AppNote 3/2012. Rapid Automated Screening of Extractable Compounds in Materials for Food Packaging, Medical or Technical Purposes KEYWORDS ABSTRACT

Similar documents
AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

AppNote 14/2012. Thermal Gravimetric Analysis/Mass Spectrometry Simulation using the GERSTEL Automated Pyrolyzer KEYWORDS ABSTRACT

AppNote 5/2008. Automation of Sample Preparation Steps using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

AppNote 8/2008. A New Purge Tool for Use with Automated Headspace Analysis KEYWORDS ABSTRACT

AppNote 4/2008. Automated Dynamic Headspace Sampling of Aqueous Samples Using Replaceable Adsorbent Traps KEYWORDS ABSTRACT

AppNote 3/2014. Analysis of Aroma Compounds in Edible Oils by Direct Thermal Desorption GC/MS using Slitted Microvials KEYWORDS ABSTRACT

TechNote 5/2002 KEYWORDS ABSTRACT. SBSE, Twister, Thermal Desorption

AppNote 1/2007. Automated Dynamic Headspace Sampling using Replaceable Sorbent Traps KEYWORDS ABSTRACT

AppNote 4/1998. New Developments in Multi-Dimensional Capillary Gas Chromatography KEYWORDS ABSTRACT

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances KEYWORDS

AppNote 15/2012. Large Volume Injection with Solvent Venting - Application to Trace Detection of Analytes in Water INTRODUCTION

Off-gassing of Rubber Particles Used for Athletic Fields using Automated Dynamic Headspace Sampling

AppNote 2/1996. Andreas Hoffmann Gerstel GmbH & Co.KG, Eberhard-Gerstel-Platz 1, D Mülheim an der Ruhr, Germany

Analysis of Food Samples using Thin Film Solid Phase Microextraction (TF-SPME) and Thermal Desorption GC/MS

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to En vi ron men tal Aqueous Samples INTRODUCTION

AromaOffi ce: Application of a Novel Linear Retention Indices Database to a Complex Hop Essential Oil

AppNote 2/2002. Discrimination of Soft Drinks using a Chemical Sensor and Principal Component Analysis KEYWORDS ABSTRACT

AppNote 6/2010. part a. Alternative Procedure for Extraction and Analysis of PAHs in Seafood by QuEChERS-SBSE-GC-MS KEYWORDS ABSTRACT

AppNote 3/2007 KEYWORDS INTRODUCTION. Beverage, Detector, Headspace, Quality Control

Thermal Desorption Unit TDU

AppNote 1/2004. Use of a Mass Spectral Based Chemical Sensor to Discriminate Food and Beverage Samples: Olive Oils and Wine as Examples KEYWORDS

AppNote 5/2002. Characterization of a Mass Spectrometer based Electronic Nose for Routine Quality Control Measurements of Flavors KEYWORDS ABSTRACT

AppNote 8/2002 KEYWORDS ABSTRACT. Electronic Nose, Fast GC, Headspace Analysis, Chemometrics,

AppNote 10/2012. Large Volume Injection with GC/MS Multi Column Switching for Direct Determination of Ethyl Carbamate in Alcoholic Beverages KEYWORDS

AppNote 8/2009. Fragrance Profi ling of Consumer Products using a Fully Automated Dynamic Headspace System KEYWORDS ABSTRACT

AppNote 8/2000. Volatile Organic Compounds from Adhesives and their Contribution to Indoor Air Problems KEY WORDS ABSTRACT

AppNote 2/2004. Analysis of Packaging Materials using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

AppNote 6/2014. Automated Online Desorption and Analysis of DNPH Derivatives of Airborne Aldehydes and Ketones KEYWORDS ABSTRACT

AppNote 2/2001. Stir Bar Sorptive Extraction: Enhancing Selectivity of the PDMS Phase KEYWORDS ABSTRACT

