TECHNICAL WORKING GROUP ITWG GUIDELINE ON IN-FIELD APPLICATIONS OF HIGH- RESOLUTION GAMMA SPECTROMETRY FOR ANALYSIS OF SPECIAL NUCLEAR MATERIAL

Similar documents
TECHNICAL WORKING GROUP ITWG GUIDELINE ON LABORATORY APPLICATIONS OF HIGH-RESOLUTION GAMMA SPECTROMETRY

Gamma Spectrum Generator for Nuclear Security

IPNDV Working Group 3: Technical Challenges and Solutions Nuclear Material (3) Technology Data Sheet

Recent improvements in on-site detection and identification of radioactive and nuclear material

Characterization of Large Structures & Components

Radioactive Waste Management

Establishing an ISO-CART Measurement Station to Meet Government Burial Regulations for Radioactive Material

NUCL 3000/5030 Laboratory 2 Fall 2013

Background Reduction Using Collimators on a Portable HPGe Nuclide Identifier

arxiv:nucl-ex/ v2 21 Jul 2005

Characterization Survey Techniques and Some Practical Feedback

CZT Gamma Spectrometry Applied to the In-situ Characterisation of Radioactive Contaminations

NUCLE A R FORENSIC S INTERN ATION A L TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON ELEMENTAL ASSAY U TITRATION

RADEAGLET. Lightweight Handheld Radioisotope Identification Device

PINS3-CF. Portable Chemical and Explosive Identification System

TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON SECONDARY ION MASS SPECTROMETRY (SIMS)

IdentiFINDER Digital Hand Held Spectrometer & Dose Rate Meter for Portable Applications

The PEAC-WMD Gamma Radiation Dose Calculator

ITWG - A Platform for International Cooperation in Nuclear Forensics

RADIOCHEMISTRY AND NUCLEAR CHEMISTRY Vol. II - Nuclear Forensics - Klaus Mayer and Maria Wallenius

Dose Reconstruction Methods and Source Term Assessment using Data from Monitoring Networks and Mobile Teams A German Approach

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

Semiconductor Detector

Identify Isotopes with the ICS-4000 Gamma Ray Spectrometer!

M E E T I n s t r u m e n t s G.m.b.H.

Radioactive Waste Characterization and Management Post-Assessment Answer Key Page 1 of 7

HALF LIFE. NJSP HMRU June 10, Student Handout CBRNE AWARENESS Module 4 1. Objectives. Student will

Department of Energy OffIce of Worker Protection Programs and Hazards Management Radiological Control Technical Position RCTP

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors

Mobile Unit for Site Characterization: Results of a Demonstration Event

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

WM 07 Conference, February 25 March 01, 2007, Tucson, AZ

A Proposal of Nuclear Materials Detection and Inspection Systems in Heavily Shielded Suspicious Objects by Non-destructive Manner.

Inaccuracy Determination in Mathematical Model of Labsocs Efficiency Calibration Program

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY

Nuclear Forensics Research and Development Prospective Directions in Azerbaijan.

ISO INTERNATIONAL STANDARD

A Primer on Gamma-ray Spectroscopy and Nuclear Forensics

Radioactivity measurements and risk assessments in soil samples at south and middle of Qatar

Seaborg s Plutonium?

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

Distinguishing fissions of 232 Th, 237 Np and 238 U with beta-delayed gamma rays

General Overview of Radiation Detection and Equipment

PINS3-CW. Portable Chemical Identification System

Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277

WM2018 Conference, March 18-22, 2018, Phoenix, Arizona, USA. PVT and LaBr3(Ce)-based Radon Express Analyzers 18164

Design and production of Scintillation Detectors

DISMANTLEMENT OF AN UNIT OF METABOLIC TREATMENTS WITH IODINE 131

Monitoring of Ionizing Radiations facilities Experience and challenge

Application Note. InSpector 1000-based CZT Package for Nuclear Power Plant Isotope Mix Analysis

Results of the 2015 national indoor radon intercomparison measurements in Serbia

Calibration & Use of a Capintec CAPTUS 3000 Portable Thyroid Uptake System for Iodine-125 Bioassay Measurements Todd W.

