Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Similar documents
Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Methanol Extraction of high level soil samples by USEPA Method 8260C

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Solid Phase Micro Extraction of Flavor Compounds in Beer

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Optimizing. Abstract: is standardd. procedures. altered to

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Maximizing Sample Throughput In Purge And Trap Analysis

Maximizing Production While Minimizing Costs

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

2017 EPA Method Update Rule and EPA Method 624.1

EPA TO-17 Volatile Organic Compounds

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Ed George and Anaïs Viven Varian, Inc.

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Optimizing Standard Preparation

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Solid Phase Micro Extraction of Flavor Compounds in Beer

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

Analytical Trap Comparison for USEPA Method 8260C

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Determination of Volatile Organic Compounds in Air

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Target Compound Results Summary

CALA Directory of Laboratories

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

STANDARD OPERATING PROCEDURES

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Associates. Project No.: October 25, 2017

Volatile Organic Compounds in Every Day Food

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Method 1624 Revision C Volatile Organic Compounds by Isotope Dilution GCMS

Copies: Dave Favero, RACER File

Volatile Organic Compounds in Water PBM

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

U.S. EPA Method 8270 for multicomponent analyte determination

Thermal Desorption Technical Support

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Study of Residual Solvents in Various Matrices by Static Headspace

Headspace Technology for GC and GC/MS: Features, benefits & applications

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Laboratory Report Number(s) Lone Tree Road (Deep) 10/5/15, 10/19/15, 6/13/16. Address Sample Date(s) Analyses Requested

Transcription:

application Note Gas Chromatography/ Mass Spectrometry Authors Ruben Garnica Dawn May PerkinElmer, Inc. Shelton, CT USA Method by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8 Introduction U.S. EPA Method Volatile Organic Compounds (VOCs) by Gas Chromatography Mass Spectrometry (GC/MS) is one of the most common environmental applications for GC/MS. This method outlines the analysis of volatile organic compounds in a variety of solid waste matrices including various air sampling trapping media, ground and surface water, soils, and sediments among others. The method requires not only demonstration of laboratory sample preparation and handling competence but instrument performance as well. The study presented here demonstrates the PerkinElmer Clarus SQ 8 GC/MS with purge and trap sample introduction both meets and exceeds the performance criteria set out in method and describes the analytical results and instrumental methodology. Experimental The PerkinElmer Clarus SQ 8C GC/MS operating in electron ionization mode with an Atomx purge and trap sample introduction system (Teledyne Tekmar, Mason, OH) was used to perform these experiments. The purge and trap conditions are presented in Table 1 and represent standard conditions for the analysis of method of VOCs by EPA Method.

