FTIR measurement of NH 3, HCN, SO 2, H 2 S and COS in pulverized lignite oxy-fuel flames Daniel Fleig, Stefan Hjärtstam and Daniel Kühnemuth

Similar documents
Gas Temperature and Radiative Heat Transfer in Oxy-fuel Flames

Combustion. Indian Institute of Science Bangalore

EXECUTIVE SUMMARY. especially in last 50 years. Industries, especially power industry, are the large anthropogenic

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Controlled FTIR measurements of acid gas (NO, mixed NO/SO 2 ) capture in water at high pressure

If we write the condensed electron configuration, then we can easily see the valence electrons.

Apportioning of Fuel and Thermal NO x

9/13/10. Each spectral line is characteristic of an individual energy transition

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur

AE 205 Materials and Energy Balances Asst. Prof. Dr. Tippabust Eksangsri. Chapter 4 Stoichiometry and MB with Reactions

Reacting Gas Mixtures

Topsøe Catalysis Forum 2009

COMBUSTION OF FUEL 12:57:42

Methodology for Analysis of Metallurgical Processes


very high temperature for excitation not necessary generally no plasma/arc/spark AAS

Chapter 4. Fundamentals of Material Balance

05/04/2011 Tarik Al-Shemmeri 2

Advanced Pharmaceutical Analysis

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

Quantity Relationships in Chemical Reactions

Measurement and Modeling of Particle Radiation in Coal Flames

VALIDATION STUDY OF FTIR-BASED EMISSIONS MEASUREMENTS AT A MUNICIPAL WASTE COMBUSTOR

Fundamentals of Combustion

1 (a) Describe a chemical test which shows the presence of water. Describe how water is treated before it is supplied to homes and industry.

C H A P T E R 3 FUELS AND COMBUSTION

Monitoring Flammable Vapors and Gases in Industrial Processes

FDE 211-MATERIAL AND ENERGY BALANCES: MATERIAL BALANCES ON REACTIVE SYSTEMS. Dr. Ilgın PakerYıkıcı Fall 2015

2. Review on Material Balances

COMBUSTION CHEMISTRY COMBUSTION AND FUELS

Chem 434 Instrumental Analysis Test 1

3 - Atomic Absorption Spectroscopy

ADVANCED DES SIMULATIONS OF OXY-GAS BURNER LOCATED INTO MODEL OF REAL MELTING CHAMBER

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry

It is the purpose of this paper to demonstrate that the detection limit is directly related to the sample size, for both TG and TG/FTIR.

ATOC 3500/CHEM 3151 Air Pollution Chemistry Lecture 1

7.1 Describing Reactions. Burning is a chemical change. When a substance undergoes a chemical change, a chemical reaction is said to take place.

Emission spectrum of H

2101 Atomic Spectroscopy

CoalGen 2009 Dynamic Control of SCR Minimum Operating Temperature

Fuel ash behavior importance of melting

LECUTRE 31:Refractory and gas radiation

UV3000. Accurate, precise, and portable ambient gas point analyzer

Atomic Absorption Spectroscopy (AAS)

Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System

Experiment #5. Empirical Formula

Thermogravimetric Analysis Advanced Techniques for Better Materials Characterisation

Balancing chemical reaction equations (stoichiometry)

MARK SCHEME for the October/November 2012 series 0620 CHEMISTRY. 0620/21 Paper 2 (Core Theory), maximum raw mark 80

ICSE Board Class IX Chemistry Paper 10 Solution SECTION I

Chapter 3: Stoichiometry

(for tutoring, homework help, or help with online classes)

1.23 Gas Calculations

Training Coarse ECOPROBE 5 CHEMISTRY RS DYNAMICS

2B Technologies, Inc. An InDevR Company

AQA Chemistry (Combined Science) Specification Checklists. Name: Teacher:

Process Chemistry Toolbox - Mixing

EXPERIMENTAL AND KINETIC

Fundamentals of Material Balances

Sustainable Power Generation Applied Heat and Power Technology. Equations, diagrams and tables

FDE 211 Material & Energy Balances. Instructor: Dr. Ilgin Paker Yikici Fall 2015

IGCSE (9-1) Edexcel - Chemistry

ammonia carbon dioxide hydrogen nitrogen electrical heat solar sound (a) In air, the two most common gases are oxygen and...

