Fast and Accurate Analysis of Fluorotelomer Alcohols and Acrylates using Triple Quadrupole GC-MS/MS

Similar documents
Fast and Accurate Quantitation of Perfluorinated Sources from Textiles using Gas Chromatography-Triple Quadrupole Mass Spectrometry

Fast Screening and Quantitation of Perfluorinated Sources from Textiles using Chemical Ionization GC-MS

Quantitative Determination of Dioxins in Drinking Water by Isotope Dilution using Triple Quadrupole GC-MS/MS

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

Research of Polychlorinated Biphenyls (PCBs) in Vegetables by GC-MS/MS

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Simultaneous determination of 74 kinds of pesticides in Chinese herbal medicine listed in Chinese pharmacopoeia (2015) by GPC-GC-MS/MS

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography

Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS

Quantitative Analysis of Opioids Using a Triple-Quadrupole GC/MS/MS

Qualitative Profiling of Co-polymer Polyethylene Terephthalate through Multifunctional Pyrolyzer-GC/MS by various Thermal Treatment Techniques

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS

Screening of pesticides in water using SPE on line

+ butanal + 2-picolineborane. + acetic acid. + butanal + 2-picolineborane. + acetic acid

Consumables Guide for the Cannabis Testing Lab

Development of 2D-LC/MS/MS Method for Quantitative Analysis of 1α,25-Dihydroxylvitamin D3 in Human Serum

Chiral Separation Using SFC and HPLC

Analysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer)

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

Benzodiazepines and Barbiturates

Development of high speed CYP cocktail inhibition assay using UHPLC-MS/MS

Quantitative Analysis of β-lactam Antibiotics in Human Plasma by High Sensitivity LC/MS/MS Method

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Highly sensitive quantitative analysis of Felodipine and Hydrochlorothiazide from plasma using LC/MS/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of 59 potential Allergens in Perfumes by twin-line fast GCMSMS

Highly Sensitive and Rugged GC/MS/MS Tool

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

Evaluation of a novel glass vial overcoming adsorption effect for pharmaceutical drugs

Protein Sequencer PPSQ-31A / 33A C297-E055B

ENVIRONMENTAL analysis

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

A Fully Validated LC-MS/MS Method for Analysis of Trace Fluorotelomer Alcohols from Water

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Validation Report 18

Pittcon P

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Definitive EtG/EtS LC-MS/MS Analysis:

Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404

Analysis of Fast Moving Consumer Goods (FMCG) using GCMS with static and dynamic headspace (HS)

Gas Chromatograph Mass Spectrometer GCMS-TQ8050 C146-E326

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Analyst Software. Automatic Optimization Tutorial

C146-E209. Headspace Samplers. HS-20 Series

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Liquid Chromatography Mass Spectrometry. Shimadzu Corporation, UK; 2 Phytocontrol, France; 3 Shimadzu France, France. Trimesium S + CH 3 ETU

Technical Report. Abstract: Pu Wang 1, Huizhong Sun 1, Qinghua Zhang 1, Feifei Tian 2, Lei Cao 2

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

Supporting Information

Application News. No. C154. Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Sulfotepp impurities in Chlorpyrifos EC formulations

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

ApplicationNOTE A STUDY OF THE ANALYSIS OF POLYBROMINATED DIPHENYL ETHER FLAME RETARDANTS BY GC/MS/MS. Introduction. Materials and Methods

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Agilent 6495 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE

Spectrophotometer for Life Science. BioSpec-nano C101-E112D

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Tecan AC Extraction Plate

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Technical Report Abstract: Keywords: RoHS, phthalate esters, Py-GC/MS, Py-Screener 1. Introduction

Appendix II- Bioanalytical Method Development and Validation

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Transcription:

POCON1647E Fast and Accurate Analysis of Fluorotelomer Alcohols and Acrylates using Triple Quadrupole GCMS/MS ASMS 2016 ThP 6 Cynthia Melanie Lahey 1, Guo Wei Elgin Ting 2, Hui Xian Crystal Yeong 1, Jackie Jackie 1, Chiew Mei Chong 1, Lai Chin Loo 1 1 Application Development & Support Centre, Shimadzu (Asia Pacific) Pte. Ltd., 7 Science Park Drive #0201/08, Singapore 118264, 2 School of Physical and Mathematical Sciences, Nanyang Technological University, 21 Nanyang Link SPMS0401, Singapore 627371