AppNote 7/2014. Automating Liquid-Liquid Extractions using a Bench-top Workstation KEYWORDS ABSTRACT. Sample Preparation, Lab Automation, LC/MS/MS

AppNote 8/2012. Rapid Automated Extraction and Confi rmation of Buprenorphine and Norbuprenorphine in Urine by DPX-LC/MS/MS KEYWORDS ABSTRACT

AppNote 4/2005. A Selectable Single or Multidimensional. Fraction Collection and Dual Detection for Trace Analysis of Complex Samples KEYWORDS

AppNote 2/2012. Determining Phenolic Compounds in Whisky using Direct Large Volume Injection and Stir Bar Sorptive Extraction KEYWORDS ABSTRACT

AppNote 6/2012. Flavor and Fragrance Analysis of Consumer Products - Dynamic Headspace Compared to Some Traditional Analysis Approaches KEYWORDS

AppNote 2/2013. A High Throughput Automated Sample Preparation and Analysis Workfl ow for Comprehensive Toxicology Urine Screenings using LC/MS/MS

AppNote 8/2011 KEYWORDS ABSTRACT INTRODUCTION

AppNote 1/2013. Determination of Barbiturates and 11-Nor-9-carboxy- 9 -THC in Urine using Automated Disposable Pipette Extraction (DPX) and LC/MS/MS

AppNote 1/2011 KEYWORDS ABSTRACT

AppNote 7/2002. Classifi cation of Food and Flavor Samples using a Chemical Sensor KEYWORDS ABSTRACT

AppNote 2/1998. Direct Injection of Distilled Spirits with PTV Matrix Removal: The Perfect Splitless Injection? KEYWORDS INTRODUCTION

AppNote 2/2007. Automated Solid Phase Microextraction using the GERSTEL MPS Prepstation and MAESTRO Software KEYWORDS ABSTRACT

AppNote 9/2012. Solid Phase Micro-Extraction (SPME) Coupled with Selectable 1 D/ 2 D GC-MS for the Determination of Food Product Flavors KEYWORDS

AppNote 5/2001. The Analysis of the Bitter and Other Flavour Compounds in Beer and Wort by Stir Bar Sorptive Extraction (SBSE) Followed by HPLC

AppNote 12/2012. Comparison of EG-Silicone-SBSE and Derivatization-PDMS-SBSE for the Analysis of Phenolic Compounds and Off-fl avors in Water KEYWORDS

GERSTEL MultiPurpose Sampler MPS. Versatile autosampler and sample preparation robot

KEYWORDS ABSTRACT AppNote 6/2002

AppNote 7/2005. Detection of Packaging Emissions using a Flexible Headspace Sampler Combined with a Multi Sensor System and a Separation Unit KEYWORDS

AppNote 12/2008 KEYWORDS ABSTRACT

AppNote 11/2012. Fully Automated Analysis of Doping Substances in Equine Urine by Disposable Pipette Extraction (DPX) coupled to GC/QqQ-MS KEYWORDS

AppNote 3/1998 KEYWORDS ABSTRACT. Beverage, Food & Flavor, PTV, Large Volume Injection, Headspace, Thermal Desorption.

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances

AppNote 2/2015 KEYWORDS ABSTRACT

AppNote 7/2013. Comprehensive Automation of the SPE- GC/MS Analysis of Opioids, Cocaine and Metabolites from Serum and Other Matrices KEYWORDS

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

Selective Sample Enrichment of Gaseous Samples using a Cooled Injection System: Trace Determination of Sulphur Components in Natural Gas

AppNote 2/1999. Automated Sample Fractionation and Analysis Using a Modular LC-GC System

AppNote 6/2010. part b. High Throughput Method for the Determination of PAHs in Seafood by QuEChERS-SBSE-GC-MS SCOPE PRINCIPLE

AppNote 2/1996. Andreas Hoffmann Gerstel GmbH & Co.KG, Aktienstrasse , D Mülheim an der Ruhr, Germany

