Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets

Similar documents
Limit-test of NDMA and NDEA in Sartans. by GC-MS (Liquid-direct-injection)

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

CORESTA Recommended Method No. 86

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

Supporting Information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Determination of Aminoglycosides in animal tissue by means of LC/MS-MS

16 Malachite green 17 Leucomalachite green

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

Psychoactive Drugs Urine LC-MS/MS Analysis Kit User Manual

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

637. Thiamethoxam. HPLC method

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

The analysis of organic acid content of additives, premix, feed, and water.

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

DISCLAIMER: This method:

ENVIRONMENTAL analysis

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Determination and quantification by GC-MS of melamine (1 mg/kg), ammeline and cyanuric acid (20 mg/kg)

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Sulfotepp impurities in Chlorpyrifos EC formulations

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

STANDARD OPERATING PROCEDURES

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

Agilent Standard Operating Procedure

Method - Determination of aromatic amines in mainstream cigarette smoke

Catecholamines Urine LC-MS/MS Analysis Kit User Manual

Agilent ESI and APCI sources: for polar to non-polar compounds

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Scope: Materials: Labeled 1.5 ml or 2 ml Eppendorf tubes Appropriate pipette tips (see Table 1) Labeled LC vials

Supplementary information 1. INSTRUCTIONS FOR STUDENTS

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Certified Reference Material - Certificate of Analysis

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Micro Volume QuEChERS kit

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Journal of Chemical and Pharmaceutical Research

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

This method describes the identification of the following prohibited colorants in cosmetic products:

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Uncontrolled Copy. Table of Contents. 1. Principle of Assay Specimens Reagents and Materials... 3

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

11 APPENDICES Appendix A. Analytical Method Dicamba on Filter Paper

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

High throughput analysis of polymer stabilizers with the Agilent 6420 Triple

The Analysis of Fluoroquinolones in Beef Kidney Using HPLC Electrospray Mass Spectrometry Application

DETERMINATION OF NINE VOLATILE NITROSAMINES AND HYDROXY-NITROSAMINES IN CIGARETTE FILLER AND MAINSTREAM TOBACCO SMOKE

Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks

New York State Department of Health - Wadsworth Center Laboratory of Organic Analytical Chemistry NYS ELAP Laboratory ID 10763

Leather Determination of ethoxylated alkylphenols Part 1: Direct method

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Soils, sludges and treated bio-waste Organic constituents - LAS by HPLC with fluorescence detection (LC-FLD) and mass selective detection (LC- MSD)

This protocol is a modified version of the DuBois assay (DuBois et al. 1956) for the quantification of total carbohydrate content of a sample.

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Transcription:

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets Contact: Oliver el-atma Chemisches und Veterinäruntersuchungsamt (CVUA) Karlsruhe, Germany (OMCL BW) Tel. 0049-71-96-609 Oliver.el-Atma@cvuaka.bwl.de Dr. Birgit Gutsche Chemisches und Veterinäruntersuchungsamt (CVUA) Karlsruhe, Germany (OCCL BW) Tel. 0049-71-96-568 Birgit.Gutsche@cvuaka.bwl.de 1 Purpose / Scope of application Detection and quantitative determination of the nitrosamines N-nitrosodimethylamine (NDMA) and N-nitrosodiethylamin (NDEA) in Sartan containing film coated tablets by UHPLC-APCI- MS/MS. Safety instructions The generally applicable guidelines of the laboratory apply, as well as in addition:.1 Workplace instructions Due to the carcinogenic properties of nitrosamines, all work must be carried out with gloves and under fume cupboard. The workplace instructions for CMR substances have to be followed.. Protective material Sample processing and preparation of standards for calibration: single use nitrile gloves: StarGuard Comfort. In case of contamination, change gloves immediately. Transfer of methanol: Chemical protection glove: Ansell Barrier. Minimization- and Substitution-Check A substitution of carcinogenic nitrosamines is not possible. To minimize the workplace concentration: Nitrosamines are needed as a standard substance. By purchasing reference solutions, the possible risk of contamination is reduced as much as possible. Dienstgebäude: Zentral Diagnostik Seite 1 von 7 Weißenburger Str., 76187 Karlsruhe (07 1) 9 6-6 11 (07 1) 9 6-55 11 Telefon Poststelle@cvuaka.bwl.de (07 1) 9 6-5 49 (07 1) 9 6-55 9 Telefax www.cvua-karlsruhe.de; www.ua-bw.de Nach ISO 1705 akkreditiertes Prüflaboratorium, DAkkS-Reg.-Nr. D-PL-18866-0-00

