CM3 PYRANOMETER INSTRUCTION MANUAL

Similar documents
SP-LITE SILICON PYRANOMETER INSTRUCTION MANUAL

INSTRUCTION MANUAL. SP Lite2 Pyranometer. August Copyright 2017 Campbell Scientific (Canada) Corp.

IRTS-P PRECISION INFRARED TEMPERATURE SENSOR INSTRUCTION MANUAL

CR10XTCR THERMOCOUPLE REFERENCE

MODEL 107 TEMPERATURE PROBE INSTRUCTION MANUAL

CR10TCR THERMOCOUPLE REFERENCE

3WHB10K 3-WIRE HALF BRIDGE TERMINAL INPUT MODULE INSTRUCTION MANUAL

CR10XTCR THERMOCOUPLE REFERENCE

WARRANTY AND ASSISTANCE

MODEL HMP45C TEMPERATURE AND RELATIVE HUMIDITY PROBE INSTRUCTION MANUAL

INSTRUCTION MANUAL. LI200X-LC Pyranometer for MetData1 4/97. Copyright (c) Campbell Scientific, Inc.

PTA-427 BAROMETRIC PRESSURE TRANSDUCER OPERATOR S MANUAL

MET ONE RAIN GAGE MODELS 380, 385, 380M, 385M

3WHB10K 3-Wire Half Bridge Terminal Input Module

INSTRUCTION MANUAL. CR10XTCR Thermocouple Reference. Revision: 1/03. Copyright (c) Campbell Scientific, Inc.

Eppley PIR Precision Infrared Radiometer

Amplified Pyranometer SP-212 & 215

Model 108 Temperature Probe Revision: 4/07

CVD20 20:1 Voltage Divider Revision: 5/12

INSTRUCTION MANUAL. CS500 Temperature and Relative Humidity Probe. Revision: 7/04. Copyright (c) Campbell Scientific, Inc.

INSTRUCTION MANUAL. CSI Model TB4 Rain Gage. Revision: 8/04. Copyright (c) Campbell Scientific, Inc.

Model ASPTC Aspirated Shield with Fine Wire Thermocouple

NOVALYNX CORPORATION MODEL 110-WS-16BP BAROMETRIC PRESSURE SENSOR INSTRUCTION MANUAL

User s Manual: MODEL 904-T CLINOMETER PAK

LP PYRA Installation and Mounting of the Pyranometer for the Measurement of Global Radiation:

LUDLUM MODEL 42-5 NEUTRON BALL CART. August 2015 Serial Number PR and Succeeding Serial Numbers

PAR Meter Instruction Manual. Product # Product # Year Warranty

Operating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2

IRTS-P Precision Infrared Temperature Sensor

Model 3024 Albedometer. User s Manual 1165 NATIONAL DRIVE SACRAMENTO, CALIFORNIA WWW. ALLWEATHERINC. COM

TB-2012M BAROMETRIC PRESSURE SENSOR USER S MANUAL

INSTRUCTION MANUAL. CS547 Conductivity and Temperature Probe and A547 Interface. Revision: 9/00. Copyright (c) Campbell Scientific, Inc.

WARRANTY AND ASSISTANCE

INSTRUCTION MANUAL. Laser Diode Sarcomere Length. Aurora Scientific Inc. 360 Industrial Parkway South, Unit 4 Aurora, Ontario, Canada L4G 3V7

3WHB10K 3-Wire Half Bridge Terminal Input Module

INSTRUCTION MANUAL. Model 207 Temperature and Relative Humidity Probe. Revision: 2/94. Copyright (c) Campbell Scientific, Inc.

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

WEATHER STATION FOR SOLAR FARM MONITORING

INSTRUCTION MANUAL. CS100 Barometric Pressure Sensor 3/05. Copyright (c) Campbell Scientific, Inc.

INSTRUCTION MANUAL. Model HFP01SC Self-Calibrating Soil Heat Flux Plate Revision: 3/14. Copyright Campbell Scientific, Inc.