AppNote 5/1994. The Use of a Multi Purpose Sampler for Headspace GC-MS Analysis of Volatile Organic Compounds in Human Urine

Determination of Formaldehyde and VOCs in Wood-based Products using an Automated Micro-Scale Chamber

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

AppNote 8/2001 KEYWORDS ABSTRACT. PAH, SPME, SBSE, GERSTEL Twister, Thermal desorption, GC-MS, PCB

Determination of volatiles in Food sample preparation strategies and techniques Dr Kathy Ridgway

Innovative Thermal Desorption From Agilent Technologies. Vivek Dhyani Application Chemist

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Determination of Volatile Substances Proof of Food Adulteration

Volatile Organic Compounds in Every Day Food

AppNote 3/2011. Using Three Types of Twister Phases for Stir Bar Sorptive Extraction of Whisky, Wine and Fruit Juice KEYWORDS ABSTRACT

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Understanding Gas Chromatography

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Dynamic headspace (DHS) technique: set-up and parameter control for GC/MS analysis of odorant formulations

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

ChromTech options for PAL systems

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010

anthracene Figure 1: Structures of Selected Polyaromatic Hydrocarbons (PAHs)

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Benzoylecgonine in Urine by SAMHSA GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

TETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Introduction to E&Ls 1

Application. Gas Chromatography February Introduction

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques

STANDARD OPERATING PROCEDURES

GERSTEL Thermal Desorption System TDS 3 The GERSTEL TDS 3 is a flexible multi-functional thermal desorption system for highly sensitive and accurate d

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Application. Gas Chromatography March 1998

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

HPLC Sample Preparation

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Transcription:

AppNote 3/2012 Rapid Automated Screening of Extractable Compounds in Materials for Food Packaging, Medical or Technical Purposes Oliver Lerch Gerstel GmbH & Co. KG, Eberhard-Gerstel-Platz 1, D-45473 Mülheim an der Ruhr, Germany KEYWORDS Extractables, Leachables, Food Packaging Materials, Medical Materials, Automated Sample Preparation, Liquid Injection, Thermal Desorption Unit (TDU), Gas Chromatography (GC) ABSTRACT This application note describes two automated methods for screening of extractable compounds from materials for food packaging, medical or technical purposes. The first method is based on automated liquid extraction performed by the GERSTEL MultiPurpose Sampler (MPS), the second involves thermal desorption of the material in question in the GERSTEL Thermal Desorption Unit (TDU). Both methods are suitable for gaining an overview of the quality and emission potential of a material and therefore useful in the search for a suitable packaging material. The methods deliver comparable qualitative results.

