Conformational Characterization of Calmodulin by Hydrogen Deuterium Exchange Mass Spectrometry

Similar documents
Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS

HDX-MS Overview. Pat Boyce Pharmaceutical and Life Sciences Marketing Team Manager Waters Europe Waters Corporation 1

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

UPLC Intact MASS Analysis Application Kit

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS

Analysis of Labeled and Non-Labeled Proteomic Data Using Progenesis QI for Proteomics

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

C HA R AC T E RIZ AT IO N O F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMOMETRIC APPROACH

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3

Routine MS Detection for USP Chromatographic Methods

PosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction

Improved Ion Mobility Separation of Protein Conformations in the Gas Phase with SYNAPT G2 HDMS

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

A Case of Pesticide Poisoning: The Use of a Broad-Scope Tof Screening Approach in Wildlife Protection

[application note] ACQUITY UPLC/SQD ANALYSIS OF POLYMER ADDITIVES. Peter J. Lee, and Alice J. Di Gioia, Waters Corporation, Milford, MA, U.S.A.

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

ImprovIng ADmE ScrEEnIng productivity In Drug DIScovEry

Hydrogen/Deuterium Exchange Mass Spectrometry: A Mini-Tutorial

PosterREPRINT AN INTERACTIVE PHYSICOCHEMICAL PROPERTY PROFILING SOFTWARE FOR EARLY CANDIDATE ANALYSIS IN DRUG DISCOVERY INTRODUCTION

Peptide Isolation Using the Prep 150 LC System

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

Q-TOF PREMIER DYNAMIC RANGE ENHANCEMENT AND ACCURATE MASS MEASUREMENT PERFORMANCE

APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEY WORDS. Simultaneous high-throughput screening of melamine (MEL) and cyanuric acid (CYA).

Solvent Flexibility for Size-Based Polymer Analysis Using the Advanced Polymer Chromatography (APC) System

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Confirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System

INTRODUCTION OVERVIEW

Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides

Probing Ca -dependent Conformational Changes in Calmodulin by H/D Exchange: Effect of Ionic Strength and Peptide Binding

POST-COLUMN ADDITION AS A TOOL TO ENHANCE PERFORMANCE IN MICROFLOW LC-MS

Stephen McDonald and Mark D. Wrona Waters Corporation, Milford, MA, USA INT RO DU C T ION. Batch-from-Structure Analysis WAT E R S SO LU T IONS

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Implementation of Methods Translation between Liquid Chromatography Instrumentation

ion mobility spectrometry IR spectroscopy

MALDI-HDMS E : A Novel Data Independent Acquisition Method for the Enhanced Analysis of 2D-Gel Tryptic Peptide Digests

[ instrument specifications ]

LC and LC/MS Column Selection Flow Chart

[ a ppl ic at ion no t e ]

Structure of the α-helix

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Multi-stage Mass Spectrometry up to MS 4 on a QTof System

SUPPLEMENTARY INFORMATION

UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations

STUDY OF THE CONFORMATION OF MYOGLOBIN ADSORBED ON NANOPARTICLES USING HYDROGEN/DEUTERIUM EXCHANGE MASS SPECTROMETRY

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Metabolic Phenotyping Using Atmospheric Pressure Gas Chromatography-MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Basic Principles for Purification Using Supercritical Fluid Chromatography

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Supplemental Information

Supporting Information

Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS)

Effects of Aqueous Sample Content and Aqueous Co-Solvent Composition on UPC 2 Separation Performance

Supplementary Figure 1

T H E ANA LYSIS O F DIOX INS AND F U R A NS USING H RG C- H IG H R E SO LU T IO N MS W IT H T H E AU T OS P EC- ULTIMA NT

[ xevo G2 tof ] Your Quantitative and Qualitative Questions Answered. Exactly.

Optimizing Peak Capacity in Nanoscale Trap-Elute Peptide Separations with Differential Column Heating

RESEARCH ARTICLE. Tuning a High Transmission Ion Guide to Prevent Gas-Phase Proton Exchange During H/D Exchange MS Analysis

UNIFI - Targeted and Untargeted Screening MS Software

Breaking the speed limit: Fast, high-resolution peptide and tryptic digest separations using fused-core particles

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Supporting Information

Staphylococcal Nuclease and Ubiquitin Local Folding Energies and Rates using PEPS-HDX-ESI- MS

EXPRESSION MORE PEPTIDES. MORE PROTEINS. MORE REPRODUCIBILITY. MORE QUANTIFICATION.

