 4.6 x 5 mm,.8 um.5.5 2 2.5,2 3 4 RRHT SB-CN 4.6 x 5mm,.8 um min.5.5 2 2.5 The C8 gives the most effective resolution and with a 4.6 x 5 mm,.8um column the initial results are achieved in just a couple of minutes, with the highest efficiency. The 5 mm length is short and fast methods can be approximated or completed in just minutes!! min

. Start Method Development with RRHT Columns: Different ZORBAX RRHT C8 Bonded Phases for Max Selectivity st choice Best Resolution & Peak Shape 2 nd choice Good alternate selectivity due to non-endcapped d 2 2 3 4 5 3 2 3 4 4 Eclipse Plus C8 StableBond SB-C8 Mobile phase: (69:3) ACN:water Flow.5 ml/min. Temp: 3 C Detector: Single Quad ESI positive mode scan Columns: RRHT 4.6 x 5 mm.8 um 2 3 4 5 3 rd choice 2 4 Eclipse Good efficiency & peak shape 3 Resolution could be achieved min Eclipse XDB-C8 Sample:. anandamide (AEA) 2. Palmitoylethanolamide l th l id (PEA) 3. 2-arachinoylglycerol (2-AG) 4. Oleoylethanolamide (OEA) 2 3 4 5 4 th choice M lti l b d d h Resolution not likely, Other choices better, for this separation. 2,3 4 Extend-C8 2 3 4 5 Multiple bonded phases for most effective method development. Match to one you are currently using.

2. Agilent Rapid Resolution HT Eclipse Plus C8 Provides Superior Peak Shape and Efficiency mau 8 4 2 ZORBAX RRHT Eclipse Plus C8 2. x 3 mm,.8 µm N: 45 6 23 bar TF:.4 Rs 4,3 :.62 36 % MeCN 64% Water w/.7 ml/l phosphoric acid 2 µl injection 22 nm Flow :.5 ml/min RRHT,.45 ml/min. (Luna) 2- Hydroxy-2-(4-Isobutylphenyl) proprionic acid 2 2-(3-Isobutylphenyl) proprionic acid 3 Ibuprofen 4 2-(4-n-butylphenyl) proprionic acid 5 Isobutyl acetophenone All at.25 mg/ml mau 2 3 4 5 6 7 8 8 Phenomenex Luna HST C8 (2) 2. x 3 mm, 2.5 µm min 6 6 bar N: 38 4 2 TF:.2 Rs 4,3 :.47 2 3 4 5 6 7 8 ZORBAX RRHT has at least 8% more efficiency than Luna HST C8(2) That means greater than baseline resolution when needed. min

2. Choose New Eclipse Plus C8 for Superior Efficiency and Peak Shape Analysis on Other C8 Poor sensitivity Due to broad peaks Peaks do not tail, but are distorted. Analysis on Eclipse Plus Better Sensitivity Sharp, Efficient Peaks Excellent peak shape

How Do We Implement These New Column Choices in the Method Development Process? Follow a method development process around standard practices ONLY when the first steps allow you to incorporate new products for fast, good results! A method development scheme based on simple strategies for optimizing selectivity and resolution can incorporate new products. The 3 key parameters must be: mobile phase mobile phase ph bonded phase

First Steps in a Method Development Scheme Choose Eclipse Plus C8 Low ph Adjust %ACN for.5 < k < 2 STEP Band spacing problems STEP 2 Change % organic Band spacing problems STEP 3 Change organic modifier Adjust % organic for.5 < k < 2 Band spacing problems STEP 4 Change bonded phase Eclipse Plus C8, SB-C8, -CN, -Phenyl, y, other RP Select a high h quality C8 or C8 bonded phase first for good peak shape, retention and resolution with typical acidic, basic and neutral samples. Optimize the organic component of the mobile phase to change selectivity it Choose alternate bonded phases to completely optimize method, if needed Choose SB-C8 for ph -2 With new RRHT column choices, these steps can be done quickly, making it possible to truly find the best method.