Program Notes Use 2:1 voltage divider on Differential Input channel 1 to get full output range from C7

WXT510 Weather Transmitter Revision: 4/07

Model HFP01 Soil Heat Flux Plate Revision: 7/12

Owner's Manual NSL100B BUILDERS LEVEL NSL500B TRANSIT LEVEL

CAUTION: BEFORE YOU BEGIN, ALWAYS MAKE SURE THE WEAPON IS UNLOADED AND SAFE TO HANDLE.

MKIII -MB USER S GUIDE 09/18/18. Rev. 3

LP NET. 1 About LP NET. 2 Working Principle. TEL r.a. FAX

INSTRUCTIONS FOR ATTACHING THE NORDIC PLOW TO THE ACTION TRACKCHAIR

INSTRUCTION MANUAL. CS616 and CS625 Water Content Reflectometers Revision: 2/14. Copyright Campbell Scientific, Inc.

LP PYRA 10 AV. 1 Introduction. 2 Working Principle

Melton Series. Owners Manual. Lab Centrifuge. Model # s A C, A C, B C, B C, C C, C C, D C, D C

Field Pro Operation and Installation Manual

FIELD GUIDE FOR HORUS SCOPES

O P E R A T I N G M A N U A L

Model 3016 Secondary Standard Pyranometer. User s Manual 1165 NATIONAL DRIVE SACRAMENTO, CALIFORNIA WWW. ALLWEATHERINC. COM

SK Series BOX PLOW INSTALLATION & OWNER S MANUAL TABLE OF CONTENTS

SKELESCOPE REFLECTOR TELESCOPE WITH TABLE TOP TRIPOD SK-100 INSTRUCTION MANUAL

Using CSIM11 ph and ORP Probes with Campbell Scientific Dataloggers

SENSOR. Weather TRANSMITTERS

WHEN BRIGHTNESS COUNTS

HFM-6 and HFM-7 OPERATOR S MANUAL

SUGAR/BRIX REFRACTOMETER INSTRUCTION MANUAL

HFM-6, HFM-7, & HFM-8 OPERATOR S MANUAL

COMPARISON OF GUNN BELLANI RADIOMETER DATA WITH GLOBAL SOLAR RADIATION SENSOR (PYRANOMETER CM6B) Author. Mungai Peter N.

INSTRUCTION MANUAL ANALOG COLORIMETER

SI-111 Precision Infrared Radiometer Revision: 7/13

Today s Weather 24 Hour Forecaster

Owner's Manual. for AUTOMATIC LEVEL. NBL Series. FOR CUSTOMER SERVICE, PARTS & REPAIR, CALL Toll Free:

DWT-10 Instruction Manual

Model 791 Calibration Plate for Tiltmeters and Inclinometers

Instruction Manual. Kite-22 Camera Crane Starter Package (KITE-22-STARTER) Exclusive of Weights

IF YOU NEED ANY FURTHER HELP WITH YOUR GALILEO TELESCOPE PLEASE CALL MON-FRI 9AM - 5PM E.S.COM

VIBRATION METER. Instruction Manual.

QUANTUM CONCEPT. Swimming User s Manual

HEAVY DUTY SWING-OUT VALVE

technical bulletin Installation Instructions: Wall-Mount Dewpoint Probe ( )

Conversion Kit for: 7620 and 7820 Swing-Out Valves ft-lbs ft-lbs 8835/ ft-lbs

CHROMA Colorimeter Model 252

SECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components:

/2" Heavy Duty Swing-Out

Instruction Manual HI Refractometer for Invert Sugar Measurements

PLOW MOUNT KIT FOR POLARIS RANGER P/N ASSEMBLY / OWNERS MANUAL

227 Delmhorst Cylindrical Soil Moisture Block Revision: 5/13

BB1 & BB2 INSTALLATION & MAINTENANCE INSTRUCTIONS

HEAVY DUTY SWING-OUT VALVE

INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements

Small Absolute Pressure Sensor

WS-7047TWC Wireless 433 MHz Weather Station With Rainfall and Temperature. Instruction Manual

CAUTION: BEFORE INSTALLATION, CONFIRM THAT THE CROSSBOW IS UNLOADED AND IS IN SAFE OPERATING CONDITION.