RADIATION SAFETY TRAINING SEALED SOURCES

Quality control of neutron-absorber materials for the nuclear fuel cycle, Principle of the JEN-3 neutron Backscattering gauge

Radionuclides in food and water. Dr. Ljudmila Benedik

RADEAGLE. Next Generation Radioisotope Identification Device. Detection and Identification: Fast, Accurate and Easy

CHEMISTRY 170. Radioisotopes

GUIDE TO LABORATORY SURVEYS. Introduction

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

INTERNATIONAL ATOMIC ENERGY AGENCY PROGRAM FOR DETECTION OF ILLICIT MATERIALS

CHEMISTRY 130 General Chemistry I. Radioisotopes

Assessment of Radiation Dose from Radioactive Waste in Bangladesh and Probable Impact on Health

An Improved U-Pu System for Field Pu Measurements

An Unmanned Aerial Vehicle-based Radiation Surveillance System. Seungwoo Lee Korea Electronics Technology Institute

ISO Radiological protection Sealed radioactive sources General requirements and classification

In-Situ Characterization of Underwater Radioactive Sludge

arxiv:nucl-ex/ v1 29 Aug 2005

Calibration of a Whole Body Counter and In Vivo measurements for Internal Dosimetry Evaluation in Chile, Two years experience.

Comparison of Several Detector Technologies for Measurement of Special Nuclear Materials i

Portable Radiation Survey Instruments

High Precision Nondestructive Assay to Complement DA. H.O. Menlove, M.T. Swinhoe, and J.B. Marlow Los Alamos National Laboratory

European Project Metrology for Radioactive Waste Management

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL

IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES

Survey Meter OD-01 Address: Phone: Fax: URL:

General Regression Neural Networks for Estimating Radiation Workers Internal Dose

Fundamentals of Radionuclide Metrology

Gamma background measurements in the Boulby Underground Laboratory

Radioactivity of the Treated Topaz

IAEA-CN-184/115. P.O. Box 2008, MS-6165 Oak Ridge, TN , USA

Gamma-Spectrum Generator

RADIOLOGICAL CHARACTERIZATION ACTIVITIES DURING THE PARTIAL DISMANTLING OF THE IRT SOFIA RESEARCH REACTOR FACILITIES

Radiological impact of rutile covered welding electrodes

TIIVISTELMÄRAPORTTI (SUMMARY REPORT)

BWXT Y-12 Y-12. A BWXT/Bechtel Enterprise SMALL, PORTABLE, LIGHTWEIGHT DT NEUTRON GENERATOR FOR USE WITH NMIS

KEYWORDS: Curie; radium; legacy; contamination; gamma spectrometry; radium-226; thorium- 232.

Unit 2. Instrumentation. Experts Teaching from Practical Experience

New methodology for source location and activity determination in preparation of repairing or decommissioning activities

Design of an Integrated Inspection System For Detection of Explosive and Illicit Materials

NGN PhD Studentship Proposal

Nucleonica Lab : An Overview of the Nuclear Science Portal with hands-on exercises

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING

High-Sensitivity Gamma-Ray Imaging With Double Sided Strip Detectors

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia

CHAPTER 25: NUCLEAR CHEMISTRY. Mrs. Brayfield

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Renewed whole-body counting chamber in STUK

Evaluation of Nondestructive Assay Characterization Methods for Pipe- Over-Pack Containers

Transcription:

NUCLE A R FORENSIC S INTERN ATION A L TECHNICAL WORKING GROUP ITWG GUIDELINE ON IN-FIELD APPLICATIONS OF HIGH- RESOLUTION GAMMA SPECTROMETRY FOR ANALYSIS OF SPECIAL NUCLEAR MATERIAL

This document was designed and printed at Lawrence Livermore National Laboratory in 2016 with the permission of the Nuclear Forensics International Technical Working Group (ITWG). ITWG Guidelines are intended as consensus-driven best-practices documents. These documents are general rather than prescriptive, and they are not intended to replace any specific laboratory operating procedures.

