Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Similar documents
Fast methods for the determination of ibuprofen in drug products

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of trace anions in concentrated hydrofluoric acid

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Dionex IonPac AS28-Fast-4µm column

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

ProPac WCX-10 Columns

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Determination of urea in ultrapure water by IC-MS/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Determination of Benzo(a)pyrene in Sausage and Preserved Ham

Automated and accurate component detection using reference mass spectra

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

APPLICATION NOTE Goal. Introduction. Keywords

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Qualification Specifications for Agilent 1100/1200 HPLC Systems

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

This method describes the identification of the following prohibited colorants in cosmetic products:

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Inline sample preparation for the determination of anions in sodium hydroxide

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Expanded capabilities in ammonium and amine detection

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

LC Technical Information

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Routine MS Detection for USP Chromatographic Methods

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

in animal feed Received: 29 June 2007 / Accepted: 20 May Wageningen Academic Publishers

HPLC Praktikum Skript

Impact factor: 3.958/ICV: 4.10 ISSN:

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Direct Determination of Small Organic Acids in Sea Water by IC-MS

CORESTA Recommended Method No. 86

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Transcription:

APPLICATION NOTE 72686 Determination of underivatized aflatoxins,, G2, and in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection Authors Sylvia Grosse, Mauro De Pra, Frank Steiner Thermo Fisher Scientific, Germering, Germany Keywords Aflatoxins, immunoaffinity solidphase extraction, FLD, Vanquish Fluorescence Detector F Goal Determination of four aflatoxins in ground hazelnut by immunoaffinity solid-phase extraction sample purification and subsequent HPLC measurements with fluorescence detection Application benefits Rapid chromatographic method for the determination of four aflatoxins in hazelnuts by fluorescence detection without derivatization Clear advantage of selective purification and enrichment of the aflatoxins on immunoaffinity solid-phase extraction cartridges for improved limit of detection and quantification Introduction Mycotoxins are naturally occurring fungal toxins that were first found in fungus aspergillus flavus. Most of them are very stable and are not destroyed during processing or cooking procedures. One common group are the aflatoxins, of which 2 naturally occurring forms are known. Aflatoxin is considered to be the most toxic to human health, but in addition the aflatoxins,, G2,, and the milk-derived derivatives M1 and M2 also have high importance. The B and G aflatoxins occur in various foods, such as nuts, grains, and spices, while the M derivatives are found in dairy products. The focus of this application is the determination of the toxins,, G2, and in ground hazelnuts. The European Commission has set various maximum levels of aflatoxins in several foods under consideration of their consumption and use. 1

The maximum level for aflatoxin ranges from 2 to 12 μg/kg for foods used for direct consumption or as an ingredient, with the exception of baby food products, which allow for a maximum level of.1 μg/kg. The limit sum of all four aflatoxins varies between 4 and 15 μg/kg. Therefore, a sensitive and accurate analytical method is required to monitor the low levels in various foods. For reliable identification and quantification, highperformance liquid chromatography (HPLC) coupled with fluorescence detection (FLD) is one of the most common techniques. Since aflatoxins and fluoresce less intensely than G2 and, it may be necessary to use derivatization reagents or mass spectrometry to enable trace level quantitation. Another common approach to improve the specificity, and therefore the limit of quantitation (LOQ), is to use immunoaffinity solid-phase extraction (SPE) to selectively purify and enrich the aflatoxins 2 prior to the HPLC analysis. With this sample purification method, low LOQ values can be achieved even in highly complex and fatty matrices. This application note describes the determination and quantification of four aflatoxins in ground hazelnuts by using AflaCLEAN Select immunoaffinity SPE for cleanup followed by HPLC-FLD analysis without derivatization. The Thermo Scientific Vanquish Flex Fluorescence Detector (FLD) provides sufficient trace level detection performance to determine aflatoxins in this matrix far below the limits defined by the European Commission. Experimental Chemicals Deionized water, 18.2 MΩ cm resistivity or higher Fisher Chemical Optima Methanol LC/MS grade (P/N 1767665) Fisher Chemical Optima Acetonitrile LC/MS grade (P/N 11334) Fisher Scientific Hexane, HPLC grade (P/N 173611) Fisher Scientific PBS buffer (phosphate buffered saline), ph 7.2 (P/N 1153546) Fisher Scientific Acetic acid LC/MS grade (P/N 18671) Fisher Scientific Sodium chloride, purris. (p.a.) (P/N 1562677) Aflatoxin mixture (purchased from a reputable vendor) Equipment Magnet stirrer, Fisher Scientific Folded filter paper, Grade 597½, Ø 185 mm, Fisher Scientific (P/N 1433141) Syringe filter, regenerated cellulose (RC), Ø 15 mm,.2 µm, Fisher Scientific (P/N 1712712) AflaCLEAN Select immunoaffinity SPE columns, 3 ml (purchased from a reputable vendor) Conical tubes (15 ml), Fisher Scientific (P/N 1137211) Vials (amber, 2 ml), Fisher Scientific (P/N 155876) Septa (silicone/ptfe), Fisher Scientific (P/N 1154818) Preparation of standards Quantification was done by standard addition calibration. To prepare the calibration standards, the stock solution containing all four aflatoxins, with a concentration of 857 µg/kg for G2 and and of 2856 µg/kg for and in methanol (stock solution 1), was diluted 1:1 using 1% acetic acid (stock solution 2). Further dilutions of stock solution 2 were made to add defined concentrations to a purified sample extract after immunoaffinity SPE (see next section). Table 1 gives an overview of the calibration levels and corresponding concentrations. Table 1. Calibration levels and added concentration in µg/kg to a purified sample extract Calibration Level Toxins G2, Concentration Toxins, Concentration 1.2.7 2.4 1.4 3.9 2.9 4 1.3 4.3 5 1.7 5.7 6 2.1 7.1 2