1 YEAR LIMITED WARRANTY TELESCOPES

Franck-Hertz Experiment

Drive electronics pq12

GHI CORRELATIONS WITH DHI AND DNI AND THE EFFECTS OF CLOUDINESS ON ONE-MINUTE DATA

LDR Series BOX PLOW INSTALLATION & OWNER S MANUAL TABLE OF CONTENTS

iweathar EDGE V Installation Guide Copyright 2012 iweathar. All rights reserved.

Route and retain the clinometer cable so that it does not interfere with moving parts where the unit is mounted.

Small Gage Pressure Sensor

Applications. Remote Weather Station with Telephone Communications. Tripod Tower Weather Station with 4-20 ma Outputs

HS515G-M. PARALOW Circle Dot Sight. User's Manual. Holosun Technologies Inc. Phone: (909) Fax: (909)

Low Pressure Sensor Amplified Analog Output SM6295-BCM-S

Transcription:

CM3 PYRANOMETER INSTRUCTION MANUAL REVISION: 10/02 COPYRIGHT (c) 1999-2002 CAMPBELL SCIENTIFIC, INC.

This is a blank page.

WARRANTY AND ASSISTANCE This equipment is warranted by CAMPBELL SCIENTIFIC (CANADA) CORP. ( CSC ) to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. ***** Batteries are not warranted. ***** CSC's obligation under this warranty is limited to repairing or replacing (at CSC's option) defective products. The customer shall assume all costs of removing, reinstalling, and shipping defective products to CSC. CSC will return such products by surface carrier prepaid. This warranty shall not apply to any CSC products which have been subjected to modification, misuse, neglect, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied, including warranties of merchantability or fitness for a particular purpose. CSC is not liable for special, indirect, incidental, or consequential damages. Products may not be returned without prior authorization. To obtain a Return Merchandise Authorization (RMA), contact CAMPBELL SCIENTIFIC (CANADA) CORP., at (780) 454-2505. An RMA number will be issued in order to facilitate Repair Personnel in identifying an instrument upon arrival. Please write this number clearly on the outside of the shipping container. Include description of symptoms and all pertinent details. CAMPBELL SCIENTIFIC (CANADA) CORP. does not accept collect calls. Non-warranty products returned for repair should be accompanied by a purchase order to cover repair costs.

Table of Contents 1. General Description...1 2. Specifications...1 3. Installation...2 4. Wiring...4 5. Example Programs...6 5.1 Average Solar Radiation...6 5.2 Total Solar Radiation...7 5.2.1 Output Format Considerations...8 6. Maintenance...9 7. Calibration...9 Figures 3-1. CM3 and CM3MT...3 3-2. 015 Pyranometer Mounting Arm...3 3-3. 025 Crossarm Stand and 019ALU Crossarm...4 3-4. UTKZ Leveling Fixture and Crossarm Mount and UT018 Tower Mounting Bracket and Crossarm...4 4-1. CM3 Wiring...5 Tables 5-1. Multipliers Required for Average Flux and Total Flux Density in SI and English Units...6 i

This is a blank page.