INTRODUCTION Methods to determine the purity of a material or, in other words, the potential of a material to emit unwanted compounds are of importance for many industries. For example: Packaging materials used in the pharmaceutical and food industries should not release harmful compounds or compounds which alter the product characteristics; materials used in medical or technical devices should not contaminate the processed media; equipment used for chemical analysis should not contaminate the sample and so on [1-4]. In packaging material analysis two terms play an important role: Leachables and extractables. The term leachables encompasses all compounds, which leach from a packaging material into the packaged product under normal storage or use conditions - as well as those formed in reactions between the packaging material and the product. The extractables found in a particular packaging material are those compounds which can be extracted from the material under extreme conditions, for example using solvent extraction. By determining the extractables in a material, the emission potential of the material can be characterized, which is useful for an initial differentiation between suitable and unsuitable materials. Most extractables can be determined by gas chromatography. This application note describes two automated methods for screening materials for extractable compounds. These easy and fast methods are currently used to check the quality of materials used in GERSTEL analytical instrumentation. These are in-house methods that do not conform to any norm or legislation. Method 1 employs a liquid extraction (LE) of the material with ethyl acetate at 45 C for 4 hrs in an agitator followed by a liquid injection into a GC/ MS. Method 2 relies on thermal desorption (TD) at temperatures between 100 and 200 C, depending on the material, followed by GC/MS analysis. EXPERIMENTAL Instrumentation. Analyses were performed using a 7890 gas chromatograph equipped with a 5975 Mass Selective Detector (Agilent Technologies). Method 1 (liquid extraction) was performed using a Multi Purpose Sampler (MPS) equipped with a heated agitator in combination with Cooled Injection System (CIS), PTV-type inlet. Method 2 (TD) was performed using a Multi Purpose Sampler (MPS) in combination with a Thermal Desorption Unit (TDU) and a Cooled Injection System (CIS), PTV-type inlet, all from GERSTEL (figure 1). AN/2012/03-2 Figure 1. GC/MS system used for the determination of extractables from materials. Materials. Samples were taken from household articles like plastic bags for food storage, plastic food wrapping film and polymer storage boxes as well as from the laboratory (SPE cartridges, disposable syringe needles etc.). The liquid extraction was performed in standard 1.5 ml vials (093640-046-00) fitted with magnetic silicone white/ptfe red caps (093640-091-00). The agitator was equipped with adapters for 1.5 ml vials (093631-002-00). Ethyl acetate and isopropanol p.a. quality were used. Thermal desorption tubes with a glass frit (013742-005-00) were used for direct thermal desorption analysis of materials in the TDU. Sample Preparation and Introduction. All materials were cut using scissors or scalpels that had been cleaned with isopropanol. Method 1 (liquid extraction, LE). Pieces of polymer film from bags or food wrap measuring approximately 1.5 x 1.5-3 cm and pieces of thicker plastic material from other samples measuring around 0.2-1 x 1-1.5 cm were cut, briefly rinsed with isopropanol to remove external contamination and placed in vials. 500 μl of ethyl acetate, the extraction solvent, to which 2.5 μg of d10-phenanthrene has been added as internal standard, was added to each vial. Ethyl acetate was chosen since it is a universal extraction solvent and since it is compatible with GC analysis. The internal standard was added in order to enable the comparison of results generated on different instruments and at different times. Blank samples were prepared using the same chemicals and equipment.

The remainder of the analysis was performed automatically by the analysis system. The MPS placed the prepared vials into the agitator in which the packaging materials were extracted for 4 hrs at 45 C and agitation speed 750 rpm. An aliquot of 1.5 μl of each sample was injected into the GC/MS. The GERSTEL MAESTRO software controls the system such that several samples can be extracted in parallel, with overlapping extraction and GC analysis for highest possible throughput and efficiency (figure 2). TDU. Following thermal desorption, analytes were cryogenically refocused in the CIS, which was subsequently heated and the analytes transferred to the GC column. In order to avoid system contamination, the maximum desorption temperature for each sample was first determined by stepwise heating of the sample until it would begin to melt. The maximum desorption temperature was then set to 20 C below the temperature at which the sample had started to melt. A blank chromatogram of each empty thermal desorption tube was recorded before using it. Figure 2. Runtime optimization through automatic multi-sample overlapping of extraction and chromatography under MAESTRO software control, illustrated in the form of the MAESTRO scheduler. The combination with liquid chromatography is possible by exchanging the extraction solvent following evaporation in the GERSTEL multi-position evaporation station (mvap) and injecting the resulting solution into LC system through the injection valve. In principle, it is possible to automate the addition of the extraction solvent using a 1 ml syringe mounted on a second autosampler tower. However, the syringe needle must penetrate the septum to add the solvent, which means the silicone layer could get into contact with the extraction solvent, increasing the danger of contaminating the sample with septum material i.e. silicone instead of only having the PTFE layer in contact with the extraction solvent. Method 2 (Thermal desorption, TD). Pieces of material measuring around 0.3 x 1-1.5 cm were cut, cleaned with isopropanol, briefly left exposed to the air in order to dry, and placed in conditioned thermal desorption tubes. The tubes were stored hermetically sealed on the MPS tray and automatically transported to the AN/2012/03-3 Figure 3. A sample extracted using method 1 (LE) inside a 1.5mL vial and a sample extracted using method 2 (TD) inside a TDU thermal desorption tube. Analysis conditions - Method 1 (LE). MPS (LE-Method): Agitator 45 C (4 h), 750 rpm Syringe 10 μl Inj. Vol. 1.5 μl TDU (TD-Method): Temperature 50 C; 250 C/min; 100-200 C (1.6 min) Pneumatics Splitless CIS (LE-Method): Temperature 40 C; 12 C/s; 280 C (20 min) Pneumatics Splitless Liner Baffled CIS (TD-Method): Temperature -120 C; 12 C/s; 280 C (20 min) Pneumatics Solvent Vent 30 ml/min, Split flow 20 ml/min @ 1 min Liner Packed with Glass beads GC: Temperature 50 C (1 min); 18 C/min; 325 C (8 min) Pneumatics 1.0 ml/min He, constant flow Column 30 m Rxi-5ms (Restek) d i = 0.25 mm, d f = 0.25 μm Detector: MSD EI mode, full scan, 35-500 amu