1. INTRODUCTION This document describes the use of gamma spectrometry in field conditions as a non-destructive tool for the characterization of nuclear material. A general description of gamma spectrometry and its use for nuclear security and nuclear forensics applications is given in the ITWG Guideline: High-resolution gamma spectrometry General overview [1]. The use of gamma spectrometry in the laboratory as a non-destructive tool for the characterization of nuclear material is described in the ITWG Guideline: Laboratory applications of highresolution gamma spectrometry for analysis of special nuclear material [2]. In-field use of gamma spectrometry for nuclear security and nuclear forensics purposes provides a means to assess the threat posed by interdicted radioactive/nuclear (RN) material. This information can then be used to determine how the material can best be transported to a safe storage location or to a nuclear forensic laboratory for detailed analysis. For the purpose of this document, it is assumed that, prior to the commencement of gammaspectrometric measurements: the RN material has been detected; the gamma and neutron dose rates are being monitored; contamination control measurements are in place; appropriate radiation protection measures have been undertaken; and appropriate measures have been put in place to secure the site. 2. USE FOR NUCLEAR FORENSICS For on-site applications, high-resolution gamma spectrometry is mostly used to quickly identify radioactive material and obtain a rough idea about the isotopic composition of nuclear material. From the point of view of on-site identification techniques and handling of the material, usually four main categories are considered: 1 Naturally Occurring Radioactive Material (NORM), 2 Medical isotopes, 3 Industrial isotopes, 4 Special nuclear material, as well as any mixture of these four. For the purposes of nuclear forensic investigations, IAEA Nuclear Security Series No. 2, Nuclear Forensic Support [3] defines the categories of nuclear and other radioactive material as follows: 1 Unirradiated direct use material. 2 Irradiated direct use material. 3 Alternative material. 4 Indirect use material. 5 Commercial radioactive sources. In some cases, specific radioisotopes can be identified using hand-held gamma spectrometers with built-in low-resolution or medium-resolution gamma detectors (based on NaI, LaBr 3, CdTe or CdZnTe crystals). These hand-held spectrometers typically employ algorithms to automatically identify radioactive material. They are simple to use and can be operated by first responders (e.g., customs and police officers, fire fighters) to detect, localize and perform initial identification of potentially illicit radioactive material. For more reliable identification, radiation experts, such as those of a Mobile Expert Support Team (MEST), can be called in to further categorize the material using a portable, high-resolution gamma spectrometer and dedicated gamma spectrum analysis software. For this purpose, high-purity germanium (HPGe) detectors are generally used, due to their good energy resolution in the energy range around 100 kev and acceptable counting efficiency at higher energies. These detectors can either be laboratory-based HPGe spectrometers with liquid-nitrogen cooling that are transported to an in-field location or more compact battery-operated HPGe detector systems that use electrical cooling and have built-in spectrum analysis algorithms. Since laboratorybased gamma spectrometers are usually portable as well, these may also be used in the field; however, there are inherent difficulties in operating a laboratory instrument in field conditions, including securing a source of liquid nitrogen, having a reliable power source, handling a fragile laboratory instrument, etc. The compact, battery- 1

operated HPGe detector systems have a great advantage in terms of ease of use, compactness, and the speed with which they can provide meaningful results. The accuracy and precision of the isotopic composition reported by these systems is not as good as with state-of-the-art laboratory systems, but, in most cases, are sufficient for performing in-field categorization of the material. One important feature to consider is the ability to use the HPGe detector remotely via a cable or a wireless connection, depending on the details of the deployment (e.g., to categorize material in an area with a higher dose rate or in order to determine the threat posed by a potential Radiological Dispersal Device or dirty bomb ). The radiation protection of the emergency workers must be considered according to national regulations and all radiation doses recorded. The radiation doses for the MEST and other deployed forces should be kept as low as reasonably achievable (ALARA) during the measurements. Every measurement should be documented with the location (including GPS coordinates if possible), date, time, equipment used, and measured dose rate. Also, a sketch of the experimental arrangement (including the distance to the sample) is helpful. A pre-printed protocol in the form of a questionnaire can help the MEST members document vital information. 3. SAMPLE REQUIREMENTS Items analyzed on-site by gamma spectrometry can have any physical shape and chemical composition. Items are usually analyzed as a whole, without subsampling. Typically, on-site measurements are taken for approximately 10-20 minutes to obtain enough information to categorize the material; such short counting times will, of course, reduce the precision of the measurements due to counting statistics. To improve the counting statistics, any unnecessary shielding between the detector and the sample should be removed. The thickness of shielding and the distance between the detector and the investigated item should be optimized in such a way as to achieve a compromise between count rate and detector dead time, resulting in the best counting statistics within the available measurement time. If possible, the investigated item should be moved away from any other sources of radiation present, and/or collimation should be used to ensure that only the item of interest is in the field of view of the detector. If more sources are to be categorized, they should be categorized individually. 4. PRO S AND CON S OF THE TECHNIQUE Pro s: Fully non-destructive method, preserving forensic evidence. Item analysis without dismantling the item. Minimal or no sample preparation requirements. Portability. Fast results available in the field. Relatively inexpensive equipment compared to destructive techniques. The operating costs are also low. A high-resolution (HPGe-based) gamma spectrometry system costs typically between 60,000 100,000 Euro, depending on detector type and additional components. Con s: For high-resolution gamma spectrometers, detector cooling is needed: liquid nitrogen or electrical cooling. The precision is often less than the precision of destructive techniques (the standard deviation for a given measurement result using high-resolution gamma spectrometry will typically be higher). With present-day commercial gamma spectrometers it is not possible to directly derive information on 242 Pu from the measured spectrum. Isotopic correlations have to be used to estimate the 242 Pu content of the investigated items. For Pu with high 242 Pu content, this causes a large uncertainty in determining the Pu isotopic composition. Note, however, that by combining gamma spectrometry with two other non-destructive methods, calorimetry and neutron coincidence counting, it is possible to derive the 242 Pu content in a fully non-destructive way. 2

