Study of the accuracy of ultrasonic flowmeters for liquid

Similar documents
ULTRASONIC FLOW MEASUREMENT WITH INTEGRATED TEMPERATURE MEASUREMENT COMPENSATION

ANALYSIS OF FREQUENCY CHARACTERISTICS ON NON-INVASIVE ULTRASONIC-DOPPLER FLOW MEASUREMENT FOR METAL PIPES

Comparison between Numerical and Experimental for UVP Measurement in Double Bent Pipe with Out-of-Plane Angle

General Principles of LEFM Time-of-Flight Ultrasonic Flow Measurements

EFFECTS OF ACOUSTIC SCATTERING AT ROUGH SURFACES ON THE

Abstract. Keywords: Non-destructive testing, ultrasonic flow meter, matched filter, time-of-flight, CFD analysis. 1. Introduction

Today s menu. Last lecture. Ultrasonic measurement systems. What is Ultrasound (cont d...)? What is ultrasound?

CHAPTER 4 BASICS OF ULTRASONIC MEASUREMENT AND ANFIS MODELLING

Application of an ultrasonic velocity profile monitor in a hydraulic laboratory

Practical Results of Ultrasonic Imaging by Inverse Wave Field Extrapolation

ACOUSTIC TRANSMISSION WITH MODE CONVERSION PHENOMENON

The Analysis of Internal Flow Field in Oil-Gas Separator

The Fluid Flow in the T-Junction. The Comparison of the Numerical Modeling and Piv Measurement

ADVANCED HYBRID TYPE ULTRASONIC FLOW METER UTILIZING STATE-OF-THE-ART PULSED-DOPPLER METHOD ALONG WITH TRADITIONAL TRANSIT TIME METHOD

EFFECT OF FLOW DISTURBANCE ON MULTI-PATH ULTRASONIC FLOWMETERS

An Essential Requirement in CV Based Industrial Appliances.

Ultrasonic Measurement of Velocity Profile on Bubbly Flow Using a Single Resonant Frequency

Research on Permanent Magnet Linear Synchronous Motor Control System Simulation *

Technical specification FLUXUS F502TE

SITRANS F flowmeters. SITRANS F US System information SITRANS F US Ultrasonic flowmeters 4/181

Orifice Plate. Stainless Steel. Flow Data. Herz Valves UK Ltd.

On the study of elastic wave scattering and Rayleigh wave velocity measurement of concrete with steel bar

Finite Element Modeling of Ultrasonic Transducers for Polymer Characterization

INDUSTRIAL APPLICATION EXPERIENCES OF NEW TYPE FLOW- METERING SYSTEM BASED ON ULTRASONIC-DOPPLER FLOW VELOCITY-PROFILE MEASUREMENT

Flow-Induced Vibration of Pipeline on Elastic Support

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

Analysis of Ultrasound Propagation in High- Temperature Nuclear Reactor Feedwater to Investigate a Clamp-on Ultrasonic Pulse Doppler

Negative refraction of ultrasonic waves in 2D phononic crystals.

Today s menu. Last lecture. Measurement of volume flow rate. Measurement of volume flow rate (cont d...) Differential pressure flow meters

Available online at ScienceDirect. Procedia Engineering 84 (2014 )

Available online at ScienceDirect. Procedia Engineering 157 (2016 )

ULTRASONIC NDE RHEALOGICAL MEASUREMENT TOOLS FOR INDUSTRIAL PROCESS CONTROL

Available online at ScienceDirect. Procedia Engineering 90 (2014 )

MODELLING AND MEASUREMENT OF BACKSCATTERING FROM PARTIALLY WATER-FILLED CYLINDRICAL SHELLS

Laser induced release of encapsulated noble gas in SCHOTT receiver

Numerical study on scanning radiation acoustic field in formations generated from a borehole

