The Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction

Similar documents
OBSERVATIONS ON THE WETTING OF REVERSED-PHASE HPLC PACKINGS

HPLC Determination of Antidepressants and their Metabolites in Plasma

[ Care and Use Manual ]

A Unique, New, Multipurpose Sorbent Engineered for Reversed-Phase SPE

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Solid Phase Extraction Method Development Tips and Tricks

CHAPTER CHROMATOGRAPHIC METHODS OF SEPARATIONS

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Dilution(*) Chromatography

SPE Method Development Tips and Tricks. Group/Presentation Title Agilent Restricted Month ##, 200X

Yuhui Yang, An Trinh, Michael Ye, and Tom Henderson 595 North Harrison Road, Bellefonte, PA T HKB

LC Technical Information

RediSep Rf C18 Flash Column Loading Techniques

EVOLUTE WCX Columns for Solid Phase Extraction of Quaternary Amines and Strongly Basic Compounds from Aqueous Samples

Development for SPE. Ron Majors Senior Chemist, CSD Joan Stevens SPE Application Chemist October 2009

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

SPE. Simply Efficient! Watchers bond Solid Phase Extraction by ISU. Ver

ProntoSIL C18-EPS Reversed-Phase HPLC Columns

CHROMABOND HLB. Introduction. Advantages of CHROMABOND HLB. Technical data

Alltech Alltima HP Introduction

A New Water-Wettable Solid Phase Extraction Sorbent for the Analysis of Drugs in Biological Fluids

penta-hilic UHPLC COLUMNS

SPE Introduction. general chromatography. solid phase extraction. How to Choose an SPE Product. Solid-Phase Processing Methods

Packings for HPLC. Packings for HPLC

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000

Sample Prep Options for GC users

Silanol. Groups Liquid Chromatography

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Chromatography. Gas Chromatography

Strategies for the Flash Purification of Highly Polar Compounds

A New Polymer-based SPE Sorbent to Reduce Matrix Effects in Bioanalytical LC-MS/MS

Chapter 23 Introduction to Analytical Separations

P R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns

RP C18 column with feature of a silanol group

High Performance Liquid Chromatography

penta-hilic UHPLC COLUMNS

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

BEKOlut - YOUR PARTNER FOR SMART SPE SOLUTIONS

Basic Principles for Purification Using Supercritical Fluid Chromatography

A guide to solid phase extraction

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Characterization of Carbon SPE for the Extraction of Polar Analytes

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Fast Separation of Vastly Different Compounds by Isocratic HPLC

Reversed Phase Solvents

ProPac WCX-10 Columns

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns

LC and LC/MS Column Selection Flow Chart

Biochemistry. Biochemical Techniques HPLC

Liquid Chromatography

Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV*

Shodex TM ODP2 HP series columns

Dual Use of a Chemical Auxiliary: Molecularly Imprinted Polymers for the Selective Recovery of Products from Biocatalytic Reaction Mixtures

Analysis - HPLC A.136. Primesep 5 µm columns B.136

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Ensure reproducible Polymeric SPE workflow for high-throughput bioanalysis applications

Pure Chromatography Consumables Pure flexibility. Pure specialization. Pure convenience.

Liquid Chromatography

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

William E. Barber, Ph.D. Applications Chemist October

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

SPEReference Manual & Users Guide

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Changing the way you think about HPLC

LIQUID CHROMATOGRAPHY

Technical Guide Thermo Scientific Syncronis HPLC Columns

HPLC Background Chem 250 F 2008 Page 1 of 24

Acclaim Mixed-Mode WCX-1

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.

Minimizing Solvent Impact on Purification of Nitrogencontaining

HICHROM. Chromatography Columns and Supplies. SAMPLE PREPARATION Solid Phase Extraction - SOLA & HyperSep. Catalogue 9.

Hypersil BDS Columns TG 01-05

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Inertsil ODS-EP Technical Information

CHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.

8. Methods in Developing Mobile Phase Condition for C18 Column

Electronic Supplementary Information. The selective response of templated polymer for the cationic drug

Orosil HPLC Columns. OroSil HPLC columns are designed for the separation of polar, semi-polar, and nonpolar compounds at low to medium ph.