Let s Look at Method Development of These Compounds propranolol pindolol dipyridamole β-blocker Anti-arrhythmic Vasodilator Ca +2 channel blocker disopyramide diltiazem

Start at Low ph, Adjust % Organic Eclipse Plus C8 Cardiac Drugs with Acetonitrile Column: ZORBAX RRHT Eclipse Plus C8, 4.6 x 5 mm,.8 μm Mobile Phase: A: 25 mm NaH 2 PO 4, ph 3. B: ACN Flow Rate: 2. ml/min Temperature: 3 C Detection: UV 24 nm 3 Sample: Cardiac Drugs.Pindolol 2. Di-isopyridamide isop 3.Propranolol 4.Dipyridamole 5. Diltiazem mau 4 2 2 4% ACN RRHT Eclipse Plus C8 allows for very fast optimization of % organic in mobile phase. mau 25 75 5 25 2 3% ACN Good resolution Fast analysis No time wasted mau 6 2% ACN 4 2-2 k= 44!! 2 4 6 8 2 min

Steroidal Compounds CH 3 OH CH 3 OH CH 3 OH CH H OH H H H H H H H H HO HO HO estriol estradiol ethinyl estradiol CH 3 OH CH 3 HO CH3 O CH 3 H CH HO CH 3 C H 3 OH HO H H dienestrol diethylstilbestrol ethinyl estradiol 3-methyl ether

Mobile Phase Optimization on Short RRHT 4.6 x 5 mm, SB-CN Column mau 25 2 5 5 ) mau 25 2 5 5 4 2 3 5 6 2 3 2 3 4 5 6 %B 55 %B A: Water R s 4,5 = 2.59 6. Estriol 2. Estradiol 3. Ethinyl Estradiol 4. Dienestrol 5. Diethylstilbestrol 6. Ethinyl estradiol methyl ether B: 3 % MeOH/7 %MeCN µl injection, 2 nm 2 3 4 5 mau 25 5 %B 2 2 3 4 5 5 R s 4,5 = 3.29 6 5 2 3 4 5 6 7 8 min RRHT columns allow for very fast optimization of % organic in mobile phase.

Comparison of Optimal Conditions On Columns with Different Dimensions Length, Particle Size, Resolution mau 4 2 4 2 3 5 3 5 8 6 4 2-2 mau 2 5 5 ) RRHT SB-CN 4.6 x 5 mm,.8um.5 ml/min µl injection 6 2.5 5 A: Water 2 3 4 5 RRHT SB-CN 4.6 x mm,.8um.5 ml/min 2 µl injection 6. Estriol 2. Estradiol 3. Ethinyl Estradiol 4. Dienestrol 5. Diethylstilbestrol 6. Ethinyl estradiol methyl ether B: 3 % MeOH/ 7 %MeCN 4 % B Detection: UV 2 nm mau 75 5 25 75 5 25 ) 2.5 5 7.5 4 2 3 5 SB-CN 4.6 x5 mm, 5um. ml/min 3 µl injection same N as 6 4.6 x 5 mm 2.5 5 7.5 RRHT! 2.5 5 7.5 RRHT Columns allow rapid resolution optimization during initial method development. Column lengths compared quickly. min

Why Do We Recommend Changing Organic Modifier Before Changing the Bonded Phase? It s easy acetonitrile and methanol are readily available! It works on any bonded phase, so it gives you a tool to possibly optimize a separation no matter the column choice!

Start at Low ph on Eclipse Plus C8 RRHT Cardiac Drugs with Methanol, Changing Organic Modifier mau 4 3 2 mau 25 75 5 25 mau Remember to use buffer! Column: ZORBAX RRHT Eclipse Plus C8, 4.6 x 5 mm,.8 μm Mobile Phase: A: 25 mm NaH 2 PO 4, ph 3. B: MeOH Flow Rate: 2. ml/min Temperature: 3 C Detection: UV 24 nm Sample: Cardiac Drugs.Pindolol 2. Di-isopyridamide isopyridamide 3.Propranolol 4.Diltiazem 5. Dipyridamole 7% MeOH 2 2 5% MeOH Adjusting [MeOH] influences retention like adjusting ACN Solvents can be compared quickly with RRHT columns MeOH changes selectivity and gives longer analysis time for these analytes 8 6 4 2 4% MeOH k= 49!! 2 4 6 8 2 min

Method Development - Change Organic Modifier Comparison of Acetonitrile and MeOH mau 4 2 8 6 4 2-2 mau 4 2 8 2 2 Column: ZORBAX RRHT Eclipse Plus C8, 4.6 x 5 mm,.8 μm Mobile Phase: A: 25 mm NaH 2 PO 4, ph 3. B: organic Flow Rate: 2. ml/min Temperature: 3 C Detection: UV 24 nm Sample: Cardiac Drugs.Pindolol 2. Diisopyridamide 3.Propranolol 4.Diltiazem 5. Dipyridamole 5 3 3 4 2 3 4 6 3% ACN 4 2-2 2 5 Resolution of pairs (2,3), (3,4), (4,5) 5% MeOH is better in MeOH. Peak 5 selectivity shift Resolution of critical pair (,2), is better in ACN