DESIGN AND PERFORMANCE OF THE CONVERGING-DIVERGING VORTEX FLOWMETER

Sławomir Grzelak, Jarosław Czoków, Marcin Kowalski, Marek Zieliński

Phased Array Inspection at Elevated Temperatures

Permanently installed and non-invasive ultrasonic flowmeter for the measurement of thermal energy and volumetric flow rate

Monitoring of Pipe-Wall Thickness and its Thinning Rate by Ultrasonic Technique

Measurements of toluene water dispersions hold-up using a non-invasive ultrasonic technique

Signal Loss. A1 A L[Neper] = ln or L[dB] = 20log 1. Proportional loss of signal amplitude with increasing propagation distance: = α d

(Total 1 mark) IB Questionbank Physics 1

ALASTAIR MCLACHLAN Applications Cameron

Design of a High Frequency Equal Width Annular Array Transducer for Medical Imaging

Basic principles of the seismic method

Characterization of high temperature solar thermal selective absorber coatings at operation temperature

Crack detection in cantilever beam by frequency based method

Numerical Simulation of H 2 O/LiBr Falling Film Absorption Process

Principle and application of ultrasonic wave

Auxiliary Fluid Flowmeter

Ultrasonic Method for Measuring Frictional Temperature Rise in Rubbing Surface

By-Pass. This voltage is proportional to the liquid level (threewire potentiometer circuit). The resistance reading can

Shear waves in solid-state materials

International Communications in Heat and Mass Transfer

Ultrasonic Non-destructive Testing and in Situ Regulation of Residual Stress

Axial Higher Order Modes Cluster (A-HOMC) Guided Wave for Pipe Inspection.

SONAR-BASED VOLUMETRIC FLOW METER FOR CHEMICAL AND PETROCHEMICAL APPLICATIONS

Ultrasonic Flow Metering Errors due to Pulsating Flow

The Relationship between the Applied Torque and Stresses in Post-Tension Structures

Lab Section Date. ME4751 Air Flow Rate Measurement

Homework Book. Wave Properties. Huijia Physics Homework Book 1 Semester 2. Name: Homeroom: Physics Class:

Nondestructive Determination of Elastic Constants of Thin Plates Based on PVDF Focusing Ultrasound Transducers and Lamb Wave Measurements

A method of the forward problem for magneto-acousto-electrical tomography

Sensitivity Estimation of Permanent Magnet Flowmeter

Available online at ScienceDirect. Procedia Engineering 174 (2017 )

UNCERTAINTIES PROPAGATION STUDY OF THE TRANSIT TIME METHODS IN ULTRASONIC ANEMOMETERS.

1817. Research of sound absorption characteristics for the periodically porous structure and its application in automobile

VersaFlow Ultrasonic clamp-on Flowmeter Specifications

Mechanical thermal expansion correction design for an ultrasonic flow meter

Calculation of Earth s Dynamic Ellipticity from GOCE Orbit Simulation Data

Capacitive Pick-Up Type DB 040

Thermal conversion of solar radiation. c =

Temperature Waves in SRF Research

Ultrasonic Testing Classroom Training Book, Second Edition

Installation effects on an ultrasonic flow meter with implications for self diagnostics

Available online at Procedia Engineering 200 (2010) (2009)

Ultrasound propagation in concentrated suspensions: shear-mediated contributions to multiple scattering

Defect Location Analysis of Tank Bottom Based on Acoustic Emission with Different Location Algorithms

NDT 2010 Conference Topics

1917. Numerical simulation and experimental research of flow-induced noise for centrifugal pumps

Optimisation using measured Green s function for improving spatial coherence in acoustic measurements

Particles Removal from a Moving Tube by Blowing Systems: A CFD Analysis

A New Calibration Method for Ultrasonic Clamp-on Flowmeters

Available online at ScienceDirect. Energy Procedia 63 (2014 ) GHGT-12

Cubic Splines MATH 375. J. Robert Buchanan. Fall Department of Mathematics. J. Robert Buchanan Cubic Splines

