Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Similar documents
Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Methanol Extraction of high level soil samples by USEPA Method 8260C

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Optimizing. Abstract: is standardd. procedures. altered to

Solid Phase Micro Extraction of Flavor Compounds in Beer

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Maximizing Production While Minimizing Costs

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Optimizing Standard Preparation

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Analytical Trap Comparison for USEPA Method 8260C

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Solid Phase Micro Extraction of Flavor Compounds in Beer

Maximizing Sample Throughput In Purge And Trap Analysis

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

EPA TO-17 Volatile Organic Compounds

2017 EPA Method Update Rule and EPA Method 624.1

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Ed George and Anaïs Viven Varian, Inc.

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Determination of Volatile Organic Compounds in Air

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Volatile Organic Compounds in Every Day Food

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Target Compound Results Summary

STANDARD OPERATING PROCEDURES

CALA Directory of Laboratories

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Copies: Dave Favero, RACER File

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Volatile Organic Compounds in Water PBM

Thermal Desorption Technical Support

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

Associates. Project No.: October 25, 2017

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

Method 1624 Revision C Volatile Organic Compounds by Isotope Dilution GCMS

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Date Issued: April 01, 2013 Expiration Date: June 30, 2013

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

Study of Residual Solvents in Various Matrices by Static Headspace

Laboratory Report Number(s) Lone Tree Road (Deep) 10/5/15, 10/19/15, 6/13/16. Address Sample Date(s) Analyses Requested

If you have any questions or concerns, please do not hesitate to contact our office at

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

317 Elm Street Milford, NH (603) Fax (603)

Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc

Headspace Technology for GC and GC/MS: Features, benefits & applications

Table 1 Soil and Groundwater Sampling Summary Table

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

METHOD 5030C PURGE-AND-TRAP FOR AQUEOUS SAMPLES

Transcription:

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler Application Note By: Teri Dattilio Introduction Purge and Trap concentration (P&T) along with Gas Chromatographic (GC) analysis is a widely used method for the analysis of Volatile Organic Compounds (VOCs). This methodology was developed to achieve the high sensitivity necessary to determine VOCs in drinking water and other matrices according to United States Environmental Protection Agency (USEPA) Method 8260 as well as other SW8000 Methods. To accommodate the requirements of USEPA Method 8260, Teledyne Tekmar has continually improved P&T technology by introducing new generations of enhanced concentrators into the analytical arena. The new Stratum PTC offers innovative U-shaped analytical and condensate traps, providing improvements over earlier P&T concentrators, and therefore demonstrating excellent data results for many current USEPA methodologies. In addition to introducing the Stratum PTC and U-shaped traps, Teledyne Tekmar has developed a proprietary #9 trap for improved chromatographic performance on early eluting compounds found in many of the aforementioned methods. In this study soil sample analysis was performed and linear calibration was demonstrated for 95 target analytes over the range of 1.0-200ppb. A 10mL purge volume was utilized, using conditions and specifications outlined in USEPA Method 8260. The Stratum PTC and SOLATek 72 in conjunction with an Agilent 6890 GC and 5973 MSD are excellent tools for the identification and quantification of VOCs in soil matrices. Experimental-Instrument Conditions An Agilent 6890/5973 GCMS, Stratum Purge and Trap Concentrator and SOLATek 72 Multi-Matrix Autosampler were used for this analysis. Results and findings were obtained through the use of a 20m X 0.18mm X 1.0µm RTX-VMS fused silica capillary column (Restek Corporation). The Mass Spectrometer Detector (MSD) scanned in the full scan mode from 35-350m/z at 5.27 scan/sec. The GC, MSD and Purge and Trap conditions are shown in Tables 1, 2 and 3 respectively. GC Parameters MSD Parameters GC: Agilent 6890 MSD: Agilent 5973 Column: Restek RTX-VMS, 20m, 0.18mm ID, 1.0µm Source: 230 C Oven Program: 35 C for 4 min; 16 C/min to 85 C for 0 min: 30 C /min to 210 C for 3 min 14.29 min runtime Quad: 150 C Inlet: 150 C Solvent Delay: 0.5 min Column Flow: 1.2mL/min Column Flow: 1.2mL/min Gas: Helium Scan Range: mz 35-350 Split: 30:1 Scans: 5.27 scans/sec Flow: 1.2mL/min Threshold: 400 Pressure: 20psi Tables 1 & 2: GC and MSD Parameters www.teledynetekmar.com