Translational Biomarker Core

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Use of High Speed/High Resolution Size-Based Chromatographic Separation of Polymeric Mixtures with Offline Infrared Detection

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Supporting information for Analysis of Cellular Responses of Macrophages to. Zinc Ions and Zinc Oxide Nanoparticles: A Combined Targeted and Proteomic

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Quality control analytical methods- Switch from HPLC to UPLC

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

[ XEVO G2-XS QTOF ] EXTRAORDINARY POSSIBILITIES FOR YOUR EVERYDAY ANALYSIS

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Transcription:

Conformational Characterization of Calmodulin by Hydrogen Deuterium Exchange Mass Spectrometry Joomi Ahn, Martha Stapels, Michael Eggertson, Keith Fadgen, and Ying Qing Yu Waters Corporation, Milford, MA, U.S. A P P L I C AT I O N B E N E F I T S INT RODUC T ION To determine changes in conformation between apo- and holo-calmodulin using hydrogen deuterium exchange combined with high resolution MS. Correct protein conformation is essential for proper biological function of protein therapeutics. Changes in protein conformation are a major concern in the biopharmaceutical industry and conformational characterization is a difficult task. Many analytical tools such as circular dichroism (CD), differential scanning calorimetry (DSC), and analytical ultracentrifugation (AUC) are used to study the higher-order structure of proteins. These methods sample global conformation and cannot determine where conformational changes occur. Nuclear magnetic resonance (NMR) and X-ray crystallography determine protein structure with high spatial resolution, but both technologies require substantial amounts of sample and are not routinely applied to biopharmaceutical products due to the significant time and effort required. Hydrogen deuterium exchange (HDX or HX MS) mass spectrometry has proven to be a useful analytical method for the study of protein dynamics and changes to protein conformation.1 Successful HDX studies require an LC/MS system that can perform rapid chromatographic separations at 0 C and make accurate mass measurements of small quantities of deuterium labeled proteins and peptides. When an HDX experiment is performed with UPLC separation technology and highresolution mass spectrometry, subtle conformational changes can be revealed and the locations of these changes are determined at the peptide level. Such innovative technology with a commercially available system solution makes conformational analyses a more practical method. Waters HDX Technology has been adopted as an Wat e r s s o lution s nanoacquity UPLC with HDX Technology Xevo QTof MS ProteinLynx Global SERVER Software k e y word s Hydrogen Deuterium Exchange Mass Spectrometry, HDX Technology, protein conformation Figure 1. nanoacquity UPLC System with HDX Technology. HDX Manager, Binary Solvent Manager (BSM), and Auxiliary Solvent Manager (ASM) are shown from top to bottom, respectively. The HDX Manager is a key component, which can perform the online pepsin digestion and maintain the low temperature at 0 C required for HDX MS. 1

E X P E R IM E N TA L Protein preparation: Bovine calmodulin purchased from Sigma Protein buffer solution: 0 mm HEPES in 100% H O ph 7.0 with and without 50 µm calcium 3 D O labeling solution: 0 mm HEPES in 99.99% D O at pd 7.0 with and without calcium 3 Quenching solution: 33 mm HCl 100 % H O ph.50 Method LC conditions: LC system: Columns: Trapping column: Desalting column: Waters nanoacquity UPLC with HDX Technology including HDX Manager ACQUITY UPLC BEH C 1 1.7-µm 1.0 x 100 mm for peptides ACQUITY UPLC BEH C 1.7-µm.1 x 50 mm for intact proteins ACQUITY VanGuard Pre-Column, BEH C 1,.1 x 5 mm MassPREP TM Desalting Column,.1 x 5 mm Column temp.: 0 C Online digestion column: Analytical flow rate: Immobilized pepsin column.1 x 30 mm from Applied Biosystems 0 µl/min Mobile phase A: 0.1 % formic acid in water, ph.1 Mobile phase B: 0.1 % formic acid in acetonitrile, ph.1 Gradient: 3 to 0 % B in 7 min Auxiliary mobile phase A: Trapping condition: MS conditions: MS system: 0.05 % formic acid in water ph.50 for digestion flow 3 min at 100 µl/min Waters Xevo QTof MS Ionization mode: ESI + Capillary voltage: Collision energy: Sampling Cone voltage: 3.0 kv V 35 V Source temp.: 0 C Desolvation temp.: 175 C Lock Mass: Glufibrinopeptide, 100 fmol/µl in 50% acetonitrile, 0.1% formic acid at 5 µl/min Data management: MassLynx.1 Software, ProteinLynx Global SERVER (PLGS) Software