The recovery experiment was carried out by adding 113 µl of the aflatoxin stock solution 1 to 2 g of the ground hazelnut sample prior to sample preparation. The resulting concentration of the recovery sample was 1.7 µg/kg for G2 and and 5.65 µg/kg for and when 1% recovery is assumed. Preparation of samples Spiked and non-spiked samples were prepared in triplicate. 2 g sodium chloride and 2 g ground hazelnut sample Add 1 ml 8% methanol Add 5 ml hexane Stir for 15 min Filter through folded filter Mix 14 ml of lower liquid phase with 86 ml PBS buffer ph 7.2 Load 5 ml onto immunoaffinity SPE cartridge Wash the cartridge with 1 ml water Elute with 2 ml methanol; let the first 1mL act for 5 min on the column Dilute the extract 1:2 with 1% acetic acid Filter through.2 µm RC membrane in HPLC vial Instrumentation Thermo Scientific Vanquish Flex UHPLC system consisting of: System Base Vanquish Flex (P/N VF-S1-A-1) Quaternary Pump F (P/N VF-P2-A-1) Split Sampler FT (P/N VF-A1-A-1) Column Compartment (P/N VH-C1-A-1) Fluorescence Detector F (P/N VF-D51-A-1) Standard Bio Flow cell, 8 µl, 2 bar (P/N 679.423) LC conditions Column: Thermo Scientific Acclaim C18, 1 3 mm, 3 µm (P/N 76186) Mobile phase: Flow rate: Isocratic mobile phase condition: Isocratic run time: 4 min A: Water B: Methanol C: Acetonitrile.5 ml/min 5% A, 3% B, 2% C Mixer volume: 35 + 5 µl Column temp.: 3 C (forced-air mode, fan speed 5) Active pre-heater temperature: 3 C Autosampler temperature: 4 C FLD Excitation wavelength: FLD Emission wavelength: 365 nm 45 nm Sensitivity: 8 Lamp mode: HighPower UV data collection rate: 1 Hz UV response time:.5 s Injection volume: 2 µl Needle wash: 9/1 water/methanol (v/v) 3

Data processing and software The data acquisition and processing was done with Thermo Scientific Chromeleon 7.2.7 Chromatography Data System (CDS) software. Results and discussion Figure 1 shows the separation of a standard aflatoxin solution on the Acclaim 12 C18 column with excitation = 365 nm and emission = 45 nm. No immunoaffinity SPE clean-up was used here. As can be seen, there are some peaks eluting before the first target aflatoxin G2. These unknown peaks are impurities in the standard solution as they were not observed in the blank (data not shown). All aflatoxin analytes are baseline separated within 4 min. 1.5e6 1.e6 5.e5.e G2 For quantification, the ground hazelnut samples and recovery samples were prepared in triplicate. Figure 2 shows the calibration curves of the standard addition calibration method for all four aflatoxins. The original sample, which contained all analytes, was set to zero, which results in a negative x-axis intercept. In this way, the calculated amount of the analytes corresponds to the absolute amount of the negative x-intercept. Linearity (R 2 ) was found to be.992.9974 for all four aflatoxins and the percentage of relative standard deviation of the retention times [% RSD RT] were all below <.2% (Table 2). The limit of detection (LOD) and the limit of quantification (LOQ) were calculated based on the sample extract diluted to a S/N ratio approximately of 3 for LOD and a S/N ratio of 1 for LOQ of each analyte and injected in triplicate. One example chromatogram of LOD and LOQ determination of aflatoxin in the diluted sample extract is given in Figure 3. Table 2 shows a full summary of calibration, linearity, LOD, and LOQ results...5 1. 1.5 2. 2.5 3. 3.5 4. Figure 1. FLD chromatogram of a standard solution of the four aflatoxins: G2,,, and at concentrations of.9 µg/kg for G2 and and 2.9 µg/kg for and G2 35 y = 13635x + 47794 3 R² =.9972 25 2 15 1 5.5 1 1.5 2 2.5 Concentration 25 2 15 1 5 y = 2681.6x + 4561.2 R² =.99466 2 4 6 8 Concentration 4 35 3 25 2 15 1 5 y = 13714x + 45381 R² =.99742.5 1 1.5 2 2.5 Concentration Figure 2. Calibration curves for all four aflatoxin analytes 4 35 3 25 2 15 1 5 y = 3267.8x + 1759 R² =.9924 2 4 6 8 Concentration 4