Table 1. Purge and Trap Instrument Conditions. Purge and Trap System: Tekmar - Atomx Trap: Tekmar #9 trap Sample Size: 5 ml Purge Parameters: Valve Oven Temp 140 C Transfer Line Temp 140 C Sample Mount Temp 90 C Water Heater Temp 90 C Sample Vial Temp 20 C Sample Equilibrate Time 0.00 min Standby Flow 10 ml/min Purge Ready Temp 40 C Condensate Ready Temp 45 C Presweep Time 0.25 min Prime Sample Fill Volume 3.0 ml Sweep Sample Time 0.25 min Sweep Sample Flow 100 ml/min Sparge Vessel Heater On Purge Time 11.00 min Purge Flow 40 ml/min Purge Temp 40 C Condensate Purge Temp 20 C Dry Purge Time 2.00 min Dry Purge Flow 100 ml/min Dry Purge Temp 20 C Desorb Parameters: Water Needle Rinse Volume 7.0 ml Sweep Needle Time 0.50 min Desorb Preheat Temp 200 C GC Start Signal Start of Desorb Desorb Time 2.00 min Drain Flow 300 ml/min Desorb Temp 200 C Bake Parameters: Number of Water Bake Rinses 2 Water Bake Rinse Volume 7.0 ml Bake Rinse Sweep Time 0.25 min Bake Rinse Sweep Flow 100 ml/min Bake Rinse Drain Time 0.40 min Bake Time 6.00 min Bake Flow 250 ml/min Bake Temp 270 C Condensate Bake Temp 200 C The Clarus SQ 8C GC/MS conditions are presented in Table 2. The heated transfer line of the concentrator was connected directly to the injector port of the Clarus 680 GC. A 2 mm id injector port liner was used inside the capillary injector. Split flow adjustments will increase or decrease the split ratio and the resulting sensitivity of the system and allow the analysts flexibility in optimizing their GC/MS system. Table 2. Clarus SQ 8C GC/MS conditions. Gas Chromatograph: PerkinElmer Clarus 680 Analytical Column: Elite 624 MS (30 m x 0.25 mm id x 1.4 μm) Injector Temperature: 220 C Carrier Gas: Helium @ 1.0 ml/min Split Flow: 80 ml/min Oven Program: Temperature Hold Time Rate 40 C 2 min 10 C/min 100 C 0 min 30 C/min 200 C 4 min End Mass Spectrometer: PerkinElmer Clarus SQ 8C GC Transfer Line Temperature: 220 C Ion Source Temperature: 300 C Function Type: Full Scan Solvent Delay: 0 0.50 min Scan Range: m/z 35-270 Scan Time: 0.20 sec Interscan Delay: 0.10 sec Calibration and performance standards were prepared from commercially available stock standards and diluted in Class-A volumetric flasks. Calibration standards were made from a combination of four multi-component standard mixes including a 76 component 8260 calibration mix, a 6 component 502.2 calibration gas mix, a 3 component VOA surrogate spike mix, and an 8260 internal standard mix. Stock standards were diluted to an intermediate concentration from which initial calibration standards were prepared. The calibration levels used in this study are presented in Table 3. Standards used for the Method Detection Limit (MDL) and precision experiments were prepared at appropriate concentration levels from intermediate standards. The internal standard mix was added to all samples such that a resulting concentration of 20 μg/l was achieved per 5 ml of sample. The tuning standard, 4-bromofluorobenzene, was prepared at 50 μg/l. Standards, stock standards, and associated QC/ QA samples were stored in a manner consistent with the guidelines set out in the method. 2

Table 3. Calibration points employed in this study. Calibration Level Concentration (μg/l) 1 0.5 2 1.0 3 2.0 4 5.0 5 10 6 25 7 50 8 100 9 200 Results and Discussion The GC conditions were optimized to provide adequate separation while keeping the analytical runtime as short as possible. Figure 1 demonstrates a mid-level (25 μg/l) total ion chromatogram (TIC) from this analysis. All of the expected analyte separation is achieved while allowing for a maximum of throughput per unit laboratory time. The injection-to-injection time is further enhanced by the performance of the Clarus 680 GC oven. The oven incorporates a dual walled design with a unique air flow path to allow the fastest cooling times of any commercially available GC oven on the market. Air flow through the Clarus 680 GC is unidirectional eliminating the mixing of heated exhaust and cool ambient air allowing for the fastest transport of heat out of the GC oven. This high speed cooling allows the user to dramatically decrease the idle time of the instrument between analytical runs and in this study the injection-toinjection time, measured to be under 30 minutes, was limited by the purge and trap system. Table 4 presents the analytical results of these experiments including the Retention Time (RT), Average Relative Response Factors (Avg RRF) and Percent Relative Standard Deviation (% RSD) of the initial calibration, the Method Detection Limits (MDL), and the method precision as Percent Recovery (% Rec) and accuracy as % RSD. All of the analytes presented meet or exceed the minimum method requirements and boast excellent detection limits, precision and accuracy. Method Detection Limits were determined by analyzing seven replicate samples at a concentration of 0.5 μg/l and 1.0 ug/l. Precision and accuracy were measured by analyzing four replicates at a concentration of 25 μg/l. The majority of compounds in this report are calibrated over the entire concentration range as presented in Table 3. Compounds whose calibration range varied from the bulk of analytes are highlighted in Table 5. Alternate calibration ranges are expected for this method especially for compounds referred to as bad actors. In all cases, however, EPA method criteria are met. Overall improved results including MDL and precision and accuracy were obtained by operating the ion source at an elevated temperature than would regularly be used. Figure 2 shows a time selection of the seven TICs from the MDL study and illustrates the excellent repeatability of the system at a low concentration level. The inset presents the seven extracted ion chromatograms of naphthalene (EIC) at m/z = 128. The % RSD of the seven naphthalene EICs was measured to be 3.8%, which includes the sample handling performed by the purge and trap. 3