Combustion Generated Pollutants

THE UNITED REPUBLIC OF TANZANIA NATIONAL EXAMINATIONS COUNCIL CERTIFICATE OF SECONDARY EDUCATION EXAMINATION. Instructions

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers

ISA QATAR : 90 th Seminar

Chemical Kinetics: NOx Mechanisms

White Paper. Overview: NDIR Definition:

AAE COMBUSTION AND THERMOCHEMISTRY

Q1. Methane and oxygen react together to produce carbon dioxide and water.

Class XI Chapter 1 Some Basic Concepts of Chemistry Chemistry

(g) 2NH 3. (g) ΔH = 92 kj mol 1

I. Measurements of soot - Laser induced incandescence, LII. spectroscopy, LIBS

Fundamentals Of Combustion (Part 1) Dr. D.P. Mishra Department of Aerospace Engineering Indian Institute of Technology, Kanpur

General Stoichiometry Notes STOICHIOMETRY: tells relative amts of reactants & products in a chemical reaction

Stoichiometry. Please take out your notebooks

Laboratory Experiment No. 3 The Empirical Formula of a Compound

Hydrogen addition to the Andrussow process for HCN synthesis

Stoichiometry. Lab. FCJJ 16 - Solar Hydrogen Science Kit. Goals. Background. Procedure + 2 O 2 CH 4 CO H2O

THE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION

Modeling Conservation of Mass

CHEMISTRY PAPER II. Q1. If ethanol reacts with oxygen it produces (a) Acetic Acid (b) Hydrocloric Acid (c) Sulphuric Acid (d) Sulphur dioxide. Ans.


RDF-LIGNITE CO-COMBUSTION IN A CFB BOILER

UV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability

Chemistry (

Class X. Exercises solution

Chapter 6. Thermochemistry

Fig. 3.2 on Page 101. Warming. Evidence for CO 2. History of Global Warming-2. Fig. 3.2 Page 101. Drilled cores from ocean floors

5.7 Quantity Relationships in Chemical Reactions (Stoichiometry)

Dr Ali Jawarneh Department of Mechanical Engineering Hashemite University

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

SCR Catalyst Layer Addition Specification and Management

The Claviature of Gas Analysis

Investigations of aerosol formation pathways during MSW combustion based on high-temperature impactor measurements

SUPPLEMENTARY TOPIC 3 ENERGY AND CHEMICAL REACTIONS

A First Course on Kinetics and Reaction Engineering Example 1.2

Usual Atomic Charges of Main Group Elements

Lecture 14: Pyro metallurgical extraction (Roasting)

Transcription:

FTIR measurement of NH 3, HCN, SO 2, H 2 S and COS in pulverized lignite oxy-fuel flames Daniel Fleig, Stefan Hjärtstam and Daniel Kühnemuth Abstract Nitrogen and sulphur compounds are investigated in the atmosphere of a high temperature oxy-fuel flame, using Fourier Transform Infrared Spectroscopy (FTIR). For this application, evaluation routines have been developed to investigate the concentrations of NH 3, HCN, SO 2, H 2 S and COS by analyzing the absorption intensity of characteristic fingerprints within the strong overlapping of the CO2 rich sample gas matrix. NH 3 and SO 2 was successfully analysed however HCN, H 2 S and where not measured. The later species had either low concentration at the specific spatial measurement point or could not be detected due to too strong overlapping from absorption bands other gases in the same wavelength. It is furthermore discussed if decomposition of HCN, H 2 S and COS on the filter of the sampling probe or elsewhere during the sampling process could cause the noticed difficulties in detectability of those gases. 1 Introduction The question of measuring the precursors for nitrogen oxides (NO x ) formation and sulphur compounds in the pulverized lignite flame of the oxy-fuel reactor at CTU was posed within the first part of the Bio-Fuel GS II course, held at the Chalmers University of Technology (CTU) from the 20 to 24 of October 2008. At CTU, one has good experience with FTIR measurements in a circulating fluidized bed boiler (CFB),