Table 2. Data of % RSD RT (n=13), calibration range, linearity, LOD, and LOQ with standard deviation (S.D.) (n=3) Compound Name % RSD RT Calibration Range R 2 LOD ± S.D. LOQ ± S.D. G2.9.2 2.1.997.75 ±.8.185 ±.17.17.7 7.1.9947.931 ±.76 1.329 ±.66.9.2 2.1.9974.14 ±.13.26 ±.17.15.7 7.1.992 1.56 ±.154 1.122 ±.61 15 1 5 A S/N 5 3 As can be seen in Figure 4, the immunoaffinity SPE cleanup results in pure extracts. Some matrix peaks can be observed in the first two minutes of the chromatogram, but there is no interference in the target analyte region from 2 to 4 min. Neither the non-spiked sample extract nor the spiked (recovery) extract, where the standard solution was added before the sample preparation, show a peak in front of the toxin G2 as it was observed for the standard solution in Figure 1. In Figure 5, which illustrates a blank injection and one calibration point, an impurity can be detected, as the standard was added after the cleanup procedure to the extract. This leads to the assumption that a clear advantage of immunoaffinity purification is that it eliminates impurities which comes from the standard solution. 15 1 B 5.e6 G2 4.e6 3.e6 2.e6 5 1.e6.e S/N 8..5 1. 1.5 2. 2.5 3. 3.5 4. Figure 4. Overlay of spiked (recovery) sample (orange) and nonspiked (blue) hazelnut sample 3 1.5e6 G2 Figure 3. A) Determination of LOD with S/N of approximately 3 of aflatoxin (triplicate injection); B) determination of LOQ with S/N of approximately 1 of aflatoxin (triplicate injection) 1.e6 5.e5.e 2. 2.5 3. 3.5 4. Figure 5. Zoomed overlaid chromatograms of processed sample spiked with the highest calibration concentration (blue) and consecutive blank injection (orange) 5

In addition, it can be shown in Figure 5 that no carryover in the blank injection was observed, even after the injection of the processed sample spiked with the highest calibration concentration with 2.1 µg/kg for G2 and and 7.1 µg/kg for and. Quantitative results with recovery rates of each compound and calculated sample amounts (corrected by recovery rate) are summarized in Table 3. Table 3. Recovery and calculated sample amount results of ground hazelnut (averaged from three preparations) Compound Name Recovery Rate [%] Calculated Sample Amount G2 1.4 72 2.2 1.3 95 3.4 The recovery rates are excellent for G2,, and but inferior for. The vendor of the immunoaffinity SPE cartridges reports a minimum recovery of 9% for in the data sheet. However, cartridge stability is limited, and the shelf life is reported to be only a few months under proper storage conditions. The cartridges used in this study were close to their expiration date, which could have caused the lower recovery rate for. The tolerated aflatoxin levels in ground hazelnuts, valid for both direct consumption and ingredient use, is defined by the European Commission Regulation to be 5 µg/ kg for and 1 µg/kg for the sum of,,, and G2. All four aflatoxins were detected in the sample. For, the amount found was 3.4 μg/kg, which is below the maximum limit. The sum of all four compounds should not exceed 1 µg/kg and was calculated to be 6.3 µg/kg. Conclusion The combination of immunoaffinity SPE purification and enrichment with FLD detection without derivatization offers a sensitive analytical method for the quantification of the aflatoxins G2,,, and in ground hazelnuts. The Vanquish Fluorescence Detector F provides sufficient trace level detection performance down to 1 µg/ml for aflatoxins and and.1 µg/ml for and G2, enabling aflatoxin analysis in ground hazelnuts far below the tolerance levels defined by the European Commission. Good selectivity, linearity, and recovery for reliable quantitative results were observed with the applied method. The method run time of less than 4 min allows a high sample throughput. References 1. COMMISSION REGULATION (EC) No 1881/26 of 19 December 26: setting maximum levels for certain contaminants in foodstuffs, ANNEX section 2: mycotoxins (M5). 2. Karsten, M.; Swart, R.; Mcleod, F.; Murphy, B.; Henderson, S.; Richter, B. Fast end Effective Determination of Aflatoxins in Grains for Food Using Accelerated Solvent Extraction followed by HPLC (HPLC 28 presentation), Chromatography-Foods- Beverages-Contaminants-Applications-Notebook-71476, page 32 33. For Research Use Only Find out more at thermofisher.com/liquidchromatography 218 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise stated. AflaCLEAN is a trademark of LCTech GmbH.This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AN72686-EN 418S