RESULTS AND DISCUSSION Comparison of the results obtained using the two methods. Blank chromatograms of both extraction methods were highly satisfactory, revealing no large peaks that could interfere with the analysis (figure 4). Figure 4. Blank chromatogram from method 1 (LE) and method 2 (TD). All peaks seen are small and do not interfere with the analysis. Extracting a material for 4 hrs at 45 C produces reliable results independent of the length of time the sample has been stored on the MPS tray before the extraction. This can be seen in figure 5 in which two chromatograms resulting from extractions of the same material are shown: One sample of the material was extracted directly after it had been prepared, the other sample was prepared at the same time, but not extracted until it had spent ten hours in the MPS tray. Figure 5. Extractables from a polymer lid extracted directly after adding ethyl acetate (upper chromatogram) and after 10 hrs in ethylacetate. The results are the same, proving that the extraction method (four hours at 45 C) is rugged. AN/2012/03-4

Figure 6 shows chromatograms resulting from extractions of a polymer lid of a plastic food storage container following method 1 (LE) and method 2 (TD). The same analytes, mainly branched n-alkanes, are extracted from the material by both methods giving the same chromatographic pattern. Under these conditions the thermal desorption method, although performed with less sample, is much more sensitive (by a factor of 10-60) than the liquid extraction. However, high boiling compounds are extracted more efficiently by LE than by TD. Both methods are well suited to give an impression of the emission potential and quality of a material. Figure 6. Extractables from the polymer lid of a plastic food storage box analyzed by method 1 (LE) and method 2 (TD). The same patterns are observed for both methods. The TD method is much more sensitive except when determining high boiling compounds. By increasing the split ratio and reducing the desorption temperature of the TD method it is possible to adjust the compound signal intensities and to get comparable chromatograms from the LE- and TD-based methods. This gives us two alternative analysis methods for a given material (figure 7). Figure 7. When the TD method parameters are adapted with lower desorption temperature and higher split ratio, the two methods give the same result for a given material and are therefore interchangeable except for the highest boiling compounds. AN/2011/03-5

Extractables found in Samples [5]. Around 70 samples were analyzed using method 1 (LE). Generally, extractables in polymer materials are mainly monomers, oligomers or breakdown products from the polymer. Also additives like plasticizers, UV-protectants and catalysts can be found. In this section some interesting findings from the examined materials are discussed. As shown above both methods can be employed with similar results. Only chromatograms resulting from liquid extraction (LE) are shown and discussed here. Most food packaging materials for single or multiple use are not very clean. They often contain hydrocarbons, organic acid esters, organic acid amides, phenolic compounds and long-chain organic acids. In some samples. toxic monomers like organic isocyanates, endocrine disrupting chemicals (EDCs), such as bisphenol A and phthalates as well as other unwanted compounds, were found. Extractables profile examples are seen in figures 8-17. Figure 8 shows an extractables profile from a film used for wrapping cheese. This profile is similar to those of many other olefine copolymers, figure 9 shows a profile from a waxed paper. Such material is frequently used in the food industry and emits large amounts of long chain alkanes. Figure 8. Extractables from a film used for wrapping cheese. Typical profile found for many olefine copolymers. Figure 9. Extractables from a waxed paper used for wrapping sausage cold cuts. AN/2011/03-6