Experimental study of CO 2 releases from a saturated liquid reservoir

Investigation of Flow Profile in Open Channels using CFD

Available online at ScienceDirect. Procedia Engineering 79 (2014 ) 49 54

Research on sound absorbing mechanism and the preparation of new backing material for ultrasound transducers

Detached Eddy Simulation on Hypersonic Base Flow Structure of Reentry-F Vehicle

General Physics (PHY 2130)

Open Access Experimental Research and Analysis of Vortex Excited Vibration Suppression of Spiral Stripe Strake

Optimization and Numerical Simulation of Multi-layer Microchannel Heat Sink

Numerical simulations of the edge tone

FROM NEAR FIELD TO FAR FIELD AND BEYOND

Generation and Analyses of Guided Waves in Planar Structures

Sound-based Sensing. Some of these notes have been adapted from Carstens J.R. Electrical Sensors and Transducers

General Physics (PHY 2130)

Transcription:

Available online at www.sciencedirect.com AASRI Procedia 3 (2012 ) 14 20 2012 AASRI Conference on Modelling, Identification and Control Study of the accuracy of ultrasonic flowmeters for liquid Liling Ma a, Jia Liu a, Junzheng Wang a * a School of Automation, Beijing Institute of Technology, Beijing 100081,China Abstract How to estimate the differential time of flight accurately is a challenging problem in the area of ultrasonic flow measurement. High time resolution is necessary, especially in a small inner diameter, short sound path measurement environment, since sound velocity in the liquid is quite fast and the order of magnitude is nanosecond. A newer method to obtain the differential time of flight is a Spline-Based algorithm. Although the calculation accuracy of a Spline-Based algorithm is higher than the common method, cross-correlation algorithm, the calculated quantity of this algorithm is huge. For this reason, an improved Spline-based algorithm is proposed to reduce the computational cost while keeping the accuracy. In addition, installation location of transducers is one of the most important factors affecting the measurement accuracy of ultrasonic flowmeter. This paper also analyzes the affection of the differential separation distance between transducers. Experiments have shown that the improved algorithm is effective and the separation distance affects the result seriously. 2012 2012 The Published Authors. Published by Elsevier by Elsevier B.V. Selection Ltd. and/or peer review under responsibility of American Applied Selection Science and/or Research peer Institute review under responsibility of American Applied Science Research Institute Keywords: Ultrasonic, Time of flight, Spline, Distance; 1. Introduction Ultrasonic flowmeters (USFM) have been rapidly developed in recent years. Due to the non-destructive testing and facilitation of installation and removal, USFMs became more and more important in fluid flow *Jia Liu. Tel.: +86-13601238491; fax: +86-010-68940761-210. E-mail address: Bryan_lj@163.com. 2212-6716 2012 The Authors. Published by Elsevier Ltd. Selection and/or peer review under responsibility of American Applied Science Research Institute doi:10.1016/j.aasri.2012.11.004

