To: Mitch Cottrell, Instructor From: Brian Garber, Crawford Binns, Jeff Hollis (Group 3, Sec. J) Subject: Project Deliverable #3 Date: 4 May 2010

Similar documents
1. Mark the correct statement(s)

Lab 1a Wind Tunnel Testing Principles & Drag Coefficients of Golf balls

ECNG3032 Control and Instrumentation I

Hands-on Lab 3. System Identification with Experimentally Acquired Data

Thermocouple Dynamic Errors Correction for Instantaneous Temperature Measurements in Induction Heating. Krzysztof Konopka 1

When fluid incompressible...flow velocity, volumetric, and mass flow are proportional.

Freescale Semiconductor

Experiment A4 Sensor Calibration Procedure

CE 321 INTRODUCTION TO FLUID MECHANICS Fall 2009 LABORATORY 1: FLOW RATE MEASUREMENT

REDUCING PROCESS VARIABLITY BY USING FASTER RESPONDING FLOWMETERS IN FLOW CONTROL

Use Authorization. Signature. Date

Lab 1f Boiling Heat Transfer Paradox

Experiment B6 Thermal Properties of Materials Procedure

SECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components:

O Plus Dry Bushing 69 kv system, 350 kv BIL, 3000 A. Table of contents

Safety Precautions WARNING If critical situations that could lead to user s death or serious injury is assumed by mishandling of the product.

Experiment # 5 5. Coupled Water Tanks

Using Microsoft Excel

Target Insertion Flowmeter

Department of Mechanical and Aerospace Engineering. MAE334 - Introduction to Instrumentation and Computers. Final Examination.

Air Flow through Woven Stainless Steel Mesh

TEMPERATURE EFFECT IN MULTIPHASE FLOW METER USING SLOTTED ORIFICE PLATE

Glossary Innovative Measurement Solutions

6. The default plot created by Excel meets all of the requirements for a formal report plot in ME 360.

MCE380: Measurements and Instrumentation Lab

Temperature Project. Lab: Section K. Masahiro Inano Christopher Casey Adam Tigue

Information and slides provided by RFD

Those data on the time required for the pitch to traverse various distances were measured in

AC : A DIMENSIONAL ANALYSIS EXPERIMENT FOR THE FLUID MECHANICS CLASSROOM

Chapter 4 DYNAMICS OF FLUID FLOW

MODEL: WB009GMFI19HLD (120V) Wall Mount DC Inverter Fan Coil Unit 9,000 BTUH

EVALUATION OF A CLOSE COUPLED SLOTTED ORIFICE, ELECTRICAL IMPEDANCE, AND SWIRL FLOW METERS FOR MULTIPHASE FLOW. A Thesis MUHAMMET CEVIK

Applying Newton s Law of Cooling When The Target Keeps Changing Temperature, Such As In Stratospheric Ballooning Missions

TS02 - Pre-Instructional Survey

Ideal Gas Law and Absolute Zero

Technical Notes. Introduction. PCB (printed circuit board) Design. Issue 1 January 2010

Evidence for the Quantization of Electric Charge and Measurement of the Fundamental Unit of Electric Charge

After successfully completing this laboratory assignment, including the assigned reading, the lab

Investigation of anomalous heat production in Ni-H systems

TI 226H Calibration of Flow Check Devices using Positive Displacement Flow Meter

Fundamentals of Particle Counting

Introduction to 1118 Labs

Remote Display Unit. Installers Handbook Copyright 2001 AirSense Technology Ltd. LM Remote Display Unit Installers Handbook Issue 1.

Department of Mechanical Engineering ME 96. Free and Forced Convection Experiment. Revised: 25 April Introduction

Thermal Cube. Custom-built Heat Flux Sensor

200 kpa On-Chip Temperature Compensated Silicon Pressure Sensors

TRACKING DYNAMIC HOLD-UP OF JUICE IN A CANE BED

Course: ENGR 3280L. Section: 001. Date: 9/6/2012. Instructor: Jim Henry. Chris Hawk 9/6/2012

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas

Speed of Light in Air

Fall 2014 Nobby Kobayashi

Experiment 5: Thermocouples (tbc 1/14/2007, revised 3/16/2007, 3/22,2007, 2/23/2009, 3/13/2011)

