for Acclaim Mixed-Mode HILIC-1 Column

Similar documents
Acclaim Mixed-Mode WCX-1

for Acclaim Mixed-Mode WCX-1

Acclaim Mixed-Mode WAX-1 Columns

Acclaim HILIC-10 Columns

for Acclaim Trinity TM P1

Thermo Scientific. Acclaim Trinity P2. Column Product Manual. P/N: October Part of Thermo Fisher Scientific

C18 Column. Care & Use Sheet

LC Technical Information

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

LC and LC/MS Column Selection Flow Chart

High Performance Liquid Chromatography

High Performance Liquid Chromatography

8. Methods in Developing Mobile Phase Condition for C18 Column


[ CARE AND USE MANUAL ]

LC Column Troubleshooting. Do you have an equivalent column for my LC column?

HPLC Columns. HILICpak VT-50 2D MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia

penta-hilic UHPLC COLUMNS

High Performance Liquid Chromatography

Packings for HPLC. Packings for HPLC

FAQ's. 1 - Which column would be the most appropriate for my application?

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules

penta-hilic UHPLC COLUMNS

ACQUITY UPLC HSS Columns

High Performance Liquid Chromatography

LC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

Open Column Chromatography, GC, TLC, and HPLC

ProPac WCX-10 Columns

Shodex TM ODP2 HP series columns

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

ACQUITY UPLC Peptide BEH C 18, 130Å and 300Å Columns

New Developments in Surfactant Analysis Using HPLC

Reversed Phase Solvents

Determination of Polymer Modifier in Asphalt

PRODUCT MANUAL. ACCLAIM OA GUARD COLUMN 4.3 x 10 mm, P/N ACCLAIM OA ANALYTICAL COLUMN 4 x 250 mm, P/N x 150 mm, P/N

637. Thiamethoxam. HPLC method

Chemistry Instrumental Analysis Lecture 31. Chem 4631

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

Biochemistry. Biochemical Techniques HPLC

IonPac Trace Cation Concentrator

It s hot! ZIC -chilic

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Determination of trace anions in concentrated hydrofluoric acid

Analysis - HPLC A.136. Primesep 5 µm columns B.136

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000

Basic principles of HPLC

Optimisil HPLC & UPLC Column catalogue

Alltech Alltima HP Introduction

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

Nexera UC Unified Chromatography

[ CARE AND USE MANUAL ] XSELECT HSS XP 2.5 µm COLUMNS I. INTRODUCTION CONTENTS I. INTRODUCTION II. GETTING STARTED III. COLUMN USE

Chromatography. Gas Chromatography

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

P R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns

[ Care and Use Manual ]

COSMOSIL Applications

IonPac SCS 1 IonPac SCG 1

Comparison of different aqueous mobile phase HPLC techniques

Determination of Caffeine by HPLC

Too Polar for Reversed Phase What Do You Do?

HPLC analysis of VFAs, Furfural and HMF

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Dionex IonPac AS28-Fast-4µm column

An HPLC column offering moderate retentivity with superb peak shape Inertsil ODS Sprint

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

GPC/SEC Practical Tips and Tricks. Thomas Dent Applications Scientist Agilent Technologies. October, 2011 Gulf Coast Conference

[ Care and Use Manual ]

HPLC Praktikum Skript

ProntoSIL C18-EPS Reversed-Phase HPLC Columns

Guide to Choosing and Using Polymer Reversed Phase Columns!

Method Development Kits

Uptisphere CS Evolution Core Shell columns for fast & highly efficient identification & quantification of small molecules

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Cation Exchange HPLC Columns

Fast Separation of Vastly Different Compounds by Isocratic HPLC

Dilution(*) Chromatography

H 3 CO H 3 CO S CH 3

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Ch.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)

Liquid Chromatography

Gemini-NX Performance paired with Axia Technology

Pure Chromatography Consumables Pure flexibility. Pure specialization. Pure convenience.