Acrylates using Triple Quadrupole GCMS/MS Introduction Fluorotelomerbased polymers (FTPs) are utilised as oil and water repellents in consumer products [1]. However, its extensive use in products such as textiles has raised environmental concerns, as residual fluorotelomer alcohols and acrylates (FTOHs and FTAs) from FTPs are considered as probable precursors of bioaccumulative and potentially toxic perfluorocarboxylic acids [1,2,3]. Traditionally, the determination of FTOHs has been performed using single quadrupole GCMS in chemical ionization (CI) mode [4,5]. Although CI can achieve sensitive detection limits (less than 20 fg/ul for FTOHs), it is sometimes avoided due to higher upkeep and running cost than electron ionization (EI) systems [5]. Furthermore, EI is a more comprehensive analytical solution than CI in applications. This motivated the use of GCMS/MS system in EI mode in this study. Methods and Materials A mixture of four fluorotelomer alcohols (FTOHs) and three fluorotelomer acrylates (FTAs) was prepared from neat standards (see Table 1). FTAcr 8:2 was purchased from Sigma Aldrich. The rest were purchased from Apollo Scientific. The internal standard compound, 2Perfluorobutyl[1,1,2,22H 4 ]ethanol (MFBET), was purchased from Wellington Laboratories. Calibration curve solutions were prepared in ethyl acetate (pesticide residue and PCB grade, from Kanto Kagaku). Textile samples were selected as test matrices for quantitation of targeted analytes. Sample extraction was carried out in ethyl acetate and preconcentrated before injection into the system. A triple quadrupole GCMS/MS, GCMSTQ8040 (Shimadzu Corporation, Japan) in EI mode by multiple reaction monitoring (MRM) acquisition was used in this work. Shimadzu AOC20i/s autosampler was utilised as the autosampler. SHStabilwax column (30 m length, mm ID, um df) was used for separation of the targeted fluorotelomers. Table 1. List of FTOHs and FTAs Analysed in This Experiment Compound 1H,1H,2H,2HPerfluorohexan1ol 1H,1H,2H,2HPerfluorooctan1ol 1H,1H,2H,2HPerfluorodecan1ol 1H,1H,2H,2HPerfluorododecan1ol 1H,1H,2H,2HPerfluorooctyl acrylate 1H,1H,2H,2HPerfluorodecyl acrylate 1H,1H,2H,2HPerfluorododecyl acrylate Acronym FTOH 4:2 FTOH 6:2 FTOH 8:2 FTAcr 6:2 FTAcr 8:2 FTAcr 10:2 CAS No. 2043472 647427 67837 865861 1752726 270545 17741605 2

Acrylates using Triple Quadrupole GCMS/MS GCMS/MS Analysis Condition Autosampler # of rinses (prerun) : 3 # of rinses (postrun) : 3 # of rinses with sample : 1 Injection volume : 1 µl GC Injection temp : 240ºC Injection mode : Splitless Flow control mode : Linear velocity Linear velocity : 43.4 cm/s High Pressure Inj : 3 kpa for 1.5 min Column temp. prog. : 40ºC (1 min) 20ºC/min 0ºC 30ºC/min 240ºC (2 min) MS/MS Ion source temp. : 200ºC Interface temp. : 240ºC Acquisition mode : MRM MRM transitions : See the table below Compound name MRM (Target) CE (V) MRM (Ref 1) CE (V) MRM (Ref 2) CE (V) MFBET (IS) >130.10 248.00>7.10 248.00>130.10 FTOH 4:2 16.00>127.10 244.00>16.00 3 FTAcr 6:2 418.00>.10.00>57.10 12 FTOH 6:2 344.00>127.10 FTAcr 8:2 518.00>.10 12.00>57.10 12 FTOH 8:2 36.00>127.10 FTAcr 10:2 618.00>.10.00>57.10 12 13>6.00 18 3