Experiment 2 Random Error and Basic Statistics

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

O Plus Dry Bushing 25 kv system, 150 kv BIL, 3000 A

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

No HOT WIRE ANEMOMETER MODEL SK-27V. SPECIFICATIONS Measurement object : Clean air at normal pressure and humidity

Electrolytic Tiltmeter Temperature Coefficients: Source, Definition and Use to Improve Accuracy

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

SCB10H Series Pressure Elements PRODUCT FAMILY SPEFICIFATION. Doc. No B

Frictional Losses in Straight Pipe

Assembly and Operation Manual. April 2016

ORDERING INFORMATION # of Ports Pressure Type Device Name

Digital display Pressure range ON/OFF output Linear output Model No Positive pressure 0 to 98 kpa (0 to 1 kgf/cm 2 ) NPN open collector o 1 to o5

Fitting PS P - B - R2

PG-200/PG-208 PRESSURE GAUGES P G G P P

Thrust Measuring Test Stand - Experiment Lab - Horizontal Forces / Vectors. SeaPerch Activity

CALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS

ED 701 General Industry Pressure Transmitter

5. DRIVE 1. SAFETY 2. PREPARATION 3. FRAME 4. TRANSMISSION 6. ROW UNIT 7. OPTIONAL EQUIPMENT

Looking at a two binary digit sum shows what we need to extend addition to multiple binary digits.

SELECCIÓN DE CAUDALIMETROS. Tablas de Fabricantes

EM375 MECHANICAL ENGINEERING EXPERIMENTATION THERMOCOUPLE LABORATORY

Steady State Operating Curve: Speed UTC Engineering 329

ME 4600:483 Lab Notes Revised 11/16/2015. Flow Measurement

Technical Data Sheet DPT & DPIT-4005 SERIES DIFFERENTIAL PRESSURE TRANSDUCERS. Features. Description

CALYS 120 IS. ATEX documenting multifunction calibrator with high accuracy 0.006%

1.4 System Accuracy* B. G. LIPTÁK (1982, 1995, 2003) DEFINITIONS OF TERMS

Virtual Instrument for Crimped Connections Temperature and Specific Losses Monitoring During. Electrical Tests

Transmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level

Theory and Design for Mechanical Measurements

Siddharth Institute of Engineering & Technology

Laboratory 1: Calibration of a High-Volume Sampler

Thermal Cube. Custom-built Heat Flux Sensor

In this experiment, the concept of electric field will be developed by

Orifice and Venturi Pipe Flow Meters

SPECIFICATIONS PARTICLE SENSOR KS-28B Higashimotomachi, Kokubunji, Tokyo , Japan

200 kpa On-Chip Temperature. Compensated Silicon Pressure Sensors. MPX2200 Series. Pressure. Application Examples. Features. MPX2200 Rev 13, 10/2008

Remember that C is a constant and ë and n are variables. This equation now fits the template of a straight line:

Week 14 The Simple Pendulum

AE 3051, Lab #16. Investigation of the Ideal Gas State Equation. By: George P. Burdell. Group E3

EXPERIMENT 30A1: MEASUREMENTS. Learning Outcomes. Introduction. Experimental Value - True Value. 100 True Value

Lab #4 Similitude: The Kármán Vortex Street CEE 331 Fall 2004

IQI IQI. Proposal of quadratic equation for prediction of flow rates versus pressure in packed beds of cement. By Jan Malczyk

The Indestructibility of Hardtack:

Calibrate Rotameter and Orifice Meter and Explore Reynolds #

Arduino Software: How long before my battery is dead?, version 2.4

Introduction to Determining Power Law Relationships

Page AC : INSTRUMENTATION FOR SHOCK AND IMPACT ANALYSIS

Transcription:

To: Mitch Cottrell, Instructor From: Brian Garber, Crawford Binns, Jeff Hollis (Group 3, Sec. J) Subject: Project Deliverable #3 Date: 4 May 2010 This projects intention is to investigate how adjusting certain parameters of a Venturi jet affect its performance. As the typical purpose of these devices is to induce a vacuum with a liquid to entrap a secondary material, we decided to use a pressure transducer. The only must for the transducer was that it read both positive and negative pressures as we were unsure of the performance of our fabricated system. The initial decision for our transducer was to be a differential pressure, diaphragm unit from Omega.com (PX143-2.5BD5V). This sensor offered several features that made it a good choice. The option for a differential reading would have given us the option to read the pressure at different locations along the venturi and the transducer could have also been piped to read the gauge pressure of the vacuum, or measuring its difference to the ambient pressure. It was also cheap (relatively). After discussing the project though, it was decided to use an already available pressure transducer. A SUNX DP-Y27 was used in the end. This sensor s range was ±100kPa, which was more than five times the range the Omega.com transducer, and turned out to be quite necessary as our device saw higher positive that we had anticipated. The SUNX also provided a digital display built into the sensor that proved quite helpful. Another feature was the ability to withstand water. The Omega would have required remote mounting using rubber hosing, and certainly greater chances for leaks in the vacuum. With the ability of the Ramco to read water pressure without faulting, the sensor could be safely mounted directly onto the system without fear. The actual Venturi was constructed out of PVC and either cemented together or utilized high viscosity grease to seal press fittings. Consistent pressures were ensured by mounting the orifice vertically, thus keeping standing water from affecting readings. The two independent variables were the size of the nozzle and the total water flow through the nozzle. The nozzle was simply created by drilling a hole in a dead end PVC cap and increasing its diameter for each test series. The flow was determined using the flow gauge built into the pump station located in room 317. The major problem that we encountered was not being able to get the SUNX to output the analog voltage signal. After spending several hours studying the wiring diagrams and trying many different configurations, the 1V-5V signal was never found. The fault is not believed to lie with the DAQ board or the lab view program. The program is an adaptation of the same program used for the class s thermocouple homework assignment. The major changes being the alteration of the signal sampling rate, which was changed in hopes of alleviating any pulsing effects from the pump, and the inclusion of the correct equation for converting the voltage input into to pressure (Psig) output. Otherwise the program followed the same format as that of the Thermocouple homework. It created a virtual channel for the analog voltage input, which was then read by the program at a given sampling rate. This data was then combine with

a time to form a 2-dimensional array. Using a Butterworth filter the noise of the signal was reduced as the data got passed through numerous arithmetic blocks to convert voltage into a psi pressure. A time stamp was created and the output in psi was saved to an Excel spreadsheet (screen shots of the front panel and block diagram are located in the Appendix). This short fall did not prevent our execution of our experiment though because of the digital display located on the sensor itself. The internal circuitry displayed the pressure in Psi with a resolution of (±0.02). We did calculate the uncertainty of the readings as if the sensor had worked correctly though. There were two components that would need to be taken into account. First, the pressure transducer had four types of uncertainty, and then the USB 6009 that was used for the analog data analysis had two types of uncertainty. The nominal value of the pressure that was used in the calculation was the maximum for the transducer, 15.96 psid. The transducers uncertainty calculations can be seen in (Table 1) in the attachments. These uncertainties are as follows: Pressure Resolution (Abs. Accuracy) = 0.02 psid Linearity = 0.1% F.S. = 0.001 * 15.96 = 0.016 psid Nonrepeatability = 0.2% F.S. + 0.02 = 0.0519 psid Temperature = 0.3% F.S. = 0.0479 psid And for the 6009: Resolution = 5/ (214-1) * 4 = 0.0012 psid, where the 5 represents the maximum voltage, and 4 represents the maximum used in the system Absolute Accuracy = 0.007 * 4 = 0.028 This produced a total uncertainty in the pressure differential of 0.080177 pisd through perturbation of all uncertainties, which is approximately 1% of the maximum pressure differential, but much more significant for lower pressure readings of interest. Despite this large setback, we believe that our experiment yielded honest and logical results. With our basic understanding of the fundamentals of a Venturi, we expected that an increase in vacuum would occur when the flow of the pipe increased but we were mistaken. Or at least that was not the case with our rudimentary set up. The highest vacuums were not found with either the larger flows or the larger nozzle sizes. Instead our highest recorded vacuum was from smallest hole size (0.132 ) at only two gallons per minute. The other nozzle sizes, ranging from (0.151 ) to (0.661 ) consistently had a lower vacuums (higher pressure) at a given flow rate than the nozzle size beneath them (see Table 2). This led us to the conclusion that one major consideration in the Venturi s performance is not the flow rate but rather the velocity of the water exciting the nozzle. This pattern is exemplified in Fig. 1 with the addition of altitude lines showing pressure inversely proportional to flow rate. The second pattern we noticed implied that an additional variable is key to the Venturi s performance. All of the tests ran at greater than 4.5 g/min resulted in positive pressures inside the system. This was not explained by the correlation between velocities of a higher flow and larger nozzle matching the velocity of a lower flow and smaller nozzle. The factor that we determined to be hindering the Venturi was the ratio between