Figure 1. TIC of volatile organic standard mix at 25 μg/l. Numbering references compound as listed in Table 4. Figure 2. Zoom of the seven MDL samples. TICs overlaid to illustrate the excellent repeatability achieved at low levels. Inset shows the naphthalene EIC at m/z = 128; the %RSD was measured to be 3.8%. 4

Table 4. Summarized initial calibration results of the analyzed VOCs from 0.5 200 μg/l for most compounds. A hyphen (-) indicates value either not provided by the method or not generated in this work. A plus sign (+) indicates calibration calculated using linear regression with R 2 presented in table. Avg Minimum % RSD MDL Precision Accuracy # Compound RT RRF RRF % RSD Criteria (mg/l) (% Rec) (% RSD) 1 Dichlorodifluoromethane 1.38 0.44 0.1 14 20 0.14 90 2.9 2 Chloromethane 1.59 0.72 0.1 0.9992 0.990 0.15 96 4.9 3 Vinyl Chloride 1.66 0.79 0.1 8.2 20 0.21 95 1.8 4 Bromomethane + 2.00 0.26 0.1 18 20 0.18 105 4.1 5 Chloroethane 2.08 0.40 0.1 4.9 20 0.24 99 1.9 6 Trichlorofluoromethane 2.29 0.57 0.1 5.9 20 0.06 109 1.2 7 Diethyl ether 2.58 0.37-6.7 20 0.11 98 4.6 8 1,1,2-Trichloro-1,2,2-trifluor 2.78 0.38 0.1 13 20 0.15 99 1.1 9 1,1-Dichloroethene 2.82 0.44 0.1 15 20 0.30 102 1.8 10 Carbon disulfide 3.02 1.46 0.1 5.2 20 0.05 93 1.4 11 Allyl Chloride + 3.21 0.17-0.9940 20 0.05 76 2.7 12 Acetonitrile + 3.29 0.20-0.9999 0.990 0.38 96 5.5 13 Methylene Chloride 3.37 0.51 0.1 16 20 0.29 94 2.5 14 trans-1,2 Dichloroethene 3.59 0.46 0.1 9.3 20 0.34 105 1.3 15 Acrylonitrile 3.67 0.44-8.8 20 0.14 97 4.6 16 1,1,-Dichloroethane 4.06 0.98 0.2 9.2 20 0.16 107 2.3 17 Chloroprene 4.11 0.70-11 20 0.25 89 1.7 18 2,2,-Dichloropropane + 4.63 0.15-0.9996 0.990-56 7.4 19 cis-1,2,-dichloroethene 4.66 0.53 0.1 9.6 20 0.24 106 1.9 20 Methyl acrylate 4.76 0.92-4.2 20 0.14 100 5.3 21 Propionitrile 4.84 0.22-4.4 20 0.12 100 5.7 22 Tetrahydrofuran 4.92 0.53-13 20 0.15 86 4.7 23 Bromochloromethane 4.92 0.26-4.5 20 0.09 107 3.6 24 Methacrylonitrile 4.96 0.83-8.9 20 0.12 90 4.7 25 Chloroform 4.99 0.69 0.2 4.2 20 0.09 103 2.1 26 1,1,1-Trichloroethane 5.14 0.55 0.1 5.5 20 0.20 107 0.9 27 Pentafluorobenzene (ISTD) 5.22 28 Carbon tetrachloride 5.27 0.38 0.1 12 20 0.13 108 1.2 29 1,1-Dichloropropene 5.32 0.43-5.9 20 0.25 89 1.6 30 Isobutyl alcohol + 5.49 0.09-11.5 20 0.22 89 5.6 31 Benzene 5.53 1.33 0.5 6.9 20 0.22 90 1.6 32 1,2-dichloroethane-d4 (Surr) 5.55 33 1,2-Dichloroethane 5.64 0.42 0.1 9.8 20 0.09 91 2.0 34 1,4-Difluorobenzene (ISTD) 6.00 35 Trichloroethene 6.25 0.32 0.2 8.9 20 0.28 103 2.5 36 1,2-Dichloropropane 6.56 0.47 0.1 6 20 0.15 96 1.8 37 Methyl methacrylate 6.67 0.39-6.2 20 0.17 91 3.9 38 Dibromomethane 6.70 0.19-4.9 20 0.10 106 2.7 39 Bromodichloromethane 6.87 0.28 0.2 7.4 20 0.17 102 1.9 40 2-Nitropropane + 7.23 0.06-0.9996 0.990-94 4.7 41 cis-1,3-dichloropropene 7.41 0.47 0.2 8.7 20 0.08 90 1.6 42 Toluene-d8 (Surr) 7.66 5