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

UPLC Method Development and Validation

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

"Theory and Practice of High Speed Chromatography for Bioanalysis" Stuart Coleman March 20, 2007

Successfully Scaling and Transferring HPLC and UPLC Methods

PYRIPROXYFEN TECHNICAL

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

HICHROM. Chromatography Columns and Supplies. LC COLUMNS Zorbax. Catalogue 9. Hichrom Limited

A C18 Silica Column With Exceptional Temperature and ph Stability

zorbax Packed by Phenomenex Material Characteristics High efficiency Guaranteed quality Phenomenex columns packed with Zorbax

M > ACN > > THF

HPLC COLUMNS WILEY-VCH. Theory, Technology, and Practice. Uwe D. Neue with a contribution from M. Zoubair El Fallah

Transcription:

for Acclaim Mixed-Mode HILIC-1 Column

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 1 of 17 Product Manual for ACCLAIM Mixed-Mode HILIC-1 Column 5µm, 4.6 x 250mm, P/N 066844 5µm, 4.6 x 150mm, P/N 066843 5µm, 2.1 x 150mm, P/N 066847 3µm, 3.0 x 150mm, P/N 070090 3µm, 3.0 x 50mm, P/N 071912 3µm, 2.1 x 150mm, P/N 070091 ACCLAIM Mixed-Mode HILIC-1 Guard 4.3 x 10mm (set of 2), P/N 066845 2.1 x 10mm (set of 2), P/N 066846 3.0 x 10mm (set of 2), P/N 071913 4.6 x 10mm (set of 2), P/N 069706 2009 Dionex Corporation Document No. 065198 Revision 02 July 2009

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 2 of 17 TABLE OF CONTENTS SECTION 1 Introduction... 3 1.1. Specifications and Recommended Operating Conditions...3 1.2. Physical...4 1.3. Acclaim Mixed-Mode HILIC-1 Products...4 SECTION 2 INSTALLATION: Step-by-Step User Guide... 5 SECTION 3 Considerations in Method Development... 7 3.1. HILIC Mode vs. Reversed-Phase Mode...7 3.2. Selection of Organic Solvents...7 3.3. Buffer Types...7 3.4. Mobile Phase ph...7 3.5. Isocratic or Gradient Method...7 SECTION 4 Column Care... 8 4.1. Column storage...8 4.2. Operating ph range: ph 2.5 to 7.5...8 4.3. Operating temperature limit: 50 ºC...8 4.4. Pressure limit < 4000 psi...8 4.5. Flow rate...8 4.6. Column washing procedure...8 SECTION 5 Frequent Asked Questions... 9 SECTION 6 Applications... 10 6.1. Dual Operation Modes: Reversed-phase and HILIC...10 6.2. Analysis of Urea...11 6.3. Complementary Selectivity: Reversed-phase and HILIC...12 6.4. Dependency of Retention on Mobile Phase Organic Content: Comparison between the Acclaim HILIC and LiChrosorb Diol...13 6.5. Analysis of NEODOL 25-12...14 6.6. Analysis of Alkylphenol Ethoxylate...15 6.7. Separation of Polyethylene Glycols...16 SECTION 7 Troubleshooting Guide... 17 7.1. High Backpressure...17 7.1.1. Finding the Source of High System Pressure...17 7.1.2. Contaminated guard or analytical column...17 7.2. High Background or noise...17 7.2.1. Preparation of Mobile Phases...17 7.3. Poor Peak Resolution...17 7.3.1. Loss of column Efficiency...17