nozzle size and pipe size. Our piping remained the same for all nozzle sizes and the steady upturn in pressure (seen in Fig. 2) means that a backpressure is being encountered downstream from the Venturi. While the couplers between joints did not prevent the lower flows from pulling a vacuum, their restriction rendered higher flows incapable of making a vacuum. While we deem this experiment a success, the sensor we used was not. Through either our fault, or a fault of the sensor, we were not able to properly utilize our LabView program or our DAQ board. From our understanding from the semesters material, we tried to eliminate any possible error that we made that would have kept the sensor from outputting properly but did not come up with any solutions besides using the attached digital display.

Appendix

Table 1 Uncertainty in Diff. Pressure Pertubation A C D F G H J K L M N Index Device Range Feature Resolution Uncert or Unc in Pdot Sig+ Sig- (uncpdot)^2 Nominal Name Res * FS Pdot 15.96 1 Sunx DP-Y27 +/- 5V out Pres. Res. ABS 0.0200 0.0200 15.9800 15.9400 0.000400 Lin +/- 0.10% FS 0.0160 0.0160 15.9760 15.9440 0.000255 Temp +/- 0.30% FS 0.0479 0.0479 16.0079 15.9121 0.002292 Nonrepeat +/-.2% FS +.02 0.0519 0.0519 16.0119 15.9081 0.002696 2 V DAQ 14 bit resolution 14 bit res 0.000305194 0.00030518 0.0012 15.9612 15.9588 0.000001 0.000305194V Abs Acc 0.007 0.0280 15.9880 15.9320 0.000784 SUM 0.006428 SQRT 0.080177 TotUnc Pdot 0.080177 Table 2 Diameter in (inches) GPM 0.132 0.151 0.171 0.203 0.215 0.233 0.25 0.278 0.312 0.376 0.611 0.5 0 0 0 0 0 0 0 0 0 0 0 1-0.08-0.34 0 0 0 0 0 0 0 0 0 1.5-0.14-0.34-0.08-0.22-0.2-0.21-0.22-0.22-0.2-0.15-0.18 2-0.38-0.32-0.24-0.2-0.18-0.2-0.19-0.14-0.13-0.08-0.08 2.5-0.3-0.2-0.14-0.09-0.15-0.1-0.01-0.03-0.04 0.03 3-0.2-0.09-0.02 0 0.02 0.08 0.06 0.1 0.14 3.5-0.08 0.08 0.06 0.2 0.23 0.25 0.29 0.28 4-0.16 0.22 0.22 0.34 0.39 0.39 0.48 0.53 4.5 0.26 0.45 0.57 0.58 0.66 0.68 0.86 5 0.52 0.72 0.92 0.9 1.06 1.13 5.5 0.58 0.98 1.02 1.26 1.2 1.4 6 1.01 1.28 1.39 1.61 1.69 6.5 1.61 2.16 1.89 2.32 7 1.73 2.52 2.42 2.44 7.5 2.72 2.75 3.04 8 3.3 3.65 8.5 3.8 4.3 9 4.28 4.5 9.5 4.5 5.18 10 5.67 10.5 6.55

0.5 1.5 2.5 3.5 4.5 5.5 6.5 7.5 8.5 9.5 10.5 Pressure (Psig) Pressure (Psig) Fig: 1 Flow vs Pressure vs Nozzle Size 8 6 4 2 0-2 6-8 4-6 2-4 0-2 -2-0 0.5 1.5 2.5 3.5 4.5 5.5 6.5 7.5 8.5 9.5 10.5 Flow (GPM) Fig: 2 Flow vs Pressure vs Nozzle Size 7 6 5 6-7 5-6 4 4-5 3 3-4 2 1 0-1 2-3 1-2 0-1 -1-0 Flow (GPM)