Key Words Q Exactive Focus, Orbitrap, veterinary drugs, small molecule HRAM quantitation, small molecule HRAM screening, UHPLC, vdia

Similar documents
Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Direct Analysis using Paper-Spray Mass Spectrometry: Method Development for the Rapid Screening of Drugs of Abuse for Forensic Toxicology

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

of mass spectrometry

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Bilag 3 Appendix B: Function tests

ADVANCED MULTI-RESIDUE SCREENING IN VETERINARY DRUG ANALYSIS USING UPLC -ToF MS

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

An Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

How to screen and identify unexpected and unwanted compounds in food

MetWorks Metabolite Identification Software

HILIC Method Development in a Few Simple Steps

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Relative quantification using TMT11plex on a modified Q Exactive HF mass spectrometer

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Targeted protein quantification

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Automated and accurate component detection using reference mass spectra

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Introduction to Fourier Transform Infrared Spectroscopy

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Multi-residue analysis of pesticides by GC-HRMS

Optimizing a Generic Approach to Analyzing PPCPs in River Water

High-Resolution Accurate-Mass (HRAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization

High-Field Orbitrap Creating new possibilities

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Quan/Qual Analyses. Unmatched Confidence for. Thermo Scientific Q Exactive Orbitrap LC-MS/MS System. Identify Quantify Confirm

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

for XPS surface analysis

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzcloud Library

Fast, Effective XPS Point Analysis of Metal Components

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

IN QUALITATIVE ANALYSIS,

EPA Method 1694: Analysis of Pharmaceuticals and Personal Care Products in Water

Key Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS

Introduction to Fourier Transform Infrared Spectroscopy

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

The key to your laboratory s success

Improving Experimental Efficiency for Drug Metabolite Profiling and Identification

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Monitoring Protein PEGylation with Ion Exchange Chromatography

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

Ultrasensitive EPA Method 1694 with the Agilent 6460 LC/MS/MS with Jet Stream Technology for Pharmaceuticals and Personal Care Products in Water

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Determination of urea in ultrapure water by IC-MS/MS

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Analysis of a Verapamil Microsomal Incubation using Metabolite ID and Mass Frontier TM

Measurement of Metabolic Stability Using SIM and Identification of Metabolites by Data Dependent Full-Scan MS/MS and CNL Scanning

Enhancement of Linearity and Response in Charged Aerosol Detection

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

High Efficiency, Broad Scope Screening of Pesticides Using Gas Chromatography High Resolution Orbitrap Mass Spectrometry

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Transcription:

Variable Data-Independent Acquisition (vdia) Delivers High Selectivity and Sensitivity in Combined Targeted and Untargeted Analyses for Small Molecules Olaf Scheibner 1, Markus Kellmann 1, Charles Yang 2, Maciej Bromirski 1, 1 Thermo Fisher Scientific, Bremen, Germany; 2 Thermo Fisher Scientific, San Jose, CA, USA Technical Note 64283 Key Words Q Exactive Focus, Orbitrap, veterinary drugs, small molecule HRAM quantitation, small molecule HRAM screening, UHPLC, vdia Goal Forty-four multi-class veterinary drugs of known concentrations were employed to demonstrate how the use of a generic variable dataindependent acquisition (vdia) method with wide MS/MS precursor isolation windows achieves sensitivity and selectivity comparable to data-dependent MS 2 acquisition (using narrow isolation windows) in quantitative and qualitative small molecule applications. In addition, a full record of MS and MS/MS data for the measured sample fit for non-targeted and unknown screening purposes is delivered. setup, five isolation windows are set, covering the entire mass range of the preceding full scan. Figure 1 diagrams a representative vdia method setup. Figure 1B shows an alternative setup, covering a wider mass range. Up to eight isolation windows can be used to span the full-scan mass range. However, earlier studies indicate that five fragmentation windows represent an optimal compromise between sensitivity and selectivity on the one hand and scan speed on the other. 1 Introduction Untargeted screening approaches need data acquisition methods that gather as much MS and MS/MS information from a sample as possible, regardless of the nature of the sample or the primary analytic purpose. To date, combinations of full-scan measurements and wide-range fragmentation techniques like all-ion fragmentation (AIF) are commonly used methods in this approach. These scan modes fragment all ions in a single fragmentation event without precursor ion isolation and detect all fragment ions in a single mixed spectrum. As a result, they suffer from limitations in sensitivity, selectivity, and dynamic range compared to data-dependent acquisition methods where detected precursors are isolated with narrow isolation windows prior to fragmentation and detection. In this technical note, a new high-resolution, accuratemass (HRAM) scan mode, termed variable data-independent acquisition (vdia), which is available on Thermo Scientific Orbitrap -based instrumentation for screening and quantitation of known and unknown samples, is described and compared to standard datadependent acquisition (DDA) methods. vdia may utilize up to eight isolation windows ranging from 50 to 800 Da, covering the entire mass range of the full preceding scan. Typically, smaller windows are used for lower mass regions to increase dynamic range and therefore sensitivity; larger windows cover higher mass regions to improve the duty cycle. In a typical vdia acquisition Figure 1. Examples for vdia setup with five fragment scan windows, covering a mass range from m/z 100 to 1000 (A) and from m/z 100 to 1500 (B). Experimental Material Forty-four multiclass veterinary drug residues, listed in Table 1, were analyzed using a single standardized chromatographic and mass spectrometric method in extracts of muscle, kidney, milk, and plasma. For absolute quantification, standard samples with known concentrations of all 44 veterinary drug residues covering eight calibration points (from 100 pg/ml (ppt) to 500 ng/ml (ppb)) were prepared. For evaluation of the method, spiked matrix samples (muscle and kidney for antibiotics, milk for avermectins, and plasma for nitroimidazoles) were analyzed by HRAM LC-MS/MS.