however not in oxy-fuel combustion. In oxy-fuel combustion the gas atmosphere differs completely from a CFB combustion atmosphere since oxygen (O 2 ) is used instead of air to create a flue gas with high concentration of carbon dioxide (CO 2 ). To reduce the combustion temperature, flue gas is recycled and, as a consequence, the combustion takes place in an atmosphere with high concentrations of combustion products. This has important implications on the combustion reactions. For example are NO x emissions reduced by reburning when recycled into the flame. To understand the reduction mechanism, it is important to achieve a good spatial resolution of NH 3 and HCN as the two most important precursors of NO x Sulphur compounds like sulphur dioxide (SO 2 ), hydrogen sulphide (H 2 S) and carbonyl sulphide (COS) on the other hand, are of interest regarding to fate of sulphur in an oxy-fuel combustion process. FTIR measurements offer a good possibility for analysis of such compounds. However high concentrations of CO 2 and water vapour (H 2 O) in an oxy-fuel flame make the evaluation of infrared spectra complicate, since the two gases have strong overlapping infrared absorbance bands over wide ranges of wave numbers. The H 2 O concentration depends on the kind of recycle (dry or wet), the oxygen concentration in the oxidiser and the stochiometric ratio. Summarized, there are two additional challenges in FTIR measurements in a pulverized lignite oxy-fuel flame compared to measurements in a CFB boiler: Considerably higher temperatures of up to about 1700K during the sampling in the flame; higher CO 2 concentrations compared to CFB combustion. This makes it necessary to evaluate the retrieved FTIR spectra individually. The aim of the work was to measure nitrogen and sulphur species in an oxy-fuel flame and to give evaluation routines for the concentrations of NH 3, HCN, SO 2, H 2 S and COS by using the software tool Grams AI.

2 Measurements 2.1 Experimental description The measurements were performed in the Chalmers 100kW oxy-fuel test unit, firing pre-dried pulverized lignite as fuel. The proximate and ultimate analyse of the fuel is given in Table 1. The heat input was P th =76 kw. That is equivalent to a lignite mass flow of 13 kg/h. Oxyfuel combustion with dry recycle, 40 vol% O 2 in the oxidizer and a stoichiometric ratio of around λ = 1.17 was established. Dry recycle was chosen as it shows lower water vapour concentrations and thereby limits the influence of water vapour on the FTIR measurements. Sampling of the gases was performed with a gas suction probe. The probe, which is shown in Figure 1, has an inner electrical heated tube. The temperature is controlled for guaranteeing a temperature higher than the acid dew-point temperature of the sampled gases. Through the outer tubes flows cooling water for lowering the temperature in the sampling gases and protect the probe from too high temperatures. Before entering the sample probe the gases have to pass through a filter of aluminium silicate at the top of the probe. Further on the gases are transported via heated pipes (above acid dew point) and a heated pump to the FTIR instrument (Bomem 9100). As the measurements were performed on about 200 C, all here given concentrations are on wet basis. Three different measurement points where approached. The measurement points were at 215 cm vertical distance from the burner inlet (position R2 in Figure 1) and at 11, 12 and 13 cm distance from the center line of the cylindrical furnace respectively. Position R2 was chosen because it is a typical position for an early stage in a flame where the high concentrations of the mentioned species are suspected. On this level, measurement points with substochiometric conditions were

searched, since in such regions the compounds of interest can be most likely available. Table 1. Analysis of the Lausitz lignite, d.a.f. = dry ash free. Proximate [wt. %, as received] Y moisture Y ash Y combustible Volatile matter Ultimate [wt. %, d.a.f.] [wt. %, d.a.f.] Y C Y H Y N Y S Y O 10.7 4.38 84.9 61.0 67.2 5.36 0.87 0.90 25.7 10.6 4.56 84.8 58.8 67.0 5.37 0.81 0.90 25.9 10.5 4.75 84.7 57.7 66.8 5.39 0.82 0.88 26.1

