Secrets of GC Column Dimensions

Similar documents
Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Selection of a Capillary GC Column

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Introduction to Capillary GC

Selection of a Capillary

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Introduction to Capillary GC

Understanding Gas Chromatography

Choosing the Correct GC Column Dimensions and Stationary Phase

Water Injections in GC - Does Water Cause Bleed?

Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns

How To Select the Correct GC Column. Simon Jones Application Engineer

Evaluation of Capillary Columns for General Performance Parameters

Trajan SGE GC Columns

So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist

Practical Fast Gas Chromatography for Contract Laboratory Program Pesticide Analyses *

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Headspace Technology for GC and GC/MS: Features, benefits & applications

Introduction and Principles of Gas Chromatography

GC Resolution Do You See What I See?

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

Application. Gas Chromatography February Introduction

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions

Activity in the FID Detection Port: A Big Problem if Underestimated

SUPELCO. Improved Performance for Cyanopropylphenyldimethylpolysiloxane (SPB-1701, SPB-1301) Capillary Columns

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

Speakers. Moderator. John V Hinshaw GC Dept. Dean CHROMacademy. Tony Taylor Technical Director CHROMacademy. Dave Walsh Editor In Chief LCGC Magazine

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Volatile organic compounds (VOCs):

Theory and Instrumentation of GC. Chromatographic Parameters

Study of Residual Solvents in Various Matrices by Static Headspace

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Author. Abstract. Introduction

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Lab 3 Guide: Gas Chromatography (GC) (Sept 8-14)

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Introduction to Gas Chromatography

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

VUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0.

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

The Effects of Carrier Gas Viscosity and Diffusion on Column Efficiency in Capillary Gas Chromatography

Column Selection. there is more to life than a boiling point column. Jaap de Zeeuw Restek Corporation, The Netherlands. Copyrights: Restek Corporation

New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Chromatography. Gas Chromatography

Principles of Gas- Chromatography (GC)

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Get Selective. By Jaap de Zeeuw

Ultra-Inert chemistry for Trace Level Analysis

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Fast and Accurate Analysis of PBDEs in a Single Run, Including BDE-209

Fast and Accurate Analysis of PBDEs in a Single Run, Including

CHAPTER 6 GAS CHROMATOGRAPHY

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Agilent J&W GC Column Selection Guide

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Analysis of BTEX in Natural Water with SPME

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Highly Sensitive and Rugged GC/MS/MS Tool

Rtx-BAC Plus 1 and Rtx-BAC Plus 2 Columns Advanced Technology for Fast, Reliable Measurement of Alcohol in Blood

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

NEW Rtx -BAC Plus 1 and Rtx -BAC Plus 2 Columns

2401 Gas (liquid) Chromatography

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

CHEM340 Tutorial 4: Chromatography

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Gas Chromatography. Introduction

Electronic Supplementary Information (ESI)

Gas Chromatography. 1. Introduction. 2. Stationary phases. 3. Retention in Gas-Liquid Chromatography. 4. Capillary gas-chromatography

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Gas Chromatography (GC)

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Supelco Ionic Liquid GC Columns Introduction to the Technology

Low Bleed Stationary Phases for Gas Chromatography

Analyzing Semi-Volatiles by GC-MS

GAS CHROMATOGRAPHY (GC)

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Agilent J&W GC Column Selection Guide SPEED YOUR SELECTION WITH THIS ONE-STOP RESOURCE

GCXGC-qMS with Total Flow Modulation for Volatile secondary metabolites analysis from organically-grown apples by HS-SPME

Transcription:

Secrets of GC Column Dimensions GC Columns and Consumables Simon Jones Application Engineer May 20, 2008 Slide 1

Secrets of GC Column Dimensions Do I have the right column phase? Resolution Equation Changes in Dimensions: Length Diameter Film Thickness Method Translation Software Carrier gas (if time) Slide 2

Start with the Right Phase DB-1 15m x 0.32mm, 0.25µm Oven: 40 C for 2 min 40-120 C at 5 C/min 0 2 4 6 8 10 12 14 Time (min.) DB-Wax 15m, 0.32mm, 0.25µm Oven: 80-190 C at 20 C/min 0 1 2 3 4 5 Time (min.) Slide 3

Variables that affect chromatography Stationary Phase Temperature Programming Carrier Gas: type and linear velocity Column Length Film Thickness Internal Diameter There s no such thing as a free lunch Slide 4

Resolution N k α 1 R = s 4 k + 1 α Efficiency Retention Selectivity N = ƒ (gas, L, r c ) k = ƒ (T, d f, r c ) α = ƒ (T, phase) L = Length r c = column radius d f = film thickness T = temperature Slide 5