effective analytical tool for biotech researchers who are studying higher-order structure of protein drugs. Characterizing higher-order structure is one of the key factors in drug discovery and early development phases and adopting this technology will offer a distinct advantage in understanding the safety and efficacy of biopharmaceuticals. In this application note, we introduce the nanoacquity UPLC System with HDX Technology (Figure 1). This system consists of a nanoacquity UPLC Binary Solvent Manager (BSM), Auxiliary Solvent Manger (ASM), and HDX Manager. The HDX Manager is capable of rapid online protein digestion, desalting, and highly resolving chromatographic separations at 0 C. Operating at 0 C is required to reduce the deuterium loss during analyses. With this system, subtle differences in deuterium uptake can be adequately determined with very reproducible HDX conditions. A Xevo QTof MS, utilizing MS E with ProteinLynx Global SERVER (PLGS) Software, provides an ideal tool for accurate mass analysis and reliable peptic peptide assignment. To demonstrate this technology, the conformational changes in bovine calmodulin (CaM) were studied. Calmodulin is an important intracellular calcium receptor and it is known that it undergoes a conformational change upon calcium binding. 3 This change regulates the biological activities of target proteins. Here we describe the HDX workflows using the nanoacquity UPLC System with HDX Technology at both whole protein and peptide levels to show how the conformational study is performed. RESULTS AND DISCUSSION Sample Holo and apo bovine calmodulin were prepared with and without calcium, respectively, in protein buffer solution. Both proteins were labeled with 0-fold dilution with D O labeling solution and incubated for 10 seconds, 1, 10, 0, and 0 minutes at room temperature. The labeling reaction was quenched by adding quenching solution to reduce the ph to.5. The quenched samples were immediately placed at 0 C. Global HDX analysis for intact protein Global HDX analysis provides information about overall conformational changes by comparing the deuterium uptake of control vs. analyte proteins. The intact HDX workflow is illustrated in Figure. First, the protein is labeled with deuterium and incubated for selected amounts of time between 10 seconds and hours. Protein in in H H O, O, ph ph 7.0 at at room temp Add 0-fold excess D O Protein labeling occurs For for various times Quench ph to.50, temp to 0 C OPTIONAL AUTOMATION Inside HDX Manager C at 0 C Quenched protein is injected into HDX manger at at ph.5 Global Analysis at Intact Level nanoacquity UPLC for fast desalting/protein separation Deuterium uptake determination ESI Q-Tof Xevo MS QTof MS Figure. Global HDX workflow for intact protein analyses. The labeled protein is quenched to ph.50, and directly injected into the HDX manager, where online desalting and separation is performed at 0 C. 3

During the exchange reaction in solution, amide hydrogens on the protein backbone exchange at a rate that depends on both solvent accessibility and hydrogen bonding. The deuterium uptake as a function of time will be different between control and analyte if there is a difference in higher-order structure. The exchange reaction is quenched with cold acidic (ph.50) buffer and then quickly injected into the HDX Manager. The temperature inside the HDX Manager is set at 0 C to minimize back-exchange. Back-exchange must be controlled because deuterium can exchange back to hydrogen in the 100 % H O LC solvents. Therefore, low temperature, acidic ph, and a rapid run time are needed to minimize back-exchange. The HDX Manager in conjunction with the nanoacquity UPLC System is designed to satisfy specific HDX requirements such as fast 0 C operation. The Waters HDX System can be fully automated for sample preparation (Figure, highlighted in dashed line). The laborious preparation steps such as labeling, time measurement, quenching, and injection are handled by robotics. This system configuration saves a significant amount of time for laboratories performing highthroughput analyses. Apo-calmodulin is known to be an open structure without bound calcium (Figure 3, left top panel). When calcium ions are bound to calmodulin (holo-), the protein adopts a more compact conformation (Figure 3, left bottom). The deuterium uptake curves display the relative deuterium level and compare the difference of apo and holo at each labeling time point (Figure 3, right panel). After structural conversion from apo to holo, a lower amount of deuteration was observed in holo calmodulin, indicating that fewer amide hydrogens were protected from exchange compared to the apo state. Figure 3. Deuterium uptake curves for intact apo and holo calmodulin with corresponding tertiary structures known from crystallography or NMR. The blue and red dots represent amide hydrogens that are available to exchange with deuterium in apo and holo, respectively. Lower deuterium uptake was observed in holo calmodulin due to its closed conformation.