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 3 of 17 SECTION 1 INTRODUCTION The Acclaim Mixed-Mode HILIC-1 column is a silica-based packing material that incorporates both Reversed-Phase (RP) and Hydrophilic Interaction Liquid Chromatography (HILIC) properties. Unlike either traditional RP or HILIC stationary phases, the packing features an alkyl long chain with a hydrophilic polar terminus, and demonstrates great potentials for separating a wide range of both highly polar and non-polar molecules, in either RP mode or HILIC mode. The Acclaim Mixed-Mode HILIC-1 columns consist of 5-µm or 3-µm high-purity, porous, spherical silica particles with 120Å diameter pores bonded with proprietary functional groups, which provide unsurpassed resolution and peak symmetry for a variety of polar and non-polar molecules. The main features of the Acclaim HILIC-1 column include 1. Operates in both reversed-phase and normal phase modes. 2. Retains highly polar molecules that would be un-retained by reversed-phase chromatography. 3. Unique selectivity, complementary to reversed-phase columns. 4. Higher hydrophobic retention compared to the conventional Diol columns. 5. Broader application range than conventional Diol columns. 1.1. Specifications and Recommended Operating Conditions Shipping Solution Storage Solution: Buffer ph Range: ph 2.5-7.5 Temperature Range < 50 C 60 / 40 Acetonitrile / 100mM Ammonium Acetate 60 / 40 Acetonitrile / 100mM Ammonium Acetate or HILIC mobile phase Stationary Phase Particle size Column Dimensions P/N Max Recommended Pressure Typical Flow Rate Mixed-Mode HILIC-1 3 µm 5 µm 2.1x150 mm 070091 5800 psi 0.2-0.5 ml/min 3.0x50 mm 071912 4500 psi 0.4-1.0 ml/min 3.0x150 mm 070090 5800 psi 0.4-1.0 ml/min 2.1x150 mm 066847 5800 psi 0.2-0.5 ml/min 4.6x150 mm 066843 5800 psi 0.8-2.0 ml/min 4.6x250 mm 066844 5800 psi 0.8-2.0 ml/min

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 4 of 17 1.2. Physical Bonding Chemistry: Proprietary alkyl diol Silica Substrate: Spherical, high-purity Particle size 5 µm and 3µm Surface area 300 m2/g Pore size 120 Å 1.3. Acclaim Mixed-Mode HILIC-1 Products Acclaim Mixed-Mode HILIC-1 Analytical Particle size 3 µm 5 µm Guard 5µm Column P/N Dimensions 2.1x150 mm 070091 3.0x50 mm 071912 3.0x150 mm 070090 2.1x150 mm 066847 4.6x150 mm 066843 4.6x250 mm 066844 4.3 x 10mm 066845 Requires Holder p/n 059456 2.1 x 10mm 066846 Requires Holder V-2 p/n 069580 3.0 x 10mm 071913 Requires Holder V-2 p/n 069580 4.6 x 10mm 069706 Requires Holder V-2 p/n 069580

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 5 of 17 SECTION 2 INSTALLATION: STEP-BY-STEP USER GUIDE Step 1 - Validating column performance Dionex recommends that you perform an efficiency test on your Acclaim Mixed-Mode HILIC column before use. The purpose of column performance validation is to ensure no damage has occurred during shipping. Test the column using the conditions described on the Quality Assurance (QA) Report enclosed in the column box. Repeat the test periodically to track the column performance over time. Note that slight variations may be obtained on tow different HPLC systems due to system electronic, plumbing, operating environment, reagent quality, column conditioning, and operator technique. Step 2 - Mobile phase preparation Obtaining reliable, consistent and accurate results require mobile phases that are free of ionic and spectrophotometric impurities. Chemicals, solvents and de-ionized water used to prepare mobile phase should be of the highest purity available. Maintaining low trace impurities and low particle levels in mobile phases helps to protect your columns and system components. DIONEX cannot guarantee proper column performance when the quality of the chemicals, solvents and water used to prepare the mobile phase has been compromised. De-ionized Water The de-ionized water used to prepare the mobile phase should be Type 1 Reagent Grade Water, or HPLC Grade Water. The de-ionized water should be free of ionized impurities, organics, microorganisms and particulate matter larger than 0.2 µm. Many commercial water purifiers are designed for HPLC applications and are suitable for these applications.! NOTE Degas the aqueous component of the mobile phase and then add the solvent component. Avoid excessive purging or degassing of mobile phases containing solvents, if possible, since the volatile solvent can be 'boiled' off from the solution. Solvents The solvents used must be free from ionic and UV-absorbing impurities. Use of ultrahigh purity solvents, HPLC grade, will usually ensure that your chromatography is not affected by impurities in the solvent. Depending on specific application, the mobile phase system consists of an organic modifier (e.g. acetonitrile or methanol) and an aqueous portion (e.g. D.I. water, or ammonium acetate or phosphate buffer). Both pre-mixed and proportioning valve generated mobile phases give satisfactory results. The use of proportioning valve provides better flexibility in method optimization, while the pre-mixed mobile phase provides more reproducible results. Mobile Phase for Column Performance Test: Pre-mixed mobile phase: mix 400 g of D.I. water and 468 g of acetonitrile. Proportioning valve generated: set up the gradient pump as 40/60 v/v D.I. water/ acetonitrile.! These two mobile phases could give slightly different results due to the ways they are prepared. NOTE Step 3 - Set up the LC system Use a standard LC system equipped with a LC pump, a column oven, a UV detector, and an injector (or an autosampler). The system should be thoroughly primed before use.