Column Dimensions Length Diameter Film Thickness Slide 6

Resolution N k α 1 R = s 4 k + 1 α Efficiency Retention Selectivity N = ƒ (gas, L, r c ) k = ƒ (T, d f, r c ) α = ƒ (T, phase) L = Length r c = column radius d f = film thickness T = temperature Slide 7

Column Length and Efficiency (Theoretical Plates) Length (m) N 15 69,450 30 138,900 60 277,800 0.25 mm ID n/m = 4630 (for k = 5) Slide 8

Column Length and Resolution R α N α L Length X 4 = Resolution X 2 t α L Upside = Cut a bunch off during routine inlet maintenance and not lose a lot of Resolution Slide 9

Column Length VS Resolution and Retention: Isothermal R=0.84 2.29 min R=1.16 4.82 min R=1.68 8.73 min 15 m 30 m 60 m Double the plates, double the time but not double the resolution Slide 10

Column Length and Cost 15m 30m 60m $ $ $ $ $ Slide 11

Length Summary If you Increase Length: Efficiency Resolution Analysis Time Pressure Cost Increase Increase Increase Increase Increase Slide 12

Resolution N k α 1 R = s 4 k + 1 α Efficiency Retention Selectivity N = ƒ (gas, L, r c ) k = ƒ (T, d f, r c ) α = ƒ (T, phase) L = Length r c = column radius d f = film thickness T = temperature Slide 13

Column Diameter and Carrier Gas Flow Lower flow rates: Smaller diameter columns Higher flow rates: Larger diameter columns Low flow rates : GC/MS High flow rates: Headspace, purge & trap Slide 14

Column Diameter Capillary Columns I.D. (mm) Common Name 0.53 Megabore 0.45 High speed Megabore 0.32 Widebore 0.20-0.25 Narrowbore 0.18 Minibore 0.10 Microbore 0.05 Nanobore Slide 15

Column Diameter - Theoretical Efficiency Total Plates I.D. (mm) n/m 5 m N ~ 112,000 0.05 23,160 10 m N ~ 112,000 0.10 11,580 0.18 6,660 20 m N ~ 112,000 0.20 5830 0.25 4630 30 m N ~ 112,000 0.32 3660 0.45 2840 k = 5 0.53 2060 Slide 16

Different Column I. D. Equal Phase Ratios Column: DB-624 30 m, 0.53 mm, 3 m Carrier: Oven: Helium, 40(cm/sec) 65 C Injection: Split Detector: FID Column: DB-624 30 m, 0.32 mm, 1.8 m 0 5 10 15 20 Time (min) Slide 17

PHASE RATIO (β) Film Thickness Column Dimensions Phase Ratio β 30 m x.53 mm x 3.0 μm 44 30 m x.32 mm x 1.8 μm 44 K C = k β β = r 2d f Slide 18

High Resolution Megabore (0.45 mm diameter) Same outer diameter as the Megabore No special hardware required Smaller inner diameter (0.45mm) Maintain phase ratio (Beta) Methods are easy to translate!

High Resolution Megabore Column Dimensions Phase Ratio β 30 m x.53 mm x 3.0 μm 44 30 m x.45 mm x 2.55 μm 44

High-SPEED Megabore Same Resolution - Faster Analysis! DB-5 30 m, 0.53 mm I.D., 0.5 µm 1 2 R = 3.33 3 4 5 36 cm/sec.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 DB-5 30 m, 0.45 mm I.D., 0.42 µm 45.9 cm/sec 1 2 R = 3.32 3 4 5 1. Benzene 2. Toluene 3. Ethylbenzene 4. m,p-xylene 5. o-xylene.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 BTEX Carrier: Helium Oven : 40 C for 3 min, 5 /min to 100 C

High SPEED Megabore Same Resolution - Faster Analysis! Increasing Sample Throughput With High-Speed Megabore Application note 5988-5271EN

Column Diameter and Capacity I.D. (mm) Capacity (ng) 0.05 1-2 0.10 6-13 0.18 25-55 Like Polarity Phase/Solute 0.25 µm film thickness 0.20 35-70 0.25 80-160 0.32 110-220 0.45 600-800 0.53 1000-2000 Slide 23

Column Diameter - Inlet Head Pressures (Helium) I.D (mm) Pressure (psig) 0.05 275-400 0.10 90-130 0.18 30-45 0.20 25-40 0.25 15-25 30 meters Hydrogen pressures x 1/2 0.32 10-20 0.45 3-7 0.53 2-4 Slide 24

Diameter Summary If you decrease the inside diameter: Efficiency Resolution Pressure Capacity Flow rate Increase Increase Increase Decrease Decrease Slide 25