Local HDX analysis for peptides To locate where deuterium has been incorporated in a protein, the labeled and quenched protein is digested into peptides with an acidic protease (pepsin). Pepsin digestion of an undeuterated sample is performed to create a peptide map. In this peptide workflow (Figure ), the HDX System performs an efficient pepsin online digestion and separates the peptides via UPLC with MS spectra collected in MS E mode from the Xevo QTof MS. Finally, ProteinLynx Global SERVER (PLGS) Software processes the MS E data to identify the peptides generated by pepsin. This UPLC/MS E analysis offers several unique advantages compared to typical HPLC/MS. Robust online pepsin digestion Rapid peptide separation at 0 C in less than 5 to minutes UPLC resolution to be able to separate complex peptide digest of large proteins Independent temperature control for digestion and separation E MS and PLGS for reliable peptic peptide ID Linear sequence coverage of 9% was observed in this study. Pepsin is active at low ph and maintains activity at low temperatures, which are ideal for HDX quench conditions. Pepsin also generates many overlapping peptides and short peptides helpful for localizing deuterium incorporation. The reproducible peptides of calmodulin by online pepsin digestion were used to construct a peptide map. These peptides shown in Figure 5 were found in all digestions in apo and holo samples. Identifying reproducible peptides is necessary because more robust analysis is possible when the deuterium uptake of the same peptide is compared in deuterated apo and holo samples. Undeuterated / HO Deuterated / DO Protein in H O, ph 7.0 at room temp. Protein labeling occurs For for various mes Quench ph to.50, temp to 0 C OPTIONAL AUTOMATION Inside HDX Manager Inside HDX Manager at C at 0 C Quenched protein is injected into HDX manger Manager at at ph ph.5.5 Local Analysis analysis at pep de Pep de level Level Online Pepsin Diges on ph.50 nanoacquity UPLC using 1.7 µm BEH column for fast separa on Pep de map PLGS Iden ty E for pep de ID Deuterium uptake HDX determina on Browser So ware ESI Q-Tof QTof MS E Figure. Peptide ID workflow. The peptide ID step is in gray and the HDX experiment is in red. The peptide ID step is needed to identify what peptides are produced during online digestion and to construct the coverage map. Figure 5. Calmodulin online pepsin digestion coverage map. Each bar under the sequence represents an identified peptide. These peptides were reproducibly observed (N=). 5

The deuterated protein was digested and separated at 0 C in same way as undeuterated protein (Figure, red arrows). Reproducible chromatographic separations were obtained throughout the labeling time-course (Figure, left panel). MS spectra showed the expected mass increase in m/z as a result of longer exposure to DO (Figure, right panel). The intensity weighted average mass of each isotopic envelope was found and compared to that of the undeuterated protein. Peptide HDX data interpretation Deuterium uptake into each peptic peptide was compared for apo and holo calmodulin. These data indicate that there was a change in deuterium uptake for certain peptides from apo to holo form. An example is shown in Figure 7. The highlighted arrows point out the different deuterium uptake in one peptide after 10 minutes of labeling. Figure. Reproducible chromatographic separations of calmodulin peptides following online pepsin digestion (left panel). The right panel shows the spectra of peptide (ADIDGDGQVNY) labeled from 10 seconds to 0 minutes. The m/z increases as a result of longer exposure to D O. Apo CaM HoloCaM 0 min 0 min Uptake from undeuterated 10 min 1 min 10 sec Undeuterated m/z 95 9 97 9 99 95 9 97 9 99 m/z Figure 7. Different deuterium uptake of the same peptide found in Apo and Holo CaM. At 10-minute labeling, this peptide (FKEAFSLF, 1-19) showed a higher deuterium uptake in Apo (left panel) than in Holo (right panel).