1. General Description This manual provides information for interfacing Kipp & Zonen s CM3 Pyranometer to a CR10(X), CR510, CR23X, CR7 or 21X datalogger. The CM3 is shipped with an instruction manual provided by Kipp & Zonen that contains information concerning the CM3 s construction, spectral sensitivity, cosine response, and a simple sensor check out procedure. Included with the sensor and manual is a calibration certificate with the sensor calibration constant and serial number. Cross check this serial number against the serial number on your CM3 to ensure that the given calibration constant corresponds to your sensor. The CM3 pyranometer is designed for continuous outdoor use. Due to its flat spectral sensitivity from 300 to 3000 nm, it can be used in natural sunlight, under plant canopies, in green houses or buildings, and inverted to measure reflected solar radiation. Two CM3s can be used in combination with an albedometer fixture (K&Z s CAF 1) to measure albedos. The CM3 can also be used to measure most types of artificial light (Xenon lamps, Halogen lamps, etc.). The CM3 pyranometer consists of a thermopile sensor, a housing, a dome, and a cable. The thermopile is coated with a black absorbent coating. The paint absorbs the radiation and converts it to heat. The resultant temperature difference is converted to a voltage by the copper-constantin thermopile. The thermopile is encapsulated in the housing in such a way that it has a field of view of 180 degrees and the angular characteristics needed to fulfill the cosine response requirements. 2. Specifications The CM3 is an ISO Second Class pyranometer. While the worst case accuracy for daily sums given by Kipp & Zonen is +10%, the typical accuracy is +5%. Tests at Campbell Scientific on one CM3 indicated an accuracy of +2% when compared to a recently calibrated Eppley PSP. ISO SPECIFICATIONS: Response Time 95%: 18 seconds Zero offset due to 200 W/m 2 thermal radiation: < 15 Wm -2 Zero offset due to temperature change of 5ºK / hr: < +4 Wm -2 Non stability (% change/year): < + 1% Non linearity (at 1000 W/m 2 ): < + 2.5% Directional error (at 1000 W/m 2 ): < + 25 Wm -2 Temperature Dependence of 1

sensitivity: Tilt response (+80º) (at 1000 W/m 2 ): < + 2% + 6% (-10 to + 40ºC) OTHER SPECIFICATIONS Expected accuracy for daily sums: + 10% Spectral range (50% points, nm): 305-2800 nm Sensitivity: 10-35 µv/wm -2 Expected signal output in atmospheric application: 0-50 mv Impedance: 79-200 (Ω) Operating Temperature: -40 to +80ºC Max. irradiance: 2000 Wm -2 Detector: Copper-constantin multi junction thermopile Cable length: 15 feet (5 m) Level accuracy: 1 degree DIMENSIONS / SHIPPING DIMENSIONS CM3: 3x3x3 in / 6x6x6 in CM3MT: 1x5x5 in / 6x6x6 in WEIGHT/SHIPPING WEIGHT CM3: 0.8 lbs / 3 lbs CM3MT: 0.6 lbs / 3 lbs 3. Installation The CM3 should be mounted such that it is never shaded by the tripod/tower or other sensors. To ensure accurate measurements, the CM3 should be mounted using the CM3MT base/leveling fixture or equivalent. The CM3MT incorporates a bubble level and three adjustment screws. Mount the CM3 Pyranometer to the CM3MT mount using the two long screws provided. The screws are slightly shorter than ideal but should provide adequate holding strength. 2

CM3 Sensor Bubble Level CM3MT PN7790(3) 015 Pyranometer Mounting Arm or 025 Pyranometer Crossarm Stand FIGURE 3-1. CM3 and CM3MT Install the CM3MT Mount on either the 025 Pyranometer Cross Arm Stand, the 015 Pyranometer Mounting Arm, or the UTKZ (not yet available, please call) before mounting them to the tower or tripod. This is done by first threading the screws through the springs and just barely through the mounting plate. This helps remove any paint that might have gotten into the threaded holes. Second, remove two of the screws/springs, slide the CM3MT Mount onto the remaining screw by slightly compressing the spring. Slide the remaining two springs between the CM3MT and the 015 or 025 mount and install the remaining two screws. Tighten the screws until all three springs have been compressed about 1/8 inch. Once the pyranometer mount has been installed on the tripod or tower, tighten the appropriate screws until the bubble indicates the sensor is level. 015ARM Pyranometer Mounting Arm Tripod Mast or Tower Support FIGURE 3-2. 015 Pyranometer Mounting Arm 3