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 6 of 17 Step 4 - Condition the column Each new column is shipped in the mobile phase used for the column performance test. Depending on different separation modes, the column should be conditioned differently. For HILIC mode (>70% acetonitrile in the mobile phase) applications, the column should be purged with 50% acetonitrile in D.I. water (v/v) for approximately 10 column volumes, followed by equilibration with the desirable mobile phase thoroughly (~20 to 50 column volumes depending on the aqueous content in the mobile phase) before any injection is made. For RP mode (<70% acetonitrile in the mobile phase) applications, the column should be purged with 50% acetonitrile in D.I. water (v/v) for approximately 10 column volumes, followed by equilibration with the desirable mobile phase thoroughly (~20 column volumes) before any injection is made. When switching to a different mobile phase, make sure that the new mobile phase is compatible with the existing mobile phase in the column to avoid column clogging due to precipitation. A good practice is to purge the column with 50% acetonitrile in D.I. water (v/v) for approximately 10 column volumes before switching to the new mobile phase. Step 5 - Reproduce the chromatogram in the Lot Validation Report Perform the column QA test using the conditions described in the QAR (Appendix), and compare the result with the reported values. The column should be fully equilibrated before any injection. At least three injections should be made until reproducible result is obtained.! Due to various reasons, such as difference of LC systems, mobile phases, oven temperature control, etc, you may observe somewhat different separation from that in the report. NOTE Once you are satisfied with the column performance report result, the column is ready for real applications.

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 7 of 17 SECTION 3 CONSIDERATIONS IN METHOD DEVELOPMENT 3.1. HILIC Mode vs. Reversed-Phase Mode The selection of separation mode depends on the type of the analyte molecules and requirement of the specific application. In general, if a highly polar molecule is the analyte of interest, HILIC mode (high organic in the mobile phase) should be considered. In addition, HILIC mode can be used to determine the degree of ethoxylation (EO) of a nonionic surfactant. For other applications, RP separation may be considered. However, it is recommended that a conventional RP column (C18, C8, or polar-embedded phase) be tried first for regular RP applications. The Acclaim HILIC column has lower hydrophobicity than C8 columns but much higher hydrophobic retention than conventional Diol columns because of its unique surface chemistry, so that selectivity different from RP columns is possible. 3.2. Selection of Organic Solvents For HILIC applications, acetonitrile is preferred as the organic modifier. For RP applications, both acetonitrile and methanol can be used depending on the application. In addition, this column is compatible with traditional normal-phase solvent, such as ethanol, iso-propanol, ethyl acetate, hexane, heptane, etc. 3.3. Buffer Types The selection of buffer depends on the detection method and ph requirement. Ammonium acetate (or formate) buffer is the preferred buffer system because of its applicability to both RP and HILIC separation modes, compatibility with UV (> 230 nm), ELS detector and MS, high solubility in organic solvent, familiarity to most chromatographers. Volatile organic acids, such as acetic acid, formic acid, and TFA, can also be used to control the mobile phase ph, and share the similar benefits of ammonium acetate (formate) buffer. Phosphate buffers are ideal for applications that require low UV background, and it usually provides somewhat better peak shapes for charged molecules, compared to acetate buffer. Although there is no problem in RP application (organic content < 70%), phosphate buffers tend to precipitate in high organic condition. 3.4. Mobile Phase ph Mobile phase ph needs to be controlled to adjust selectivity and obtain reproducible results of charged molecules. 3.5. Isocratic or Gradient Method Isocratic methods are suitable for simple and/or well-defined application. When dealing with unknown samples, or a sample consisting of molecules with dramatically different hydrophilicity or hydrophobicity, a gradient method is often advantageous.