Resolution N k α 1 R = s 4 k + 1 α Efficiency Retention Selectivity N = ƒ (gas, L, r c ) k = ƒ (T, d f, r c ) α = ƒ (T, phase) L = Length r c = column radius d f = film thickness T = temperature Slide 26

Film Thickness and Retention: Isothermal Thickness (µm) Retention Change 0.10 0.40 0.25 1.00 1.0 4.00 3.0 12.0 5.0 20.0 Constant Diameter Normalized to 0.25 µm Slide 27

Film Thickness and Resolution When solute k < 5 d f R (early eluters) or T When solute k > 5 (later eluters) d f or T R Slide 28

Other Retention - Adsorption Analysis of Noble & Fixed Gases Using HP PLOT MoleSieve 2 4 5 Column: Carrier: Oven: Sample: HP-PLOT/MoleSieve 30 m x 0.53 mm x 50 m HP part no. 19095P-MS0 Helium, 4 ml/min 35 C(3min) to 120 C (5 min) at 25 C/min 250 l, split (ratio 50:1) 6 1 3 1.Neon 2. Argon 3. Oxygen 4. Nitrogen 5. Krypton 6. Xenon 2 4 6 8 Time (min) Slide 29

Film Thickness and Capacity Thickness (µm) Capacity (ng) 0.10 50-100 0.25 125-250 0.50 250-300 1 500-1000 3 1500-3000 0.32 mm I.D. Like Polarity Phase/Solute 5 2500-5000 Slide 30

Film Thickness and Bleed More stationary phase = More degradation products Slide 31

Film Thickness and Inertness 0.25 1.0 3.0 active inactive active inactive active inactive Slide 32

Film Thickness Summary If you increase the film thickness: Retention Increase Resolution (k<5) Increase Resolution (k>5) Decrease Capacity Increase Bleed Increase Inertness Increase Efficiency Decrease Slide 33

GC Column Dimensions Examples. Slide 34

Method 8270 30m x 0.25mm ID, 0.50 μm 25 min run time Slide 35

Fast 8270 Semivolatile Analysis 12.5m X 100 μm ID HP-5ms column 7.5 min run time But.is this practical? See Agilent application note #5989-0207EN for more details. Slide 36

Running Samples on 100 μm ID Columns Practical? Environmental type samples = high contaminant residue potential. Smaller ID columns have reduced capacity for matrix contaminants due to lower surface area, shorter length (less forgiving). Less surface area a 10 m x.10 mm column has 7.5X less overall surface area than a 30 m x 0.25 mm ID column. A more robust solution might be to switch to a 20 m x 0.18 mm ID column (nice middle-ground between 0.25 and 0.10). These columns could be used in splitless or in split mode, whereas the 0.10 mm ID columns are practically limited to split introduction. Slide 37

Fast 8270 Semivolatile Analysis 20m x 0.18mm ID x 0.36um, DB-5.625 13.5 min run time Flow program or faster temperature program? Slide 38

What this method optimization means to your lab Before Optimization GC/MS 1 GC/MS 2 GC/MS 3 19 Spls/day 19 Spls/day 19 Spls/day After Optimization GC/MS 1 GC/MS 2 GC/MS 3 29 Spls/day 28 Spls/day Idle Slide 39

Regular Unleaded Gasoline California Phase I 23 4 5 6 7 8 9 10 12 Normal 18 19 17 21 25 26 28 Column: DB-PETRO 100 100m x 0.25 mm I.D., 0.5 µm Carrier: H 2, 24 psig, 31 cm/s Oven: 35 C// 9.5 min// 13.3 /min// 45 // 11 min// 1.4 /min// 60 // 11min// 2.7 /min// 220 // 3.6 min Injector: Split 1:200, 0.2 µl Detector: FID @ 300 C 11 13 14 15 16 20 22 23 24 27 29 30 31 32 34 33 35 36 37 38 39 0 20 40 60 80 Time (min.) 23 4 5 6 7 8 9 10 11 12 13 16 17 18 19 High Speed 14 24 15 20 22 23 21 Column: DB-1 40 m x 0.10 mm I.D., 0.20 µm Carrier: H 2, 78 psig, 34.8 cm/s Oven: 35 C// 3.6 min // 36.1 /min// 45 C// 4.15 min // 3.91 /min// 60 C// 4.15 min//6.9 / min// 220 C// 1.38 min Injector: Split 1:400, 250 C, 0.2 µl Detector: FID @ 300 C 0 10 20 30 Time (min.) 25 26 27 28 29 30 31 32 34 35 33 36 37 38 39