Figure 10 shows a profile from a polymer film-based sausage packaging. Large amounts of compounds were found, including some that could affect the health of the consumer, such as dioctylphthalate and diphenyl methanediisocyanate. Tributylacetylcitrate, a plasticizer used as substitute for phthalates was also detected. Figure 10. Extractables from polymer film based packaging for sausages. Figure 11 shows the extractables profile of a microwave box. Among other compounds, drometrizole, a UV stabilizer, which, according to the FDA, may be used in food packaging material, and bisphenol A, a known EDC, were found. Figure 11. Extractables from a microwave box. AN/2011/03-7

The manufacturer of a polymer foam used as seal in mineral water bottle closures (figure 12) has apparently substituted phthalate plasticizers with more modern substances such as tributylacetylcitrate and 1,2- cyclohexanedicarboxylic acid diisononyl ester. The latter was present in significant amounts. These compounds are also found in toys. Figure 12. Extractables from a polymer foam used as seal in mineral water bottle closures. The toxic monomer toluene diisocyanate was found in a film-based closure used for packaging of fresh food (figure 13). Figure 13. Extractables from a film-based closure used for packaging fresh food. AN/2012/03-8

The lid of a microwave box, which was examined was of very high quality and emitted almost no compounds (figure 14). Unfortunately the only significant peak seen was dichlorobenzene, which may in fairness have been introduced as a contamination during transport since 1,4-dichlorobenzene is generally used as a moth repellant on clothes. Figure 14. Extractables from a lid of a microwave box. Only a few packaging materials were found to be clean, one is shown in figure 15. Figure 15. Extractables from polymer film-based packaging used for sausage cold cuts. A comparison between materials from a polypropylene lunch box and from a polypropylene SPE cartridge showed that the polymer of the lunch box emits significantly more compounds. The chromatogram resulting from extraction of the SPE cartridge material shows that it is possible to produce clean polypropylene with good characteristics at a reasonable price (figure 16). AN/2012/03-9

Figure 16. Extractable profiles for material from a polypropylene lunch box (black trace) and from a polypropylene SPE cartridge (red trace). The quality of materials for medical or laboratory devices is generally high. On the other hand, the extractables profile of an examined children s drinking bottle was found to be similar to, or even significantly worse than, that of a garbage bag we examined, emitting far more than the clean material used for medical devices (figure 17). This is astonishing and alarming since children are especially susceptible to contaminants and since most consumers use food packaging materials on a daily basis while they are rarely exposed to medical devices. It must be kept in mind that this study deals with extractables and not with leachables, which means it is a worst case simulation. Also, most of the extracted compounds are regarded as non-toxic but exposure to some of these may still result in chronic health effects. Moreover, fat containing foodstuffs can be expected to extract organic compounds from packaging material quite efficiently. Figure 17. Extractables profiles of from top: 1) Children s drinking bottle, 2) Garbage bag, 3) SPE cartridge and 4) Plastic part of a disposable syringe needle. The profile of the children s drinking bottle is more similar to the profile of the garbage bag than to those of the SPE cartridge or disposable syringe. AN/2012/03-10