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 8 of 17 SECTION 4 COLUMN CARE 4.1. Column storage The column can be stored in the mobile phase for short period of time (e.g. overnight). If not in use for longer than one week, it is recommended to store the column in a solution with higher organic content, such as 70/30 v/v acetonitrile (or methanol)/20-100 mm ammonium acetate or D.I. water at a ph between 3.5 and 5.5. 4.2. Operating ph range: ph 2.5 to 7.5 The column lifetime depends heavily on the chromatographic condition. To obtain better column lifetime, it is highly recommended to use "silica friendlily" mobile phases, such as a buffer at a ph between ph 3 to 7. Although compatible with 100% aqueous mobile phase, it is not recommended to use this column extensively in such condition. The mobile phase ph should be controlled between 3.5 and 6 if more than 80% buffer is present in the mobile phase. If used in HILIC conditions, the column can be used in ph 2 to 8 range. 4.3. Operating temperature limit: 50 ºC Although our experimental results indicate that the column can be used at 50 C, a lower operating temperature (20-40 C) is recommended for routine separations. 4.4. Pressure limit It is extremely important not to impose sudden column pressure surge. The maximum recommended pressures are listed in section 1.1, the use of lower operating pressures will extend column lifetime. 4.5. Flow rate The typical operating flow rates are listed in section 1.1, higher flow rate can be used for faster analysis as long as the pressure limit is not exceeded. 4.6. Column washing procedure All samples should be pre-treated and filtered before being injected on the column. In the event of column washing practice is needed, the following procedure can be used as a guideline: For a 4.6 mm i.d. column: 1. Wash the column with D.I water /acetonitrile v/v 50/50 for 10 column volumes at a flow rate between 0.5 to 1 ml/min. 2. Wash the column with 0.1 M ammonium acetate ph 4 to 5 /acetonitrile v/v 50/50 for 20 column volumes at a flow rate between 0.5 to 1 ml/min. 3. Wash the column with 0.1 M ammonium acetate ph 4 to 5 /acetonitrile v/v 10/90 for 20 column volumes at a flow rate between 0.5 to 1 ml/min. 4. Wash the column with 0.1 M ammonium acetate ph 4 to 5 /acetonitrile v/v 50/50 for 20 column volumes at a flow rate between 0.5 to 1 ml/min. 5. Before any injection is made, the column should be equilibrated with the mobile phase thoroughly.! For a 2.1 mm i.d. column, the flow rate should be reduced to 20% of that for 4.6 mm i.d column. NOTE! If above treatment fails to revive the column, the column should be replaced. NOTE