Brief description The homogenized sample is suspended in methanol, extracted in an ultrasonic bath and diluted with water. The membrane-filtered sample is separated by ultra-high performance liquid chromatography (UHPLC) and analyzed by chemical ionization under atmospheric pressure and tandem mass spectrometry (UHPLC-APCI-MS/MS) in a multiple reaction monitoring mode (MRM). The quantification of NDMA and NDEA is performed using isotopic labeled internal standards according to the internal standard method. The qualitative confirmation of each substance is provided by ion ratios of two mass transitions. 4 Chemicals and materials 4.1 Reference materials CMR / toxic Substance Abbreviation CAS-No. X N-nitrosodimethylamine NDMA 6-75-9 X N-nitrosodiethylamine NDEA 55-18-5 4. Internal Standard CMR / toxic Substance Abbreviation CAS-No. X N-nitrosodimethyl-d 6 -amine NDMA-d6 1789-05-9 X N-nitrosodiethyl-d 10 -amine NDEA-d10 119794-54- 4. Further Chemicals CMR / toxic Substance Abbreviation CAS-No. X Formic acid LC/MS grade HCOOH 64-18-6 X Methanol HPLC grade MeOH 67-56-1 4.4 Required Solutions for LC-MS/MS Eluent A: Eluent B: 0.1 vol-% HCOOH in water dissolve 1 ml HCOOH in 1000 ml of ultrapure water MeOH HPLC grade 5 Devices 5.1 LC-MS/MS - UHPLC-System coupled with a tandem mass spectrometer with APCI source (i.e. Agilent Infinity 190 UHPLC + Agilent 6460 APCI-QQQ-MS) - Column: Waters HSS-T 100 x.0mm, 1.8 µm, 100 Å 5. Laboratory devices - brown glass GC-Vials - syringe attachment filter with 0.0 µm PET membrane - analytical balance, 0.1mg precision - vortex mixer Az.: vom 10.10.018 Seite von 7

- ultrasonic bath - 1-15 ml centrifuge tubes with plain bottom and plastic screw cap - 5 ml disposable syringes - different piston and/or direct displacement pipets for volumes of µl up to 10.000 µl - laboratory glassware (volumetric flasks, cylinders, solvent bottles etc.) 6 ProcedureCalibration 6.1.1 Stability / Storage Solution Storage Stability stock solutions refrigerator at least 1 year standard-/mix-/additional solutions refrigerator 1 year calibration and sample solutions room temperature at least 1 week internal standards (ISTD) refrigerator unlimited 6.1. Reference substances and solutions Substances and stock solutions NDEA and NDMA are purchased as solution (= stock solution): 5000 µg/ml in MeOH Standard solution From stock solutions: 0 µl of each stock solution / 10 ml MeOH (c = 10 µg/ml) Calibration solution: From standard solution: 500 µl / 10 ml of water (c = 500 ng/ml) 6.1. ISTD ISTD stock solutions NDMA-d6: approx. 10 mg / 10 ml MeOH (c = approx. 1000 µg/ml) NDEA-d10: approx. 10 mg / 10 ml MeOH (c = approx. 1000 µg/ml) ISTD solution From ISTD stock solutions: 5 µl of each stock solution/ 50 ml MeOH (c = 500 ng/ml) 6.1.4 Calibration working solutions Working range from 0 ppm up to.0 ppm Description Blank value + ISTD calibrationsolution ISTD-solution MeOH water C NDEA [ng/ml] C NDMA [ng/ml] 0 00 9500 0 0 K1 4 00 9496 0. 0. K 10 00 9490 0.5 0.5 Az.: vom 10.10.018 Seite von 7

K 0 00 9480 1 1 K4 40 00 9460 K5 00 900 10 10 K6 400 00 9100 0 0 K7 1000 00 8500 50 50 K8 0 00 7500 100 100 Concentration of internal standard: approx. 10 ng/ml in each case Working range from.0 ppm up to 0.0 ppm Description calibrationsolution ISTD-solution MeOH water C NDMA [ng/ml] Blank value + ISTD 0 0 0 950 0 K1 40 0 0 910 0 K 100 0 0 850 50 K 0 0 750 100 K4 00 0 0 650 150 K5 400 0 0 550 K6 600 0 0 50 00 Concentration of internal standard: approx. 10 ng/ml in each case NDEA concentration is not specified in that table although contained in the solutions as the method is not validated for this NDEA working range. 6. Sample preparation 6..1 Blank sample All equipment used for sample preparation and handling, e.g. centrifuge tubes, pipet tips, membrane filters, have to be checked for possible contamination with nitrosamines. Therefore a blank sample using all solutions and equipment (i.e. sample preparation according to 6.. without sample weighing) has to be analyzed. The prepared solution is transferred into a vial for the subsequent measurement to obtain a blank value. 6.. Sample preparation approx. 100 mg of a homogenized sample of the finished product are weighed into a plastic centrifuge tube addition of µl of ISTD solution addition of 00 µl of MeOH vortexing, followed by treatment for 5 minutes in an ultrasonic bath addition of 9.5 ml of ultrapure water vortexing, followed by treatment for 5 minutes in an ultrasonic bath ultracentrifugation of the sample followed by membrane filtration into a GC vial A quality assurance sample is regularly treated in the same way. Az.: vom 10.10.018 Seite 4 von 7