Stratum PTC and SOLATek 72 Parameters Variable Value Variable Value Rinse Water Temp 90 C Sample Preheat Time 1.00 min Sample Cup Temp: 40 C Sample Temp 40 C Sample Needle Temp 60 C Purge Temp 0 C Transfer Line Temp 125 C Purge Flow 0mL/min Soil Valve Temp 125 C Condenser Ready Temp 40 C Sample Sweep Time 0.50 min Condenser Purge Temp 20 C Needle Rinse Volume 7mL Dry Purge Time 3.00 min Needle Sweep Time 0.75 min Dry Purge Temp 20 C Sample Preheat Time 0.00 min Dry Purge Flow 50mL/min Preheat Stir Off GC Start Start of Desorb Preheat Stir Mode Spin Desorb Preheat Temp 245 C Preheat Stir Speed 1 Desorb Drain On Purge Time 11 min Desorb Time 2.00 min Purge Stir On Desorb Temp 250 C Purge Stir Mode Spin Desorb Flow 300mL/min Purge Stir Speed 5 Bake Time 4.00 min Valve Oven Temp 150 C Bake Temp 270 C Transfer Line Temp 150 C Bake Flow 400mL/min Sample Mount Temp 90 C Condenser Bake Temp 175 C Purge Ready Temp 45 C Focus Temp -150 C Standby Flow 25mL/min Inject Time 1.00 min Pre-Purge Time 0.50 min Inject Temp 180 C Pre-Purge Flow 40mL/min Standby Temp 100 C Sample Heater Off Table 3: Stratum PTC/SOLATek 72 Parameters Stratum PTC Parameters are indicated with the blue background Calibration A working standard was prepared in methanol at a final standard concentration of 50ppm. Calibration standards were prepared in 100mL volumetric flasks filled to volume with deionized water over the calibration range of 1.0-200ppb and 10mL aliquots were transferred directly to VOA vials for analysis. Internal Standards (IS) were added at 5µL using the Internal Standards Addition Module of the SOLATek 72 Autosampler to hold at a constant concentration of 25ppb. Calibration data was processed using Agilent ChemStation software. All analytes and their corresponding calibration data were evaluated using both the %RSD of the relative response factor and by calibration curve linearity. For most compounds the %RSD was <12% over the entire 1.0-200ppb-calibration range. For those target compounds with %RSD >15%, linear regression was employed with acceptance at 0.995 or greater, indicating linear response for all target analytes. The calibration data meets all USEPA Method 8260 performance criteria. Calibration data along with results of a 50ppb Continuing Calibration Verification (CCV) standard are presented in Table 4. Figure 1 shows the total ion chromatogram of a 50ppb standard. Method Detection Limit (MDL) Determination A study was performed to statistically determine the Method Detection Limits (MDL s) according to the procedure in USEPA Method 8260. Seven aliquots of a 1.0ppb standard were analyzed and the data processed to determine the MDL s for the compounds listed in Table 4 below. The detection limit results for most of the compounds were 0.5µg/L or less. The data collected met system performance criteria for Method 8260. www.teledynetekmar.com Page 2 of 5