Due to on-site limitations (e.g. relatively short spectrum collection times) and due to the inability to fully control the measurement conditions on-site, the accuracy and precision of in-field measurements is typically lower than that in a laboratory setting. 5. FAQ COST, DO S AND DON T S, COMMON MISTAKES How much does a portable gamma spectrometry system cost? Compact, hand-held, low-resolution spectrometers (RadioIsotope Identifiers) cost between 16,000 30,000 Euro Compact, portable high-resolution (electrically cooled HPGe-based) spectrometers cost typically about 70,000-100,000 Euro. Portable (but not compact) spectrometers can be assembled from separate parts, resulting in a system cost which is roughly 50-60% of the above costs. For detectors that are cooled by liquid nitrogen, the cost of the liquid nitrogen supply should be considered. If the detector is to be kept at operating temperature all of the time, then an automated liquid nitrogen filling system and additional work safety measures could be cost effective. The cost of suitable calibration sources and their registration and storage should be included. How reliable are various gamma spectrometers in detecting nuclear material masked by the radiation of another radioactive source? Low resolution (NaI and LaBr3) spectrometers often fail to correctly identify nuclear material masked by another radioactive source emitting similar gamma energies (the masking limit is very low). Medium resolution spectrometers (CdTe and CdZnTe) have a better success rate, while high-resolution gamma spectrometers (HPGe) can identify nuclear material in most masking scenarios (the masking limit is quite high). How long does it take to cool down a warm highresolution detector, before it can be used in the field? For standard liquid-nitrogen cooled systems at least 4-6 (but preferably more) hours of cool-down time is recommended. These detectors can be kept cold for days without refilling the liquid nitrogen. For some new portable, liquid-nitrogen cooled HPGe detectors a cool-down time of two to six hours is specified by the manufacturers. Once cold and filled with liquid nitrogen, they can be kept cold for ~20 hours up to three days (depending on LN2-storage volume) without refilling. State-of-the-art compact, electrically cooled systems can reach their operating temperature within about three to four hours. To be on the safe side, detectors to be used in the field should always be kept at operating temperature, in order to be able to carry them on-site immediately once analysis is required. How long does it take to categorize nuclear material by high-resolution gamma spectrometry on-site? Usually not more than 15 minutes (plus a background measurement). For a first evaluation of the radiological situation by the MEST, typically around 30 minutes are required. Common mistakes HPGe detectors tend to lose liquid nitrogen during transport, so the level of filling should be checked before, and after, transport (e.g. via weighing). The temperature of the HPGe detector must be checked before a voltage bias is applied. In most detectors this occurs automatically. The function of the HPGe detector (both liquid nitrogen and electrically cooled) should be checked regularly (ideally on a weekly basis for a detector that is to be used in the field). The calibration of the detector should also be checked as part of the QA for the equipment. For in-field use, it may be necessary to take calibration sources into the field. 3

Due to the fact that low energy gamma radiation can be shielded more easily compared to high energy gamma radiation, it is possible that, for a given radionuclide, low energy gamma lines will be missing in the spectrum obtained by a HPGe detector. In this case, care should be taken when interpreting results from the automatic analysis of the spectrum. In all cases, it is recommended to view the spectrum to check for this kind of effect and other effects that can influence the automatic analysis software. 6. REFERENCES 1. ITWG Guideline: High-resolution gamma spectrometry General overview. 2. ITWG Guideline: Laboratory applications of highresolution gamma spectrometry for the analysis of special nuclear material. 3. IAEA Nuclear Security Series No. 2, Nuclear Forensic Support, International Atomic Energy Agency, Vienna, 2006. Also available online at : http://wwwpub.iaea.org/mtcd/publications/pdf/pub1241_ web.pdf. 7. CONTRIBUTORS This document was prepared on the basis of contributions from: J. Zsigrai Institute for Transuranium Elements European Commission C. Larsson Defense R&D Canada Canada E. Kroeger BfS Germany M. Thomas BfS Germany M. Wallenius Institute for Transuranium Elements European Commission N. Cong Tam IKI Hungary J. Safar Hungarian Atomic Energy Agency Hungary DOCUMENT REVISION HISTORY Version No. Version Date 1 March 2013 Document INFL-GSIF Description of Changes Initial Draft Changes made by J. Zsigrai 4