Table 4 continued Avg Minimum % RSD MDL Precision Accuracy # Compound RT RRF RRF % RSD Criteria (mg/l) (% Rec) (% RSD) 43 Toluene 7.74 0.67 0.4 6.5 20 0.22 100 1.5 44 trans-1,3-dichloropropene 8.11 0.39 0.1 17 20 0.11 93 2.5 45 Ethyl methacrylate 8.16 0.66-11 20 0.07 101 2.6 46 1,1,2-Trichloroethane 8.31 0.28 0.1 4.3 20 0.06 96 1.9 47 Tetrachloroethene+ 8.35 0.43 0.2 20 20 0.17 123 4.6 48 1,3-Dichloropropane 8.50 0.65-4.4 20 0.06 94 3.1 49 Chlorodibromomethane 8.69 0.28 0.1 12 20 0.10 108 3.7 50 Ethylene Dibromide 8.81 0.40 0.1 3.6 20 0.07 111 2.8 51 Chlorobenzene-d5 (ISTD) 9.21 52 Chlorobenzene 9.24 1.08 0.5 4.7 20 0.23 102 1.0 53 Ethylbenzene 9.31 1.34 0.1 2.2 20 0.20 104 1.4 54 1,1,1,2-Tetrachloroethane 9.31 0.27-7.8 20 0.19 105 1.9 55 p,m-xylene 9.41 1.09 0.1 4.4 20 0.23 101 1.6 56 o-xylene 9.73 1.15 0.1 2.5 20 0.19 103 2.2 57 Styrene 9.76 1.09 0.3 3.8 20 0.18 106 1.8 58 Bromoform 9.91 0.22 0.1 0.9996 0.990 0.10 104 2.8 59 Isopropylbenzene 10.01 1.61 0.1 2.8 20 0.18 106 1.8 60 cis-1,4-dichloro-2-butene 10.12 0.22-0.9986 0.990 0.09 68 2.2 61 Bromofluorobenzene (Surr) 10.15 62 Bromobenzene 10.25 0.98-8.5 20 0.15 98 2.3 63 1,1,2,2-Tetrachloroethane 10.28 1.17 0.3 5.9 20 0.16 93 9.9 64 n-propylbenzene 10.30 3.21-4.8 20 0.20 101 1.1 65 1,2,3-Trichloropropane 10.32 1.76-5.8 20 0.10 91 3.3 66 trans-1,4-dichloro-2-butene 10.33 0.25-0.9991 0.990 0.05 71 2.2 67 2-Chlorotoluene 10.38 1.88-4.4 20 0.19 102 0.9 68 1,3,5-trimethylbenzene 10.43 3.00-6.3 20 0.21 104 1.1 69 4-Chlorotoluene 10.47 1.99-7.6 20 0.15 100 3.2 70 tert-butylbenzene 10.64 2.57-9.2 20 0.18 94 1.3 71 1,2,4-Trimethylbenzene 10.69 2.94-4.8 20 0.19 105 1.2 72 sec-butylbenzene 10.79 3.48-5.1 20 0.20 107 1.7 73 4-Isopropyltoluene 10.88 2.98-4.5 20 0.22 97 1.0 74 1,3-Dichlorobenzene 10.89 1.64 0.6 9 20 0.22 98 0.7 75 1,4-Dichlorobenzene-d4 (ISTD) 10.93 76 1,4-Dichlorobenzene 10.95 1.65 0.5 9.4 20 0.18 97 1.4 77 n-butylbenzene 11.14 2.10-6.5 20 0.17 98 1.2 78 1,2-Dichlorobenzene 11.19 1.60 0.4 4.5 20 0.12 102 1.5 79 1,2-Dibromo-3-chloropropane 11.69 0.33 0.05 9.8 20 0.08 98 3.6 80 Nitrobenzene+ 11.84 0.06-0.9999 0.990-102 5.4 81 1,2,4-Trichlorobenzene 12.24 1.19 0.2 9.3 20 0.16 92 1.3 82 Hexachlorobutadiene 12.32 0.50-16 20 0.23 91 1.3 83 Naphthalene 12.46 3.60-7.3 20 0.06 102 2.4 84 1,2,3-Trichlorobenzene 12.65 1.13 0.2 7.2 20 0.09 94 1.6 6