Acrylates using Triple Quadrupole GCMS/MS Sample Preparation Individual neat standard was mixed and diluted in ethyl acetate to prepare calibration standard mixtures of 1, 2, 5, 10, 20, 50, 100, and 200 ng/ml. Calibration curves were in the range of 1 100 ng/ml, except those of FTOH 6:2 and FTOH 8:2 (1 200 ng/ml). The peak shape of FTOH 4:2 was poorer at low concentration, than that of the other compounds. Hence, the calibration curve of FTOH 4:2 was generated in the range of 5 200 ng/ml. Internal standard (MFBET) was spiked to all standard mixtures at the concentration of 100 ng/ml. Each textile sample was cut into dimensions of 2 mm x 2 mm, after which 1 g of the sample was placed into a vial. Subsequently, 10 ml of ethyl acetate was added into the vial and capped. The sample vial was heated at 60 ºC for 2 hours. The extract was subjected to a 0.22 µm nylon filter after heating. Finally, the filtered extract was concentrated 10 times and spiked with 100 ng/ml MFBET (Internal Standard), before injection to the GCMS/MS. Results Calibration Curves Linearity (R 2 ) of the internal standard calibration curve of each compound is greater than 0. (refer to Figure 1). 1.50 1.50 1.50 1.25 1.25 1.25 5200 ng/ml R 2 = 0.818 0.0 0.5 1.0 Conc. Ratio 1100 ng/ml R 2 = 0.7211 Conc. Ratio 1200 ng/ml R 2 = 0.665 0.0 0.5 1.0 Conc. Ratio Conc. Ratio FTOH 4:2 FTAcr 6:2 FTOH 6:2 FTAcr 8:2 1100 ng/ml R 2 = 0.7634 4 3 1200 ng/ml R 2 = 0.8352 0.0 0.5 1.0 Conc. Ratio 1100 ng/ml R 2 = 0.7811 Conc. Ratio FTOH 8:2 FTAcr 10:2 2 1 1100 ng/ml R 2 = 0. 0 Conc. Ratio Figure 1: Calibration Curves 4

Acrylates using Triple Quadrupole GCMS/MS Mass Chromatograms ISTD >130.10 3.1e5 Q 16.00>127.10 1.38e4 Q 418.00>.10 1.84e4 Q 344.00>127.10 1.52e4 10 10 10 10 4.2 4.3 4.4 4.5 4.6 4.7 4.2 4.3 4.4 4.5 4.6 4.7 4.2 4.3 4.4 4.5 4.6 4.7 4.6 4.7 4.8 4. 5.0 5.1 MFBET (IS), 100 ng/ml FTOH 4:2 FTAcr 6:2 FTOH 6:2 Q 518.00>.10 2.87e4 Q 36.00>127.10 e4 Q 618.00>.10 1.e4 Q 7.83e4 10 10 10 10 4.8 4. 5.0 5.1 5.2 5.3 5.1 5.2 5.3 5.4 5.5 5.6 5.5 5.6 5.7 5.8 5. 6.0 5.7 5.8 5. 6.0 6.1 6.2 FTAcr 8:2 FTOH 8:2 FTAcr 10:2 Figure 2: Mass Chromatgrams of 5 ng/ml Calibration Standard Instrument Detection Limit and Method Precision The Instrument detection limit (IDL) of each compound was determined by injecting 0.5 ng/ml standard mixture. IDLs were estimated to be between 0.1 0.5 ng/ml, whereby the S/N ratio is 3. However, due to the poor peak shape of FTOH 4:2 at low concentration, its IDL was estimated to be 2 ng/ml. FTAcr 6:2 at 0.5 ng/ml at 0.5 ng/ml Q 418.00>.10.04e2 Q 7.77e3 10 10 4.2 4.3 4.4 4.5 4.6 4.7 5.7 5.8 5. 6.0 6.1 6.2 LOD = 0.5 ng/ml LOD = 0.1 ng/ml Figure 3: FTAcr 6:2 and at 0.5 ng/ml for Calculation of IDLs 5

Acrylates using Triple Quadrupole GCMS/MS Method precision was evaluated by calculating the repeatability at the low and midcalibration levels by injecting seven replicates of standard mixtures at 1 ng/ml (except FTOH 4:2, 5 ng/ml) and 20 ng/ml, as displayed in Table 2. All RSD were below 10. Table 2. Repeatability of Area Ratio (n=7) at Low and MidCalibration Levels Repeatability of Area Ratio (n=7) Compound name FTOH 4:2 FTAcr 6:2 FTOH 6:2 FTAcr 8:2 FTOH 8:2 FTAcr 10:2 RSD of low calibration level (1 ng/ml)* 4.7 5.5 6.1 3.2 8.0 5.6 3.4 RSD of mid calibration level (20 ng/ml) 3.2 2.4 0.8 0.5 2.7 1.6 1.4 *Except FTOH 4:2, which was 5 ng/ml spike Matrix Effect Matrix effect (ME ) of the MRM method was calculated by comparing the peak area ratios of postspiked standards in the sample extracts to those in the diluent. The results are shown in Table 3. This was performed in three replicates. Most of the matrix effects calculated are more than 100, and especially greater for low concentration spikes. Hence, a matrixmatch calibration method for more accurate quantification needs to be considered. Table 3. Matrix Effect (ME ) from Different Textile Samples (n=3) Blue Cloth Matrix Effect (ME ) Black Cloth Matrix Effect (ME ) Red Cloth Matrix Effect (ME ) Compound name Post spiked with 1 ng/ml* Postspiked with 20 ng/ml Post spiked with 1 ng/ml* Postspiked with 20 ng/ml Post spiked with 1 ng/ml* Postspiked with 20 ng/ml FTOH 4:2 121 8 124 11 142 125 FTAcr 6:2 180 107 162 171 207 171 FTOH 6:2 30 113 286 2 182 6 FTAcr 8:2 172 118 204 15 18 12 FTOH 8:2 388 124 428 177 373 17 FTAcr 10:2 187 130 201 221 203 218 15 118 1 174 204 182 *Except FTOH 4:2, which was 5 ng/ml spike 6