MDL according to 40CFR 136, Appendix B, Revision 1.11 Spike level Compound %RSD MDL (ug/l) Pentafluorobenzene (IS) 25.0 0.000 50ppbCCV (%DEV) % Carryover Dichlorodifluoromethane 1.00 12.65 0.404 89 0 Chloromethane 1.00 10.77 0.330 89 0.12 Vinyl Chloride 1.00 4.66 0.357 97 0.03 Bromomethane 1.00 10.33 0.330 94 0.20 Chloroethane (Ethyl Chloride) 1.00 11.47 0.471 96 0.03 Trichlorofluoromethane 1.00 9.19 0.400 93 0.02 Diethyl Ether 1.00 3.87 0.381 99 0.00 1,1-Dichloroethene 1.00 5.11 0.484 95 0 Carbon Disulfide 1.00 9.49 0.310 94 0.31 1,1,2-Trichlorofluoroethane 1.00 11.21 0.501 93 0.01 (F ) Iodomethane 1.00 14.71 0.228 113 0.28 Allyl Chloride 1.00 13.81 0.404 90 0.08 Methylene Chloride 1.00 8.6 0.357 92 0.06 Acetone 1.00 0.9999 0.710 99 0.30 trans-1,2-dichloroethene 1.00 7.65 0.388 92 0.16 Methyl Acetate 1.00 7.89 0.431 91 0.13 MTBE 1.00 3.63 0.335 97 0.00 TBA 1.00 9.31 0.384 94 0 Diisopropyl Ether 1.00 7.21 0.313 97 0 Chloroprene 1.00 7.33 0.346 96 0.05 1,1-Dichloroethane 1.00 5.02 0.366 95 0.01 Acrylonitrile 1.00 10.07 0.417 93 0.26 Vinyl acetate 1.00 12.17 0.496 104 0.02 ETBE 1.00 2.64 0.292 98 0 cis-1,2-dichloroethene 1.00 4.48 0.384 95 0.08 2,2-Dichloropropane 1.00 9.33 0.400 92 0 Bromochloromethane 1.00 6.25 0.303 92 0.05 Chloroform 1.00 4.12 0.435 94 0.01 Carbon Tetrachloride 1.00 2.5 0.228 97 0.00 1,1,1-Trichloroethane 1.00 6.68 0.469 95 0.00 THF 1.00 6.3 0.388 96 0 Dibromofluoromethane 1.00 9.03 0.466 94 0.00 (S t ) Methyl Acrylate 1.00 5.34 0.356 95 0 1,1-Dichloropropene 1.00 2.71 0.304 98 0.05 2-Butanone (MEK) 1.00 10.2 0.391 88 0 Benzene 1.00 6.97 0.420 94 0.04 Propionitrile 1.00 4.47 0.317 94 0.10 tert Amyl Methyl Ether 1.00 2.84 0.113 98 0.00 (TAME) 1,2-Dichloroethane 1.00 10.7 0.511 92 0.08 Isobutyl Alcohol 1.00 11.11 0.324 83 0 Isopropyl Acetate 1.00 7.96 0.299 106 0.00 Trichloroethene 1.00 5.87 0.388 104 0.11 1,4-Difluorobenzene (IS) 25.00 0.000 Dibromomethane 1.00 8.63 0.149 95 0.06 1,2-Dichloropropane 1.00 5.06 0.205 96 0.02 Bromodichloromethane 1.00 3.68 0.222 95 0.02 Methyl Methacrylate 1.00 0.9999 0.737 99 0 Stratum PTC Purge and Trap Concentrator SOLATek 72 Multi-Matrix Autosampler www.teledynetekmar.com Page 3 of 5