Table 5. Alternate calibration range compounds. Calibration Concentration Range Compound 0.5 100 μg/l Tetrachloroethane 1.0 200 μg/l Bromomethane Allyl Chloride Acetonitrile Isobutyl Alcohol 2.0 200 μg/l 2,2-Dichloropropane 2-Nitropropane Nitrobenzene Design advancements in the Clarus SQ 8 mass spectrometer allow such a wide operating range with the Clarifi detector providing much of the advantage. The detector s enhanced sensitivity enables full scan mass spectrometry methods to be performed resulting in library searchable spectra at the lowest possible concentration levels. With the added sensitivity of SIFI (selected ion full ion) or SIM (selected ion monitoring) data acquisition at even lower limits are possible for specific compounds. The voltage setting on the Clarifi detector was important in developing the method for such a wide concentration range and deserves mention here. The UltraTune (Standard DFTPP/BFB) function built into the TurboMass v6.0 software is designed to set the voltage such that the m/z = 69 peak of the tune gas FC43 is approximately 80% in the tune window. This setting is appropriate for general instrument analyses however users are encouraged to fine tune this setting to suit their specific analytical needs. In our case this was achieved by analyzing two samples, the high and low concentration level of our calibration, at voltage settings below that set during UltraTune. The optimum voltage setting produced measurable signal at the lowest level while not saturating the detector at the highest level. In our case the initial UltraTune setting was 1570V with test measurements performed at -50V and -100V of this value. The analysis was performed at 1520V. The stability of the Clarus SQ 8 system allows this setting to be used for extended periods of time and need only be reproduced periodically. The UltraTune (Standard DFTPP/BFB) function also produces a satisfactory tune which meets the Tune Evaluation requirements set out in the method. Figure 3 demonstrates the passing BFB Tune Evaluation Sample. Figure 4 shows the mass spectrum utilized in the Tune Evaluation Test and all of the necessary peak intensity comparisons are achieved. Figure 3. TurboMass 6.0 EPA BFB tune evaluation results. Figure 4. Mass spectrum of EPA BFB tune evaluation used for test in Figure 2. Conclusion The analysis of VOCs by purge and trap GC/MS following EPA Method is easily achievable over a wide concentration range using the Clarus SQ 8 GC/MS system. Satisfactory analytical results were achieved for a concentration range from 0.5 200 μg/l for a majority of analyte compounds using the full scan approach providing library searchable spectra at all concentration levels. Excellent detection limits are reported as are accuracy and precision measurements. A number of technological advances make the Clarus SQ 8 GC/MS the ideal systems for laboratories wishing to perform high throughput and sensitivity analyses with an ease of operation currently unmatched. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/contactus Copyright 2012, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 010017_01