Unique Selectivity: The Power of Ionic Liquid GC Columns

Similar documents
Supelco Ionic Liquid GC Columns Introduction to the Technology

Journal of Chromatography A

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

Application of an Ionic Liquid Column to the Analysis of Flavor and Fragrance Ingredients

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis

Katherine K. Stenerson, Michael Ye, Michael Halpenny, Olga Shimelis, and Leonard M. Sidisky. Supelco, Div. of Sigma-Aldrich Bellefonte, PA USA

Headspace Technology for GC and GC/MS: Features, benefits & applications

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Optimizing. Abstract: is standardd. procedures. altered to

Glove Selection Chart

Carbon Molecular Sieves and Spherical Graphitized Polymer Carbons for Sample Preparation Processes and Bulk Scale Purification Processes

10/27/10. Chapter 27. Injector typically 50 C hotter than oven

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

N6,6 Posidyne Membranes

EPA TO-17 Volatile Organic Compounds

Study of Residual Solvents in Various Matrices by Static Headspace

Get Selective. By Jaap de Zeeuw

Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010

Secrets of GC Column Dimensions

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Gas Chromatography. 1. Introduction. 2. Stationary phases. 3. Retention in Gas-Liquid Chromatography. 4. Capillary gas-chromatography

Introduction and Principles of Gas Chromatography

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

SUPELCO. Improved Performance for Cyanopropylphenyldimethylpolysiloxane (SPB-1701, SPB-1301) Capillary Columns

Principles of Instrumental Analysis

Colin F. Poole Department of Chemistry Wayne State University USA

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

SUPELCO. The Analysis of all 209 PCB Congeners on the SPB -Octyl and MDN -5S Capillary Columns. Katherine K. Stenerson and Leonard M.

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Solid Phase Micro Extraction of Flavor Compounds in Beer

Maximizing Production While Minimizing Costs

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Optimizing Standard Preparation

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Target Compound Results Summary

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

The Suite for Environmental GC Analysis

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

CHEM340 Tutorial 4: Chromatography

Thermal Desorption Technical Support

Determination of Hydrocarbon Components in Petroleum Naphthas

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Recent advances in Inverse Gas Chromatography - IGC - measurements of HSP. RALF DUEMPELMANN, ERIC BRENDLE and STEVEN ABBOTT

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Tube O Reactor. (Micro: Cat. # K, K, K) (Medi: Cat. # K, K, K)

GAS CHROMATOGRAPHY (GC)

Solvents and solvent selection for chromatography. Prof. Colin F. Poole Department of Chemistry Wayne State University USA

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

Loudon Chapter 8 Review: Alkyl Halides, Alcohols, etc. Jacquie Richardson, CU Boulder Last updated 8/24/2017

SPE Introduction. general chromatography. solid phase extraction. How to Choose an SPE Product. Solid-Phase Processing Methods

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

GL Chromatodisc. Usage Type Description. GL Chromatodisc Type A GL Chromatodisc Type P GL Chromatodisc Type N 13AI

CALA Directory of Laboratories

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Experiment 1: Thin Layer Chromatography

Methanol Extraction of high level soil samples by USEPA Method 8260C

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

D2245

Experiment 1: Extraction and Thin Layer Chromatography

Chemistry Instrumental Analysis Lecture 28. Chem 4631

Gas chromatography. Flow measurement. Flow control. Injection methods. Flow measurement

8. Methods in Developing Mobile Phase Condition for C18 Column

High Purity Chromasolv Solvents

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Electronic Supplementary Information (ESI)

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

How To Select the Correct GC Column. Simon Jones Application Engineer

Speakers. Moderator. John V Hinshaw GC Dept. Dean CHROMacademy. Tony Taylor Technical Director CHROMacademy. Dave Walsh Editor In Chief LCGC Magazine

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Industrial Applications with a New Polyethylene Glycol-Based GC Column

For Chromatography, you might want to remember Polar Dissolves More, not like dissolves like.

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Chromatography. Gas Chromatography

Chemical Compatibility Guide and Life Expectancy for the SolVac Filter Holder

Gloves: choice and use

Electronegativity Scale F > O > Cl, N > Br > C, H

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Transcription:

Unique Selectivity: The Power of Ionic Liquid GC Columns Leonard M. Sidisky, Jamie L. Desorcie, Gustavo Serrano Izaguirre, Daniel L. Shollenberger; Greg A. Baney, Katherine K. Stenerson, and Michael D. Buchanan Supelco, Div. of Sigma-Aldrich Bellefonte, PA 1823 USA T415083 sigma-aldrich.com/analytical

Abstract Choosing a stationary phase is the most critical step in column selection, more important than the column s I.D., film thickness, or length. This is because the stationary phase determines the selectivity of the column, and that selectivity influences resolution. Changing stationary phase may be an effective way to increase resolution. 1 Beginning in 200, extensive evaluations of columns manufactured with ionic liquid stationary phases have occurred. Their main strength was discovered to be unique selectivity. These columns have the ability to perform many of the same applications as columns made with polysiloxane polymer or polyethylene glycol stationary phases of similar polarity, but with slight elution order changes. Many times this results in increased resolution and/or shorter run times. 2