Figure 1. Sketch of the oxy-fuel combustion furnace with top-fired pulverized coal burner at CTU. Measurements were performed in port R2 (marked in red). 1620 mm 4 mm Cooling water outlet Heated tube: to prevent condensation Gas suction tube Cooling water inlet A A Section A-A Ø45,0/41,0 Ø33,0/29,0 Ø18,0/14,0 Ø8,0/6,0 m Figure 2. Gas suction probe 2.2 Data Acquisition Before the measurements in the flame, a background spectrum (0 spectrum) from the used purge gas was taken (nitrogen atmosphere). The FTIR was set up in that way that for each of the three measurement

positions a series of ten single interferograms was taken. These retrieved interferograms of the measured flue gas composition are via Fast Fourier Transform (FFT) transformed into single beam spectra respectively. Of the retrieved ten single beam spectra, a time average spectrum is formed. These average single beam spectra are then converted into absorption spectra, by applying the Bouguer-Lambert- Beer law. Consequently each measurement point is presented by one absorption spectrum. The spectra are then used to perform spectral subtractions in order to determine the concentration of HCN and NH 3 SO 2, H 2 S and COS. The three measurement points in level R2 (se Figure 1) are named R2_11, which is 11cm from the centreline, R2_12 and R2_13 which are 12cm and 13cm from the centreline respectively. Here, mainly the evaluation of the latter one is presented as the position with the most reducing conditions. In series with the FTIR instrument the concentrations of CO 2 O 2, CO, NO and SO 2 were measured by standard gas analysis instruments. 2.3 Spectra Subtraction The concentrations of HCN, NH 3, SO 2, H 2 S and COS, probably present in reducing atmosphere of the flame zone, are determined by successive spectral subtraction of several reference spectra of interfering gas species from the absorbance spectrum of the measured gas sample. As the concentration of species of interest in the sample generally differs from that in the available reference gases, each reference gas spectrum is multiplied by a scaling factor F. A A F = 0 (1) [Åmand et al.] s R

Were A S is the absorbance of the sample and A R is the absorbance of the reference gas. Following this method, the characteristic fingerprint of the searched gas should be retrieved on a clear baseline. Combustion gases are a mixture of many different gases. Especially H 2 O and CO 2 have absorption bands over wide ranges of wave numbers. Already in flue gases from air combustion, the high CO 2 and H 2 O concentrations have wide bands with total absorption, narrowing the possible bands for evaluation significantly down. But even in the more open bands, overlapping CO 2 and H 2 O absorption bands give the need for subtraction of reference spectra for these two species. Otherwise equation (1) does not give zero and thus the accuracy of the evaluation is reduced. 3 Results 3.1 NH 3 Ammonia has a characteristic absorption band, or fingerprint, mainly consisting of two distinct absorption peaks at 965 and 930 cm -1 respectively. This fingerprint, blue line in Error! Reference source not found., can be clearly identified in the green sample spectrum. In order to be able to evaluate the concentration of NH 3 that is corresponding to the intensity of these absorption peaks, absorption spectra of CO 2, H 2 O and C 2 H 4 are subtracted from the initial sample spectrum. Error! Reference source not found. shows the subtraction of a calculated (Formula 1) 76 vol% CO 2 (blue spectrum) from the sample spectrum (green). This subtraction is done with a CO 2 reference spectrum of 89.7 vol% CO 2. A factor F = 0.886 was determined on the interval between 997.6 and 909.7 cm -1 using the auto-subtract function within the FTIR software Grams AI.

Figure 3. Subtraction of a CO 2 reference spectrum [76.0 vol% (F=0.84729)] (blue) from the flue gas absorption spectrum (green) between 997.6 and 909.7 cm -1 (R2_13) The result of this first substraction is shown in Error! Reference source not found. as green spectrum. The blue absorption spectrum corresponds to a water sepectrum with 11 vol% water. In this wavenumber region the absorbance of water is low. The subtraction of the water spectrum from the measured spectrum (green) has therefore no significant change. The received spectrum (green) is shown in Error! Reference source not found..

Figure 4. Subtraction of a H2O reference spectrum [11.0 vol% (F=0.94988)] (blue) from the CO2 subtracted flue gas absorption spectrum (green) between 997.6 and 909.7 cm-1 (R2_13) The peak between the two NH 3 absorption peaks at 965 and 930 cm -1 in Error! Reference source not found. is an absorption peak of ethylene, C 2 H 4. The subtraction of 136.6 ppm C 2 H 4 (blue spectrum in Error! Reference source not found.) leads finally to an absorption spectrum, showing the characteristic NH 3 peaks on a clear baseline (blue spectrum in Figure 6)