Peptide HDX analysis typically generates hundreds of peptic peptides depending on the size of protein, therefore manually determining the deuterium uptake of each peptide is a time-consuming process. With Waters software, data interpretation is automated, and the data processing time is significantly reduced. The measured deuterium incorporation can be plotted as deuterium uptake curves for each peptide (Figure ). These charts show where the difference /no differences in deuterium uptake occur for all peptides. Figure 9 is one of the representations of HDX results. The relative percent deuterium uptake was color coded on 3D structure, thus overall dynamics of apo and holo calmodulin can be easily compared over time. Three dimensional (3D) structures confirmed that those peptides with significant differences in deuterium uptake were located primarily the calcium binding region. 1-10 1-13 Relative Deuterium Level (Da) 1 APO HOLO 10 0 0.10 1.00 10.00 100.00 1000.00 Time (min) Relative Deuterium Level (Da) 10 APO HOLO 9 7 5 3 1 0 0.10 1.00 10.00 100.00 1000.00 Time (min) Yes No Is deuterium uptake different? 37-7 35-5 Relative Deuterium Level (Da) 9 APO HOLO 7 5 3 1 0 0.10 1.00 10.00 100.00 1000.00 Time (min) Relative Deuterium Level (Da) 9 APO HOLO 7 5 3 1 0 0.10 1.00 10.00 100.00 1000.00 Time (min) Figure. The Example of deuterium uptake curves for apo and holo calmodulin. Upper curves (overlapping peptides covering 1-10 and 1-13) illustrate that there was distinct change in deuterium uptake between apo and holo calmodulin. The lower curves (peptides 37-7 and 35-5) showed no change in deuterium uptake between apo and holo calmodulin. Every peptide represented in Figure 5 has its own uptake curve. 7

For example, the highlighted box on the left panel in Figure 9 indicated 3% deuterium uptake for apo calmodulin whereas the same peptide region in holo form only showed % deuterium uptake at 10 seconds of labeling time. Figure 10 is another example of data representation with corresponding 3D structures. This HDX heat map overlays the deuteration on the sequence. 3D structures obtained from X-ray crystallography or NMR are complementary to the HDX results if available. Apo Holo 10 s 1 min 10 min 1 hour hours 10 s 1 min 10 min 1 hour hours Figure 9. HDX heat map display for assessing conformational difference in localized region ( - 5). Two sets of colored bands of the heat map represent the apo (top band) and holo (bottom band) calmodulin of % deuterium uptake. Cooler color represents less exchange, thus greater protection from exchange. The 3D structures with black arrows are the corresponding regions of the displayed heat map (-5). Represented holo region shows much cooler color in mostly purple and dark blue in early time points (for example at 10 seconds) than in apo calmodulin. Figure 10. Deuterium uptake information represented on calmodulin apo and holo 3D structures (apo = 1CFD.pdb, holo = 1PRW.pdb). At 10 seconds labeling, two regions of calmodulin were compared in highlighted boxes. Overall, the apo structure shows more rapid exchange due to greater protein flexibility without calcium binding.

C O N C LUSIONS The Waters nanoacquity UPLC System with HDX Technology, when combined with high-resolution MS, was effectively used to determine changes in calmodulin conformation upon calcium binding. This system provides unique features that improve the quality of HDX analysis. HDX workflow for global and local HDX analyses. Robust online pepsin digestion Rapid reproducible UPLC separation at 0 C E Accurate identification of peptic peptides by MS and PLGS Automated HDX data processing Waters HDX Technology facilitates the study of protein conformation and protein dynamics at global and local levels. The information obtained helps researchers understand protein function and related biological activity. The HDX System is the informative analytical tool for biopharmaceutical drug discovery and development. References 1. Engen JR. Analysis of Protein Conformation and Dynamics by Hydrogen/ Deuterium Exchange MS. Anal. Chem. 009: 1; 770-75.. Wales TE, Fadgen KE, Gerhardt GC, Engen JR. High-Speed and High-Resolution UPLC Separation at Zero Degrees Celsius. Anal. Chem. 00: 0; 15-0. 3. Zhu MM, Rempel DL, Zhao J, Giblin DE, Gross ML. Proving Ca+-Induced Conformational Changes in Porcine Calmodulin by H/D Exchange and ESI-MS: Effect of Cations and Ionic Strength. Biochemistry. 003: ; 153-97. Solutions in practice Hydrogen Deuterium Exchange MS Analysis for Calmodulin Protein Conformation. Waters Technology Brief. 010: 70003en. Investigating Conformational Changes In Oxidazed Interferon Alpha B by Hydrogen/Deuterium Exchange Mass Spectrometry. Waters Poster. 010: 70003550en. Hydrogen/Deuterium Exchange Mass Spectrometry System for Capturing Conformational Changes of Biotherapeutics. Waters Poster. 010: 7000333en. Waters, UPLC, nanoacquity UPLC, VanGuard, and Xevo are registered trademarks of Waters Corporation. ProteinLynx Global SERVER, MassLynx, MassPREP, and The Science of What s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 011 Waters Corporation. Produced in the U.S.A. February 011 70003en AG-PDF Waters Corporation 3 Maple Street Milford, MA 01757 U.S.A. T: 1 50 7 000 F: 1 50 7 1990 www.waters.com