6. Chromatographic conditions The specified parameters are the default parameters for this method: o Column: HSS-T 100 x,0 mm, 1.8 µm, 100 Å o Column oven temperature: 0 C o Injection volume: 0 µl Elution gradient: Time [min] Flow [ml/min] Eluent A [%] Eluent B [%] 0 0.5 95 5.0 0.5 95 5 8.0 0.5 40 60 9.0 0.5 5 95 1 0.5 5 95 Stop Time: Post Time: 1 min 4 min Divert Time setting for the switchover valve: 0 min: to waste valve 1.5 min: to MS valve for NDMA and NDMA-d6 detection (time segment ) 4.0 min: to MS valve for NDEA and NDEA-d10 detection (time segment ) 8.5 min: to waste valve 6.4 Ionization conditions and data acquisition The mass spectrometer settings may vary depending on the used device and the device condition. Thus, examples for optimized settings are given subsequently. For the ionization of NDMA and NDEA separate source parameters are used and applied to the time segments and respectively. APCI source parameter: Parameter Time segment (NDMA) Time segment (NDEA) Gas Temp. 00 C 00 C APCI Heater 50 C 50 C Gas Flow 6 l/min 5 l/min Nebulizer 55 psi 45 psi Capillary 0 V 1500 V Corrona 8 µa 8 µa Az.: vom 10.10.018 Seite 5 von 7

MRM method: Name RT [min] precursor ion / product ion pair (transition) [m/z] Resolution Dwell time [ms] Fragmentor [V] Collision energy [V] CAV [V] NDMA.6 75 / 58* 75 / 4 75 / 44 7 7 7 9 17 1 NDMAd6.5 81 / 64 81 /46* 0 0 1 18 NDEA 7. 10 / 75* 10 / 47 10 / 9 76 76 76 9 17 1 NDEAd10 7.1 11 / 81 11/ 4* 81 81 9 17 *=Quantifier (in case of interferences, it is also possible to quantify by using another precursor ion / product ion pair (transition)) Comment: transition m/z 75-4 usually has a higher background (noise) 7 Data interpretation 7.1 Evaluation of the measured data 7.1.1 General Analysis is carried out by the integration of the peak areas of the respective mass traces and calculation according to the method of internal standard. 7.1. Qualifier At least a second precursor ion / product ion pair (transition) is used to verify the results (qualifier). The relative intensity of the quantifier / qualifier (qualifier ratio) from the calibration measurements is compared to the qualifier ratio of the samples (software determines the intensity ratio and issues the qualifier ratio ). The maximum accepted relative ion intensity tolerance is set as follows (taken from the Commission Decision (/657/EG) amending Directive 96//EG): Deviation +/- 0 % (Qualifier Ratio) 7.1. Calculation The NDMA and NDEA contents in the sample are calculated based on the following formula: nitrosamine [mg/kg] = X DF W 1000 Az.: vom 10.10.018 Seite 6 von 7

X = ng nitrosamine per ml measuring solution DF = dilution factor W = sample weight in g Furthermore the recovery rate obtained from the validation data is included in a final calculation step. 8 Remarks The method is validated for two working ranges for NDMA (0 ppm and 0 ppm) and one working range for NDEA (0 1.5 ppm). The limit of quantification (LoQ) and the limit of detection (LoD) is related to tablet mass, without considering measurement uncertainty: LoQ of NDMA: 0. ppm (working range 0 -.0 ppm). LoD of NDMA: 0.08 ppm (working range 0 -.0 ppm) LoQ of NDEA is 0.04 ppm (working range 0-1.5 ppm) LoD of NDEA is 0.0 ppm (working range 0-1.5 ppm) The validation data were obtained with valsartan film coated tablets as matrix. Az.: vom 10.10.018 Seite 7 von 7