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

for Acclaim Mixed-Mode HILIC-1 Column

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Chiral Flash Columns

Optimisil HPLC & UPLC Column catalogue

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Determination of trace anions in concentrated hydrofluoric acid

Fluophase and Fluofix Columns

Chemical Analysis Problem

Too Polar for Reversed Phase What Do You Do?

for Acclaim Mixed-Mode WCX-1

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

NEVIRAPINE ORAL SUSPENSION Final text for addition to The International Pharmacopoeia (February 2009)

HPLC COLUMNS WILEY-VCH. Theory, Technology, and Practice. Uwe D. Neue with a contribution from M. Zoubair El Fallah

(HILIC) Bill Champion Agilent Technologies, Inc opt 3/opt3/opt 2 HILIC - Agilent Restricted

Capital HPLC Columns and HPLC Cartridges. Optimal HPLC Range

Part 2. Overview of HPLC Media

Bulletin 910. Guide to Solid Phase Extraction. Introduction 1

Transcription:

The Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction Edouard S. P. Bouvier, Randy E. Meirowitz and Uwe D. Neue Waters Corporation, 34 Maple Street, Milford, MA 01757 USA Presented at HPLC'97 (Birmingham, UK) Poster #P-112/A

Abstract Solid phase extraction (SPE) is a routinely used chromatographic technique for sample enrichment and sample clean-up. Processing aqueous samples by SPE is predominantly performed using cartridges containing C18-bonded silica. Before a sample can be applied onto the cartridge, the sorbent requires a conditioning step with a water-miscible solvent. The purpose of this step is to wet the surface of the highly porous stationary phase support. This wetting depends on the contact angle of the solvent/sorbent/air system. Recently, we have introduced a porous polymeric sorbent, poly(divinylbenzene-co-n-vinylpyrrolidone) for reversed-phase solid phase extraction. This sorbent possesses greater water-wetting properties than traditional C18-silica sorbents. We have measured wetting properties for several hydro-organic mixtures on both C18-silica and the polymeric resin. We have correlated the results to performance of the sorbents in solid phase extraction.

Introduction Solid phase extraction (SPE) and thin layer chromatography (TLC) are two examples of techniques in which a fluid is passed through a column or bed initially containing dry porous particles. In SPE, a conditioning step is typically performed to fully wet the porous sorbent before the sample is applied. One common theory that has been postulated to explain why this is necessary is that methanol "solvates" the alkyl-bonded phase. As a result, these C8 or C18 groups, which now have methanol associated with them, become fully extended and can thus interact better with the aqueous sample. Another theory, postulated here, is that aqueous liquids cannot penetrate into the pores of reversed-phase sorbents because of the unfavorable contact angle that the liquid has with the sorbent. Over 99.9% of the surface area of a porous silica resides within the pores, so failure to penetrate into the pores will prevent adequate adsorption of analyte. Liquid can only enter these pores if (1) the contact angle between the liquid and the surface is less than 90, or (2) the external pressure is great enough to overcome the capillary pressure. When performing SPE, a vacuum (~0.4 bar) is typically used to effect flow. Under such conditions, a non-wetting liquid (<90 ) will not penetrate the pores.

Contact Angle Definition θ θ θ < 90 wetted surface θ > 90 non-wetted surface A surface is wetted by a solvent when the contact angle is less than 90. Wetting of pores occurs without external pressure when θ < 90.

Pore wetting on C18-silica for water and methanol Equation of Young and Laplace: Pc = 4 d γ cosθ where : Pc = intrusion pressure γ = surface tension d = capillary diameter θ = contact angle water methanol For water on C18-silica, θ > 90. Water does not penetrate pores. For methanol on C18-silica, θ < 90. Methanol fully wets pores.

Pressure Required for Water to Penetrate Pores at Different Contact Angles 500 Capillary pressure required (bar) 400 300 200 100 95 100 120 110 0 90 0 50 100 150 200 Pore size [Å]

SPE Procedure Important considerations: Conditioning solvent must fully wet pores (i.e., intrusion contact angle must be < 90 ). Equilibration solvent, sample and wash solvents must not allow liquid to extrude from pores (i.e.,extrusion contact angle must be < 90 ). [NOTE: Due to hysteresis, contact angle for extrusion may be less than contact angle for intrusion.] If conditioning solvent is removed from pores before application of equilibration solvent, then equilibration solvent must have intrusion contact angle < 90 to re-wet pores. Condition cartridge with water-miscible organic solvent (e.g., methanol or acetonitrile) Equilibrate cartridge with water or aqueous buffer Load sample onto cartridge Wash off weakly retained interferences with weak solvent Elute product with strong solvent Evaporate, reconstitute and analyze

Waters Oasis HLB Sorbent: A Water-Wettable Reversed-Phase Copolymer N O Hydrophilic monomer Lipophilic monomer water loving Provides wetting properties Reduces contact angle with water fat loving Provides reversed-phase property for analyte retention

Effect of Drying on Recovery: C 18 % Recovery for C18 Cartridge 100 90 80 70 60 50 40 30 20 10 0 0 0 1 2 3 4 5 6 7 8 9 0 2 4 6 8 10 Drying Time [min] Drying Time [min] Procainamide Acetaminophen Ranitidine Propranolol Doxepin vs. Oasis HLB Cartridges Vacuum drying after conditioning removes methanol. Water cannot wet pores. Recovery is adversely affected. % Recovery for Oasis HLB Cartridge Cartridges were conditioned with 1 ml methanol, and then either (a) kept wetted, or (b) dried under vacuum for 10 minutes (0.4 bar), prior to equilibration with 1 ml water. 1 ml spiked serum was then loaded, followed by 1 ml water. Analytes were eluted with 1 ml methanol, and analyzed by HPLC. 100 80 60 40 20 Water can wet pores even if methanol is removed. Recovery is unaffected by drying.