OK, Test Time Fusel Oil Simple Standard 1 2 DB-624 25 m x.53 mm I.D. x 3.0 μm Inlet: FID: Carrier: Oven: 250 o C, split 300 o C H 2, 50 cm/sec 40 o C for 5 min. 10 o C/min to 250 o C 1. acetaldehyde 2. methanol 3. 3-methyl-butanol (isoamyl alcohol) 4. 2-methyl-butanol (active amyl alcohol) How would you try to get better R for this? k, α, N? 3 4 C1099 0 2 4 6 8 10 Time (min) Slide 41

Need Plates? Length AND Column Diameter Column Dimensions Theoretical Plates 25 m x.53 mm x 3.0 μm 34,530 60 m x.25 mm x 1.4 μm 181,860 Slide 42

Fusel Oil Standard DB-624 60 m x.25 mm I.D. x 1.4 μm Inlet: FID: Carrier: Oven: 250 o C, split 300 o C H 2, 50 cm/sec 40 o C for 5 min. 10 o C/min to 250 o C 1 2 3 4 5 1. acetaldehyde 2. methanol 3. ethanol 4. acetone 5. 1-propanol 6. ethyl acetate 7. isobutanol 8. 1-butanol 9. 3-pentanol (IS) 10. 3-methyl-butanol (isoamyl alcohol) 11. 2-methyl-butanol (active amyl alcohol) 12. hexanol 13. phenylethanol Rs = 1.8 7 8 9 11 6 10 12 13 C1100 0 5 10 15 20 Time (min) Slide 43

Method Translation Software Input Screen Available at www.chem.agilent.com/en-us/support/downloads/pages/default.aspx or use the key words GC Method Translator in the quick search box at the Agilent.com website. Slide 44

CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention time Slide 45

RESOLUTION VS. LINEAR VELOCITY Helium Resolution of 1.5 = baseline resolution 4.50 3.84 3.36 R = 1.46 R = 1.31 R = 0.97 30 cm/sec 35 cm/sec 40 cm/sec 4.4 psig 5.1 psig 5.8 psig DB-1, 15 m x 0.32 mm ID, 0.25 um 60 C isothermal 1,3- and 1,4-Dichlorobenzene Slide 46

VAN DEEMTER CURVE OPGV = Optimum Practical Gas Velocity 1.00 H 0.75 0.50 0.25 u opt OPGV 10 20 30 40 50 60 u (cm/sec) Slide 47

μ opt and OPGV μ opt : Maximum efficiency OPGV: Optimal practical gas velocity Maximum efficiency per unit time 1.5-2x u opt Slide 48

COMMON CARRIER GASES Nitrogen Helium Hydrogen Slide 49

VAN DEEMTER CURVES 1.00 N 2 H 0.75 0.50 Small 0.25 Large He H 2 10 20 30 40 50 60 u (cm/sec) Page 50 Group/Presentation Title Agilent Restricted Month ##, 200X

CARRIER GAS Helium vs. Hydrogen Helium (35 cm/sec) Hydrogen (73 cm/sec) 1 3 2 1 4 7 8 2 3 6 5 9 4 6 7 8 5 9 0 2 4 6 8 10 12 Time (min.) 0 2 4 6 8 Time (min.) DB-1, 15 m x 0.25 mm i.d., 0.25 µm 50 C for 2 min, 50-110 C at 20 /min 10.5 min 7.8 min Slide 51

CARRIER GAS Gas Advantages Disadvantages Nitrogen Cheap, Readily available Long run times Helium Good compromise, Safe Expensive Hydrogen Shorter run times, Cheap Explosive* *Hydrogen is difficult to explode under GC conditions Slide 52

Conclusions Make sure what you are doing make sense Try not to make a big change in diameter; take small steps A good place to start is to switch from 0.53 to 0.45 mm id, or 0.32 to 0.25 mm Remember that with a decrease in diameter you will also have a decrease in capacity and flow Avoid 0.100 mm id columns unless you are proficient in 0.18 mm id Use Method Translator Utilize Technical Support! Slide 53

Thank you! Agilent technical support can be reached at: 1-800-227-9770.4.1 Questions? Slide 54

Wrap-up E-Seminar Questions Thank you for attending Agilent e-seminars. Our e-seminar schedule is expanding every week. Please check our website frequently at: www.agilent.com/chem/education Or register for Stay current with e-notes to receive regular updates Slide 55

Upcoming GC e-seminars Techniques for Making Your GC Analysis More Repeatable and Robust June 20, 2008 1:00 p.m. EDT Techniques, Tips and Tricks of Troubleshooting Capillary GC Systems July 15, 2008 2:00 p.m. EDT