Currently, aliphatic and aromatic hydrocarbons from mineral oil (MOSH/MOAH) contained in recycled cardboard and transferred to foodstuff through packaging are widely discussed [6]. Obviously this discussion needs to be expanded to address the quality of food packaging material in general. It is fair to say that packaging materials should be monitored more and just maybe new legislation is needed. CONCLUSIONS Two alternative methods for the qualitative determination of extractable compounds in materials were developed. The following was achieved: - Two automated, rapid and easy methods for screening of extractables in packaging materials were developed. - Comparable extractables profiles were obtained using the developed methods. - The thermal desorption method was shown to be far more sensitive, but it could be adjusted to give similar intensities as the liquid extraction for better comparability. - Liquid extraction was shown to be more sensitive for high boiling compounds. - The quality of solvents and the emission potential of the extraction container were shown to be key factors that must always be checked carefully prior to the determination of extractables in a material. REFERENCES [1] CHMP/CVMP. Guideline on Plastic Immediate Packaging Materials. European Medicines Evaluation Agency, London, UK, 1 December 2005 [2] Center for Drug Evaluation and Research. Guidance for Industry: Container Closure Systems for Packaging Human Drugs and Biologics - Chemistry, Manufacturing, and Controls Documentation. US Food and Drug Administration: Rockville, MD, May 1999 [3] 21 CFR Part 177 Indirect Food Additives: Polymers, US Food and Drug Administration, 1 April 2011 [4] E. Pfannkoch, J. Whitecavage: Stir Bar Sorptive Extraction from Food Simulating Solvents: Preliminary Studies, GERSTEL AppNote 3/ 2002 [5] www.wikipedia.org [6] R. Lorenzini, K. Fiselier, M. Biedermann, M. Barbanera, I. Braschi, K. Grob: Saturated and Aromatic Mineral Oil Hydrocarbons from Paperboard Foodpackaging: Estimation of Longterm Migration from Contents in the Paperboard and Data on Boxes from the Market, Food Additives & Contaminants: Part A, 27 (2010) 1765-1774 Regarding the analytical results of the extractables screening the following statements can safely be made: - Most of the examined food packaging materials showed a medium to high extractables profile. - Hydrocarbons, organic acid esters, organic acid amides, phenolic compounds and long chain organic acids were the most widely found extractables. - In some packaging materials, very critical compounds like isocyanates, phthalates, dichlorobenzene and bisphenol A were found. - Examined materials for use in medical or laboratory devices were found to be clean. It is, in other words, possible to produce clean, high quality polymers at a reasonable price. - Fat-containing foodstuffs are quite likely to be contaminated by compounds that are leached from the packaging material. AN/2012/03-11

GERSTEL GmbH & Co. KG Eberhard-Gerstel-Platz 1 45473 Mülheim an der Ruhr Germany +49 (0) 208-7 65 03-0 +49 (0) 208-7 65 03 33 gerstel@gerstel.com www.gerstel.com GERSTEL Worldwide GERSTEL, Inc. 701 Digital Drive, Suite J Linthicum, MD 21090 USA +1 (410) 247 5885 +1 (410) 247 5887 sales@gerstelus.com www.gerstelus.com GERSTEL AG Wassergrabe 27 CH-6210 Sursee Switzerland +41 (41) 9 21 97 23 gerstelag@ch.gerstel.com www.gerstel.ch GERSTEL K.K. 1-3-1 Nakane, Meguro-ku Tokyo 152-0031 SMBC Toritsudai Ekimae Bldg 4F Japan +81 3 5731 5321 +81 3 5731 5322 info@gerstel.co.jp www.gerstel.co.jp GERSTEL LLP Level 25, North Tower One Raffles Quay Singapore 048583 +65 6622 5486 +65 6622 5999 SEA@gerstel.com www.gerstel.com GERSTEL Brasil Av. Pascoal da Rocha Falcão, 367 04785-000 São Paulo - SP Brasil +55 (11)5665-8931 +55 (11)5666-9084 gerstel-brasil@gerstel.com www.gerstel.com.br Information, descriptions and specifications in this Publication are subject to change without notice. GERSTEL, GRAPHPACK and TWISTER are registered trademarks of GERSTEL GmbH & Co. KG. Copyright by GERSTEL GmbH & Co. KG Awarded for the active pursuit of environmental sustainability