Change Mobile Phase - Acetonitrile vs. Methanol on ZORBAX Eclipse XDB-CN with Estrogens Column: ZORBAX Eclipse XDB-CN Column Dimensions: 4.6 x 5 mm, 5 μm Mobile Phase: As shown Flow Rate: 2. ml/min Injection Volume: 2. μl Column Temperature: 25 C Detector: UV, 2 nm 2 3 4 5 Mobile Phase: 35:65 ACN:Water Solvent comparison with all samples can yield optimized methods. 6 7 2 4 6 8 2 4 6 8 4 Mobile Phase: 47.5:52.5 MeOH:Water min Sample:. Estriol (.3 μg/μl), 2. β-estradiol (.3 μg/μl), 3. Ethinyl Estradiol (.47 μg/μl), 4. Dienestrol (.23 μg/μl), 5. Diethylstilbestrol (.28 μg/μl) 6. Ethinylestradiol 3-methyl ether (.3 μg/μl) 7. Ethynodiol Diacetate t (.39 μg/μl) 2 3 5 6 7 2 4 6 8 2 4 6 8 min

Why Is Changing the Bonded Phase Effective? Differences in interactions between polar and non-polar compounds. Other types of interactions with a bonded phase can be exploited (pi-pi interactions etc.) These all change with bonded phase! Changing the bonded phase can improve selectivity/ resolution, reduce analysis time When you use RRHT (.8um) columns, the comparison of bonded phases can be done quickly!! You must have multiple RRHT column choices available

Bonded Phase Selectivity Differences in 3% ACN mau mau mau mau mau 2 2 2 2 RRHT SB-CN 4.6 x 5 mm,.8 μm RRHT SB-Phenyl 4.6 x 5 mm,.8 μm RRHT SB-Aq 4.6 x 5 mm,.8 μm RRHT Eclipse Plus C8 4.6 x 5 mm,.8 μm RRHT SB-C8 4.6 x 5 mm,.8 μm 5 Different bonded phases compared Analysis time of each run is only 2 minutes Comparison among phases done in optimum % organic for SB-CN The fast runs mean a comparison can be done even if you have a good separation on the C8 More chances to optimize!

Selectivity/Bonded Phase Choices: Eclipse Plus C8 vs. C8 and RRHT vs. RR RRHT Eclipse Plus C8, 4.6 x 5mm,.8 um. Oxybenzone 2. Internal Std. 3. Octylmethoxycinnamate 247 bar N: 88 Pw:.4 4 min. TF:.2 N: 6 Pw:.47 47 min. TF:.6 N: 8 Pw:.93 93 min. TF:.2 ) 2 3 4 Rapid Resolution Eclipse Plus C8, 4.6 x 5mm, 3.5 um N: 5 Pw:.9 min. TF:.9 N: 66 Pw:.37 min. TF:.99 N: 63 Pw:.6 min. TF:.98 2.2 min. 2 3 4 min 88 bar Conditions: Mobile Phase.: Water: Acetonitrile (3:7) Flow Rate: 2. ml/min. Detection: UV 23nm Temperature: 3 C Sample: Lip balm extract in ACN (melted at C ACN, cooled and.45 um filtered) Less retention can save significant time the C8 is a good choice here. The RRHT column is delivering the efficiency and resolution expected, but the C8 bonded phase may be the better choice. min

Selectivity of Polar Phases Provides Optimum Separation of Steroids Versus Non-Polar C8/C8 2 3 Column: ZORBAX Eclipse XDB-CN Mobile Phase: 35:65 ACN:Water 4 5 6 7 5 5 2 min 45 4,5 Column: ZORBAX Eclipse XDB-Phenyl 2 Mobile Phase: 48:52 ACN:Water 3 6 7 5 5 2 5 Column: ZORBAX Eclipse XDB-C8 2 3 4 Mobile Phase: 54:46 ACN:Water 6 7 5 5 2 4 3 5 Column: ZORBAX Eclipse XDB-C8 2 Mobile Phase: 6:4 ACN:Water 6 7 5 5 2 minutes. Column Dimensions: 4.6 x 5 mm, 5 μm Flow Rate: 2. ml/min Injection Volume: 2. μl Column Temperature: 25 C Detector: UV, 2 nm Sample:. Estriol (.3 μg/μl), 2. β-estradiol (.3 μg/μl), 3. Ethinyl Estradiol (.47 μg/μl), 4. Dienestrol (.23 μg/μl), 5. Diethylstilbestrol (.28 μg/μl) 6. Ethinyl estradiol 3-methyl ether (.3 μg/μl) 7. Ethynodiol Diacetate (.39 μg/μl) Eclipse XDB-CN is the optimum bonded phase the only one that can resolve all the components in under 2

So then Why is ph Critical to Method Development? What compounds are ph sensitive? How does ph affect retention and resolution? How does ph influence method development? How does ph influence column choice?