GC Column Polarity Scale A test was devised to experimentally determine which, if any, non-ionic liquid columns have similar polarity to each ionic liquid column. The procedure we adopted was proposed by Prof. Luigi Mondello (University of Messina, Italy). Each column was first characterized using a series of five probes (benzene, n-butanol, 2-pentanone, nitropropane, and pyridine). N-alkane markers were used to determine the retention index for each probe. McReynolds Constants were then calculated using the retention index data of each column relative to the retention index data for the same five probes on squalane, the most nonpolar GC stationary phase. The five McReynolds Constants were then summed to obtain Polarity (P) values, which were then normalized to SLB-IL100 (set at P.N. = 100) to obtain Polarity Number (P.N.) values. Figure 1 shows the tabulated results, whereas a visual representation is displayed in Figure 2. 3

Figure 1. Tabulated Results McReynolds Constants Column Benzene N-Butanol 2-Pentanone Nitropropane Pyridine P P.N. SPB -Octyl 17-20 19 28 1 Equity -1 11 10 33 0 1 130 3 SLB -5ms 33 30 55 91 43 252 SPB-20 7 79 104 17 109 535 12 Equity-1701 82 131 150 233 13 732 1 SPB-35 175 113 151 225 175 839 19 SPB-50 154 134 17 2 218 948 21 SPB-225 233 342 342 501 375 1793 40 PAG 27 459 320 508 428 1991 45 SUPELCOWAX 10 334 509 375 01 505 2324 52 SLB-IL59 338 505 549 49 583 224 59 SLB-IL0 32 492 525 7 54 222 59 SLB-IL1 371 551 51 24 48 2710 1 SP -2330 49 3 08 859 712 3311 75 SLB-IL7 45 90 43 845 745 3379 7 SP-2331 495 74 22 85 735 3382 7 SP-250 510 724 52 913 773 3572 81 SLB-IL82 532 7 701 921 808 338 82 TCEP 22 871 772 1072 957 4294 97 SLB-IL100 02 853 884 1017 1081 4437 100 SLB-IL111 7 930 957 1192 1093 4938 111 4

Figure 2. Visual Representation 5

Polarity and Selectivity A quick look at Figure 2 reveals the following regarding column polarity: SLB-IL59, SLB-IL0, and SLB-IL1 are somewhat similar in polarity (slightly more polar) to columns made with a polyethylene glycol (PEG) stationary phase. SLB-IL7 and SLB-IL82 are similar in polarity to columns made with a polysiloxane polymer stationary phase that contains a high percentage of cyanopropyl pendent groups. SLB-IL100 is very close in polarity to columns made with the TCEP stationary phase; TCEP = 1,2,3-tris(2-cyanoethoxy)propane. SLB-IL111 has a polarity unlike any other GC column. Polarity comparisons are a good start, but do not tell the whole story. Recent solvation parameter model (SPM) evaluations indicate that only the ionic liquid columns are capable of simultaneously providing intense H-acceptor and intense H-donor interactions, along with dipolar and - interactions. 2 These multiple interactions result in unique selectivity, and may produce better resolution and/or faster analysis.

Examples Several examples of unique selectivity resulting in better resolution and/or faster analysis are illustrated in Figure 3 through Figure 9. 7

Figure 3. Fusel Alcohols; 90 ºC Isothermal SLB-IL0 30 m x 0.25 mm I.D., 0.20 µm PEG 30 m x 0.25 mm I.D., 0.25 µm 2 1. 2-Methyl-1-Butanol (Active Amyl Alcohol) 2. 3-Methyl-1-Butanol (Isoamyl Alcohol) 1,2 1 0 2 4 Time (min) 1.0 2.0 3.0 4.0 5.0.0 Time (min) 8

Figure 4. Anilines; 75 ºC (2 min), 10 ºC/min to 250 ºC SLB-IL59 30 m x 0.25 mm I.D., 0.20 µm 3 4 1 5 7 1. Aniline 2. 3-(Chloromethyl)pyridine 3. 2-Aminotoluene 4. 4-Aminotoluene 5. 3-Aminotoluene. 2,-Dimethylaniline 7. 2,4-Dimethylaniline 2 11.0 12.0 13.0 PEG 30 m x 0.25 mm I.D., 0.25 µm 1 3 4 5 7 2 12.0 13.0 14.0 9

Figure 5. Aromatics in Reformulated Gasoline; Last Analyte to Elute is 1-Methylnaphthalene (1 mn) SLB-IL111 (1 mn elutes at 14.7 min) 0 m x 0.25 mm I.D., 0.20 µm B TCEP (1 mn elutes at 39.4 min) 0 m x 0.25 mm I.D., 0.44 µm 10