Figure 5. Subtraction of a C2H4 reference spectrum [136.6 ppm (F=0.94878)] (blue) from the CO2 and H2O subtracted flue gas absorption spectrum (green) between 997.6 and 909.7 cm-1 (R2_13) The final substraction shown in Figure 6 of 477.5 ppm NH 3 corresponding to the blue spectrum in Figure 6 leads to a clear basline (red spectra), denoting that the substractions where correctly done. Figure 6. Subtraction of a NH3 reference spectrum [477.5 ppm (F=0.99485)] (blue) from the CO2, H2O and C2H4 subtracted flue gas absorption spectrum (green) between 997.6 and 909.7 cm-1. Resulting subtracted spectrum in red. (R2_13) 3.2 HCN The strongest absorption peak of HCN is at 714 cm -1. Unfortunately the overlapping of CO 2 absorption bands is in this region so strong that the signal back from this region is not linear according to the Bouguer- Lambert-Beer law and therefore no evaluation of the concentration of HCN is possible. An other band for HCN quantification can be found

between 3375 and 3200cm -1. It is less intense than the peak at 715 cm -1 but the interference with CO 2 and H 2 O is not that severe. In Error! Reference source not found. the substraction of a absorption spectrum of 13.5 vol% H 2 O (blue spectrum) is shown. This value differs from that one obtained by water subtraction between 997.6 and 909.7 cm -1 (see Error! Reference source not found.) and denotes the uncertainty of determining the water content. Figure 7. Subtraction of a H2O reference spectrum [13.5 vol% (F=0.96159)] (blue) from the flue gas absorption spectrum (green) between 3400 and 3200 cm-1 (R2_13) The subtraction of 74.8 vol % CO 2, shown in Error! Reference source not found. is done in the spectral range of 1100 to 1010 cm -1 as the absorption band of CO 2 is in that region more defined, and better

subtraction results are obtained. Figure 8. Subtraction of a CO 2 reference spectrum [74.8 vol% (F=0.83376] (blue) from the H 2 O substracted flue gas absorption spectrum (green) between 1100 and 1010 cm -1 (R2_13)

The final substraction of HCN is though still hard to define, as the noice of other absorption bands, most probably resulting from imperfektion of the water band subtraction. The result of the substraction is that only 15 ppm HCN are supposed to be found in the sample. This can be seen by the fact that the pattern of the blue marked HCN absorption band in Error! Reference source not found. is hardly to be found in the sample spectrum in green. The uncertainty is high due to the low subtraction factor of F=0.17364. Figure 9. Subtraction of a HCN reference spectrum [15 ppm (F=0.17364)] (blue) from the H2O, CO2 and C2H2 subtracted flue gas absorption spectrum (green) between 3400 and 3200 cm-1 (R2_13) 3.3 SO 2, H 2 S and COS In general it was well possible to analyse the SO 2 content in the gas sample with help of the FTIR because of its relatively high absorbance between 1230 and 1010 cm-1. For example in R2_11 around 2240ppm SO 2 (wet basis) were detected. The procedure for SO2 analysis is shown in Figure 10 to 12.

Figure 10 shows the subtraction of a 14.2% H 2 O reference spectrum (green) with a scaling factor F of 1.251 from the measured spectrum. This yields in a water reference spectrum with theoretical 17.76%. The blue curve is the resulting measured spectrum without water bands. Figure 10. Subtraction of H 2 O reference spectrum [14.1% x 1.251] (green) from the sampled spectrum (red) between 1230 and 1016cm -1 (R2_11), resulting measured spectrum blue Figure 11 shows the subtraction of a 89.7% CO 2 reference spectrum with a scaling factor F=0.8733 from the resulting spectrum in Figure 10 between 1100cm -1 and 1010cm -1. The blue curve denotes the new resulting measured spectrum.

Figure 11. Subtraction of CO 2 reference spectrum [89.7% x 0.8733] (green) from the resulting spectrum from Figure 10 between 1100 and 1010cm -1 (R2_11), resulting measured spectrum (blue) Figure 12 shows the subtraction of a 1994ppm SO 2 spectrum (green) with a scaling factor F=1.1187, from the resulting measured spectrum in Figure 11 between 1230 and 1016 cm-1. This corresponds to a SO 2 concentration of 2231ppm on wet basis. If also a subtraction of a methane reference spectrum was done a SO 2 concentration of 2242ppm was gotten.