Liling Ma et al. / AASRI Procedia 3 ( 2012 ) 14 20 15 measurement [1]. These meters can be used as a spool-piece meter or as a clamp-on meter. But the transit-time clamp-on USFMs are not very robust. Many papers aim to improve the accuracy of the flowmeter. The fluid velocity distribution in the pipe can be obtained using the computational fluid dynamic (CFD) modeling approach [2]. The relationships between correction factor K and Reynolds number for different ultrasound paths can be obtained through numerical calculation [3]. The accuracy of measuring flow rates changes with the incidence angles [4] and upstream installations [5]. Especially, absolute propagation time (the upstream and downstream transit times) always contains the delays caused by emitting and receiving circuits and the transit time when the signal spreads in the wall and acoustic wedge. Thus this paper attempts to address this problem. One of the most important parameters is the differential of transit time. We can get it by several ways. The first one is the cross-correlation algorithm. The peak-searching algorithm identifies the peak of the discrete function of the cross-correlation using Cosine interpolation algorithm [6], Gaussian interpolation algorithm [7] or Parabolic interpolation algorithm. The second one is the sample tracking algorithm. In this algorithm, the time shift of each sample in a delayed echo signal is measured with respect to a continuous, interpolated representation of the reference echo signal [8]. The third one is Spline-Based algorithm. The algorithm uses cubic splines to produce a continuous time representation of a reference signal, and then computes an analytical matching function between this reference and a delayed signal. This method is about 500 times better than normalized cross-correlation and entails a substantial computational cost [9]. The proposed method improves the Spline-Based algorithm by combining with the cross-correlation algorithm. The improved algorithm reduces the computational cost without reducing the accuracy. This proposed method is evaluated in experiments. In addition, an important part of the installation process is the transducer separation distance previously described. If the signal received by a transducer when set up at the wrong position, the calculation result would be definitely wrong. This paper analyzes the affection of the differential separation distance between transducers to show the importance to determine the correct position. Experiments have shown that the separation distance affects the result seriously. 2. Proposed Method The flow velocity is calculated by Eq.(1) v 2D t sin 2 ( t t ) ( t t ) up delay dn delay (1) t dn and t up in (1) represent the downstream and upstream transit times. t denotes the differential time of Fig.1.Ultrasonic transducers are mounted in accordance with V-shape.

16 Liling Ma et al. / AASRI Procedia 3 ( 2012 ) 14 20 flight. t delay is the transit time that signal takes when it is not in the liquid. D is the diameter of the steel pipes. is the refraction angle when the signal passes through the steel-water interface. Ultrasonic transducers are mounted in accordance with the V-shape shown in Fig.1. Transducer X emits ultrasonic waves, and transducer Y receive them. On the contrary, Transducer Y emits ultrasonic waves, and transducer X receive them. 2.1. Improved Spline-Based Algorithm The differential time of flight, t, is a key parameter, and its order of magnitude is nanoseconds. Since sound velocity in the water is quite fast, so high time resolution is necessary, especially in a small inner diameter, short sound path measurement environment. Typically, the cross-correlation algorithm is the most widely used algorithm. And this method is good at restraining white noise. But simulation results show that the calculation accuracy of the Spline-Based algorithm is higher than the cross-correlation algorithm. Assume S 1 and S 2 are discrete signals of reference signal and delayed signal (continuous time signal), and each group signal has 2048 points. Extract the kernel signal from S 1 and S 2 starts from the same point. The kernel signal is the effective part of the ultrasonic. Assume S 1 [n] and S 2 [n] represent the kernel signal. The length of S 1 [n] is N and The length of S 2 [n] is M, where M<N. As is shown in Fig.2, extracting kernel signal is depending on experience. If M and N are too big, the kernel signal will include useless points. In contrast, if M and N are too small, the useful points will be missed. Fig.2. Kernel signal extraction