Figure. Esters and Ethers; 40 ºC (4 min), 8 ºC/min to 200 ºC (5 min) 1 1 2 3 2, 3 4 4 1 5 5 SLB-IL0 30 m x 0.25 mm I.D., 0.20 µm 7 PEG 30 m x 0.25 mm I.D., 0.25 µm 7 8 9 8 2 4 8 10 Time (min) 9 10 10 11 11 C 12 12 1. Methyl formate 2. Ethyl formate 3. Methyl acetate 4. Tetrahydrofuran 5. Ethyl acetate. Isopropyl acetate 7. n-propyl acetate 8. Isobutyl acetate 9. 1,4-Dioxane 10. n-butyl acetate 11. Isoamyl acetate 12. n-amyl acetate C. Contaminant 2 4 8 10 Time (min) 11

Figure 7. Polychlorinated Biphenyls (PCBs) SLB-IL82 30 m x 0.25 mm I.D., 0.20 µm; 50 ºC (2 min), 5 ºC/min to 20 ºC Poly(cyanopropylphenyl)siloxane 30 m x 0.25 mm I.D., 0.20 µm 0 ºC (1 min), 8 ºC/min to 230 ºC 12

2 1 3 4 Figure 8. Fatty Acid Methyl Esters (FAMEs); 170 ºC, 1 ºC/min to 225 ºC 5 7 SLB-IL0 30 m x 0.25 mm I.D., 0.20 µm 8 9 10 11 12 13 14 15 18 1 17 20 19 21 23 28 22 24 25 2 27 29 30 31 32 33 34 10 20 30 Min PEG 30 m x 0.25 mm I.D., 0.25 µm 18 1 17 20 19 21 37 35 3 3 min 38 1. C4:0 2. C:0 3. C8:0 4. C10:0 5. C11:0. C12:0 7. C13:0 8. C14:0 9. C14:1 10. C15:0 11. C15:1 12. C1:0 13. C1:1 14. C17:0 15. C17:1 1. C18:0 17. C18:1n9c 18. C18:1n9t 19. C18:2nc 20. C18:2nt 21. C18:3n 22. C18:3n3 23. C20:0 24. C20:1n9 25. C20:2 2. C20:3n 27. C21:0 28. C20:3n3 29. C20:4n 30. C20:5n3 31. C22:0 32. C22:1n9 33. C22:2 34. C23:0 35. C22:5n3 3. C24:0 37. C22:n3 38. C24:1n9 13

Figure 9. FAMEs in B20 Biodiesel; 50 ºC, 13 ºC/min to 270 ºC (5 min) SLB-IL111 30 m x 0.25 mm I.D., 0.20 µm No overlap of hydrocarbon and FAME fractions PEG 30 m x 0.25 mm I.D., 0.25 µm 14

Figure 10. Chlorinated Solvents; 40 ºC (4 min), 8 ºC/min to 200 ºC (5 min) 2 1 5 SLB-IL0 30 m x 0.25 mm I.D., 0.20 µm 4 3 1 5 4 9 7 10 8 11 2 4 8 10 11 12 Time (min) 2 3, 4 3 12 PEG 30 m x 0.25 mm I.D., 0.25 µm 14 1. 1,1-Dichloroethylene 2. trans-1,2-dichloroethylene 3. Carbon tetrachloride 4. 1,1,1-Trichloroethane 5. 1,1-Dichloroethane. Methylene chloride 7. Trichloroethylene 8. Tetrachloroethene 9. Chloroform 10. 1,2-Dichloroethane 11. 1,1,1,2-Tetrachloroethane 12. 1,1,2,2-Tetrachloroethane 5 7 10 8 11 12 9 2 4 8 10 12 Time (min) 14 1 15

Figure 11. Native Spearmint Essential Oil; 75 ºC (4 min), 4 ºC/min to 200 ºC (10 min) SLB-IL0 30 m x 0.25 mm I.D., 0.20 µm 0 10 20 30 40 Min PEG 30 m x 0.25 mm I.D., 0.25 µm 0 10 20 30 Min 1

Summary Ionic liquid stationary phases are something totally new and completely different in the world of GC. Because their wide range of interaction mechanisms translates to different selectivity options, they have the opportunity to positively impact current GC and GC-MS practices. Columns can be engineered with completely unique selectivity to non-ionic liquid columns, while producing good peak shape and resolution for compounds of varying functionality. Equity, SLB SP, SPB, and SUPELCOWAX is a registered trademarks of Sigma-Aldrich Co. LLC. 17

References 1. Barry E.F., Columns: packed and capillary; column selection in gas chromatography. In: Grob R.L., Barry E.F., editors. Modern Practice of Gas Chromatography, Fourth Edition. New Jersey: John Wiley & Sons, Inc.; 2004. p 5 191. 2. Rodríguez-Sánchez S., Galindo-Iranzo P., Soria A.C., Sanz M.L., Quintanilla-López J.E., Lebrón-Aguilar R., Principle component analysis (PCA) evaluation of seven commercial ionic liquid capillary GC columns. Supelco Reporter/2015; 33.1: 3-4. Equity, SLB, SP, SPB, and SUPELCOWAX are registered trademarks of Sigma-Aldrich Co. LLC. 18

Thank You 19