Acrylates using Triple Quadrupole GCMS/MS (x10,000) 13>6.00 5.0 2.5 After 24 injections Without guard column (x10,000) 13>6.00 7.5 5.0 2.5 5.85 5.0 5.5 6.00 5.75 5.80 5.85 5.0 (x10,000) 13>6.00 5.0 2.5 After 54 injections With guard column (x10,000) 13>6.00 7.5 5.0 2.5 6.35 6.40 6.45 6.50 6.30 6.35 6.40 6.45 Figure 4: in Spiked Textile Samples (1 ng/ml) After injecting the sample extracts, retention times of the analytes gradually shifted and the peak shapes of FTAcr 10:2 and deteriorated. This phenomena did not occur with the calibration standards (prepared in diluent), before the sample extracts were analysed. As such, the sample matrix might be the cause of column deterioration. It was observed that the use of a 5 m guard column in front of the analytical column helped to prevent the peak shape deterioration, as displayed in Figure 4. Better sample extraction method should be explored to extend the column lifetime. Conclusion Analysis of FTOHs and FTAs was developed using triple quadrupole GCMS/MS within 7 minutes. Excellent calibration curve linearity and repeatability were obtained from this experiment. The sample extracts in general gave high matrix effect (ME ) and deterioration to the column, therefore the use of guard column is recommended. Future study might include quantification by matrixmatch calibration and improvement of sample extraction method to reduce column deterioration. 7

Acrylates using Triple Quadrupole GCMS/MS References [1] Takemine, S.; Matsumura, C.; Okada, Y.; Suzuki, M.; Yoshida, M.; Tsurukwa, M.; Nakano, T. Investigation of Fluorinated Telomer Alcohols and Related Compounds in Fluorinated Oil and Water Repellants. Organohalogen Compounds. 200, 71, 18501853. [2] Yoo, H.; Washington, J. W.; Jenkins, T. M.; Ellington, J. J. Quantitative Determination of Perfluorochemicals and Fluorotelomer Alcohols in Plants from BiosolidAmended Fields using LC/MS/MS and GC/MS. Environ. Sci. Technol. 2011, 45, 78570. [3] Henderson, W. M.; Weber, E. J.; Duirk, S. E.; Washington, J. W.; Smith, M. A. Quantification of FlurotelomerBased Chemicals in Mammalian Matrices by Monitoring Perfluroalkyl Chain Fragments with GC/MS. J. Chromatogr. B. 2007, 846, 5161. [4] Ellington, J. J.; Washington, J. W.; Evans, J. J.; Jenkins, T. M.; Hafner, S. C.; Neill, M. P. Analysis of Flurotelomer Alcohols in Soils: Optimization of Extraction and Chromatography. J. Chromatogr. A. 200, 1216, 53475354. [5] Washington, J. W.; Ellington, J. J.; Jenkins, T. M.; Evans, J. J.; Yoo, H.; Hafner, S. C. Degradability of An AcrylateLinked, Fuorotelomer Polymer in Soil. Environ. Sci. Technol. 200, 43, 66176623. Disclaimer: Disclaimer: The products and applications in this presentation are intended for Research Use Only (RUO). Not for use in diagnostic procedures. First Edition: June, 2016 For Research Use Only. Not for use in diagnostic procedure. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. www.shimadzu.com/an/ The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. Company names, product/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation or its affiliates, whether or not they are used with trademark symbol TM or. Thirdparty trademarks and trade names may be used in this publication to refer to either the entities or their products/services. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. Shimadzu Corporation, 2016