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 9 of 17 SECTION 5 FREQUENT ASKED QUESTIONS 1. What is the difference between the Acclaim Mixed-Mode HILIC-1 column and conventional HILIC (Diol) column? The Acclaim HILIC-1 column is based on a new silica-based mixed-mode stationary phase that incorporates both RP and HILIC properties. Unlike either traditional RP or HILIC stationary phases, the new packing features an alkyl long chain with a hydrophilic polar terminus, and demonstrates broader applications for separating a wide range of both highly polar and nonpolar molecules, in either RP mode or HILIC mode (Figure 1). By comparison, conventional Diol columns have significantly decreased hydrophobicity (Figure 2), and have much narrower application range. 2. Why do I need an Acclaim Mixed-Mode HILIC-1 column? The Acclaim HILIC-1 column combines both RP and HILIC characteristics. It can be used in either HILIC conditions or RP conditions the same way as conventional Diol columns or RP columns, without special requirements. More importantly, the optimal balance between the hydrophilic and hydrophobic portions on the column surface leads to unique selectivity that neither conventional Diol columns nor RP columns can provide. 3. When do I need an Acclaim Mixed-Mode HILIC-1 column? You should consider using an Acclaim HILIC-1 column when 1) Regular RP (e.g. C18, C8, polar-embedded phase, etc) and HILIC (e.g. Diol, CN, amino, silica) columns fail to provide satisfactory results; 2) You are dealing with highly polar molecules, or very hydrophobic molecules; 3) You need orthogonal selectivity to complement the primary method; 4) You are analyzing ethoxylated surfactants (Figures 6-8). 4. What factors should I consider for method development using this column? During method development, the following factors should be considered: 1) Separation mode 2) Type of organic modifier 3) Aqueous content in the mobile phase 4) Mobile phase ph control 5) Temperature 5. What mobile phases should I use with this column? The new column is compatible with any mobile phases for HILIC or Normal-Phase separations provided that the ph requirement is met. The new column is compatible with any mobile phases for Revered-Phase separations provided that the ph requirement is met. However, for typical RP applications, a RP column (C18 or C8) should be tried first. Please refer to "Section 3 Considerations in Method Development" for more details. 6. What should I do before starting using Acclaim Mixed-Mode HILIC-1 column? Read this User Guide carefully, and contact Dionex Technical Support if you have any questions regarding using this column. 7. Can I use this column in normal-phase mode? Yes. The column is compatible with traditional normal-phase solvent, such as ethanol, iso-propanol, ethyl acetate, hexane, heptane, etc. 8. How to store the column? The column can be stored in the mobile phase for short period of time (e.g. overnight). If not in use for longer than one week, it is recommended to store the column in a solution with higher organic content, such as 70/30 v/v acetonitrile (or methanol)/20-100 mm ammonium acetate or D.I. water at a ph between 3.5 and 5.5.

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 10 of 17 SECTION 6 APPLICATIONS 6.1. Dual Operation Modes: Reversed-phase and HILIC 1 Reversed Phase Mode 2 AU 2 1 HILIC Mode 0 2 4 6 8 10 Minutes Column: Acclaim Mixed-Mode HILIC-1, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: CH 3 CN/0.1 NH 4 OAc, ph5.2 v/v 52/48 for RP mode; v/v 92/8 for HILIC mode Temperature: 30 C Flow Rate: 1 ml/min Inj. Volume: 10 µl Detection: UV @ 254 nm Peaks: (100 ppm each) 1. Cytosine 2. Naphthalene

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 11 of 17 6.2. Analysis of Urea 95/5 v/v CH 3 CN/0.1 M NH 4 OAc, ph4.5 mv 97/3 v/v CH 3 CN/0.1 M NH 4 OAc, ph4.5 97.5/2.5 v/v CH 3 CN/0.1 M NH 4 OAc, ph4.5 0 2 4 6 8 10 Minutes Column: Acclaim Mixed-Mode HILIC-1, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: See chromatograms for details Temperature: 30 C Flow Rate: 1 ml/min Inj. Volume: 10 µl Detection: Sedex 85 ELS detector (gain - 9, 50ºC) Sample: Urea (0.1% in 90% CH 3 CN)

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 12 of 17 6.3. Complementary Selectivity: Reversed-phase and HILIC 2 Acclaim 120 C8 CH 3 CN/0.1 NH 4 OAc, ph5.2 v/v 60/40 3 1 AU 1 2 Acclaim Mixed-Mode HILIC-1 CH 3 CN/0.1 NH 4 OAc, ph5.2 v/v 97.5/2.5 3 0 4 8 12 16 20 Minutes Column: 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: CH 3 CN/0.1 NH 4 OAc, ph5.2 Temperature: 30 C Flow Rate: 1 ml/min Inj. Volume: 5 µl Detection: UV @ 254 nm Peaks: 1. Toluene 2. Caffeine 3. Benzoic acid

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 13 of 17 6.4. Dependency of Retention on Mobile Phase Organic Content: Comparison between the Acclaim HILIC and LiChrosorb Diol Benzoic Acid Caffeine Toluene 20 20 40 Retention (minutes) 16 12 8 4 Retention (minutes) 16 12 8 4 Retention (minutes) 35 30 25 20 15 10 5 0 0 20 40 60 80 100 0 0 20 40 60 80 100 0 0 20 40 60 80 100 %CH 3 CN %CH 3 CN %CH 3 CN Column: Acclaim Mixed-Mode HILIC-1, LiChrosorb Diol, 5 μm Mobile Phase: CH 3 CN/0.1 NH 4 OAc, ph5.2 Flow Rate: 1 ml/min Detection: UV @ 254 nm Dimensions: 4.6 x 150 mm Temperature: 30 C Inj. Volume: 5 µl