025STAND Crossarm Stand 019ALU Crossarm 3/4 x 3/4 NU-RAIL 3/4 x 1 NU-RAIL 3/4 x 3/4 NU-RAIL FIGURE 3-3. 025 Crossarm Stand and 019ALU Crossarm UTKZ Leveling Fixture and Crossarm Mount (includes CM3MT) CM3 UT018 Tower Mounting Bracket UT018 Crossarm Tower Support FIGURE 3-4. UTKZ Leveling Fixture and Crossarm Mount and UT018 Tower Mounting Bracket and Crossarm 4. Wiring Use Differential Voltage measurement instruction 2 to measure the CM3. The CM3 wiring diagram is shown in Figure 4-1. The red lead is connected to the high side (H) of any differential channel. The blue lead is connected to the corresponding low (L) side of the differential channel. On a CR10(X) or CR510, the white lead is connected to an analog ground (AG) and the clear to ground (G). On the CR23X, 21X or CR7 both the white and clear leads are connected to ground ( ). While a differential measurement is better and preferred, the CM3 can be measured on a single-ended channel using Instruction 1 if the power induced voltages discussed in the following paragraph are prevented. If a 21X is used to measure the CM3 and it powers a 12 VDC sensor or 12 VDC radio, the current drawn off the 12 VDC supply may cause a difference in 4

ground potential between the 21X ground terminals and the reference ground point in the datalogger. This ground potential results in an offset on single ended measurements. This offset can be as large as +60 mv. Thus, single ended measurements should be avoided. The offset does not, however, affect differential measurements. While the 21X is the logger most susceptible to this condition, the other dataloggers can be affected if the amount of power is too large or the power return line is connected to analog ground (AG). White = Signal Ground Blue = Signal Reference Red = Signal Clear = Shield FIGURE 4-1. CM3 Wiring Color Function CR10(X), 21X/CR7 CR23X CR510 Red Signal Differential Channel - H Differential Channel - H Differential Channel - H Blue Signal Reference Differential Channel - L Differential Channel - L Differential Channel - L White Signal Ground AG Clear Shield G 5

5. Example Programs Solar radiation can be reported as an average flux density (W m -2 ) or daily total flux density (MJ m -2 ). The appropriate multipliers are listed in Table 5-1. Programming examples are given for both average and daily total solar radiation. The output from the CM3 varies from 10-35 x 10-6 V / W m -2. Given a maximum solar radiation of 1500 W m -2, the maximum sensor output voltage will be 15-52.5 mv. Example: (21.87 x 10-6 V W -1 m 2 ) (1500 W m -2 ) = 0.03281 V or 32.81 mv TABLE 5-1. Multipliers Required for Average Flux and Total Flux Density in SI and English Units UNITS MULTIPLIERS W m -2 1 C 10 3 MJ m -2 kj m -2 cal cm -2 min -1 14333. C 10 6 cal cm -2 (average) t (total) C 10 9 t (total) C 10 6 (average) t 0. 02389 (total) C 10 6 C=CM3 calibration: eg. 21.87x10-6 V / W m -2 t = datalogger execution interval in seconds 5.1 Average Solar Radiation Example 1 shows the program instructions used by a CR10X to measure the signal from the CM3. A sixty-minute average is calculated and stored in final storage. 6