When Does ph affect Resolution? Compound Type Comparison ph 3 non-polar polar ionizable naphthalene benzyl alcohol benzoic acid acenaphthene nitrobenzene benzanilide 2 3 4 5 6 7 8 9 2 3 4 5 6 7 8 2 3 4 5 6 7 8 ph 7 2 3 4 5 6 7 8 9 2 3 4 5 6 7 8 2 3 4 5 6 7 8 Time (min) Time (min) Time (min) Ionizable compounds acids and bases can change retention and selectivity the most with changes in ph

Change in Retention with ph for Ionizable Compounds is Key to Method Development Non-charged analytes have better retention (i.e. acids at low ph and bases at high ph) Silanols on silica ionize at mid-ph, increasing retention of basic analytes (i.e possible ion-exchange interactions) Choose mobile phase ph to optimize retention and selectivity during method development Eclipse Plus can be used over a wide ph range Other choices exist for high ph

Change in Retention with ph for Ionizable Compounds is Compound Dependent Non-charged analytes have better retention (i.e. acids at low ph and bases at high ph) 2 retentio on time (min ) 8 6 4 2 Acetylsalicylic acid pka 3.5 Pyridine pka 5.2 Codeine pka 8. Procainamide pka 9.2 Amphetamine pka 9.9 pka 4 Caffeine Mobile Phase: 45% MeOH: 55% 2 mm Phosphate Buffer ph 2.5 ph 6.5 ph 8 ph.5 ph

Retention vs. ph for Basic Antihistamines - Less Variation in Retention at Low ph Re etention Volume ml 5 4 3 2 Column: Eclipse XDB-C8 Doxylamine pka 44 4.4, 92 9.2 4.6 x 5 mm, 5 μm Chlorpheniramine pka 9. Mobile Phase: 75% 25 mm phosphate Triprolidine pka 6.5 buffer Diphenhydramine pka 9. 25% Acetonitrile No change in retention in this range 3 4 5 6 7 ph Slight changes in ph may dramatically change retention and selectivity Always evaluate retention changes with ph during method development

Therefore ph Should be Considered a Key Method Development Parameter A single column can be used over a wide ph range making selectivity comparisons quick and reliable when ph is changed. May have better selectivity in the ph range 3 8. Other bonded phases can be reconsidered for mid ph, just like at low ph

Method Development Scheme Adding Mid ph From low ph ZORBAX Eclipse Plus C8 ph 7 (6-9) 2-5 mm buffer, Adjust %ACN for.5 < k < 2 STEP 5 Band spacing problems STEP 6 Change % organic Band spacing problems STEP 7 Change organic modifier (MeOH) Adjust % organic for.5 < k < 2 Restart at STEP 6 Band spacing problems STEP 8 Try Eclipse Plus Phenyl-Hexyl, Eclipse XDB- CN, XDB-Phenyl or Bonus-RP Restart at STEP 5 Mid ph can provide better selectivity It may be more compatible with your sample The process for investigating mid ph is the same as for low ph Eclipse Plus delivers outstanding performance at mid ph Alternate bonded phases should also be considered if improved selectivity is desired

Eclipse Plus Columns Are Ideal for Method Development at Low and Mid ph Can be used from ph 2-9 Accommodates method development in low and mid ph ranges Allows ph sensitive separations to be optimized Excellent peak shape and efficiency achieved with new Eclipse Plus material

Eclipse Plus Can Be Used Over a Wide ph Range (ph 2 9) to Optimize Selectivity ph 2.7.% formic acid/acn 4 6 3 2 5 7. procainamide 2. buspirone 3. pioglitazone 4. eletriptan 5. dipyridamoleid 6. diltiazem 7. furosemide 2 4 6 8 ph 4.8 NH 4 OAc/ACN 2 6 min Selectivity and resolution can change with ph 7 5 3 4 2 4 6 8 ph 7 NaH 2 PO 4 /ACN 7 3 2 6 min Eclipse Plus can be used with many mobile phases and phs 4 5 2 4 6 8 Conditions: Column: Eclipse Plus C8 4.6 x mm, 5um Gradient: 9% in minutes Detection: UV 254 nm