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 14 of 17 6.5. Analysis of NEODOL 25-12 Acclai m HILIC C12 C13 C14 C15 Acclaim 120 C8 C12 C13 C14 C15 0 4 8 12 16 20 R(OCH 2 CH 2 ) n OH R = C 12 to C 15 n ~ NEODOL 25 12 Minutes Column: 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: A - CH 3 CN, B - D.I. H 2 O Gradient: 30% to 80% A in 15 min,then hold for 5 min Temperature: 30 C Flow Rate: 1.5 ml/min Inj. Volume: 10 µl Detection: ELS Detector Sample: NEODOL 25-12 (0.2%)

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 15 of 17 6.6. Analysis of Alkylphenol Ethoxylate Acclaim 120 C8 AU LiChrosorb Diol C 8 H 17 (OCH 2 CH 2 ) n OH Acclaim HILIC 0 6 12 18 24 30 Minutes n~9 IGEPAL CA-630 Column: See chromatograms for details, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: 99/1 v/v CH 3 CN/0.1 M NH 4 OAc, ph5.2 Temperature: 30 C Flow Rate: 1.0 ml/min Inj. Volume: 10 µl Detection: UV @ 225 nm Sample: IGEPAL CA-630 (0.1%)

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 16 of 17 6.7. Separation of Polyethylene Glycols PEG-6000 mv PEG-3400 PEG-2000 PEG-400 PEG-1000 0 4 8 12 16 20 Minutes Column: Acclaim HILIC, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: A - MeOH, B - D.I. H 2 O Gradient: 20% to 95% A in 20 min Temperature: 30 C Flow Rate: 1 ml/min Inj. Volume: 25 µl Detection: ELSD Samples: Various PEGs (0.04% each)

Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 17 of 17 SECTION 7 TROUBLESHOOTING GUIDE The following instruction should help you to locate and eliminate problems traceable to hardware and chemistry issues. Please keep in mind that some problems may be due to other reasons, such as sample contamination, poor water quality, etc. If you cannot solve your problems with the help of this manual, please contact you local Dionex customer support specialist. 7.1. High Backpressure 7.1.1. Finding the Source of High System Pressure If the system pressure is excessively high, determine the cause of the high pressure. The system should be used with a highpressure in-line filter for mobile phases. The filter should be positioned between the gradient pump pressure transducer and the injection valve. Make sure you have a high-pressure in-line filter in place and that it is not contaminated. A. Make sure that the pump is set to the correct flow rate. Higher than recommended mobile phase flow rates will cause higher pressure. Measure the pump flow rate if necessary with a stop watch and graduated cylinder. B. Determine which part of the system is causing the high pressure. It could be a piece of tubing that has plugged, collapsed tubing walls from over tightening, an injection valve with a plugged port, a column with particulates plugging the bed support, a plugged High-pressure In-line filter, or detector cell. To identify which part of the chromatographic system is causing the problem, disconnect the pump fluid line from the injection valve and turn the pump on. Watch the pressure; it should not exceed 50psi (0.34MPa). Continue adding the system components back into the fluid path, one by one, while watching the system pressure. 7.1.2. Contaminated guard or analytical column If the column is the cause of the back pressure, replace the guard column. 7.2. High Background or noise 7.2.1. Preparation of Mobile Phases A. Ensure the mobile phase is made correctly. B. Ensure the eluents are made from chemicals with the recommended purity. 7.3. Poor Peak Resolution Poor peak resolution can be due to any or all of the following factors. 7.3.1. Loss of column Efficiency A. Extra-column system effects can result in sample band dispersion, decreasing peak efficiencies. Make sure you are using tubing with an i.d. of no greater than 0.010" (0.007" for 2.1mm) to make all liquid line connections between the injection valve and the detector cell inlet on standard bore (4-mm) systems. Check for leaks. B. Guard column / column needs to be replaced.