Example 1 ;{CR10X} ; *Table 1 Program 01: 10 Execution Interval (seconds) 1: Volt (Diff) (P2) 1: 1 Reps 2: 24 250 mv 60 Hz Rejection Range 3: 1 DIFF Channel 4: 1 Loc [ W_m2 ] 5: 45.725 Mult ;multiplier = (1 / 0.02187 mv / W/m2) 6: 0.0 Offset ;Set negative values to zero. ; 2: If (X<=>F) (P89) 1: 1 X Loc [ W_m2 ] 2: 4 < 3: 0 F 4: 30 Then Do 3: Z=F x 10^n (P30) 1: 0.0 F 2: 00 n, Exponent of 10 3: 1 Z Loc [ W_m2 ] 4: End (P95) 5: If time is (P92) 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 10 Set Output Flag High (Flag 0) 6: Real Time (P77) 1: 1220 Year,Day,Hour/Minute (midnight = 2400) 7: Average (P71) 1: 1 Reps 2: 1 Loc [ W_m2 ] *Table 2 Program 02: 0.0000 Execution Interval (seconds) *Table 3 Subroutines End Program 5.2 Total Solar Radiation In Example 2 a CR10X is used to record daily total flux density. This total flux density is in MJ m -2 day -1 to avoid the need for high resolution discussed in 5.2.1. Negative values are set to zero before they are added to the running total. 7

5.2.1 Output Format Considerations If the solar radiation is totalized in units of kj m -2, there is a possibility of over-ranging the output limits. The largest number that the datalogger can output to final storage is 6999 in low resolution and 99999 in high resolution (Instruction 78, Set Resolution). Assume that the daily total flux density is desired in kj m -2. Assume an irradiance of 0.5 kw m -2, the maximum low-resolution output limit will be exceeded in just under four hours. This value was found by taking the maximum flux density the datalogger can record in low resolution and dividing by the total hourly flux density. 39. hr = 6999kJm 2 2 1 1 ( 05. kjm s )( 3600shr ) To circumvent this limitation, record an average flux (see Example 1). Then, during post processing, multiply the average flux by the number of seconds in the output interval to arrive at an output interval flux density. Sum the output interval totals over a day to find a daily total flux density. Another alternative is to record total flux using the high-resolution format (Instruction 78, see Datalogger manual for details). The disadvantage of the high-resolution format is that it requires four bytes of memory per data point, consuming twice as much memory as low resolution. (Given the larger amount of memory in today s dataloggers, this may not matter.) Example 2 ;{CR10X} ; *Table 1 Program 01: 10 Execution Interval (seconds) 1: Volt (Diff) (P2) 1: 1 Reps 2: 24 250 mv 60 Hz Rejection Range 3: 1 DIFF Channel 4: 1 Loc [ MJ_m2 ] 5: 0.45725 Mult ;multiplier = [10s / (21870 mv / MJ/m2)] Step 1 of 2 6: 0.0 Offset 2: Z=X*F (P37) 1: 1 X Loc [ MJ_m2 ] 2: 0.001 F Step 2 of 2 3: 1 Z Loc [ MJ_m2 ] ;Set negative values to zero. ; 3: If (X<=>F) (P89) 1: 1 X Loc [ MJ_m2 ] 2: 4 < 3: 0 F 4: 30 Then Do 8

4: Z=F x 10^n (P30) 1: 0.0 F 2: 00 n, Exponent of 10 3: 1 Z Loc [ MJ_m2 ] 5: End (P95) 6: If time is (P92) 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 10 Set Output Flag High (Flag 0) 7: Real Time (P77) 1: 1220 Year,Day,Hour/Minute (midnight = 2400) 8: Totalize (P72) 1: 1 Reps 2: 1 Loc [ MJ_m2 ] *Table 2 Program 02: 0.0000 Execution Interval (seconds) *Table 3 Subroutines End Program 6. Maintenance 7. Calibration On a monthly basis the level of the pyranometer should be checked. Any dust or debris on the sensor window should be removed. The debris can be removed using water (de-ionized or distilled) or alcohol. Recalibration is suggested every two years. Calibrations can be done in one of two ways. One method is to calibrate the sensor in the field by placing a transfer standard (a sensor that has been calibrated against a secondary standard ) next to the sensor being calibrated. Preferably daily totals of several days should be compared. The calibration factor could be corrected if results differ by more than six percent. Another method would be to send the sensor to a facility that has a secondary standard. Contact Kipp & Zonen (www.kippzonen.com) for the nearest calibration facility, or send it to Campbell Scientific and they will have it recalibrated. 9

This is a blank page. 10