2 Table 1. Components used for evaluation of the vdia approach. Mass Spectrometry Method A generic full scan with wide-isolation variable dataindependent acquisition (FS-vDIA) MS/MS method on a Compound Compound Abamectin* Marbofloxacine Thermo Scientific Q Exactive Focus MS system was Amoxicillin Ampicillin Cefalexin Metronidazole Metronidazole-OH Moxidectin used for all samples: Full Scan Resolution setting 70,000 (FWHM) at m/z 200 Cefalonium Nafcillin Mass range (m/z) 100-1000 Cefaperazone Oxacillin vdia Resolution setting 17,500 (FWHM) at m/z 200 Cefapirim Penicillin G Isolation windows (m/z) 100 205, 195 305, 295 405, Cefquinome Penicillin V 395 505, 495 1000 Chlorotetracycline Ronidazol Spray voltage 4.4 kv Ciprofloxacin Sarafloxacine Sheath gas 30.0 arb. Cloxacillin Sulfadiazine Aux gas 5.0 arb. Danofloxacin Sulfadimethoxin Capillary temp. 250 C Dapsone Sulfadimidin / Sulfamethazine Heater temp. 300 C Difloxacin Sulfadoxin RF-lens level 50 Dimetridazol Sulfamerazin HCD collision energy 35 ev Doramectin* Sulfamethoxazole Data Processing Doxycyclin Sulfamethoxypyridazine Data processing was performed using Thermo Scientific Enrofloxacin Sulfathiazole TraceFinder processing software version 3.2. For generation of extracted ion chromatograms, an extraction Eprinomectin Tetracycline window of 5 ppm was used. For non-targeted screening, a Erythromycine Flumequine Ipronidazol-OH Thiamphenicol Trimethoprim Tylosine built-in component and fragment m/z values database was used together with a spectral library, consisting of 1500 components each. Liquid Chromatography Method A generic LC method was run on a Thermo Scientific UltiMate 3000 XRS Quaternary Rapid Separation LC (RSLC) and used for all samples: Column Mobile phase A Mobile phase B Gradient Flow rate Total chromatographic cycle Thermo Scientific Accucore C18 aq 100 x 2.1 mm, 2.6 µm particle size (p/n 17326-102130) Water + 0.1% formic acid Acetonitrile + 0.1% formic acid 6 min gradient from 5% B to 95% B 300 µl/min 15 min Results and Discussion In the vdia approach, a full scan is followed by two to eight wide-isolation fragmentation scans, which together cover the same isolation range as the preceding full scan. Figure 1 shows two examples of a vdia method with five fragmentation scans. For each the applied isolation and scan ranges are shown. Example A covers a mass range from m/z 100 to 1000 and was used in this study. Example B shows an alternative setup, covering a scan range from m/z 100 1500 using the same number of scan windows. Earlier work has shown that the selectivity of the method increases as the size of the isolation ranges for the fragment scans is decreased, resulting in a higher number of scan ranges for the same mass range covered. 1 Since the overall cycle time for this setup increases with the number of fragmentation scan ranges, the same work showed that a setup with five fragmentation windows maintains enough scan speed to be suitable for fast chromatography. The setup used in this study resulted in an overall cycle time of 650 ms. The vdia approach described above bridges the gap between full scan-data-dependent MS 2 (FS-ddMS 2 ) experiments and full-range fragmentation scan modes such as AIF. FS-ddMS 2 experiments, where MS 2 scans are performed on targets of interest (present on an inclusion list) upon their detection in the full scan, are known to be very selective and sensitive with respect to the fragment ion information obtained. Retrospective FS-ddMS 2 data analysis for additional compounds of interest, however, is limited to full-scan quantitation by accurate mass without confirmation of identity by MS/MS. To maintain scan