MDL according to 40CFR 136, Appendix B, Revision 1.11 Spike level 50ppbCCV % Compound %RSD MDL (ug/l) (%DEV) Carryover n-propyl Acetate 1.00 13.2 0.132 108 0.00 2-Cleve 1.00 0.9999 0.147 105 0 cis-1,3-dichloropropene 1.00 10.91 0.127 104 0.04 Toluene-d8 (surr) 1.00 10.94 0.213 100 0.03 Toluene 1.00 13.19 0.164 98 0.04 Tetrachloroethene 1.00 11.93 0.503 100 0.07 4-methyl2-pentanone 1.00 12.07 0.394 104 0.00 1,1,2-Trichloroethane 1.00 8.69 0.404 94 0.00 Ethyl Methacrylate 1.00 9.77 0.779 106 0 Dibromochloromethane 1.00 5.39 0.561 101 0.01 1,3-Dichloropropane 1.00 7.23 0.212 96 0.02 1,2-Dibromoethane 1.00 4.34 0.112 97 0.03 n-butyl Acetate 1.00 11.56 0.643 107 0.00 2-Hexanone 1.00 11.42 0.521 103 0.1 Chlorobenzene-d5 (IS) 25.00 0.000 Chlorobenzene 1.00 11.4 0.236 95 0.09 Ethylbenzene 1.00 13.58 0.196 100 0.05 1,1,1,2-Tetrachloroethane 1.00 6 0.269 98 0 M&P Xylene 1.00 10.42 0.129 92 0.05 Ortho Xylene 1.00 13.68 0.165 109 0.03 Styrene 1.00 12.41 0.069 100 0.04 Bromoform 1.00 5.23 0.239 103 0 Isopropylbenzene 1.00 11.72 0.137 105 0.02 n-amyl Acetate 1.00 9.55 0.149 107 0 BFB (surr) 1.00 8.24 0.090 99 0.08 n-propylbenzene 1.00 10.28 0.163 97 0.04 cis-1,4-dichloro-2-butene 1.00 5.15 0.430 99 0 Bromobenzene 1.00 9.87 0.186 97 0.09 1,1,2,2-Tetrachloroethane 1.00 12.35 0.204 90 0.00 1,3,5-Trimethylbenzene 1.00 0.9999 0.157 128 0.03 2-Chlorotoluene 1.00 11.29 0.173 100 0.04 trans-1,4-dichloro-2-butene 1.00 5.15 0.368 99 0.04 4-Chlorotoluene 1.00 12.25 0.134 97 0.09 Tertbutylbenzene 1.00 14.55 0.231 104 0.03 1,2,4-Trimethylbenzene 1.00 13.96 0.175 106 0.02 sec-butylbenzene 1.00 10.29 0.185 98 0.03 p-isopropyltoluene 1.00 12.88 0.167 104 0.02 1,3-Dichlorobenzene 1.00 12.17 0.199 93 0.13 1,4-Dichlorobenzene-d4 (IS) 25.00 0.000 1,4-Dichlorobenzene 1.00 13.18 0.187 93 0.17 n-butylbenzene 1.00 12.07 0.145 101 0.07 1,2-Dichlorobenzen 1.00 11 0.175 95 0.08 1,2-Dibromo-3-Chloropropane 1.00 14.98 0.302 107 0 Nitrobenzene 1.00 12.12 1.046 107 0 Hexachlorobutadiene 1.00 11.26 0.227 96 0.13 1,2,4-Trichlorobenzene 1.00 9.58 0.160 102 0.16 Napthalene 1.00 10.83 0.235 105 0.09 1,2,3-Trichlorobenzene 1.00 12.34 0.302 94 0.14 Table 4: Calibration %RSDs and Statistically Determined Method Detection Limits for 8260 Target Compounds Stratum PTC Newly Designed U- shaped Analytical and Condensate Traps www.teledynetekmar.com Page 4 of 5

Carryover Evaluation Figure 1: Total Ion Chromatogram of a 50ppb Calibration Standard Data was collected and evaluated for carryover of target analytes in subsequent blanks following a 200ppb calibration standard. The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler performed remarkably well. Most target compounds had carryover <0.4% in the first blank for all 95-target analytes evaluated. This data is also presented in Table 4. Conclusion The need for water removal from Purge and Trap analysis has been present since the introduction of Purge and Trap technology. The Stratum PTC is equipped with an innovative U-shaped condensate trap. The unique geometry of the trap aids in the removal of water that is typical in Purge and trap analysis. The new condensate trap offers improved water management and therefore a great replacement to early generation concentrators. Increasing demands for low-level sensitivity for VOC analysis has led the need for improved Purge and Trap technology. The new proprietary Tekmar #9 analytical trap performs well on troublesome early eluting compounds as well as difficult to retain compounds like 2-chlorovinyl ether and nitrobenzene. Using the # 9 trap, analysis of volatile organic compounds by USEPA Method 8260 was demonstrated. Linear calibration was also demonstrated over the range of 1.0-200ppb for 95 target compounds. The new Siltek coated sample pathway proves to be the optimal choice for the reduction of carryover for pathway sensitive compounds such as halogenates and others. The new Stratum PTC and SOLATek 72 Multi-Matrix Autosampler prove to be excellent analytical instruments. Teledyne Tekmar once again continues to offer excellent water management, improved analytical performance and a reduction in carryover in subsequent samples all while satisfying the requirements of multiple EPA and other analytical methods. www.teledynetekmar.com Page 5 of 5