Selectivity Differences at ph 2 and ph 7 Can Be Dramatic. Acetaminophen 2. Caffeine 3. Acetylsalicylic acid 4. unknown Eclipse Plus C8 4.6 x 5mm, 5 um 3 ph 2.7 2 4.5.5 2 2.5 3 3.5 3 4 2 2 min. ph 7. min Column: 4.6 x 5, 5 um PN 959946-96 Gradient: -6% B/3 min. ph 2.7: A:.% formic acid B:.% fa in ACN ph 7.: A 2 mm Na phosphate B: ACN Sample: generic Excedrin tablet.5.5 2 2.5 3 3.5 min Selectivity can change dramatically from ph 2 7 with many samples, such as for the ingredients in this analgesic tablet. Eclipse Plus can be used under both conditions

Selectivity Differences at Mid ph Key to Resolution Eclipse Plus Performs well at Low and Mid ph 2 Eclipse Plus C8 4.6 x 75 mm 3.5 µm 2 mm phosphate buffer, ph 3. 3 N=57 TF:.28 2 3 4 5 6 7. Famotidine 2. Cimetidine 3. Pirenzipine Eclipse Plus C8 4.6 x 75 mm 3.5 µm 2 mm phosphate buffer, ph 7. 2 N=8 TF:.99 3 min Columns: as listed Mobile Phase: 2% MeOH, 8% 2 mm phosphate, ph as listed Flow Rate: ml/min. Detection: UV 23 semi micro flow cell N=79 TF:.5 2 4 6 8 2 4 6 For this sample resolution is not possible at ph 3, but is at ph 7 Eclipse Plus was used to experiment at both phs min

Try an Alternate Selectivity at Mid ph if Needed O One choice for alternate t selectivity it is Bonus-RP. The alkyl-amide bonded phase provides very different selectivity at mid-ph Bonus-RP Eclipse XDB-C8 3 2 4 3 4 2 Mobile Phase: 25% MeOH Flow Rate:. ml/min Temperature: Detection: UV 254 nm Injection Vol: 3 μl Sample: Cephalosporins. Cephalexin 2. Cephaclor 3. Cephuroxime 4. Cephoxitin 75% mm Na Ambient 2 4 6 8 Time (min) Mid ph was the second ph region to consider. High ph can also be considered if low or mid ph do not resolve all analytes.

Method Development at High ph a Third Choice From Mid ph ZORBAX Extend-C8 STEP 9 ph.5 (9-2) 5 mm ammonia, or TEA, or 5 mm organic or borate buffers T = 25 C (ambient 4 C) Adjust MeOH for.5 < k < 2 Reasons to Consider High ph Increase retention of basic compounds by analyzing them in non-charged form Improve selectivity it Band spacing problems STEP Change organic modifier (ACN or THF) Adjust for.5 < k <2 Try different HPLC mode

High ph Increases Retention of Antihistamines Extend-C8 Column: ZORBAX Extend-C8, 4.6 x 5 mm, 5 μ m Mobile Phase: See Below Flow Rate:. ml/min Temperature: RT Detection: UV 254 nm Sample:l:. Maleate 2. Scopolamine 3. Pseudoephedrine 4. Doxylamine 5. Chlorpheniramine 6. Triprolidine 7. Diphenhydramine ph 7 ph 3% 2 mm Na 2 HPO 4 3% 2 mm TEA (triethylamine) 7% MeOH 7% MeOH 2,3 4 7 5 3 t 5 R = 8.5 t R =.4 4 7 6 2 6 5 Time (min) 5 Time (min) The retention of the basic compounds in this sample increases at high ph.

Conclusions New column choices, like.8um RRHT columns and Eclipse Plus bonded phases can enhance typical approaches to method development. These columns provide fast analysis, allowing more options to be considered quickly, and with superior performance, the best results are readily achievable. Since many compounds are ionizable, and retention and selectivity of these compounds are greatly impacted by ph, a method development scheme for LC methods should encourage varying ph. Any method development scheme should also include the use of different bonded phases. Different bonded phases and columns can be selected by ph as well. The third key aspect of method development is mobile phase and organic modifier and this is an effective tool for many separations.

Wrap-up e-seminar Questions Thank you for attending today s Agilent e-seminar. Our Seminar schedule is expanding regularly. Please check our web site frequently at: www.agilent.com/chem/eseminars Or register for Stay current with enotes to receive regular updates.