Liling Ma et al. / AASRI Procedia 3 ( 2012 ) 14 20 17 Signal S 1 [n] is processed by cubic splines. The spline interpolant is piecewise continuous over an interval [0 1], is sampling period, as Eq.(2). Signal S 2 [n] is still discrete signal. f ( t) a ( t/ T) b ( t/ T) c ( t/ T) d.0 t 1 3 2 1 1 1 1 1 1 1 1 S () t f ( t) a ( t/ T) b ( t/ T) c ( t/ T) d...( i 1) t i 3 2 1 i i i i i 1 i i f ( t) a ( t/ T ) b ( t/ T ) c ( t/ T ) d...( N 2) t ( N 1) 3 2 N 1 N 1 N 1 N 1 N 1 N 1 N 1 N 1 (2) f i (t) is the i-th piecewise function. There are N-1 segments in total.. If T i =i-1 and t=t/ -T i, the Eq.(2) simplifies to Eq.(3) f ( t ) a ( t ) b ( t ) c ( t ) d 0 t 1 ' ' 3 ' 2 ' ' 1 1 1 1 1 f t a t b t c t d t ' ' 3 ' 2 ' ' i( ) i ( ) i ( ) i ( ) i.0 1 f ( t ) a ( t ) b ( t ) c ( t ) d 0 t 1 ' ' 3 ' 2 ' ' N 1 N 1 N 1 N 1 N 1 (3) Eq.(3) shows abscissa of each segment starts from 0. The value of t obtained by Eq.(4) that minimizes the (t ) is called local delay estimate here. M ' ' 2 i 2 i 1 ( t ) ( f ( t ) S [ i ]) (4) The subscript of f i (t ) is different from Eq.(3). Signal S 2 [n] has only M points, so f i (t ) is the i th segment of M segments which involved in computation. First time, k=0, points of signal S 2 [n] compute with the first to the M th segment of S 1 [t]. Second time, k=1, points of signal S 2 [n] compute with the second to the M+1 th segment of S 1 [t], and so on. Finally, there are (N-M) t-values. And there must be a t lying between 0 and 1, which called global delay estimate. Differential time of flight t is calculated by Eq.(5). ' t ( t k ) (5) The calculated quantity of the Spline-Based algorithm is huge. The proposed method can reduce the running time of Spline-Based algorithm while keeping the accuracy. Before calculating Eq.(4), we can find the maximum point,m i, of the cross-correlation function of discrete signals S 1 and S 2 by Fast Fourier Transform. If m i-1 <m i+1, then k=i; Else if m i-1 >m i+1, k=i-1. k is the parameter we need in Eq.(5). Since k is a known parameter, we can get t by computing only once. Points of signal S 2 [n] compute with the k+1 to the k+m th segment of S 1 [t] by Eq.(4). It is clear that N-M-1 times calculate quantities are removed. And the speed of algorithm will be increased. 2.2. Initial position of transducers Installation location of transducers is one of the most important factors affecting the measurement accuracy

18 Liling Ma et al. / AASRI Procedia 3 ( 2012 ) 14 20 of ultrasonic flowmeter. When the wall thickness, the wall material, and the fluid medium are all known quantities and the acoustic path is determined in accordance with the V-shape, the separate distance between each transducer should be definite. Assume transducer acoustic wedge is organic glass, and the material of the wall is steel. The fluid medium is water and incidence angle of ultrasonic wave is 37 degree. The Snell law is given in Eq.(6) and is used to calculate the refraction angle in different medium. sin / C1 sin / C2 sin / C 3 (6) is the incident angle when signal passes through the organic glass-steel interface. is the refraction angle when signal passes through the organic glass-steel interface or the incident angle when signal passes through the steel-water interface. is the refraction angle when signal passes through the steel-water interface. C1 C2 and C3 are sound velocities in organic glass, steel wall and water respectively. =37 o, and longitudinal wave velocity in organic glass is 2730m/s. If shear wave velocity in steel wall is 3230m/s, we can get =45 o. Longitudinal wave velocity in water is 1480m/s at 20 o C, then =19 o.the differential distance between transduser X and transducer Y is calculated based these angles. L 2 d tan 2 d tan AB t (7) In Eq.(7), d t is wall thickness, d is inner diameter. At the same time, the value of L AB in turn affect the value of in Eq(1). Assume the position of transducer Y shifts to the right axially 2mm from the correct position. This will cause to refraction angle,, 3 degrees bigger than the true value. As a result, the calculation error of water velocity will be up to 11% based on Eq.(1). Moreover, this impact is more pronounced to small diameter measurement environment. So it is necessary to fine tune the transducer location. 3. Experiment results The experiments are carried out using the clamp-on meter mounted on a stainless-steel pipe which outer diameter is 23 mm and wall thickness is 3 mm. The fluid medium is water. Control chip is FPGA of EP2C8Q208. 3.1. Calculating the differential time of flight by proposed method. Assume each group signal has 2048 points. The kernel signal, S 1 [n], is the 1000 th to the 1800 th points of reference signal, N=801; The kernel signal, S 2 [n], is the 1200 th to the 1700 th points of reference signal, M=501. Compared with improved algorithm, the Spline-Based algorithm will compute M-N=300 times more.