SPE Recoveries - Effect of Cartridge Drying C 18-100 mg DVB - 50 mg Porapak RDX - 50 mg HLB - 50 mg Procainamide Acetaminophen Ranitidine Caffeine Toluamide sorbent kept wet sorbent dried 10 min Toluidine Propranolol Dipropylphthalate 0 20 40 60 80 100 0 20 40 60 80 100 0 20 40 60 80 100 0 20 40 60 80 100 Percent Recovery Cartridges were conditioned with 1 ml methanol, and then either (a) kept wetted, or (b) dried under vacuum for 10 minutes (0.4 bar), prior to equilibration with 1 ml water. 1 ml spiked phosphate-buffered saline solution was then loaded, followed by 1 ml water. Analytes were eluted with 1 ml methanol, and analyzed by HPLC.

Comparison of Dye Retention: HLB vs. C 18 Shown in picture are two cartridges that have been conditioned with 1 ml of methanol and then dried for 5 minutes with vacuum (0.4 bar). 1 ml of 20 mm phosphate buffer, ph 7.0, containing phenol red was then loaded onto each cartridge. Cartridge on left (Oasis HLB) retains dye. Cartridge on right (C18) does not. Oasis HLB C 18

Can an SPE Sorbent be Too Hydrophobic? After wetting a sorbent with methanol and equilibrating with water or buffer, one of two phenomena can occur. Either (1) the pores in the hydrophobic sorbent remain wetted with water, or (2) the water extrudes from the sorbent's pores. We postulate that the extrusion contact angle is the driving factor here. Because of hysteresis (e.g., due to H-bonding with surface silanols), the extrusion contact angle may be significantly less than the intrusion contact angle. If the extrusion contact angle for water is <90, then the pores will remain wetted. In this case, SPE performance will be unaffected if the sorbent dries on the vacuum manifold out after applying the aqueous equilibration solvent. On the other hand, if the extrusion contact angle is greater than 90, the water will be extruded from the pores, even if the sorbent bed is kept wet. In this case, low recoveries will result. We have synthesized several different C 8 -silica bonded phases to measure performance in SPE. They vary in degree of surface coverage. Physical properties and surface coverage are shown in the next slide. Also included are chromatographic retention comparison for several compounds. For this data, 3.9x150 mm columns were slurry packed, and each standard was individually injected.

Sorbent # C 8 -Silica Phases Tested: Effect of Bonding Density and Endcapping %MeOH for wetting (θ<90 ) Chemistry Endcap? Coverage µmol/m² 1 C8 no 2.7 55-60 2 C8 yes 3.3 55-60 3 C8 no 3.4 55-60 4 C8 yes 3.6 60-65 C8 bondings were carried out on 37-55 µm Porasil. Methanol wetting was determined by observing if the sorbent wetted after mixing with different methanol/water mixtures; a wetted sorbent sinks, a non-wetted sorbent floats. For SPE, 1 ml/50 mg cartridges were conditioned with 1 ml methanol, followed by 1 ml water. After loading with 1 ml sample, and washing with 1 ml water, analytes were eluted with 1 ml methanol. NOTE: The sorbent was not allowed to dry out during the SPE procedure. Chromatographic retention factors (k) in 50 mm potassium phosphate, ph 7.0 with 30% methanol. Sorbent # Chemistry Acetaminophen Procainamide Ranitidine Caffeine Toluamide 1 C8 0.51 1.76 1.73 1.99 3.16 2 C8 0.41 0.27 0.26 1.02 3.10 3 C8 0.38 0.24 0.24 0.94 2.95 4 C8 0.38 0.24 0.24 0.86 3.04

SPE Recovery on C8-Silica: Effect of Bonding Coverage and Endcapping Percent recovery 100 80 60 40 20 0 Increasing hydrophobicity Sorbent coverage endcap? #4 C8 high yes #3 C8 high no #2 C8 low yes #1 C8 low no Toluamide Caffeine Ranitidine Procainamide Acetaminophen The unendcapped C8-silica sorbent ( lowest coverage) gives the best performance in SPE, even though the chromatographic data indicates that retention is comparable to the more hydrophobic phases, particularly for neutral compounds.

Conclusions Wetting properties play an important role in SPE performance. A contact angle less than 90 is required for liquid to penetrate pores. For C18-silica and water, θ > 90. SPE performance will be adversely affected if the sorbent dries out after the initial conditioning step. A novel reversed-phase sorbent has been designed that has θ < 90 with water. SPE performance is unaffected if sorbent dries out. RP-silica sorbents with high bonding coverage may give poor SPE performance. Water may extrude from pores, even if pores have been previously wetted.