Figure 12. Subtraction of SO 2 reference spectrum [1994ppm x 1.1187] (green) from the resulting spectrum from Figure 11 between 1230 and 1016cm -1 (R2_11), resulting spectrum after subtraction (blue) Correspondingly, a SO 2 concentration in the same range was measured in the flue gas (furnace outlet). However in the measurement point R2_13 only around 300ppm SO 2 was measured. A reference spectrum with 1017ppm and a scaling factor of 0.2934 was used. The relatively low scaling factor gives a higher uncertainty compared to the scaling factor used in point R2_11. Anyway the SO 2 concentration was significant lower compared to that in point R2_13. H 2 S and COS are in general difficult to analyse because of their weak absorbance. The absorbance takes place between around 2080 and 2010cm-1. It was not possible to analyse H 2 S and COS in the gas samples. For example in R2_11 neither H 2 S nor COS was noticeable. Figure 13 shows the subtraction of a H 2 S reference spectrum from the measured spectra (H 2 O, CO 2 and CO reference spectra are already subtracted). A scaling factor of -4.3 was received which means that H 2 S is not detectable. In Figure 14 the subtraction of COS from the measured spectra is shown. Here a scaling factor of -0.06 was received which means that COS is not detectable.

Figure 13. Unsuccessful subtraction of H 2 S reference spectrum [842ppm x -4.3] (green) from the measured spectrum in R2_11 (red), resulting spectrum after subtraction (blue) Figure 14. Unsuccessful subtraction of COS reference spectrum [108ppm x - 0.6] (green) from the measured spectrum in R2_11 (red), resulting spectrum after subtraction (blue) In the measurement point R2_13 the overlapping of other gases (probably CO) was too high so that no subtraction was possible, see Figure 15.

Figure 15. Disturbance in the spectral range where COS or H2S could be analysed, red: measured spectrum, green: COS reference spectrum (R2_13) 4 Discussion and conclusions The FTIR measurements at the Chalmers pulverized lignite boiler have shown that it is possible to identify small amounts of NH 3. On the specific measurement point, a peak with 134.1 ppm was clearly identified. This was the fact even though the matrix in the sample gases from oxy-fuel combustion is from a data evaluation point of view much more complicated than in samples from air combustion because of the high CO 2 concentrations. Overlapping of absorption bands is even at the very high CO 2 of the oxy-fuel flue gases not so severe that it would interfere with the evaluation of the characteristic NH 3 absorption peaks at 930 and 965 cm -1. Only insignificantly low concentrations of HCN were detected. Beside data evaluation problems as high interference with the water absorption bands, there seems to be a problem with the data acquisition as well. One would expect that at reducing conditions as found in the measurement point P24, where about 477.5 ppm NH3 and several percent CO have been found, some HCN should be present as well. The fact that in the sampled gases almost no HCN was found could be explained with two different hypotheses:

There is now HCN present in the specific measurement points (In this case the choice of sample positions should be reaccessed) The HCN present in the flue gases is destroyed during sampling, for example due to catalytic reaction with the sample probe material or the filter cake of char at the tip of the sampling probe The second hypotheses is basing on the findings of Bassiakis et al. They describe that once formed HCN can react with coal hydrogen towards NH3, if the contact time between gas and particles is long enough, and if little oxygen is present to compete against the HCN on the coal hydrogen. It was always possible to analyse the SO 2 concentration in the gas sample. In measurement point R2_11 around 2240ppm SO 2 was measured and in point R2_13 only around 300ppm SO 2 was measured. Such low SO 2 concentrations can denote substochiometric conditions and other sulphur compounds like H 2 S and COS could be expected. However it was not possible to detect H 2 S or COS in the point R2_13 because of too strong absorbance bands from other gases, probably CO. In the measurement point R2_11 no H 2 S or COS were seen in the spectrum. For the last case similar questions like for HCN can be asked. References Åmand. L-E. et al. The theory behind FTIR analysis, Application Exapmples Form Measurement at the 12 MW Circulating Fluidized Bed Boiler at Chalmers (course material, Lund university)

Bassilakis, R et al. Sulfur and nitrogen evolution in the Argonne coals. Experiment and modeling. Energy & Fuels (1993), 7(6), 710-20.