Liling Ma et al. / AASRI Procedia 3 ( 2012 ) 14 20 19 Fig.3. Time consuming of Spline-Based algorithm and improved algorithm. In Fig.3, SSE represents the Spline-Based algorithm. Time consuming of Spline-Based algorithm is about 130ms. Time consuming of improved algorithm is approximately 8ms. Obviously, the improved method can reduce the running time while keeping the accuracy. 3.2. Effect of axial movement of transducers Experiment results are shown in Fig.4. The water velocity is calculated by Eq.(1). Moving the transducer B axially to the right from the correct position will lead to increasing of, then water velocity will decrease. The velocity is in an inverse ratio of. Despite other unknown factors, we only consider the error of separate distance axially. Assume zero point is the correct position. Fig.4. The relationship between water velocity and refraction angle when the transducer position is changed.

20 Liling Ma et al. / AASRI Procedia 3 ( 2012 ) 14 20 4. Conclusion The improved Spline-Based method can reduce the running time effectually, and its stability is slightly better than the cross-correlation algorithm as is shown in the environment results. Furthermore, mounting position of transducers should be paid more attention since it could affect the accuracy of result seriously. Acknowledgement The authors wish to acknowledge the support of members of the faculty in our laboratory, the Beijing University of Technology for financial support, and would also like to thank the anonymous reviewers. Reference [1] L. C. Lynnworth, Y. Liu, Ultrasonic flowmeters: Half-century progress report, 1955 2005, Ultrasonics, vol. 44, pp.1371 1378, 2006. [2]DandanZheng, PengyongZhang, TianshiXu, Study of acoustic transducer protrusion and recess effects on ultrasonic flow meter measurement by numerical simulation, Flow Measurement and Instrumentation, vol.22, pp.488-493, 2011. [3]H. J. Dane, Ultrasonic measurement of unsteady gas flow, Flow Measurement and Instrumentation, vol. 8, pp. 183-190, 1997. [4]Yuto Inoue, Hiroshige Kikura, Hideki Murakawa, et al, A study of ultrasonic propagation for ultrasonic flow rate measurement, Flow Measurement and Instrumentation, vol.19, pp 223-232,2008. [5]C. Ruppel, F. Peters, Effects of upstream installations on the reading of an ultrasonic flowmeter, Flow Measurement and Instrumentation, vol.15, pp.167-177, 2004. [6] S. Foster, P. Embree, W. O Brien, Flow velocity profile via time-domain correlation: Error analysis and computer simulation, IEEE Transactions On Ultrasonics, Ferroelectrics, And Frequency Control, vol. 37, pp.164-175, 1990. [7] P. G. de Jong, T. Arts, A. P. Hoks, R. S. Reneman, Determina-tion of tissue motion velocity by correlation interpolation of pulsed ultrasonic echo signals, Ultrasonic. Imaging, vol. 12, pp. 84-98, 1990. [8] Reza Zahiri-Azar, Septimiu E. Salcudean, Time-Delay Estimation in Ultrasound Echo Signals Using Individual Sample Tracking. IEEE Transactions On Ultrasonics, Ferroelectrics, And Frequency Control, vol. 55, pp.2640-2650, 2008. [9] Francesco Viola, William F. Walker, Computationally Efficient Spline-Based Time Delay Estimation. IEEE Transactions On Ultrasonics, Ferroelectrics, And Frequency Control, vol. 55, pp.2084-2091, 2008.