speed, this method has to work with dynamic exclusion, so only one survey MS 2 scan is triggered per compound. As a result, no elution profiles can be extracted for the confirming fragment ions. Full-range fragmentation experiments like AIF, where fragments from all species present in the full scan are detected in a single MS 2 scan, have the advantage of collecting all possible full scan and MS/MS information for the sample. Thus, they are fully suitable for retrospective data analysis. However, dynamic range, selectivity, and achieved detection limits are limited as the number of ions fragmented per species is lower due to the combined nature of the analyses. The vdia experiment combines the advantages of both approaches by using smaller isolation windows for the fragmentation scans while still fragmenting all precursors from the preceding full scan. With this, a complete record of MS and MS/MS data is kept, so all compounds of interest can be processed even in retrospective manner and full elution profiles of all confirming ions for all components can be extracted. To evaluate the power of the new vdia approach, animal product samples (pig muscle, pig plasma, and cow milk), spiked with 44 veterinary drugs (Table 1), were measured with the FS-vDIA method described. The spike levels were 5 μg/kg for antibiotics in pig muscle matrix, 1 μg/kg for nitroimidazoles in pig plasma matrix, and 1 μg/kg for avermectins in cow milk matrix. As shown in Figure 2, the FS-vDIA method shows very good selectivity at low concentration levels, all analyte ions are confirmed with multiple fragment ions matching both with high mass accuracy as well as retention time of elution profiles. In contrast to data-dependent MS 2 methods, full elution profiles of all fragments are available, providing an additional option for data quality control since confirmation results can easily be checked for false positive hits on random spikes or elevated background signals. 3 Figure 2. Selectivity of selected component in matrix; A: ampicillin in pig muscle at 5 μg/kg; B: sulfadiazin in pig kidney at 5 μg/kg; C: ronidazol in pig plasma at 1 μg/kg; D: moxidextin in cow milk at 1 μg/kg.

4 In addition, a dilution series of a standard mixture of 44 veterinary drugs was analyzed to evaluate the linearity of calibration. The range of concentration levels was set from 0.01 μg/kg to 500 μg/kg. Linear calibration curves were obtained over the entire calibration range for all 44 compounds. Figure 3 shows the calibration curves of ampicillin, cefalonium, dimetridazol, and metronidazole as examples using the vdia method. Analytical samples are often used for both known screening as well as unknown screening purposes. To conduct a wide-range unknown screening, the data was processed against a 1500 component accurate-mass database and spectral library contained fully within TraceFinder software. In addition to the known spiked components, 30 components could be identified and confirmed by means of accurate mass, isotope pattern match, fragment ion match, and library search identification. As an example, cortisol (hydrocortisone) was identified with all four methods of identification and confirmation. Figure 4 shows the screening result in TraceFinder software for a plasma sample. Figure 3. Selected examples of linearity for different compounds in the calibration range of 0.01 500 μg/kg.

Technical Note 64283 Figure 4. Identification of cortisol (hydrocortisone) in a pig meat sample with all four stages of confirmation. Conclusion The vdia scan mode on the Q Exactive Focus MS serves as a generic acquisition method for diverse analytes, delivering a complete data record of the measured samples. It is perfectly suited to acquiring data for quantitative purposes while additionally offering uncompromised performance for non-targeted and unknown screening applications. Overcoming the limitations of full-range fragmentation techniques, vdia provides selectivity and sensitivity comparable to datadependent MS 2 measurements while generating a complete record of full scan and fragmentation data for each of the measured samples. References 1. Vogler, B., Master s Thesis, Institute of Organic Chemistry, University of Zürich, Zürich, Switzerland, 2013, 76. www.thermofisher.com 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 TN64283-EN 0716S Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 45 453 9100 Korea +82 2 3420 8600 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 Singapore +65 6289 1190 Spain +34 914 845 965 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752